Page 1: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

LCGC - RADWAG WEIGHING SOLUTIONS P. LTD

Laboratory Balances

Page 2: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

MYA/U series of microbalances series Y have been designed tomeet the high requirements of mass measurements with thehighest precision. Measurement reliability and accuracy isensured by internal calibration.

Microbalances consist of two major parts (an electronic system anda precise mechanical measurement system in a separate enclosure).This solution eliminates the temperature influence and separatesfrom shocks and vibrations caused by users operating software.All the elements of the balance are made of glass and steel whicheliminates the influence of electrostatics on the weighingprocess.

www.lcgcindia.com

[email protected]

MYA/U ULTRA-MICROBALANCES

Technical Data :

Max capacity

Readability

Repeatability

LinearityPan size

Weighing chamber dimensions

Stabilization time

Calibration

Working temperature

Interface

Power supply

Display

MYA 2/U

2 g

0,1�g

0,25�g

±0,5�gø 16 mm

ø 90 × 90 mm

10-20 s

automatic (internal)

+18 ° - +30 °C

RS 232, 2×USB, Ethernet

230V / 11V AC

5,7'’ touch screen

Additional Equipment:

ANTI-VIBRATION TABLE

"KAFKA" THERMAL PRINTER

SET FOR THE DETERMINATION OF AIR

FOOT TARE AND PRINT BUTTONS

ULTRASONIC AIR IONIZER WITH HUMIDIFIER

STANDARD MASS

PW-WIN, RAD-KEY, REC-FS COMPUTER SOFTWARE

PROFESSIONAL WEIGHING TABLE

ISO9001

%

GLP

Checkweighing

Filling

Percentage

Statistics

Infraredsensors

GLPprocedures

Electronic level indicator

- ALARM function- graphic level indicator- programmable acceptable tilts

Infrared proximity sensors

- PRINT function- TARE function- opening weighing chambers- sensors’ sensitivity adjustment

Page 3: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

MYA MICROBALANCES

Technical Data :

Additional Equipment:

ISO9001

%

GLP

Checkweighing

Filling

Percentage

Statistics

Infraredsensors

GLPprocedures

MYA series of microbalances series Y have been designed to meet thehigh requirements of mass measurements with the highest precision.Measurement reliability and accuracy is ensured by internal calibration.

Microbalances consist of two major parts (an electronic system and aprecise mechanical measurement system in a separate enclosure). This solution eliminates the temperature influence and separates from shocks and vibrations caused by users operating software.All the elements of the balance are made of glass and steel whicheliminates the influence of electrostatics on the weighing process.

ANTI-VIBRATION TABLE

"KAFKA" THERMAL PRINTERSET FOR THE DETERMINATION OF AIR

FOOT TARE AND PRINT BUTTONSADDITIONAL ATTACHMENT FOR PIPETTES CALIBRATION

ULTRASONIC AIR IONIZER WITH HUMIDIFIER

STANDARD MASS

PW-WIN, RAD-KEY, REC-FS COMPUTER SOFTWARE

PROFESSIONAL WEIGHING TABLE

Max capacity

Readability

Repeatability

Linearity

Pan size

Weighing chamber dimensions

Stabilization time

Calibration

Working temperatureInterfacePower supply

Display

MYA 55 g

1�g

2�g

±3�g

MYA 11

11 g

1�g

3�g

±5�g

MYA 21

21 g

1�g

5�g

±10�g

MYA 31

31 g

1�g

5�g

±10�g

ø 30 mm

ø 90 × 90 mm

10 s

automatic (internal)

+18 ° - +30 °CRS 232, 2×USB, Ethernet230V / 11V AC

5,7'’ touch screen

www.lcgcindia.com

[email protected]

Air buoyoancycorrection

Infrared proximity sensors

- PRINT function- TARE function- opening weighing chambers- sensors’ sensitivity adjustment

Electronic level indicator

- ALARM function- graphic level indicator- programmable acceptable tilts

Page 4: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

ANALYTICAL BALANCES XA/Y

Technical Data :

Additional Equipment:

ISO9001M

GLPprocedures

%

GLP

Air buoyoancycorrection

Filling

Statistics

PipetteCalibration

Countingpieces

AnimalWeighing

Densitydetermination

Checkweighing

Percentage

Infraredsensors

Formulation

Analytical balances of type has been designed on thebasis of new electronic modules and up-to-date technology.

