INVITED PAPER Microelectrodes, Microelectronics,and ......INVITED PAPER Microelectrodes,...
Transcript of INVITED PAPER Microelectrodes, Microelectronics,and ......INVITED PAPER Microelectrodes,...
INV ITEDP A P E R
Microelectrodes,Microelectronics, andImplantable NeuralMicrosystemsProgress in development of tiny electrodes, cables, circuitry, signal processors
and wireless interfaces promises to advance understanding of the
human nervous system and its disorders.
ByKensall D. Wise, Life Fellow IEEE, Amir M. Sodagar, Member IEEE, Ying Yao, Member IEEE,
Mayurachat Ning Gulari, Gayatri E. Perlin, and Khalil Najafi, Fellow IEEE
ABSTRACT | Lithographically definedmicroelectrode arrays now
permit high-density recording and stimulation in the brain and
are facilitating new insights into the organization and function of
the central nervous system. They will soon allow more detailed
mapping of neural structures than has ever before been possible,
and capabilities for highly localized drug-delivery are being
added for treating disorders such as severe epilepsy. For chronic
neuroscience and neuroprosthesis applications, the arrays are
being used in implantable microsystems that provide embedded
signal processing and wireless data transmission to the outside
world. A 64-channel microsystem amplifies the neural signals by
60 dB with a user-programmable bandwidth and an input-
referred noise level of 8 �Vrms before processing the signals
digitally. The channels can be scanned at a rate of 62.5 kS/s, and
signals above a user-specified biphasic threshold are transmitted
wirelessly to the external world at 2 Mbps. Individual channels
can also be digitized and viewed externally at high resolution to
examine spike waveforms. The microsystem dissipates 14.14 mW
from 1.8 V and measures 1.4 � 1.55 cm2.
KEYWORDS | BioMEMS; implant; microelectrodes; microstimu-
lation; microsystems; neural probes; neural recording
I . INTRODUCTION
Of all the parts that make up the human body, the nervous
system is by far the least understood and its disorders are the
most difficult to treat. Research on the use of electrical
stimulation to restore movement to paralyzed limbs goes
back at least 250 years to the work of Franklin [1] and
others, but it was midway through the last century thatphysiologists began to use electrical recording to try to
understand the nervous system at the cellular level [2].
Electrolytically sharpened metal wire electrodes [3] were
used for extracellular recording, while glass micropipettes
[4] were used to probe individual cells. Significant progress
was made in understanding the behavior of single neurons
and some sensory areas of the brain, but it became
increasingly clear that understanding the organization andsignal-processing techniques used at the system level would
require simultaneous recording and stimulation from many
sites having known spatial locations in tissueVsomething
that existing technology did not then permit.
From the 1950s onward, there was also growing
interest in developing electronic interfaces to the brain
for use in prosthetic devices for treating various disorders.
Work on visual prostheses began with Brindley’s earlyefforts [5], but by the late 1960s, a program to develop a
cortical prosthesis for the blind was also under way under
Dobelle at the University of Utah [6], [7]. Initial work on
cochlear prostheses for the deaf began about the same time
[8]–[10]. Most early work involved experiments using
wire-based stimulating electrode arrays, but it was quickly
recognized that better technology was needed, not only for
Manuscript received June 18, 2007; revised January, 30 2008. This work was
supported by the Engineering Research Centers Program, National Science
Foundation, under Award EEC-9986866.
The authors are with the Engineering Research Center for Wireless Integrated
MicroSystems, Department of Electrical Engineering and Computer Science,
The University of Michigan, Ann Arbor, MI 48109-2122 USA.
Digital Object Identifier: 10.1109/JPROC.2008.922564
1184 Proceedings of the IEEE | Vol. 96, No. 7, July 2008 0018-9219/$25.00 �2008 IEEE
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
forming the arrays themselves but also for the implantedmicrosystems that would be required for any practical
device. In addition to chronically implanted electrodes,
reliable leads, signal-processing circuitry, telemetry inter-
faces, and power sources would be needed. The required
technology did not exist at that time, but it was coming.
Microelectronics had taken a significant step forward
in 1959 with the invention of the lithographically based
planar process for integrated circuit fabrication, and by themid-1960s, the use of this process for fabricating
extracellular microelectrode arrays for neurophysiology
was under way [11]–[13]. This work drew on the selective
silicon etching technology being developed at Bell
Telephone Laboratories [14] and established the basis for
subsequent work on micromachined integrated sensors,
microelectromechanical systems (MEMS), and the micro-
systems of today. After 40 years, most of the technologyneeded for the realization of chronically implantable
neuroelectronic microsystems is now in place [15].
Advanced electrode design and wireless implantable
microsystems are hot topics today [16], with entire
conferences and journals devoted to them. Optimism is
being fueled by the spectacular successes achieved by
cochlear implants for the deaf and by deep brain
stimulation (DBS) in the subthalamic nucleus forParkinson’s Disease. Over 125 000 cochlear microsystems
have now been implanted worldwide, allowing profoundly
deaf patients to function normally in a hearing world, and
the use of DBS for the suppression of Parkinson’s tremor
has been remarkably successful, with very few side effects
when the electrodes are positioned correctly. Improved
prostheses for deafness and Parkinson’s Disease are beingdeveloped [17]–[19], along with new devices aimed at
treating blindness [20], [21], paralysis [22], [23], epilepsy
[24], and other disorders. Extracellular microelectrodes,
which provide the critical interface between the nervous
system and microelectronics, are discussed first below,
followed by a look at the present state-of-the-art in
implantable microsystems, including the interconnect
cables, site-interface circuitry, embedded signal processor,and wireless link to the external world.
II . MICROELECTRODE TECHNOLOGY
Extracellular microelectrodes record the voltages produced
by ionic current flow around neurons as their cell
membranes depolarize as the result of inputs received
from other cells. These neural Bspike[ potentials representthe electrical half of an electrochemical system, with
amplitudes as high as several hundred microvolts and a
frequency content extending up to perhaps 10 kHz. Almost
40 years ago, authors questioned whether this ionic current
flow was effectively isotropic or whether it was constrained
to flow only in narrow clefts between the surrounding
nonneuronal cells [3], making the observation of a cell
discharge (Bsingle unit[) largely a matter of luck. Thisspeculation was partly a result of the observation that in a
single pass through tissue, an electrode typically Bsees[ onlya few neurons, far fewer than the number known to exist
anatomically. However, based on many years of recording
with multisite probes, there is no evidence that electrode-
cell coupling is anisotropic. Cortical neurons can typically
Fig. 1. Basic structure of a micromachined multielectrode probe for recording or stimulation in the central nervous system.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1185
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
be seen electrically above background noise as far as 100 �maway, and in areas such as the cochlear nucleus, where cells
are smaller, recording fields are 40–50 �m in extent. Thus,
in order to record simultaneously from essentially all
neurons within a single block of tissue, sites on 80–200 �mcenters are required. This is possible today using the
lithographic techniques developed for integrated circuits.
The use of integrated circuit technology to create dense
arrays of thin-film electrodes for single-unit recording inthe nervous system began at Stanford University in 1966.
The resulting neural probes [11]–[13] had the general
structure shown in Fig. 1, which is typical of virtually all
lithographically defined electrode arrays reported since
that time. The probe structure consists of a selectively
etched (Bmicromachined[) substrate, with conducting
leads insulated above and below by inorganic dielectrics
and recording/stimulating sites formed by an area ofexposed metal. In 1970, it was difficult to fabricate such
devices reproducibly, but as MEMS technology developed
in the 1970s and 1980s, diffused boron etch-stops, reactive
ion etching (RIE), and silicon-on-insulator wafer technol-
ogy allowed probes to be realized reproducibly with high
yield using single-sided processing of wafers having normal
thickness [25].
