January 3, 2008

43
Cosmic RAy Telescope for the Effects of Radiation CRaTER Pre-Ship Review (PSR) Instrument Calibration Science Requirements Compliance Justin C Kasper Smithsonian Astrophysical Observatory January 3, 2008

description

CRaTER Pre-Ship Review (PSR) Instrument Calibration Science Requirements Compliance Justin C Kasper Smithsonian Astrophysical Observatory. January 3, 2008. Outline. Calibration Relate the ADU of the Pulse Height Analysis to the original energy deposited for each detector - PowerPoint PPT Presentation

Transcript of January 3, 2008

Page 1: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER Pre-Ship Review(PSR)

Instrument CalibrationScience Requirements Compliance

Justin C KasperSmithsonian Astrophysical Observatory

January 3, 2008

Page 2: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Outline

• Calibration– Relate the ADU of the Pulse Height Analysis to the original energy

deposited for each detector– Models of energy deposition– Description of Massachusetts General Hospital Proton Facility– Example of MGH observations– Description of calibration method– Demonstration of success with EM– Discussion of alternative methods for redundant confirmation of

calibration• Requirements

– Review– Results– Details

Page 3: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Relevant Documents

• Project Controlled– ESMD-RLEP-0010 (Revision A effective November 30 2005)

– LRO Mission Requirements Document (MRD) – 431-RQMT-00004

– CRaTER Data ICD – 431-ICD-000104

• CRaTER Configuration Controlled Documents– 32-01205 Instrument Requirements Document

– 32-01206 Performance and Environmental Verification Plan

– 32-01207 Calibration Plan

– 32-03010 CRaTER Digital Subsystem Functional Specification

– 32-05001 Detector Specification Document

– 32-05002 CRaTER Functional Instrument Description and Performance Verification Plan

Page 4: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER Layout

Page 5: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER Telescope Layout

Telescope in cross-sectionA single detector (D5 for EM)

A pair of thin and thick detectors (D5 and D6 for EM)

Page 6: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

6

CRaTER Functional Diagram

Page 7: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Response Function

• Calibration Requirement– Relate ADU to Energy Deposited at 0.5% level

• Experience with EM and both FM– System is linear at 0.1% level

– Noise level is less than 0.15%

– Temperature dependence is less than 0.1%

– Drift with time is less than 0.02%/month

• Sufficient Response Function– Ei = Gi * (Ci – Oi)

– Ei is energy deposited [MeV]

– Gi is gain [MeV/ADU]

– Ci is PHA amplitude [ADU]

– Oi is an offset [ADU]

Page 8: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Energy Loss of Protons in Silicon

Simulations of energy deposit

Observations of scattering

Page 9: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Signature of Spread Energy Protons in Detector PairSimulation (Units of MeV) Observation (Units of ADU)

Page 10: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

MGH Beam Testing with EM

Page 11: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Algorithm for Calibrating Instrument

Page 12: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Observations and Simulations

Page 13: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

FM S/N 01 Calibration Parameters(CRA0120071216130951L0 CRA0120071216133252L0 CRA0120071216135457L0)

Detector Var Value Error Units % ErrorD1 Gain 78.360688 0.13793621KeV/ADU 0.18

Offset -0.45246847 0.096347427ADU 21.29

D2 Gain 22.676271 0.058317026KeV/ADU 0.26 Offset -7.4135601 0.56938238ADU 7.68

D3 Gain 81.768613 0.099649736KeV/ADU 0.12

Offset -0.19174824 0.053966765ADU 28.14

D4 Gain 22.58814 0.040475074KeV/ADU 0.18 Offset 0.70068405 0.11857201ADU 16.92D5 Gain 83.6859 0.17280925KeV/ADU 0.21

Offset -0.68360328 0.068337712ADU 10.00

D6 Gain 22.299163 0.048207834KeV/ADU 0.22 Offset 3.2444374 0.10117969ADU 3.12

Page 14: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Calibration and Derived Parameters

Flight Model S/N 02(Selected for delivery to NASA)

Flight Model S/N 01

Var Unit D1 D3 D5 D2 D4 D6 D1 D3 D5 D2 D4 D6Offset ADU -0.3 -0.5 -0.4 0.9 -2.2 0.9 -0.5 -0.2 -0.7 -7.4 0.7 3.2

GainKeV/ADU 81.6 82.9 80.6 22.2 22.4 22.2 78.4 81.8 83.7 22.7 22.6 22.3

Min E KeV 435.5 453.5 436.0 90.8 162.5 91.5 427.3 424.5 475.6 281.5 97.1 39.1

Max EMeV/um 334.1 339.6 330.2 91.0 92.0 90.9 321.0 334.9 342.8 93.1 92.5 91.3

Thick um 148 149 149 1000 1000 1000 152 147 148 933 933 932Min LET

KeV/um 2.9 3.0 2.9 0.1 0.2 0.1 2.8 2.9 3.2 0.3 0.1 0.0

Max LET

MeV/um 2.3 2.3 2.2 0.1 0.1 0.1 2.1 2.3 2.3 0.1 0.1 0.1

Max/Min # 24847 14028 24227 20289 21890 55148

Page 15: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

LET Range for Both Instruments

Flight Model S/N 02

(Selected for delivery to

NASA)

