Schedule of Accreditation issued by
United Kingdom Accreditation Service
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
0147
Accredited to
ISO/IEC 17025:2005
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
610 Wharfedale Road
Winnersh Triangle
Wokingham
Berkshire
RG41 5TP
United Kingdom
Contact: Contact Centre
Tel: +44 (0) 118 927 6201
E-Mail: [email protected]
Website: www.keysight.com
Calibration performed by the Organisations at the locations specified overleaf
The Calibration and Measurement Capabilities (CMCs) presented in this Schedule represent the smallest uncertainties for which the laboratory is accredited and are obtained at the Winnersh facility. Measurement systems are also available at the locations listed overleaf and include some or all of the quantities included in this Schedule. Customers’ equipment will be calibrated at a location with a capability appropriate for that particular type of equipment. The assigned uncertainties will be such as to properly demonstrate the performance of the equipment and may be greater than the CMCs stated in this Schedule. Where more than one location is able to calibrate a specific item of equipment, the quoted uncertainties will normally be identical. Fort further information regarding Calibration and Measurement Capabilities, see the last page of this Schedule. Coverage factor The coverage factor k stated herein is the number by which the combined standard uncertainty u(y) is multiplied to obtain the expanded uncertainty U for a coverage probability of 95 %. The coverage factor is provided so that the user can determine u(y) for use in subsequent uncertainty budgets. For certain calibrations, Monte Carlo Simulation (MCS) is used for uncertainty evaluation, based on the processes described in JCGM 101:2008, Evaluation of measurement data - Supplement 1 to the “Guide to the expression of uncertainty in measurement” - Propagation of distributions using a Monte Carlo method. If an MCS evaluation reveals that an asymmetric probability density function is associated with the estimate y of the output quantity Y, the value of k may not be relevant or available. In such cases, the stated CMCs represent the endpoints of the 95 % coverage interval and the following information will be provided in calibration certificates:
• The combined standard uncertainty u(y). This may be used directly in subsequent uncertainty budgets without the need for a divisor or, where one is required, with a divisor of 1.
• The endpoints of the 95 % coverage interval. This information may be used, for example, to assist in evaluating compliance with specification limits or to otherwise assess fitness for purpose.
0147
Accredited to
ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 2 of 60
Laboratory locations:
Location details Activity
Address Keysight Technologies UK Limited 610 Wharfedale Road Winnersh Triangle Wokingham Berkshire RG41 5TP
Local contact Contact Centre Tel: +44 (0) 118 927 6201 E-mail: [email protected]
Calibration DC and LF electrical quantities RF and Microwave electrical quantities
Address Keysight Technologies Deutschland GmbH Servicezentrum Herrenberger Straße 130 71034 Böblingen Germany
Local contact Contact Centre Tel: +49 (0) 7031 464 6333 E-mail: [email protected]
Calibration DC and LF electrical quantities RF and Microwave electrical quantities
Address Keysight Technologies Deutschland GmbH Keltenring 13 82041 Oberhaching Germany
Local contact Contact Centre Tel: +49 (0) 7031 464 6333 E-mail: [email protected]
Calibration DC and LF electrical quantities RF and Microwave electrical quantities
Address Keysight Technologies Deutschland GmbH Lise-Meitner-Allee 27 D-44801 Bochum Germany
Local contact Contact Centre Tel: +49 (0) 7031 464 6333 E-mail: [email protected]
Calibration Calibration of Battery Test Systems and similar devices.
0147
Accredited to
ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 3 of 60
Site activities performed away from the locations listed above:
Customers’ sites or premises Calibrations may be performed in an air-conditioned mobile laboratory that is taken to the customers’ sites, or in suitable areas within the customers’ premises. The customers’ premises must be appropriate for the nature of the particular calibrations undertaken and will be the subject of contract review arrangements between the laboratory and the customer.
Local contact Contact Centre (UK) Tel: +44 (0) 118 927 6201 E-mail: [email protected] Contact Centre (Germany) Tel: +49 (0) 7031 464 6333 E-mail: [email protected]
Calibration DC and LF electrical quantities RF and Microwave electrical quantities. Calibration of Battery Test Systems and similar devices.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 4 of 60
DETAIL OF ACCREDITATION
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
DC RESISTANCE
Specific Values 100 μ 280 ppm Measurement of resistors with negligible power dissipation.
The CMCs up to 10 M are
obtained at 20 C in a temperature controlled oil bath.
1 m 25 ppm
10 m 14 ppm
100 m 5.0 ppm
1 2.0 ppm
10 1.5 ppm
100 1.5 ppm
1 k 1.5 ppm
10 k 0.50 ppm
100 k 2.4 ppm
1 M 5.7 ppm
10 M 6.8 ppm
100 M 19 ppm
1 G 270 ppm
19 kΩ 3.0 ppm As measured in air 23 °C ± 1°C
190 kΩ 5.2 ppm
1.9 MΩ 12 ppm
3 MΩ 12 ppm
19 MΩ 12 ppm
30 MΩ 12 ppm
33 MΩ 12 ppm
109 MΩ 60 ppm
290 MΩ 70 ppm
Other Values 0 ±10 µΩ 0.40 µΩ Sourcing only, relative to a
10 μ to 100 μ
100 μ to 1 m
1 m to 10 m
10 m to 100 m
0.1 to 1
1 to 10
10 to 100
100 to 1 k
1 k to 10 k
10 k to 120 k
120 k to 1.2 M
1.2 M to 12 M
12 M to 120 M
120 M to 1.2 G
650 ppm
60 ppm
60 ppm
9.0 ppm
6.5 ppm
6.5 ppm
2.0 ppm
2.0 ppm
1.7 ppm
11 ppm + 74 m
15 ppm + 2.4
53 ppm + 110
500 ppm + 3.9 k
0.50 % + 300 k
reference established using a low thermal emf, 4-terminal short circuit.
10 G
100 G
1 T
10 T
0.10 %
0.10 %
0.10 %
0.20 %
Calibration of resistance measuring equipment and standard resistors at test voltages of 100V, 250V and 500 V.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 5 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
DC VOLTAGE
Specific Values 1 V to 1.1 V 10 V
1.2 μV 0.80 ppm
Standard cells and solid-state reference standards can be measured on a fully automated system.
10 mV 0.1 V 1 V 10 V 19 V 100 V 1 kV
23 ppm 2.5 ppm 0.80 ppm 0.80 ppm 1.2 ppm 0.80 ppm 1.0 ppm
Other Values -10 μV to +10 μV 50 nV
10 μV to 1 V 0.60 ppm + 0.43 μV
1 V to 10 V 1.3 ppm
10 V to 100 V 3.5 ppm
100 V to 1000 V 4.0 ppm
DC CURRENT 0 A 1.0 pA For calibration of precision digital multimeters (open circuit input).
5 nA to 10 nA 10 nA to 100 nA 100 nA to 1.2 μA 1.2 μA to 12 μA 12 μA to 120 μA 120 μA to 1.2 mA 1.2 mA to 12 mA 12 mA to 100 mA 100 mA to 120 mA 120 mA to 1 A 1 A to 10 A 10 A to 20 A 20 A to 200 A
180 ppm 180 ppm to 39 ppm 40 ppm to 14 ppm 14 ppm to 13 ppm 8.0 ppm to 5.0 ppm 7.0 ppm to 5.0 ppm 7.0 ppm to 5.0 ppm 7.0 ppm to 5.0 ppm 39 ppm to 7.0 ppm 39 ppm to 11 ppm 42 ppm to 21 ppm 42 ppm 180 ppm to 71 ppm
Current probe calibration Simulated current using multi-turn coil arrangement.
Current difference method 10 mA to 1000 A 0.40 %
Absolute current method 5 mA to 500 A 0.54 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 6 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
DIFFERENTIAL VOLTAGE AND CURRENT
These capabilities are for measurement of short-term differences in DC voltage or current and will be typically applied to load and line regulation checks on DC power supplies and similar devices.
Differential DC Voltage For measured voltages in the ranges:
0.1 V to 10 V 0.50 V to 1.0 V
10 V to 100 V 10 V to 60 V
100 V to 1000 V 0.60 mV to 1.5 mV
Differential DC Current For measured currents in the ranges:
10 mA to 1 A 0.5 A to 1.0 A
1 A to 10 A 5.0 A to 10 A
10 A to 100 A 50 A to 100 A
100 A to 200 A 0.5 mA to 1.0 mA
AC VOLTAGE
Three reference standards are available for dissemination of AC voltage measurement traceability, each with different frequency ranges, voltage ranges and CMCs. For clarity, each system is described separately in this Schedule; these capabilities are listed on the following five pages.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 7 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Specific Values (System 1)
Voltage Calibration and Measurement Capability in ppm of value for specific values of AC voltages at the frequencies shown
10 Hz 40 Hz 1 kHz 30 kHz 50 kHz 100 kHz 200 kHz 500 kHz 700 kHz 1 MHz
100 mV 46 52 54 54 290 750
300 mV 43 38 38 39 40 41 220 170 540 450
1 V 41 32 30 32 33 34 41 110 150 440
3 V 52 39 39 39 40 40 140 160 390 390
10 V 48 32 32 32 33 33 34 130 160 390
20 V 120 360 360
30 V 44 41 32 32 34 37 83
100 V 48 41 36 36 37 38
10 Hz 40 Hz 1 kHz 30 kHz 50 kHz 100 kHz
300 V 48 48 48 48 48
500 V 190
700 V 150
1000 V 44 44 44 44
Other values (System 1)
Voltage range
Calibration and Measurement Capability in ppm of value for other values of AC voltages over the frequency ranges shown
10 Hz to 100 Hz 100 Hz to 30 kHz 30 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz
1 mV to 3.3 mV 1300 1100 1600 2900 5900
3.3 mV to 11 mV 410 260 580 1700 4000
11 mV to 33 mV 320 200 420 1300 3000
33 mV to 110 mV 250 120 230 690 1900
10 Hz to 40 Hz 40 Hz to 30 kHz 30 kHz to 200 kHz 200 kHz to 500 kHz 500 kHz to 1 MHz
0.1 V to 0.33 V 42 42 86 350 900
0.33 V to 1.1 V 39 39 82 350 900
1.1 V to 3.3 V 39 39 82 350 890
3.3 V to 11 V 39 39 82 350 890
10 Hz to 40 Hz 40 Hz to 30 kHz 30 kHz to 100 kHz 100 kHz to 500 kHz 500 kHz to 1 MHz
10 V to 30 V 48 40 82
10 V to 20 V 350 900
30 V to 110 V 48 40 84
10 Hz to 40 Hz 40 Hz to 20 kHz 20 kHz to 100 kHz
100 V to 330 V 57 57 140
300 V to 1.1 kV 57 63 200
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 8 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Specific Values (System 2)
Voltage Calibration and Measurement Capability in ppm of value for specific values of AC voltages at the frequencies shown . The maximum voltage is constrained by the V.Hz product of 108 10 Hz 20 Hz 40 Hz 500 Hz 1 kHz 10 kHz 20 kHz 50 kHz 100 kHz 200 kHz 500 kHz 1 MHz
2 mV 1200 1200 1200 1200 1200 1200 1200 1200 1200 1200 1800 3300
6 mV 270 190 170 170 170 170 170 170 180 320 640 2100
20 mV 220 150 130 130 130 130 130 130 140 230 500 1600
60 mV 160 93 89 89 88 89 89 93 110 190 380 1100
100 mV 150 70 55 51 51 51 51 51 84 180 340 1100
200 mV 140 70 52 48 48 49 49 68 78 160 330 1100
600 mV 140 58 41 41 41 41 42 42 68 140 310 1100
1 V 140 56 36 36 36 36 36 37 63 130 280 970
2 V 140 56 36 36 36 36 36 37 63 130 280 970
6 V 140 57 36 36 36 36 36 39 74 170 400 1300
10 V 140 57 37 36 36 36 36 39 74 170 400 1300
20 V 140 57 38 37 37 37 37 40 76 170 400 1300