XA/Y

Measurement reliability and accuracy is assured by internalcalibration triggered by time flow or temperature conditions.

Balances are equipped with spacious weighing chamber withautomatically opened side glass doors. Balances areoperated by using a touch panel covering a 5.7” colourgraphic display. Y scales have implemented new easilyoperated software!

www.lcgcindia.com

[email protected]

Max capacity

Minimal load

Readability

Tare range

Working temperature

Repeatability

Linearity

Sensitivity drift

Stabilization time

Power supply

Calibration

Pan size

XA 60/220/Y

60/220 g

1 mg

0,01/0,1 mg

-220 g

0,02/0,1 mg

±0,05/0,2 mg

5 s / 3 s

85 mm

XA 100/Y

100 g

10 mg

0,1 mg

±0,2 mg

3 s

XA 160/Y

160 g

10 mg0,1 mg

-160 g-100 g

0,1 mg0,1 mg

±0,2 mg

3 s 3 s

XA 220/Y

220 g

10 mg

0,1 mg

-220 g

0,1 mg

±0,2 mg

XA 52/Y

52 g

1 mg

0,01 mg

-52 g

0,01 mg

±0,02 mg

5 s

85 mmø

XA 110/Y

100 g

1 mg

0,01 mg

-100 g

0,025 mg

±0,05 mg

5 s

85 mmø ø ø

XA 310/Y

310 g

10 mg0,1 mg

-310 g

0,2 mg

±0,3 mg

3 s

100 mmø100 mmø100 mmø100 mm

230V / 11V AC or 120V / 11V AC

2 ppm/°C in temperature +18 ° - +30 °C

+10° - +50°C

ANTI-VIBRATION TABLE (MILD STEEL OR STAINLESS STEEL)

PRINTER KAFKA, EPSON

DENSITY KIT FOR SOLIDS AND LIQUIDS

PC KEYBOARD

ADDITIONAL LCD DISPLAY

EXTERNAL RECHARGEABLE BATTERY WITH CHARGER

COMPUTER SOFTWARE: PW-WIN 2004, RAD-KEY 2000, PIPETTES

FOOT BUTTON FOR TARE OR PRINT FUNCTIONS

internal (automatic)

Infrared proximity sensors

- PRINT function- TARE function- opening weighing chambers- sensors’ sensitivity adjustment

Electronic level indicator

- ALARM function- graphic level indicator- programmable acceptable tilts

Page 5: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

www.lcgcindia.com

[email protected]

ANALYTICAL BALANCES XA / X

Additional Equipment:

GLPprocedures

%

GLP

Air buoyoancycorrection

Filling

Statistics

PipetteCalibration

Countingpieces

AnimalWeighing

Densitydetermination

Checkweighing

Percentage

Formulation

ISO9001M

XA/X balances are equipped with big backlit graphic display with extendedmenu,12-keys membrane keyboard, big weighing chamber with sliding upperglass door and side glass door.Balance accuracy is guaranteed by automatic internal calibration dependant ontime and temperature changes.

Balances have possibility of disconnecting the head from weighing chamber.This solution considerably reduces vibrations transferred to weighing chamberand decreases dimensions of device.

XA/X balances enable to generate GLP procedures, these balances have RS 232socket, PS/2, possibility of connecting additional display and possibility ofweighing loads outside the main weighing platform(so called under-hook weighing).

New electronics and technological solutions!New intuitive operation and user friendly software of balances!

XA/X series- professional level analitycal balances

Technical Data :

Max capacity

Minimal load

Readability

Tare range

Working temperature

Repeatability

Linearity

Sensitivity drift

Stabilization time

Power supply

Calibration

Pan size

XA210/X

210 g

1 mg 1 mg

0,01 mg

-210 g

0,03 mg 0,02/0,1 mg

±0,1 mg

5 s

85 mm

XA 60/220/X

60/220 g

0,01/0,1 mg

±0,05/0,2 mg

5 s / 3 s

XA 100/X

100 g

10 mg0,1 mg

-160 g-220 g -100 g

0,1 mg±0,2 mg

3 s

XA 160/X

160 g

10 mg

0,1 mg

-220 g

0,1 mg

±0,2 mg

XA 52/X

52 g

1 mg

0,01 mg

-52 g

0,01 mg

±0,02 mg

5 s

85 mmø

XA 110/X

100 g

1 mg

0,01 mg

-100 g

0,025 mg

±0,05 mg

5 s

85 mmø ø 85 mmø ø ø

XA 220/X XA 310/X

220 g 310 g

10 mg 10 mg0,1 mg 0,1 mg

-310 g

0,1 mg 0,2 mg

±0,2 mg ±0,3 mg

3 s3 s 3 s

100 mmø100 mmø100 mm 100 mm

230V / 11V AC or 120V / 11V AC

2 ppm/°C in temperature +18 ° - +30 °C

+10° - +50°C

internal (automatic)

