K13250 LICO 100 Portable Color Measurement System

28
1 K13250 LICO 100 Portable Color Measurement System Operating Instructions Koehler Instrument Company, Inc. 1595 Sycamore Ave Bohemia, New York 11716 http://www.koehlerinstrument.com e-mail: [email protected]

Transcript of K13250 LICO 100 Portable Color Measurement System

1

K13250 LICO 100 Portable

Color Measurement System

Operating Instructions

Koehler Instrument Company, Inc.

1595 Sycamore Ave

Bohemia, New York 11716

http://www.koehlerinstrument.com

e-mail: [email protected]

2

Dear client,

Congratulations for the purchase of this high quality instrument from Koehler Instrument Company.

LICO® 100 is a modern and sophisticated instrument for your lab. LICO® 100 includes state-of-the-

art technology and a very efficient user surface for easy familiarization. Still we recommend

reading these operating instructions thoroughly, as they give important information you should know

to handle this product adequately.

Thank you for your confidence, and enjoy working with your new LICO® 100.

Yours,

Koehler Instrument Company, Inc.

Concise Operating Instructions

1) Connect LICO® 100 and printer LD500, if desired, to the mains.

Then plug the printer cable that came with the printer into the

RS232-port of LICO® 100. Turn LICO® 100 on by pressing key

OK and the printer by toggling the green switch to the right.

2) After zero measurement wait for the display of LICO® 100 to

prompt you to insert the calibration cuvette.

3) Fill a cuvette with distilled water and put it into the cuvette

compartment. The display reads e.g.:

4) Take the water cuvette out and put one containing sample liquid

into the cuvette compartment.

5) After measuring, the color values of the sample are indicated

in the configured color system.

6) Now press the -key until the desired color system is

displayed.

Zero meas.m.

Please wait

Pleas ins. cuv.

for calibration

Calibration

Please wait

Pleas ins. cuv.

for sample

70.5

Iodine color value

ÐDIM ¢Conf 11°

3

CONTENTS

A Instrument DescriptionA1 General............................................................................................................................. 4A2 Instrument Design........................................................................................................... 5A3 Instrument Set-Up ........................................................................................................... 6A4 Reference Standards ...................................................................................................... 6

B OperationB1 Control And Functional Elements .................................................................................. 7B2 Initial Start-Up of the Instrument .................................................................................... 9B3 Printer Start-Up................................................................................................................ 9B4 Battery Mode ................................................................................................................. 10B5 Turning-Off .................................................................................................................... 11B6 Photometric Preparation............................................................................................... 11

C Color MeasurementC1 Calibration...................................................................................................................... 12C2 Color Measurement....................................................................................................... 14C3 Iodine Number ............................................................................................................... 15C4 Hazen Color Number (APHA-Method) .......................................................................... 15C5 Gardner Color Number.................................................................................................. 16C6 Mineral Oil Color Number ASTM D 1500 And ISO 2049 ............................................... 17C7 Saybolt Color Number (ASTM D 156) .......................................................................... 17C8 Configuration................................................................................................................. 18

D Technical NotesD1 Error Messages ............................................................................................................. 22D2 Service Offices of Dr. Lange GmbH & Co. KG............................................................. 22D3 Maintenande and Test Media Control .......................................................................... 23D4 Maintenance Contract................................................................................................... 23D5 Lamp Exchange ............................................................................................................ 24D6 Program-Module Card (BIT-BOX) of LICO® 100 ........................................................... 25D7 RS232C-Interface of LICO® 100 .................................................................................... 26D8 Technical Data............................................................................................................... 26

E AnnexE1 Spares and Accessories............................................................................................... 27E2 Regulations and Standards......................................................................................... 28

Koehler Instrument Company, Inc. LICO® 100 K13250

4

A Instrument Description

A1 General

LICO® 100 is a universal colorimeter for liquids suitable for manifold industrial applications. It is easy

to handle and has a sturdy design.

But what is really the reason why objective color measurement of liquids is so important? The human

eye is able to perceive about 100 spectral colors in approximately 100 degrees of brightness and

saturation, which results in about 1 million perceptible shades. Therefore, hues are difficult to describe

by subjective visual comparison.

