Instrumentation Important Points to Remember

download Instrumentation Important Points to Remember

of 25

Transcript of Instrumentation Important Points to Remember

  • 8/12/2019 Instrumentation Important Points to Remember

    1/25

    INSTRUMENTATION IMPORTANT POINTS TO REMEMBER

    CHAPTER: ANALOG VOLTMETERS AND AMMETERS REF: SAHANEY

    PMMC type meters can be used for only to measure dc. Induction type meters are used to measure only ac. Pmmc is most accurate type of dc current measurements. Electro dynamo meter type instruments are useful to measure both ac and dc and scale

    is common for both, hence it is useful as transfer instruments.

    Thermal instruments have advantage that their calibration is same for both ac and dcthey are particularly suited for alternating type currents because they depends upon theheating effect of the alternating current. Their readings are independent of frequency

    or waveform of the current, and any stray magnetic fields which may exist in the vicinity

    of the meter.

    The induction principle is more generally used for watt hour meters than for ammetersand voltmeters owing to their comparatively high cost and inaccuracy of induction

    instruments of latter types.

    The most serious error is error produced by heat generated in the instrument which canchange the resistance of coil. In ammeters it is not a big problem but in case of

    voltmeters it is a serious problem.

    Hence resistance of the coil must be constant A high swamping resistance of whose resistance temp coefficient is small is connected in

    series with the coil, so that , although the resistance of the coil change considerably the

    overall change in resistance of circuit is small.

    PERMANENT MAGNET MOVING COIL (PMMC)

    The pmmc instrument is most accurate type for dc measurement. TORQUE EQUATION: deflecting torque Td = NBLDI =GI Where G is a constant = NBLD The spring control provides a restoring or control torque Tc= k For final steady deflection Tc=Td $=( G/K)I As the deflection is directly propotional to the current passing through the meter we get

    a uniform scale for the instrument.

  • 8/12/2019 Instrumentation Important Points to Remember

    2/25

    Errors in PMMC meters:- weakening of permanent magnets due to aging oftemperature effects

    Weakening of springs due to ageing of temperature effects Change of resistance of the moving coil with temperature Advantages and disadvanges of PMMC: scale is uniform Power consumption is very low, between 25 micro watt to 200 micro watts Disadvantages are costlier than moving iron type and cannot be used for ac

    measurements.

    OHMMETERS:-

    Series type: these ohmmeters will read infinite resistance at 0 current position and 0

    resistance at full scale current position.

    Rh = R1+ R2||Rm

    Shunt type : zero mark on the left side and infinite mark on right side

    Rh= R1||Rm

    MOVING IRON TYPE: the most common ammeters and voltmeters for laboratory or switch

    board use at power frequencies are the moving iron instruments.

    Cheap in cost as compared with any other type of ac instrument of same accuracy and

    ruggedness.

    This is attraction type instrument mostly.

    Deflection is proportional to square of current. The deflection torque is therefore unidirectional

    whatever may be the polarity of the current.

    Advantages and disadvantages:- universal use , cheap ,robust ,accuracy, 240 degree scale

    Due to hysteresis loss dc measurements are generally avoided.

    Waveform errors :- scale is rms , hence waveforms highly distorted can measure satisfactorily.

    ELECTRODYNAMO METER TYPE INSTRUMENTS:-

    A transfer instrument is one that may be calibrated with a dc source and then used withoutmodification to measure ac this requires a transfer type instrument to have same accuracy for

    both dc and ac , which the electro-dynamo meter instruments have.

    Advantages and disadvantages:-

    Because of air core these are free from eddy current errors and hysteresis errors.

  • 8/12/2019 Instrumentation Important Points to Remember

    3/25

    Can use both ac and dc and measures rms value irrespective of waveform

    Disadv: low torque/weight ratio can cause increased frictional losses and low sensitivity

    They are more expensive compared to pmmc and moving iron type.

    Sensitive to overloads and mechanical impacts should be handle with great care.

    Have non uniform scale.

    High power consumption because the current under measurement has to provide the field flux

    HOT WIRE INSTRUMENTS:

    Calibrated to measure rms value

    .instable due to stretching of wire, lack of ambient temperature compensation, sluggish

    response and high power consumption and inability to withstand overloads and mechanical

    shocks

    THERMO COUPLE INSTRUMENTS:

    When two metals having different work functions are placed together , a voltage is generated

    at the junction which is nearly proportional to the temperature. This junction is called thermo

    couple.