ANTI-VIBRATION TABLE (MILD STEEL OR STAINLESS STEEL)

PRINTER KAFKA

PRINTER EPSON

DENSITY KIT FOR SOLIDS AND LIQUIDS

PC KEYBOARD

ADDITIONAL LCD DISPLAY

EXTERNAL RECHARGEABLE BATTERY WITH CHARGER

COMPUTER SOFTWARE: PW-WIN 2004, RAD-KEY 2000, PIPETTES

FOOT BUTTON FOR TARE OR PRINT FUNCTIONS

Under-hookweighing

Page 6: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

ANALYTICAL BALANCES AS / X

Technical Data :

Additional Equipment:

www.lcgcindia.com

[email protected]

GLPprocedures

%

GLP

Air buoyoancycorrection

Filling

Statistics

Countingpieces

AnimalWeighing

Densitydetermination

Checkweighing

Percentage

Under-hookweighing

Formulation

Balances series epresent standard level of analytical balances. They are equippedwith backlit graphic display. Accuracy and precise measurement of the balances aredetermined by automatic internal calibration, dependant on time and temperaturechanges.

AS/X r

Analytical balances AS/X series are offered with following capacities: 110, 160, 220,310 g with readability of 0,1 mg, 60 g with readability of 0,01 mg and 60/220 g withreadability of 0,01/0,1 mg.

AS/X balances are equipped with optional pipette calibration as well as automatic dooropening system. GLP procedure of these balances is a report from calibration process ofthe balance, and this information has unchangeable shape (no modification possible).User has access to big measuring chamber with sliding doors and sliding top glass.Each balance is equipped with RS 232 output and possibility of connection ofadditional display as standard.

Balances are equipped with possibility of weighing loads outside the main weighingplatform (so called under hook weighing). This means of mass measuring is analternative for loads with non-standard dimensions and shapes and those whichcreate magnetic field. Under hook weighing is also applied for density determinationprocedures.

ANTI-VIBRATION TABLE (MILD STEEL OR STAINLESS STEEL)

PRINTERS: KAFKA 1, KAFKA 1Z, KAFKA SQ

PRINTER EPSON: TM-U220D, TM-U220B, TM-U220A

DENSITY KIT FOR SOLIDS AND LIQUIDS

RACK FOR UNDER-HOOK WEIGHING

ADDITIONAL LCD DISPLAY

EXTERNAL RECHARGEABLE BATTERY WITH CHARGER

COMPUTER SOFTWARE: PW-WIN 2005, RAD-KEY 2000, REC-FS 2000

ISO9001M

AS 110/X110 g

10 mg

0,1 mg

-110 g

0,1 mg

± 0,2 mg

AS 160/X

160 g

10 mg

0,1 mg

-160 g

0,1 mg

± 0,2 mg

AS 220/X220 g

10 mg

0,1 mg

-220 g

0,1 mg

± 0,2 mg

AS 310/X310 g

10 mg

0,1 mg

-310 g

0,2 mg

± 0,3 mg

ø85 mm ø85 mm ø85 mm ø85 mm

Max capacity

Minimal load

Readability

Tare range

Working temperature

Repeatability

Linearity

Sensitivity drift

Stabilization time

Power supplyCalibration

Pan size

+10° - +45°C

3,5 s

2 ppm/°C in temperature +15° - +35°C230V / 11V AC or 120V / 11V AC

internal (automatic)

Page 7: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

ANALYTICAL BALANCES AS

Technical Data :

Additional Equipment:

ISO9001M

%

Countingpieces

Densitydetermination

Percentage

www.lcgcindia.com

[email protected]

Balances have possibility of weighing loads under the weighing pan, i.e. under-hookweighing. In such case the mass of load is measured under the pan. It is an alternative forloads with non-standard dimensions, shapes and loads which create magnetic fieldUnder-hook weighing is also applied for density determination of materials.

