ITL Project Ref.: S101851.02 Page 1 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086 V3.0 12/09/2004 TRF originator: VDE COMPONENT TEST REPORT EN 61010–1:2001 (2nd Edition), IEC 61010-1:2001 (2nd Edition) Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements Report Reference No. .................... : S101851.02 Tested by (name and signature).......: Avi Miles Approved by (name and signature) ..: Shmuel Gnatt Date of issue .....................................: 15 February 2012 Contents ...........................................: Pages 62 Testing Laboratory .........................: ITL (Product Testing) Ltd. Address .............................................: 1 Bat-Sheva St., POB 87 Lod 71100, Israel Testing location/procedure ...............: PM120 Address .............................................: As above Applicant’s name ............................: Galcon Ltd, Address .............................................: Kfar Blum 12150, Israel Test specification: Standard ...........................................: EN 61010–1:2001 (2nd Edition), IEC 61010-1:2001 (2 nd Edition) Test procedure..................................: PM120 Non-standard test method ................: Test Report Form No. .....................: IEC61010_C TRF Originator ..................................: VDE Master TRF .......................................: Dated 01-07-27 Copyright © 2001 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE), Geneva, Switzerland. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context. Test item description......................: Irrigation Control System Manufacturer………………………….: Galcon Ltd. Trademark ........................................: Galcon Model/Type reference .......................: 7001, 7001D, 7001DS, 7001D-C, 7001D-CS, 7101, 7101D, 7101DS, 7101D-C, 7101D-CS Rating(s) ...........................................: Battery operated 9V

Transcript of ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf ·...

Page 1: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 1 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

COMPONENT TEST REPORTEN 61010–1:2001 (2nd Edition), IEC 61010-1:2001 (2nd Edition)Safety requirements for electrical equipment for measurement,

control, and laboratory usePart 1: General requirements

Report Reference No. .................... : S101851.02Tested by (name and signature).......: Avi Miles

Approved by (name and signature) ..: Shmuel Gnatt

Date of issue.....................................: 15 February 2012

Contents ...........................................: Pages 62

Testing Laboratory .........................: ITL (Product Testing) Ltd.

Address.............................................: 1 Bat-Sheva St., POB 87 Lod 71100, Israel

Testing location/procedure ...............: PM120

Address.............................................: As above

Applicant’s name ............................: Galcon Ltd,

Address.............................................: Kfar Blum 12150, Israel

Test specification:

Standard ...........................................: EN 61010–1:2001 (2nd Edition), IEC 61010-1:2001 (2nd Edition)

Test procedure..................................: PM120

Non-standard test method ................: —

Test Report Form No. .....................: IEC61010_C

TRF Originator ..................................: VDE

Master TRF.......................................: Dated 01-07-27

Copyright © 2001 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE),Geneva, Switzerland. All rights reserved.This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged ascopyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting fromthe reader's interpretation of the reproduced material due to its placement and context.

Test item description......................: Irrigation Control System

Manufacturer………………………….: Galcon Ltd.

Trademark ........................................: Galcon

Model/Type reference.......................: 7001, 7001D, 7001DS, 7001D-C, 7001D-CS, 7101, 7101D,7101DS, 7101D-C, 7101D-CS

Rating(s) ...........................................: Battery operated 9V

Page 2: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 2 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

Test item particulars:

Type of item tested……………………………..: Irrigation system

Description of equipment function…………….: Irrigation system to control water timing irrigation

Installation/overvoltage category……………...: N/A

Pollution degree………………………………...: III

Environmental rating…………………………..: The product located outdoors

Equipment mobility…………………………….: Fixed

Connection to mains supply………………….. Battery operated

Operating conditions…………………………..: Continuous

Overall size of the equipment (L x W x H)…..: 140x60x60mm

Mass of the equipment (kg)…………………..: Aprox. 1kg

Marked degree of protection to IEC 60529…: IP55

Accessories and detachable parts included inthe evaluation…………………………………..:

N/A

Option…………………………………………: N/A

Test case verdicts:

Test case does not apply to the test object…..: N

Test object does meet the requirement………: P (Pass)

Test object does not meet the requirement….: F (Fail)

Testing:

Date of receipt of test item…………………….: 23/2/2011

Date (s) of performance of tests………………: 23-24//2/2011

Page 3: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 3 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

General remarks:

This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory andappended to a CB Test Certificate issued by an NCB in accordance with IECEE 02.This report shall not be reproduced, except in full, without the written approval of the issuing testinglaboratory.The test results presented in this report relate only to the item(s) tested."(see remark #)" refers to a remark appended to the report."(see Annex #)" refers to an annex appended to the report."(see Form A.#)" refers to a table appended to the report.

Throughout this report a point is used as the decimal separator.

Page 4: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 4 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

Copy of marking plate:

Appendix 1 - Summary of test results (information/comments):

This Amendment Test Report S101851.02 adds models 7001, 7001D-C, 7001D-CS, 7101, 7101DS,7101D-C, 7101D-CS to the originally investigated models 7001D and 7001DS (Project S101851.01). Newmodels incorporate different software or are partial version of 7001DS. No testing deemed necessary.

EUT Description:

The EUT is an irrigation control system, powered by 9Vdc battery alkaline battery regarded inherently

limited energy source. The EUT is made to control water irrigation timing at field conditions.

Page 5: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 5 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDE

List of Critical Components

TABLE: 3 - List of components and circuits relied on for safety

No. Object/part no. Manufacturer/Trademark

Type/model Technical data Standard Mark(s) ofconformity

1 Plastic Enclosure A.D.D.PLAST

BI/DI-4 IP 65 ManufacturerDeclaredCompliance

2 Printed circuit boards Galcon Rated FV-1 min ManufacturerDeclaredCompliance

3 Battery Any Any Rated: 9VDC,Alkaline, type 9-Volt 6LR61

---

Page 6: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 6 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

5 MARKING AND DOCUMENTATION —

5.1.1 General —

Required equipment markings are: ---

Equipment shall be marked in English language. P

Visible: P

From the exterior; or N

After removing a cover; or N

Opening a door N

After removal from a rack or panel N

Not put on parts which can be removed by anOPERATOR

N

Letter symbols (IEC 60027) used N

Graphic symbols (IEC 61010-1: Table 1) used N

5.1.2 Identification —

Equipment is identified by: —

5.1.2a) Manufacturer’s or supplier’s name or trademark P

5.1.2b) Model number, name or other means P

Manufacturing location identified N

5.1.3 Mains supply —

Equipment is marked as follows: —

5.1.3a) Nature of supply: —

1) a.c. RATED mains frequency or range offrequencies............................................................:

N

2) d.c. with symbol 1 P

5.1.3b) RATED supply voltage(s) or range: .......................... P

5.1.3c) Max. RATED power (W or VA) or input current:....... P

The measured value not more than 110 % Not applicable for ‘component’type product.

N

If more than one voltage range: —

Separate values marked; or N

Values differ by less than 20 % N

5.1.3d) OPERATOR-set for different RATED supply voltages: —

Indicates the equipment set voltage N

PORTABLE EQUIPMENT indication is visible from theexterior

N

Changing the setting changes the indication N

5.1.3e) Accessory mains socket-outlets accepting standardmains plugs are marked:

Page 7: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 7 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

With the voltage if it is different from the mainssupply voltage .......................................................:

N

For use only with specific equipment N

If not marked for specific equipment it is markedwith:

The maximum RATED current or power; or N

Symbol 14 with full details in the documentation N

5.1.4 Fuses —

OPERATOR replaceable fuse marking(see also 5.4.5): .....................................................

