45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
R. J. Peláez, J. L. Doménech, V. J. Herrero, B. Alemán, C. Bermúdez, S. I. Ramírez,E. Moreno, K. Lauwaet, G. Santoro,
J. A. Martín‐Gago, J. R. Pardo, J. Cernicharo, F. Beltrán, J. M. HernándezP. de Vicente, J. D. Gallego and the Team of Yebes Observatory
Funding: Spanish grants CSD2009‐00038 “ASTROMOL”, FIS2016‐77726‐C3‐1‐PEU grant ERC‐2013‐Syg 610256 “NANOCOSMOS”
[email protected], IEM‐CSIC, Spain
2
• Interstellar molecular clouds are the place of star formation and molecules play a key role in this process.
• Molecules cool efficiently the gas heated by gravity collapse and reduce the ionization degree much more efficiently than atoms, contributing to decouple the gas from electromagnetic fields.
• Molecular species are the product of a rich chemistry, and their study allows to find out relevant properties of these regions.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
INTRODUCTION
• Since the discovery of CH and CN radicals in ISM (1940), more than 200 molecular species have been found by spectroscopic methods, containing up to 70 atoms.
• 80% have been detected with radio astronomy, that is,mm wavelengths, by their pure rotational transitions.
• Valid for NON symmetric species, with dipole moment ≠ 0.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
3
• Radio astronomy is very useful for the characterization of very cold IS regions, where only molecular levels very close to the ground state are populated.
• It provides very high spectral resolution.
• It is particularly convenient to study molecular clouds, as comparedwith visible and UV, due to the different opacities of the media to be crossed by the radiation.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
4
Dark molecular clouds are composed mainly of H2 and Dust(15% mass of galaxies)
UV and visible ( large opacity )
FIR and Radio( good transparency )
Dark molecular cloud Barnard 68 , T 16 K
Unlike UV and visible light, scattered or absorbed by interstellar dust, longer wavelengths can pass through the microscopic particles.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
5
Earth’s Atmosphere presents high transmittance for radio waves, which increases markedly with increasing altitude and dryness
Wavelength
Atmosph
ericTran
smission
300 250 200 150 100 50
Frequency (GHZ)
1.0
0.8
0.6
0.4
0.2
0.0
South PoleAtacama desert, Chile (very dry)Hanoi, Vietnam(very humid)
O2 O2
H2O
O3
6
Nowadays, the most recent generation of radio‐telescopes, like ALMA, provides more sensitivity and angular resolution than all the previous
ones to search molecules in space.
ALMA (Atacama Large Mm/sub‐mm Array)Atacama Desert (Chile), 2400 m altitude
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
7
PRESENT WORK: USE OF RADIO ASTRONOMY TECHNIQUES
TO DETECT ROTATIONAL EMISSIONS IN LABORATORYCOLD PLASMAS AT ROOM TEMPERATURES.
40 m Radio Telescope of Yebes Observatory, Guadalajara, Spain
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
8
102 103 104 105 106 107 108 109
10-2 10-1 100 101 102 103 104 1051010
1015
1020
1025
1030
1010
1015
1020
1025
1030
Ne (
m- 3
)
E (eV)
T (K)
Fusion reactor core
Flame
Laser focus
SolidsLiquids
Interplanetaryspace
Solar corona
Solar photosphere
Solar core
Jupiter core
Si crystal
Lightning
Fusion plasma edge
ArcGlow
Discharges
PLASMA
ColdPlasmas
IonospheresAuroras
IntestellarMedia (ISM)
9
Glow Discharge ReactorsLength 100 cm
Density 2.5 x 1015 cm‐3 (0.1 mbar)
Column density in line of sight:Stable species 10 17 cm‐2
Radicals (1%) 1015 cm‐2
Ions (0.01%) 1013 cm‐2
Molecular CloudsLength 1019 cm
Density 104‐105 cm‐3
Column density in line of sight
H2 1022 cm‐2
CO 1018 cm‐2
HCN, HCO+, C2H, CN, CS 1014 cm‐2
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
10
BASIC EXPERIMENTAL SCHEME
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
13.56 MHz Generator
+ MatchingREACTOR
Hot Load300 K
Diff. PumpQuadr.
Mass Spec.
Gas Input
COLD LOAD
Gas Output
UV & visibleEmission
Coil
PLASMAEMISSION
BackgroundEmission
• Versatile Reactor devoted to study Gas Phase, Plasma Chemistry in ICP discharges & Photochemistrywith UV lamps.
• HEMT Heterodine Receivers for broad band, high resolution, rotational emission spectroscopy.
• Cold Load at the backend to minimize mm background radiation (+ movable Hot Load for calibration).