Balances series represent standard level of analytical balances. They areequipped with backlit LCD display. Accuracy and precise measurement of thebalances are determined by internal calibration, dependant on time andtemperature changes.

AS

Analytical balances AS series are offered with following capacities: 110, 160, 220,310 g with readability of 0,1mg and 60/220 with readability 0,1/0,01 mg and100 g with readability 0,01 mg.

GLP procedure of these balances is a report from calibration process of the balance,and this information has unchangeable shape (no modification possible). User hasaccess to big measuring chamber with sliding doors and sliding top glass. Eachbalance is equipped with RS 232 output and possibility of connection of additionaldisplay as standard.

ANTI-VIBRATION TABLE (MILD STEEL OR STAINLESS STEEL)

PRINTERS: KAFKA 1, KAFKA 1Z, KAFKA SQ

PRINTER EPSON: TM-U220D, TM-U220B, TM-U220A

DENSITY KIT FOR SOLIDS AND LIQUIDS

RACK FOR UNDER-HOOK WEIGHING

ADDITIONAL LCD DISPLAY

EXTERNAL RECHARGEABLE BATTERY WITH CHARGER

COMPUTER SOFTWARE: PW-WIN 2005, RAD-KEY 2000, REC-FS 2000

Max capacity

Minimal load

Readability

Tare range

Working temperature

Repeatability

Linearity

Stabilization time

Sensitivity drift

Power supply

Calibration

Pan size

+10° - +45°C

AS 60/C/2

60 g

1 mg

0,01 mg

-60 g

AS 60/220/C/2

60 g/220 g

1 mg0,01 mg/0,1 mg

-220 g

AS 110/C/2

110 g

10 mg0,1 mg

-110 g

AS 160/C/2

160 g

10 mg0,1 mg

-160 g

AS 220/C/2

220 g

10 mg

0,1 mg

-220 g

AS 310/C/2 *

310 g

10 mg0,1 mg

-310 g

0,02 mg

± 0,05 mg

6 s

0,02/0,1 mg

± 0,05/0,2 mg

6 s / 3,5 s

0,1 mg

± 0,2 mg

3,5 s

0,1 mg

± 0,2 mg

3,5 s

0,1 mg

± 0,2 mg

3,5 s

0,2 mg

± 0,3 mg

3,5 s

2 ppm/°C in temperature +15° - +35°C

230V / 11V AC or 120V / 11V AC

internal (automatic)

ø85 mm* non-verified balance

Page 8: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

PRECISION BALANCES PS / X

Technical Data :

Additional Equipment:

1www.lcgcindia.com

[email protected]

%

Filling

Statistics

Countingpieces

AnimalWeighing

Densitydetermination

Checkweighing

Percentage

Formulation

Balances series are new types of precision balances with large graphic backlitdisplay and extended user menu. The balances comprise all the features andfunctions of WPX balances and also the newest electronic and mechanicalsolutions and upgraded software.

PS/X

The electronic and mechanical structure is similar to balances series PS. They areequipped with 12 key membrane keyboard, which ensures easier more intuitiveoperation for the user. Pan dimension in this series is 195×195 or 128×128 mmwith a draught shield over the weighing pan . All the balances series PS/X featureinternal calibration system. Each balance is equipped with RS 232 and RS485 outputs,PS/2 socket and additional LCD display plug. GLP procedure is available in a form ofa report from balance calibration. shape of the printout is modifiable. User name,project name, print of date and time, and print frame are user defined onprintout.

Balances are equipped with optional weighing of loads outside the weighing pan, so called under hook weighing. it is an alternative for weighing loads with nonstandard dimensions or made of magnetic substances. Under hook weighing isadditionally applied in case of density determination procedure.