N

5.1.5 TERMINALS, connections and operating devices —

Where necessary for safety, indication of purposeof TERMINALS, connectors, controls and indicatorsmarked

N

If insufficient space, symbol 14 used N

5.1.5.1 TERMINALS N

Mains supply TERMINALS identified N

Other TERMINAL marking: ........................................... N

5.1.5.1a) FUNCTIONAL EARTH TERMINALS (symbol 5 used) N

5.1.5.1b) PROTECTIVE CONDUCTOR TERMINALS: —

Symbol 6 is placed close to or on the TERMINAL; OR N

Part of appliance inlet N

5.1.5.1c) TERMINALS of measuring and control circuits(symbol 7 used)

N

5.1.5.1d) HAZARDOUS LIVE TERMINALS supplied from theinterior

Standard MAINS socket outlet; or N

RATINGS marked; or N

Symbol 14 used P

5.1.5.1e) ACCESSIBLE FUNCTIONAL EARTH TERMINALS: —

Self-evident; or N

Indication (symbol 8 acceptable) N

5.1.5.2 Measuring circuit TERMINALS —

For TERMINALS other than those permanentlyconnected and not ACCESSIBLE:

RATED voltage or current marked N

Unless clear indication that below limits: —

Maximum RATED voltage to earth is marked; or N

For specific connection to other equipmentTERMINALS only, and means for identifying provided

N

Page 8: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 8 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Appropriate measurement category marked(CAT II, CAT III or CAT IV); or

N

No measurement category marked (CAT I) N

Required markings are adjacent to TERMINALS; OR N

If insufficient space: —

On the RATING plate or scale plate; or N

TERMINAL is marked with symbol 14 N

5.1.6 Switches and circuit breakers —

If disconnecting device, on or off position marked N

5.1.7 Equipment protected by DOUBLE INSULATION orREINFORCED INSULATION

Protected throughout (symbol 11 used) N

Only partially protected (symbol 11 not used) N

5.1.8 Field-wiring TERMINAL boxes —

If TERMINAL or ENCLOSURE exceeds 60 C: —

Cable temperature RATING marked N

Marking visible or beside TERMINAL N

5.2 Warning markings —

Visible when ready for NORMAL USE N

Are near or on applicable parts N

Symbols and text correct dimensions and colour N

If necessary marked with symbol 14 N

Statement to isolate or disconnect N

5.3 Durability of markings —

The required markings remain clear and legible inNORMAL USE

(see Form A.4) P

5.4 Documentation —

5.4.1 General —

Equipment is accompanied by documentationwhich includes:

5.4.1a) Intended use P

5.4.1b) Technical specification P

5.4.1c) Instructions for use P

5.4.1d) Name and address of manufacturer or supplier P

5.4.1e) Information specified in 5.4.2 to 5.4.5 —

5.4.1f) If marking of TERMINALS required, definition ofmeasurement category

N

5.4.1g) If CAT 1: —

Page 9: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 9 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Warning N

RATINGS N

Warning statements and a clear explanation ofwarning symbols:

Provided in the documentation; or N

Information is marked on the equipment N

5.4.2 Equipment RATINGS —

Documentation includes: —

5.4.2a) Supply voltage or voltage range N

Frequency or frequency range N

Power or current RATING N

5.4.2b) Description of all input and output connections N

5.4.2c) RATING of insulation of external circuits,when such circuits are nowhere ACCESSIBLE

N

5.4.2d) Statement of the range of environmental conditions N

5.4.2e) Degree of protection (IEC 60529) N

5.4.3 Equipment installation —

Documentation includes instructions for: —

5.4.3a) Assembly, location and mounting P

5.4.3b) Protective earthing N

5.4.3c) Connections to supply N

5.4.3d) PERMANENTLY CONNECTED EQUIPMENT: —

1) Supply wiring requirements N

2) If external switch or circuit-breaker, requirementsand location recommendation

N

5.4.3e) Ventilation requirements N

5.4.3f) Special services (e. g. air, cooling liquid) Outdoor installation. N

5.4.3g) Maximum sound power level Outdoor installation. N

5.4.3h) Instructions about sound pressure Outdoor installation. N

5.4.3i) Permanently connected measuring TERMINALS: —

Measurement category N

RATED MAXIMUM WORKING VOLTAGE or current N

5.4.4 Equipment operation —

Instructions for use include: —

5.4.4a) Identification of operating controls P

5.4.4b) Positioning for disconnection N

5.4.4c) Interconnection N

5.4.4d) Specification of intermittent operation limits N

Page 10: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 10 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

5.4.4e) Explanation of symbols used P

5.4.4f) Replacement of consumable materials N

5.4.4g) Cleaning and decontamination (see 11.2) N

5.4.4h) Listing of any poisonous or injurious gases andquantities

N

5.4.4i) Risk-reduction procedures relating to flammableliquids

N

A statement about protection impairment if used ina manner not specified by the manufacturer

N

5.4.5 Equipment maintenance —

Instructions include: —

Sufficient preventive maintenance and inspectioninformation

P

Replacement of hoses, etc. N

Specific battery type P

Any manufacturer specified parts N

RATING and characteristics of fuses N

6 PROTECTION AGAINST ELECTRIC SHOCK (see Form A.5) —

6.1 General —

6.1.1 Requirements —

ACCESSIBLE parts not HAZADOUS LIVE in NORMALCONDITION and SINGLE FAULT CONDITION

N

Conformity is checked by the determination of 6.2and 6.3 followed by the tests of 6.4 to 6.11

6.1.2 Exceptions No exceptions —

Capacitance test N

Parts not HAZARDOUS LIVE 10 s after interruption ofsupply

N

6.2 Determination of ACCESSIBLE parts —

6.2.1 General examination N

6.2.2 Openings above parts that are HAZARDOUS LIVE No openings above parts thatare HAZARDOUS LIVE

N

6.2.3 Openings for pre-set controls No such openings N

6.3 Permissible limits for ACCESSIBLE parts —

6.3.1 Values in NORMAL CONDITION N

6.3.2 Values in SINGLE FAULT CONDITION N

6.4 Protection in NORMAL CONDITION(see 6.2, 6.3.1, 6.7, 6.8 and 8.1)

N

6.5 Protection in SINGLE FAULT CONDITION —

Additional protection is provided by: —

Page 11: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 11 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

One or more of 6.5.1 to 6.5.3; or N

Automatic disconnection of the supply (6.5.4) Not used N

6.5.1 Protective BONDING —

ACCESSIBLE conductive parts: —

Separated by DOUBLE INSULATION or REINFORCEDINSULATION; or

N

Bonded to the PROTECTIVE CONDUCTOR TERMINAL; or N

Separated by screen or BARRIER bonded toPROTECTIVE CONDUCTOR TERMINAL from parts whichare HAZARDOUS LIVE

N

6.5.1.1 Integrity of PROTECTIVE BONDING —

6.5.1.1a) PROTECTIVE BONDING consists of directly connectedstructural parts or discrete conductors or both; andwithstands thermal and dynamic stresses

N

6.5.1.1b) Soldered connections: —

Independently secured N

Not used for other purposes N

Screw connections are secured N

6.5.1.1c) PROTECTIVE BONDING not interrupted N

6.5.1.1d) Any moveable connection specifically designed,and meets 6.5.1.3

N

6.5.1.1e) No external metal braid of cables used N

6.5.1.1f) If MAINS supply passes through: —

Means provided for passing protective conductor; N

Impedance meets 6.5.1.3. N

6.5.1.1g) Protective conductors bare or insulated,if insulated, green/yellow

Power supply cords providedby end-user.

