Freezing Point Osmometer User Manual V3700 K-7400S Semi-Micro Osmometer

Transcript of K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe...

Page 1: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Freezing Point OsmometerUser ManualV3700

K-7400S Semi-Micro Osmometer

Page 2: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

2

Note: For your own safety, read the manual and always observe the warnings and safety information on the device and in the manual

Manuel en français: Si jamais vous préfériez un manuel en français pour ce produit, veuillez vous contacter le support technique (Technical Support) par email ou par fax avec le no. de série. Merci beaucoup.

Technical Support: Phone: +49 30 809727-111 (9-17h, Central European Time)Fax: +49 30 8015010E-Mail: [email protected]: German, English

Publisher: KNAUER Wissenschaftliche Geräte GmbHHegauer Weg 38D-14163 BerlinPhone: +49 30 809727-0Fax: +49 30 8015010Internet: www.knauer.netE-Mail: [email protected]

Version information: Article number: V3700Version number: 1.0Release date: 2017-03-14Translation of the original edition

The information in this document is subject to change without prior notice. For latest version of the manual, check our website:http://www.knauer.net/knowledge/user-manuals.html

Copyright: © KNAUER Wissenschaftliche Geräte GmbH 2017All rights reserved.

Page 3: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

3 Table of Contents

Table of ContentsProduct Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Scope of Delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Safety for Users. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Signal Words . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Decontamination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

Symbols and Signs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Work Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Unpacking the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Connecting the Measuring Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

Stand-by Position of Measuring Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Measurement Principle and Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Switch-on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

Menu Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Preparing a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Measurement Preparation Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Pre-calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Measurement Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

Measuring Sample Solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Performing a Sample Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Important Operating Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

Functionality Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20System Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Self-Freezing of the Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20

Non-Freezing of the Sample . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Using the EuroOsmo® 7400 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Hardware Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Main Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

Run Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Sample List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Working Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

K-7400S user manual V3700

Page 4: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Table of Contents 4

Report & Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Search Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Maintenance and Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27Cleaning the Stirring Wire and the Thermistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Function Checks of Cooling Capacity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

Reference Thermometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Main Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Repeat Orders. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Legal Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Transport Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Warranty Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

KNAUER Software License Agreement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

K-7400S user manual V3700

Page 5: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

5 Product Information

Product InformationNote: Only use the device for applications that fall within the range of the intended use. Otherwise, the protective and safety equipment of the device could fail.

ViewsFront View

Keypad

Legend1 Measuring head

2 LCD display

3 Keypad

Fig. 1 Front view

Buttons Function

START/STOP Starts a measurement or calibration run. If this button has no specific function on the screen, it causes to return on the main screen.Note: With this button, every running measurement can be interrupted.

VIB Activates the stirring wire. During RUN and CALIBRATE, the button is without any function.

Arrows Cursor navigation buttons for parameter or value selec-tion

1

2

3

K-7400S user manual V3700

Page 6: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Scope of Delivery 6

Rear View

Scope of DeliveryPlease check whether the accessories are complete in accordance with the scope of delivery mentioned below. If a part is missing despite our compre-hensive exit controls, please contact our customer support.Note: Only use original parts and accessories made by KNAUER or a com-pany authorized by KNAUER. Freezing point osmometer K-7400S Power cable User manual (German/English) Installation Qualification document ("IQ", English) RS-232 Connection cable Measuring head Pack of 1000 sample vials 400 mOsmol/kg calibration solution (10 vials) 850 mOsmol/kg calibration solution (12 vials) Pack of cotton swabs

Safety for UsersProfessional Group The user manual addresses persons who are qualified as chemical laboratory

technicians or have completed comparable vocational training.The following knowledge is required: Knowledge regarding the health risks of chemicals Participation during an installation of a device or a training by the com-

pany KNAUER or an authorized company.Safety Equipment When working with the device, take measures according to lab regulations

and wear protective clothing: Safety glasses with side protection Protective gloves Lab coat

What must be taken into account?

All safety instructions in the user manual

Legend1 RS-232 connector

2 Measuring head connector

3 Ventilator

4 Power switch

5 Power connector

Fig. 2 Rear view

5

4

3

2

1

K-7400S user manual V3700

Page 7: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

7 Safety for Users

The environmental, installation, and connection specifications in the user manual

National and international regulations pertaining to laboratory work Original spare parts and tools made or recommended by KNAUER Good Laboratory Practice (GLP) Accident prevention regulations published by the accident insurance

companies for laboratory work Power cable: Defective power cables are not to be used to connect the

device and the power supply system. Power strip: If several devices are connected to one power strip, always

consider the maximum power consumption of each device. Power supply: Only connect devices to voltage sources, whose voltage

equals the device's voltage.Where is use of the device

prohibited?Never use the system in potentially explosive atmospheres without appro-priate protective equipment. For further information, contact the Technical Support of KNAUER.

Secure decommissioning At any time, take the device completely out of operation by either switching off the power switch or by pulling the power plug.

Opening the Device The device may be opened by the KNAUER Technical Support or any com-pany authorized by KNAUER only.

Signal WordsPossible dangers related to the device are divided into personal and material damage in this user manual.

Lethal injuries will occur.

Serious or moderate injuries can occur.

Minor injuries can occur.

Device defects can occur.

DecontaminationContamination of devices with toxic, infectious or radioactive substances poses a hazard for all persons during operation, repair, sale, and disposal of a device.

Life-threatening injuriesHealth danger if getting in contact with toxic, infectious or radio-active sub-stances. Before disposing of the device or sending it away for repair, you are

required to decontaminate the device in a technically correct manner.All contaminated devices must be properly decontaminated by a specialist company or the operating company before they can be recommissioned, repaired, sold, or disposed of. All materials or fluids used for decontamina-tion must be collected separately and disposed of properly.

Decontamination Report Devices without a completed Decontamination Report will not be repaired. If you would like to return a device to KNAUER, make sure to enclose a com-pleted Decontamination Report with the device: http://www.knauer.net/en/downloads/service.html

K-7400S user manual V3700

Page 8: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Symbols and Signs 8

Symbols and SignsThe following symbols and signs can be found on the device or in the user manual:

InstallationThis chapter describes all preparatory steps prior to start-up.