ANTI-VIBRATION TABLE

"KAFKA" THERMAL PRINTER

NEEDLE PRINTER "EPSON"

CITIZEN LABEL PRINTER

DENSITY DETERMINATION KIT FOR SOLIDS AND LIQUIDS

PC KEYBOARD

FOOT BUTTON FOR TARE OR PRINT FUNCTIONS

RACK FOR UNDER HOOK WEIGHING

LCD ADDITIONAL DISPLAY

EXTERNAL RECHARGEABLE BATTERY PACK

SUITCASE FOR BALANCE

COMPUTER SOFTWARE: PW-WIN 2005, RAD-KEY 2000

ISO9001M

Max capacity

Minimal load

Readability

Tare range

Linearity

Repeatability

Pan size

Stabilization time

Sensitivity drift

Working temperaturePower supply

Net weight/Gross weight

Packaging size

PS 200/2000/X*

200/2000 g

20 mg

1/10 mg

-2000 g

±2/10 mg

1/10 mg

2 s / 1,5 s

5/7,1 kg

PS 250/X*

250 g

20 mg

1 mg

-250 g

±2 mg

1 mg

2 s

4,1/6 kg

PS 450/X*

450 g

20 mg

1 mg

-450 g

±2 mg

1 mg

2 s

4,4/6,3 kg

PS 750/X*

750 g

20 mg

1 mg

-750 g

±3 mg

1,5 mg

2 s

4,4/6,3 kg

PS 1000/X

1000 g

20 mg

1 mg

-1000 g

±3 mg

1,5 mg

2 s

4,4/6,3 kg

PS 1500/X

1500 g

500 mg

10 mg

-1500 g

±20 mg

10 mg

1,5 s

5/7,1 kg

PS 2500/X

2500 g

500 mg

10 mg

-2500 g

±20 mg

10 mg

1,5 s

5/7,1 kg

PS 4500/X

4500 g

500 mg

10 mg

-4500 g

±20 mg

10 mg

1,5 s

5/7,1 kg

PS 6000/X

6000 g

500 mg

10 mg

-6000 g

±30 mg

15 mg

1,5 s

5/7,1 kg

PS 8000/X**

8000 g

500 mg

10 mg

-8000 g

15 mg

±20 mg

1,5 s

3,4/5,3 kg

128×128 mm 195×195 mm

2 ppm/°C in temperature +18 - +30 °C

+15 - +30 °C

230V / 11V AC or 120V / 11V AC

570×325×225 mm* scales verified from May 2009** external calibration

Page 9: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

www.lcgcindia.com

[email protected]

PRECISION BALANCES PS / C1

Additional Equipment:

ISO9001M

Technical Data :

Technical Data :

%

Countingpieces

SummingFunction

Densitydetermination

Checkweighing

Percentage

Balances series PS are equipped with 12 key front panel and LCD backlit display. Pan dimension inthis series is 195×195 or 128×128 mm with a glass shield over the weighing pan which protects theload against possible breeze. Each balance has RS 232 and additional display outputs as standard.GLP procedure is available in a form of a report from balance calibration. Shape of the printout ismodifiable. User name, project name, print of date and time, and print frame are user defined on printout.Balances marked as C/1 are equipped with system of external calibration. Balances are equippedwith optional weighing of loads outside the weighing pan, so called under hook weighing. It is analternative for weighing loads with non standard dimensions or made of magnetic substances. Under hook weighing is additionally applied in case of density determination procedure.

ANTI-VIBRATION TABLE

"KAFKA" THERMAL PRINTERNEEDLE PRINTER "EPSON"