As electrode technology developed, a wide variety ofmaterials were explored for use as the probe substrate
[26]–[30], including silicon, metals, glass, sapphire, and
polymers. While some of the resulting devices recorded
from neurons acutely, shaping the substrate, the strength
of that substrate [31], and the lack of suitable encapsulat-
ing dielectrics for chronic use were all important
problems. Most present approaches to probe formation
use silicon substrates [32]–[35]. Silicon can be shaped witha precision that probably exceeds that of any other material
[36] and taps directly into a plethora of processing
knowledge and equipment developed for the microelec-
tronics industry. It is compatible with the use of high-
quality silicon dioxide and silicon nitride dielectrics along
with some polymers, as well as with the formation of on-
chip circuitry [37]. It is also compatible with the
realization of probe structures that are very small (a fewmicrometers or even submicrometers in width).
The probe substrate must be realizable in reasonable
volumes with high yield, and that implies the ability to
define it lithographically in a batch process. Since silicon
wafers are typically 500–800 �m thick, it is important that
the probe shanks be formed using some form of etch stop,
either as a diffused-boron layer [25] or a buried-oxide layer
[34], [35]. Either approach can be effective, althoughboron etch-stops permit the use of multiple substrate
thicknesses and produce a smooth surface that is probably
helpful in minimizing tissue damage (Fig. 2). The boron
etch-stop also avoids any possibility of silicon substrate
dissolution in-vivo. It should be noted that not all silicon
electrode arrays are formed lithographically in a planar
process. The two-dimensional silicon depth arrays devel-
oped at the University of Utah [38], [39] are based on glass
reflow, sawing, and etching. They have produced some ofthe longest chronic implants reported to date.
Using silicon as the substrate and thermally grown
silicon dioxide (fused quartz) and low-pressure chemical-
vapor-deposited (LPCVD) silicon nitride as the insulating
dielectrics, conductors from the sites to bonding pads,
cables, or circuitry at the rear of the probe can be formed
using refractory metals, metal silicides, or polysilicon. In
spite of its relatively high resistance (10 �/square),polysilicon is adequate for most recording electrodes. For
stimulating probes, however, lower resistance materials
are needed to reduce drive voltages and the resulting
Fig. 2. Scanning electron microscope views of the tips of neural
probes fabricated using a boron-etch stop for the silicon substrate.
(Top) Side view of the substrate only. (Middle) Perspective view of a
probe tip. (Bottom) The lateral boron diffusion has been removed
using an RIE field etch.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1186 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
voltage-stress on the encapsulating dielectrics. Typical cell
thresholds for stimulation are 10–20 �A, but stimulating
currents can reach 100 �A or more. Since the conducting
substrate below the leads and the extracellular fluid above
them both act as ground planes, interchannel crosstalk is
virtually negligible on such probes, and the small area of
the lithographically defined leads minimizes shunt capac-
itance so that signal attenuation is negligible as well. Siteimpedances are dominated by the series capacitance of the
metal-electrolyte double layer. Recording sites are usually
formed using gold or platinum, although anodically formed
iridium oxide [40], [41] is increasingly used. It produces
significantly lower recording impedances than other
materials and is essential for stimulating sites, permitting
more than 20 times the charge delivery to tissue than
platinum or gold at the same voltage [15], [42].Sites are thought to average the potential field seen in
tissue over their area. For recording from small cells, site
areas of 100 �m2 have often been used, although larger
areas of 300–400 �m2 have been more typical recently to
reduce thermal electrode noise, which is inversely related
to area. For stimulating sites, 1000 �m2 is suitable for
current levels of about �70 �A at supply voltages of �5 V,
limited by the spreading resistances in tissue near thesite [43].
Fig. 3 shows several neural probes on the back of a U.S.
penny. More than 7500 such devices have been supplied to
the neuroscience community by the University of
Michigan and have changed research directions in systems
neurophysiology. One of the additional benefits of using
silicon as the substrate material is that circuitry for site
Fig. 3. Several different probe designs shown on the back of a U.S. penny.
Fig. 4. A 64-site probe on a U.S. postage stamp. The sites are spaced
100 �m in depth with shanks on 200 �m centers. The probe is
configured for mounting in a 3-D array.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1187
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
selection, signal amplification, stimulus current genera-tion, command decoding, and self-test can be integrated in
the probe substrate itself. Fig. 4 shows a 64-site eight-
channel recording probe on a U.S. postage stamp [44],
[45], while Fig. 5 shows a 64-site four-channel stimulating
array [46]. Several such probes can be microassembled
[47] to form three-dimensional (3-D) electrode arrays, as
in the 256-site four-probe 16-channel array also shown in
Fig. 5. This array has sites on 400 �m centers in threedimensions. Three-dimensional arrays of recording and
stimulating probes are poised to launch a revolution in our
understanding of neural systems, allowing computer-
controlled mapping of the connections between differentareas of brain in a few minutes that would be impossible
using discrete wire electrodes.
Lithographically defined probes can also support more
localized studies at the cellular level. By omitting the
boron diffusion in selected areas of the substrate, back-
looking sites can be realized as well as top-looking sites and
combinations of the two. Fig. 6 shows a diagram and
photograph of an 18-�m-wide probe containing top, back,and double-sided sites separated in depth by 40 �m. The
5-�m-diameter back-looking sites are separated from the
conducting substrate by less than 2 �m and yet record
Fig. 5. A 64-site eight-channel stimulating probe with sites on 400 �m centers and with CMOS electronics for stimulus generation,
recording, and self-test. Below, four such probes are mounted in a micromachined silicon platform to form a 256-site 32-channel array.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1188 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
well, with the substrate appearing as an insulator to the
extracellular current [48]. Fig. 7 shows recordings made
with a 50-�m-wide probe having top-, double-, and back-
sided sites on 20 �m centers as it advances in guinea pig
inferior colliculus. Cells can clearly be handed off from one
site to the next, but the substrate here is screening some
cells from some sites. The double-sided sites average the
potentials seen on their two surfaces, but as shank width
Fig. 6. Diagram and backside view of an 18-�m-wide probe shank having top-, double-, and back-looking sites. The sites are
separated in depth by 40 �m.
Fig. 7. Recordings made with a 50-�m-wide neural probe having top-, double-, and back-sided sites on 20 �m centers as it advances
in depth through guinea pig inferior colliculus.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1189
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
decreases, screening becomes less pronounced. For explor-
ing questions in neuroscience using acute extracellular
recording and stimulation, the necessary electrode array
technology is here now and is sufficient to allow rapid
progress in our understanding of neural structures. The
challenges are to provide high-density two-dimensional(2-D) and 3-D electrode arrays commercially and to develop
ways of asking the right questions of the recorded data using
appropriate algorithms embedded in smart multichannel
data acquisition systems.
For studies of plasticity and learning and for practical
neural prostheses based on recording, viable interfaces
with tissue for months, years, or even decades will be
required. Long-term (years) stimulation and shorter term(weeks) recording is possible now, but single-unit
recording for years will require further progress at the
probe–tissue interface. Probes implanted chronically are
encapsulated by a layer of glia a few micrometers thick.
Stimulating currents can penetrate this layer, but it can
shield small recording sites from extracellular potentials
that may be present. A variety of approaches are being
explored to increase recording life in-vivo from weeks ormonths to years [49]. These include coating the sites with
materials that prevent protein adsorption or attract
neuronal growth along with forming projections on the
sites that extend beyond the glial sheath.