60 V 140 59 47 43 43 43 43 44 93
100 V 140 60 47 43 43 43 43 50 93
200 V 150 60 47 43 43 43 43 50 100
600 V 150 81 56 50 50 50 50 130 500
1000 V 150 81 58 51 51 51 54 130 510
Other values (System 2)
Voltage range
Calibration and Measurement Capability in (ppm of value + μV) for other values of AC voltages over the frequency ranges shown. The maximum voltage is constrained by the V.Hz product of 108
10 Hz to 20 Hz
20 Hz to 40 Hz
40 Hz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 300 kHz
300 kHz to 500 kHz
500 kHz to 1 MHz
0.6 mV to 2.2 mV
380 + 7.5 310 + 3.6 290 + 2.4 310 + 2.8 320 + 3.1 460 + 4.3 880 + 8.6 2700 + 12
2.2 mV to 7 mV
260 + 7.5 180 + 3.6 160 + 2.4 170 + 2.8 180 + 3.1 320 + 4.3 600 + 8.6 2100 + 12
7 mV to 22 mV
210 + 7.5 140 + 3.6 120 + 2.4 130 + 2.8 130 + 3.1 230 + 4.3 480 + 8.6 1600 + 12
22 mV to 70 mV
150 + 7.0 82 + 3.7 71 + 2.5 79 + 2.8 96 + 3.1 180 + 4.3 350 + 8.6 1100 + 12
70 mV to 220 mV
140 + 7.0 67 + 3.7 51 + 2.5 47 + 2.8 82 + 3.1 160 + 4.3 340 + 8.6 1100 + 12
220 mV to 700 mV
140 + 7.0 62 + 3.7 39 + 2.5 38 + 2.8 69 + 3.1 150 + 4.3 310 + 8.6 1100 + 12
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 9 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Other values (System 2, continued)
Voltage range
Calibration and Measurement Capability in ppm of value for other values of AC voltages over the frequency ranges shown. The maximum voltage is constrained by the V.Hz product of 108
10 Hz to 20 Hz
20 Hz to 40 Hz
40 Hz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 300 kHz
300 kHz to 500 kHz
500 kHz to 1 MHz
0.7 V to 2.2 V
140 56 36 37 63 130 280 970
2.2 V to 7 V
140 57 36 39 74 170 400 1300
7 V to 22 V 140 57 38 40 76 170 400 1300
22 V to 70 V
140 58 43 44 91 900
+ 16 mV
70 V to 220 V
150 60 47 50 100
220 V to 700 V
150 81 56 130 500
700 V to 1050 V
150 81 58 130 510
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 10 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Specific Values (System 3)
Voltage Calibration and Measurement Capability in ppm of value for specific values of AC voltages at the frequencies shown
10 Hz 20 Hz 100 Hz 1 kHz 10 kHz 20 kHz 50 kHz 100 kHz
300 kHz
500 kHz
1 MHz
2 mV 1200 1200 1200 1200 1200 1200 1200 1200 1200 1800 3600
6 mV 140 110 110 110 110 110 110 130 270 560 2100
10 mV 72 110
20 mV 120 88 74 74 74 74 86 120 230 500 1600
60 mV 140 88 72 68 73 76 95 150 290 510 1200
100 mV 46 47
200 mV 91 48 42 43 42 42 56 110 200 320 970
600 mV 84 55 40 40 40 40 45 62 180 290 950
1 V 120 51 35 36 36 36 39 58 150 260 920
2 V 82 45 35 35 35 35 38 58 170 260 920
6 V 83 45 36 35 35 35 43 69 190 390 1200
20 V 82 46 36 36 36 36 39 70 190 390 1200
60 V 83 48 39 39 39 39 49 74 180
200 V 91 49 41 39 39 39 45 77
600 V 43 40 40 42 120 490
1000 V 43 41 41 41
Other values (System 3)
Voltage range
Calibration and Measurement Capability in (ppm of value + μV) for other values of AC voltages over the frequency ranges shown
10 Hz to 20 Hz
20 Hz to 40 Hz
40 Hz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 300 kHz
300 kHz to 500 kHz
500 kHz to 1 MHz
0.6 mV to 2.2 mV
230 + 7.5 200 + 3.6 190 + 2.4 200 + 2.8 230 + 3.1 410 + 4.3 830 + 8.6 3000 + 12
2.2 mV to 7 mV
130 + 7.5 87 + 3.6 85 + 2.4 89 + 2.8 120 + 3.1 260 + 4.3 540 + 8.6 2100 + 12
7 mV to 22 mV
110 + 7.5 77 + 3.6 60 + 2.4 74 + 2.8 110 + 3.1 230 + 4.3 480 + 8.6 1600 + 12
22 mV to 70 mV
130 + 7.0 72 + 3.7 59 + 2.5 80 + 2.8 150 + 3.1 290 + 4.3 490 + 8.6 1100 + 12
70 mV to 220 mV
120 + 7.0 49 + 3.7 32 + 2.5 37 + 2.8 59 + 3.1 150 + 4.3 310 + 8.6 970 + 12
220 mV to 700 mV
76 + 7.0 52 + 3.7 32 + 2.5 39 + 2.8 61 + 3.1 170 + 4.3 300 + 8.6 940 + 12
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 11 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Other values (System 3, continued)
Voltage range
Calibration and Measurement Capability in ppm of value for other values of AC voltages over the frequency ranges shown The maximum voltage is constrained by the V.Hz product of 108
10 Hz to 20 Hz
20 Hz to 40 Hz
40 Hz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 300 kHz
300 kHz to 500 kHz
500 kHz to 1 MHz
0.7 V to 2.2 V
82 45 35 38 58 170 260 920
2.2 V to 7 V
83 45 36 43 72 190 390 1200
7 V to 22 V 82 46 36 39 70 190 390 1200
22 V to 70 V
83 48 39 49 80 900
+ 16 mV
70 V to 220 V
91 49 41 45 77
220 V to 700 V
43 120 490
700 V to 1050 V
43
Voltage range
Calibration and Measurement Capability in (% of value + μV) for other values of AC voltages over the frequency ranges shown
500 kHz to 1.2 MHz 1.2 MHz to 2 MHz 2 MHz to 10 MHz 10 MHz to 20 MHz 20 MHz to 30 MHz
0.6 mV to 2.2 mV
0.21 + 1.0 0.21 + 1.0 0.28 + 1.0 0.43 + 1.0 0.87 + 2.0
2.2 mV to 7 mV
0.12 + 1.0 0.12 + 1.0 0.16 + 1.0 0.27 + 1.0 0.50 + 1.0
7 mV to 22 mV
0.11 + 0 0.11 + 0 0.15 + 0 0.26 + 0 0.50 + 0
22 mV to 70 mV
0.080 + 0 0.080 + 0 0.14 + 0 0.24 + 0 0.48 + 0
70 mV to 220 mV
0.080 + 0 0.080 + 0 0.14 + 0 0.24 + 0 0.47 + 0
220 mV to 700 mV
0.080 + 0 0.080 + 0 0.14 + 0 0.24 + 0 0.47 + 0
700 mV to 2.2 V
0.080 + 0 0.080 + 0 0.14 + 0 0.24 + 0 0.47 + 0
2.2 V to 7 V 0.080 + 0 0.080 + 0 0.14 + 0 0.24 + 0 0.48 + 0
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 12 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC VOLTAGE (continued)
Specific Values Square wave Sine wave Triangle wave
1 kHz 1.8 mV 11.9 mV, 21.9 mV, 21.9 mV, 22 mV, 56 mV, 89.9 mV, 90 mV, 155 mV, 219 mV, 220 mV, 560 mV, 899 mV, 900 mV, 3.75 V, 6.59 V, 6.6 V and 30.8 V
1.0 % 1.4 % 1.7 % 0.10 %
Source output impedance
nominally 50 Ω or 1 M. Source output impedance
nominally 1 M. Calculated peak-to-peak voltage from rms measurement by DMM and appropriate to calibration of an oscilloscope calibrator output of a multifunction calibrator.
10 Hz, 100 Hz, 1 kHz and 10 kHz 55 V
0.10 %
1 kHz Source output impedance
nominally 50 .
6.4 mV, 10.9 mV, 11 mV, 28 mV, 44.9 mV, 45 mV, 78 mV, 109 mV, 110 mV, 280 mV, 449 mV, 450 mV, 780 mV, 1.09 V, 1.1 V, 1.8 V and 2.5 V
0.40 %
LF AC Voltage system
Calibration and Measurement Capability in % of value for LF AC voltages at the frequencies and voltages shown. Calibrations may be conducted at intermediate values with the larger of the adjacent uncertainties. The sine wave distortion must not exceed 1 %.
Voltage 1 Hz and 3 Hz 9.5 Hz 10 Hz 45 Hz 1 kHz
3 mV 0.013 %
10 mV 0.021 % 0.016 %
30 mV 0.0063 % 0.0024 % 0.0027 % 0.0080 %
33 mV 0.0033 %
70 mV 0.0023 %
300 mV 0.0018 % 0.0017 % 0.0021 % 0.0070 %
330 mV 0.0026 %
3 V 0.0018 % 0.0016 % 0.0019 % 0.0070 %
3.3 V 0.0042 %
5 V 0.0026 % 0.0054 % 0.0032 % 0.0076 %
30 V 0.0043 % 0.0025 % 0.0033 %
33 V 0.0033 %
300 V 0.0040 %
330 V 0.0052 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 13 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
AC PHASE ANGLE
Specific Values
Voltage : Voltage Phase angle uncertainty in degrees
3 V : 3 V At 0° At 60° At 90°
65 Hz 0.020° 0.020° 0.020
400 Hz 0.020° 0.020° 0.020°
1 kHz 0.030° 0.030° 0.030°
5 kHz 0.15° 0.090° 0.070°
10 kHz 0.15° 0.15° 0.15°
30 kHz 0.41° 0.37° 0.40°
30 V : 3 V At 0° At 60° At 90°
65 Hz 0.020°
50 V : 3 V At 0° At 60° At 90°
65 Hz 0.020°
Voltage : Current Phase angle uncertainty in degrees
30 mV : 300 mA At 0° At 60° At 90°
65 Hz 0.010° 0.020°
1 kHz 0.030°
30 kHz 0.36
200 mV : 2 A At 0° At 60° At 90°
65 Hz 0.020° 0.020°
50 mV : 5 A At 0° At 60° At 90°
65 Hz 0.020°
400 Hz 0.020°
200 mV : 20 A At 0° At 60° At 90°
65 Hz 0.020°
400 Hz 0.020°
3.3 V : 300 mA At 0° At 60° At 90°
65 Hz 0.030° 0.030°
3.3 V : 2 A At 0° At 60° At 90°
65 Hz 0.010° 0.030°
3.3 V : 5 A At 0° At 60° At 90°
65 Hz 0.020°
400 Hz 0.020°
3.3 V : 20 A At 0° At 60° At 90°
65 Hz 0.030°
400 Hz 0.020°
33 V : 300 mA At 0° At 60° At 90°
65 Hz 0.020° 0.020°
33 V : 2 A At 0° At 60° At 90°
65 Hz 0.020° 0.020°
33 V : 5 A At 0° At 60° At 90°
65 Hz 0.030°
400 Hz 0.030°
33 V : 20 A At 0° At 60° At 90°
65 Hz 0.030°
400 Hz 0.030°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 14 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC CURRENT
Specific Values From 10 Hz to 5 kHz:
10 μA 100 μA 1 mA 10 mA 100 mA
0.087 % 0.033 % 0.033 % 0.033 % 0.033 %
Voltage/resistance method
From 20 Hz to 10 kHz:
10 mA 100 mA 1 A 10 A
0.011 % 0.011 % 0.015 % 0.025 %
Thermal transfer method
20 µA
1 kHz 0.014 %
10 kHz 0.075 %
33 µA 1 kHz 10 kHz and 30 kHz
0.018 % 0.045 %
190 µA, 200 µA, 329 µA, 330 µA, 1.9 mA, 2 mA, 3.29 mA, 3.3 mA, 19 mA, 20 mA, 32.9 mA, 33 mA, 190 mA, 200 mA, 329 mA and 330 mA 10 Hz to 10 kHz 30 kHz
0.0040 % 0.0050 %
1.09 A, 2 A, 2.09 A, 2.99 A and 3.3 A 10 Hz to 10 kHz
0.0070 %
10 A and 10.9 A 40 Hz to 10 kHz
0.010 %
20 A 45 Hz to 5 kHz
0.010 %
Other Values 9 μA to 220 μA
10 Hz to 20 Hz 250 ppm + 16 nA
20 Hz to 40 Hz 160 ppm + 10 nA
40 Hz to 1 kHz 120 ppm + 8.0 nA
1 kHz to 5 kHz 280 ppm + 12 nA
5 kHz to 10 kHz 1100 ppm + 65 nA
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 15 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC CURRENT (continued)
Other values (continued) 220 μA to 2.2 mA
10 Hz to 20 Hz 250 ppm + 40 nA
20 Hz to 40 Hz 160 ppm + 35 nA
40 Hz to 1 kHz 120 ppm + 35 nA
1 kHz to 5 kHz 200 ppm + 110 nA
5 kHz to 10 kHz 1100 ppm + 650 nA
2.2 mA to 22 mA
10 Hz to 20 Hz 250 ppm + 400 nA
20 Hz to 40 Hz 160 ppm + 350 nA
40 Hz to 1 kHz 120 ppm + 350 nA
1 kHz to 5 kHz 200 ppm + 550 nA
5 kHz to 10 kHz 1100 ppm + 5.0 μA
22 mA to 220 mA
10 Hz to 20 Hz 250 ppm + 4.0 μA
20 Hz to 40 Hz 160 ppm + 3.5 μA
40 Hz to 1 kHz 120 ppm + 2.5 μA
1 kHz to 5 kHz 200 ppm +3.50 μA
5 kHz to 10 kHz 1100 ppm + 10 μA
220 mA to 2.2 A
20 Hz to 1 kHz 260 ppm + 35 μA
1 kHz to 5 kHz 450 ppm + 80 μA
5 kHz to 10 kHz 0.70 % + 160 μA
2.2 A to 11 A
40 Hz to 1 kHz 460 ppm + 0.17 mA
1 kHz to 5 kHz 950 ppm + 0.38 mA
5 kHz to 10 kHz 0.36 % + 0.75 mA
AC RESISTANCE
Specific Values CMC (% of value)
10 , 100 and 1 k 10 100 1 k For the calibration of meters and standards with 4-terminal pair BNC connectors. Other configurations can be calibrated but the uncertainties may be increased. * 0.031 % at 100 kHz
20 Hz to 100 Hz 0.043 0.051 0.053
100 Hz to 1 MHz 0.094 0.050 0.039 *
1 MHz 0.096 0.030 0.031
2 MHz 0.090 0.040 0.031
3 MHz 0.10 0.050 0.031
4 MHz 0.10 0.050 0.041
5 MHz 0.12 0.050 0.051
10 MHz 0.40 0.20 0.20
13 MHz 0.60 0.30 0.30
10 k
20 Hz to 100 kHz 100 kHz 100 kHz to 500 kHz 1 MHz
0.022 % 0.021 % 0.031 % 0.031 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 16 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AC RESISTANCE
(continued) 100 k
20 Hz to 100 kHz 100 kHz 100 kHz to 500 kHz 1 MHz
0.017 % 0.0070 % 0.017 % 0.0070 %
CAPACITANCE
CMC (% of value)
Specific Values 1 pF, 10 pF and 100 pF 1 pF 10 pF 100 pF For the calibration of meters and standards with 4-terminal pair BNC connectors. Other configurations can be calibrated but the uncertainties may be increased.