Flight Model S/N 01

(Now flight spare)

Page 16: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

16

Level 1 Requirements and Data Products

RLEP-LRO-M10

RLEP-LRO-M20

Provide Linear Energy Transfer (LET) spectra of cosmic rays

(particularly above 10 MeV), most critically important to the

engineering and modeling communities to assure safe, long-term, human presence in space.

Provide LET spectra behind different amounts and types of areal density, including tissue equivalent

plastic.

The LRO shall characterize the deep space radiation environment at energies in excess of 10 MeV in lunar orbit, including neutron albedo.

The LRO shall measure the deposition of deep space radiation on human equivalent tissue while in the lunar orbit environment.

Level 1 Requirements Level 1 Data Products

Page 17: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Flow of Level 1 to Level 2 Requirements

RLEP-LRO-M10Characterize deep

space radiation environment

RLEP-LRO-M20Radiation effects on human equivalent

tissue

Measure the Linear Energy Transfer (LET) spectrum

Minimum pathlength through total TEP > 60 mm

Two asymmetric TEP components 1/3 and 2/3 total length

Measure change in LET spectrum through Tissue Equivalent Plastic (TEP)

Minimum LET measurement 0.2 keV per micron

Maximum LET measurement 2 MeV per micron

Energy deposition resolution < 0.5% max energy

Minimum full telescope geometrical factor 0.1 cm2 sr

01

03

04

02

05

06

07

08

Level 2 Instrument Requirements

Level 1 Parent Requirements

Page 18: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Flow of Requirements

01

03

04

02

05

06

07

Thin and thick detector pairs 140 & 1000 μm

Minimum energy < 250 keV

Nominal instrument shielding > 1524 μm Al

Nadir and zenith field of view shielding < 762 μm

Telescope stack

Pathlength constraint < 10% for D1D6

Zenith field of view < 33o

Nadir field of view < 70o

Calibration system

Event selection

Maximum transmission rate > 1000 events/second

01

02

03

04

05

06

07

08

09

10

11

Level 2 Parent Requirements

Level 3 Instrument Requirements

Page 19: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Minimum LET measurement 0.2 keV per micron

Maximum LET measurement 2MeV per micron

Energy deposition resolution < 0.5% max energy

• Berkeley National Laboratory 88” Cyclotron (Ions, 20 MeV/nuc)

• Massachusetts General Hospital Proton Therapy Center (0-300 MeV p)

• Brookhaven National Laboratory (Gev/nuc Fe)

MGH Proton Accelerator

Testing Validation Techniques with the CRaTER Prototype

Page 20: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Level 2 Requirements Verification Outline

Item Requirement Quantity MethodCRaTER-L2-01 Measure the Linear Energy Transfer

(LET) spectrumLET     A

CRaTER-L2-02 Measure change in LET spectrum through Tissue Equivalent Plastic (TEP)

TEP     A

CRaTER-L2-03 Minimum pathlength through total TEP > 60 mm I    

CRaTER-L2-04 Two asymmetric TEP components 1/3 and 2/3 I    

CRaTER-L2-05 Minimum LET measurement < 0.25 keV per micron

  T  

CRaTER-L2-06 Maximum LET measurement > 2 MeV per micron

  T  

CRaTER-L2-07 Energy deposition resolution < 0.5% max energy   T  

CRaTER-L2-08 Minimum D1D6 geometrical factor > 0.1 cm2 sr I    

Page 21: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Level 2 Verification

Item Requirement Quantity Method S/N 2 S/N 1CRaTER-L2-01 Measure the Linear

Energy Transfer (LET) spectrum

LET     A Verified instrument measures LET using energetic particle beams, radioactive sources, models

CRaTER-L2-02 Measure change in LET spectrum through Tissue Equivalent Plastic (TEP)