N

Exceptions: —

1) earthing braids; N

2) internal protective conductors etc.; N

Green/yellow not used for other purposes N

6.5.1.1h) TERMINAL suitable, and meets 6.5.1.2 N

6.5.1.2 PROTECTIVE CONDUCTOR TERMINAL —

6.5.1.2a) Contact surfaces are metal N

6.5.1.2b) Appliance inlet used N

6.5.1.2c) For rewireable cords and PERMANENTLY CONNECTEDEQUIPMENT, PROTECTIVE CONDUCTOR TERMINAL isclose to MAINS supply TERMINALS

N

6.5.1.2d) If no MAINS supply is required, any PROTECTIVECONDUCTOR TERMINAL:

Page 12: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 12 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Is near TERMINALS of circuit for which protectiveearthling is necessary

N

External if other TERMINALS external N

6.5.1.2e) Equivalent current-carrying capacity to MAINSsupply TERMINALS

N

6.5.1.2f) If plug-in, makes first and breaks last N

6.5.1.2g) If also used for other bonding purposes, protectiveconductor:

Applied first; N

Secured independently; N

Unlikely to be removed by servicing; or N

Warning marking requires replacement ofprotective conductor

N

6.5.1.2h) Protective conductor of measuring circuit: N

1) Current RATING; N

2) PROTECTIVE BONDING: —

Not interrupted; or N

Indirect bonding used (see 6.5.1.5) N

6.5.1.2i) FUNCTIONAL EARTH TERMINALS allow independentconnection

N

6.5.1.2j) If a binding screw: —

Suitable size for bond wire N

Not smaller than M 4 (No. 6) N

At least 3 turns of screw engaged N

Contact pressure not capable of reduction bydeformation of materials

N

Passes tightening torque test N

6.5.1.3 Impedance of PROTECTIVE BONDING of plug-connected equipment

N

6.5.1.4 Bonding impedance of PERMANENTLY CONNECTEDEQUIPMENT

N

6.5.1.5 Indirect bonding for measuring and test equipment N

6.5.2 DOUBLE INSULATION and REINFORCED INSULATION(see 6.7, 6.8 and 6.9.2)

6.5.3 PROTECTIVE IMPEDANCE Not used N

6.5.3a) HIGH-INTEGRITY single component used (s. 14.6); or N

6.5.3b) A combination of components used; or N

6.5.3c) A combination of BASIC INSULATION and current- orvoltage-limiting device used

N

Page 13: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 13 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Components, wires and connections are RATED asrequired

N

6.5.4 Automatic disconnection of the supply Not used N

If used, it meets : —

6.5.4a) Supplied with the equipment; or N

Specified by installation instruction N

6.5.4b) RATED disconnecting time within limit specified N

6.5.4c) RATED for maximum RATED LOAD N

6.6 Connections to external circuits —

6.6.1 General —

Connections do not cause ACCESSIBLE parts of thefollowing to become HAZARDOUS LIVE in NORMALCONDITION or SINGLE FAULT CONDITION:

6.6.1a) The external circuits N

6.6.1b) The equipment As above N

Separation of circuits provided; or As above N

Short circuit of separation does not cause a Hazard As above N

Instructions or markings include: —

1) RATED conditions for TERMINAL Not required for SELV N

2) Required RATING of external circuit insulation Not required for SELV N

6.6.2 TERMINALS for external circuits —

TERMINALS which receive a charge from an internalcapacitor are not HAZARDOUS LIVE

No such terminals N

High voltage TERMINALS energized from the interiorare:

Not ACCESSIBLE if connected; or No such terminals N

Unmated HAZARDOUS LIVE TERMINALS notACCESSIBLE ; or

No such terminals N

marked with symbol 12 No such terminals N

6.6.3 Circuits with TERMINALS which are HAZARDOUS LIVE —

These circuits are: —

Not connected to ACCESSIBLE conductive parts; or No such terminals N

Connected to ACCESSIBLE conductive parts, but arenot MAINS CIRCUITS and have one TERMINAL contactat earth potential

No such terminals N

No ACCESSIBLE conductive parts are HAZARDOUSLIVE

No such terminals N

6.6.4 ACCESSIBLE TERMINALS for stranded conductors —

6.6.4a) No risk of accidental contact because: —

Located or shielded No such terminals N

Page 14: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 14 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Self-evident or marked whether connected toACCESSIBLE conductive parts

No such terminals N

6.6.4b) ACCESSIBLE TERMINALS will not work loose No such terminals N

6.7 CLEARANCES and CREEPAGE DISTANCES N

6.8 Procedure for dielectric strength tests N

6.9 Constructional requirements for protectionagainst electric shock

6.9.1 General —

If a failure could cause a HAZARD: —

6.9.1a) Security of wiring connections N

6.9.1b) Screws securing removable covers N

6.9.1c) Accidental loosening N

Easily damaged materials not used N

Non-impregnated hydroscopic materials not used N

6.9.2 ENCLOSURES of equipment with DOUBLE INSULATIONor REINFORCED INSULATION

ENCLOSURE surrounds all metal parts except forsmall metal parts which are separated

N

ENCLOSURES or parts made of insulating material N

Protection for metal ENCLOSURES or parts by: —

6.9.2a) An insulating coating or BARRIER on the inside; or N

6.9.2b) CLEARANCES and CREEPAGE DISTANCES cannot bereduced by loosening of parts or wires

N

6.9.3 Over-range indication —

Unambiguous N

6.10 Connection to MAINS supply source andconnections between parts of equipment

6.10.1 MAINS supply cords —

6.10.1a) RATED for maximum equipment current (see 5.1.3c) N

Cable complies with IEC 60227 or IEC 60245 N

6.10.1b) Heat-resistant if likely to contact hot parts N

6.10.1c) Temperature RATING (cord and inlet) N

6.10.1d) Green/yellow used only for connection toPROTECTIVE CONDUCTOR TERMINALS

N

Detachable cords with IEC 60320 MAINSconnectors:

Conform to IEC 60799; or N

Have the current RATING of the MAINS connector N

6.10.2 Fitting of non-detachable MAINS supply cords —

Page 15: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 15 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Non-detachable cord protection: —

6.10.2a) Inlet or bushing smoothly rounded; or N

6.10.2b) Insulated cord guard protruding ≥5D N

The protective earth conductor is the last to takethe strain

N

6.10.2 Cord anchorages: —

6.10.2a) Cord is not clamped by direct pressure from ascrew

N

6.10.2b) Knots are not used N

6.10.2c) Cannot push the cord into the equipment to causea hazard

N

6.10.2d) No failure of cord insulation in anchorage withmetal parts

N

6.10.2e) Compression bushing: —

1) Clamps all types and sizes of MAINS cords; and N

2) Is suitable: —

For connection to TERMINALS provided; or N

It is designed for screened MAINS cord N

6.10.2f) Cord replacement does not cause a HAZARD andmethod of strain relief is clear

N

Push-pull test N

6.10.3 Plugs and connectors —

6.10.3a) Plugs, connectors and socket-outlets for connectingequipment to the MAINS supply, includingappliance couplers used to connect detachableMAINS supply cords, shall conform to AS/NZS3112, AS/NZS 3109.1 or AS/NZS 3123 asappropriate.CONFORMITY IS CHECKED BY INSPECTION. FOR PLUGSRECEIVING A CHARGE FROM AN INTERNAL CAPACITOR,THE MEASUREMENTS OF 6.3 ARE MADE TO ESTABLISHTHAT THE LEVELS OF 6.3.1 C) ARE NOT EXCEEDED.(MAINS supply plugs, connectors etc., conform withrelevant specifications)