100 Hz to 1 kHz 1 kHz to 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz 10 MHz 13 MHz
- 0.26 0.041 0.053 0.071 0.091 0.25 0.37
- 0.023 0.019 0.019 0.019 0.020 0.023 0.025
0.046 0.021 0.021 0.022 0.023 0.025 0.039 0.054
1000 pF
20 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz 1 kHz to 1 MHz 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz 10 MHz 13 MHz
0.078 % 0.067 % 0.032 % 0.065 % 0.033 % 0.031 % 0.039 % 0.051 % 0.067 % 0.19 % 0.28 %
CMC (% of value)
10 nF and 100 nF At 10 nF At 100 nF
120 Hz 0.017 0.017
1 kHz 0.017 0.017
10 kHz 0.017 0.017
100 kHz 0.017 0.017
CMC (% of value)
1 μF and 10 μF At 1 μF At 10 μF
120 Hz 0.018 0.017
1 kHz 0.018 0.017
10 kHz 0.018 0.023
100 kHz 0.019 0.072
Measurement At 1 kHz Measurement by capacitance meter.
220 pF 0.20 %
390 pF, 480 pF, 600 pF, 1 nF and 2 nF 0.15 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 17 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CAPACITANCE (continued ) At 1 kHz Measurement by capacitance meter.
7 nF, 10.9 nF, 20 nF, 70 nF, 109 nF, 200 nF and 300 nF 0.090 %
At 100 Hz
700 nF, 1.09 µF, 2 µF, 3 µF, 7 µF, 10.9 µF, 20 µF and 30 µF 0.090 %
At 50 Hz
70 µF and 109 µF 0.17 %
200 µF, 300 µF, 0.33 mF, 0.7 mF, 1.09 mF, 1.1 mF, 2 mF, 3 mF, 3.3 mF, 10.9 mF, 20 mF, 30 mF, 33 mF and 110 mF 0.025 %
Capacitance calculated from measured voltage and time after application of a known direct current (ΔV / ΔT method).
Generation 0 pF 5.8 pF For calibration of the capacitance function on digital multimeters (open circuit input).
Specific values 10 nF 0.030 nF The CMCs apply to the capacitance ranges of digital multimeters that use current charge/discharge time to calculate the applied capacitance.
100 nF 0.27 nF
1 µF 2.7 nF
10 µF 27 nF
100 µF 430 nF
1 mF 4.3 µF
10 mF 45 µF
Other values 0.19 nF to 3.3 nF 0.40 % + 0.010 nF The CMCs apply to the capacitance ranges of digital multimeters that use current charge/discharge time to calculate the applied capacitance.
3.3 nF to 11 nF 0.20 % + 0.010 nF
11 nF to 110 nF 0.20 % + 0.080 nF
110 nF to 330 nF 0.20 % + 0.23 nF
0.33 µF to 1.1 µF 0.20 % + 0.78 nF
1.1 µF to 3.3 µF 0.20 % + 2.3 nF
3.3 µF to 11 µF 0.20 % + 7.8 nF
11 µF to 33 µF 0.34 % + 23 nF
33 µF to 110 µF 0.38 % + 78 nF
110 µF to 330 µF 0.38 % + 230 nF
0.33 mF to 1.1 mF 0.38 % + 0.78 µF
1.1 mF to 3.3 mF 0.38 % + 2.3 µF
3.3 mF to 11 mF 0.38 % + 7.8 µF
11 mF to 33 mF 0.60 % + 23 µF
33 mF to 110 mF 0.86 % + 78 µF
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 18 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
DISSIPATION FACTOR D (tan δ) values between 0 and 0.01 at the following nominal values of capacitance:
For the calibration of meters and standards with 4-terminal pair BNC connectors. Other configurations can be calibrated but the uncertainties may be increased.
CMC (tan δ)
1 pF and 10 pF At 1 pF At 10 pF
1 kHz to 1 MHz 0.0024 0.00024
2 MHz 0.00044 0.00011
3 MHz 0.00041 0.00011
4 MHz 0.00042 0.00011
5 MHz 0.00023 0.00011
10 MHz 0.00058 0.00012
13 MHz 0.00084 0.00013
100 pF
100 Hz to 1 kHz 0.00043
1 kHz to 1 MHz 0.000090
2 MHz 0.000090
3 MHz 0.000090
4 MHz 0.00010
5 MHz 0.00010
10 MHz 0.00018
13 MHz 0.00025
1000 pF
20 Hz to 100 Hz 100 Hz to 1 kHz 1 kHz 1 kHz to 1 MHz 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz 10 MHz 13 MHz
0.00054 0.00046 0.00008 0.000090 0.000090 0.00010 0.00013 0.00017 0.00022 0.00059 0.00085
CMC (tan δ)
10 nF and 100 nF 10 nF 100 nF
120 Hz 0.000060 0.000070
1 kHz 0.000060 0.000060
10 kHz 0.000060 0.000060
100 kHz 0.000060 0.000070
1 μF and 10 μF 1 μF 10 μF
120 Hz 0.000070 0.000070
1 kHz 0.000060 0.000070
10 kHz 0.000070 0.00029
100 kHz 0.000080 0.00070
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 19 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
DISTORTION From 20 Hz to 20 kHz: 0 % to 0.05 % 0.05 % to 10 %
0.0070 % absolute 14 % of reading
From 20 kHz to 50 kHz: 0 % to 0.1 % 0.1 % to 10 %
0.027 % absolute 27 % of reading
From 50 kHz to 100 kHz: 0 % to 0.18 % 0.18 % to 10 %
0.050 % absolute 27 % of reading
FREQUENCY
Specific Values 1 MHz, 5 MHz and 10 MHz 1.0 x 10-12
Other Values 0.1 Hz to 500 MHz 500 MHz to 1 GHz 1 GHz to 2 GHz 2 GHz to 40 GHz
6.0 in 1011 + 0.020 nHz 2.0 in 109 1.0 in 109 5.0 in 1010
PULSE CHARACTERISTICS
Time Interval 20 ps to 1 s 0.11 % + 8.5 ps Including pulse period, width and transition time
Pulse amplitude 25 mV to 100 mV 3.2 mV
100 mV to 200 mV 4.0 mV
200 mV to 400 mV 5.4 mV
400 mV to 750 mV 7.8 mV
750 mV to 10 V 1.5 % to 2.0 %
10 V to 70 V 1.5 %
Aberrations Relative to pulse amplitdes as above.
0.30 % of pulse amplitude
AMPLITUDE MODULATION
Carrier Frequency Rate AM Depth The uncertainties are with
respect to the modulation depth and are appropriate for signals with harmonic distortion not exceeding -40 dBc. The uncertainties may be increased for higher levels of distortion at modulation depths less than 30%.
150 kHz to 10MHz 50 Hz to 10 kHz 40 % to 99 % 2.0 % of rdg. + 0.10% AM
5 % to 40% 2.0 % of rdg. + 0.010% AM
20 Hz to 50 kHz 40 % to 99% 3.0 % of rdg. + 0.10% AM
0 % to 40 % 3.0 % of rdg. + 0.010% AM
10 MHz to 1.3 GHz 50 Hz to 50 kHz 40 to 99% 1.0 % of rdg. + 0.10% AM
5 % to 40 % 1.0 % of rdg. + 0.010% AM
20 Hz to 50 Hz 40 % to 99 % 3.0 % of rdg. + 0.10% AM
and
50 kHz to 100 kHz 0 % to 40 % 3.0 % of rdg. + 0.010% AM
1.3 GHz to 26.5 GHz 50 Hz to 50 kHz 40 % to 99 % 1.5 % of rdg. + 0.10% AM
Measured via 11793A 5 % to 40 % 1.5 % of rdg. + 0.010% AM
10 MHz to 26.5 GHz 20 Hz to 100 kHz 40 % to 99 % 3.0 % of rdg. + 0.10% AM
0 % to 40% 3.0 % of rdg. + 0.010% AM
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 20 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
AMPLITUDE MODULATION
8901/2 calibrator AM Accuracy
Depth Rate CMC
33 % 10 kHz 0.25 %
8901/2 AM Accuracy
Modulation depth Rate 8901/2 IF Setting
CMC
80 % 50 kHz 1.5 MHz 0.26 %
80 % 100 kHz 1.5 MHz 0.31 %
80 % 50 Hz 1.5 MHz 0.12 %
80 % 20 Hz 1.5 MHz 0.59 %
80 % 20 Hz 455 kHz 0.29 %
80 % 50 Hz 455 kHz 0.12 %
80 % 10 kHz 455 kHz 0.12 %
FREQUENCY MODULATION
Carrier Frequency Rate FM deviation
150 kHz to 10 MHz 20 Hz to 10 kHz 4 kHz to 40 kHz 1.9 % + 10 Hz
0 kHz to 4 kHz 1.9 % + 1.0 Hz
10 MHz to 1.3 GHz 50 Hz to 100 kHz 40 kHz to 400 kHz 1.0 % + 100 Hz
4 kHz to 40 kHz 1.0 %.+ 10 Hz
0 kHz to 4 kHz 1.0 % + 1.0 Hz
20 Hz to 50 Hz 40 kHz to 400 kHz 4.8 % + 100 Hz
and 4 kHz to 40 kHz 4.8 % + 10 Hz
100 kHz to 200 kHz 0 kHz to 4 kHz 4.8 % + 1.0 Hz
10 MHz to 26.5 GHz 50 Hz to 100 kHz 40 kHz to 400 kHz 1.0 % + 100 Hz
Measured via 11793A
4 kHz to 40 kHz 1.0 % + 10 Hz
0 kHz to 4 kHz 1.0 % + 1.0 Hz
20 Hz to 200 kHz 40 kHz to 400 kHz 4.8% + 100 Hz
4 kHz to 40 kHz 4.8% + 10 Hz
0 kHz to 4 kHz 4.8% + 1.0 Hz
8901/2 FM calibrator
Peak Deviation Rate CMC
34 kHz 10 kHz 0.050 %
8901/2 FM Accuracy
Peak Deviation Rate 8901/2 IF Setting
CMC
100 kHz 50 kHz 1.5 MHz 0.12 kHz
100 kHz 100 kHz 1.5 MHz 0.17 kHz
100 kHz 50 Hz 1.5 MHz 0.14 kHz
100 kHz 20 Hz 1.5 MHz 0.14 kHz
10 kHz 20 Hz 455 kHz 36 Hz
10 kHz 10 kHz 455 kHz 18 Hz
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 21 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
RF POWER
Reference sources
Output power 1 mW 50 MHz
0.22 %
For sources with female type-N connectors
Source reflection coefficient 0 to 0.1 0.012 Where the output reflection phase is known to be 0° ± 40° or 180° ± 40°.