Tresco [50] has also found that 100–150 �m-wide probe
shanks produce a 30–40% cell loss within 100 �m of the
shank (throughout the recording field of a typical site) and
elevated levels of astrocytes considerably beyond this. Onepossible explanation for these far-field effects is mechanical
motion of the probe in tissue, which is exacerbated by any
tethering of the probe to the skull. Designing probes with
shank widths that approach cellular dimensions and with
back-ends that fold over flat on the cortical surface [51] to
allow the dura to be replaced over the implant is one
approach to reducing such motion. Ultimately, scaling probe
size to smaller dimensions will be limited by strength rather
than technology, especially if penetrating the pia arachnoid
is required. Fig. 8 shows a probe whose shank consists of a10-�m-wide lattice of supporting silicon but which is
otherwise open, allowing cellular processes to regrow
through the shank. Histological studies are now under
way to determine the efficacy of such structures. The probes
still have sufficient strength for cortical use.
Fig. 8. Back side of a multisite lattice recording probe on a human hair, with an inset of the tip of the probe.
Fig. 9. Diagram of a probe containing devices for both electrical
and chemical recording and stimulation.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1190 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
Fig. 10. View of a 50-�m-wide probe shank containing a microchannel and an integrated thermal flowmeter. Insets show two such
probes on a U.S. penny and the cross-section of a channel before etch-back to form the channel wall [57].
Fig. 11. Neural spike counts in the presence of ten 100 ms injections of AMPA (an excitatory drug) delivered over a 10-s period [57].
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1191
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
Considerable progress has been made recently in
addressing the chemical aspects of the electrochemical
neural system, with the goal of realizing probes that, in
addition to stimulating/recording electrodes, have a
complete set of microfluidic components: drug delivery
channels, flowmeters, valves, and chemical sensors. Fig. 9
shows the diagram of such a probe. Probes having fluidicdrug-delivery channels [52], shutters [53], and flowmeters
[54], [55] have been demonstrated, and integrated
microvalves [56] are being developed. Fig. 10 shows
probes having drug-delivery channels and integrated
thermal flowmeters. Such probes can resolve flows below
200 pL/s, have thermal time constants of about 200 �s,and are vacuum-shielded to limit any temperature rise in
tissue to less than 1 �C. Fig. 11 shows 10 �M AMPA, anexcitatory neural agonist, being delivered in ten 100-ms-
long pulses over a 10-s interval [57]. The neural spike rate
increases after delivery, gradually returning to pre-AMPA
levels. The use of voltammetry to measure in-vivo chemical
concentrations (e.g., dopamine) has also been recently
demonstrated [58]. By combining electrical recording and
stimulation with chemical recording and stimulation, new
windows are being opened in neuropharmacology, with
new prospects for the treatment of disorders such as severe
epilepsy by providing therapies at point of need.
Fig. 12. A neural microsystem consisting of 2-D and 3-D arrays of cortically implanted electrodes with ribbon cables connecting
them to a subcutaneous electronics package containing circuitry for amplification, spike detection and encoding,
and the wireless transmission of power and bidirectional data.
Fig. 13. Block diagram of a wireless implantable microsystem for
cortical single-unit recording along with its external user interface.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1192 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
III . FULLY IMPLANTABLE NEURALMICROSYSTEMS
Chronic recording and/or stimulation in neuroscience and
in neural prostheses requires more than electrodes.Electronic site selection, stimulus generation, signal
amplification, spike detection and encoding, and the
wireless transmission of power and bidirectional data are
also required, in addition to any microfluidic functions
needed. Fig. 12 shows one possible configuration for such a
system, where two- and three-dimensional electrode arrays
interface with the cortex and are connected to a
subcutaneous electronics package using ultraflexible rib-bon cables. Within this package, the neural signals are
amplified, interpreted, and wirelessly interfaced to the
outside world. The remote location of the electronics with
respect to the cortex here allows more flexibility in power
dissipation since, in cortex, any temperature rise must be
less than 2 �C to avoid damage. It also allows better
shielding of the front-end neural signals from the wireless
power link and permits satellite platforms for antennas orfluidic reservoirs when needed. The tissue displaced by the
electrode array must be small enough to avoid any
significant disruption of the physiological system, and its
vertical rise above the cortical surface should be less
than 1 mm to allow the array to remain free of the skull
and float in tissue. Based on this implant architecture, the
various components of the microsystem will now be
discussed.
A. Interconnect Cables and Site-Interface CircuitryThe recent integration of parylene films, deposited and
patterned at wafer level prior to probe release, into the
probe process [59] is important not only because it allows
the use of parylene over the upper surface of the probe butalso because it permits the fabrication of parylene ribbon
cables as an integral part of the probe structure. Such
cables are more robust than silicon alternatives [60] and
potentially solve the mechanical interconnect problems
between the electrodes and the electronics package.
Accessing a few leads from an acute electrode array is
relatively simple, but for more than 30–40 sites, the back-
ends of such probes become prohibitively large. Forchronic implants, the situation is even worse, and since the
first silicon probes were produced in the 1960s, the
integration of circuitry on the probe itself has been an
important goal. The primary reason for on-chip circuitry is
to reduce the number of external leads (through site
selection and/or multiplexing) and permit the use of more
sites, but other benefits are to reduce impedance levels,
increase signal amplitudes, and decrease sensitivity tocable leakage.
The design of on-chip amplifiers for neural probes has
progressed from open-loop designs [44], [61] to capaci-
tively coupled closed-loop designs [62], [63]. Reported
amplifiers provide closed-loop gains of 1000, equivalent
input noise levels of 8 �Vrms, an upper cutoff frequency of
10 kHz, and lower cutoffs below 100 Hz [62]. Program-
mable low-frequency cutoffs have recently been demon-strated [64] to include or reject slow-wave activity,
depending on the application. State-of-the-art amplifiers
[45] dissipate 75 �W from �1.5 V and fit in an area of
0.07 mm2 in 0.5 �m complementary metal–oxide–
semiconductor (CMOS). Recognizing that most sites will
not be near neurons of interest, a front-end site selector
allows the user to choose from a large number of sites on
Fig. 14. Block diagram of a simple multichannel neural recording system.
Fig. 15. Functional block diagram of a 32-channel digital spike
detector [76].
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1193
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
the probe (electronic site positioning). The external probe
interface then typically requires seven leads (three power,
data in, data out, clock, and strobe), independent of the
number of sites. For a stimulating probe, the desired
current amplitudes and site addresses are entered serially,
and the generated currents are steered to the sites through
Fig. 16. Functional block diagram of the signal processor [77].
Fig. 17. The formation of data packets by time-division multiplexing the spike detector outputs.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1194 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
the site selector while providing recording access to anysite [46], [65]. Microsystems using both active (multi-
plexed and nonmultiplexed) and passive electrode arrays
have been reported, but as these systems evolve, more will
likely incorporate on-chip circuitry to reduce lead counts
and improve system reliability.