Additional equipment:
• Quadrupole Mass Spectrometry • Visible Emission Spectroscopy
11
Very weak signalP 10‐15 W, F 30‐300 GHz
HeterodyneDetectionAmplification
Frequency Down ConversionProcessing by FFTS
Gain 120 dB IF 2 GHz
RadioReceiver
Low NoiseAmplifier Mixer IF
Amplifier
LocalOscillators( 1 … n )
IFoutput
IF FilterFast Fourier TransformSpectrometer
15 K, He Cryostat
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Careful noise reduction is essential
12
TWO CAMPAIGNS OF EXPERIMENTS TWO INDEPENDENT EQUIPMENTS
1.‐ Proof of concept in the 40 m Radio Telescope of YebesObservatory (Spain) (2016/2017).
2.‐ New set‐up devoted to Laboratory Astrophysics in the Yebes Observatory location (2018).
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
13
1. PROOF OF CONCEPT IN THE 40 m RADIO TELESCOPE OF YEBES OBSERVATORY
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
100 steps
14
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Prototype Reactor
• Designed to be used in the 40 m telescope • Dimensions: 40 cm length x 25 cm diameter• Gas flow or static conditions 10‐2 ‐ 10‐1 mbar• RF discharge: 13.56 MHz, 5‐60 W• Windows: 9.6 mm thickness Fused Silica or 70 m Upilex polyimide film, non parallel between them.
Plasma Chemistry
Photochemistry
15
HEMT Heterodine Q band Receiverused in the 40 m Radio Telescope
• Spectral Range = 30‐50 GHz • Instantaneous Bandwidh = 2 GHz• 2 FFTS (65536 channels/u) Resolution = 38 kHz
Two Sampling Modes for Noise Reduction• ON/OFF technique (sample in/sample out) the common one in Astronomy
• Frequency Switching ±24 MHz (for plasma experiments)
Reactor
Mirrors System
Receiver
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
300 250 200 150 100 50
Frequency (GHZ)
Atmosph
ericTran
smission
1.0
0.8
0.6
0.4
0.2
0.0South Pole, Atacama, ChileHanoi, Vietnam
Q band 16
Cold Load
• 40 m Antenna towards the Zenith (89o)(clear blue sky) TB = 42 K at 45 GHz
• Load of liquid N2 (cloudy or rainy weather) TB = 77 K
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Mirrors
N2 Load
RemovableMirror
Reactor
Mirrors
Signal fromAntenna
17
70 m UpilexWindows
Implosion Risk!
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Tests of interference effects of the windows on the background signal
70 m Upilex
9.6 mm thickness Fused Silica
• 9.6 mm Fused Silica: similar thickness to Q band wavelengths (6‐9 mm).• 70 m Upilex: much lower interferences, but dangerous for warping at
vacuum and giving rise occasionally to loud implosions.
18
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
• The discharge didn’t induce any coupling with the radio receivers. • Astronomical observation of a SiO maser signal from the star TX
Cam through the reactor showed identical results with PLASMA ON/OFF.
Test of interferences between Receivers and RF discharge
TX CamSiO, v=1
PLASMA ON
PLASMA OFF
19
• 80% of molecular species discovered in the ISM contain at least one carbon atom Study focused on different carbon containing molecules of astrophysical interest.
Particular showcases:
• Rotational emissions of OCS and its plasma products. • Species produced with O2 plasmas used for reactor cleaning
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
EXPERIMENTAL RESULTS
20
Line profile and intensity vs. OCS gas pressure
PartialPressure(mbar)0.300.150.060.007
Tempe
rature
(K)
Voigt Profile: Pressure
broadening
Doppler Profile: Temperaturebroadening
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Low working pressures lead to significantly higher resolution, with not so large intensity loss
21
OCS 0 J=4‐3
OCS 0 J=4‐30.50 mbar
Plasma Off
0.15 mbar Plasma Off
On/Off
Frequency switching 0.15 mbar
Plasma On(5 W)
OCS, 0.5 mbar, no plasma• On/Off Acquisition • The 2 J=4‐3 doublet of OCS
and OC34S ( 0.025 mbar) were detected.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
22
OCS, 0.15 mbar, plasma (5 W) • Frequency Switching (±24 MHz)• OCS dissociation 70%• 2 /0 intensities evince a rotational temperature change:
Plasma Off (300 K) vs. Plasma On (450 K)
CS radical produced in OCS plasma, 0.15 mbar, 50 W
3
2
1
0
T B(K)
‐5 0 5(‐0) (MHz)
0 = 48990.957 MHz
CS J=1‐0
During the discharge, Sulphur was deposited on the walls, which got very dirty and smelly
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
23
O2 cleaning discharge after Sulphur & Carbondeposition by the former OCS plasma: SO2 generation
• 0.1 mbar O2, 60 W.
• SO2 (40 bar), detected in 2 min.