For the calibration of sources and receivers.
2 MHz to 26.5 GHz:
0 dBm to -50 dBm -50 dBm to -90 dBm -90 dBm to -127 dBm
0.15 dB 0.25 dB 0.35 dB
For EMC receivers the results may also be presented in
terms of dBμV in a 50 system.
CALIBRATION OF RF POWER SOURCES
The following Calibration and Measurement Capability is for the calibration 75 RF power sources fitted with Type N coaxial connectors. The uncertainties are for the measurement of sources with an output VSWR of 1.2:1 or less and with connectors that are in good condition. Instruments with a higher VSWR, fitted with different connectors to those shown, or which have connectors that are in poor condition will be assigned larger uncertainties.
75 Type N coaxial system
-30 dBm to -20 dBm -20 dBm to +10 dBm +10 dBm to +20 dBm Remarks
100 kHz to 2 GHz 0.23 dB to 0.26 dB 0.054 dB to 0.15 dB 0.13 dB to 0.17 dB Using Model 8483A sensor
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 22 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CALIBRATION OF RF POWER SOURCES (continued)
The following Calibration and Measurement Capabilities are for the calibration of 50 RF power sources fitted with Type N, APC-2.4 or APC-3.5 coaxial connectors. The uncertainties are for the measurement of sources with an output VSWR of 1.2:1 or less (up to 26.5 GHz) or 1.6:1 or less (26.5 GHz to 50 GHz), with connectors that are in good condition. Instruments that are not well matched, fitted with different connectors to those shown, or which have connectors that are in poor condition will be assigned larger uncertainties.
50 Type N coaxial system
-30 dBm to -20 dBm -20 dBm to +10 dBm +10 dBm to +20 dBm
Remarks
10 MHz to 18 GHz 0.22 dB to 0.23 dB 0.055 dB to 0.090 dB 0.16 dB to 0.18 dB Using Model 8481A sensor
50 Type N coaxial system
-30 dBm to -20 dBm -20 dBm to +10 dBm +10 dBm to +20 dBm
Remarks
100 kHz to 4.2 GHz 0.23 dB to 0.24 dB 0.052 dB to 0.098 dB 0.13 dB to 0.15 dB Using Model 8482A sensor
50 Type N coaxial system
-70 dBm to -60 dBm -60 dBm to -30 dBm -30 dBm to -20 dBm Remarks
10 MHz to 18 GHz 1.3 dB 0.057 dB to 0.15 dB 0.076 dB to 0.11 dB Using Model 8481D sensor
50 Type N coaxial system
-60 dBm to -50 dBm -50 dBm to 0 dBm 0 dBm to +20 dBm Remarks
10 MHz to 18 GHz 3.0 dB 0.080 dB to 0.38 dB 0.066 dB to 0.11 dB Using Model E9300A sensor
50 Type N coaxial system
-60 dBm to -50 dBm -50 dBm to 0 dBm 0 dBm to +20 dBm Remarks
10 MHz to 6 GHz 3.1 dB 0.080 dB to 0.27 dB 0.067 dB to 0.080 dB Using Model E9301A sensor
50 Type N coaxial system
-60 dBm to -50 dBm -50 dBm to +20 dBm Remarks
9 kHz to 6 GHz 3.0 dB 0.070 dB to 0.26 dB Using Model E9304A sensor
50 Type N coaxial system
-10 dBm to +25 dBm +25 dBm to +35 dBm Remarks
10 MHz to 18 GHz 0.070 dB to 0.10 dB 0.19 dB to 0.20 dB Using Model 8481H sensor
50 Type N coaxial system
-10 dBm to +25 dBm +25 dBm to +35 dBm Remarks
100 kHz to 4.2 GHz 0.080 dB 0.19 dB to 0.20 dB Using Model 8482H sensor
50 Type N coaxial system
-70 dBm to -60 dBm -60 dBm to +20 dBm Remarks
10 MHz to 18 GHz 3.1 dB 0.17 dB to 0.27 dB Using Model E4412A sensor
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 23 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CALIBRATION OF RF POWER SOURCES (continued)
50 Type N coaxial system
-35 dBm to -20 dBm -20 dBm to +20 dBm Remarks
100 kHz to 6 GHz 0.46dB 0.056 dB to 0.12 dB Using Model N8482A sensor
50 APC-3.5 coaxial system
-70 dBm to -60 dBm -60 dBm to +20 dBm Remarks
50 MHz to 26.5 GHz 3.8 dB 0.17 dB to 0.29 dB Using Model E4413A sensor
50 APC-3.5 coaxial system
-30 dBm to -20 dBm -20 dBm to +10 dBm +10 dBm to +20 dBm
Remarks
50 MHz to 26.5 GHz 0.29 dB to 0.31 dB 0.055 dB to 0.14 dB 0.16 dB to 0.20 dB Using Model 8485A sensor
50 APC-3.5 coaxial system
-70 dBm to -60 dBm -60 dBm to -30 dBm -30 dBm to -20 dBm Remarks
50 MHz to 26.5 GHz 1.3 dB 0.075 dB to 0.18 dB 0.13 dB to 0.18 dB Using Model 8485D sensor
50 APC-2.4 coaxial system
-30 dBm to -20 dBm -20 dBm to +10 dBm +10 dBm to +20 dBm
Remarks
50 MHz to 50 GHz 0.29 dB to 0.39 dB 0.055 dB to 0.27 dB 0.13 dB to 0.31 dB Using Model 8487A sensor
50 APC-2.4 coaxial system
-70 dBm to -60 dBm -60 dBm to -30 dBm -30 dBm to -20 dBm Remarks
30 MHz to 50 GHz 0.94 dB to 1.1 dB 0.060 dB to 0.27 dB 0.12 dB to 0.29 dB Using Model 8487D sensor
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 24 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CALIBRATION FACTOR
RF Power Sensors
The table overleaf shows the Calibration and Measurement Capability in % calibration factor for the various RF power sensors listed below. These are examples and are not intended to be restrictive; other sensors of similar frequency range, power range etc. may also be calibrated.
The uncertainties are for devices with values of voltage reflection coefficient (VRC) that typically do not exceed 0.01. The uncertainties will be increased for devices with significantly greater VRC.
The uncertainties apply to a measurement procedure that includes an evaluation of the connector repeatability of the device being calibrated. An alternative measurement procedure is available based on a single measurement of the device and generic data for the connector repeatability. This alternative procedure will normally be used, unless requested otherwise by the customer, and the reported uncertainty will be larger.
Column
Sensor type
Impedance
Connector
Maximum power
Frequency range
Example type
A
Thermocouple
50
Type N
100 mW
10 MHz to 18 GHz
8481A
B
Thermocouple
50
Type N
3 W
10 MHz to 18 GHz
8481H
C
Thermocouple
50
Type N
100 mW
100 kHz to 4.2 GHz
8482A
D
Thermocouple
50
Type N
3 W
100 kHz to 4.2 GHz
8482H
E
Thermocouple
75
Type N
100 mW
100 kHz to 3 GHz
8483A
F
Diode
50
Type N
10 μW
10 MHz to 18 GHz
8481D
G
Thermocouple
50
3.5 mm
100 mW
50 MHz to 26.5 GHz
8485A
H
Diode
50
3.5 mm
10 μW
50 MHz to 26.5 GHz
8485D
I
Thermistor Mount
50
Type N
10 mW
10 MHz to 18 GHz
8478B
J
Thermistor Mount
50
Type N
10 mW
1 MHz to 1 GHz
478A-H55
K
Thermocouple
50
Type N
100 mW
100 kHz to 2.6 GHz
11722A
L
Thermocouple
50
Type N
100 mW
10 MHz to 18 GHz
11792A opt. 001
M
Thermocouple
50
Type N
100 mW
10 MHz to 26.5 GHz
11792A
N
Diode
50
Type N
320 mW
9 kHz to 6 GHz
E9304A
See page 24
Thermocouple
50
2.4 mm
300 mW
50 MHz to 50 GHz
8487A
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 25 of 60
Calibration and Measurement Capability in % calibration factor for the power sensor types described on Page 22
Frequency A B C D E F G H I J K L M N
Lo
catio
n: S
ee
Pa
ges 2
an
d 3
9 kHz 0.67
30 kHz 0.67
50 kHz 0.67
100 kHz 0.72 1.5 2.3 0.73 0.67
300 kHz 0.62 1.4 1.2 0.62 0.67
500 kHz 0.61 1.4 0.62 0.67
1 MHz 0.61 1.4 1.1 0.59 0.61 0.67
3 MHz 0.61 1.4 1.0 0.58 0.61 0.67
5 MHz 0.61 1.4 0.58 0.61 0.67
10 MHz 0.64 1.55 0.57 1.4 1.0 0.75 0.55 0.55 0.57 0.63 0.88 0.75
30 MHz 0.58 1.4 0.57 1.4 1.0 0.67 0.47 0.55 0.57 0.58 0.67 0.76
50 MHz 0.49 1.4 0.48 1.4 1.0 0.50 0.52 0.53 0.46 0.42 0.49 0.49 0.52 0.48
100 MHz 0.58 1.4 0.57 1.4 1.1 0.66 0.64 0.66 0.46 0.58 0.57 0.57 0.64 0.63
300 MHz 0.68 1.5 0.67 1.5 1.1 0.74 0.64 0.66 0.58 0.58 0.67 0.68 0.64 0.63
500 MHz 0.68 1.5 0.67 1.5 1.1 0.75 0.65 0.66 0.58 0.76 0.68 0.63
800 MHz 0.51 1.4 0.74 0.69 0.63
900 MHz
1 GHz 0.68 1.5 0.67 1.5 1.1 0.75 0.65 0.66 0.58 0.77 0.68 0.68 0.65 0.63
1.2 GHz 0.68 1.5 0.75 0.63
1.5 GHz 0.68 1.5 0.68 1.5 1.1 0.75 0.74 0.75 0.59 0.68 0.68 0.74 0.63
2 GHz 0.69 1.5 0.68 1.5 1.1 0.76 0.75 0.76 0.60 0.68 0.68 0.75 0.64
2.5 GHz 0.68 1.5 1.2
2.6 GHz 0.68 1.5 0.69
3 GHz 0.69 1.5 0.69 1.5 1.3 0.77 0.79 0.80 0.61 0.68 0.79 0.66
3.5 GHz 0.70 1.5
3.7 GHz 0.70 1.5
4 GHz 0.69 1.5 0.71 1.5 0.77 0.79 0.80 0.63 0.68 0.79 0.66
