KUST ELEKTRONIK:LM1010 Precission LCR Meter
LM1010 is a precision LCR meter with high accuracy, good stability, and wide measurement range. Controlled by a 16 bits MPU, LM1010 can be used for evaluating LCR compo-nents, materials and semiconductor devices over a wide range of frequencies (50 Hz to 100kHZ) and test signal levels (50mV to 1V, with 10mV resolution ). With its powerful functions, excellent performance, easy to read LCD display and easy menu operation, LM1010 is suitable for high speed measurement needed on product line and high accuracy and stability measurement need in laboratory. By using the Handler interface, IEEE488 interface (optional), and RS232C interface, LM1010 ca be easily implemented to automatic test system or computer remote control.• AC Signal: 20mV - 1V• Up to 40 measurment/sec.• 4.3” LCD Display• Test Signal Level Monitor• 30Ω or 100Ω output Impedance• Parameters: L,C, R, |Z|, D, X, θ, Q, DCR• Basic Accuracy: 0.05%• Comms: RS232, HANDLER,USB host, GPIB(optional)• LED Indication of the result Green - PASS / Red - FAIL• Beeper - result sound indication• Trigger Function: Manual, Internal, External, BUS• Internal Memory for 100 Instrument Settings• Key-Lock with password function• Correction/Compensation Function
2 KUST Elektronikwww.kust .de | www.kust-e lektronik .com
Measurement functionLCR test parameters | Z |, | Y |, C, L, X, B, R, G, D, Q, θ, DCR
Test frequency:
LM1010A50,60,80,100,120,150,200,250,300,400,500,600,800,1K, 1.2K, 1.5K, 2K, 2.5K, 3K, 4K, 5K, 6K, 8K, 10K, 12K, 15K, 20K, 25K, 30K, 40K, 50K, 60K, 70K, 80K, 90K, 95K 100K, (Hz), total of 37
Basic accuracy 0.05%
The equivalent circuit Series, parallel
Mathematical functions The absolute value of the deviation, percentage deviation
Range mode Automatic, hold, manually select
Trigger Internal, manual, external, bus
Measurement speed (≥1kHz) Fast: The fastest 40 times / sec speed: 11 times / sec, Slow: 3 times / sec.
The average number of 1-255
Delay time 0-60s, with 1ms step
Correction: Open / short / load
Test side configuration Five end
Display mode Direct reading, Δ, Δ%, V / I (measured voltage / current monitoring)
Display 5 resolution, 4.3-inch LCD Monitor
Test signal
Test frequency 50 Hz - 100 kHz
Output Impedance 30 Ω, 100Ω selectable
Test signal level 20mV – 1V Accuracy: 10%
Display range
| Z |, R, X 0.01mΩ - 99.999 MΩ
| Y |, G, B 0.0001μS - 99.999 S
C 0.0001 pF - 9.9999 F
L 0.0001 μH - 999.99 H
D 0.0001 - 9.9999
Q 0.0001 --99999
θ (DEG) -179.99º - 179.99 º
θ (RAD) 3.14159 - 3.14159
DCR 0.01mΩ - 99.999 MΩ
OtherOutput 4 Terminals – 4 wire Connection. Ground terminal.
Front USB Screen Shots, Set-up files storage, Firmware Update
Comparator function Sorting functionResult Indication Green / Red LED, soundMonitor Voltage & CurrentComparator function Sorting functionMemory 100 internal instrument settings for storage / recall, external 500 groups
Interface Standard RS232C, HANDLER, USB HOST, optional GPIB
LM1010 requires the fol-lowing power source:Voltage: 198 to 242 VacFrequency: 47 to 66 HzPower: 50 VA maximum
LM1010 must be oper-ated under within the following environment conditions:Temperature: 0 to 40 Humidity: less then 90% RH
KA1002 :4 terminal Kelvin test leads
KA1001: 4 terminal test fixture
KA1008: Gilded shorting plate
Every LM1010 LCR Meter is supplied with main’s lead and factory calibration.