B. Neural Signal ProcessorsFig. 13 shows a block diagram of the complete
microsystem required for wireless cortical recording,
either for neuroscience or for neuroprosthesis applica-
tions. After amplification of the neural signals, here using16- or 64-channel amplifier chips housed in the
electronics package, the signals are fed to a neural signal
processor (NPU). In its simplest form, such a processor
could be realized by a simple time-domain analog
multiplexer as in Fig. 14. This approach keeps the system
complexity low as long as the number of recording
channels is small. For instance, the three-channel analog
recording system reported in [66] easily fits on a 2.2 �2.2 mm2 chip in 1.5 �m CMOS technology. As soon as the
number of channels grows, however, the amount of neural
Fig. 18. (a) Time-domain data compression in an eight-channel spike detector module by using a local memory and (b) forming the
outgoing data packets out of the neural activity retrieved from four such modules.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1195
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
information to be wirelessly transmitted will increase tosuch an extent that it cannot be easily handled. Thus, the
signal processor in Fig. 13 must take a more active role in
processing and compressing the recorded data. Much of
the emphasis in future generations of implantable
microsystems will focus on improving both the quantity
and quality of neural recordings. Work on the former will
focus primarily on increasing the number of channels,
while quality will focus on preserving as many features ofthe neural signal as possible, starting from the spike width
above a user-supplied threshold and progressing to a high-
resolution digital record of the spike waveform. In this
evolution, tradeoffs among system capability, power, and
size will be important. Sophisticated digital signal-
processing (DSP) algorithms for spike detection and
classification [67]–[73] require too much power and size at
present, so reported spike detectors have used user-programmable thresholds to provide information on spike
occurrence. This is adequate for some applications [74], [75]
and significantly reduces the amount of transmitted data.
A simplified block diagram of a mixed-signal neural
processor [76] is shown in Fig. 15. It receives four time-
multiplexed analog signal channels, each containing the
neural signals recorded on eight sites. After 5-bit analog-to-
digital conversion, the resulting signal amplitudes areseparated into 32 individual channels. A special-purpose
digital spike processor then computes the signal averages
and standard deviations for the individual channels and
calculates appropriate biphasic thresholds based on a user-
specified number of standard deviations. The amplitude of
each channel is then compared to its threshold to detectneural spikes. If the channel is active (contains a spike), the
sample is tagged with the associated channel address and
put in a buffer to be transmitted wirelessly at 2.5 Mbps.
This NPU provides full waveform information on neural
spikes above threshold, ignores subthreshold noise, saves a
factor of about 12 in output bandwidth, and increases the
number of allowable channels from 25 to 312.
The 32-channel signal processor [77] shown in Fig. 16supports two operational modes.
• In Scan Mode, all neural channels are scanned for
the occurrence of neural spikes. The addresses of
the active channels, i.e., channels with above-
threshold neural activity, are sorted, packed, and
sent to the outside world through a reverse
(outgoing) telemetry link. The threshold polarity
(positive, negative, or biphasic) and its level are setby the user for each channel individually.
• In Monitor Mode, a neural channel is selected,
sampled at high resolution, and transmitted to the
outside world.
This processor is equipped with additional circuitry
that allows it to be used as a 32-channel module in a 64-
channel neural processing architecture. Intrachip modu-larity (implemented by using eight-channel spike detectormodules (SD-8) in parallel) helps achieve high channel
scan rates. The number of channels handled by each
module, the number of parallel modules, and the size of
the local memory on each module are parameters that
can be optimized according to the targeted recording
Fig. 19. Data packaging [80]: (a) Incoming data packet, (b) outgoing data packet in Scan Mode, and (c) outgoing data packet in Monitor Mode.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1196 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
capacity and speed. Interchip modularity allows expansionof the system by simply configuring one processor as the
master and another one (or more) processor(s) as the
slave(s).
C. Data CompressionPerhaps the simplest scheme for sending the detected
neural activity to the outside world is to use time-division
multiplexing, as illustrated in Fig. 17. The binary outputsof all the spike detectors are periodically scanned and
sent to the external interface after data packet formation
[78]. However, since the extracellularly recorded action
potentials from cortical neurons have typical durations of
approximately 1 ms with firing rates from G 1 to 150
spikes per second [79], the output of each spike detector
contains useful information (neural activity) only a small
Fig. 20. Wireless transfer of data packets along with a synchronized clock [80]: (a) functional block diagram and
(b) actual operating waveforms.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1197
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
portion of time. Hence, transmitting the outputs of all thespike detectors (independent of their contents) wastes
significant recording bandwidth. A more efficient ap-
proach is to transmit only spike activity information
from active channels [77]. Here, when a spike is
detected on a given channel, the address of that channel
is sent to the external host. As shown in Fig. 16, the
local memory in each SD-8 module temporarily stores
the active channel addresses. The memory space is sharedamong all eight channels. The data fusion core polls the
four SD-8 modules to fetch the active channel addresses,
completing the process of spike detection, channel-
address tagging, spike sorting, and spike queuing. The
addresses retrieved from the four parallel queues will then
be used to make an outgoing data packet, as illustrated in
Fig. 18.
D. Bidirectional Wireless InterfacesImplantable recording microsystems need to be
powered and programmed through a forward telemetry
link in order to perform long-term wireless recording via a
reverse wireless path, as shown in Fig. 13. To support
these requirements, a major building block in such
systems is a bidirectional wireless interface that contains
at least two parts:/ a forward telemetry front-end that retrieves an
incoming stream of data packets containing setup
commands and data and provides the microsystem
with regulated supply voltages derived from the
modulated inductively coupled radio-frequency
carrier.
/ a reverse telemetry back-end that prepares the out-
going stream of neural data packets, and transmitsthe recorded information to the external world.
E. Data Transfer and PackagingWireless digital data transfer between the implantable
microsystem and the external host is basically asynchro-
nous, so to simplify detection and minimize power, a
synchronized clock is embedded in the data stream. To
facilitate recognition of a data packet and help the receiverseparate it from the preceding one, a start pulse (usually a
predefined pattern of 0s and 1s) is sent ahead of each
packet, and parity bits are added to each packet to allow
error detection. Fig. 19 shows incoming and outgoing data
packets [80], designed to program the microsystem and
transmit the recorded neural information, respectively. In
the incoming data packet, B0 and B1 encode four
commands, and B2–B9 carry the channel address ofinterest for either spike detection setting or for Monitor
Mode operation, or to convey required settings to the
channel of interest for operation in Scan Mode. The
outgoing data packet format in Scan Mode reports the
detected neural activity on one channel per each SD-8
using four bits. Every 8 bits are accompanied by a parity
bit. In Monitor Mode, each data packet carries two
consecutive 8-bit amplitude samples, along with two paritybits. The chip address bit (CAB), seen in the outgoing data
packet in both modes, is used when two 32-channel neural
processors of the same type (shown in Fig. 16) are
connected in a master–slave configuration to realize a 64-
channel processor. The CAB represents the chip that has
Fig. 21. Wireless recording in Monitor Mode.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1198 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
prepared the data packet (B1[ for the master chip and B0[for the slave).
Fig. 20 presents the functional details of the reverse
telemetry link. First, the data packets are Manchester
encoded in order to incorporate the system clock. Then,
the encoded bit stream is on–off key (OOK) modulated
and is wirelessly transmitted to the external system. On
the external receiver side, after amplification, the reverse
process consisting of OOK demodulation and Manchesterdecoding is performed in order to retrieve the received
data and a synchronized clock.
F. Wireless RecordingWireless implantable recording microsystems have
recently been demonstrated [78], [80]. Fig. 21 shows in-
vivo recordings obtained in Monitor Mode from the guinea
pig auditory cortex using one of these systems [80]. TheBoriginal signal[ here is the amplified neural signal before
analog-to-digital conversion on the microsystem side, and
the Breconstructed signal[ is the retrieved signal after
passing through the wireless link, external receiver, and
postprocessing software in the host computer (see Fig. 13).
