MEASURING INSTRUMENTS MASTER LIST
CONTROL NO INSTRUMENT NAME RANGE MAKER FOR USAGE REMARKS
1 SAPI 01 Vernier Caliper 0 to 200 mm Baker Distance,ID & OD Once in12 Month QC
2 SAPI 02 Point Micrometer 0 to 25 mm Mitutoyo Thickness Once in12 Month QC
3 SAPI 03 Screw Ring Gauge M6 (GO) Baker M6 Bolt checking Once in12 Month QC
4 SAPI 04 Screw Ring Gauge M8 (GO) Baker M8 Bolt checking Once in12 Month QC
5 SAPI 05 Screw Ring Gauge M6 (NO GO) Baker M6 Bolt checking Once in12 Month QC
6 SAPI 06 Screw Ring Gauge M8 (NO GO) Baker M8 Bolt checking Once in12 Month QC
7 SAPI 07 Screw Plug Gauge M6 (GO/NOGO) Baker M6 Nut checking Once in12 Month QC
8 SAPI 08 Screw Plug Gauge M8 (GO/NOGO) Baker M8 Nut checking Once in12 Month QC
9 SAPI 09 Micrometer 0 to 25 mm Baker OD &Thickness Once in24 Month QC
10 SAPI 10 Height Gauge 0 to 600 mm Baker Distance & Height Once in12 Month QC
11 SAPI 11 Measuring Pin - Advance OD Once in12 Month QC
12 SAPI 12 Ring Gauge 20 mm Unique ID Once in12 Month QC
Note OD : Outer Diameter , ID : Inner Diameter
SL.NO
CABLIRATIONPERIOD
USERDEPT
MEASURING INSTRUMENTS CALIBRATION PLAN /ACTUAL
DOC NO: PREPARED
DOC DT & REV NO:
PLAN
ACTUAL
INSTRUMENT RANGE MAKER FOR USAGE
CABLIRATION PLAN/ACTUAL 2015
2015 (YEAR)
JAN FEB MAR APR MAY JUN JULY AUG SEP OCT NOV
1 SAPI 01 Vernier Caliper 0 to 200 mm Baker Once in12 Month QC
PLAN
ACTUAL
2 SAPI 02 Point Micrometer 0 to 25 mm Mitutoyo Thickness Once in12 Month QC
PLAN
ACTUAL
3 SAPI 03 Screw Ring Gauge M6 (GO) Baker Once in12 Month QC
PLAN
ACTUAL
4 SAPI 04 Screw Ring Gauge M8 (GO) Baker Once in12 Month QC
PLAN
ACTUAL
5 SAPI 05 Screw Ring Gauge M6 (NO GO) Baker Once in12 Month QC
PLAN
ACTUAL
6 SAPI 06 Screw Ring Gauge M8 (NO GO) Baker Once in12 Month QC
PLAN
ACTUAL
7 SAPI 07 Screw Plug Gauge M6 (GO/NOGO) Baker M6 Nut checking Once in12 Month QC
PLAN
ACTUAL
8 SAPI 08 Screw Plug Gauge M8 (GO/NOGO) Baker M8 Nut checking Once in12 Month QC
PLAN
ACTUAL
9 SAPI 09 Micrometer 0 to 25 mm Baker OD &Thickness Once in12 Month QC
PLAN
ACTUAL
10 SAPI 10 Height Gauge 0 to 600 mm Baker Once in12 Month QC
PLAN
ACTUAL
11 SAPI 11 Measuring Pin - Advance OD Once in12 Month QC
PLAN
ACTUAL
12 SAPI 12 Ring Gauge 20 mm Unique ID Once in12 Month QC
PLAN
ACTUAL
SL.NO
CONTROL NO
CABLIRATIONPERIOD
USERDEPT
PLAN VS
ACTUAL
Distance,ID & OD
M6 Bolt checking
M8 Bolt checking
M6 Bolt checking
M8 Bolt checking
Distance & Height
CHECKED VERIFIED
CABLIRATION PLAN/ACTUAL 2015
REMARKS2015 (YEAR)
NOV DEC
HISTORY CARD1 CONTROL NO SAPI 01
PHOTO
2 INS NAME VERNIER CALIPER
3 MAKE BAKER
4 -
5 RANGE 0~200MM
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Certificate Referance No Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO: 00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
LaboratoryName
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 02
PHOTO
2 INS NAME POINT MICROMETER
3 MAKE MITUTOYO
4 -
5 RANGE 0~25MM
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 03
PHOTO
2 INS NAME SCREW RING GAUGE
3 MAKE BAKER
4 -
5 RANGE M6 (GO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 04
PHOTO
2 INS NAME SCREW RING GAUGE
3 MAKE BAKER
4 -
5 RANGE M8 (GO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 05
PHOTO
2 INS NAME SCREW RING GAUGE
3 MAKE BAKER
4 -
5 RANGE M6 (NOGO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 06
PHOTO
2 INS NAME SCREW RING GAUGE
3 MAKE BAKER
4 -
5 RANGE M8 (NOGO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 07
PHOTO
2 INS NAME SCREW PLUG GAUGE
3 MAKE BAKER
4 -
5 RANGE M6 (GO/NOGO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 08
PHOTO
2 INS NAME SCREW PLUG GAUGE
3 MAKE BAKER
4 -
5 RANGE M8 (GO/NOGO)
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 08
PHOTO
2 INS NAME MICROMETER
3 MAKE BAKER
4 -
5 RANGE 0~25MM
6 06.06.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 08
PHOTO
2 INS NAME HEIGHT GAUGE
3 MAKE BAKER
4 -
5 RANGE 0~600MM
6 06.09.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 08
PHOTO
2 INS NAME MEASURING PIN
3 MAKE ADVANCE
4 -
5 RANGE -
6 06.09.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD1 Control No SAPI 08
PHOTO
2 INS NAME RING GAUGE
3 MAKE UNIQUE
4 -
5 RANGE 20MM
6 06.09.2013
7 BOMBAY TOOLS
8 USER DEPT QC
9 INCHARGE SIGN
NO Laboratory Name Certificate Referance Number Judgement
1
2
3
4
5
6
7
8
9
10
NO Problem Content Action Judgement
1
2
3
4
5
REV NO:00 DATE: 01.10.2013
MANUFACTURERID NO
PURCHASEDATE
PURCHASECOMPANY
Date of Calibration
CalibrationPeriod
Due Date of Calibration
Date of Problem
LABORATORY NAME
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY
HISTORY CARD
PHOTO
Incharge Approved
Incharge Approved
PROBLEM RECORD STATUSPROBLEM RECORD STATUS
CALIBRATION HISTORYCALIBRATION HISTORY