TEP     A MGH Beam Run September MGH Beam Run December

CRaTER-L2-03 Minimum pathlength through total TEP

> 60 mm I     80.947 mm as measured +/- 0.001 mm

80.946 mm as measured +/- 0.001 mm

CRaTER-L2-04 Two asymmetric TEP components

1/3 and 2/3 I     26.972 mm and 53.992 mm sections of TEP used, both

+/- 0.001 mm measured with micrometer

26.979 mm and 53.967 mm sections of TEP used, both +/-

0.001 mm measured with micrometer

CRaTER-L2-05 Minimum LET measurement

< 0.25 keV per micron

  T   2.9, 3.0, 2.9, 0.09, 0.16, 0.09 2.8, 2.8, 3.2, 0.3, 0.1, 0.04

CRaTER-L2-06 Maximum LET measurement

> 2 MeV per micron

  T   2.2, 2.2, 2.2, 0.09, 0.09, 0.09 2.1, 2.2, 2.3, 0.09, 0.09, 0.09

CRaTER-L2-07 Energy deposition resolution

< 0.5% max energy

  T   <0.1% electronics from external pulser, <0.06% detectors using width of alpha source, <0.25% gain uncertainty from

beam calibrations at MGH

CRaTER-L2-08 Minimum D1D6 geometrical factor

> 0.1 cm2 sr I     0.57 cm^2 sr derived from mechanical drawings

Page 22: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Level 3 Requirements Verification OutlineItem Requirement Quantity Method

CRaTER-L3-01 Thin and thick detector pairs

140 and 1000 microns I    

CRaTER-L3-02 Minimum energy < 250 keV   T  CRaTER-L3-03 Nominal instrument

shielding1524 microns Al I    

CRaTER-L3-04 Nadir and zenith field of view shielding

762 microns Al I    

CRaTER-L3-05 Telescope stack Shield, D1D2, A1, D3D4, A2, D5D6, shield

I    

CRaTER-L3-06 Pathlength constraint 10% for D1D6 I    CRaTER-L3-07 Zenith field of view <34 degrees D1D4 I    CRaTER-L3-08 Nadir field of view <70 degrees D3D6 I    CRaTER-L3-09 Calibration system Variable rate and gain   T  CRaTER-L3-10 Event selection 64-bit mask   T  

CRaTER-L3-11 Maximum event transmission rate

1,000 events/sec   T  

Page 23: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Item Requirement Quantity Method S/N 2 S/N 1CRaTER-L3-01 Thin and thick

detector pairs140 and 1000 microns

I     148, 149, 149, 1000, 1000, 1000 152, 147, 148, 933, 933, 932

CRaTER-L3-02 Minimum energy

< 250 keV   T   435.4, 453.4, 436.0, 90.8, 162.4, 91.4

427.2, 424.52, 475.6, 281.4, 97.1, 39.1

CRaTER-L3-03 Nominal instrument shielding

1524 microns Al I     Verified by inspection of instrument and mechanical drawings

CRaTER-L3-04 Nadir and zenith field of view shielding

762 microns Al I     Measured to be 812.8 microns zenith and 810.3 microns nadir

Measured to be 812.8 microns zenith and 810.3 microns nadir

CRaTER-L3-05 Telescope stack Shield, D1D2, A1, D3D4, A2, D5D6, shield

I     Verified by inspection of instrument and mechanical drawings

CRaTER-L3-06 Pathlength constraint

10% for D1D6 I     Using geometry of telescope and uniform distribution on sky < 5%

CRaTER-L3-07 Zenith field of view

<34 degrees D1D4

I     33 degrees from mechanical drawing

CRaTER-L3-08 Nadir field of view

<70 degrees D3D6

I     69 degrees from mechanical drawing

CRaTER-L3-09 Calibration system

Variable rate and gain

  T   Verified by use of the internal calibration system

CRaTER-L3-10 Event selection 64-bit mask   T   Verified by testing in a beam and by using ambient cosmic ray muons

CRaTER-L3-11 Maximum event transmission rate

1,200 events/sec   T   Verified using the internal pulser at 2 kHz and the MGH beam facility

Level 3 Verification

Page 24: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-01 Measure the LETSpectrum

• Requirement– The fundamental measurement of the CRaTER instrument shall be of the linear energy transfer

(LET) of charged energetic particles, defined as the mean energy absorbed (∆E) locally, per unit path length (∆l), when the particle traverses a silicon solid-state detector.

Simulation (Units of MeV) Observation (Units of ADU)MGH proton measurements

Page 25: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-02 Measure LET Spectrum after Passing through TEP

• Requirement– The LET spectrum shall be measured before entering and after propagating though a

compound with radiation absorption properties similar to human tissue such as A-150 Human Tissue Equivalent Plastic (TEP).

Breakup of iron into fragments after passing through TEP measured at BNL with EM

Iron nuclei enter D1

Iron nuclei enter D3

Iron breaks up within first section of TEP

Fragments pass

through te

lescope

Page 26: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-03 Minimum Pathlength through total TEP

• Requirement– The minimum pathlength through the total amount of TEP in the telescope shall be at least

60 mm.

Page 27: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-04 Two asymmetric TEP components

• Requirement– The TEP shall consist of two components of different length, 1/3 and 2/3 the total length of

the TEP. If the total TEP is 61 mm in length, then the TEP section closest to deep space will have a length of approximately 54 mm and the second section of TEP will have a length of approximately 27 mm.