The signals shown in Fig. 21 are a small slice of a 30-s
recording in which two separate units were identified
based on postrecording analysis. For the first unit, thesignal/noise ratio (SNR) decreased from 11.21 on the
system side to 8.77 after recovery on the external side;
the second unit experienced an SNR reduction from 4.56
to 3.35 in passing through the same path. This degradation
in the signal quality comes from analog/digital conversion
with 8-bit resolution. Wireless data transfer does not
contribute to SNR reduction because digital data
modulation/transfer is used. Averaged over 24 886 datapackets, the packet error rate was 0.33%, which is excellent
performance for wireless data transfer in this application.
For a 2 MHz clock, the channel scan rate for spike
detection in this system is 62.5 kS/s and the total system
power dissipation at 1.8 V is 14.4 mW. The implantable
version of the microsystem measures 1.4 � 1.55 cm2,fitting on a U.S. penny.
IV. CONCLUSIONS
The technology for creating neural probes capable of
acutely measuring the neuronal activity throughout a
volume of tissue is now in place. Using three-dimensional
arrays of stimulating and recording electrodes, detailedmapping of connections in the nervous system should soon
be possible. Such studies will give important insights into
the signal-processing techniques used in the nervous
system and into the disorders that disrupt them. For
chronic studies, further advances are needed in the
electrode–tissue interface, and these needs are driving
electrode size toward cellular dimensions and below.
Beyond the electrodes themselves, substantial progresshas been made in the cables, site-selection and amplifica-
tion circuitry, embedded neural signal processors, and
wireless interfaces needed for chronic investigations in
neuroscience and for neural prostheses. The first com-
pletely implantable neural microsystems are now emerging
and should stimulate substantial progress in both areas. The
coming decade should see some dramatic breakthroughs in
our understanding of the nervous system and in our abilityto treat its disorders. h
Acknowledgment
The authors would like to thank Dr. F. Terry Hambrecht
and Dr. W. J. Heetderks of the Neural Prosthesis Program,
National Institute of Neurological Disorders and Stroke, for
their vision and support of this work over many years, aswell as the many faculty, students, and staff at the
University of Michigan who contributed to this research
in so many important ways. The support and encourage-
ment provided by P. V. Anderson of Cupertino, CA, is also
gratefully acknowledged.
REFERENCES
[1] B. Franklin, BAn account of the effects ofelectricity in paralytic cases,[ Phil. Trans.Lond., vol. L, pp. 481–483, 1975.
[2] M. Verseano and K. Negishi, BNeuronalactivity in cortical and thalamic networks,[ J.Gen. Physiol., vol. 43, pp. 177–195, 1960.
[3] D. A. Robinson, BThe electrical properties ofmetal microelectrodes,[ Proc. IEEE, vol. 56,pp. 1065–1071, Jun. 1968.
[4] O. F. Schanne, M. Lavallee, R. Laprade, andS. Gagne, BElectrical properties of glassmicroelectrodes,[ Proc. IEEE, vol. 56,pp. 1072–1082, Jun. 1968.
[5] G. S. Brindley and W. S. Lewin, BThesensations produced by electrical stimulationof the visual cortex,[ J. Physiol. (Lond.),vol. 196, pp. 479–493, 1968.
[6] W. H. Dobelle, M. G. Mladejovsky, andJ. P. Girvin, BArtificial vision for the blind:Electrical stimulation of visual cortex offershope for a functional prosthesis,[ Science,vol. 183, pp. 440–444, Feb. 1974.
[7] BAn electronic link to the visual cortex may letblind Fsee_,[ Electronics, Dec. 20, 1974.
[8] F. B. Simmons, C. J. Mongeon, W. R. Lewis,and D. A. Huntington, BElectrical stimulationof acoustical nerve and inferior colliculus:Results in man,[ Arch. Otolaryngol., vol. 79,pp. 559–567, 1964.
[9] F. A. Spelman, BThe past, present, and futureof cochlear prostheses,[ IEEE Eng. Med. Biol.Mag., pp. 27–33, May 1999.
[10] W. F. House and K. I. Berliner, BCochlearimplants: From idea to clinical practice,[ inCochlear Implants: A Practical Guide,H. Cooper, Ed. San Diego, CA: Singular,1991, pp. 9–33.
[11] K. D. Wise, J. B. Angell, and A. Starr, BAnintegrated circuit approach to extracellularmicroelectrodes,[ in Dig. 8th Int. Conf. Eng.Med. Biol., Jun. 1969, p. 14.5.
[12] K. D. Wise and J. B. Angell, BAmicroprobe with integrated amplifiers forneurophysiology,[ in Dig. IEEE Int. Solid-StateCircuits Conf., Philadelphia, PA, Feb. 1971,pp. 100–101.
[13] K. D. Wise and J. B. Angell, BAlow-capacitance multielectrode probe forneurophysiology,[ IEEE Trans. Biomed. Eng.,vol. BE-22, pp. 212–219, May 1975.
[14] M. P. Lepselter, BBeam-lead technology,[Bell Syst. Tech. J., vol. 45, pp. 233–254,Feb. 1966.
[15] K. D. Wise, D. J. Anderson, J. F. Hetke,D. R. Kipke, and K. Najafi, BWirelessimplantable microsystems: Electronicinterface to the nervous system,[ Proc. IEEE(Special Issue on Biomedical Applications forMEMS and Microfluidics), pp. 76–97,Jan. 2004.
[16] R. Harrison and K. D. Wise, BImplantable andprosthetic devices: Life-changing circuits,[ inDig. IEEE Int. Solid-State Circuits Conf.,San Francisco, CA, Feb. 2007, pp. 376–377.
[17] J. Wang, M. Gulari, and K. D. Wise, BAnintegrated position-sensing system for aMEMS-based cochlear implant,[ in Dig. IEEEElectron Devices Meeting, Washington, DC,Dec. 2005, pp. 129–132.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1199
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
[18] P. T. Bhatti, S. Lee, and K. D. Wise, BA 32-site4-channel cochlear electrode array,[ IEEE J.Solid-State Circuits, pp. 2965–2973,Dec. 2006.
[19] P. Limousin, P. Krack, P. Pollack,A. Benazzouz, C. Ardouin, D. Hoffmann, andA. Benabid, BElectrical stimulation of thesubthalamic nucleus in advanced Parkinson’sdisease,[ New Eng. J. Med., vol. 339,pp. 1105–1111, 1998.
[20] J. D. Weiland and M. S. Humayun, BAbiomimetic retinal stimulating array,[ IEEEEng. Med. Biol. Mag., vol. 24, pp. 14–21,Sep. 2006.
[21] E. Margalit, M. Maia, J. D. Weiland,R. J. Greenberg, G. Y. Fujii, G. Torres,D. V. Piyathaisere, T. M. O’Hearn, W. Liu,G. Lazzi, G. Dagnelie, D. A. Scribner,E. de Juan, Jr., and M. S. Humayun, BRetinalprosthesis for the blind,[ Surv. Ophthal.,vol. 47, pp. 335–356, 2002.
[22] D. M. Taylor, S. I. Tillery, and A. B. Schwartz,BDirect cortical control of 3-Dneuroprosthetic devices,[ Science, vol. 296,pp. 1829–1832, Jun. 2002.
[23] L. R. Hochberg, M. D. Serruya, G. M. Friehs,J. A. Mukand, M. Saleh, A. H. Caplan,A. Branner, D. Chen, R. D. Penn, andJ. P. Donoghue, BNeuronal ensemble controlof prosthetic devices by a human withtetraplegia,[ Nature, vol. 442, pp. 164–171,Jul. 13, 2006.