    Can be used to measure both ac and dc measurements and this type of instruments are very

    usefull for measurements at RF frequencies.

    ADVANTAGES: indicates rms value irrespective of waveform, not effected by stray magnetic

    fields, free from frequency errors can be used at microwave frequencies also.

    The overload capacity of these instruments is small compared to other type of instruments.

    RECTIFIER TYPE INSTRUMENTS:

    Convert ac to pulsating dc and uses PMMC ,and this is attractive method because pmmc has

    good sensitivity compared to other instruments.

    The rectifier type of instruments are calibrated for rms values of sinusoidal voltages or

    currents . in order that a rectifier type instrument should read true rms value of thesinusoidal input , the scale markings are made with using a multiplying factor of 1.11 in case

    of instruments using full wave rectifier 2.22 in case of instruments using halfwave rectifiers.

    Waveform errors are present in case of other than sinusoidal measurements.(see pg no 295

    for an example).

  • 8/12/2019 Instrumentation Important Points to Remember

    4/25

    Sensitivity for ac signals is lower than that of dc signals.

    METER TYPE SUITABILITY MAJOR USES

    PMMC DC Most widely used for dc in low and medium

    impedance ckts.

    MOVING IRON

    AC OR DC

    Inexpensive type used for rough indication of

    currents and voltages. Widely used in

    indicating type applications such as panels.

    ELECTRO DYNAMO METER Widely used for precise ac current and

    voltage measurements at power frequencies.

    Transfer instrument

    ELECTROSTATIC High voltages and little currents can be

    supplied by ciruit under measurement

    THERMO COUPLE Rf signals.

    RECTIFIER Medium sensitivity and medium impedance

    ckts.

    Note: electrostatic type instruments do not take any current , they measure rms value of

    voltage.

  • 8/12/2019 Instrumentation Important Points to Remember

    5/25

    CATHODE RAY OSCILLOSCOPE

    The normal form of a cro uses a horizontal input which is an internally generated rampvoltage called TIMEBASE

    Electro static deflection = D= LldEd/2dEa . L= distance between screen and center of the deflecting plates. Ed = potential between deflection plates. D= distance between deflecting plates. ld length of deflecting plates, Ea voltage of pre

    accelerating anode.

    DEFLECTIN SENSITIVITY S= D/Ed and deflection factor = 1/S.

    Graticule is a grid of lines that serves as a scale for measurements purpose AQUADAG: the bombarding electrons , striking the screen release secondary emission

    electrons are collected by an aqueous solution of graphite called aquadag which is

    connected to the second anode, collection of secondary electrons is necessary to keep

    the crt screen in a state of electrical equilibrium.

    Vertical amplifier is the principal factor in determining the sensitivity and bandwidth ofan oscilloscope.

    The gain of the vertical amplifier determines the smallest signal that the oscilloscopecan satisfactorily reproduce on the CRT screen, the sensitivity of an oscilloscope is

    directly proportional to gain of the vertical amplifier that is, as gain of the vertical

    amplifier increases sensitivity increases, which allows us to observe smallest amplitude

    signals.

    HORIZONTAL AMPS:- the horizontal amplifiers basically serves two purposes. 1. When the oscilloscope being used in the ordinary mode of operation to display a

    signal applied to the vertical input, the horizontal amplifier will amplify the sweep

    generator output.

    2. when the oscilloscope is being used in the X-Y mode , the signal applied to thehorizontal input terminal will be amplified by the horizontal amplifier.

    MEASUREMENTS OF PHASE AND FREQUENCY(LISSAJOUS PATTERNS)

  • 8/12/2019 Instrumentation Important Points to Remember

    6/25

    Fy/Fx = number of horizontal tangencies/ number of vertical tangencies.

    ans ()

    NOTE:

  • 8/12/2019 Instrumentation Important Points to Remember

    7/25

    DUAL TRACE OSCILLOSCOPE: there are two separate vertical channels

    There are two common operating modes alternate mode and chop mode.

    DUAL BEAM OSCILOOSCOPE: the dual trace oscilloscope cannot capture two fast transient

    events, as it cannot switch quickly enough between traces

    The dual beam cro has two separate electron beams and two separate vertical channels

    SAMPLING OSCILLOSCOPES: these are to examine very fast signals

    The horizontal amplifier should be designed for high amplitude and slow rise timesignals.

    Post acceleration is needed in a CRO if the frequency of the signal is greater than10MHZ.

    The source of emission of electrons in a CRT is a barium and strontium oxide coatedcathode.