PreparationsNote: To ensure structural stability, the device is not intended to be stacked on or below other devices.

Work LocationThe intended use be ensured only if the requirements for ambient condi-tions of the operating environment are met. You will find the ambient con-ditions in the chapter "Technical Data" (see page 28).

General Purpose Position the device on a level surface. Protect the device against direct exposure to sunlight. Set up the device at a location not exposed to air drafts such as air condi-

tioning systems. Do not set up the device near to other machines that cause floor vibra-

tions. Keep the the devices away from high frequency sources. High frequen-

cies may compromise measurement values.Space requirements At least 5 cm, if there is another device on one side.

At least 10 cm, if there are devices set up on both sides. At least 15 cm to the cooler fan on the rear.

Ambient temperature The device should not be exposed to major temperature fluctuations during measurement.

Symbol Meaning

Warning Signs Electric shock hazard

Electrostatic discharge hazard, damages to system, device, or components can occur.

Biohazard

CE Mark A device or system marked with CE fulfills the product specific requirements of European directives. This is confirmed in a Declaration of Conformity.

K-7400S user manual V3700

Page 9: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

9 Installation

Device defectThe device overheats at exposure to sunlight and insufficient air circulation. Device failures are very likely. Set up the device in such a way that it is protected against exposure to

direct sunlight. Keep at least 15 cm clear at the rear and 5–10 cm at each side for air cir-

culation.

Power SupplyUse only the enclosed power cable to connect the device to the mains to make sure that the specifications stated in Technical Data are met. But check beforehand to use power cables admitted for use in your country. Replace defective power cables only with accessories from KNAUER. Detachable power cables must not be replaced with different cable types.The maximum power input is 70 watts.

Electronic defectElectronic hazard when using an identically constructed power adapter from another manufacturer. Contact the Technical Support of KNAUER. Do not connect the device to the power connection.Note: The nominal capacity of the connected devices must be maximum 50 % of the power supply to account for larger inrush currents when switching on the modules.

Prerequisites The electrical power supply at the installation site must be connected directly to the nearest main power line.

The power must be free from ripple, residual current, voltage peaks and electromagnetic interference.

The connectors for the mains voltage are grounded accordingly. The device receives sufficient power with reserve capacity.

Power plug The device is intended for use with AC power networks of 100 – 240 V. Make sure that the power plug on the rear of the device is always accessi-

ble, so that the device can be disconnected from the power supply. To disconnect the device from the main power, disconnect the power

plug. The osmometer needs a three-core power plug.

Electric shockMake sure that the power cable is not exposed to high stresses. The power cable should not be bent or clamped. Make sure that the power cable does not come in contact with liquids

and heat sources.

Unpacking the DeviceAt the factory, all KNAUER devices are carefully packed for safe transport. After unpacking, check all device parts and accessories for signs of damage during transport and if necessary assert claims for damages immediately to the transport company.The complete listing of consumable and spare parts can be found in section "Accessories and spare parts" on page 29.

Prerequisite Check packaging for damage caused during transportation. If necessary, put forward any claim for damages to the carrier.

Tools Utility knife

K-7400S user manual V3700

Page 10: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Installation 10

Bruising dangerDamage to the device by carrying or lifting it on protruding housing parts. The device may fall and thus cause injuries. Lift the device only centrally on the side of the housing.

Process 1. Set up the package in such a way that you can read the label. Using the utility knife, cut the adhesive tape and open the packaging.

2. Remove the foam insert. Take out the accessory kit and the manual. 3. Open the accessory kit and check the scope of delivery. In case any parts

are missing, contact the Technical Support. 4. Clasp the device from below, lift it out of the packaging and place it on

its feet.5. Check the device for signs of damage that occurred during transport. In

case you notice any damage, contact the Technical Support.6. Place the device in its site of operation and remove protective foil.

Next steps Store packaging and keep the included packing list for repeat orders.

Start-upNote: The device is not suitable for constant operation because it would reduce the lifespan of the Peltier elements.

Risk of InfectionRisk of infection by injuries caused by a contaminated stirring wire. Wear safety gloves when handling the measuring head.

Skin InjuryThe cooling compartment reaches temperatures up to -30 °C during the running measurement. Contact with unprotected hands may lead to skin injuries. Avoid direct contact with the cooling compartment. For cleaning or drying, use cotton swabs or a lint-free tissue.

For information and safety instructions for power supply, see page 9.

Electric shockHigh voltages build up inside the device, the housing serves as a protective cover. Never put the device into operation when the housing is open. First close the housing, then plug in the power plug and switch the

device on.

Electronic defectDamage to the electronics if the device is turned on while connecting or interrupting the power connection. Switch off the device beforehand.

Prerequisites The device is switched off. The power cable is detached from the power supply.

Procedure 1. Attach the power cable to the power connector on the rear of the device. 2. Switch the device on by using the "ON/OFF" switch.

Next Steps Bring the device into operation. To put the device out of operation, switch off the device and detach the power plug from the power supply.

K-7400S user manual V3700

Page 11: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

11 Operation

OperationConnecting the Measuring HeadParts of the measuring head are an inbuilt motor with a stirring wire and the thermistor as temperature sensor. The measuring head cable is to be plugged into the connector on the back of the device.The sample vial, filled with the exact volume of calibration or sample solu-tion (50 – 150 μl) must be mounted onto the adapter.

Note: Any change of the measuring head causes the need for a precalibra-tion of the device to adapt the device to the measuring head (see “Pre-cali-bration” on page 16).

Legende1 Stirring wire2 Thermistor3 Foot

Fig. 3 Measuring head

Fig. 4 Sample vial

1 2 3

K-7400S user manual V3700

Page 12: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Operation 12

The measuring head is placed on the housing plate with the sample vial ris-ing into the cooling cavity. The device is now ready for measurement.Note: The sample volume can be reduced down to a volume of 50 μl. It is important that all calibrations must be performed with exactly the same vol-ume as the later sample.