The specific color of a product is often regarded as a sign of purity and quality, although that is not

necessarily true. Clients often interpret deviations from the shade of an earlier production as defective

product or supply and reject it preventively. In order to avoid such trouble, objective color assessment,

a uniform colorimetric system and reliable instrument are required for production and quality control of

transparent colored liquids.

New Standard EN 1557

EN 1557 (Colorimetric characterization of transparent colored liquids) now also defines color

measurement of liquids on the basis of the time-tested colorimetric standards of DIN 5033. The

principle is to measure the X-, Y- and Z-transmissions. The sample is illuminated with a halogen lamp

corresponding to standard illuminant C. On this basis, Koehler developed the LICO® 100 for exact and

reproducible measurements of transparent yellow-cast liquids.

Koehler Instrument Company, Inc. LICO® 100 K13250

5

A2 Instrument Design

LICO® 100 measures transmissions of all transparent media. It operates with a halogen lamp

producing standard illuminant C and uses the 2° standard observer as defined by DIN-standard 5033.

A reference beam path compensates measuring values modified by driftings of lamps and

temperatures during operation. Therefore, LICO® 100 provides an excellent long-term constancy and

reproducible measuring results. The universal cuvette compartment suitable for 50 x 10mm square

cuvettes and 11mm round glass cuvettes permits an optimum adaptation of LICO® 100 to the

measurement tasks.

In just one measurement, LICO® 100 evaluates by to the latest ISO-standards giving e.g. Iodine, Hazen

or Gardner values. The basical instrument includes one optional color number with the possibility of

adding-on a random combination of color numbers later or at once. These color systems are

available: Iodine number, Hazen number, Gardner number, Saybolt number and ASTM D 1500 number.

To add-on more color systems later, a program-module-box (bit-box) is supplied. The description of

this bit-box is given in chapter D6.

Fig. 1: Measuring principle of LICO® 100

Koehler Instrument Company, Inc. LICO® 100 K13250

6

A3 Instrument Set-Up

Unwrap the instrument carefully and check it for missing parts (delivery note). Please keep the packing

box to pack the instrument safely in case it has to be transported again later.

LICO® 100 should be placed on a level desk surface. The instrument comes with a power supply unit

and adapter for the socket (100 – 240 Volt / 47 – 63 Hz) and is turned on by either pressing the OK-

key or the grey menu key.

Do not turn on and off too fast. Wait for 20 seconds before turning on again not to damage the

electronic or mechanic system of the instrument.

The blue lamp protection cover on the upper left part of the instrument serves to carry the heat off

the lamp and must not be covered.

As LICO® 100 is a sophisticated optical instrument, avoid the following at any rate:

• Do not use the instrument in direct sunlight.

• Do not allow water, inflammable or metallic objects to enter the instrument.

• Do not put filled cuvettes or heavy objects on the housing.

• Do not use the instrument in excessively moist, dusty or vibrating surroundings.

• Do not open the instrument. There are no parts in the housing for the user to check.

• If you have any questions of problems, please contact Koehler’s Technical Service

Department.

A4 Reference Standards

In general, an absolutely transparent medium similar to the sample serves as a calibration or reference

standard for colorimetry on the transmission principle. To measure the color of liquids, distilled water is

filled into a cuvette similar to the measuring cuvette and used as a reference standard. In some

exceptional cases where the density of the substance to be measured differs from that of the distilled

water, calibration is made against a colorless measuring medium. Calibration assigns the reference

standard a transmission of 100% (or Iodine number = 0, Hazen = 0, etc.). As this assignment never

changes, 100 no calibration value needs to be set for LICO®. Instead, just fill a cuvette with the

reference standard and put it into the cuvette compartment to perform calibration.

Koehler Instrument Company, Inc. LICO® 100 K13250

7

B Operation

B1 Control And Functional ElementsFigures 2 and 3 show the control and functional elements of LICO® 100.