DENSITY DETERMINATION KIT FOR SOLIDS AND LIQUIDS

RACK FOR UNDER HOOK WEIGHING

LCD ADDITIONAL DISPLAYEXTERNAL RECHARGEABLE BATTERY PACKSUITCASE FOR BALANCE

COMPUTER SOFTWARE: PW-WIN 2005, RAD-KEY 2000, REC-FS 2000

Stabilization time

Stabilization time

Pan size

Pan size

Max capacity

Max capacity

Minimal load

Minimal load

Readability

Readability

Tare range

Tare range

Working temperature

Working temperature

Repeatability

Repeatability

Linearity

Linearity

Sensitivity drift

Sensitivity drift

Power supply

Power supply

Net weight/Gross weight

Net weight/Gross weight

Packaging size

Packaging size

PS 200/2000/C/1

200/2000 g20 mg1/10 mg

-2000 g1/10 mg

±2/10 mg2 s / 1,5 s

3,3/5,2 kg

PS 110/C/1*110 g20 mg1 mg

-110 g1 mg

±2 mg2 s

3,2/5,1 kg

* verified balances

PS 210/C/1*210 g20 mg1 mg

-210 g1 mg

±2 mg2 s

3,2/5,1 kg

PS 360/C/1360 g20 mg1 mg

-360 g1 mg

±2 mg2 s

3,2/5,1 kg

PS 510/C/1510 g20 mg1 mg

-510 g1 mg

±2 mg2 s

3,2/5,1 kg

PS 750/C/1750 g20 mg1 mg

-750 g1,5 mg

±3 mg2 s

3,2/5,1 kg

PS 1000/C/1

1000 g20 mg1 mg

-1000 g1,5 mg

±3 mg2 s

3,2/5,1 kg

128×128 mm2 ppm/°C in temperature +18 - +30 °C

+15 - +30 °C230V / 11V AC or 120V / 11V AC

570×325×225 mm

* verified balances570×325×225 m

PS 1200/C/1*

1200 g500 mg10 mg

-1200 g10 mg

±20 mg1,5 s

3,2/5,1 kg

PS 2100/C/1*

2100 g500 mg10 mg

-2100 g10 mg

±20 mg1,5 s

3,4/5,3 kg

PS 3500/C/1

3500 g500 mg10 mg

-3500 g10 mg

±20 mg1,5 s

3,4/5,3 kg

PS 4500/C/1

4500 g500 mg10 mg

-4500 g10 mg

±20 mg1,5 s

3,4/5,3 kg

PS 6000/C/1

6000 g500 mg10 mg

-6000 g15 mg

±20 mg1,5 s

3,4/5,3 kg

PS 8000/C/1

8000 g500 mg10 mg

-8000 g15 mg

±20 mg1,5 s

3,4/5,3 kg230V / 11V AC or 120V / 11V AC

195×195 mm2 ppm/°C in temperature +18 - +30 °C

+15 - +30 °C

Page 10: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

PRECISION BALANCES PS / C2

www.lcgcindia.com

[email protected]

ISO9001M

Additional Equipment:

Technical Data :

Technical Data :

Balances series PS are equipped with 12 key front panel and LCD backlit display. Pan dimension isthis series is 195×195 or 128×128 mm with a glass shield over the weighing pan which protects theload against possible breeze. Each balance has RS 232 and additional display outputs as standard.GLP procedure is available in a form of a report from balance calibration. shape of the printout ismodifiable. User name, project name, print of date and time, and print frame are user defined onprintout.Balances marked as C/2 are equipped with system of internal automatic calibration.Balances are equipped with optional weighing of loads outside the weighing pan, so called underhook weighing. it is an alternative for weighing loads with non standard dimensions or made ofmagnetic substances. Under hook weighing is additionally applied in case of density determination procedure.

Anti-vibration Table

"kafka" Thermal PrinterNeedle Printer "epson"