Stand-by Position of Measuring HeadNote: Generally the measuring head should be stored with attached sample vial. While working with the osmometer, use the four columns on top of the device for temporarily set down the measuring head without sample vial. In the measurement position, it is not necessary that the feet of the measuring head rest on the device.To avoid sample spilling into the measuring head, it must be held upright.

Measurement Principle and ProcessThe osmotic pressure in a solution is proportional to the freezing point depression of that solution. An aqueous solution, with an osmotic pressure corresponding to an ideal 1-molal solution, freezes at -1.858 °C. An aqueous solution with this freezing point has a concentration of 1 Osmol/kg. To perform a measurement, the sample solution is cooled down first with-out stirring. Even pure water can be super-cooled down to about -5 through -8 °C without freezing. The freezing is initiated by an automatically start of the stirring wire at a set temperature. The temperature reaches the freezing point, 0 °C in case of pure water.

Fig. 5 Set-down of the measuring head without sample vial

K-7400S user manual V3700

Page 13: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

13 Operation

The freezing point of a solution is below 0 °C. The freezing point depression ΔT is a measure for the osmolality of a solution. It can be read directly from the display given in mOsmol/kg.

During a measurement, the actual temperature is displayed. The device detects the maximum temperature reached, stores it, and if crystallization occurred satisfactory, displays the result. After finishing the measurement, the sample vial is warmed up to 20°C (stand-by temperature of the device). The measuring head can remain there before and after measurement as long as desired.

Switch-onWhile switching on, the device performs an automatic self-test. During this time, the display presents the start screen with the current firmware version for a short time.

Fig. 6 Temperature-time-curve for pure water

Fig. 7 Temperature-time-curve for a sample solution

KNAUER V01.XX Freezing Point Osmometer

K-7400S user manual V3700

Page 14: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Operation 14

After completion of the self-test, the display shows the measurement screen (see Fig. 8 screen 1). The cursor is placed on the rhombus field. As shown in Fig. 8, it will take you to the relevant menu points.

Menu GuideStarting with the cursor on the rhombus field, further menus become acces-sible in an endless loop by pushing the vertical arrow buttons . By pushing the horizontal arrow buttons , the cursor moves from field to field. Only those fields are accessible which are allowed to change the set-ting. Other fields, like the actual measured temperature, will be skipped. On any changeable field the possible settings can be scrolled using the ver-tical arrow buttons . The displayed values will be accepted and stored when leaving the field with one of the horizontal arrow buttons . In this way, you can e. g. choose between printer and PC for data output (screen 2) or display the measurement data in mOsmol or °C (screen 1).

Fig. 8 Menu path

Screen Explanation

1 Main menu

2 Data Output Controls the connection of the Osmometer with a printer or PC.

3 TEMP Displays the current temperatures of the sample and the cooling cavity. Additionally, units can be changed from [°C] to [°F] with the buttons. If the sample temperature is displayed with "----", the measuring head is not connected.

4 INFO Shows the serial number and the current firmware version.

5 DATE Setting for the date and time. The setting remains saved when switching off the device.

K-7400S user manual V3700

Page 15: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

15 Operation

Note: Calibration and measurement must be performed at the same set-tings (freeze & cooling limit). Additionally, an identical sample volume must be used for measuring and calibration to ensure exact and reproducible data. As a result, every change of the temperature settings or the sample volume demands a new calibration.

Preparing a MeasurementNote: The device should be switched on at least 5 minutes before use.To check the stirring wire, press the <VIB> button.

Stab InjuryRisk of injury from vibration wire during function test performance. Perform the function test with attached measuring vial.Note: If you want to print out the measurement results, you may connect a printer via RS-232 connector. The result of each measurement will be printed automatically in the follow-ing format:

After a calibration run, you get the following print:

If a measurement is cancelled due to an error, the screen displays the corre-sponding error message. If a printer is connected, it prints the error mes-sage, without the specification about the type of error.

Measurement Preparation ProcessPrerequisites The device is switched on and calibrated.

Process 1. Pipet 50 – 150 μl sample or calibration solution into a clean, dry sample vial.

6 CALIBRATE Displays the calibration points which are currently in use. There are 3 possible calibration points. The first field is fixed at "0000". At least 2 calibration points are necessary for regular measurements. From this screen, the calibrations will be started (see "Performing the Calibration" on page 17).

7 COOL Sets following parameters, in 0,5°C steps: Initiation point "freeze" within the range from

-13,0 to -4,0°C. Minimum temperature "cooling limit" within

the range from -18.0 to -12.0 °C.

8 CALIBRATE-T Enables the pre-calibration to adapt the osmome-ter to the measuring head, see "Pre-calibration" on page 16.

Screen Explanation

Measured 0400 mOsmol 23.03.15 08:24:35

Calibrated 1234 mOsmol 23.03.15 07:36:17

(NO) FREEZE 23.03.15 09:06:04

K-7400S user manual V3700

Page 16: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Operation 16

Note: The sample volume can be freely chosen in the mentioned range. It is important that you use exactly the same amount for calibration and mea-surement.Practical Tip: Please consider, the higher the volume, the better the preci-sion of the measurement.2. Put the filled sample vial on the thermistor, so that the thermistor and

the stirring wire dip in the solution. The meniscus of the liquid must be aligned horizontally.

3. Place the measuring head on top of the device, so that the sample vial is inserted into the cooling cavity. The feet of the measuring head do not have to rest on the device.

Next Steps The device is now ready for measurement.

CalibrationPre-calibrationA pre-calibration is necessary when: the device is taken into operation the first time. the measuring head has been replaced.

Prerequisites Prepare a measurement with water (HPLC quality).Process 1. Select CALIBRATE-T in the osmometer menu (screen 8 in Fig. 8).

2. Push the <START> button. While the pre-calibration is running, the screen displays the sample temperature and the elapsed time. The run can be interrupted at any time by pressing the <STOP> button.