4.2 GHz 0.76 1.5
5 GHz 0.73 1.5 0.81 0.79 0.80 0.73 0.72 0.79 0.67
6 GHz 0.72 1.5 0.83 0.91 0.92 0.72 0.72 0.91 0.68
7 GHz 0.73 1.5 0.85 0.94 0.95 0.73 0.72 0.94
8 GHz 0.73 1.5 0.89 0.99 0.99 0.80 0.73 0.99
9 GHz 0.81 1.5 0.97 1.1 1.1 0.89 0.81 1.1
10 GHz 0.82 1.5 1.1 1.1 1.1 0.92 0.82 1.1
11 GHz 0.82 1.5 1.2 1.2 1.2 0.99 0.82 1.2
12 GHz 0.83 1.5 1.1 1.2 1.2 1.00 0.82 1.2
12.4 GHz 0.87 1.6 1.1 1.3 1.3 0.99
13 GHz 0.87 1.6 1.0 1.3 1.3 0.93 0.86 1.3
14 GHz 0.87 1.6 0.93 1.3 1.3 0.87 0.87 1.3
15 GHz 0.89 1.6 1.0 1.3 1.3 0.87 0.88 1.3
16 GHz 0.98 1.6 1.3 1.4 1.4 1.0 0.98 1.4
17 GHz 1.3 1.8 1.7 1.4 1.4 1.4 1.2 1.4
18 GHz 1.7 2.1 2.2 1.4 1.4 1.9 1.6 1.4
18.5 GHz 1.6 1.7 1.6
19 GHz to 21.5 GHz in 0.5 GHz steps
1.7 1.7
1.7
22 GHz 1.6 1.7 1.6
22.5 GHz 1.7 1.8 1.7
23 GHz to 25.5 GHz in 0.5 GHz steps
1.8 1.9
1.8
26 GHz and 26.5 GHz
1.9 1.9 1.9
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 26 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CALIBRATION FACTOR (continued)
RF Power Sensors (continued)
Calibration and Measurement Capability in % calibration factor for the power sensor types described on Page 22
Frequency CMC Frequency CMC
0.05 GHz 0.6 24 GHz 2.1
0.1 GHz 0.7 25 GHz 2.2
0.3 GHz 0.8 26 GHz 2.2
0.5 GHz 0.8 26.5 GHz 2.3
1 GHz 0.8 27 GHz 2.5
2 GHz 0.8 28 GHz 2.6
3 GHz 0.8 29 GHz 2.7
4 GHz 0.8 30 GHz 2.9
5 GHz 0.8 31 GHz 3.0
6 GHz 0.9 32 GHz 2.9
7 GHz 0.9 33 GHz 2.7
8 GHz 0.9 34 GHz 2.5
9 GHz 1.0 34.5 GHz 2.5
10 GHz 1.0 35 GHz 2.4
11 GHz 1.0 36 GHz 2.4
12 GHz 1.0 37 GHz 2.5
12.4 GHz 1.0 38 GHz 2.5
13 GHz 1.0 39 GHz 2.5
14 GHz 1.1 40 GHz 2.4
15 GHz 1.1 41 GHz 2.5
16 GHz 1.2 42 GHz 2.6
17 GHz 1.2 43 GHz 2.6
18 GHz 1.2 44 GHz 2.6
19 GHz 1.9 45 GHz 2.6
20 GHz 1.9 46 GHz 2.6
21 GHz 1.9 47 GHz 2.6
22 GHz 1.9 48 GHz 2.6
23 GHz 2.0 49 GHz 2.6
50 GHz 2.6
Power Sensor Calibration System An additional single-box power sensor calibration system is available for calibration of certain models of power sensor. This system provides calibration of calibration factor, VSWR and, for suitable types of sensor, transition time. The CMCs associated with this system are shown overleaf.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 27 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
Power Sensor Calibration System - calibration factor
Frequency Calibration and Measurement Capability
Type N sensors 3.5 mm sensors 2.4 mm sensors
10 MHz 0.61 % 0.87 % 1.1 %
30 MHz 0.57 % 0.80 % 0.77 %
50 MHz reference reference reference
100 MHz 0.54 % 0.62 % 0.63 %
300 MHz 0.62 % 0.67 % 0.66 %
500 MHz to 1.5 GHz 0.62 % 0.67 % 0.68 %
2 GHz 0.62 % 0.69 % 0.68 %
3 GHz 0.63 % 0.70 % 0.68 %
4 GHz 0.63 % 0.75 % 0.75 %
5 GHz 0.70 % 0.75 % 0.75 %
6 GHz 0.72 % 0.76 % 0.74 %
7 GHz 0.74 % 0.76 % 0.74 %
8 GHz 0.72 % 0.98 % 0.97 %
9 GHz 0.77 % 0.98 % 0.97 %
10 GHz 0.78 % 0.98 % 0.97 %
11 GHz 0.78 % 0.99 % 0.97 %
12 GHz 0.79 % 1.0 % 0.97 %
12.4 GHz 0.82 % 1.0 % 1.0 %
13 GHz to 14 GHz 0.81 % 1.1 % 1.1 %
15 GHz 0.83 % 1.1 % 1.0 %
16 GHz 0.92 % 1.1 % 1.1 %
17 GHz 1.0 % 1.1 % 1.1 %
18 GHz 1.1 % 1.1 % 1.1 %
19 GHz to 26 GHz 1.5 % 1.5 %
26.5 GHz 1.6 % 1.6 %
27 GHz to 31 GHz 1.7 % 1.7 %
32 GHz 1.7 % 1.7 %
33 GHz 1.8 % 1.7 %
34 GHz to 40 GHz 1.7 %
41 GHz to 44 GHz 2.4 %
45 GHz to 48 GHz 2.5 %
49 GHz to 50 GHz 2.4 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 28 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
Power Sensor Calibration System - VSWR
Range and frequency Calibration and Measurement Capability (VSWR)
1.0:1 to 1.5:1 Type N sensors 3.5 mm sensors 2.4 mm sensors
10 MHz 0.0061 0.0055 0.011
30 MHz to 100 MHz 0.0061 0.0054 0.011
300 MHz to 500 MHz 0.0061 0.0054 0.0040
1 GHz to 2 GHz 0.0061 0.0035 0.0040
3 GHz to 5 GHz 0.0075 0.0057 0.0053
6 GHz to 7 GHz 0.0076 0.0057 0.0053
8 GHz 0.0075 0.0057 0.0053
9 GHz to 10 GHz 0.0081 0.0061 0.0053
11 GHz to 18 GHz 0.010 0.0074 0.0080
19 GHz to 20 GHz 0.0074 0.0080
21 GHz to 26.5 GHz 0.0098 0.013
27 GHz to 33 GHz 0.020 0.013
34 GHz to 40 GHz 0.013
41 GHz to 50 GHz 0.022
Power Sensor Calibration System - Linearity
At 50 MHz:
Reference level 0 dBm
+20 dBm 0.23 %
Type N, 3.5 mm and 2.4 mm sensors
+19 dBm to +16 dBm 0.22 %
+15 dBm to +10 dBm 0.21 %
+9 dBm 0.16 %
+8 dBm to +6 dBm 0.15 %
+5 dBm to -1 dBm 0.14 %
-2 dBm to -4 dBm 0.25 %
-5 dBm to -6 dBm 0.24 %
-7 dBm to -10 dBm 0.29 %
-11 dBm 0.26 %
-12 dBm to -14 dBm 0.32 %
-15 dBm to -21 dBm 0.31 %
-22 dBm 0.36 %
-23 dBm to -29 dBm 0.35 %
-30 dBm to -34 dBm 0.39 %
-35 dBm to -36 dBm 0.40 %
Power Sensor Calibration System - Transition time
Rise time 6 ns to 15 ns 0.30 ns N1921/2A and U2021/2XA sensors only. Fall time 6 ns to 15 ns 0.34 ns
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 29 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
RF VOLTAGE CMC at Stated Nominal voltage
Thermal voltage converter 0.5 V 1 V 3 V
Calibration of 50 Ω TVCs with BNC connectors. The uncertainty may be increased for other connector types
calibration 1 MHz to 2 MHz 0.060 % 0.060 % 0.060 %
5 MHz 0.10 % 0.10 % 0.10 %
7 MHz 0.10 % 0.10 % 0.13 %
10 MHz 0.10 % 0.10 % 0.13 %
20 MHz 0.13 % 0.13 % 0.17 %
30 MHz 0.15 % 0.16 % 0.21 %
40 MHz 0.24 % 0.24 % 0.32 %
50 MHz 0.30 % 0.29 % 0.32 %
60 MHz 0.36 % 0.36 % 0.42 %
70 MHz 0.51 % 0.51 % 0.44 %
80 MHz 0.51 % 0.51 % 0.47 %
90 MHz 0.76 % 0.56 % 0.47 %
100 MHz 0.76 % 0.56 % 0.47 %
Calibration of voltmeters and sources
0.25 V to 5 V Generation Measurement
(10 MHz max. for 3 V to 5 V)
20 Hz to 1 MHz 0.15 % 0.32 %
1 MHz to 10 MHz 0.31 % 0.39 %
10 MHz to 30 MHz 0.31 % 0.59 %
30 MHz to 60 MHz 0.50 % 0.70 %
60 MHz to 80 MHz 0.82 % 0.89 %
80 MHz to 100 MHz 0.82 % 1.3 %
5 V to 10 V
20 Hz to 3 MHz 0.16 % 0.16 %
3 MHz to 10 MHz 0.26 % 0.26 %
10 MHz to 1 GHz
5 mV to 10 mV 25 mV to 100 mV 0.25 V to 1.3 V 3.4 V to 5.5 V
1.0 % 1.2 % 1.0 %
1.3 %
Sine wave peak-to-peak voltage calculated from measured RF Power in 50 Ω.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 30 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
See P
ages 2
and 3
RF ATTENUATION
Specific Values Incremental attenuation
30 MHz and 50 MHz
60 MHz [1] For the calibration of spectrum analysers, receivers, network analysers, signal generators etc. [1] Includes step attenuators
10 dB 0.0020 dB 0.010 dB
20 dB 0.0020 dB 0.010 dB
30 dB 0.0030 dB 0.010 dB
40 dB 0.0030 dB 0.012 dB
50 dB 0.0040 dB 0.012 dB
60 dB 0.0040 dB 0.012 dB
70 dB 0.0050 dB 0.013 dB
80 dB 0.0050 dB 0.013 dB
90 dB 0.0050 dB 0.014 dB
100 dB 0.0050 dB 0.016 dB
110 dB 0.0060 dB 0.024 dB
120 dB 0.0060 dB -
30 dB
50 MHz 0.0070 dB For the calibration of Model 11708A fixed attenuators
Further attenuation capabilities are described in the VECTOR NETWORK ANALYSIS and the iPIMMS sections of this Schedule.
SPECTRUM ANALYSIS
Calibration of the spectral content, including harmonic distortion, of signal sources in a 50 Ω coaxial system.
0 dBc to -50 dBc
10 Hz to 3 GHz 1.4 dB
3 GHz to 20 GHz 1.8 dB
20 GHz to 40 GHz 2.0 dB
40 GHz to 50 GHz 2.3 dB
-50 dBc to -90 dBc
10 Hz to 3 GHz 1.8 dB
3 GHz to 20 GHz 2.1 dB
20 GHz to 40 GHz 2.3 dB
40 GHz to 50 GHz 2.5 dB
-90 dBc to -150 dBc
10 Hz to 3 GHz 4.0 dB
3 GHz to 6.6 GHz 4.3 dB
6.6 GHz to 22 GHz 4.6 dB
22 GHz to 26.5 GHz 5.0 dB
26.5 GHz to 31.1 GHz 4.5 dB
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 31 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
DIRECTIVITY of VRC bridges
The capability described below is for the measurement of directivity of 50 VRC bridges and similar devices. The uncertainties are given in linear quantities (VRC) where the range applies to the range of measured directivity. The values and uncertainties may be reported in terms of return loss (dB), calculated from the linear values and uncertainties.