  • 8/12/2019 Instrumentation Important Points to Remember

    8/25

    In sweep mode operation of cro fsignal > = fsweep in order to get stable output and reasonable

    accuracy.

    MEASUREMENT OF RESISTANCE:-

    Low resistance : order of 1 ohm or below that MEDIUM RESISTANCE:- 1- 0.1 M Ohm

    High resistance :- above 0.1 M ohm.

    Measurement of medium resistance:- 1. Ammetervoltmeter method, 2. Substitution method

    3. wheatstone bridge 4. Ohmmeter method.

    Wheatstone bridge:- bridge sensitivity is defined as the deflection of the galvanometer per unit

    fractional change in unknown resistance.

    SB= SVE/(P/Q + 2+ S/R)

    SB = bridge sensitivity and Sv = galvanometer voltage sensitivity.

    Q-Meter:

    q-meter is designed to measure some coil properties and capacitor properties.

    Operation based on series-resonance. At resonance Xc= XL then V=IR

    Q= XL /R = XC /R = Vc /V since at resonance VC =Q V

    If a fixes voltage applied to the circuit, a voltmeter across the capacitor can be calibrated to

    read Q directly.

    SERIES CONNECTION:- Low- impedance components such as low value resistors , small coils

    and large capacitors, are measured in series with the measuring circuit.

    PARALLEL CONNECTION:- high impedance components such as high value resistors ,, certain

    inductors and small capacitors are measured by connecting them in parallel with the

    measuring circuit.

    SOURCES OF ERROR:- distributed capacitance or self capacitance of measuring circuit.

    And insertion resistance of Q-meter.

  • 8/12/2019 Instrumentation Important Points to Remember

    9/25

    ERRORS IN MEASUREMENT AND THEIR STATISTICAL ANALYSIS

    Relative limiting error =

    .

    Combination of quantities with limiting errors:

    SUM OF TWO QUANTITIES:- X= X1+X2 then the relative limiting error is

    = (

    +

    )

    If a limiting error is mentioned in % then convert it into limiting error form and proceed.

  • 8/12/2019 Instrumentation Important Points to Remember

    10/25

    Previous ies questions:-

    Measurement of an unknown voltage with a dc potentiometer loses its advantage ofopen-circuit measurement when Voltage is larger than the range of the

    potentiometer.

    Harmonic distortion analyzer measures the rms value of all the harmonic componentexcept the fundamental frequency component.

    Load cells utilize elastic member as the primary transducer and strain gauges assecondary transducer.

    IES 2000 EC

    PMMC ------- linear scale over the entire range.Moving IRON -------- Square law type scale because Torque I

    2.

    Thermo couple ---------suitable for high frequency measurements.

    Electrostatic type -----------voltmeter.

    Wagners earth device removes all the earth capacitances from the bridge network. When temperatures being measured are high and physical contact with the process to

    be measured is impossible , use is made of thermal radiation methods or optical

    pyrometers are used, these pyrometers are used under conditions where corrosive

    vapours or liquids would destroy thermo couple and thermometers and thermistors if

    made to come in contact with the measured medium, this type of pyro meters are used

    for rapidly moving objects. Temperature range is 12000C to 1500

    0C.

    In the pH cell, one of the electrodes, called reference electrode is at constant potentialregardless of pH value.

    Pirani gauge:- the operation of pirani gauge depends on variation of the thermalconductivity of a gas with pressure. Pirani gauge is useful for pressures ranging from

    10-1

    mm to 10-3

    mm

    The CRO uses electrostatic type of focusing as compared to a TV picture tube which iselectromagnetic focusing.

    IES 2001 EC

    Flash type A/D converter is fastest, dual slope converter is slow but most accurate. Maxwell bridge can be used to measure Q< 10, and hays bridge Q>10, and

    disadvantage of Anderson bridge is complex than its proto type Maxwell bridge. But

    using Anderson bridge we can measure both self and mutual inductance.

  • 8/12/2019 Instrumentation Important Points to Remember

    11/25

    Anemometeris used to measure gas flow or gas velocity and stroboscopeis used tomeasure rotational speed.

    Piezoelectric crystal sensitivity is field developed per unit stress. Potentiometric accelerometer:-major drawback is low resolution and limited to low

    frequency measurements.

    LVDT accelerometer:- much smaller mass and thus a higher natural frequency .therefore can be used for high frequency applications. Lvdt offers low resistance to the

    motion than the potentiometric and is capable of much smaller resolutions, no problem

    of moving contacts.

    Piezo-electric accelerometer:- useful for high input frequencies , response is poor atfrequencies less than 10Hz.