Fig. 2: Front view

Fig. 3: Back view

Bit-Box

compartmentBattery cover Mains socket

Printer

port

Display

Lamp cover

Cal- and Measure key

Cuvette compartment

Menu key

OK-key

Cursor keys

Koehler Instrument Company, Inc. LICO® 100 K13250

8

LICO® 100 is operated with 8 keys. The following table shows the keys and displays in the operating

instructions.

Definition Representation

Cursor keys Press left, right, upper or lower cursor key { } $ &

Calibration and

Measure key

Press blue key (Calibration)

Press green key (Measurement) CAL calibration key (blue)

Meas measure key (green)

Menu key Turn instrument on

Quit submenu option Menu key (grey)

OK-Key Turn instrument on

Confirm or call the selected actionOK

Display and input

area

Display of current values, data, submenus and input

areas of the single options. The input area is

marked by a window inviting inputs.

Aut.shut off:

30 min

Menu selection Selecting an option with the cursor keys or key OK

e.g..:ÐConfig, another window opens offering

more options:

Switch Off Config Printer

Error messages It is necessary to acknowledge every error

message by pressing the grey menu key, or the

instrument stops operating.

Min. limit: 10

Koehler Instrument Company, Inc. LICO® 100 K13250

9

B2 Initial Start-Up of the InstrumentFor start-up plug-in the mains plug and press the OK-key or grey menu key. The port for the power

supply unit is at the rear of the instrument. When the instrument is on, the display indicates instrument

type and number of EPROM-version while the self-check is running. When self-check and zero

measurement are over (approx. 20 sec.) you are prompted to insert a calibration or sample cuvette.

B3 Printer Start-UpThe supply voltage for the printer is turned on and off with the green toggle switch �� on the right of

the printer’s front. The printer is on when the black key �� is in neutral position (1) and off when key ��

is at (0). Paper feed is done pressing the '↑' –part of key �� next to the printing unit.

B3.1 Connection to LICO® 100

The printer is connected by plugging the supplied connection cable to the RS232-port on the rear side

of LICO® 100.

B3.2 Change paper

Turn the printer on and open the flap �� behind the printing unit. Put a new reel of paper into the paper

container and unroll approx. 10cm of it. Then lead the free beginning of the paper into the slot until

perceiving a slight resistance. Press the paper feed key and push the paper on carefully. Hold the key

down until the paper passes the tear bar. Finally close the flap again.

LICO 100V1.00

Pleas ins. cuv.

for sampleLICO 100

V1.00 15:20 30.06.00

Koehler Instrument Company, Inc. LICO® 100 K13250

10

B4 Battery ModeLICO® 100 can be used with power supply unit or rechargeable batteries. For the latter mode, open

the cover of the battery compartment and insert the batteries as follows:

1. Insert the battery with the switch to the right into the holder until only one contact remains visible.

2. Shift the battery from the left side over the contacts, passing a slight resistance.

3. A beep indicates that the batteries are placed and ready for operation (only when the instrument

is not connected to the mains).

Attention: The contacts must not get into contact with metal or greasy parts.

To recharge the batteries, insert them into the compartment and connect LICO® 100 to the mains with

the power supply unit. We recommend a continuous loading time of 14 hours before the first use. To

exploit the battery capacity best, charge and discharge it always entirely. The display of LICO® 100

informs about the current discharge degree of the battery when the instrument is turned on. When the

battery is discharged, a beep and a display message are given. The message must be acknowledged

Fig. 4: Insertion of rechargeable batteries

Contact

Battery switch

Battery

Koehler Instrument Company, Inc. LICO® 100 K13250

11

with the grey menu key to go on working for a short while. The display message “battery empty” means

that the battery is definitely discharged and needs recharging.

B5 Turning-OffLICO® 100 turns off automatically by a timer function. Manual turning-off is possible, too.

Manual Turning-Off

Press Cursor-Key & in the measuring menu and select option

„Switch Offf“ with the OK-key. Then the instrument turns off.

Turning-Off Via Timer

On delivery, the timer of LICO® 100 is set to automatic turning-off

time of 30 minutes. The automatic turning-off is only active if no

measurement was made within the given period and no cuvette is

in the compartment. The setting can be modified within the tolerances of 10 to 60 minutes.