N

6.10.3b) If equipment supplied at voltages below 6.3.2.a) orfrom a sole source:

Plugs of supply cords do not fit MAINS socketsabove RATED supply voltage

N

MAINS-type plugs used only for connection to MAINSsupply

N

610.3c) Plug pins which receive a charge from an internalcapacitor

N

6.10.3d) Accessory MAINS socket outlets: —

Page 16: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 16 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

1) Marking if accepts a standard MAINS plug(see 5.1.3e)

N

2) Input has a protective earth conductor if outlethas earth TERMINAL contact

N

6.10.3e) Plug portion of direct plug-in equipment shallconform to Appendix J of AS/NZS 3112.Conformity is checked by tests specified inAppendix J

6.11 Disconnection from supply source —

6.11.1 General —

Disconnects all current carrying conductors N

6.11.1.1 Exceptions —

6.11.1.1a) Equipment supplied by low energy source; or N

6.11.1.1b) Equipment connected to impedance protectedsupply; or

N

6.11.1.1c) Equipment constitutes an impedance protectedload

N

6.11.2 Requirements according to type of equipment —

6.11.2.1 PERMANENTLY CONNECTED EQUIPMENT and multi-phase equipment

Not Permanently connectedand not multiphase typeequipment.

Employs switch or circuit-breaker N

If switch or circuit-breaker is not part of theequipment, documentation specifies:

6.11.2.1a) Switch or circuit-breaker to be included in buildinginstallation

N

6.11.2.1b) Location No such equipment N

6.11.2.1c) Marking No such equipment N

6.11.2.2 Single-phase cord-connected equipment —

Equipment is provided with: —

6.11.2.2a) Switch or circuit-breaker; or N

6.11.2.2b) Appliance coupler (disconnectable without TOOL); or N

6.11.2.2c) Separable plug (without locking device) N

6.11.2.3 HAZARDS arising from function —

Emergency switch No such equipment N

Emergency switch 1 m from the moving part No such equipment N

6.11.3 Disconnecting devices —

Electrically close to the supply N

6.11.3.1 Switches and circuit-breakers —

When used as disconnection device: —

Page 17: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 17 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Meets IEC 60947-1 and IEC 60947-3 N

Marked to indicate function N

Not incorporated in MAINS cord N

Does not interrupt protective earth conductor N

If has other contacts meets separationrequirements of 6.6 and 6.7

N

6.11.3.2 Appliance couplers and plugs —

Where an appliance coupler or separable plug isused as the disconnecting device (see 6.11.2.2):

Readily identifiable and easily reached by theOPERATOR

N

Single-phase PORTABLE EQUIPMENT cordlength 3 m

N

Protective earth conductor connected first anddisconnected last

N

7 PROTECTION AGAINST MECHANICAL HAZARDS No such mechanical movingparts.

7.1 General —

Conformity is checked by 7.2 to 7.6 N

7.2 Moving parts —

Moving parts not able to crush, etc.(see also 6.11.2.3)

No such moving parts N

If OPERATOR access permitted: —

7.2a) Access requires TOOL No such function N

7.2b) Statement about training No such operation N

7.2c) Warning markings or symbol 14 No such operation N

7.3 Stability —

Marking of non-automatic means N

Conformity tests: —

7.3a) 10° tilt test N

7.3b) multi-directional force test N

7.3c) downward force test N

7.4 Provisions for lifting and carrying —

Handles or grips withstand four times weight No such handles or grips N

Equipment >18 kg: —

Has means for lifting or carrying; or Less than 18 kg N

Directions in documentation Less than 18 kg N

7.5 Wall mounting —

Mounting brackets withstand four times weight No such brackets. N

Page 18: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 18 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

7.6 Expelled parts —

Equipment contains or limits the energy N

Protection not removable without the aid of a TOOL N

8 MECHANICAL RESISTANCE TO SHOCK ANDIMPACT

After the tests of 8.1 to 8.2: —

Voltage tests N

Inspections: —

8a) HAZARDOUS LIVE parts not accessible N

8b) ENCLOSURE shows no cracks (hazard) N

8c) CLEARANCES not less than their permitted values N

8d) BARRIERS not damaged or loosened N

8e) No moving parts exposed, except permitted by 7.2 N

8f) No damage which could cause spread of fire N

9 PROTECTION AGAINST THE SPREAD OF FIRE —Conformity for protection against spread of fire shallbe tested as below or by an alternative testprescribed in Annex ZA.As an exception to the above requirements, fluid-containing parts of refrigeration systems are tomeet the relevant requirements of AS/NZS60335IEC 60335.

(Conformity for each source of HAZARD or area ofthe equipment is checked by one of the following:)

9a) Fault test of 4.4; or N

9b) Application of 9.1 (eliminating or reducing thesources of ignition); or

N

9c) Application of 9.2 (containment of fire within theequipment)

N

9.1 Eliminating or reducing the sources of ignitionwithin the equipment

9.1a) 1) Limited-energy circuit (see 9.3); or Alkaline battery 9V wasregarded inherently limitedenergy source

P

2) Insulation meets the requirements for BASICINSULATION; OR

Not Applied method N

Bridging the insulation does not cause ignition Not Applied method N

9.1b) Surface temperature of liquids and parts (see 9.4.a) No such liquids N

9.1c) No ignition in circuits designed to produce heat No such heat circuits N

9.2 Containment of the fire within the equipment,should it occur

Page 19: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 19 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

9.2a) Energizing of the equipment is controlled by anOPERATOR held switch

N

9.2b) Enclosure is conform with constructionalrequirements of 9.2.1; and

Not Applied method N

Requirements of 9.4b) or c) are met Not Applied method N

9.2.1 Constructional requirements —

9.2.1a) Insulated wires have flammability classification FV1or better

N

Connectors and insulating material haveflammability classification FV2 or better

N

9.2.1b) The enclosure is constructed as follows: —

1) Bottom constructed with: —

No openings; or N

Extent as specified in figure 7; or N

Baffles as specified in figure 6; or Not Applied N

Perforated as specified in Table 12; or Not Applied N

Metal screen with a mesh Not Applied N

2) Sides have no openings as specified in figure 7 N

3) Material of ENCLOSURE and any baffle or flamebarrier is made of:

Metal (except magnesium); or N

Non metallic materials have flammabilityclassification FV1 or better

N

4) ENCLOSURE and any baffle or flame barrier haveadequate rigidity

N

9.3 Limited-energy circuit —

9.3a) Potential not more than 30 r.m.s. and 42.4 V peak,or 60 V dc

N

9.3b) Current limited by one of following means: —

1) Inherently or by impedance; or Not Applied method N

2) Overcurrent protective device; or Not Applied method N

3) A regulating network limits also in SINGLE FAULTCONDITION

N

9.3c) Is separated by at least BASIC INSULATION Not Applied method N

If overcurrent protective device used: —

Fuse or a non adjustable electromechanical device Not Applied method N

9.4 Requirements for equipment containing or usingflammable liquids

No such liquids N

Flammable liquids contained in or specified for usewith equipment do not cause spread of fire

No such liquids N

Page 20: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 20 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Risk is reduced to a tolerable level : No such liquids —

9.4a) The temperature of surface or parts in contact withflammable liquids is 25 °C below fire point