Type N connectors 16 dB to 50 dB
10 MHz to 2 GHz 0.017 to 0.0060
2 GHz to 18 GHz 0.019 to 0.010
3.5 mm connectors 16 dB to 50 dB
10 MHz to 3 GHz 0.018 to 0.0060
3 GHz to 18 GHz 0.018 to 0.0080
18 GHz to 26.5 GHz 0.019 to 0.0080
7 mm connectors 16 dB to 50 dB
10 MHz to 18 GHz 0.017 to 0.0040
2.4 mm connectors 16 dB to 50 dB
10 MHz to 4 GHz 0.019 to 0.0090
4 GHz to 20 GHz 0.019 to 0.010
20 GHz to 36 GHz 0.021 to 0.013
36 GHz to 40 GHz 0.023 to 0.016
40 GHz to 50 GHz 0.026 to 0.020
PHASE NOISE
Signal analysers at 1 GHz
Offset from carrier Analyser phase noise
specification dBc/Hz
10 Hz > -80 0.51 dB
100 Hz > -107 0.51 dB
1 kHz > -126 0.40 dB
10 kHz > -140 0.37 dB
30 kHz > -142 0.36 dB
100 kHz > -144 0.53 dB
1 MHz > -153 0.54 dB
6 MHz > -155 0.54 dB
9.9 MHz and 10 MHz > -155 0.66 dB
Single-sided phase noise for signal sources
Minimum -150 dBc/Hz Carrier: 250 kHz to 26.5 GHz Offsets 1 Hz to 100 kHz
+2.1 dB; -∞ dB The quoted uncertainty may be increased depending on relative noise level between the source being calibrated and the specification of the reference source.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 32 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CISPR16 PULSE RESPONSE PRF
Quasi-peak detector response 100 Hz, 60 Hz, 10 Hz, 5 Hz, 2 Hz and 1 Hz 0.12 dB
Band A (9 kHz to 150 kHz)
20 Hz, 10 Hz, 2 Hz and 1 Hz 0.12 dB Band B (150 kHz to 30 MHz)
20 Hz, 10 Hz, 2 Hz and 1 Hz 0.12 dB Band C (150 kHz to 300 MHz)
20 Hz, 10 Hz, 2 Hz and 1 Hz 0.12 dB Band D (300 MHz to 1 GHz)
Quasi-peak to Peak detector relative response ratio
25 Hz 0.11 dB Band A
100 Hz 0.11 dB Bands B, C and D
Quasi-peak to Average detector relative response ratio
25 Hz 0.11 dB Band A
500 Hz 0.11 dB Band B
5 kHz 0.11 dB Bands C and D
Quasi-peak to RMS detector relative response ratio 25 Hz 0.11 dB Band A
1000 Hz 0.11 dB Bands B, C and D
Quasi-peak absolute amplitude 25 Hz 0.40 dB Band A
100 Hz 0.39 dB Band B
100 Hz 0.56 dB Band C
NOISE VOLTAGE AND CURRENT
Noise voltage (peak to peak) Bandwidth 20 MHz or 1 GHz
0.07 mV to 2 mV 0.070 mV
For PARD (ripple and noise) measurements on power supplies and similar devices.
2 mV to 2 V 2.0 %
Noise voltage (rms) Bandwidth 20 MHz
0.01 mV to 0.5 mV 21 %
0.5 mV to 1 mV 6.5 %
1 mV to 1 V 2.0 %
Probe noise current Bandwidth 20 MHz
Noise measurements on current probes.
0.01 mA to 0.5 mA 21 %
0.5 mA to 1 mA 6.5 %
1 mA to 1 A 2.0 %
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 33 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS
This section of the Schedule presents the CMCs for a range of vector network analysis systems. Unless otherwise indicated, each system provides measurements of reflection magnitude, reflection phase, transmission magnitude and transmission phase. The transmission magnitude capabilities are expressed in dB as this aspect of the capability will normally be used for the calibration of attenuators. Reflection magnitude is expressed in terms of voltage reflection coefficient (VRC) however this may be reported in terms of voltage standing wave ratio (VSWR) or return loss (dB). Measurements are made in a 50 Ω coaxial system, with the exception of the 8753ES Opt 075 system, which uses a 75 Ω coaxial system.
E8364B system (2.4 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 20 GHz 0.0053 to 0.0058
20 GHz to 40 GHz 0.0094 to 0.0098
40 GHz to 50 GHz 0.012
0.05 to 1
50 MHz to 20 GHz 0.0056 to 0.017
20 GHz to 40 GHz 0.0097 to 0.029
40 GHz to 50 GHz 0.012 to 0.035
Reflection phase 0° to ±180°
VRC 0.001 to 0.008
50 MHz to 50 GHz 180°
VRC 0.008 to 0.01
50 MHz to 20 GHz 180° to 59°
20 GHz to 50 GHz 180°
VRC 0.01 to 0.05
50 MHz to 20 GHz 59° to 10°
20 GHz to 50 GHz 180° to 20°
VRC 0.05 to 1
50 MHz to 20 GHz 10° to 1.2°
20 GHz to 50 GHz 20° to 3.1°
Transmission magnitude 0 dB to 40 dB
50 MHz to 50 GHz 0.028 dB to 0.19 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.067 dB to 3.3 dB
500 MHz to 50 GHz 0.064 dB to 0.47 dB
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 34 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
E8364B system (2.4 mm) 50 Ω system
(continued)
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.30° to 4.4°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.4° to 47°
Attenuation 0 dB to 70 dB
100 MHz to 2 GHz 0.33° to 3.3°
2 GHz to 20 GHz 0.45° to 4.0°
20 GHz to 50 GHz 2.7° to 9.1°
E8364B system (3.5 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 26.5 GHz 0.0034 to 0.0054
0.05 to 0.5
50 MHz to 26.5 GHz 0.0037 to 0.012
0.5 to 1
50 MHz to 26.5 GHz 0.0081 to 0.028
Reflection phase 0° to ±180°
VRC 0.001 to 0.005
50 MHz to 26.5 GHz 180°
VRC 0.005 to 0.01
50 MHz to 26.5 GHz 180° to 33°
VRC 0.01 to 0.05
50 MHz to 26.5 GHz 33° to 6.7°
VRC 0.05 to 1
50 MHz to 26.5 GHz 1.2° to 11°
Transmission magnitude 0 dB to 40 dB
50 MHz to 26.5 GHz 0.043 dB to 0.16 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.11 dB to 3.2 dB
500 MHz to 26.5 GHz 0.099 dB to 0.36 dB
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 35 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
E8364B system (3.5 mm) 50 Ω system
(continued)
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.46° to 4.4°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.5° to 47°
Attenuation 0 dB to 70 dB
100 MHz to 2 GHz 0.51° to 3.3°
2 GHz to 26.5 GHz 1.0° to 6.5°
E8364B system (7 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0020 to 0.0026
0.05 to 1
50 MHz to 18 GHz 0.0023 to 0.016
Reflection phase 0° to ±180°
VRC 0.001 to 0.003
50 MHz to 18 GHz 180°
VRC 0.003 to 0.01
50 MHz to 18 GHz 180° to 19°
VRC 0.01 to 0.05
50 MHz to 18 GHz 19° to 4.2°
VRC 0.05 to 0.3
50 MHz to 18 GHz 6.3° to 1.1°
VRC 0.3 to 1
50 MHz to 18 GHz 0.99° to 3.0°
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.037 dB to 0.16 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.10 dB to 3.2 dB
500 MHz to 18 GHz 0.094 dB to 0.22 dB
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 36 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
E8364B system (7 mm) 50 Ω system
(continued)
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.40° to 4.4°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.4° to 47°
Attenuation 0 dB to 70 dB
100 MHz to 18 GHz 0.44° to 4.0°
E8364B system (Type N) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0031 to 0.0061
0.05 to 1
50 MHz to 18 GHz 0.0033 to 0.026
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.008
50 MHz to 2 GHz 180° to 38°
2 GHz to 18 GHz 180°
VRC 0.008 to 0.009
50 MHz to 9 GHz 180° to 34°
9 GHz to 18 GHz 180°
VRC 0.009 to 0.02
50 MHz to 18 GHz 180° to 15°
VRC 0.02 to 0.05
50 MHz to 18 GHz 30° to 6.2°
VRC 0.05 to 1
50 MHz to 18 GHz 13° to 1.0°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 37 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
E8364B system (Type N) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.040 dB to 0.16 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.080 dB to 3.2 dB
500 MHz to 18 GHz 0.077 dB to 0.29 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.43° to 4.4°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.5° to 47°
Attenuation 0 dB to 70 dB
100 MHz to 18 GHz 0.47° to 4.2°
N5230C system (2.4 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 20 GHz 0.0053 to 0.0058
20 GHz to 40 GHz 0.0094 to 0.0098
40 GHz to 50 GHz 0.012
0.05 to 1
50 MHz to 20 GHz 0.0056 to 0.017
20 GHz to 40 GHz 0.0097 to 0.029
40 GHz to 50 GHz 0.012 to 0.035
Reflection phase 0° to ±180°
VRC 0.001 to 0.008
50 MHz to 50 GHz 180°
VRC 0.008 to 0.01
50 MHz to 20 GHz 180° to 58°
20 GHz to 50 GHz 180°
VRC 0.01 to 0.05
50 MHz to 50 GHz 180° to 10°
VRC 0.05 to 1
50 MHz to 20 GHz 10° to 1.2°
20 GHz to 50 GHz 26° to 3.1°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 38 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
N5230C system (2.4 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 50 GHz 0.028 dB to 0.25 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.11 dB to 2.6 dB
500 MHz to 50 GHz 0.082 dB to 3.5 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.31° to 4.0°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.4° to 34°
Attenuation 0 dB to 70 dB
100 MHz to 2 GHz 0.34° to 4.2°
2 GHz to 8 GHz 0.46° to 5.1°
8 GHz to 30 GHz 1.1° to 17°
30 GHz to 50 GHz 3.6° to 57°
N5230C system (3.5 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 26.5 GHz 0.0034 to 0.0054
0.05 to 0.5
50 MHz to 26.5 GHz 0.0036 to 0.012
0.5 to 1
50 MHz to 26.5 GHz 0.0082 to 0.028
Reflection phase 0° to ±180°
VRC 0.001 to 0.005
50 MHz to 26.5 GHz 180°
VRC 0.005 to 0.01
50 MHz to 26.5 GHz 180° to 33°
VRC 0.01 to 0.05
50 MHz to 26.5 GHz 33° to 6.7°
VRC 0.05 to 1
50 MHz to 26.5 GHz 1.2° to 13°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 39 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
N5230C system (3.5 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 26.5 GHz 0.043 dB to 0.17 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.11 dB to 2.6 dB
500 MHz to 8 GHz 0.097 dB to 0.41 dB
8 GHz to 26.5 GHz 0.10 dB to 1.2 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.47° to 4.1°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.5° to 34°
Attenuation 0 dB to 70 dB
100 MHz to 2 GHz 0.51° to 4.3°
2 GHz to 8 GHz 1.0° to 5.2°
8 GHz to 26.5 GHz 1.8° to 16°
N5230C system (7 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0020 to 0.0026
0.05 to 1
50 MHz to 18 GHz 0.0022 to 0.016
Reflection phase 0° to ±180°
VRC 0.001 to 0.003
50 MHz to 18 GHz 180°
VRC 0.003 to 0.01
50 MHz to 18 GHz 180° to 19°
VRC 0.01 to 0.05
50 MHz to 18 GHz 19° to 4.2°
VRC 0.05 to 0.3
50 MHz to 18 GHz 6.3° to 1.1°
VRC 0.3 to 1
50 MHz to 18 GHz 0.85° to 3.1°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 40 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
N5230C system (7 mm) (continued) 50 Ω system
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.038 dB to 0.20 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.11 dB to 3.1 dB
500 MHz to 8 GHz 0.098 dB to 0.48 dB
8 GHz to 18 GHz 0.075 dB to 1.4 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.42° to 4.9°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.9° to 44°
Attenuation 0 dB to 70 dB
100 MHz to 8 GHz 0.45° to 5.9°
8 GHz to 18 GHz 1.3º to 17º
N5230C system (Type N) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0031 to 0.0061
0.05 to 1
50 MHz to 18 GHz 0.0033 to 0.027
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.008
50 MHz to 2 GHz 180° to 38°
2 GHz to 18 GHz 180°
VRC 0.008 to 0.009
50 MHz to 8 GHz 180° to 34°
8 GHz to 18 GHz 180°
VRC 0.009 to 0.02
50 MHz to 18 GHz 180° to 15°
VRC 0.02 to 0.05
50 MHz to 18 GHz 30° to 6.1°
VRC 0.05 to 1
50 MHz to 18 GHz 13° to 1.0°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 41 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
N5230C system (Type N) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.041 dB to 0.17 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.073 dB to 2.6 dB
500 MHz to 8 GHz 0.072 dB to 0.41 dB
8 GHz to 18 GHz 0.11 dB to 1.1 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.45° to 4.1°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 2.5° to 34°