B6 Photometric PreparationThe photometric preparation requires utmost care, because careless sample preparation is the major

source of trouble, and not the handling of the instrument itself.

Cleanliness is of vital importance in photometry. Therefore, clean the glass cuvettes immediately after

use. Employ only optically perfect samples for measuring. They must be clean and not show any signs

of opaqueness.

Air bubbles in the sample or at the cuvette sides influence the measuring result. They can be avoided

by filling the liquid slowly into the cuvette. Pasty or solid products should be melted and filled-in without

bubbles, if they cannot be measured as they are.

B6.1 Sampling And Sample Preparation

Take a sample from the product you want to measure in accordance with DIN 53242 part 1. Eliminate

turbidities by heating or filtration. Heat partly solid samples before measuring in order to dissolve the

solid share in the liquid one. The preparation must not cause any chemical changes in the sample.

During the measurement there should be no bubbles in the sample.1

• There are several cuvette types available for color measurement with LICO® 100, differing by

material (glass, polystyrene or PMMA2) and path length (11mm and 50mm).

1 Acc- to DIN 6162, ed. 12.81, para. 5 a. 6

LICO 100V1.00

14:5017.02.00

Switch Off Config Printer

Koehler Instrument Company, Inc. LICO® 100 K13250

12

• The program calculates automatically Iodine, Hazen, Gardner, Saybolt and ASTM D 1500-

numbers under consideration of the cuvette type employed.

• For the 11mm disposable round glass cuvettes, a dry thermostat is available to warm up to 15

cuvettes to 100°C or 148°C.

C Color Measurement

Whenever the instrument is turned on, a zero measurement is made, taking the empty value (i.e. the

pure transmission values without cuvette inserted) of the optical beam path and comparing it with the

empty value obtained in the previous zero measurement. If the difference between both empty values

is near zero, you can start measuring without calibration. Calibration is only required in case of

considerable deviations of the empty value due to e.g. fluctuations of temperature. The calibration

remains valid until the next recalibration. LICO® 100 features a calibration memory for both path

lengths employed (50mm square cuvette and 11mm round cuvette). If you have to use two different

path lengths, owing to the nature of your samples (e.g. bright samples with Hazen values below 50 or

strongly colored samples with greater Gardner values), it is possible to calibrate both path lengths and

then toggle between them. LICO® 100 uses the corresponding set of calibration data automatically.

C1 Calibration

LICO® 100 features a calibration memory for 11mm round cuvettes and 50mm square cuvettes. To be

on the safe side, calibrate LICO® 100 at least every day. In the calibration, the employed medium is

assigned a transmission of 100% over the whole spectral range. Observe the following rules for

calibration:

• The cuvette type selected for calibration must be identical to the one used for measurement,

otherwise wrong Iodine, Hazen and Gardner values are given.

• Use always clean cuvettes and fresh distilled water for calibration.

2 Polymethylmethacrylate

Koehler Instrument Company, Inc. LICO® 100 K13250

13

Instrument calibration for color measurement

1. Turn-on instrument and printer, if desired. After self-check and

zero measurement, LICO® 100 might ask for calibration. If

LICO® 100 is already on, make sure the cuvette compartment is

empty and press the blue CAL - key. The display reads:

2. Fill a clean cuvette (material and path length to suit your

samples) with distilled water and put it into the measuring

compartment.

3. LICO® 100 detects the cuvette type automatically.

4. Now press the blue CAL - key again to trigger calibration.

5. When the calibration is over and the printer is on, date, time and selected cuvette type are

printed, e.g. :

Then LICO® 100 prompts „Please insert sample cuvette“. Now you can insert your sample cuvette

directly. As soon as you have done so, LICO starts measuring automatically. To calibrate one more

path length, remove the cuvette and press the blue CAL -key again.

Please calibrate carefully to avoid mismeasurements caused

by faulty calibration.

Zero meas.m.

Please wait

Pleas ins. cuv.

for calibration

Calibration

Please wait

Calibration with 11 mm circular cuv. ok!

