Introduction to Medical EMC Tutorial
Session Chair: Darryl P. Ray
Darryl Ray EMC Consulting, LLC
IEEE EMC 2014 -‐ Introduc2on to Medical EMC Tutorial -‐ Darryl Ray EMC Consul2ng, LLC 1
![Page 2: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/2.jpg)
Topics For Today
• Basics of Medical EMC & IEC 60601-‐1-‐2, 4th Edi:on Darryl P. Ray -‐ Darryl Ray EMC Consul2ng LLC, Carlsbad, CA,USA
• Tes:ng Considera:ons Harald Buchwald -‐ CSA Group, Strasskirchen, Germany
• EMC Aspects for Ac:ve Implants Curt Sponberg -‐ Medtronics, Minneapolis, MN, USA
• ACernoon Break (3:00PM – 3:30PM)
IEEE EMC 2014 -‐ Introduc2on to Medical EMC Tutorial -‐ Darryl Ray EMC Consul2ng, LLC 2
![Page 3: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/3.jpg)
Topics For Today (cont.)
• Tes:ng Wireless on Medical Devices Greg Kiemel -‐ NWEMC, Hillsboro, OR, USA
• Risk Management Keith Armstrong -‐ Cherry Clough Consultants, Brocton, United Kingdom
• An FDA Perspec:ve on EMC of Medical Devices Jeffrey L. Silberberg -‐ Food & Drug Administra2on, Silver Spring, MD, USA
• Open Panel Discussion – All
IEEE EMC 2014 -‐ Introduc2on to Medical EMC Tutorial -‐ Darryl Ray EMC Consul2ng, LLC 3
![Page 4: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/4.jpg)
Darryl P. Ray
Principal Consultant Darryl Ray EMC Consul:ng, LLC
Carlsbad, CA, USA
darryl.ray@dray-‐emc.com +1-‐925-‐487-‐0072
www.dray-‐emc.com
Basics of Medical EMC
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 4
![Page 5: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/5.jpg)
Overview
• IEC 60601 series of standards
• Complying with FDA & EU requirements
• Worldwide regulatory issues
• Design challenges
• What is “Essen:al Performance” & “Basic Safety”
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 5
![Page 6: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/6.jpg)
IEC 60601/80601 Series of Standards
• Safety standards • Pa:ent • Operator • Environment
• Essen:al Performance covered
• Over 80 standards & growing
• 80601 = Joint IEC/ISO standard
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 6
![Page 7: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/7.jpg)
IEC 60601/80601 Series Hierarchy
Medical Electrical Equipment Standards
• Par:cular -‐ Product specific (vert.) IEC 80601-‐2-‐X
• Collateral -‐ Parallel (horizontal) IEC 60601-‐1-‐X
• General safety standard IEC 60601-‐1
• Basic standards (e.g. 61000-‐4-‐x)
Precedence
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 7
![Page 8: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/8.jpg)
Medical EMC Requirements
• Requirements are generally per IEC 60601-‐1-‐2:2007
• EN 60601-‐1-‐2:2007 is equivalent
• Par:cular standards (IEC 80601-‐2-‐X) may modify levels and acceptance criteria
• IEC 60601-‐1-‐2 4th edi:on published February 2014
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 8
![Page 9: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/9.jpg)
Item US EU Japan Brazil Canada Korea China
Version of 60601-‐1-‐2
3rd Edi:on (IEC)
2007 (EN)
2nd Edi:on (JIS)
3rd Edi:on (IEC)
3rd Edi:on (IEC)
3rd Edi:on (IEC)
3rd Edi:on (YY0505)
Accept Manufacturers Data
yes yes yes no yes yes
no -‐ in country test
required
Lab Accredita:on Required no no no no no no no
CB EMC Report Required no no no no no no no
Unique Test Condi:ons possible possible yes
(100 V) yes
(117/220 V) no Yes (220 V) no
Unique Test Report Requirements
yes – see FDA web
site no no no no no no
Medical EMC Regulatory Requirements by Country
Note: Requirements subject to change – Check with your RA staff for updates IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 9
![Page 10: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/10.jpg)
Reading Tips
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 10
Small Caps = Defined Term
![Page 11: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/11.jpg)
* = Ra:onale (usually Annex A)
Reading Tips
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 11
![Page 12: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/12.jpg)
Complying with FDA Requirements
• Design controls required per 21CFR
• Quality requirements – Need detailed design requirements, test plan and report
• Explain why decisions were made
• Product compliance required throughout the lifecycle
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 12
![Page 13: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/13.jpg)
Complying with FDA Requirements
• Monitor design changes – retest as required
• Can it withstand an audit?