3. After finishing the calibration, the screen displays the message "success-ful". The previous calibration values will be deleted.

Device DefectIf the sample vial is removed from the measuring head while the sample or the calibration solution is frozen, the thermistor or the stirring wire can be damaged. When the message "next sample" appears on the display, the sample vial

is safe to remove.Note: Any performed pre-calibration deletes the existing calibration values.4. If you get the message "not successful" on the screen at the end of the

pre-calibration, the sample is not frozen. Make sure if the sample is 0 mOsmol/kg a valid measuring head is proper connected to the device.5. Repeat the pre-calibration if necessary.

Measurement CalibrationNormally a Two-Point Calibration is sufficient.Note: Calibrations must be performed very carefully. A confusion of the cal-ibration solutions will affect all later measurements and leads to incorrect results.Up to 3 calibration points are possible. The desired calibration concentration can be selected on any of the fields. After that, the calibration run is started by pressing the <START> button.The first field is fixed on "0000". On the other fields the choices are ----, %%%%, 50, 100, 150, up to 1900, 1950, 2000 in steps of 50 mOsmol/kg.Selecting "----", the corresponding calibration point will be deactivated.

thermistor > +9.99°C< 00:56

K-7400S user manual V3700

Page 17: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

17 Operation

The selection of "%%%%" deactivates all other calibration points (except "0000"). Simultaneously the displayed measurement results will be shown in % of this calibration point instead of "mOsmol/kg". The measurement results are now displayed in % of this calibration value.Note: At least two calibration points must be active, the "0000" and one of the others.The calibration point remains active until a new valid value for the selected field is created on the CALIBRATE screen. If a calibration run is not success-fully finished the last calibration value becomes reactivated.At least 2 calibration points are necessary for regular measurements. To determine the measurement results, it will be interpolated between the clos-est calibration points, or extrapolated from there.Note: It is recommended to choose the calibration solutions so that the expected measurement values are situated between the calibration points.

Performing the CalibrationPrerequisites Prepare a measurement with water (HPLC quality, see "Measurement Prepa-

ration Process" on page 15).Process 1. Activate the CALIBRATE screen and set the first calibration field to "0000".

2. Push the <START> button. The run can be interrupted at any time by pressing the <STOP> button. While the calibration is running, the screen displays the sample temperature and the elapsed time.

3. After the measurement is successfully completed, the measurement value is displayed.

Push the <START> button for confirmation. An earlier measured calibra-tion will be overwritten. By pushing any other button, the new value will be abandoned and the CALIBRATE menu will be shown again.

4. Prepare a measurement with a calibration solution (see "Measurement Preparation Process" on page 15). Enclosed to the osmometer are fitting solutions for calibrations with an osmolality of 400 or 850 mOsmol/kg.

5. Activate the second calibration field on the CALIBRATE screen and set the osmolality for the calibration solution.

6. Repeat step 2 to 3.For special applications, especially in the pharmaceutical industry, a Three-Point Calibration is required. If this is the case, proceed according to the process above and repeat the process for the calibration field 3.

Re-calibrationA re-calibration is necessary after replacing the measuring head changing the temperature settings (see fig. 8 screen 7) changing the sample volume

0000 mOsmol > +5.43 °C1:23

0000 mOsmol > -0.00 °C o.k.? <start>

K-7400S user manual V3700

Page 18: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Operation 18

Measuring Sample SolutionsPerforming a Sample Measurement

Prerequisites Prepare a measurement with a sample solution (see "Measurement Prepara-tion Process" on page 15).

Process 1. To perform measurements, choose screen 1 (Fig. 8) with the but-tons and push the <START> button. While the measurement is running, the screen displays the sample temperature and the elapsed time.

The run can be interrupted any time by pressing the <STOP> button.2. The device starts the crystallisation automatically and detects the freezing

point depression. If the crystallisation occurred properly, it displays the corresponding osmolality on the display:

The printer output is given in the following format:

3. Before the next measurement, wait for the sample to unfreeze. This is shown by the message "next sample" on the display.

If no crystallisation occurs after the automatical start of the stirring wire, the measurement will be stopped with the following error message:

If the crystallisation occurs before the start of the stirring wire, the measure-ment will be stopped with the following error message:

Device DefectIf the sample vial is removed from the measuring head while the sample or the calibration solution is frozen, the thermistor or the stirring wire can be damaged. When the message "next sample" appears on the display, the sample vial

is safe to remove.Note: The sample melting can be quickened by warming up with the fin-gers. When"next sample" appears below the measurement value, the sam-ple is unfrozen, and it is save to remove the vial.

Important Operating Notes1. The device is only suitable for aqueous solutions.2. Only use sample vials which were obtained from KNAUER or recom-

mended by KNAUER. Sample vials should not be reused.3. To ensure a fixed sample volume, always introduce the solutions with a

clean and dry pipette tip into the sample vial. Make sure that there are no air bubbles in the liquid or drops on the vial wall above the liquid. The sample solution should stand upright in the sample vial.

RUN > -1.23°C 1:23

>>> 1234 mOsmol <<< <next sample>

Measured 1234 mOsmol 19.03.15 16:05:35

> ERROR < NO FREEZE

> ERROR < FREEZE

K-7400S user manual V3700

Page 19: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

19 Functionality Tests

4. Handle the thermistor always very carefully. Do not use abrasive materials for cleaning. The thermistor should be cleaned occasionally with a soft tensidic cleaning agent (dilution: 1:10 with water (HPLC quality) about 40±5 °C). After immersing, activate the stirring function 3 – 5 times for 5 seconds and rinse with water (HPLC quality). The cleaning frequency depends on the used sample matrix. It is sufficient for pure saline solu-tions to rinse the thermistor with the following sample solution. A dirty thermistor does not give accurate measurement results.

5. Cleaning and drying the thermistor with alcohol or acetone is not recom-mended. High-frequency sources may compromise measurement values. Remaining solvent vapours could cause false results.

6. Solutions containing proteins, such as sera, can only be measured once. Freezing causes denaturation of the protein. A repetition of the measure-ment with the same sample would result in increased values.