    IES 2002:-

    Ventri tube is used to measure flow.

    Rochelle salt is a piezo-electric material. Period measurements are preferred over frequency measurements at lower

    frequencies due to gating error of 1 count ambiguity.

    Ramp type ADC-------------->> integrating type Dual slope ADC---------------------->> voltage dependent conversion time. Successive approximation ------------>> null balance type. Microwave oven is based on the principle of dielectric heating.

    IES 2003

    Range of thermistor -600 C to 150 C Hot-wire anemometer is used to measure gas velocities. Photoelectric threshold of a metal is E=

    where is work function of

    metal.

    In a metal conductivity increases with temperature at low temperatures and constantat some temperatures and at high temperatures conductivity decreases with

    increasing temperature.

    THERMISTORS:- Thermistor means thermal resistor, most of the thermistors have negative temperature

    coefficients.

    Theve temperature coefficient of resistance can be as large as several % per degreecelsius. This allows the thermistor circuits to detect very small change in temperature

    which could not be observed with an RTD or Thermo couple.

    Operating range -600 cC to 150 C and the resistance of thermistor ranges from 0.5 ohmto 0.75 Mega Ohms.

    Thermistor is highly sensitive device but highly non-linear.

  • 8/12/2019 Instrumentation Important Points to Remember

    12/25

    Thermo couples:- it works on the principle of seebeck effect, seebeck discovered that when

    heat applies to junction of two dissimilar metals form an electric circuit and current flows as a

    result of generated EMF.

    BASEMETAL couples RARE METAL COUPLES.

    Positive

    wire

    copper iron Chromel

    90% Cr

    10% Ni

    chromel Platinum

    90%

    rhodium

    10%

    Tungsten

    95%

    Rhenium

    5%

    Rhodium

    Iridium

    Negative

    wire

    Constantan

    40% Ni,

    60% CU

    constantan Alumel

    94% Ni,

    2% Al, +Si

    and Mn

    constantan platinum Tungsten

    72%

    Rhenium

    26%

    Iridium.

    Temp.

    Range0C

    -2500

    to +

    4000

    C

    -200 to +

    850

    -200 to

    +1100

    -200 to

    +850

    0 to

    +1400

    0 to

    + 2600

    0 to

    + 2100

    Note:- it should be home in mind that thermocouples are active transducers unlike RTD and

    thermistors which are passive.

    Advantages:- cheaper than the resistance thermometers, suitable for rapid changes in

    temperature.

    Disadvantage:- low accuracy and hence they cannot be used for precision work.

    To prevent contamination of thermocouple it should be kept in protecting tube or closed end

    metal.

    When temperatures near ambient are to be measured with a thermocouple and it is

    inconvenient to use a fixed reference junction and therefore compensating circuits must beemployed in the measuring system

    RESISTANCE THERMOMETER :-

    The resistance of a conductor changes when its temperature is changed. This property is

    utilized for measurement of temperature.

  • 8/12/2019 Instrumentation Important Points to Remember

    13/25

    Platinum is best suited for RTD applications as it can withstand for high temperatures and

    being noble metal it shows limited susceptibility to contamination

    THE REQUIREMENTS OF CONDUCTOR MATERIAL TO BE USED IN RTD:-

    Change in resistance of material per unit change in temperature is as large as possible The material should have high value of resistivity so that minimum volume of material is

    used for the construction of RTD.

    The resistance of material should have a continuous and stable relationship withtemperature .

    Gold and silver are rarely used for construction of RTD on account of low resistivities Tugsten has relatively high resistivity but is reserved for high temperature applications

    and it is extremely brittle and difficult to work.

    Copper is occasionally used for RTD its low resistivity forces conductor to be longer thanplatinum but copper RTDs are economical alternatives to platinum one

    Most pf the RTDs are made either platinum or nickel or nickel alloys.

    The best use of RTD is in temperature range between 6000 C to 11000 C.IES 2004:-

    Lloyd-fisher square is used to measure iron losses.ENERGY METERS:-

    Motor meters:- motor meters are used for both ac and dc circuits, for dc circuits metermay be ampere-hour meter or watt hour meter

    In motor meters the moving system is revolving continuously unlike the indicatingmeters .

    In a motor meter the speed of the moving system is controlled by braking system . thebraking system consists of a permanent magnet so placed that it induces eddy currents

    produce a braking torque which is proportional to speed of moving system.