No such liquids N

9.4b) The quantity of liquid is limited No such liquids N

9.4c) Flames are contained within the equipment No such liquids N

Detailed instructions for risk-reduction provided No such liquids N

9.5 Overcurrent protection N

Devices not in the protective conductor N

Fuses or single-pole circuit-breakers not fittedin neutral (multi-phase)

N

9.5.1 PERMANENTLY CONNECTED EQUIPMENT Not this type equipment N

Overcurrent device: Not this type equipment —

Fitted within the equipment; or Not this type equipment N

Specified in manufacturer's instructions Not this type equipment N

9.5.2 Other equipment N

Protection within the equipment N

10 EQUIPMENT TEMPERATURE LIMITS ANDRESISTANCE TO HEAT

10.1 Surface temperature limits for protection againstburns

Easily touched surfaces within the limits No such hot surfaces N

Heated surfaces necessary for functional reasonsexceeding specified values:

Are recognizable as such by appearance orfunction; or

N

Are marked with symbol 13 N

Guards are not removable without TOOL N

10.2 Temperatures of windings No excessive temperatures N

Limits not exceeded in: —

NORMAL CONDITION N

SINGLE FAULT CONDITION N

10.3 Other temperature measurements N

Following measurements conducted if applicable: —

10.3a) Value of 60 °C of field-wiring TERMINAL box notexceeded

Not applicable test N

10.3b) Surface of flammable liquids and parts in contactwith this liquids

Not applicable test N

10.3c) Surface of non-metallic ENCLOSURES N

10.3d) Parts made of insulating material supporting partsconnected to MAINS supply

N

Page 21: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 21 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

10.3e) TERMINALS carrying a current more than 0.5 A N

10.4 Conduct of temperature test N

10.5 Resistance to heat N

10.5.1 Integrity of CLEARANCE and CREEPAGEDISTANCES

N

10.5.2 Non-metallic ENCLOSURES N

After treatment: N

No HAZARDOUS LIVE parts ACCESSIBLE; N

Tests of 8.1 and 8.2 N

In case of doubt, tests of 6.8(without humidity preconditioning)

N

10.5.3 Insulating material N

10.5.3a) Parts supporting parts connected to MAINS supply N

10.5.3b) TERMINALS carrying a current more than 0.5 A N

Examination of material data; or Not required N

in case of doubt:: —

1) Ball pressure test; or Not required N

2) Vicat softening test of ISO 306 Not required N

11 PROTECTION AGAINST HAZARDS FROMFLUIDS

No such fluids —

11.1 General N

11.2 Cleaning N

11.3 Spillage N

11.4 Overflow N

11.5 Battery electrolyte —

Battery electrolyte leakage presents no hazard N

11.6 Specially protected equipment N

11.7 Fluid pressure and leakage —

11.7.1 Maximum pressure —

Maximum pressure of any part does not exceedPRATED

N

11.7.2 Leakage and rupture at high pressure N

Test to IEC 60335 (refrigeration only) N

11.7.3 Leakage from low-pressure parts N

11.7.4 Overpressure safety device —

Does not operate in NORMAL USE N

Meets ISO 4126-1; and N

Page 22: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 22 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

It is conform with: —

11.7.4a) Connected as close as possible to parts intended tobe protected

N

11.7.4b) Easy access for inspection, maintenance and repair N

11.7.4c) Adjustment only with TOOL N

11.7.4d) No discharge towards person N

11.7.4e) No HAZARD from deposit of discharged material N

11.7.4f) Adequate discharge capacity N

11.7.4g) No shut-off valve between overpressure safetydevice and protected parts

N

12 PROTECTION AGAINST RADIATION,INCLUDING LASER SOURCES, AND AGAINSTSONIC AND ULTRASONIC PRESSURE

No such radiation or sonicsources.

12.1 General —

Equipment provides protection N

12.2 Equipment producing ionising radiation N

12.2.1 Ionising radiation N

12.2.2 Accelerated electrons N

12.3 Ultra-violet (UV) radiation —

No unintentional and HAZARDOUS escape of UVradiation

N

12.4 Micro-wave radiation —

Power density does not exceed 10 W/m2: N

12.5 Sonic and ultrasonic pressure —

12.5.1 Sound level N

12.5.2 Ultrasonic pressure N

12.6 Laser sourcesEquipment employing laser sources shall meet therequirements of IEC 60825-1.Conformity is checked as specified in IEC 60825-1.NOTE – For the purpose of this Standard,compliance with AS/NZS 2211.1 is deemedcompliant with IEC 60825-1.

(Laser sources (IEC 60825-1))

N

13 PROTECTION AGAINST LIBERATED GASES,EXPLOSION AND IMPLOSION

13.1 Poisonous and injurious gases N

Attached data/test reports demonstrate conformity N

13.2 Explosion and implosion —

13.2.1 Components —

Page 23: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 23 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

Components liable to explode: —

Pressure release device provided; or No explosion hazard N

Apparatus incorporates OPERATORprotection (see also 7.6)

N

Pressure release device: —

Discharge without danger No such device N

Cannot be obstructed N

13.2.2 Batteries and battery charging —

If explosion or fire hazard could occur: —

Protection incorporated in the equipment; or Alkaline battery type 9-Volt6LR61 was regardedinherently safe

N

Instructions specify batteries with built-in protection N

In case of wrong type of battery used: —

No HAZARD; or N

Warning by marking and within instructions P

Equipment with means to charge rechargeablebatteries:

Warning against the charging of non-rechargeablebatteries; and

N

Type of rechargeable battery indicated; or N

Symbol 14 used N

Battery compartment design N

Single component failure N

Polarity reversal test N

13.2.3 Implosion of cathode ray tubes —

If maximum face dimensions > 160 mm: ............... —

Intrinsically protected and correctly mounted; or No such device N

ENCLOSURE provides protection: N

If non-intrinsically protected: —

Screen not removable without TOOL N

If glass screen, not in contact with surface of tube N

13.2.4 Equipment RATED for high pressure (See 11.7) N

14 COMPONENTS

14.1 General

Page 24: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 24 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

If safety is involved, components shall be used inaccordance with their specified RATINGS unless aspecific exception is made. They shall conform toone of the following:

The requirements of this standard and, wherenecessary for the application, any additionalapplicable safety requirements of the relevant IECcomponent standard; or the relevant AustralianStandard

(Where safety is involved, components meetrelevant requirements)

N

14.2 Motors —

14.2.1 Motor temperatures —

Does not present a HAZARD when stopped orprevented form starting; or

No such devices. N

Protected by overtemperature or thermal protectiondevice conform with 14.3

N

14.2.2 Series excitation motors —

Connected direct to device, if overspeeding causesa HAZARD

No such devices. N

14.3 Overtemperature protection devices N

Devices operating in a SINGLE FAULT CONDITION N

14.3a) Reliable function is ensured N

14.3b) RATED to interrupt maximum current and voltage N

14.3c) Does not operate in NORMAL USE N

14.4 Fuse holders N

No access to HAZARDOUS LIVE parts N

14.5 Mains voltage selecting devices N

Accidental change not possible N

14.6 HIGH INTEGRITY components N

Used in applicable positions (see Table 3) N

Conforms with IEC publications N

Single electronic device not used N

14.7 Mains transformers tested outside equipment N

14.8 Printed circuit boards P

Data shows conformity with FV-1 of IEC 60707 orbetter; or

Critical Printed circuit boardsrated FV-1 min.

