INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The...

11
IMPORTANT: Your Unit is Preprogrammed Place your unit in the desired location. Plug the unit in and allow it to cool and become stable for a minimum of 24 hours before logging temperature or stocking products. Be Careful When Setting or Changing Temperatures WARNING: Changing some controller parameters can result in potentially damaging your unit and/or losing your products. K2 will not be held responsible for losses due to improper controller programming. Unfamiliar with the operation of a K2 controller? Use the video tutorials on our website resources page or call us for assistancee with special parameters. Some K2 units can be changed from Celsius to Fahrenhiet. We do not recommend changing your unit to Farhenheit. Your controller has operational parameters that rely on data in Celsius to maintain proper function. Thank you for your purchase. This manual contains: set-up instructions, unit specs, safety information, controller operation and maintenance steps. Evco Controller Operating Manual K212, K225, K249, K272

Transcript of INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The...

Page 1: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

IMPORTANT: Your Unit is PreprogrammedPlace your unit in the desired location. Plug the unit in and allow it to cool and become stable for a minimum of 24 hours before logging temperature or stocking products.

Be Careful When Setting or Changing TemperaturesWARNING: Changing some controller parameters can result in potentially damaging your unit and/or losing your products. K2 will not be held responsible for losses due to improper controller programming.

Unfamiliar with the operation of a K2 controller?Use the video tutorials on our website resources page or call us for assistancee with special parameters.

Some K2 units can be changed from Celsius to Fahrenhiet. We do not recommend changing your unit to Farhenheit. Your controller has operational parameters that rely on data in Celsius to maintain proper function.

Thank you for your purchase. This manual contains: set-up instructions, unit specs, safety information,

controller operation and maintenance steps.

Evco Controller Operating Manual

K212, K225, K249, K272

Page 2: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

CONTENTS1. Front Panel Commands 2. General Operations3. Parameters

4. Alarms5. Electrical Connection 6. Temperature Logging

Front Panel Commands 1

Evco Operating Manual

Page 3: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

• Make sure the keypad is unlocked and press the DOWN key for 4s.

• Scroll through the menu labels with the UP and DOWN keys.

• CH label: displaying compressor operating hours.

• rCH label: compressor operating hours reset.

• nS1 label: compressor star-up time.

• To access the label press SET.

•In order to reset the compressor operating hours once selected the rCH label, insert the password “149” using the UP and DOWN keys then confirm touching the SET key.

•Touch the ON/STANDBY key to exit the procedure.

General Operations2

Evco Operating Manual

Keypad Unlocking

Compressor Opperational Time

Date and Time

Buzzer

Cabinet Light

• If the parameter Loc=1 (default) after 30s* without any keys of the display has been pressed, the display will show the label “Loc” and the keypad will lock automatically.

• To unlock the keypad, touch the SET key for 1s. Then the display will show the label “UnL”.

• If the parameter is u1=0

• Touch the ON/STAND-BY key.

• Ensure the keypad is unlocked then press the DOWN key for 4s.

• Scroll through the menu’s labels UP or DOWN key up to the “rtc” label then press the SET key.

• The moniter will display “yy” followed by the two last digits of the year.

• Set the yeat using the UP or DOWN key then press the SET key to confirm.

• Repeat the operation for the following labels:

• n: month (01 to 12)

• d: day (01 to 31)

• h: time (00 to 23)

• n: minutes (00 to 59)

• If the parameters u1=3 and u4=1 touch any key to shut down the buzzer alarm.

Displaying Temperature Probes

Manual Defrost• First, check the keypad is not locked (and in case unlock it) and the overcooling cycle is not activated.

• Press the UP key, holding it for 2s.

• If the parameter P3=1 and the evaporator temperature value is lower than the parameter d2, defrost cycle will start.

• Ensure the keypad is unlocked then press the DOWN key for 4 sec.

• Scroll through the menu’s labels by the UP or DOWN key.

•Pb1: cabinet temperature probe

(if parameter P4=0,1 or 2); inlet air

temperature probe (if parameter P4=3).

•Pb2: Evaporator temperature probe (if parameter P3=1 or 2)

•Pb3: Auxiliary temperature probe

(If P4=1, 2 or 3).

• To access the label press SET.

• Touch the ON/STANDBY key to exit the procedure.

Temperature Settings• If the keypad is locked, first unlock it.

• Touch the SET key then set the desired temperature by pressing the UP or DOWN keys within 15s according with the limits range of the set point (parameters r1 and r2)

• Press SET to confirm.

*s=seconds

Page 4: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Alarms4

5Electrical Connection

Page 5: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Parameters3

Accessing Operational Parameters• Touch the SET key for 4 sec, the moniter will display the label “PA”

• Press the SET key and insert the password “-19”

• Press SET key to confirm

• Scroll through the parameters list using the UP or DOWN key

• For modifying a parameter value, press SET key at the parameter label then adjust the value by the UP or DOWN key

• Press SET key to confirm the changing

•Press SET key for 4 sec or do not operate for 60 sec to exit the procedure.

WARNING:Do not modify prameters in this unit without fully understanding their function. Please consult K2 Scientific support. k2sci.com/kb

N Parameter Default Setpoint

1 SP 0.0 setpoint

8 P3 1 evaporator probe function

13 r0 2.0 setpoint differential

14 r1 -50 minimum setpoint

15 r2 50.0 maximum setpoint

33 d0 8 automatic defrost interval

34 d1 0 defrost type

35 d2 8.0 threshold for defrost end

36 d3 30 defrost duration

51 AA 0 select value for high/low temperature alarms

52 A1 -10.0 threshold for low temperature alarm

53 A2 2 low temperature alarm type

54 A4 10.0 threshold for high temperature alarm

55 A5 2 high temperature alarm type

74 i0 5 door switch input section

75 i1 0 door switch input activation

76 i2 30 open door alarm delay

Basic parameters

Page 6: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Full list of parameters

Page 7: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Page 8: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Page 9: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Page 10: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

Page 11: INSTRUCTION MANUAL · Read this manual carefully and follow the instructions contained herein. The user assumes full responsibility in case of operations carried out without observing

Evco Operating Manual

ServiceK2 Scientific want to make sure you are happy with your purchase. There are several ways for you to contact us with questions or service needs. Be sure to include your four digit order number or at least your model number handy to speed up the process.

1; Contact us via our chat feature at www.k2sci.com2: Email [email protected]: Call 800-218-7613

When storing vaccines you may be required to preform a field validation test. A NIST callibrated external data logger may be used for this purpose. A logger with text, email or online access is an added layer of protection for your product load in the event of a temperature excursion. K2 offers NIST calibrated data loggers to match your unit at a very affordable rate.

Temperature Logging6