12:00 18.08.00

70.5

Iodine color value

ÐDIM ¢Conf 11°

Koehler Instrument Company, Inc. LICO® 100 K13250

14

C2 Color Measurement

The colors of all transparent liquids can be measured, using 50mm square cuvettes or 11mm round

glass cuvettes, depending on their transmittances. If required, select another color system for printer

output in the configuration menu (cp. page Error! Bookmark not defined.).

Instrument preparation

1. Turn instrument and printer on.

2. Calibrate the instrument as described in chapter. C1.

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment.

Measurement starts automatically.

2. After measuring, the color values of the sample will appearon the display

and be printed, e.g.

2.6 Iodine color value

507 Hazen color value

3. Now press the } -key (DIM) until the desired color system

appears on the display.

4. The color system will be displayed during the measurements

to follow.

2.6

Iodine color value

ÐDIM ¢Conf 11°

507

Hazen color value

ÐDIM ¢Conf 11°

Koehler Instrument Company, Inc. LICO® 100 K13250

15

C3 Iodine Number

DIN 6162 defines the Iodine color number as mg of iodine per 100ml of potassium iodide solution.

Color matching by the Iodine number serves to judge the shade depth of clear liquids like e.g. tensides,

plasticizers, resins, oils and fatty acids, whose colors resemble that of an equally layered solution of

potassium iodide. For Iodine numbers less than or about 1, the Hazen number according to DIN-ISO

6271 should be preferred.3 Sampling and preparation are described in chapter B6. A satisfactory

precision of the Iodine value is achieved with a 11mm round glass cuvette.

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment. Measurement starts auto-

matically.

2. After the measurement, the color value is indicated on the

display.

3. Now press the } -key (DIM) until the Iodine number appears on the display.

C4 Hazen Color Number (APHA-Method)

The Hazen color number (DIN-ISO 6271, also known as "APHA-Method” or “Platinum-Cobalt-scale”)

is defined as mg of platinum per 1L of solution. The Hazen parent solution is composed of 1.246g

potassium-hexachloroplatinate (IV) and 1.00g cobaltous (II)-chloride dissolved in 100ml of hydrochloric

acid and filled up with distilled water until reaching 1000ml. The Hazen color scale serves to determine

the colors of nearly water-clear products. It has a narrower gradation in the light yellow range than the

Iodine scale and reaches even water-clear casts.

Attention:

Samples with Hazen values between 50 and 1000 can be measured with sufficient

accuracy in 11mm round cuvettes.

Hazen values < 50 should be mesured in 50mm cuvettes,

Hazen values < 10 should be measured with the cuvette which was used for

calibration, because nearly water-clear products are susceptible to giving faulty measuring

results due to cuvette tolerances.

3 acc. to DIN 6162, Edition 12.81, para. 1

70.5

Iodine color value

ÐDIM ¢Conf 11°

Koehler Instrument Company, Inc. LICO® 100 K13250

16

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment. Measurement starts auto-

matically.

2. After the measurement, the color value is indicated on the display.

3. Now press the } -key (DIM) until the Hazen number appears on the display.

C5 Gardner Color Number

The Gardner color number is defined in DIN ISO 4630. The lighter Gardner color standards (1 to 8)

are based on potassium-chloroplatinate solutions, color numbers 9 to 18 on ferric(III)-chloride,

cobaltous(II)-chloride and hydrochloric acid solutions. LICO® 100 determines the Gardner number at

all cuvette path lengths (in 50mm cuvettes only values between 0 and 4), with 11mm cuvettes providing

sufficient accuracy.

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment.

2. After the measurement, the color value is indicated on the

display.

3. Now press the } -key (DIM) until the Gardner number

appears on the display.

20

Hazen color value

ÐDIM ¢Conf 50°

6.5

Gardner color value

ÐDIM ¢Conf 11°

Koehler Instrument Company, Inc. LICO® 100 K13250

17

C6 Mineral Oil Color Number ASTM D 1500 And ISO 2049

The Mineral Oil Color Scale is used for matching the colors of mineral oil products like lubricating oil,

fuel oil, diesel oil and paraffins. The color scale begins at 0 for water-clear, colorless oils and ends at

8 for very dark brown oils. Visual matching is in steps of 0.5 (0.5 1.0 1.5...). LICO® displays the results

in steps of 0.1. For samples of stronger hue than color glass 8,

color value D8 is displayed.