P

Test shows conformity with FV-1 of IEC 60707 orbetter; or

N

Thin film flexible PCB with limited-energy circuitused

N

Page 25: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 25 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDEITL086 V3.0 12/09/2004

14.9 Circuits or components used as transientovervoltage limiting devices

After test, no sign of overload or degradation No such devices. N

15 PROTECTION BY INTERLOCKS —

15.1 General —

Interlocks are designed to remove a hazardbefore OPERATOR exposed

N

15.2 Prevention of reactivation N

15.3 Reliability —

Single fault unlikely to occur; or N

Cannot cause a HAZARD N

16 TEST AND MEASUREMENT EQUIPMENT No such measuringequipment.

N

16.1 Current measuring circuits N

16.2 Multifunction meters and similar equipment N

No HAZARD from: —

RATED input voltage combinations N

Settings of functions N

Settings of range controls N

ANNEX F ROUTINE TESTS N

Manufacturer’s declaration N

Page 26: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 26 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDE

4.4.2 TABLE: Summary of SINGLE FAULT CONDITIONS Form A.1 N

Sub clause Title Does notapply

Carriedout

Comments

4.4.2.1 PROTECTIVE IMPEDANCE X

4.4.2.2 Protective conductor X

4.4.2.3 Equipment or parts for short-term orintermittent operation

X

4.4.2.4 Motors X

4.4.2.5 Capacitors X

4.4.2.6 Mains transformersAttach drawing of MAINS Txs showing allprotective devices(See Appendix 1 for Test Results).

X

4.4.2.7 Outputs X

4.4.2.8 Equipment for more than one supply X

4.4.2.9 Cooling X– air holes closed X– fans stopped X– coolant stopped X

4.4.2.10 Heating devices X– timer overridden X– temperature controller overridden X– loss of cooling liquid X– overfilled or empty or both X

4.4.2.11 Insulation between circuits and parts X

4.4.2.12 Interlocks X

List below all SINGLE FAULT CONDITIONS not covered by 4.4.2.1 to 4.4.2.12:

Supplementary information:(see Form A.2 for details of tests)

Page 27: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 27 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_C TRF originator: VDE

4.4 TABLE: Testing in single FAULT CONDITION – Results Form A.2 N

Testsub

clause

FaultNo.

Fault description Td 4.4.3(NOTE)

How was testterminatedComments

Meets4.4.4

NOTE Td = Test duration in h:min:sRecord dielectric strength test on Form A.14 and temperature tests on Form A.20.Record in the comments column for each test whether carried out during or after SINGLE FAULT CONDITION.

Page 28: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 28 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

5.1.3c) TABLE: Mains supply Form A.3 N

Marked rating........................: V —

Phase .....................................: —

Frequency ............................: Hz —

Current .................................: A —

Power ..................................: W —

Power ..................................: VA —

Test Voltage Frequency Current Power in Power in Comments

No. V Hz A W VA

Note: Measurements are only required for marked ratings.

Supplementary information:

Page 29: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 29 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

5.3 TABLE: Durability of markings Form A.4 N

Marking method (see NOTE) Agent

1) A Water

2) B Isopropyl alcohol

3) C (specify agent)

4) D (specify agent)

5) E (specify agent)

NOTE – Where applicable include print method, label material, ink or paint type,fixing method, adhesive and surface to which marking is fixed.

Marking location Marking method (see above)

Identification (5.1.2) -Pass

N - Mains supply (5.1.3)

N - Fuses (5.1.4)

N - TERMINALS and operating devices (5.1.5.1)

N - Measuring circuit TERMINALS (5.1.5.2)

N - Switches and circuit breakers (5.1.6)

N - DOUBLE/REINFORCED equipment (5.1.7)

N - Field wiring TERMINAL boxes (5.1.8)

Warning marking (5.2) -Pass

N - Battery charging (13.2.2)

Method Test agent Remains legible Label loose Curled edges Comments

Verdict Verdict Verdict

Page 30: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 30 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6 TABLE: Protection against electric shock - Form A.5 N

Pollution degree ...... : Installation category (overvoltage category) ...... :

Location or Insulationtype

Maximumworking

CREEPAGE DISTANCE(NOTE 3)

CLEARANCE(NOTE 3)

Testvoltage

Comments

description (NOTE 1) voltage(NOTE 2)

PWBmm

CTI Othermm

CTI mm (NOTE 2)V

NOTE 1 – Type of insulation: NOTE 2 - Types of voltage NOTE 3 - INSTALLATION CATEGORIES (OVERVOLTAGE CATEGORIES)BI = BASIC INSULATION Peak impulse test voltage (pulse) or POLLUTION DEGREES which differ fromDI = DOUBLE INSULATION r.m.s. these should be shown under "Comments".PI = PROTECTIVE IMPEDANCE d.c.RI = Reinforced INSULATION peakSI = Supplementary INSULATION

Supplementary Information:

Page 31: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 31 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.2 TABLE: List of ACCESSIBLE parts Form A.6 N

6.1.2 Exceptions —

6.2 Determination of accessible parts —

Item Description Determination method(NOTE 5)

Exception under 6.1.2(NOTE 4)

NOTE 1 – Test fingers and pins are to be applied without force unless a force is specified (see 6.2.1)NOTE 2 – Special consideration should be given to inadequate insulation and high voltage parts (see 6.2)NOTE 3 – Parts are considered to be ACCESSIBLE if they could be touched in the absence of any covering which is not considered to

provide suitable insulation (see note to paragraph 1 of 6.4).NOTE 4 – Capacitor test may be required (see Form A.7).NOTE 5 – The determination methods are: visual; rigid test finger; jointed test finger; pin 3 mm diameter; pin 4 mm diameter.

Supplementary information

Page 32: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 32 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6 TABLE: Values in NORMAL CONDITION Form A.7 N

6.1.1 Exceptions 11.2 Cleaning and decontamination —6.3.1 Values in NORMAL CONDITION (see NOTE 1) 11.3 Spillage —6.6.2 Terminals for external circuit 11.4 Overflow —6.10.3 Plugs and connections —

Item Voltage Current Capacitance 10 s test (NOTE 2) Comments

(seeFormA.6)

Vr.m.s.

Vpeak

Vd.c.

Testcircuit

A1/ A2/A3

mAr.m.s.

mApeak

mAd.c. C mJ V C mJ

NOTE 1 – The requirements of 6.3.1 include drying out (if specified). For permanently connected equipment, the current values are 1,5times the specified values.NOTE 2 – A 5 s test is specified in 6.10.3c).

Page 33: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 33 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.3.2 TABLE: Values in SINGLE FAULT CONDITION Form A.8 N

Item Sub clauseand

Voltage Transient(see NOTE)

Current Capacitance

(SeeFormA.6)

fault No.(see

FormA.2)

Vr.m.s.

Vpeak

Vd.c. V s

Testcircuit

A1/ A2/A3

mAr.m.s

mApeak

mAd.c.

F(NOTE)

Comments

NOTE – Transient voltages must be below the limits given from Figure 1 and the capacitance below the limits from figure 2 of IEC 61010-1.