[24] D. J. Anschel, E. L. Ortega, A. C. Kraus, andR. S. Fisher, BFocally injected adenoseneprevents seizures in the rat,[ Exp. Neurology,vol. 190, pp. 544–457, 2004.
[25] K. Najafi, K. D. Wise, and T. Mochizuki,BA high-yield IC-compatible multichannelrecording array,[ IEEE Trans. Electron Devices,vol. ED-32, pp. 1206–1211, Jul. 1985.
[26] G. A. May, S. A. Shamma, and R. L. White,BA tantalum-on-sapphire microelectrodearray,[ IEEE Trans. Electron Devices,vol. ED-26, pp. 1932–1939, Dec. 1979.
[27] M. Kuperstein and D. A. Whittington,BA practical 24-channel microelectrode forneural recording in-vivo,[ IEEE Trans. Biomed.Eng., vol. BE-28, pp. 288–293, Mar. 1981.
[28] O. J. Prohaska, F. Olcaytug, P. Pfundner, andH. Dragaun, BThin-film multiple electrodeprobes: Possibilities and limitations,[ IEEETrans. Biomed. Eng., vol. BE-33, pp. 223–229,Feb. 1986.
[29] N. A. Blum, B. G. Carkhuff, H. K. Charles,R. L. Edwards, and R. A. Meyer, BMultisitemicroprobes for neural recordings,[ IEEETrans. Biomed. Eng., vol. 38, pp. 68–74,Jan. 1991.
[30] P. J. Rousche, D. S. Pellinen, D. P. Pivin, Jr.,J. C. Williams, R. J. Vetter, and D. R. Kipke,BFlexible polyimide-based intracorticalelectrode arrays with bioactive capability,[IEEE Trans. Biomed. Eng., vol. 48,pp. 361–371, Mar. 2001.
[31] K. Najafi and J. F. Hetke, BStrengthcharacterization of silicon microprobes inneurophysiological tissues,[ IEEE Trans.Biomed. Eng., vol. 37, pp. 474–481, May 1990.
[32] T. Kewley, M. D. Hills, D. A. Borkholder,I. E. Opris, N. I. Maluf, C. W. Storment,J. M. Bower, and G. T. A. Kovacs,BPlasma-etched neural probes,[ Sens.Actuators A, vol. 58, pp. 27–35, 1997.
[33] T. H. Yoon, E. J. Hwang, D. Y. Shin, S. I. Park,S. J. Oh, S. C. Jung, H. C. Shin, and S. J. Kim,BA micromachined silicon depth probe formultichannel neural recording,[ IEEE Trans.Biomed. Eng., vol. 47, pp. 1082–1087,Aug. 2000.
[34] P. Norlin, M. Kindlundh, A. Mouroux,K. Yoshida, and U. G. Hofmann, BA 32-siteneural recording probe fabricated by DRIE ofSOI substrates,[ J. Micromech. Microeng.,vol. 12, pp. 414–419, 2002.
[35] K. C. Cheung, K. Djupsund, Y. Dan, andL. P. Lee, BImplantable multichannelelectrode array based on SOI technology,[IEEE J. Microelectromech. Syst., vol. 12,pp. 179–184, Apr. 2003.
[36] K. E. Petersen, BSilicon as a mechanicalmaterial,[ Proc. IEEE, vol. 70, pp. 420–457,May 1982.
[37] K. Najafi and K. D. Wise, BAn implantablemultielectrode array with on-chip signalprocessing,[ IEEE J. Solid-State Circuits,vol. SC-21, pp. 1035–1045, Dec. 1986.
[38] E. M. Maynard, C. T. Nordhausen, andR. A. Normann, BThe Utah intracorticalelectrode array: A recording structure forpotential brain-computer interfaces,[Electroenceph. Clin. Neurophys., vol. 102,pp. 228–239, 1997.
[39] R. A. Normann, D. J. Warren, J. Ammermuller,E. Fernandez, and S. Guillory,BHigh-resolution spatio-temporal mapping ofvisual pathways using multi-electrode arrays,[Vision Res., vol. 41, pp. 1261–1275, 2001.
[40] J. D. Weiland and D. J. Anderson, BChronicneural stimulation with thin-film, iridiumoxide electrodes,[ IEEE Trans. Biomed. Eng.,vol. 47, pp. 911–918, 2000.
[41] X. Beebe and T. L. Rose, BCharge injectionlimits of activated iridium oxide electrodeswith 0.2 ms pulses in bicarbonate bufferedsaline,[ IEEE Trans. Biomed. Eng., vol. 35,pp. 494–495, Jun. 1988.
[42] S. J. Tanghe, K. Najafi, and K. D. Wise, BAplanar IrO multichannel stimulating electrodefor use in neural prostheses,[ Sens. Actuators,vol. B1, pp. 464–467, Jan. 1990.
[43] C. Kim and K. D. Wise, BA 64-site multi-shankCMOS low-profile neural stimulating probe,[IEEE J. Solid-State Circuits, vol. 31,pp. 1230–1238, Sep. 1996.
[44] J. Ji and K. D. Wise, BAn implantableCMOS circuit interface for multiplexedmicroelectrode recording arrays,[ IEEE J.Solid-State Circuits, vol. 27, pp. 433–443,Mar. 1992.
[45] G. E. Perlin, A. M. Sodagar, and K. D. Wise,BNeural recording front-end designs for fullyimplantable neuroscience applications andneural prosthetic microsystems,[ in Dig. IEEEInt. Conf. Eng. Med. Biol., New York,Sep. 2006, pp. 2982–2985.
[46] M. D. Gingerich, J. F. Hetke, D. J. Anderson,and K. D. Wise, BA 256-site 3D CMOSmicroelectrode array for multipointstimulation and recording in the centralnervous system,[ in Dig. Int. Conf. Solid-StateSens. Actuators (Transducers’01), Munich,Germany, Jun. 2001.
[47] Q. Bai and K. D. Wise, BA high-yieldmicroassembly structure forthree-dimensional microelectrode arrays,[IEEE Trans. Biomed. Eng., pp. 281–289,Mar. 2000.
[48] G. E. Perlin and K. D. Wise, BThe effect of thesubstrate on the extracellular neural activityrecorded with micromachined siliconmicroprobes,[ in Dig. IEEE Conf. Eng. Med.Biol., San Francisco, CA, Sep. 2004,pp. 2002–2005.
[49] X. Cui, V. A. Lee, Y. Raphael, J. A. Wiler,J. F. Hetke, D. J. Anderson, and D. C. Martin,BSurface modification of neural recordingelectrodes with conducting polymer/biomolecule blends,[ J. Biomed. Mater. Res.,vol. 56, pp. 261–272, 2001.
[50] R. Biran, D. C. Martin, and P. A. Tresco,BNeuronal cell loss accompanies the braintissue response to chronically implantedsilicon microelectrode arrays,[ Exp.Neurology, Jul. 2005.
[51] Y. Yao, M. N. Gulari, S. Ghimire, J. F. Hetke,and K. D. Wise, BA low-profile three-dimensional silicon/parylene stimulatingelectrode array for neural prosthesisapplications,[ in Dig. IEEE Conf. Eng. Med.Biol., Shanghai, China, Sep. 2005.
[52] J. Chen, K. D. Wise, J. F. Hetke, andS. C. Bledsoe, Jr., BA multichannel neuralprobe for selective chemical delivery at thecellular level,[ IEEE Trans. Biomed. Eng.,vol. 44, pp. 760–769, Aug. 1997.