GG Owing to the intensive color gradation of the ASTM4 D 1500 color

number, this is only determined for the 11mm round cuvette. There

is no determination of path length 50mm.

Measuring procedure

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment.

2. After the measurement, the color value is indicated on the

display.

3. Now press the } -key (DIM) until the ASTM- number

appears on the display.

C7 Saybolt Color Number (ASTM D 156)

The Saybolt color scale is used to match colors of refined oil products like fuel and kerosene or also

petroleum waxes and pharmaceutical white oils. Its colorimetric characteristics are comparable with

those of the Hazen scale (APHA). The Saybolt scale begins at +30 (brightest coloration,

corresponding to approx. 8-10 Hazen) and ends at -16 (stongest coloration, corresponding to approx.

350 Hazen).

FF Measurements can be made with cuvettes of 11mm or 50mm path lengths. The greater the path

length, the better the measuring accuracy (50mm recommended).

4 American Society for Testing and Materials

4.5

ASTM D 1500

ÐDIM ¢Conf 11°

24

Koehler Instrument Company, Inc. LICO® 100 K13250

18

1. Fill the sample liquid into a cuvette and insert it into the

corresponding compartment.

2. After the measurement, the color value is indicated in the

configured color system on the display.

3. Now press the } -key (DIM) until the Saybolt number

appears on the display.

C8 Configuration

The configuration serves to select the color systems for the print-out and general instrument settings

like e.g. date/time, language, etc.

Press cursor-key & for configuration.

The selection menu shown at the right opens. Move the bar to

the menu options with cursor-key & or $ . The option in

reverse presentation can be selected by pressing key OK .

C8.1 Turning-off

After selecting option Turn-off and pressing key OK , LICO® 100

is turned off.

C8.2 Configure

After selecting option Config. and pressing key OK , another menu window opens. The small arrow

↓↓ in the frame window at the right indicates that there are more options to follow.

Saybolt color value

ÐDIM ¢Conf 50°

LICO 100V1.00

14:5017.02.00

Switch Off Config Printer

LICO 100V1.00

14:5017.02.00

Switch Off Config Printer

LICO 100V1.00

14:5017.02.00

Switch Off Config Printerer Language

Date Timer Display

Koehler Instrument Company, Inc. LICO® 100 K13250

19

Language selection

These languages are at choice: German (D), English

(GB), French (F), Swedish (S), Spanish (E), Dutch (NL), Italian

(I), Polish (PL) and Danish (DK).

Move the bar to the desired language and select it with key OK

.

Date and time

After selecting option Date and pressing key OK , a window

opens to set date and time. Move the blinking cursor to the digit

you want to change by cursor-key { or } . Cursor-keys $ or

& increase or decrease the blinking digit by one. Finish the

setting with the OK -key.

LICO 100 Turn-off Config Printer

Language select: D . GB F S E

LICO 100V1.00

14:5017.02.00

Turn-off Config Printer

Date Time

17.02.00 14:50

Koehler Instrument Company, Inc. LICO® 100 K13250

20

Timer

Selecting option Timer and pressing key OK opens a window to

set timer functions "Aut.shut off" and "delay meas.". Set

the desired value with cursor-keys $ and & .

Auto shut off time

Selecting option "Aut.shut off" and pressing key OK

opens a window to set the turn-off time. Move the blinking cursor

to the digit you want to change by cursor-key { or } Cursor-

keys $ or & increase or decrease the blinking digit by one.

Finish the setting with the OK -key.

Measurement delay

Selecting option "delay meas." and pressing key OK opens

a window to set the Measurement delay.

Move the blinking cursor to the digit you want to change by

cursor-key { or } . Cursor-keys $ or & increase or

decrease the blinking digit by one. Finish the setting with the

OK -key.

LICO 100

14:5017.02.00

Turn-off Config Printer

Adjust timer: Aut.shut off delay. meas.

LICO 100V1.00

14:5017.02.00

Aut.shut off:

30 min

LICO 100V1.00

14:5017.02.00

delay. meas.