Page 34: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 34 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.5.1.1 TABLE: Cross-sectional area of bonding conductors Form A.9 N

Conductor location Cross-sectional areamm2

Verdict

6.5.1.2 TABLE: Tighting torque test

Conductor location Size of Screw Tighting torqueNm

Verdict

Page 35: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 35 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.5.1.3 TABLE: Bonding impedance of plug connected equipment Form A.10 N

ACCESSIBLE part under testTest

currentA

Voltage attainedafter 1 min

V

Calculated resistance(maximum allowed 0,1 )

Verdict

Supplementary information:

6.5.1.4 TABLE: Bonding impedance of PERMANENTLY CONNECTED EQUIPMENT N

ACCESSIBLE part under testTest

currentA

Voltage attained after 1 min(maximum 10 V)

V

Verdict

Supplementary information:

Page 36: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 36 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.5.1.5 TABLE: Indirect bonding for measuring and test equipment Form A.11 N

ACCESSIBLE part under test Voltageattained

s

Time for voltage to drop toallowable levels

s

Verdict

a) Voltage limiting device — — —

Supplementary Information:

ACCESSIBLE part under test Voltageapplied

V

Time for device to trip

s

Verdict

b) Voltage-sensitive tripping device

Supplementary Information:

Page 37: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 37 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.5.3 TABLE: PROTECTIVE IMPEDANCE Form A.12 N

A high INTEGRITY single component

Component Location Comments

A combination of components

Component Location Comments

A combination of BASIC INSULATION and a current or voltage limiting device

Component Location Comments

Supplementary information:

Page 38: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 38 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.7 TABLE: CLEARANCES and CREEPAGE DISTANCES Form A.13 N

8 Mechanical resistance to shock and impact10.5.1 Integrity of CLEARANCES and CREEPAGE DISTANCES

Location Measured(initial – 6.7) Verdict Mechanical tests (note) Test at

max.Measured after test

(if required) Verdict

(see FormA.5)

CREEPAGEDISTANCE

CLEARANCE Appliedforce

Rigidity(8.1)

Drop(8.2)

RATEDambient

CREEPAGEDISTANCE

CLEARANCE Comments

mm mm (6.7) N StaticDynamic Normal Hand-held/Plug-in

(10.5.1) mm mm

NOTE – Refer to Form A.12 for dielectric strength tests following the above tests.

Page 39: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 39 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.8 TABLE: Dielectric strength tests Form A.14 N

4.4.4.1 b) Conformity after application of fault conditions1

6.4 Protection in NORMAL CONDITION

6.5.2 DOUBLE INSULATION and REINFORCED INSULATION

6.6.1 Connections to external circuits6.7.3.1 c) CLEARANCE values – General: reduced CLEARANCES for homogeneous construction6.10.2.5) Fitting of non-detachable MAINS SUPPLY cords1

8 Mechanical resistance to shock and impact9.1 a) 2) Eliminating or reducing the sources of ignition within the equipment9.3 c) Limited-energy circuit11.2 Cleaning1

11.3 Spillage1

11.4 Overflow1

11.6 Specially protected equipment1

1 Record the fault, test or treatment applied before the dielectric strength test

Test site altitude ................................................: m —

Test voltage correction factor (see Table 10)....: —

Location orreferences from

Forms A.2 and A.5

Clause orsub-clause

Humidity

Yes/No

Workingvoltage

V

Test voltager.m.s./peak/d.c

V

Comments Verdict

Supplementary information:

Page 40: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 40 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

6.10.2 TABLE: Cord anchorage Form A.15 N

Location Masskg

PullN

Verdict TorqueNm

Verdict Comment

Supplementary information:

Page 41: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 41 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

9 TABLE: Protection against the spread of fire Form A.16 N

ItemSource of HAZARD or area of theequipment considered (circuit,

component, liquid etc.)

ProtectionMethod

(9a, 9b or 9c)Protection details Verdict

Supplementary information:

Page 42: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 42 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

9.2.1 TABLE: Constructional requirements Form A.17 N

14.8 Printed circuit boards

Material tested....................................................... : —

Generic name........................................................ : —

Material manufacturer ........................................... : —

Type ...................................................................... : —

Colour.................................................................... : —

Conditioning details............................................... : —

Sample 1 Sample 2 Sample 3

Thickness of specimen mm

Duration of flaming after first Application s

Duration of flaming plus glowingAfter second application

s

Specimen burns to holding clamp Yes/No

Cotton ignited Yes/No

Sample result Pass/Fail

Supplementary information:

Page 43: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 43 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

9.3 TABLE: Limited-energy circuit Form A.18 N

Item 9.3 a) 9.3 b) Current and power limitation 9.3 c) Decision

orLocation

(see FormA.16)

Maximum potential incircuit voltage

r.m.s./ d.c.V

Maximumavailablecurrent

A

Maximumavailable

powerVA

Overloadprotectionafter 120 s

A

Circuitseparation Yes/No Comments

Supplementary information:

Page 44: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 44 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

9.4 TABLE: Requirements for equipment containing or using flammable liquids Form A.19N

Type of liquid 9.4 Flammable liquids Verdict

b) quantity c) Containment

Supplementary information:

Page 45: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 45 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

10. TABLE : Temperature Measurements Form A.20A P

10.1 Surface temperature limits - NORMAL CONDITION and / or SIGNLE FAULT CONDITION

10.2 Temperature of windings- NORMAL CONDITION and / or SIGNLE FAULT CONDITION

10.3 Other temperature measurements

Operating conditions: Model 7001D – Normal conditions

Frequency ............... : DC Hz Test room ambient temperature (ta) .....: 23.2 °C

Voltage .................... : 9 V Test duration ........................................: 2.5 h min

Part / Location tmC

tcC

tmax

CVerdict Comments

Display 23.2 41.2 - Pass 23.2-22+40

battery compartment 23.4 41.4 - Pass 23.4-22+40

Capacitor C10 23.3 41.3 - Pass 23.3-22+40

Capacitor C9 23.3 41.3 - Pass 23.3-22+40

NOTE 1 - tm = measured temperaturetc = tm corrected (tm–ta+ 40 C or max. RATED

ambient)tmax = maximum permitted temperature

NOTE 2 - See also 14.1 with reference to component operating conditionsNOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessaryNOTE 4 - See Form A.20B for details of winding temperature measurements

Supplementary information:

Page 46: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 46 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

10. TABLE : Temperature Measurements Form A.20A P

10.1 Surface temperature limits - NORMAL CONDITION and / or SIGNLE FAULT CONDITION

10.2 Temperature of windings- NORMAL CONDITION and / or SIGNLE FAULT CONDITION

10.3 Other temperature measurements

Operating conditions: Model 7001DS – Normal conditions

Frequency ............... : DC Hz Test room ambient temperature (ta) .....: 25.3 °C

Voltage .................... : 9 V Test duration ........................................: 2.5 h min

Part / Location tmC

tcC

tmax

CVerdict Comments

Display 25.3 41.3 - Pass 25.3-24+40

battery compartment 25.5 41.5 - Pass 25.5-24+40

Capacitor C10 25.2 41.2 - Pass 25.2-24+40

Capacitor C9 25.2 41.2 - Pass 25.2-24+40

NOTE 1 - tm = measured temperaturetc = tm corrected (tm–ta+ 40 C or max. RATED

ambient)tmax = maximum permitted temperature

NOTE 2 - See also 14.1 with reference to component operating conditionsNOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessaryNOTE 4 - See Form A.20B for details of winding temperature measurements

Supplementary information:

Page 47: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 47 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

10.2 TABLE: Temperature of windingsResistance method Temperature Measurements

Form A.20B N

4.4.2.6 MAINS Transformers

14.2.1 Motor temperatures

Operating conditions:

Frequency ............... : Hz Test room ambient temperature (ta1/ta2) ... : / °C (initial / final)

Voltage .................... : V Test duration ............................................ : h min

Part / Designation Rcold

Rwarm

CurrentA

trK

tcC

tmaxC

Verdict Comments

NOTE 1- Rcold = initial resistance Rwarm = final resistancetr = temperature rise tc = tr corrected (tc= tr - { ta2 - ta1} + [40 C or max RATED ambient])tmax = maximum permitted temperature

NOTE 2 - Indicate insulation class (IEC 85) under comments (optional)NOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessary

Supplementary information:

Page 48: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 48 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

10.5.2 TABLE: Resistance to heat of non-metallic enclosures Form A.21 N

Test method used: —

Non operative treatment .........................................: [ ]

Empty ENCLOSURE...................................................: [ ]

Operative treatment................................................: [ ]

Temperature during tests .......................................: —

ENCLOSURE samples tested were ...........................: —

Description Material Comments Verdict

Dielectric strength test (6.8)....................................: V r.m.s./peak/d.c.