Page 28: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-05 Minimum LET measurement

• Requirement– At each point in the telescope where the LET spectrum is to be observed, the minimum LET

measured shall be no greater than 0.25 keV/ micron in the Silicon.

Page 29: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-06 Maximum LET measurement

• Requirement– At each point in the telescope where the LET spectrum is to be observed, the maximum

LET measured shall be no less than 2 MeV/ micron in the Silicon.

Page 30: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-07 Energy deposition resolution

• Requirement– The pulse height analysis of the energy deposited in each detector shall have an energy

resolution better than 1/200 the maximum energy measured by that detector.

Upper limit on system noise is less than 0.15% < 0.5%

Page 31: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L2-08 Geometrical Factor

• Requirement– The geometrical factor created by the first and last detectors shall be at least 0.1 cm2 sr.

2

22

1

2222

21

226

21

2 42

1rrzrrzrrG

r1 = 1.75 cmr2 = 1.75 cmZ = 12.71 – 0.25 = 12.46 cmG = 0.57 > 0.1 cm2 sr

Page 32: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-01 Thin and thick detector pairs

• Requirement– The telescope stack shall contain adjacent pairs of thin and thick Silicon detectors. The

thickness of the thin detectors will be approximately 140 microns and the thick detectors will be approximately 1000 microns.

Page 33: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-02 Minimum Energy

• Requirement– The Silicon detectors shall be capable of measuring a minimum energy deposition of 250

keV or lower.

Page 34: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-03 Nominal instrument shielding

• Requirement– The equivalent shielding of the CRaTER telescope outside of the zenith and nadir fields of

view shall be no less than 1524 microns (0.060 inches) of aluminum.

Verified by inspection of mechanical drawings

Page 35: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-04 Nadir and zenith field of view shielding

• Requirement– The zenith and nadir fields of view of the telescope shall have no more than 762 microns

(0.030) of aluminum shielding.

EM built to 762 microns

FM S/N 2 end caps specified to be 0.030” +/- 0.002”, or 762 +/- 51 micronsMeasured to be 812.8 microns zenith and 810.3 microns nadir

812 and 810 microns > 762 microns

Within the design tolerance of the mechanical drawing, but nonetheless is slightly greater than the requirement specified. We have filed a Non-conforming Material Report (NMR).

The science team feels that this difference in thickness does not affect the quality of the measurements made, since it is still sufficiently thin to allow protons to enter the telescope over the desired energy range, as long as the thickness is measured so it can be fed into the models.

Page 36: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-05 Telescope stack

• Requirement– The telescope shall consist of a stack of components labeled from the zenith side as zenith

shield (S1), the first pair of thin (D1) and thick (D2) detectors, the first TEP absorber (A1), the second pair of thin (D3) and thick (D4) detectors, the second TEP absorber (A2), the third pair of thin (D5) and thick (D6) detectors, and the final nadir shield (S2).

Verified by direct inspection

Page 37: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-06 Full telescope pathlength constraint

• Requirement– The root mean squared (RMS) uncertainty in the length of TEP traversed by a particle that

traverses the entire telescope axis shall be less than 10%.

RMS from Monte Carlo is 0.78% < 10%

Page 38: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-07 Zenith field of view

• Requirement– The zenith field of view, defined as D2D5 coincident events incident from deep space using

the naming convention in CRaTER-L3-04, shall be less than 34 degrees full width.

By inspection 33 < 34 degrees

Page 39: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-08 Nadir field of view

• Requirement– The nadir field of view, defined as D4D5 coincident events incident from the lunar surface,

shall be less than 70 degrees full width.

By inspection, 69 < 70 degrees

Page 40: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-09 Calibration system

• Requirement– The CRaTER electronics shall be capable of injecting calibration signals at with

different amplitudes and rates into the measurement chain.

Page 41: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-10 Event selection

• Requirement– A command capability shall exist to allow specification of detector coincidences

that will be analyzed and sent to the spacecraft for transmission to the ground.

Shown that this works using proton beam at MGH

Page 42: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

CRaTER-L3-11 Maximum event rate

• Requirement– CRaTER will be capable of transmitting primary science data, namely the energy deposited

in each of the detectors, on at least 1000 events per second.

Verified by exposure to proton and iron beams, and by running internal calibration system at 2kHz

Page 43: January 3, 2008

Cosmic RAy Telescope for the Effects of Radiation

Conclusions

• We have conducted multiple calibration activities for both flight models

– External pulse generator

– Radioactive sources

– Proton beam at MGH

• Absolute calibration of each detector

– Linear response

– Solve for gain and offsets for each detector by fitting to numerical simulation

• Verification

– Performance requirements verified by inspection or testing

– Results from beam calibration propagated to determine minimum energy, energy resolution, min and max LET