• SO2 was produced first, and quickly disappeared in favor of CO and CO2
SO2142,12 – 133,1
47913.427 MHz)
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
24
Concentrations are estimated with MADEX Code (Cernicharo 2012)
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Proof of concept published in Astronomy & Astrophysics
25
Lessons learned from the former experimentsin the radio telescope
• Radio‐astronomy techniques can be used successfully to perform rotational spectroscopy and chemical simulations in laboratory.
• Estimated sensitivity can be as good as 0. 1 bar of CS in the Q band. Better sensitivity and broader spectral range is expected in the W band.
• Special care is needed to get the lower background temperatures and best observing procedures.
• Windows material should provide as low interferences and noise as possible and allow safe operation => Teflon seems the best choice
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
26
2.‐ NEW SET‐UP FOR LABORATORY ASTROPHYSICS Yebes Observatory
Former Campaignsin 40 m Telescope
New ExperimentsLocation
Control room of an 13.5 m old telescope covered with spherical dome (1980’s)
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
27
EQUIPED WITH Q AND W BAND RECEIVERS A LARGER REACTOR AND CRYOGENIC COLD LOADS
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
New Reactor (1 m x 0.6 m)16 FFTS
Q & W band Radio‐Receiversin a single housing
Mirrors
FRONT ENDSIDE VIEW
RF Source & Matching
Gas Mixing
VacuumPumps
28
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Teflon corrugated windows
optimized for W band
New Reactor
Cold Load: He Cryostat at 17 K
Mirrors
BACK END
Visible Spectrometer
MassSpectrometer
29Plasma
RECEIVERS
• Two receivers in a single cryostat set‐up: Q band = 32‐50 GHz ( = 9.1 – 6 mm)W band = 72‐116 GHz ( = 6.4 – 2.7 mm) with 3 sub‐bands: L (72‐90 GHz), M(84‐102 GHz), H (98‐116 GHz)
• Q band and one of W sub‐bands can operate simultaneously.Each one is connected to 8 FFTS2.5 GHz, 1.5 x 106 channels
Instantaneous Bandwidth = 18 GHz Resolution = 38 kHz
30
300 250 200 150 100 50
Frequency (GHZ)
Atmosph
ericTran
smission
1.0
0.8
0.6
0.4
0.2
0.0
Q ban
d
W ban
d
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
1. OFF/ON technique: Cell empty / Cell full 2. Frequency Switching (± 37 MHz) 3. Load switching : without cell evacuation, but using the hot
load and an additional cold load next to the receivers as references
31
SAMPLING MODES FOR NOISE REDUCTION
Mobile Mirror to switch between Hot Load and Internal Cold Load
in front of the Receiver
EXPERIMENTAL RESULTS (May‐June 2018)
Precursors and plasma products of pure gases and mixtures containing carbon atoms, like for example:
• CH3CN• CH4 + N2• O2 (used for reactor cleaning)
Very low pressures ( tens of bar) can be used at stable plasma conditions, reducing line‐widths and improving signal detection.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
32
CH3CN without discharge, 25 bar, 11 min/sub‐band
All vibrational excited states < 0.1 eV detected
13 C/12 C = 1% 0.25 bar CH3
13CN15 N/14 N = 0.4% 0.1 bar CH3C15N
33
Remaining CH3CN(0.5 bar) > 80% precursor dissociation
Plasma of CH3CN (3 µbar, 20 W, 80 s)HC3N (0.006 bar) & HCN (0.4 bar) produced
34
Reactor cleaning with 20 µbar, O2 plasma, after CH3CN discharge
HCN and CO produced in the discharge were detected.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
35
HCN (0.085 bar) = 2.98 D
CO (8 bar) = 0.12 D Dirtiness deposited
on the reactor walls
Plasma of 13 µbar CH4 + 19 µbar N2 , 20 W (40 min)Main component of Titan atmosphereCH3CN, HCN and HC3N produced in the discharge were detected.
36
HCN (0.4 bar)CH3CN (0.01 bar)
HC3N (0.015 bar)
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
Reactor cleaning with O2 plasma after the CH4 + N2 dischargeCO (5 µbar) , HCN (0.1 µbar) and HNCO (0.3 µbar) were detected
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
37
(0.3 µbar)
0.1 µbar
(5 µbar)
1. The technique is very useful to measure frequencies of molecular species with high accuracy and sensitivity, and to derive their rotational constants with high precision.
2. It allows to detect and identify minor products in cold plasmas, even with very complex precursor mixtures, and contribute to the identification of the kinetic processes.
3. These experiments open new and very encouraging opportunities to laboratory astrophysics and to possible collaborations between astronomers and plasma specialists.
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague
38
Thank you very muchfor your attention!
39
45th Conf. Plasma PhysicsJuly 2‐6 2018, Prague