7. Please note the difference between ideal and real osmolality. Only real osmolality can be measured. It is not possible to prepare other calibration solutions by dilution of a calibration solution since the activity coefficient of the solution changes with dilution.

8. Water which is stored in polyethylene bottles over a longer period could be difficult to be super-cooled. If such difficulties may occur, we recom-mend a storage in glass bottles.

9. Make sure that the sample material is homogeneous. If the measurement result has to be reproduced with the same sample, it has to be mixed with the stirring wire before each further measurement. As ice does not contain salines and floats on top while unfreezing, it results in a concen-tration gradient which has to be removed before measurement.

10. Evaporation can increase the concentration of dissolved particles which leads to increase of osmolality of a solution. Calibration solutions should be used not longer than 30 - 60 min after the vials were opened. We rec-ommend using KNAUER calibration solutions which are mentioned in chapter "Reorders" (see page 29).

11. After finishing all measurements, the osmometer should be switched off, because a permanent operation reduces the lifespan of the device. If the device is switched on but unused for a longer period, it may occur that some water from room air is condensed in the cooling cavity. It should be removed carefully with a dry, lint-free tissue prior to the next mea-surement.

12. The stirring wire triggers the reproducible freezing and is set optimal in the factory. Make sure that you do not bend the wire while attaching the sample vials and cleaning.

Functionality TestsInstallation Qualification

(IQ)

The customer may request the Installation Qualification, which is free of charge. In case of a request, the Technical Support of KNAUER or from a provider authorized by KNAUER performs this functionality test during the installation. The Installation Qualification is a standardized document that comes as part of the delivery and includes the following: confirmation of flawless condition at delivery check if the delivery is complete certification on the functionality of the device

Operation Qualification (OQ)

The Operation Qualification includes an extensive functionality test accord-ing to KNAUER standard OQ documents. The Operation Qualification is a standardized document and free of charge. It is not part of the delivery, please contact the Technical Support in case of request.

K-7400S user manual V3700

Page 20: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Troubleshooting 20

The Operation Qualification includes the following: definition of customer requirements and acceptance terms documentation on device specifications device functionality check at installation site

Test Intervals To make sure that the device operates within the specified range, you should test the device regularly. The test intervals are dependent on the usage of the device.

Execution The test can be carried out either by the Technical Support of KNAUER or from a provider authorized by KNAUER (for a fee).

TroubleshootingSystem MessagesAny error message will be deleted simply by pressing any button. The main screen will be displayed.In any case the first row shows > ERROR < which is specified in the second row.Following messages can appear:

Self-Freezing of the SampleIf the sample does not super-cool but freezes spontaneously without initia-tion, there are three possible causes:If the following solutions do not solve the problem, the super-cooling tem-peratures are too low. The freezing temperatures in the device must be set higher. The cooling temperature can be changed in screen 7 COOL (Fig. 8 "Menu path" on page 14).

Error message Cause Remedy

FREEZE The crystallization took place or an error occurred before the active triggering of the freezing process.

see “Self-Freezing of the Sample” on page 20

NO FREEZE No sample crystallization took place after the activation of the stirring wire.

see “Non-Freezing of the Sample” on page 21

OUT OF RANGE The measured value is outside the acceptable range.

Usually a malfunction of the thermistor. Contact the Technical Support.

NOT CALIBRATED No calibration points are available. See section "Measurement Calibra-tion" on page 16

COOLER TIMEOUT Timeout due to too slow cooling: >5 minutes from START

Contact the Technical Support.

MEASUREMENT TIME-OUT

Timeout due to too long measure-ment: >5 minutes from START

Contact the Technical Support.

Cause Remedy

Too little sample solution is in the sample vial. A very cold zone, which causes self-freezing of the sample, develops above the solution on the sample vial wall.

Use exact the required sample vol-ume: 50 – 150 μl

K-7400S user manual V3700

Page 21: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

21 Using the EuroOsmo® 7400 Software

Note: Calibration and measurement have to be performed with the same temperature settings and equal volume.

Non-Freezing of the SampleIf the sample does not freeze, one of the following causes may be responsi-ble:

If these solutions are not effective, the freezing temperature is too high and has to be lowered. The freezing temperature can be changed in screen 7 COOL (Fig. 8 "Menu path" on page 14).Note: Calibration and measurement have do be performed with the same temperature settings and equal volume.

Using the EuroOsmo® 7400 SoftwareThis chapter will explain the basic knowledge needed for working with K-7400S and the EuroOsmo® 7400 software.Note: The License Agreement should be read before using the EuroOsmo®7400 software (see page 30). Using the software, the user indi-cates that he has read the License Agreement and agrees with the regula-tions.

InstallationThe hardware requirements (processor, working memory) are very low for the installation of EuroOsmo® 7400.The program runs under Windows 7, 8 and 10, but also with older operat-ing systems, back to Windows XP.

Process 1. Insert the EuroOsmo CD into the CD ROM drive.2. Wait for 30 seconds. The set-up starts automatically. You will be guided

through the installation via self-explanatory windows.3. If your CD ROM drive does not have AutoPlay: Double-click on <My Computer> Right-mouse click on your CD ROM drive. Select <AutoPlay> from the right-click menu.

The sample vial or the thermistor is scratched or contaminated. Both could act as crystallization nuclei causing premature freezing.

Use a new sample vial and care-fully clean the thermistor.

Only use sample vials which were obtained from KNAUER or recommended by KNAUER.

Solutions saturated with air or gases are sometimes difficult to super-cool.

Degas the sample solution by ultrasound.

Cause Remedy

The cooling rate is not sufficient. Ensure that the air inlets and out-lets at the device are free.

Too much sample solution is in the sample vial. This leads to an uncooled zone in the upper part of the solution. The stirring causes a mixing of warm and cold sample layer.

Use exact the required sample vol-ume: 50 – 150 μl

Cause Remedy

K-7400S user manual V3700

Page 22: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Using the EuroOsmo® 7400 Software 22

4. If there is no AutoPlay menu, select Explore from the right-click menu, go to EuroOsmo® 7400/Disk1 and double-click on <Setup.exe>.