Attenuation 0 dB to 70 dB
100 MHz to 8 GHz 0.48° to 5.1°
8 GHz to 18 GHz 1.7° to 14°
8722D system (2.4 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 20 GHz 0.0053 to 0.0058
20 GHz to 40 GHz 0.0094 to 0.0098
0.05 to 1
50 MHz to 20 GHz 0.0055 to 0.019
20 GHz to 40 GHz 0.0097 to 0.031
Reflection phase 0° to ±180°
VRC 0.001 to 0.008
50 MHz to 40 GHz 180°
VRC 0.008 to 0.01
50 MHz to 20 GHz 180° to 58°
20 GHz to 40 GHz 180°
VRC 0.01 to 0.05
50 MHz to 40 GHz 180° to 10°
VRC 0.05 to 1
50 MHz to 20 GHz 10° to 1.4°
20 GHz to 40 GHz 20° to 4.4°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 42 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8722D system (2.4 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 40 GHz 0.046 dB to 0.25 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.091 dB to 5.5dB
500 MHz to 40 GHz 0.052 dB to 0.79 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.49° to 8.1°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 4.1° to 180°
Attenuation 0 dB to 70 dB
100 MHz to 4 GHz 0.49° to 8.0°
4 GHz to 40 GHz 1.1° to 11°
8722D system (3.5 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 26.5 GHz 0.0034 to 0.0053
0.05 to 0.5
50 MHz to 26.5 GHz 0.0036 to 0.013
0.5 to 1
50 MHz to 26.5 GHz 0.0079 to 0.030
Reflection phase 0° to ±180°
VRC 0.001 to 0.005
50 MHz to 26.5 GHz 180°
VRC 0.005 to 0.01
50 MHz to 26.5 GHz 180° to 33°
VRC 0.01 to 0.05
50 MHz to 26.5 GHz 33° to 6.7°
VRC 0.05 to 1
50 MHz to 26.5 GHz 1.4° to 13°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 43 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8722D system (3.5 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 26.5 GHz 0.061 dB to 0.25 dB
40 dB to 70 dB
50 MHz to 500 MHz 0.067 dB to 5.5dB
500 MHz to 26.5 GHz 0.067 dB to 0.84 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 500 MHz 0.65° to 8.2°
Attenuation 50 dB to 70 dB
500 MHz to 100 MHz 4.1° to 180°
Attenuation 0 dB to 70 dB
500 MHz to 7 GHz 0.74° to 5.7°
7 GHz to 26.5 GHz 1.7° to 12°
8722D system (7 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0020 to 0.0026
0.05 to 1
50 MHz to 18 GHz 0.0023 to 0.018
Reflection phase 0° to ±180°
VRC 0.001 to 0.003
50 MHz to 18 GHz 180°
VRC 0.003 to 0.01
50 MHz to 18 GHz 180° to 19°
VRC 0.01 to 0.05
50 MHz to 18 GHz 19° to 4.2°
VRC 0.05 to 0.3
50 MHz to 18 GHz 6.3° to 1.2°
VRC 0.3 to 1
50 MHz to 18 GHz 1.0° to 3.2°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 44 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8722D system (7 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 100 MHz 0.056 dB to 0.25 dB
100 MHz to 18 GHz 0.056 dB to 0.072 dB
40 dB to 70 dB
50 MHz to 100 MHz 0.38 dB to 5.5 dB
100 MHz to 18 GHz 0.079 dB to 0.62 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.60° to 8.1°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 4.1° to 180°
Attenuation 0 dB to 70 dB
100 MHz to 7 GHz 0.64° to 8.1°
7 GHz to 18 GHz 1.4° to 8.3°
8722D system (Type N) 50 Ω system
Reflection magnitude 0.001 to 0.05
50 MHz to 18 GHz 0.0031 to 0.0061
0.05 to 1
50 MHz to 18 GHz 0.0033 to 0.029
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.01
50 MHz to 18 GHz 180° to 30°
VRC 0.01 to 0.05
50 MHz to 18 GHz 30° to 6.1°
VRC 0.05 to 0.3
50 MHz to 18 GHz 6.1° to 1.6°
VRC 0.3 to 1
50 MHz to 18 GHz 1.2° to 4.2°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 45 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8722D system (Type N) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.058 dB to 0.25 dB
40 dB to 70 dB
50 MHz to 100 MHz 0.38 dB to 5.5 dB
100 MHz to 18 GHz 0.081 dB to 0.62 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 50 dB
50 MHz to 100 MHz 0.63° to 8.2°
Attenuation 50 dB to 70 dB
50 MHz to 100 MHz 4.1° to 180°
Attenuation 0 dB to 70 dB
100 MHz to 7 GHz 0.66° to 8.1°
7 GHz to 18 GHz 1.5° to 8.4°
E5071C system (Type N) 50 Ω system
Reflection magnitude 0.001 to 0.05
9 kHz to 8.5 GHz 0.0028 to 0.0091
0.05 to 0.1
9 kHz to 8.5 GHz 0.0031 to 0.0096
0.1 to 1
9 kHz to 8.5 GHz 0.0034 to 0.029
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
9 kHz to 8.5 GHz 180°
VRC 0.004 to 0.01
9 kHz to 8.5 GHz 180° to 27°
VRC 0.01 to 0.05
9 kHz to 8.5 GHz 27° to 5.7°
VRC 0.05 to 1
9 kHz to 8.5 GHz 18° to 1.2°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 46 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
E5071C system (Type N) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
9 kHz to 8.5 GHz 0.044 dB to 0.13 dB
40 dB to 70 dB
9 kHz to 300 kHz 0.12 dB to 2.1 dB
300 kHz to 8.5 GHz 0.085 dB to 0.33 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 40 dB
9 kHz to 300 kHz 0.54° to 1.3°
300 kHz to 1 GHz 0.47º to 0.99º
1 GHz to 8.5 GHz 0.56° to 2.2°
Attenuation 40 dB to 60 dB
9 kHz to 1 GHz 0.91° to 8.1°
1 GHz to 8.5 GHz 0.99° to 2.9°
Attenuation 60 dB to 70 dB
9 kHz to 1 GHz 1.6° to 26°
1 GHz to 8.5 GHz 1.6° to 4.3°
8753ES system (3.5 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
30 kHz to 6 GHz 0.0029 to 0.0047
0.05 to 1
30 kHz to 6 GHz 0.0032 to 0.023
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
30 kHz to 6 GHz 180°
VRC 0.004 to 0.01
30 kHz to 6 GHz 180° to 29°
VRC 0.01 to 0.05
30 kHz to 6 GHz 29° to 5.8°
VRC 0.05 to 1
30 kHz to 6 GHz 9.2° to 1.2°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 47 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8753ES system (3.5 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
30 kHz to 100 kHz 0.041 dB to 0.20 dB
100 kHz to 6 GHz 0.041 dB to 0.096 dB
40 dB to 70 dB
30 kHz to 100 kHz 0.20 dB to 4.3 dB
100 kHz to 6 GHz 0.070 dB to 0.84 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 40 dB
30 kHz to 6 GHz 0.44° to 2.1°
Attenuation 40 dB to 60 dB
30 kHz to 6 GHz 0.78° to 19°
Attenuation 60 dB to 70 dB
30 kHz to 100 kHz 19° to 180°
100 kHz to 6 GHz 1.9° to 9.9°
8753ES system (7 mm) 50 Ω system
Reflection magnitude 0.001 to 0.05
30 kHz to 6 GHz 0.0020 to 0.0023
0.05 to 1
30 kHz to 6 GHz 0.0022 to 0.011
Reflection phase 0° to ±180°
VRC 0.001 to 0.003
30 kHz to 6 GHz 180°
VRC 0.003 to 0.01
30 kHz to 6 GHz 180° to 19°
VRC 0.01 to 0.05
30 kHz to 6 GHz 19° to 4.1°
VRC 0.05 to 1
30 kHz to 6 GHz 4.9° to 0.82°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 48 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8753ES system (7 mm) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
30 kHz to 6 GHz 0.036 dB to 0.19 dB
40 dB to 70 dB
30 kHz to 100 kHz 0.19 dB to 4.3 dB
100 kHz to 6 GHz 0.065 dB to 0.84 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 40 dB
30 kHz to 6 GHz 0.38° to 2.0°
Attenuation 40 dB to 60 dB
30 kHz to 6 GHz 0.72° to 19°
Attenuation 60 dB to 70 dB
30 kHz to 100 kHz 19° to 180°
100 kHz to 6 GHz 1.9° to 9.9°
8753ES system (Type N) 50 Ω system
Reflection magnitude 0.001 to 0.05
30 kHz to 6 GHz 0.0025 to 0.0072
0.05 to 1
30 kHz to 6 GHz 0.0027 to 0.033
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
30 kHz to 6 GHz 180°
VRC 0.004 to 0.01
30 kHz to 6 GHz 180° to 24°
VRC 0.01 to 0.05
30 kHz to 6 GHz 180° to 5.0°
VRC 0.05 to 1
30 kHz to 3 GHz 7.0° to 0.84°
3 GHz to 6 GHz 14° to 3.2°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 49 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8753ES system (Type N) 50 Ω system
(continued)
Transmission magnitude 0 dB to 40 dB
30 kHz to 6 GHz 0.036 dB to 0.19 dB
40 dB to 70 dB
30 kHz to 100 kHz 0.19 dB to 4.3 dB
100 kHz to 6 GHz 0.065 dB to 0.84 dB
Transmission phase 0° to ±180°
Attenuation 0 dB to 40 dB
30 kHz to 6 GHz 0.39° to 2.0°
Attenuation 40 dB to 60 dB
30 kHz to 6 GHz 0.73° to 19°
Attenuation 60 dB to 70 dB
30 kHz to 100 kHz 19° to 180°
100 kHz to 6 GHz 1.9° to 9.9°
8753ES system (Type N) 75 Ω system
Reflection magnitude 0.001 to 0.05
30 kHz to 3 GHz 0.0027 to 0.0050
0.05 to 1
30 kHz to 3 GHz 0.0031 to 0.11
Reflection phase 0° to ±180°
VRC 0.001 to 0.004
30 kHz to 3 GHz 180°
VRC 0.004 to 0.01
30 kHz to 3 GHz 180° to 26°
VRC 0.01 to 0.05
30 kHz to 3 GHz 47° to 5.8°
VRC 0.05 to 1
30 kHz to 3 GHz 3.1° to 19°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 50 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8515A system (3.5 mm) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 26.5 GHz 0.0056 to 0.0091
0.1 to 1
50 MHz to 26.5 GHz 0.0065 to 0.044
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 26.5 GHz 180°
VRC 0.004 to 0.05
50 MHz to 26.5 GHz 180° to 7.2°
VRC 0.05 to 0.1
50 MHz to 26.5 GHz 15° to 3.7°
VRC 0.1 to 1
50 MHz to 26.5 GHz 1.2° to 10°
Transmission magnitude 0 dB to 40 dB
50 MHz to 26.5 GHz 0.053 dB to 0.19 dB
40 dB to 70 dB
50 MHz to 26.5 GHz 0.077 dB to 3.2 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 26.5 GHz 0.21° to 6.3°
40 dB to 70 dB
50 MHz to 26.5 GHz 0.75° to 31°
8510-8515A system (7 mm) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 18 GHz 0.0042 to 0.0054
0.1 to 1
50 MHz to 18 GHz 0.0049 to 0.025
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 51 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8515A system (7 mm) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.05
50 MHz to 18 GHz 180° to 5.4°
VRC 0.05 to 0.1
50 MHz to 18 GHz 9.1° to 2.9°
VRC 0.1 to 1
50 MHz to 18 GHz 0.73° to 6.5°
Transmission magnitude 0 dB to 40 dB
50 MHz to 20 GHz 0.048 dB to 0.084 dB
40 dB to 70 dB
50 MHz to 20 GHz 0.067 dB to 1.3 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 20 GHz 0.18° to 4.0°
40 dB to 70 dB
50 MHz to 20 GHz 0.72° to 13°
8510-8515A system (Type N) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 18 GHz 0.0048 to 0.010
0.1 to 1
50 MHz to 18 GHz 0.0056 to 0.042
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 52 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8515A system (Type N) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.05
50 MHz to 18 GHz 180° to 6.1°
VRC 0.05 to 0.1
50 MHz to 18 GHz 15° to 3.2°
VRC 0.1 to 1
50 MHz to 18 GHz 0.83° to 9.2°
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.051 dB to 0.11 dB
40 dB to 70 dB
50 MHz to 18 GHz 0.075 dB to 1.3 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 18 GHz 0.21° to 4.2°
40 dB to 70 dB
50 MHz to 18 GHz 0.73° to 13°
8510-8517A system (2.4 mm) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 50 GHz 0.0085 to 0.020
0.1 to 1
50 MHz to 50 GHz 0.0087 to 0.057
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 53 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8517A system (2.4 mm) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.008
50 MHz to 50 GHz 180°
VRC 0.008 to 0.05
50 MHz to 50 GHz 180° to 10°
VRC 0.05 to 0.1
50 MHz to 50 GHz 32° to 5.1°
VRC 0.1 to 1
50 MHz to 50 GHz 1.2° to 20°
Transmission magnitude 0 dB to 40 dB
50 MHz to 50 GHz 0.031 dB to 0.92 dB
40 dB to 70 dB
50 MHz to 40GHz 0.085 dB to 4.7 dB
40 GHz to 50 GHz 2.4 dB to 12 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 50 GHz 0.28° to 15°
40 dB to 60 dB
50 MHz to 50 GHz 1.1° to 83°
60 dB to 70 dB
50 MHz to 50 GHz 5.7° to 180°
8510-8517A system (3.5 mm) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 26.5 GHz 0.0057 to 0.0091
0.1 to 1
50 MHz to 26.5 GHz 0.0063 to 0.044