00 sec

Koehler Instrument Company, Inc. LICO® 100 K13250

21

Display

Selecting option "Display“ and pressing key OK opens a

window for display settings of "Contrast" and "Lighting".

Make your choice with the cursor-keys $ and & and confirm it

with key OK .

Printer

At this option, you can set more color systems for printer output.

The system selected at display is marked n and will be printed in

the first line. To add more color systems, move the bar to the

desired system with cursor-keys $ and & and confirm your

choice with the OK - key.

When the cursor passes the upper or lower margin of the menu, it

pages on and shows more options. Here you can add date and

time to the measurement.

To finish the configuration of the printer output, select "back" and

press key OK .

The program returns to the beginning of the configuration.

LICO 100V1.00

14:5017.02.00

Turn-off Config Printer

Adjust display:Contrast....Lighting

LICO 100V1.00

14:5017.02.00

Turn-off Config Printer

†Iodine number ¡Hazen number †Gardner no. †Saybolt no.

LICO 100V1.00

14:5017.02.00

Turn-off Config Printer

†ASTM D 1500 †Date / Time back

Koehler Instrument Company, Inc. LICO® 100 K13250

22

C8.3 Instrument check

LICO® 100 can be checked with a certified Koehler test solution set addista®-color in compliance with

ISO 9000ff.

D Technical Notes

D1 Error MessagesThe following display messages indicate errors of operator or instrument.

Result outside the measuring range.

There is a cuvette left in the cuvette compartment.

Put new reel of paper into LD 500.

The lamp is probably broken.

If one of these error messages appears, please check whether e.g. the printer is connected properly,

calibration is correct and the sample cuvette at its place. If you still cannot eliminate the error, contact

our service department or field service.

D2 Service Office of Koehler Instrument Company

Koehler Instrument Company1595 Sycamore Avenue

Bohemia, NY 11716

tel: 800-878-9070 / 631-589-3800

fax: 631-589-3815

e-mail: [email protected]

***

Please remove

cuvette!

Insert paper

Error!CalibrationNo SignalLamp ?

Koehler Instrument Company, Inc. LICO® 100 K13250

23

D3 Maintenance and Test Media ControlEven a perfected product like LICO® 100 from Koehler requires careful handling. Use a dry cloth or a

mild household cleaner for the housing, but never solvents like naphta, acetone or the like. The optical

elements should be cleaned by our maintenance staff.

A certified set of test solutions is available for test media control according to ISO 9000.

The test set comprises six bottles

containing 50ml of test solution each, a

quality control certificate according to DIN

55350 part 18, 4.1.2 and a test protocol

for the documentation of reference, actual

and tolerance values.

LICO® 100 is tested as described in

enclosed test instruction. The instrument

should be tested in regular intervals of 1 to

2 weeks, depending on its application and

frequency of use.

D4 Maintenance ContractWithin one year after delivery date you can sign a maintenance contract for LICO® 100, extending the

guaranteed period to up to 5 years. The maintenance job is done twice a year, fixing an appointment at

your premises by telephone before.

The maintenance contract comprises:

1. General check-up and cleaning

2. Mechanical and electrical check-up

3. Function test and calibration

4. Exchange of defective parts

5. Technical advice

6. Delivery of an instrument on loan, if requested

7. Free software updates

Fig. 5: Test solution addista-color

Koehler Instrument Company, Inc. LICO® 100 K13250

24

D5 Lamp ExchangeThe lamp compartment is left of the display and covered with a slotted cap on the upper side of the

housing. If the lamp is broken, it should be exchanged by Koehler service staff, because we

recommend a general instrument check-up after an operating period of 1-2 years. If you insist on

doing it yourself, proceed as follows:

R e m o v e p o w e r s u p p l y u n i t o r b a t t e r y!

Remove the grey cover of the bit-box. Loosen the

fixture of the bit-box by pushing the locking clamp ��

and put it apart. Unplug contact �� from the lamp.

Loosen the metal clamps �� at the upper end one

be one from their stay and hinge them away. Take

the lamp �� off and put in a new one with the

marking pointing to the left.

Attention: Do not touch the glass bulb of the

lamp.