Hardware InstallationThe device has to be connected with a defined port of the computer (e.g. COM 1) prior to starting the program.

Main ScreenAfter starting the program you will see the main screen of the software.

If the osmometer is not connected or switched off, you will get the follow-ing message.

Fig. 10 Error message for an unconnected osmometer

Icons

Fig. 9 Main screen of EuroOsmo® 7400 Software

Command Explanation

Prints the result list.

Displays a preview of the actual report.

Starts the search function.

K-7400S user manual V3700

Page 23: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

23 Using the EuroOsmo® 7400 Software

Run Settings

If the sample has already been measured, the following warning will be dis-played.

Fig. 11 Warning prior to overwriting existing results

Sample ListThis list is used to name the samples and codes. The measurement results will be filled automatically and can be saved. Using the scroll bar at the bot-tom, a <Comments> column becomes accessible for detailed information.The samples will be measured in the order of the list and the results are filled in and saved.

Fig. 12 New sample list

Pastes the data of a selected field into following lines of the chosen column.

After each bar code scan, the cursor switches to the next sample line.

Command Explanation

Select the number of the sample to be measured from the sample list.

Start button for any measurement or calibration run. While starting the run it changes to a stop button to enable a manually aborting of the run.

Command Explanation

K-7400S user manual V3700

Page 24: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Using the EuroOsmo® 7400 Software 24

Double clicking on <No.> of an already measured sample displays the corresponding temperature sequence. The path of the already stored sample list is shown on top of the graph.

Fig. 13 Measurement window with the temperature sequence of a run

Fig. 14 Setup area of the main window

SetupArea for editing general settings for measurement sequences and corre-sponding reports.

Command Explanation

Operator Enter the name of the user.

Communication Port Select the COM port to which your osmometer is connected.

K-7400S user manual V3700

Page 25: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

25 Using the EuroOsmo® 7400 Software

Working Parameters

Fig. 15 Error message in case of not valid parameters

Fig. 16 Report design area

Command Explanation

Monitoring view The entered value defines the scale of the monitor by its final value in minutes. The temperature pat-tern is shown continuously for the defined period.

Measurement view The entered value defines the scale of the measure-ment by its final value in minutes.

Freezing Level The initiation point FREEZE can be set within the range from -13.0 to -4.0 °C. Any decimal value will be rounded in steps of 0.5°C.

Cooling Limit The minimum temperature can be set within the range from -18.0 to -12.0 °C. Any decimal valuewill be rounded in steps of 0.5°C.

Units Select the units valid for the measurement sequence. The choice is mOsmol, °C, °F, and %. If °F is selected, the values of the freezing level and cooling limit are automatically changed to °F.

By clicking on the <Apply> button, the entered val-ues are sent to the osmometer. If any value is out of the allowed range a corresponding message will be displayed. The setting of the osmometer remains unchanged.

K-7400S user manual V3700

Page 26: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Using the EuroOsmo® 7400 Software 26

Report & Graph

A click on the <print> button opens a printer setup window prior to printing the report. A preview of the actual report will be displayed when clicking on the <preview> button.

Search Function

Fig. 17 Search function

Clicking on the <...> button, an explorer window appears to enter the search path. Additional you have the option to search also in subfolders. Search subjects are the operator, the sample name, and/or the sample code. As wildcard characters can be used ? for single characters and * for unde-fined areas before and/or after to the defined search characters.

Command Explanation

Report Settings Report design area.

Logo Bitmap Clicking on the <...> button, an explorer window appears to select the logo.

Change Caption Clicking on the <...> button, an explorer window appears to edit the caption. Clicking on the <Font...> button to format the caption.

Change SubTitle Clicking on the <...> button, an explorer window appears to edit the subtitles. Additionally, you can format the subtitles here.

Graph Settings Area for settings of the graph design. A click on one of the four buttons will open the colour selection window where the colours of the curve, axes, grid lines, and the back ground can be defined separately.

GridOn This option field is to show or to hide the grid lines of the graphs.

K-7400S user manual V3700

Page 27: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

27 Maintenance and Care

If necessary the search results can be reduced by advanced options. These are the measurement values and/or the date of the measurement.

Fig. 18 Search function window, advanced options

Maintenance and CareCleaning the Stirring Wire and the Thermis-torNote: Stirring wire and thermistor should not be touched.

Risk of InfectionRisk of infection by injuries caused by a contaminated stirring wire. Wear safety gloves when handling the measuring head.

Stab InjuryRisk of injury from stirring wire when detaching it by hand. Detach the stirring wire only with attached measuring vial.The measuring head parts which are wetted from the sample can be cleaned easily by the user by attaching a sample vial with a suitable cleaning solution to the measuring head (see "Important Operating Notes" on page 18).Note: The supply and the appropriate use of these cleaning solutions is in the responsibility of the user.

Function Checks of Cooling CapacityThe cooling capacity can be checked by measuring the temperature in the cooling cavity (-13 to -16 °C during cooling down and ~ 20 °C in standby position).

Prerequisites The device must be calibrated. Process 1. Connect the cable of the measuring head with the device and place the

head on top of it (see page 12). 2. Press the <START/STOP> button to start the measurement and wait 5

minutes to get the error message <MEASUREMENT TIMEOUT> on the display.

3. Press quickly twice the button to read out the cooling cavity value on the COOL menu (see page 14).

4. The value must be in the range of -13 to -16 °C.5. If the value is not reached, please check that the cooling limit is set to

-16 °C.

K-7400S user manual V3700

Page 28: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Technical Data 28

6. Retry the test and press quickly twice the button after 5 minutes with-out displaying the error message.

7. If you change the cooling limit, you must perform a new calibration after this test.

Reference ThermometerAdditionally, you can check the device with a calibrated reference thermom-eter. This thermometer should display temprature changes quickly.

Process 1. Fill the cooling cavity close to the top with water (HPLC quality) and insert the thermometer.