Supplementary information

Page 49: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 49 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

10.5.3 TABLE: Insulating Materials Form A.22 N

10.5.3a) Ballpressure test

Max. allowed impression diameter .........................: 2 mm —

Part Test temperature°C

Impression Diameter(mm)

Verdict

Supplementary information:

10.5.3b) Vicat softening test (ISO 306)

Part Vicat softening temperature°C

Thickness of sample(mm)

Verdict

Supplementary information:

Page 50: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 50 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

8 TABLE: Mechanical resistance to shock and impact Form A.23 N

11 Protection against hazards from fluids NVoltage tests can be carried out once after performing the tests of clause 8 and clause 11. However, if voltage tests are carried outseparately after each set of tests, two forms can be used.

Clause 8 tests Clause 11 tests

Location(see form

A.5)

Static Dynamic Normal HandheldPlug-in

Cleaning(11.2)

Spillage(11.3)

Overflow(11.4)

IEC 60529(11.6)

Workingvoltage

V

Testvoltage

V

Verdict Comments

NOTE – Use r.m.s., d.c. or peak to indicate the used test voltage.

Page 51: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 51 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

11.7.2 TABLE: Leakage and rupture at high pressure Form A.24 N

PartMaximum

permissibleworkingpressure

MPa

Testpressure

MPa

Leakage

YES / NO

Burst

YES / NO

Comments

Supplementary information:

11.7.3 Leakage from low-pressure parts

PartTest

pressureMPa

Leakage

YES / NOComments

Supplementary information:

Page 52: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 52 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

12.2.1 TABLE: Ionising radiation Form A 25 N

Locations testedMeasured values

µSv/hVerdict

Comments

Supplementary information:

Page 53: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 53 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

12.5.1 TABLE: Sound level Form A.26 N

Locations tested Measured valuesdBA

Calculated maximum soundpressure level

At operator’s normal positionand at bystanders’ positions

a)

b)

c)

d)

e)

Supplementary information:

12.5.2 Ultrasonic pressure

Locations tested Measured values Comments

dB kHz

At OPERATOR’S normalposition

At 1 m from the ENCLOSURE

a)

b)

c)

d)

e)

NOTE – No limit is specified at present, but a limit of 110 dB above the reference pressure value of 20 Pa is under consideration forapplicable frequencies between 20 kHz and 100 kHz.

Supplementary information:

Page 54: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 54 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

13.2.2 TABLE: Batteries.Remark: Only rapid discharge test conducted asbattery has no charge across it.

Form A.27 N

Battery load and charging circuit diagram:

Battery type.............................................................: —

Battery manufacturer/model/catalogue No.............: —

Battery ratings.........................................................: —

Reverse polarity instalment test

Single component failures Verdict

Component Open circuit Short circuit

Supplementary information:

Page 55: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 55 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

14.3 TABLE: Overtemperature protection devices Form A.28 N

Reliability test

Component Type(note)

Verdict Comments

NOTE:NSR = non-self-resetting (10 times)NR = non-resetting (1 time)SR = self-resetting (200 times)

Supplementary information:

Page 56: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 56 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

4.4.2.6 TABLE: Mains transformer Form A.29 N

4.4.2.6.1 Short circuit

14.7.1 MAINS transformers tested outside equipment

Type ............................: —

Manufacturer ...............: —

Test in equipment

Test on bench

Test repeated inside equipment (see 14.7)

Optional – Insulation class (IEC 60085) of the lowest RATED winding ............ : —

Winding identification

Type of Protector for winding (Note 1)

Elapsed time

Current, A primarysecondary

Winding temperature, °C primary(see Note 2) secondary

Tissue paper / cheesecloth OK ?(Pass / Fail)

Voltage tests (see Note 3)

primary to secondary ______ V ____

primary to core ______ V ____

secondary to secondary ______ V ____

secondary to core ______ V ____

VerdictNote 1: Primary fuse - PF / (

)

A

Secondary fuse - SF / (

)

A

Overtemperature protection - OP / (

)

°C

Impedance protection - ZNote 2: Indicate method of measurement TC = with thermocouple R = resistance method

If resistance method is used,record resistance in cold and warm condition in FormA.20B!Note 3: Record the voltage applied and the type of voltage (r.m.s. / d.c. / peak) and for

results use NB = no breakdown or B = breakdown

Supplementary information:

Page 57: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 57 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

4.4.2.6 TABLE: Mains transformer Form A.30 N

14.7.2 Overload tests (for mains transformers)

Type ............................: —

Manufacturer ...............: —

Test in equipment

Test on bench

Test repeated inside equipment (see 14.7)

Optional – Insulation class (IEC 60085) of the lowest RATED winding ............ : —

Winding identification

Type of Protector for winding (Note 1)

Elapsed time

Current, A primarysecondary

Winding temperature, °C primary(see Note 2) secondary

Tissue paper / cheesecloth OK ?(Pass / Fail)

Voltage tests (see Note 3)

primary to secondary ______ V ____

primary to core ______ V ____

secondary to secondary ______ V ____

secondary to core ______ V ____

VerdictNote 1: Primary fuse - PF / (

)

A

Secondary fuse - SF / (

)

A

Overtemperature protection - OP / (

)

°C

Impedance protection - ZNote 2: Indicate method of measurement TC = with thermocouple R = resistance method

If resistance method is used,record resistance in cold and warm condition in FormA.20B!Note 3: Record the voltage applied and the type of voltage (r.m.s. / d.c. / peak) and for

results use NB = no breakdown or B = breakdown

Supplementary information:

Page 58: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 58 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

16.1 TABLE: Current measuring circuits Form A.31 N

These tests are performed with all types and models of current transformers without internal protection,and which are specified by the manufacturer for use with the equipment

a) Current transformers

Type/Model RATED currentA

Test currentA

InterruptYes / No

Verdict Comments

Supplementary information:

b) Range changing switches

Type / Model Maximum rated currentof switch

A

Cycling testVerdict

Comments

Supplementary information:

Page 59: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 59 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

16.2 TABLE: Multifunctional meters and similar equipment Form A. 32 N

Operating conditions .............................................. : —

Maximum RATED voltage applied (V)...................... : —

Measurement category........................................... : —

Test source limit (KVA)........................................... : —

Function Range Verdict

Supplementary information:

Page 60: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 60 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

PHOTOGRAPHS

Photograph No. 17001D - Top side view with open panel

Page 61: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 61 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

Photograph No. 27001D – Bottom View with battery compartment open

Page 62: ITL Project Ref.: S101851.02 Page 1 of 62galconc.com/.../06/Galcon-7001-Testing-Compl-Ltr.pdf · ITL Project Ref.: S101851.02 Page 3 of 62 IEC/EN 61010-1 TRF No.: IEC 61010_C ITL086

ITL Project Ref.:S101851.02 Page 62 of 62

IEC/EN 61010-1

TRF No.: IEC 61010_CITL086 V3.0 12/09/2004 TRF originator: VDE

Photograph No. 37001D – Internal View

END OF TEST REPORT