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 54 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8517A system (3.5 mm) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 26.5 GHz 180°
VRC 0.004 to 0.02
50 MHz to 26.5 GHz 180° to 18°
VRC 0.02 to 0.1
50 MHz to 26.5 GHz 31° to 3.7°
VRC 0.1 to 1
50 MHz to 26.5 GHz 1.1° to 11°
Transmission magnitude 0 dB to 40 dB
50 MHz to 26.5 GHz 0.058 dB to 0.19 dB
40 dB to 70 dB
50 MHz to 26.5 GHz 0.12 dB to 3.2 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 26.5 GHz 0.27° to 6.9°
40 dB to 60 dB
50 MHz to 26.5 GHz 1.1° to 10°
60 dB to 70 dB
50 MHz to 26.5 GHz 3.1° to 27°
8510-8517A system (7 mm) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 18 GHz 0.0043 to 0.0054
0.1 to 1
50 MHz to 18 GHz 0.0047 to 0.025
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 55 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8517A system (7 mm) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.05
50 MHz to 18 GHz 180° to 5.4°
VRC 0.05 to 0.1
50 MHz to 18 GHz 9.6° to 2.8°
VRC 0.1 to 1
50 MHz to 18 GHz 0.76° to 7.0°
Transmission magnitude 0 dB to 40 dB
50 MHz to 20 GHz 0.051 dB to 0.14 dB
40 dB to 70 dB
50 MHz to 20 GHz 0.079 dB to 3.2 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 20 GHz 0.22° to 4.3°
40 dB to 70 dB
50 MHz to 20 GHz 1.0° to 27°
8510-8517A system (Type N) 50 Ω system
Reflection magnitude 0.002 to 0.1
50 MHz to 18 GHz 0.0049 to 0.010
0.1 to 1
50 MHz to 18 GHz 0.0055 to 0.042
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 56 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
VECTOR NETWORK ANALYSIS (continued)
8510-8517A system (Type N) 50 Ω system
(continued)
Reflection phase 0° to ±180°
VRC 0.002 to 0.004
50 MHz to 18 GHz 180°
VRC 0.004 to 0.05
50 MHz to 18 GHz 180° to 6.1°
VRC 0.05 to 0.1
50 MHz to 18 GHz 15° to 3.1°
VRC 0.1 to 1
50 MHz to 18 GHz 0.85° to 9.3°
Transmission magnitude 0 dB to 40 dB
50 MHz to 18 GHz 0.055 dB to 0.14 dB
40 dB to 70 dB
50 MHz to 18 GHz 0.11 dB to 3.2 dB
Transmission phase 0° to ±180°
0 dB to 40 dB
50 MHz to 18 GHz 0.26° to 4.3°
40 dB to 70 dB
50 MHz to 18 GHz 1.0° to 27°
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 57 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Industry Primary Impedance Measurement System (iPIMMS)
iPIMMS is a measurement tool supplied and maintained by the UK’s National Physical Laboratory. Measurements are performed using a Microwave Network Analyser with enhancements and corrections being applied via NPL software. NOTES For the linear voltage reflection and transmission coefficient measurands (i.e. complex-valued S-parameters) described in this section of the schedule, the Calibration and Measurement Capability is shown as an interval of values, where a selected value within the interval represents an expanded uncertainty at a level of confidence of approximately 95%. Furthermore, a selected value within the interval will represent the uncertainty applied equally and simultaneously to both the Real and Imaginary parts of the S-parameter. The uncertainty value therefore defines a circular region of uncertainty, in the appropriate Sparameter plane, centred on the measured, quoted, mean value with a radius equal to the stated expanded uncertainty. For Voltage Reflection Coefficients (VRCs), the stated uncertainty is assumed here to be independent of the nominal |VRC|, so a single interval is presented applicable for all |VRC| in the range 0 ≤ |VRC| ≤ 1. For Voltage Transmission Coefficients (VTCs), the stated uncertainty is dependent on the nominal |VTC|, so uncertainty intervals are presented for selected, representative, values of |VTC| in the range 0 ≤ |VTC| ≤ 1. The measured values and uncertainties may also be reported as dB, VSWR, or VRC, with the angle reported in degrees or radians. Typical examples are presented in this schedule for clarification purposes.
Se
e P
age
s 2
an
d 3
iPIMMS LINEAR COMPLEX VOLTAGE REFLECTION COEFFICIENT
(VRC) IN 50 COAXIAL LINE
45 MHz to 18 GHz:
-1 Re(VRC) +1
-1 Im(VRC) +1 constrained by:
0 |VRC| 1
0.0020 to 0.0070 For devices fitted with 7 mm or Type-N connectors only.
45 MHz to 26.5 GHz:
-1 Re(VRC) +1
-1 Im(VRC) +1 constrained by:
0 |VRC| 1
0.0060 to 0.012 For devices fitted with 3.5 mm connectors.
iPIMMS LINEAR COMPLEX VOLTAGE TRANSMISSION COEFFICIENT
(VTC) IN 50 COAXIAL LINE
45 MHz to 18 GHz:
-1 Re(VTC) +1
-1 Im(VTC) +1 constrained by:
0 |VTC| 1
|VTC| = 0 |VTC| = 0.1 |VTC| = 1
0.0000050 to 0.000050 0.00015 to 0.00050 0.0015 to 0.0050
For devices fitted with 7 mm or Type-N connectors only.
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 58 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
iPIMMS LINEAR COMPLEX VOLTAGE TRANSMISSION COEFFICIENT
(VTC) IN 50 COAXIAL LINE
Typical examples expressed in terms of attenuation and transmission phase
Attenuation from 0 dB to 40 dB Attenuation at 50 dB Attenuation at 60 dB Attenuation at 70 dB Attenuation at 80 dB
0.013 dB to 0.043 dB 0.090° to 0.28° 0.014 dB to 0.14 dB 0.090° to 0.91° 0.043 dB to 0.42 dB 0.29° to 2.9° 0.14 dB to 1.3 dB 0.91° to 9.1° 0.42 dB to 3.5 dB 2.9° to 30°
For devices fitted with 7 mm or Type-N connectors only.
45 MHz to 26.5 GHz:
-1 Re(VTC) +1
-1 Im(VTC) +1 constrained by:
0 |VTC| 1
For devices fitted with 3.5 mm connectors.
|VTC| = 0 |VTC| = 0.1 |VTC| = 1
0.0000050 to 0.00010 0.00015 to 0.00050 0.0015 to 0.0050
Typical examples expressed in terms of attenuation and transmission phase
Attenuation from 0 dB to 30 dB Attenuation at 40 dB Attenuation at 50 dB Attenuation at 60 dB Attenuation at 70 dB Attenuation at 80 dB
0.013 dB to 0.043 dB 0.09° to 0.29° 0.014 dB to 0.086 dB 0.09° to 0.57° 0.014 dB to 0.27 dB 0.09° to 1.8° 0.043 dB to 0.83 dB 0.29° to 5.7° 0.14 dB to 2.4 dB 0.91° to 18° 0.42 dB to 6.0 dB 2.9° to 90°
For devices fitted with 3.5 mm connectors
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 59 of 60
Measured Quantity Instrument or Gauge
Range
Calibration and Measurement
Capability (CMC) Expressed as an
Expanded Uncertainty (k = 2)
Remarks
Locatio
n
Se
e P
age
s 2
an
d 3
CALIBRATION OF BATTERY TEST SYSTEMS
DC Voltage 1 V to 6 V 55 ppm to 18 ppm
6 V to 20 V 79 ppm to 30 ppm
20 V to 80 V 21 ppm
80 V to 600 V 18 ppm
600 V to 1000 V 20 ppm
DC Current 0 A to 60 A 29 mA
60 A to 160 A 50 ppm + 26 mA
160 A to 600 A 140 ppm + 35 mA
Temperature simulation -200 C to 0 C 0.075 C
PT100(385) equivalent resistances applied.
0 C to 100 C 0.090 C
100 C to 400 C 0.11 C
400 C to 630 C 0.13 C
630 C to 800 C 0.23 C
END
0147
Accredited to ISO/IEC 17025:2005
Schedule of Accreditation issued by
United Kingdom Accreditation Service 2 P ine Trees , Cher t sey Lane, S ta i nes -upon-Thames , TW 18 3HR, UK
Keysight Technologies UK Limited
Issue No: 066 Issue date: 23 July 2018
Calibration performed by the Organisation at the locations specified
Assessment Manager: CA Page 60 of 60
Appendix - Calibration and Measurement Capabilities Introduction The definitive statement of the accreditation status of a calibration laboratory is the Accreditation Certificate and the associated Schedule of Accreditation. This Schedule of Accreditation is a critical document, as it defines the measurement capabilities, ranges and boundaries of the calibration activities for which the organisation holds accreditation. Calibration and Measurement Capabilities (CMCs) The capabilities provided by accredited calibration laboratories are described by the Calibration and Measurement Capability (CMC), which expresses the lowest uncertainty of measurement that can be achieved during a calibration. If a particular device under calibration itself contributes significantly to the uncertainty (for example, if it has limited resolution or exhibits significant non-repeatability) then the uncertainty quoted on a calibration certificate will be increased to account for such factors. The CIPM-ILAC definition of the CMC is as follows: A CMC is a calibration and measurement capability available to customers under normal conditions: (a) as published in the BIPM key comparison database (KCDB) of the CIPM MRA; or (b) as described in the laboratory’s scope of accreditation granted by a signatory to the ILAC Arrangement. The CMC is normally used to describe the uncertainty that appears in an accredited calibration laboratory's schedule of accreditation and is the uncertainty for which the laboratory has been accredited using the procedure that was the subject of assessment. The CMC is calculated according to the procedures given in M3003 and is normally stated as an expanded uncertainty at a coverage probability of 95 %, which
usually requires the use of a coverage factor of k = 2. An accredited laboratory is not permitted to quote an uncertainty that is smaller than the
published CMC in certificates issued under its accreditation. The CMC may be described using various methods in the Schedule of Accreditation: As a single value that is valid throughout the range. As an explicit function of the measurand or of a parameter (see below). As a range of values. The range is stated such that the customer can make a reasonable estimate of the likely uncertainty at any point within the range. As a matrix or table where the CMCs depend on the values of the measurand and a further quantity. In graphical form, providing there is sufficient resolution on each axis to obtain at least two significant figures for the CMC. Expression of CMCs - symbols and units In general, only units of the SI and those units recognised for use with the SI are used to express the values of quantities and of the associated CMCs. Nevertheless, other commonly used units may be used where considered appropriate for the intended audience. For example, the term “ppm” (part per million) is frequently used by manufacturers of test and measurement equipment to specify the performance of their products. Terms like this may be used in Schedules of Accreditation where they are in common use and understood by the users of such equipment, providing their use does not introduce any ambiguity in the capability that is being described. When the CMC is expressed as an explicit function of the measurand or of a parameter, this often comprises a relative term (e.g., percentage) and an absolute term, i.e. one expressed in the same units as those of the measurand. This form of expression is used to describe the capability that can be achieved over a range of values. Some examples, and an indication of how they are to be interpreted, are shown below. DC voltage, 100 mV to 1 V: 0.0025 % + 5.0 μV: Over the range 100 mV to 1 V, the CMC is 0.0025 %∙V + 5.0 μV, where V is the measured voltage. Hydraulic pressure, 0.5 MPa to 140 MPa: 0.0036 % + 0.12 ppm/MPa + 4.0 Pa Over the range 0.5 MPa to 140 MPa, the CMC is 0.0036 %∙p + (0.12∙10-6∙p∙10-6) + 4.0 Pa, where p is the measured pressure in Pa. It should be noted that the percentage symbol (%) simply represents the number 0.01. In cases where the CMC is stated only as a percentage, this is to be interpreted as meaning percentage of the measured value or indication.
Thus, for example, a CMC of 1.5 % means 1.5 ∙ 0.01 ∙ i, where i is the instrument indication.