Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip)...

4
Introducers, Stylets and Guides The key to difficult airway access

Transcript of Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip)...

Page 1: Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip) - used in difficult airway scenarios. An ET tube is advanced over the introducer.

Introducers, Stylets and Guides

The key to difficult airway access

Page 2: Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip) - used in difficult airway scenarios. An ET tube is advanced over the introducer.

Smiths Medical has manufactured the reusable tracheal tube introducer known as a gum elastic bougie for over 30 years. This product has been accepted as the gold standard for difficult airways.2 The Portex® Single Use Bougie, offers several simple solutions for the difficult airway with a comprehensive range of single use and reusable devices.

THE KEY TO DIFFICULT AIRWAY ACCESS

Single Use Bougie (Coudé Tip)

The Portex® Single Use Bougie gives you virtually the same performance as the reusable tracheal tube introducer, with additional benefits:

• Single use Reduces the risk of cross infection

• Portex® ivory color Easy to identify from our reusable product

• 700 mm length 100 mm longer than the reusable product

• Clear graduated markings Helps to identify landmarks

• Atraumatic coudé tip Reduces the potential risk of trauma

Reusable Introducers (Bougie)

For many years, anesthetists have favored the technique of tracheal intubation using the Portex® Venn tracheal tube introducer. It has an excellent reputation for convenience and “ease-of-use” particularly in difficult situations. 1, 4-7

General belief attributes its high success rates to include:

• Superior memory

• Good flexibility

• Coudé tip that can easily identify tracheal rings

Single Use Bougie (Coudé Tip) - used in difficult airway scenarios.

An ET tube is advanced over the introducer. The coudé angled tip confirms correct placement in a difficult airway scenario through bumping (clicks) on the tracheal rings. The coudé tip has minimal resistance on the ET tube upon removal.

Introducers (Bougie / Coudé Tip) - used in difficult airway scenarios.

The test was completed on a 3 second bend test using the new single use bougie against our reusable gum elastic bougie.3

*Competitor product 1 second.

1 sec 2 sec 3 sec

5 sec

*

Coudé angled tip

Coudé angled tip

Page 3: Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip) - used in difficult airway scenarios. An ET tube is advanced over the introducer.

Stylets

The Portex® intubation stylets have a shepherd’s hook design and can be inserted into the ET tube prior to intubation. It is made of malleable aluminium with a polyethylene outer sleeve that allows the stylet to be adapted to any desired shape.

• Single use so reduces risk of cross infection

• Choice of two lengths for adult intubation

• Choice of three sizes to fit most adult and pediatric tracheal tubes

Guides (Exchange guides)

The Portex® tracheal tube guides are available in both reusable (15CH and 10CH) and single use (15FR, 10CH and 5CH).

A guide is used when an ET tube is exchanged for a new or speciality tube and helps to guarantee correct placement when a difficult airway is known or suspected. It is left in the trachea when the initial tube is removed and allows the new tube to advance over the guide into place. The guide is then removed allowing the ET tube to be connected to the ventilator or anesthetic machine.

• Choice of reusable or single use.

• Recommended for use in cases where a tube exchange is necessary.

• Available in 5CH (pediatric , 10CH and 15CH).

DIFFICULT INTUBATION IS ONE OF THE MOST CHALLENGING SITUATIONS FACED BY ANAESTHESIOLOGISTS1

Stylets - for pre-forming the tracheal tube to a desired shape.

The Portex® intubation stylets are single use and designed for pre-forming the tracheal tube to facilitate intubation. Made from coated malleable aluminium, the stylets are easily adapted to the desired shape.

Guides (exchange guides) – for when an existing tracheal tube needs to be exchanged.

The guides do not have a coudé tip and the straight design offers negligible resistance on removal from an ET tube.

Shepherd’s hook design

Straight tip

Straight tip

Page 4: Introducers, Stylets and Guides - Smiths Medical · 2019. 11. 6. · Single Use Bougie (Coudé Tip) - used in difficult airway scenarios. An ET tube is advanced over the introducer.

PRODUCT(S) DESCRIBED MAY NOT BE LICENSED OR AVAILABLE FOR SALE IN CANADA AND OTHER COUNTRIES

Smiths Medical ASD, Inc.6000 Nathan Lane NorthMinneapolis, MN 55442, USATel: 1-614-210-7300Toll-Free USA: 1-800-258-5361www.smiths-medical.com

Find your local contact information at: www.smiths-medical.com/customer-support

Smiths Medical is part of the global technology business Smiths Group plc. Please see the Instructions for Use/Operator’s Manual for a complete listing of the indications, contraindications, warnings and precautions. Portex and the Smiths Medical design mark are trademarks of Smiths Medical. The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries. All other names and marks mentioned are the trade names, trademarks or service marks of their respective owners. ©2018 Smiths Medical. All rights reserved. GA194028EN-052018

MHYTCA-1273

FOR MORE INFORMATION VISIT OUR WEBSITE WWW.SMITHS-MEDICAL.COM

Introducers, Stylets and Guides

References1. The Eschmann Tracheal Tube Introducer, Viswanathan S et al. Anaesthesiology Review 1992;19(6):29-34 Hodzovic I. Latto I.P. 2. Rational choice of tracheal tube introducer (bougie). 15th Anaesthesia Update 26-30 January 2004.3. Data from bench tests held on file at Smiths Medical.4. Successful difficult intubation. Dogra S, Falconer R, Latto I. Anaesthesia 1990;45:774-65. Difficult Intubation. Cossham PS. British Journal of Anaesthetists 1985;57:2396. A flexible solution for emergency intubation difficulties. Nocera A. Annals of Emergency Medicine 1996;5:665-77. Successful difficult intubation. Kidd JF, Dyson A, Latto I. Anaesthesia 1988;43:437-8

SU = Single Use R = Reusable

Ordering Information

Product Code Description Recommended Use

Recommended tube I.D. (mm) Size (CH/mm) Length (in/mm) Units

per Case

14-504-17 Tracheal Tube Introducer (Bougie) R 6.0-11.0 15/5.0 23.6/600 1

100/123/515 Single Use Bougie SU 6.0-11.0 15/5.0 27.55/700 10

103004 Intubation Stylet (Pediatric) SU 2.5-4.5 6/2.0 11.75/298 10

103005 Intubation Stylet (Adult) SU 5.0-7.0 8/2.6 13.75/349 10

103006 Intubation Stylet (Large Adult) SU 7.5-10.0 10.5/3.5 15.75/400 10

14-504-76 Tracheal Tube Guide R 4.0-6.0 10/3.3 27.55/700 1

14-504-68 Tracheal Tube Guide R 6.0-11.0 15/5.0 27.55/700 1

100/125/005 Tracheal Tube Guide (Pediatric) SU 2.5-3.5 5/1.7 19.68/500 10

100/125/010 Tracheal Tube Guide SU 4.0-6.0 10/3.3 27.55/700 10

100/123/015 Single Use Exchange Guide (Adult) SU 6.0-11.0 15/5.0 27.55/700 10