• Unique standards acceptance & report requirements (see FDA web site)
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 13
![Page 14: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/14.jpg)
FDA & EU – Dates of Compliance
• FDA u:lizes a predicate rule – legacy devices may not need to comply with newer standards
• European Union • MDD encourages “state of the art”
• Typical 3 year transi:on period for revised standards • Actual dates are stated in the Official Journal
• No:fied Body may specify dates of compliance
• MDD takes precedence over the EMC direc:ve
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 14
![Page 15: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/15.jpg)
What is Essential Performance (EP)?
• Pertains only to safety performance (freedom From unacceptable risk)
• Defined by: • IEC 60601-‐1 edi:on 3.1 • IEC/ISO 80601-‐2-‐X • Manufacturers may define using risk analysis
• Some devices have no essen:al performance • Relevance: Immunity acceptance criteria is linked to Essen:al Performance (and Basic Safety)
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 15
![Page 16: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/16.jpg)
Basic Safety & Essential Performance
From IEC 60601-‐1:2005 + A1:2012
3.27 Essen2al Performance performance of a clinical func:on, other than that related to basic safety where loss or degrada:on beyond the limits specified by the manufacturer results in an unacceptable risk
3.10 Basic Safety freedom from unacceptable risk directly caused by physical hazards when me equipment is used under normal condi:on and single fault condi:on
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 16
![Page 17: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/17.jpg)
Essential Performance - Example
From IEC/ISO 80601-‐2-‐72 (DraC)
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 17
![Page 18: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/18.jpg)
Basic Safety & Essential Performance
• Some Devices Do Not Have Specified Essen:al Performance
• All Devices Must Comply with Basic Safety
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 18
![Page 19: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/19.jpg)
Intended & Normal Use Definitions
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 19
From IEC 60601-‐1, Edi:on 3.1 3.44 Intended Use use for which a product, process or service is intended according to the specifica:ons, instruc:ons and informa:on provided by the manufacturer 3.71 Normal Use opera:on, including rou:ne inspec:on and adjustments by any operator and stand-‐by according to the instruc:ons for use
![Page 20: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/20.jpg)
Intended & Normal Use Definitions
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 20
From IEC 60601-‐1, Edi:on 3.1
NOTE Intended use should not be confused with normal use. While both included the concept of use as intended by the manufacturer, intended use focuses on the medical purpose while normal use incorporates not only the medical purpose, but maintenance, transport, etc. as well
![Page 21: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/21.jpg)
Some Unique Design Challenges
• Analog circuit suscep:bili:es
• Leakage current restric:ons
• Interference from mobile transmirers
• ESD immunity on connectors
• I/O ports – safety tes:ng per IEC 60601-‐1
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 21
![Page 22: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/22.jpg)
Analog Circuits – Can Be Very Susceptible
• 2Hz modula:on (from 3rd edi:on) is a challenge
• Many devices have high gain amplifiers
• Filter amplifiers as much as possible
• Locate in a shielded region
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 22
![Page 23: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/23.jpg)
Leakage Current Restrictions
• IEC 60601-‐1 2nd Edi:on has low leakage current limits
• IEC 60601-‐1 3rd Edi:on is less stringent, aligns with IEC 60950 – watch the version!
