Instruction Sheet smarTROLL Calibration Kit Instruction Sheet Learn ...

2
Technical Note Innovations in Water Monitoring Instruction Sheet Innovations in Water Monitoring Page 1 Page 1 Kit Contents (0080830) Battery-powered air pump Large bubbler cup, tubing, check valve, and pinch valve for calibrating the 100% dissolved oxygen saturation point. Restrictor to accommodate the sub 2-inch TROLL 9500 instrument 2 alkaline D-cell batteries Sodium sulfite solution, 500 mL for calibrating the 0% dissolved oxygen point In addition to these items, you will need the plastic calibration cup that came with the instrument for calibrating the 0% dissolved oxygen point. Kit Contents (0048580) Battery-powered air pump Large bubbler cup, tubing, check valve, and pinch valve for calibrating the 100% dissolved oxygen saturation point. 2 alkaline D-cell batteries Sodium sulfite solution, 500 mL for calibrating the 0% dissolved oxygen point Small calibration cup for calibrating the 0% dissolved oxygen point. Calibration Kit for the Aqua TROLL ® 400, SMARTROLL™ MP and TROLL ® 9500 RDO ® Sensor 0080830 – for Aqua TROLL 400 Instrument and TROLL 9500 Instrument (sub 2-inch), SMARTROLL™ MP 0048580 – for TROLL 9500 Instrument (sub 4-inch)

Transcript of Instruction Sheet smarTROLL Calibration Kit Instruction Sheet Learn ...

Page 1: Instruction Sheet smarTROLL Calibration Kit Instruction Sheet Learn ...

Technical Note

Innovations in Water Monitoring

Instruction Sheet

Innovations in Water Monitoring

Page 1 Page 1

Kit Contents (0080830)

• Battery-poweredairpump• Largebubblercup,tubing,checkvalve,andpinch

valveforcalibratingthe100%dissolvedoxygensaturationpoint.

• Restrictortoaccommodatethesub2-inchTROLL9500instrument

• 2alkalineD-cellbatteries• Sodiumsulfitesolution,500mLforcalibratingthe

0%dissolvedoxygenpoint• Inadditiontotheseitems,youwillneedtheplastic

calibrationcupthatcamewiththeinstrumentforcalibratingthe0%dissolvedoxygenpoint.

Kit Contents (0048580)

• Battery-poweredairpump• Largebubblercup,tubing,checkvalve,andpinch

valveforcalibratingthe100%dissolvedoxygensaturationpoint.

• 2alkalineD-cellbatteries• Sodiumsulfitesolution,500mLforcalibratingthe

0%dissolvedoxygenpoint• Smallcalibrationcupforcalibratingthe0%

dissolvedoxygenpoint.

Calibration Kit for the Aqua TROLL® 400, SMARTROLL™ MP and TROLL® 9500 RDO ® Sensor

0080830 – for Aqua TROLL 400 Instrument and TROLL 9500 Instrument (sub 2-inch), SMARTROLL™ MP0048580 – for TROLL 9500 Instrument (sub 4-inch)

Page 2: Instruction Sheet smarTROLL Calibration Kit Instruction Sheet Learn ...

Page 2

Informationsubjecttochangewithoutnotice.In-Situ,In-Situlogo,BaroMerge,BaroTROLL,HERMIT,iSitu,RDO,Pocket-Situ,RuggedCable,RuggedReader,TROLL,andWin-SituaretrademarksorregisteredtrademarksofIn-SituInc.©2012.Allrightsreserved.

For more information, contact In-Situ Inc. 221 East Lincoln Avenue, Fort Collins, CO 80524 1-800-446-7488 (toll-free in U.S.A. & Canada)1-970-498-1500 (international & domestic)www.in-situ.com 0080832 Rev. 004 04/13

Application

TheRDO®Sensorisextremelystable,whicheliminatestheneedforfrequentcalibration.Standardoperatingprocedures,environmentalconditions(e.g.,potentialforfouling,abrasiveconditions)maynecessitateroutinecalibration.

AfterinstallinganewRDOSensor,In-Situ®recommendsthatyouperformacalibration.Forbestresults,performa2-pointcalibrationat100%and0%dissolvedoxygen(DO)saturationbeforeusingthesensor.

100% Oxygen Saturation Calibration

1. Installthebatteriesintheportableairpumpasshowninthediagramontheinsidecover.

2. Fillthelargecalibrationcuptowithin4cm(1.5in.)ofthetopwithcleanwater.

Note: Use tap water, distilled water, or your sample matrix if it is freshwater, not brackish or saline.

3. Turnonthebubbler,openthepinchvalveonthetubing.Runatfullflowratefor10minutestofullysaturatethewaterwithoxygen.After10minutesyoucanusethecontrolvalvetomatchthewaterturbulenceconditionsexpectedinthefield.

4. PlacetheinstrumentinthewaterensuringthatthesensorandthetemperaturethermistorarecompletelysubmergedandthattheRDOsensingmaterialisnotinthedirectpathofthebubbles.

5. Connecttheinstrumenttoacomputerandestablishaconnectiontothecalibrationsoftware.

6. Followthecalibrationwizardprovidedinthesoftware.

Note: Video tutorials and operator’s manuals are available on the In-Situ website www.in-situ.com in the Tech Resources section.

0% Oxygen Saturation Calibration

1. Fillthesmallercalibrationcuptotheupperfilllinewithsodiumsulfite(NA2SO3)solution.

2. Placetheinstrumentinthesolutionandallowtimefortemperaturestabilization.

3. EnsuretherearenoairbubblesonthesensingsurfaceoftheRDOSensor.

4. Followthecalibrationwizardprovidedinthesoftware.

Note: Video tutorials and operator’s manuals are available on the In-Situ website www.in-situ.com in the Tech Resources section.

5. Whenyouarefinishedrinsethoroughlyincleanwater.

Replacement Parts

Sodium sulfite solution, 500 mL 0017670