Hinge the metal clamps back to their places and fix

them to their stay. Connect the new lamp with the

plug-in contact �� and do the same steps as for

dismantling, but in reverse sense. Only when the

instrument is mounted completely, turn it on and test

it.

Fig. 6: Removing the fixture of the bit-box

Fig. 7: Lamp exchange

Koehler Instrument Company, Inc. LICO® 100 K13250

25

D6 Program-Module Card (BIT-BOX) of LICO® 100

The program module card (BIT-BOX) permits fast and comfortable updating of the internal software

and an easy extension of color systems. The slot for the program module card is at the rear side of

the housing. The instrument must be off when the bit box installed. Program updates or extensions are

made by inserting the bit box first and then turning the instrument on. Follow the display prompts. Turn

the instrument off after updating and remove the bit-box again. For color number extension, the bit-box

can be used but once.

Fig. 8: bit-box

Koehler Instrument Company, Inc. LICO® 100 K13250

26

D7 RS232C-Interface of LICO® 100

LICO® 100 has a serial interface at its rear side corresponding to RS 232 C and V24-standard.

D8 Technical Data

Viewing geometry: 0°/180° (transmission)

Display/operation: Backlit LC-display (128 x 64 Pixels)

Reference standard: Distilled water

Spectral range: 8 spectral wavelengths ranged from 400nm to 700nm

Wavelength accuracy: ± 2 nm

Diffuse light: < 1 %

Light source: Krypton lamp

Warming-up: approx. 2 minutes

Reproducibility: ± 2 Hazen at 50mm path width

Documentation: External printer

Control: Integated microprocessor

Measuring time: approx. 10 sec.

Program functions: Iodine number 0 ... 120

Hazen-(APHA) numberl 0 ... 1000

Gardner color scale 0 ... 18

ASTM D 1500 number 0 ... 8

Saybolt-number +30 ... -16

Cuvettes compartment: 50mm x 10mm square cuvette

11 mm round cuvette with automatic detection

Control elements: 8 Foil keys

Data interfaces: RS 232 C for LD 500-Printer

Mains connection: Power supply unit 100-240 V, 47/63 Hz

Dimensions : 260mm x 90mm x 200mm (WxHxD)

Weight: approx. 1350g

Subject to technical modifications.

Koehler Instrument Company, Inc. LICO® 100 K13250

27

E Annex

E1 Spares and Accessories

Identification: Quantity/Pcs.

Bit Box "Iodine number" 1

Bit Box "Hazen number" 1

Bit Box "Gardner number" 1

Bit Box "Saybolt number" 1

Bit Box "ASTM D 1500 number"

Square plastic cuvettes 10mm * 50mm 50

Square glass cuvette 10mm * 50mm 1

Round glass cuvettes 11mm 494

Cuvette holder for 16 round cuvettes 1

Printer paper 4 reels

Exchange lamp 1

Test solutions addista-color 1

Rechargeable battery 1.4Ah 1

Koehler Instrument Company, Inc. LICO® 100 K13250

28

E2 Regulations5 and Standards

EN 1557 Colorimetric characterization of optically clear colored liquids (products) as X, Y, Z -tristimulus values in transmission, 10.'92 (DIN 53995)

DIN 5033 Color measurement (ASTM E 179, ISO 7724)

DIN 6174 Colorimetric evaluation of color differences (ASTM E 308)

DIN 6162 Determination of iodine color number

ISO 6271 Clear liquids;Estimation of color by the platinum-cobalt-scale (Hazen color number,APHA color number, also ASTM D 1045-58, ASTM D 268-49, ASTM D 1209-62, BS2690:1956.)

ISO 4630 Estimation of color of clear liquids by the Gardner color scale, also ASTM D 1544-80

ASTM D1500 Standard Test Method for ASTM Color of Petroleum Products (ASTM Color Scale),also DIN/ISO 2049.

ASTM D156 Standard Test Method for Saybolt Color of Petroleum Products (DIN 51411).

ASTM D1925 Yellowness-Index

ASTM D 6045 Standard Test Method for Color of Petroleum Products by the Automatic TristimulusMethod