• Y-‐capacitor values constrained
• Select medical grade power supplies
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 23
![Page 24: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/24.jpg)
ESD Testing
• Consider transient suppression on accessible connector pins
• 3rd edi:on allows for connector exemp:on; • Allowed only if the port cannot be filtered • Requires labeling & staff training
• Connector test method differs between IEC 60601-‐1-‐2, 3rd and 4th edi:ons
• Increased :me between successive pulses allowed
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 24
![Page 25: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/25.jpg)
IEC 60601-1-2
• Applies only to medical electrical equipment and systems
• Requirements for emissions
• Requirements for immunity
• Labeling requirements; • Device marking • Accompanying documents
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 25
![Page 26: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/26.jpg)
IEC 60601-1-2 3rd Edition
• Immunity acceptance criteria based on 11 specific pass/fail criteria “as related to Basic Safety & Essen:al Performance”
• Separa:on of life suppor:ng vs. non life suppor:ng equipment
• Allow manufacturers to use lower immunity test levels (based only on significant physical, technological or physiological limita2ons)
• Interpreta:on Sheet (I-‐SH) published in 2010 (clarifies AC input voltage requirements)
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 26
![Page 27: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/27.jpg)
Immunity Compliance Criteria 3rd Edition
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 27
![Page 28: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/28.jpg)
Declaring Compliance to IEC 60601-1-2
• All applicable tests must pass
and
• Labeling requirements apply (suggest using a labeling checklist)
IEEE EMC 2014 -‐ Basics of Medical EMC -‐ Darryl Ray EMC Consul2ng, LLC 28
![Page 29: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/29.jpg)
IEC 60601-1-2, 4th Edition
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 29
![Page 30: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/30.jpg)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 30
Published February 25, 2014
![Page 31: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/31.jpg)
4th Edition Motivation
• Create safety standard w/ respect to EM disturbances
• Drawbacks with the 3rd edi:on • Basic Safety and Essen:al Performance aspects – not adequately addressed
• Test levels in the current standard are 13+ years old (new EM environments unaccounted for, e.g. cell phones)
• Mobile device usage restric:ons are now generally ignored
• Devices in the same intended use loca:on meet different immunity levels
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 31
![Page 32: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/32.jpg)
The Real World!
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 32
![Page 33: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/33.jpg)
4th Edition Philosophy
• Requirements based on the intended use environment (not the device type)
• Immunity levels based on the reasonably foreseeable maximum
• Suscep:bility from mobile transmirers must be addressed
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 33
![Page 34: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/34.jpg)
4th Edition Philosophy (cont.)
• Use Risk Management to tailor requirements
• Immunity requirements defined by port (like CISPR24)
• Improve user friendliness of standard • Use of tables • Numerous annexes with supplemental informa:on
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 34
![Page 35: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/35.jpg)
Changes From 3rd Edition - General
Requirements now based on three use environments; • Professional Healthcare (hospital & small clinic)
• Home Healthcare (most loca2ons outside the hospital/small clinic)
• Special (determined on a case by case basis)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 35
![Page 36: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/36.jpg)
Intended Use Environment Examples
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 36
![Page 37: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/37.jpg)
Changes From 3rd Edition - General
• AC input voltage requirements clarified • Streamlined to one voltage/frequency (excep2on: Voltage Dips & Interrupts)
• Regulators may insist on na:onal voltages
• Tes:ng of I/O ports (SIP/SOPS) clarified