Density Determination Kit For Solids And Liquids

Rack For Under Hook Weighing

Lcd Additional DisplayExternal Rechargeable Battery PackSuitcase For Balance

Computer Software: Pw-win 2005, Rad-key 2000, Rec-fs 2000

Stabilization timePan size

Max capacityMinimal loadReadabilityTare range

Working temperature

RepeatabilityLinearity

Sensitivity drift

Power supplyNet weight/Gross weightPackaging size

Stabilization timePan size

Max capacityMinimal loadReadabilityTare range

Working temperature

RepeatabilityLinearity

Sensitivity drift

Power supplyNet weight/Gross weightPackaging size

PS 200/2000/C/2200/2000 g20 mg1/10 mg

-2000 g1/10 mg

±2/10 mg2 s / 1,5 s

4,3/6,2 kg

PS 210/C/2*210 g20 mg1 mg

-210 g1 mg

±2 mg2 s

3,5/5,4 kg

PS 360/C/2*360 g20 mg1 mg

-360 g1 mg

±2 mg2 s

3,5/5,4 kg

PS 600/C/2*600 g20 mg1 mg

-600 g1 mg

±2 mg2 s

3,7/5,6 kg

PS 750/C/2750 g20 mg1 mg

-750 g1,5 mg

±3 mg2 s

3,7/5,6 kg

PS 1000/C/21000 g20 mg1 mg

-1000 g1,5 mg

±3 mg2 s

4,3/6,2 kg

PS 110/C/2*110 g20 mg1 mg

-110 g1 mg

±2 mg2 s

3,5/5,4 kg

* verified balances

230V / 11V AC or 120V / 11V AC

570×325×225 mm

128×128 mm2 ppm/°C in temperature +18 - +30 °C

+15 - +30 °C

PS 1200/C/2*1200 g500 mg10 mg

-1200 g10 mg

±20 mg1,5 s

4/5,9 kg

PS 2100/C/2*2100 g500 mg10 mg

-2100 g10 mg

±20 mg1,5 s

4,6/6,5 kg

PS 3500/C/2*3500 g500 mg10 mg

-3500 g10 mg

±20 mg1,5 s

4,6/6,5 kg

PS 4500/C/2*

4500 g500 mg10 mg

-4500 g10 mg

±20 mg1,5 s

4,6/6,5 kg

PS 6000/C/2*6000 g500 mg10 mg

-6000 g15 mg

±20 mg1,5 s

4,6/6,5 kg

195×195 mm

* verified balances

2 ppm/°C in temperature +18 - +30 °C+15 - +30 °C

230V / 11V AC or 120V / 11V AC

570×325×225 mm

%

Countingpieces

SummingFunction

Densitydetermination

Checkweighing

Percentage

Page 11: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

MOISTURE ANALYZER MAC

Technical Data :

Additional Equipment:

ISO9001

www.lcgcindia.com

[email protected]

Moisture analyzers are measuring devices specially designed fordetermination of moisture content of relatively small samples of variousmaterials.

· easy access due to backlit LCD display· drying profile (standard, mild, step, rapid).· finish mode (manual, humidity stabilization, automatic, time defined).· GLP/GMP printouts and reports· halogen or infrared lamps· standard and non-standard applications· optimization of work due to halogen lamps mode

Maximal capacity of moisture analyzer series MAC is 210 g /1 mg. Moisture content is measured with accuracy 0,01% (0,001% for samples up to 1,5g). Maximal drying temperature equals 160°C (an extra cost option is moisture analyzer for drying with temperature 250°C).

Moisture analyzer MAC enables:

ANTI-VIBRATION TABLE

"KAFKA" THERMAL PRINTER

NEEDLE PRINTER "EPSON"

PC KEYBOARD

DISPOSABLE PANS

CONTROL TERMOMETER

CALIBRATION WEIGHT

COMPUTER SOFTWARE: PW-WIN 2005, RAD-KEY 2000

Max capacity

Reading unit

Tare rangeMax mass of sampling

Accuracy of moisture reading

Repeatability

Drying chaber dimensions

Pan size

Range of drying temperature

Drying modes

Auto switch off options

Aditional functions

Power of heating device

Working temperature

Power suply

Display

+/-0,1% (sample 2g), +/-0,02% (sample 10g)

120×120×20 mm

ø 90 mm

max. 160° C (optionally max. 250° C)4 drying modes (standard, quick, stepped, mild)

3 modes(manual, automatic, time defined)

sample identification

400 W

+15° - +40 °C

230V AC or 120V AC

LCD (backlit)

MAC 50/1

50 g

0,1 mg

-50 g

50 g

MAC 50

50 g

1 mg

-50 g

50 g

MAC 110

110 g

1 mg

-110 g

110 g

MAC 210

210 g

1 mg

-210 g

210 g

0,01 / 0,001 % 0,01 %

Page 12: Laboratory Balances - LCGCINDIAlcgcindia.com/images/LCGC RADWAG FINAL BOOK.pdf weighing@lcgcindia.com Balances have possibility of weighing loads under the weighing pan, i.e. under-hook

LCGC - RADWAG WEIGHING SOLUTIONS P. LTD

Web: www.lcgcindia.com

Chennai: 93802 14500E-mail: [email protected]

Chandigarh: 93574 48808E-mail: [email protected]

Head Office:Hyderabad:

Phone: 040-64594253 /52E-mail: [email protected]

Main Branch: Mumbai

E-mail: [email protected]: 022-24082070/80/90

Ahmedabad: 98253 17798E-mail: [email protected]

Bangalore: 98440 21071E-mail: [email protected]

New Delhi: 98103 87404E-mail: [email protected]

Goa: 98232 35457E-mail: [email protected]

Indore: 98260 84686Email: [email protected]

Pune: 93730 15242E-mail: [email protected]

Surat: 98241 49149E-mail: [email protected]

Kolkatta : [email protected]

Plot No: 57, Road No.5,Aleap Industrial Estate, Nr.Pragati Nagar,Kukatpally,Hyderabad - 500 072

Chandigarh

Delhi

Ahmedabad

IndoreKolkatta

Surat

Mumbai

Pune

Goa

Hyderabad

Bangalore