    Creeping:-in some meters a slow but continuous rotation is obtained even when thereis no current flowing through the current coil and only pressure coil is energized . this is

    called creeping.

    ENERGY METERS FOR AC CIRCUITS:- Induction type of energy meters are universally used for measurement of energy in

    domestic and industrial ac circuits. Induction type of meters posses lower friction and

    higher torque /weight ratio . also induction type meters are inexpensive and accurate

    and retain their accuracy over a wide range of loads and temperature conditions.

    LVDT:- Advantages:- high range, can be measured from 1.25mm to 250mm, and LVDT is

    Friction less device.

    The LVDT gives a high output and many a times there is no need for amplification . thetransducer produces a high sensitivity which is typically about 40 V/cm.

  • 8/12/2019 Instrumentation Important Points to Remember

    14/25

    LVDTS show low hysteresis and hence repeatability is good in all conditions. Most of the LVDTs consume less than 1Watt. Disadvantages:- Relatively large displacements are needed to get appreciable differential output. There are sensitive to stray magnetic fields but shielding is possible. This is done by

    providing magnetic shields with longitudinal slots.

    Many a times the transducer performance is affected by vibrations. Piezo-resistive effect:change in the value of resistivity of a conductor when it is

    strained.

    IES 2005:-

    Rotameter is used to measure flow rate.

    IES 2006:-

    Among the following meters which does not exhibit square law response?

    Moving coil, moving iron, electro dynamo meter, hot wire instrument.

    Ans:- PMMC

    The operation of a pirani gauge depends on variation of the thermal conductivity of a gas with

    pressure.

    Resistive hygrometer:- some hygroscopic salts exhibit a change in resistivity with humidity .the most common is lithium chloride.

    Capacitive hygrometer:- these exhibit a change in dielectric constant with humidity changes.

    Microwave refractometer:-for secondary standards , a system is consisting of two cavities ,

    each of which is coupled to a klystron may be employed , one cavity contains dry air , the other

    contains the mixture to be measured , the change in dielectric constant changes the frequency

    of one oscillator.

    Crystal hygrometer: it uses quartz crystal

    Which one of the following is most sensitive device?

    Thermo couple, RTD, thermistor , pyrometerANS:- THERMISTER

    Ies 2007:-

    Potentiometric transducer is a zero order displacement transducer.

  • 8/12/2019 Instrumentation Important Points to Remember

    15/25

    ACE MOCK TESTS:-

    Problem:- a fullwave bridge rectifier ac voltmeter is used to measure a dc voltage of10V , the amount of error in reading will be?

    Ans:- vrms indicated = 1.11 * 10 V= 11.11 volts

    Error is +11 %.

    Problem :- a half wave type rectifier voltmeter is measuring 10V DC what is the

    measured value?

    Ans:- Vrms= 2.22 Vdc. 22.22V.

    TRANSDUCERS:-

    When actuated transforms energy from one form to other form. Passive transducers:- passive transducers derive power required for transduction from

    an auxiliary power source. They also derive part of the power required for conversion

    from the physical quantity under measurement. Typical examples of passive transducers

    are resistive, inductive, capacitive.

    Active transducers:- active transducers those which do not require an auxiliary powersource to produce their output . velocity, temperature , light intensity and force can be

    transduced with the help of active transducers include tacho-generators, thermo

    couples, photo voltaic cells and piezoelectric crystals.

    Capacitive transducers:- these type transducers are commonly used to measuredisplacements. These transducers use following effects

    1. Change in capacitance due to change in overlapping area of plats and2. Change in capacitance due to change in distance between the two plates.

    Advantages:- extremely sensitive, good frequency response till 50kHzand hence

    they are very useful for dynamic studies. High i/p impedance hence no loading,

    Disadvantages:- non linear, output impedance is high and can cause loading effect

    on next stage devices.

  • 8/12/2019 Instrumentation Important Points to Remember

    16/25

  • 8/12/2019 Instrumentation Important Points to Remember

    17/25

  • 8/12/2019 Instrumentation Important Points to Remember

    18/25

  • 8/12/2019 Instrumentation Important Points to Remember

    19/25

  • 8/12/2019 Instrumentation Important Points to Remember

    20/25

  • 8/12/2019 Instrumentation Important Points to Remember

    21/25

  • 8/12/2019 Instrumentation Important Points to Remember

    22/25

  • 8/12/2019 Instrumentation Important Points to Remember

    23/25

  • 8/12/2019 Instrumentation Important Points to Remember

    24/25

  • 8/12/2019 Instrumentation Important Points to Remember

    25/25