2. Perform the test as described above and observe the temperature pro-gression.

3. After the test, it is necessary to dry completely the cooling cavity with a tissue.

The cooling capacity is usually sufficient for a sample volume of 150 μl in the sample vial if the freezing turbulence in the solution is initialized by a start of the stirring wire within 150 seconds.If the temperature of -13 °C cannot be obtained, the device must be returned for repair.The thermistor should always be placed in the sample centre 0.5 to 1 mm above the bottom of the sample vial.Note: The shape of the stirring wire is essential for a precise and reproduc-ible measurement. Therefore, take care to not bend it. The lower end of the wire should contact the vial wall.

Technical DataMain Features

Communication

General

Sample volume 50 – 150 μl

Solvent Water (HPLC quality)

Measurement time ca. 2 min

Measurement range 0 – 2000 mOsmol/kg

Precision SD ≤ 4 mOsmol/kg [0 – 400 mOsmol/kg]

RSD ≤ 1.0 % [400 – 2000 mOsmol/kg]

Linearity ± 1 % [0 – 1500 mOsmol/kg]

± 1.5 % [0 – 2000 mOsmol/kg]

Calibration Two-Point Calibration

(0 and one free selectable osmolality)

optional:

Three-Point Calibration

(0 and 2 free selectable osmolalities)

Measuring procedure Completely automatic after sample insertion

Port RS-232

Control Keypad (LED screen, 2 rows with 24 caracters), software

Power supply 100 – 240 V, 50 – 60 Hz, 70 W

Dimensions 160 x 182 x 340 mm

K-7400S user manual V3700

Page 29: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

29 Repeat Orders

Repeat OrdersThis list for repeat orders is valid for the time the document has been pub-lished. Deviations afterwards are possible.Note: For repeat orders of spare parts use the enclosed packing list. Contact the Technical Support in case there are any questions on spare parts or accessories.

Further Information Further information on spare parts and accessories can be found online: www.knauer.net

Legal InformationTransport DamageThe packaging of our devices provides the best possible protection against transport damage. Check the devices for signs of transport damage. In case you notice damages, contact the Technical Support and the forwarder com-pany within three workdays.

Warranty ConditionsThe factory warranty for the device is stipulated by contract. During the warranty period, any components with material or design-related defects will be replaced or repaired by the manufacturer free of charge. Please con-nect to our website for further information on terms and conditions.All warranty claims shall expire in the event that any unauthorized changes are made to the device. This warranty also excludes the following:

Weight 4.5 kg (device + measuring head)

Ambient conditions Temperature range: 10 – 35 °C

Rel. humidity: 20 – 80 % (non-condensing)

Name Order num-ber

Freezing point osmometer K-7400S A0006AC

Power cable 230V M1479

RS-232 Connection cable M1702

Plain printer A3711

User manual V3700

Measuring head for plastic sample vials A0840-2

Pack of 500 plastic sample vials A0272

Pack of 1000 plastic sample vials A0720

300 mOsmol/kg calibration solution (12 vials) A01240

400 mOsmol/kg calibration solution (10 vials) A01241

850 mOsmol/kg calibration solution (12 vials) A01250

2000 mOsmol/kg calibration solution (12 vials) A01248

Pack of cotton swabs A0469

Paper roll for plain printer V7013

K-7400S user manual V3700

Page 30: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Legal Information 30

accidental or willful damage damage or errors caused by third parties that are not contractually

related to the manufacturer at the time the damage occurs wear parts, fuses, glass parts, columns, light sources, cuvettes and other

optical components damage caused by negligence or improper operation of the device and

damage caused by clogged capillary packaging and transport damageIn the event of device malfunctions, directly contact the manufacturer.KNAUER Wissenschaftliche Geräte GmbH Hegauer Weg 38 14163 Berlin, Germany Phone: +49 30 809727-111 Telefax: +49 30 8015010 e-mail: [email protected] Internet: www.knauer.net

Declaration of ConformityThe Declaration of Conformity is part of the delivery and accompanies the product as a separate document.

DisposalHand in old devices or disassembled old components at a certified waste facility, where they will be disposed of properly.

AVV Marking in Germany According to the German "Abfallverzeichnisverordnung" (AVV) (January, 2001), old devices manufactured by KNAUER are marked as waste electrical and electronic equipment: 160214.

WEEE Registration KNAUER as a company is registered by the WEEE number DE 34642789 in the German "Elektroaltgeräteregister" (EAR). The number belongs to cate-gory 8 and 9, which, among others, comprise laboratory equipment.All distributors and importers are responsible for the disposal of old devices, as defined by the WEEE directive. End-users can send their old devices man-ufactured by KNAUER back to the distributor, the importer, or the company free of charge, but would be charged for the disposal.

KNAUER Software License AgreementbetweenKNAUER Wissenschaftliche Geräte GmbH, Hegauer Weg 38, D-14163 Berlin(hereinafter Licensor or KNAUER)and the software user (hereinafter Licensee)for the software packageEuroOsmo®7400 V1.5 and handbooks.In return for acquiring a license to use the software ("EuroOsmo®7400 V1.5") and related documentation, Licensee agrees to the following terms and conditions:1. License. This agreement grants the Licensee a license to(a)use the EuroOsmo®7400 V1.5 on a single computer system or, in the case of a multi-user or networked system which permits access to the EuroOsmo®7400 V1.5 by more than one user at the same time, at one (1) single working location, and