[53] D. Papageorgiou, S. C. Bledsoe, M. Gulari,J. F. Hetke, D. J. Anderson, and K. D. Wise,BA shuttered probe with in-line flowmetersfor chronic in-vivo drug delivery,[ in Dig.IEEE Microelectromech. Syst. Conf., Interlaken,Jan. 2001, pp. 212–215.
[54] D. Papageorgiou, S. Shore, S. Bledsoe, andK. D. Wise, BA shuttered probe with in-lineflowmeters for chronic in-vivo drug delivery,[IEEE J. Microelectromech. Syst., vol. 15,pp. 1025–1033, Aug. 2006.
[55] Y. Li, K. Baek, M. Gulari, and K. D. Wise, BAdrug-delivery probe with an in-line flowmeterbased on trench refill and chemicalmechanical polishing techniques,[ in IEEESensors Conf., Atlanta, GA, Oct. 2007.
[56] K. Baek, Y. Li, M. N. Gulari, and K. D. Wise,BA pneumatically-actuated microvalve forspatially-selective chemical delivery,[ in Dig.North Amer. Conf. Solid-State Sens., Actuators,Microsyst., Hilton Head, SC, Jun. 2006,pp. 155–158.
[57] Y. Li, BAn integrated drug-delivery probe withan in-line flowmeter,[ Ph.D. dissertation,Univ. of Michigan, Ann Arbor, 2006.
[58] M. D. Johnson, R. K. Franklin, K. A. Scott,R. B. Brown, and D. R. Kipke, BNeural probesfor concurrent detection of neurochemicaland electrophysiological signals in vivo,[ inProc. IEEE Eng. Med. Biol. Conf., Shanghai,China, Sep. 2005, pp. 7325–7328.
[59] J. Wang, M. N. Gulari, and K. D. Wise, BAparylene-silicon cochlear electrode array withintegrated position sensors,[ in Dig. IEEE Int.Conf. Eng. Med. Biol., New York, Sep. 2006,pp. 3170–3173.
[60] J. F. Hetke, K. Najafi, and K. D. Wise,BFlexible miniature ribbon cables forlong-term connection to implantablesensors,[ Sens. Acuators, vol. A23,pp. 999–1002, Apr. 1990.
[61] Q. Bai and K. D. Wise, BSingle-unit recordingwith active microelectrode arrays,[ IEEETrans. Biomed. Eng., pp. 911–920, Aug. 2001.
[62] R. H. Olsson, III, M. N. Gulari, andK. D. Wise, BA fully-integrated bandpassamplifier for extracellular neural recording,[in Proc. 1st Int. IEEE EMBS Conf. Neural Eng.,Capri, Italy, Mar. 2003, pp. 165–168.
[63] R. R. Harrison and C. Charles, BA low-powerlow-noise CMOS amplifier for neuralrecording applications,[ IEEE J. Solid-StateCircuits, vol. 38, pp. 958–965, Jun. 2003.
[64] R. H. Olsson, III, D. Buhl, A. M. Sirota,G. Buzsaki, and K. D. Wise, BBand-tunableand multiplexed integrated circuits forsimultaneous recording and stimulation withmicroelectrode arrays,[ IEEE Trans. Biomed.Eng., vol. 52, pp. 1303–1311, Jul. 2005.
[65] Y. Yao, M. N. Gulari, J. F. Hetke, andK. D. Wise, BA low-profile three-dimensionalneural stimulating array with on-chip currentgeneration,[ in Dig. IEEE Conf. Eng. Med. Biol.,San Francisco, CA, Sep. 2004, pp. 1994–1997.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1200 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
[66] P. Mohseni and K. Najafi, BWirelessmulti-channel biopotential recording using anintegrated FM telemetry circuit,[ IEEE Trans.Neural. Syst. Rehabil. Eng., vol. 13,pp. 263–271, Sep. 2005.
[67] C.-W. Ko, Y.-D. Lin, H.-W. Chung, andG.-J. Jan, BAn EEG spike detection algorithmusing artificial neural network withmulti-channel correlation,[ in Proc. IEEE Eng.Med. Biol. Conf., 1998, vol. 4, pp. 2070–2073.
[68] T. Kalayci and O. Ozdamar, BWaveletpreprocessing for automated neural networkdetection of EEG spikes,[ IEEE Eng. Med. Biol.Mag., vol. 14, pp. 160–166, Mar. 1995.
[69] Y. Oomori, F. Yano, and S. P. Ninomija, BAcomputer-assisted spike detecting systemusing normalized method for epilepsydiagnosis,[ in Proc. 21st Ann. Hawaii Int. Conf.Syst. Sci., 1988, vol. 4, pp. 51–54.
[70] L. S. Pon, M. Sun, and R. J. Sclabassi, BThebi-directional spike detection in EEG usingmathematical morphology and wavelettransform,[ in Proc. 6th Int. Conf. SignalProcess., 2002, vol. 2, pp. 1512–1515.
[71] S. Kim, J. McNames, and K. Burchiel, BActionpotential detection with automatic templatematching,[ in Proc. Int. Conf. Eng. Med. Biol.,San Francisco, CA, Sep. 2004, pp. 41–44.
[72] M. Adjouadi, D. Sanchez, M. Cabrerizo,M. Ayala, P. Jayakar, I. Yaylali, andA. Marreto, BInterictal spike detection usingthe walsh transform,[ IEEE Trans. Biomed.Eng., vol. 51, pp. 868–872, May 2004.
[73] Z. Nenadic and J. W. Burdick, BSpikedetection using the continuous wavelettransform,[ IEEE Trans. Biomed. Eng., vol. 52,pp. 74–87, Jan. 2005.
[74] A. B. Schwartz, BCortical neural prostheses,[Annu. Rev. Neurosci., vol. 27, pp. 487–507,2004.
[75] D. M. Taylor, S. I. H. Tillery, andA. B. Schwartz, BInformation conveyedthrough brain-control: Cursor versus robot,[IEEE Trans. Neural Syst. Rehabil. Eng., vol. 11,pp. 195–199, Jun. 2003.
[76] R. Olsson, III, and K. D. Wise, BAthree-dimensional neural recordingmicrosystem with implantable data
compression circuitry,[ IEEE J. Solid-StateCircuits, vol. 40, pp. 2796–2804, Dec. 2005.
[77] A. M. Sodagar, K. D. Wise, and K. Najafi,BA fully-integrated mixed-signal neuralprocessing module for implantablemulti-channel cortical recording,[ IEEETrans. Biomed. Eng., vol. 54, pp. 1075–1088,Jun. 2007.
[78] R. Harrison, P. T. Watkins, R. J. Kier,R. O. Lovejoy, D. J. Black, B. Greger, andF. Solzbacher, BA low-power integratedcircuit for a wireless 100-electrode neuralrecording system,[ IEEE J. Solid-State Circuits,vol. 42, pp. 123–133, Jan. 2007.
[79] M. A. L. Nicolelis, A. A. Ghazanfar,B. M. Faggin, S. Votaw, and L. M. O. Oliveira,BReconstructing the engram: Simultaneous,multisite, many single neuron recordings,[Neuron, vol. 18, pp. 529–537, Apr. 1997.
[80] A. M. Sodagar, G. E. Perlin, Y. Yao,K. D. Wise, and K. Najafi, BAn implantablemicrosystem for wireless multi-channelcortical recording,[ in Proc. Int. Conf.Solid-State Sens., Actuators, Microsyst.(Transducers’07), Lyon, France, Jun. 2007.
