Transporting Chemicals In and Out of the Glove Box Jessica A. Haak (Designer) Jeffrey S. Quesnel (Subject Matter Expert)
Department of Education, Concordia University, Montreal, Quebec
Introduction
1. The Glove Box Computer
2. Bringing Chemicals In
3. Bringing Chemicals Out
✪ Never open the outer antechamber door when the inner antechamber door is open or vice versa. ✪ Make sure your chemicals is sealed inside a glove box certified container. ✪ Always do three full vacuum and refill cycles before bringing chemicals into the glove box. ✪ Check the glove box computer regularly while working in the glove box.
4 Important Glove Box Pointers
Ideal Conditions
Cause for Concern
✪ Should any of these conditions occur contact a senior lab member right away.
Step 1: Ensure that the antechamber gauge is at 0 psi.
a. Resolve any deviation by turning the antechamber valve to the “refill” position.
b. Wait until the vacuum gauge reaches 0 psi.
c. Turn the antechamber valve to the marked “closed” position..
Step 2: Open the outer antechamber door.
Procedure for Model A a. Turn the antechamber door knob
counterclockwise until the door feels loose.
b. Turn the antechamber door counterclockwise 25°.
c. Detach the antechamber door and place it knob down on a flat surface.
Procedure for Model B a. Pull the antechamber handle to
unlatch the antechamber hinge. b. Pull the door open to a 90° angle..
Step 3: Load a chemical in a sealed container into the antechamber.
a. Slide the antechamber tray out of the glove box by one foot.
b. Place the chemical in a sealed container on the tray.
c. Push the antechamber tray completely into the antechamber.
Step 4: Close the outer antechamber door.
Procedure for Model A a. Pickup the outer antechamber door. b. Fit the outer antechamber door grooves into the
door frame. c. Turn the outer antechamber door clockwise 25°. d. Turn the outer antechamber door knob
clockwise, tightening only using your thumb, index finger, and middle finger. (called the 3 finger rule), until it is difficult to continue turning.
Procedure for Model B a. Push the outer antechamber door towards the
glove box until it is touching the door frame. b. Line the outer antechamber handle with the
outer antechamber latch and pull down on the antechamber handle to seal the door.
Antechamber Vacuum Gauge
At 0 psi.
Model B
Model A
✪ Before bringing any solid chemicals into the antechamber make sure that the lid is on, but loosened by half a turn. This will allow room for the air to escape when it goes into the glove box.
✪ Did you know that not all chemicals can go into the glove box? When in doubt talk to your trainer or a senior lab member.
✪ Be careful when tightening model A doors. Make sure that the next user will be able to comfortably open and close the door.
✪ Always use the “3 finger rule” when tightening model A doors: your thumb, index finger and middle finger.
Cycles 1&2 a. Turn the antechamber valve to
the marked “evacuate” position. b. Turn the antechamber cycle
wheel to “1”. c. Wait five minutes. d. Turn the antechamber valve to
the marked “refill” position until between 5 psi and 15 psi is reached on the antechamber vacuum gauge to complete cycle 1.
e. Repeat steps a-d again, this time turning the cycle wheel to “2”
Cycle 3 a. Turn the antechamber valve to
the marked “evacuate” position.
b. Turn the antechamber cycle wheel to “3”.
c. Wait five minutes. d. Turn the antechamber valve to
the marked “refill” position until 0 psi on the antechamber vacuum gauge is reached to complete cycle 3.
e. Put the antechamber valve to the “closed” position.
f. Turn the antechamber cycle wheel to “N2”.
✪ Always use the cycle wheel to indicate what cycle you are in, this will avoid confusion when there are multiple users.
4. Preparing for the Next User
Step 6: With your hands in the gloves, Open the inner antechamber door. Step 7: With your hands in the gloves pick up the sealed chemicals and place in the glove box. Step 8: With your hands in the gloves, close the inner anti-chamber door.
Step One: Open the inner antechamber door.
Step Two: Place the sealed chemical in the antechamber.
Step Three: Close the inner antechamber door.
Step Four: Open the outer antechamber door.
Step Six: Close the outer Antechamber door.
Step Five: Remove the sealed chemical.
✪ Label your chemicals. ✪ Tidy up chemicals and equipment. ✪ Sweep the bench. ✪ Place all dirty or broken glassware, empty
bottles, dirty cloths, and debris in the glove box garbage.
✪ Clean the inside of the gloves with a cloth & solvent.
✪ Clean the outside of the gloves with a cloth & water.
O2: Between 0.1-5 ppm H2O: Between 0.1-5 ppm Pressure: Between 1-6 mbar Four Circulator Icons: Green Alarm Sign: Off
O2: Above 5 ppm H2O: Above 5 ppm Pressure: Above 6 mbar Four Circulator Icons: Red Alarm Sign: On
Step 5: Complete three full vacuum cycles using the antechamber valave.
Evacuate
Refill
Antechamber Valve
Evacuate
Refill
Closed
Antechamber Valve