K-7400S user manual V3700

Page 31: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

31 Legal Information

(b)make one copy of the EuroOsmo®2000 V1.2 in machine readable form solely for back-up purposes provided your reproduce KNAUER´s copyright proprietary legends and mark the back-up copy with the express wording "back-up copy Knauer GmbH, Hegauer Weg 38, D-14163 Berlin".2. Restrictions. Licensee may not distribute copies of the EuroOsmo®7400 V1.5 to others or electronically transfer the EuroOsmo®7400 V1.5 from one computer to another over a network. Licensee may not use the EuroOsmo®7400 V1.5 from multiple locations of a multi-user or networked system at any time. The EuroOsmo®7400 V1.5 contains trade secrets and in order to protect them the Licensee may not decompile, reverse engineer, disassemble, or otherwise reduce the Software to a human-perceivable form: LICENSEE MAY NOT MODIFY, ADAPT, TRANSLATE, RENT, LEASE, LOAN, RESELL FOR PROFIT, DISTRIBUTE, NETWORK OR ANY DERIVATIVE WORKS BASED UPON THE SOFTWARE OR ANY PART THEREOF. The Licensee must not remove from the program firm names, trademarks, logos, copyright notices and other markings or devices, which serve the protection of the program.3. Ownership of EuroOsmo®7400 V1.5.Licensee owns the media upon which the EuroOsmo®7400 V1.5 is recorded or fixed, but KNAUER, Licensor, retains title and ownership of the EuroOsmo®7400 V1.5 recorded on the original media and all subsequent copies of the EuroOsmo®7400 V1.5, regardless of the form or media in which or on which the original and other copies may exist. This license is not a sale of the EuroOsmo®7400 V1.5 or any copy.4. Confidentiality. Licensee agrees to maintain the EuroOsmo®7400 V1.5 in confidence and to not disclose the EuroOsmo®7400 V1.5 to any third party without the express written consent of KNAUER. Licensee further agrees to take all rea-sonable precautions to preclude access of unauthorized persons to the EuroOsmo®7400 V1.5.The Licensee must not pass copies of the program or the handbook to third persons.5. Term. This license is effective until January 1, 2030, unless terminated earlier. Licensee may terminate the license at any time by destroying the Software (including the related documentation) together with all copies or modifica-tions in any form. KNAUER will have the right to terminate the Licensee's license immediately if the Licensee fails to comply with all copies or modifi-cations in any form. The termination does not raise Licensee's claim for refunding of the license fee. 6. Limited Warranty. KNAUER warrants only that the media upon which the EuroOsmo®7400 V1.5 is furnished will be free from defects in material or workmanship under normal use at the time of delivery. KNAUER DOES NOT AND CANNOT WARRANT THE PERFORMANCE OR RESULTS LICENSEE MAY OBTAIN BY USING THE SOFTWARE OR DOCUMENTATION. THE FOREGOING STATES THE SOLE AND EXCLUSIVE REMEDIES KNAUER WILL PROVIDE FOR BREACH OF WARRANTY. EXCEPT FOR THE FOREGOING LIMITED WARRANTY, KNAUER MAKES NO WARRANTIES; EXPRESSED OR IMPLIED, AS TO NON-INFRINGEMENT OF THIRD PARTY RIGHTS, MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. Some countries do not allow the exclusion of implied warranties or limitations on how long an implied warranty may last, so the above limitations many not apply to the Licensee. This warranty gives the Licensee specific legal rights and the Licensee may also have other rights which vary from country to country.7. Limitation of Liability.

K-7400S user manual V3700

Page 32: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

Legal Information 32

IN NO EVENT WILL KNAUER BE LIABLE TO LICENSEE FOR ANY SPECIAL DAMAGES, INCLUDING ANY LOST PROFITS, LOST SAVINGS, OR OTHER INCIDENTAL OR CONSEQUENTIAL DAMAGES - EVEN IF KNAUER HAS BEEN ADVISED OF THE POSSIBILTY OF SUCH DAMAGES, OR FOR ANY CLAIM BY ANY OTHER PARTY. Some countries do not allow the exclusion of limitation of special, incidental, or consequential damages, so the above limitation, or exclusion may not apply to the Licensee. Any liability is - in any case - lim-ited to the license fee.8. Limitation of Remedies. KNAUER´s entire liability and Licensee's exclusive remedy shall be:(a)the replacement of any media not meeting KNAUER´s limited warranty which is returned to KNAUER, or (b)if KNAUER or its distributor is unable to deliver replacement media which is free of defects in materials or workmanship, the Licensee may terminate this Agreement by returning the EuroOsmo 7400 V1.5 and the Licensee's money will be refunded. 9. This license is not assignable. The Licensee does not have the right to assign rights or grant sublicenses.10. License Fee / Royalty. For the license to use the contractual EuroOsmo®7400 V1.5 in the above defined scope and within the restrictions the Licensee pays to the Licensor a fee. The amount becomes due and payable upon delivery of the program discs. The license to use of the program commences with payment of the royalty.11. If a provision of this Agreement should be invalid or become invalid, then the legal effect of the other provisions shall remain unchanged. Instead of an invalid provision, a valid provision shall be deemed to be agreed which comes closest to what the parties intended economically.This document is the complete Agreement between KNAUER and Licensee; it supersedes all and any prior agreement or offer. All and any modifications of this agreement have to be made in writing.For all controversies out of this Agreement, the District Court Berlin shall have exclusive jurisdiction. The applicable law shall be that of the Federal Republic of Germany (German Civil Code) excluding the UN-Code on the international sale of goods.

K-7400S user manual V3700

Page 33: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable

K-7400S user manual V3700

33 Index

IndexAAccessories 6Ambient conditions 8AVV marking 30

CCalibrationPre-calibration 16

calibration 16computer connection 10contamination 7Ddecontamination 7Device Overview 5disposal 7IInstallation 8Installation Site 8IQ 19

Llabel 10Legal Information 29

Mmarkings 6OOQ 19

Ppacking list 10power cable 7

power strip 7power supply 7professional group 6RReference Thermometer 28repair 7repeat orders 29

Ssafety 6safety equipment 6scope of delivery 10setup 9solventself-ignition point 7

space requirements 8spare parts 6symbols 6TTechnical Data 28testInstallation Qualification 19Operation Qualification 19

transportpackaging 9

transport damage 29

Uunpacking 9Wwarning signs 8warranty 29

Page 34: K-7400S Semi-Micro Osmometer Freezing Point …flowspek.ch/download/osmometer-K-7400S-en.pdfThe osmometer needs a three-core power plug. Electric shock Make sure that the power cable