ABOUT THE AUTHORS
Kensall D. Wise (Life Fellow, IEEE) received the
B.S.E.E. degree (with highest distinction) from
Purdue University, West Lafayette, IN, in 1963
and the M.S. and Ph.D. degrees in electrical
engineering from Stanford University, Stanford,
CA, in 1964 and 1969, respectively.
From 1963 to 1965 and from 1972 to 1974, he
was a Member of Technical Staff with Bell
Telephone Laboratories, where his work was
concerned with the exploratory development of
integrated electronics for use in telephone communications. From 1965
to 1972, he was a Research Assistant and then a Research Associate and
Lecturer in the Department of Electrical Engineering, Stanford, working
on the development of micromachined solid-state sensors. In 1974, he
joined the Department of Electrical Engineering and Computer Science,
University of Michigan, Ann Arbor, where he is now the J. Reid and Polly
Anderson Professor of Manufacturing Technology and Director of the
Engineering Research Center for Wireless Integrated MicroSystems. His
present research focuses on the development of integrated microsys-
tems for health care and environmental monitoring. In 2002, he was
named the William Gould Dow Distinguished University Professor at the
University of Michigan, where he also held the 2007 Henry Russel
Lectureship.
Dr. Wise is a member of the U.S. National Academy of Engineering. He
organized and was the first Chairman of the Technical Subcommittee on
Solid-State Sensors of the IEEE Electron Devices Society (EDS). He was
General Chairman of the 1984 IEEE Solid-State Sensor Conference, IEEE-
EDS National Lecturer (1986), and Technical Program Chairman (1985)
and General Chairman (1997) of the IEEE International Conference on
Solid-State Sensors and Actuators. He received the Paul Rappaport
Award from the EDS (1990), a Distinguished Faculty Achievement Award
from the University of Michigan (1995), the Columbus Prize from the
Christopher Columbus Fellowship Foundation (1996), the SRC Aristotle
Award (1997), and the 1999 IEEE Solid-State Circuits Field Award.
Amir M. Sodagar (Member, IEEE) received the
B.S. degree from K. N. Toosi University of Tech-
nology (KNTU), Tehran, Iran, in 1992 and the M.S.
and Ph.D. degrees from Iran University of Science
and Technology (IUST), Tehran, in 1995 and 2000,
respectively, all in electrical engineering.
He was with S. Rajaee University as a Lecturer
from 1995 to 2000, with Iran Telecommunication
Research Center (ITRC) as a Design Engineer
during 1996–1997, with VLSI Circuits and Systems
Laboratory, University of Tehran, as a Research Engineer from 1997 to
1998, and with EMAD Semicon as a Senior Design Engineer from 1998 to
2000. He was a Guest Lecturer at several institutions during this time and
was a member of the Electrical and Electronics Engineering Technical
Committee of the International Kharazmi Youth Innovation Festival
during 1995–2000. In 2000, he joined the National Science Foundation
Engineering Research Center for Wireless Integrated MicroSystems
(WIMS), University of Michigan, Ann Arbor, as a Research Fellow, where
he worked on integrated microsystems for electric nerve stimulation. In
2002, he joined KNTU as an Assistant Professor. Since 2004, he has been
with WIMS as a Visiting Associate Research Scientist and subsequently as
the Technical Director for Biomedical Microsystems. His research
interests are in mixed-signal integrated circuits and biomedical implant-
able microsystems. He is the author of Analysis of Bipolar and CMOS
Amplifiers (Boca Raton, FL: CRC Press/Taylor & Francis Group, 2007).
Ying Yao (Member, IEEE) received the B.S.
degree in electrical engineering from Wuhan
University, China, in 1997 and the M.S. and Ph.D.
degrees in electrical engineering from the Univer-
sity of Michigan, Ann Arbor, in 2000 and 2005,
respectively.
Since 1999, she has been a Research Assistant
and then a Research Fellow with the Engineering
Research Center for Wireless Integrated Micro-
Systems, University of Michigan. Her research
interests focus on the development of implantable wireless integrated
microsystems for neural stimulating and recording with applications in
brain–machine interfaces, neuroscience, and neuroprostheses.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
Vol. 96, No. 7, July 2008 | Proceedings of the IEEE 1201
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.
Mayurachat Ning Gulari received the B.S. degree
in chemical technology and the M.S. degree in
polymer science and engineering from Petroleum
and Petrochemical College, Chulalongkorn Uni-
versity, Bangkok, Thailand, in 1993 and 1995,
respectively.
She was a Planning Engineer with Star Petro-
leum Refining Company (a joint venture of
Chevron and Texaco), where she worked on
optimization of feedstock and products. She is
currently a Research Engineer in the Engineering Research Center for
Wireless Integrated MicroSystems, University of Michigan, Ann Arbor,
where her research activities involve designing and fabricating micro-
machined silicon probes for drug delivery, recording, and stimulation.
Gayatri E. Perlin received the B.S.E. and M.S.E.
degrees in electrical engineering from the Univer-
sity of Michigan, Ann Arbor, in 2001 and 2003,
respectively, where she is currently pursuing the
Ph.D. degree in electrical engineering.
Her thesis is focused on the development of a
fully implantable microsystem for neural prosthe-
ses. Her research interests include microfabrica-
tion, microelectromechanical devices, and
integrated circuits for biomedical and other
applications.
Khalil Najafi (Fellow, IEEE) received the B.S., M.S.,
and Ph.D. degrees from the Department of Elec-
trical Engineering and Computer Science, Univer-
sity of Michigan, Ann Arbor, in 1980, 1981, and
1986, respectively, all in electrical engineering.
He is the Schlumberger Professor of Engineer-
ing in the Electrical Engineering and Computer
Science Department, University of Michigan. He
was Director of the Solid-State Electronics Labo-
ratory from 1998 to 2005. He has been Director of
National Science Foundation (NSF)’s National Nanotechnology Infra-
structure Network (NNIN) since 2004 and Deputy Director of the NSF
Engineering Research Center on Wireless Integrated Microsystems
(WIMS) at the University of Michigan. His research interests include
micromachining technologies, micromachined sensors, actuators, and
microelectromechanical systems; analog integrated circuits; implantable
biomedical microsystems; micropackaging; and low-power wireless
sensing/actuating systems. He has been active in the field of solid-state
sensors and actuators for more than 20 years. He has been involved in
several conferences and workshops dealing with microsensors, actua-
tors, and microsystems, including the International Conference on Solid-
State Sensors and Actuators, the Hilton Head Solid-State Sensors and
Actuators Workshop, and the IEEE/ASME Micro Electromechanical
Systems Conference. He is an Associate Editor of the Journal of
Micromechanics and Microengineering and an Editor for Sensors and
Materials.
Dr. Najafi is a Fellow of AIBME. He received a National Science
Foundation Young Investigator Award from 1992 to 1997. He is an
Associate Editor of the IEEE JOURNAL OF MICROELECTROMECHANICAL
SYSTEMS. He was an Associate Editor of the IEEE JOURNAL OF SOLID-
STATE CIRCUITS from 2000 to 2004, the Editor for Solid-State Sensors of
the IEEE TRANSACTIONS ON ELECTRON DEVICES from 1996 to 2006, and an
Associate Editor of the IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
from 1999 to 2000.
Wise et al. : Microelectrodes, Microelectronics, and Implantable Neural Microsystems
1202 Proceedings of the IEEE | Vol. 96, No. 7, July 2008
Authorized licensed use limited to: National University of Singapore. Downloaded on October 6, 2008 at 6:15 from IEEE Xplore. Restrictions apply.