• The Poten:al Equaliza:on Conductor must be connected during tes:ng – if applicable
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 37
![Page 38: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/38.jpg)
Potential Equalization Conductor
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 38
![Page 39: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/39.jpg)
Changes From 3rd Edition - Emissions
• Standby mode should be considered
• Requirements of CISPR 14-‐1* & CISPR 15 eliminated (*except for switching devices and motors)
• ITE equipment must meet CISPR 32 (not CISPR 22)
• X-‐Ray generators allowed 20dB relaxa:on
• New test: Pa:ent cable common mode emissions (informa2ve)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 39
![Page 40: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/40.jpg)
Changes From 3rd Edition - Immunity
• Immunity pass/fail criteria is based on Essen:al Performance and Basic Safety only
• Specific failure arributes from 3rd edi:on eliminated
• Immunity levels are based on use loca:on (not the device type)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 40
![Page 41: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/41.jpg)
Changes From 3rd Edition - Immunity
• Higher immunity levels in some instances
• Standby mode should be considered
• ESD; • Increased ESD test levels • Modified ESD test method on connectors
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 41
![Page 42: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/42.jpg)
Changes From 3rd Edition - Immunity
• Transmirer Exclusion Band eliminated (Radiated Immunity Only)
• Tes:ng on DC input port condi:onally required (Permanent Cables >3m)
• Provision for test samples damaged during tes:ng
• Two new immunity tests added
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 42
![Page 43: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/43.jpg)
Changes From 3rd Edition - Immunity
• New test: Close Field Proximity • 15 Specific test frequencies • 9 V/m to 28 V/m • Mostly pulse modula:ons • Future communica:on services should be considered
• New test: Surge for devices connected to 12V vehicle power
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 43
![Page 44: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/44.jpg)
Changes From 3rd Edition - Immunity
• Conducted immunity levels increased in some cases
• Modified Voltage Dips & Interrupts tes:ng
• Magne:c Immunity test levels significantly increased
• Ar:ficial Hand tes:ng requirements clarified
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 44
![Page 45: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/45.jpg)
Changes From 3rd Edition - Documentation
• CISPR 11 Class A emissions warning statement (text differs from CISPR 11)
• Labeling requirements modified (EMC tables deleted)
• EMC test plan required -‐ recommended content
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 45
![Page 46: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/46.jpg)
Changes From 3rd Edition - Documentation
• EMC test report required -‐ minimum content defined
• Risk management process – numerous EMC considera:ons required
• Referenced standards -‐ dated references specified
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 46
![Page 47: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/47.jpg)
Phenomenon IEC 60601-‐1-‐2: 3rd Edi:on
IEC 60601-‐1-‐2: 4th Edi:on
Prof. Healthcare Environment
Home Healthcare Environment
ESD 8 kV Air Discharge (max.) 6 kV Contact Discharge
15 kV Air Discharge (max.) 8 kV Contact Discharge
EFT/Burst 2 kV -‐ AC Mains 1 kV -‐ I/O Ports
5 kHz or 100 kHz PRR
2 kV AC Mains 1 kV I/O Ports 100 kHz PRR
Surges (AC Mains) 2 kV 2 kV
Bold = Changes From the 3rd edi4on
Comparison of Immunity Levels
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 47
![Page 48: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/48.jpg)
Phenomenon IEC 60601-‐1-‐2: 3rd Edi:on
IEC 60601-‐1-‐2: 4th Edi:on
Prof. Healthcare Environment
Home Healthcare Environment
Magne:c Immunity (50/60 Hz) 3 A/M 30 A/M
Conducted Immunity 3 V (0.15-‐ 80 MHz) 10V ISM Bands (Life Support)
3 V (0.15 -‐ 80 MHz)
6 V (ISM Bands)
3 V (0.15 -‐ 80 MHz)
6 V (ISM + Amateur)
Voltage Dips & Interrupts
• UT < 5%, 0.5 periods
• UT = 40 %, 5 periods
• UT = 70%, 25 periods
• UT < 5%, 5 seconds
• UT = 0%, 0.5 cycle (0, 45, 90, 135, 180, 225, 270 and 315°)
• UT = 0 %; 1 cycle UT = 70%; 25/30 cycles (@ 0 degrees)
• UT = 0%; 250/300 cycle
