Technical Note
Innovations in Water Monitoring
Instruction Sheet
Innovations in Water Monitoring
Page 1 Page 1
Kit Contents (0080830)
• Battery-poweredairpump• Largebubblercup,tubing,checkvalve,andpinch
valveforcalibratingthe100%dissolvedoxygensaturationpoint.
• Restrictortoaccommodatethesub2-inchTROLL9500instrument
• 2alkalineD-cellbatteries• Sodiumsulfitesolution,500mLforcalibratingthe
0%dissolvedoxygenpoint• Inadditiontotheseitems,youwillneedtheplastic
calibrationcupthatcamewiththeinstrumentforcalibratingthe0%dissolvedoxygenpoint.
Kit Contents (0048580)
• Battery-poweredairpump• Largebubblercup,tubing,checkvalve,andpinch
valveforcalibratingthe100%dissolvedoxygensaturationpoint.
• 2alkalineD-cellbatteries• Sodiumsulfitesolution,500mLforcalibratingthe
0%dissolvedoxygenpoint• Smallcalibrationcupforcalibratingthe0%
dissolvedoxygenpoint.
Calibration Kit for the Aqua TROLL® 400, SMARTROLL™ MP and TROLL® 9500 RDO ® Sensor
0080830 – for Aqua TROLL 400 Instrument and TROLL 9500 Instrument (sub 2-inch), SMARTROLL™ MP0048580 – for TROLL 9500 Instrument (sub 4-inch)
Page 2
Informationsubjecttochangewithoutnotice.In-Situ,In-Situlogo,BaroMerge,BaroTROLL,HERMIT,iSitu,RDO,Pocket-Situ,RuggedCable,RuggedReader,TROLL,andWin-SituaretrademarksorregisteredtrademarksofIn-SituInc.©2012.Allrightsreserved.
For more information, contact In-Situ Inc. 221 East Lincoln Avenue, Fort Collins, CO 80524 1-800-446-7488 (toll-free in U.S.A. & Canada)1-970-498-1500 (international & domestic)www.in-situ.com 0080832 Rev. 004 04/13
Application
TheRDO®Sensorisextremelystable,whicheliminatestheneedforfrequentcalibration.Standardoperatingprocedures,environmentalconditions(e.g.,potentialforfouling,abrasiveconditions)maynecessitateroutinecalibration.
AfterinstallinganewRDOSensor,In-Situ®recommendsthatyouperformacalibration.Forbestresults,performa2-pointcalibrationat100%and0%dissolvedoxygen(DO)saturationbeforeusingthesensor.
100% Oxygen Saturation Calibration
1. Installthebatteriesintheportableairpumpasshowninthediagramontheinsidecover.
2. Fillthelargecalibrationcuptowithin4cm(1.5in.)ofthetopwithcleanwater.
Note: Use tap water, distilled water, or your sample matrix if it is freshwater, not brackish or saline.
3. Turnonthebubbler,openthepinchvalveonthetubing.Runatfullflowratefor10minutestofullysaturatethewaterwithoxygen.After10minutesyoucanusethecontrolvalvetomatchthewaterturbulenceconditionsexpectedinthefield.
4. PlacetheinstrumentinthewaterensuringthatthesensorandthetemperaturethermistorarecompletelysubmergedandthattheRDOsensingmaterialisnotinthedirectpathofthebubbles.
5. Connecttheinstrumenttoacomputerandestablishaconnectiontothecalibrationsoftware.
6. Followthecalibrationwizardprovidedinthesoftware.
Note: Video tutorials and operator’s manuals are available on the In-Situ website www.in-situ.com in the Tech Resources section.
0% Oxygen Saturation Calibration
1. Fillthesmallercalibrationcuptotheupperfilllinewithsodiumsulfite(NA2SO3)solution.
2. Placetheinstrumentinthesolutionandallowtimefortemperaturestabilization.
3. EnsuretherearenoairbubblesonthesensingsurfaceoftheRDOSensor.
4. Followthecalibrationwizardprovidedinthesoftware.
Note: Video tutorials and operator’s manuals are available on the In-Situ website www.in-situ.com in the Tech Resources section.
5. Whenyouarefinishedrinsethoroughlyincleanwater.
Replacement Parts
Sodium sulfite solution, 500 mL 0017670