Bold = Changes From the 3rd edi4on
Comparison of Immunity Levels (cont.)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 48
![Page 49: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/49.jpg)
Phenomenon IEC 60601-‐1-‐2: 3rd Edi:on
IEC 60601-‐1-‐2: 4th Edi:on
Prof. Healthcare Environment
Home Healthcare Environment
Radiated Immunity
3 V/m -‐ Non Life Support 10 V/m -‐ Life Support 80 MHz – 2.5 GHz
80%@2 Hz (or 1 kHz) AM Modula:on
3 V/m
80 MHz – 2.7 GHz
80%@ 1 kHz
AM Modula:on
10 V/m
80 MHz – 2.7 GHz
80%@ 1 kHz
AM Modula:on
Proximity Field from Wireless Transmirers (New Test)
N/A 9 V/m to 28 V/m
15 specific frequencies
Bold = Changes From the 3rd edi4on
Comparison of Immunity Levels (cont.)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 49
![Page 50: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/50.jpg)
Test Frequency (MHz)
Test Level (Volts/meter)
Modula:on (@ 50% duty cycle)
Communica:on Service
(par2al list)
385 27 18 Hz TETRA 400
450 28 FM (5 kHz devia2on)
GMRS/FRS
710, 745, 780 9 217 Hz LTE
810, 870, 930 28 18 Hz GSM 800
1720, 1845, 1970 28 217 Hz GSM 1800
2450 28 217 Hz RFID
5240, 5500, 5785 9 217 Hz WLAN
Close Field Proximity Test Levels
Based on 0.3m Separa2on Distance
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 50
![Page 51: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/51.jpg)
Phenomenon IEC 60601-‐1-‐2: 3rd Edi:on
IEC 60601-‐1-‐2: 4th Edi:on
Prof. Healthcare Environment
Home Healthcare Environment
Electrical Transients -‐ Vehicle
12 Volt Powered (New Test)
N/A N/A ISO 7637-‐2 Pulses -‐ 600V max.
Bold = Changes From the 3rd edi4on
Comparison of Immunity Levels (cont.)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 51
![Page 52: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/52.jpg)
Non Patient Connected I/O Ports
• Related to the use of the port
• Intended Use – no tes:ng (except ESD)
• Normal Use – tes:ng is required
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 52
![Page 53: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/53.jpg)
ESD Testing on Connectors
Connector Shell Discharge Requirements Connector Usage
Metallic Contact Discharge to Shell Only Intended Use
Non Metallic Air Discharge to Shell & Air Discharge to Pins*
Intended Use & Normal Use
• if Reachable by the IEC test finger
• INTENDED USE – Medical purpose use only
• NORMAL USE – Intended use plus maintenance, standby mode, transport, etc.
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 53
![Page 54: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/54.jpg)
Integration with Non Medical Equipment
• Must not compromise the system Essen:al Performance and Basic Safety
• Non medical equipment must comply with relevant standards
• ITE (mul:media) must comply with CISPR 32 (not CISPR 22)
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 54
![Page 55: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/55.jpg)
When Do We Have to Comply with the 4th Edition?
United States European Union Other Regions
Legacy Devices; Never*
New Submirals;
June, 2017
All Devices;
Late 2018 (es2mated)
All Devices;
Varies With -‐ The Country** Part 2 Standards
Notes * Per FDA predicate scheme ** Some countries may not presently accept the 4th edi:on
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 55
![Page 56: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/56.jpg)
Impact on Legacy Products
• Retes:ng • w/ Poten:al Equaliza:on Conductor arached (if applicable)
• Emissions & immunity tes:ng in standby mode
• Possible modified immunity pass/fail criteria
• Modified immunity test requirements per clause 8
• New Tests – Close Field Proximity & Surge (vehicles)
• Labeling modifica:ons • Update Risk Management file – may impact tes:ng
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 56
![Page 57: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/57.jpg)
EMC Performance Standard – IEC 60601-4-2
• Don’t confuse with “Essen:al Performance”
• Pass/Fail criteria may be more stringent than essen:al performance & basic safety
• Expected publica:on 2014/2015 – assumes posi:ve vote
• IEC 60601-‐4-‐2 – regulatory connec:on unclear
IEEE EMC 2014 -‐ IEC 60601-‐1-‐2, 4th Edi2on -‐ Darryl Ray EMC Consul2ng, LLC 57
![Page 58: Introduction to Medical EMC Tutorialb-dig.iie.org.mx/Bibdig2/P14-0270/PDF/Workshop/041_3889.pdf · 2014. 7. 1. · 6060112 3rd& Edion (IEC)& 2007 (EN) 2nd& Edion (JIS)& 3rd& Edion](https://reader036.fdocuments.net/reader036/viewer/2022071517/613acab1f8f21c0c8268a32a/html5/thumbnails/58.jpg)
The End