Installation and configuration
description
Transcript of Installation and configuration
Code P010 Date: 20.2.20071
Installation and configuration
Clarity User Training
Code P010 Date: 20.2.20072
Software protection
Software installation
HW installation
Software configuration
Installation
Code P010 Date: 20.2.20073
Software protection by hardware key Hardware key (dongle) required for each station USB standard (LPT optional on request) The software is independent of hardware (A/D boards) User code (16 position alfanumeric code) combined with the
apropriate dongle (S/N yyyy) is necessary to operate the Clarity Datastation (S/N xx-zyyyy)
30 day trial version available
IMPORTANT !! Please note that a loss of a dongle (hardware key) is considered as a loss of the license right to use Clarity SW.
Software Protection
Code P010 Date: 20.2.20074
Install the software first before any hardware! Insert the CD
Start install.exe (if not started automatically)
Follow the dialog screens
Set the installation directory
Software Installation
Code P010 Date: 20.2.20075
Software Installation
Set the program folder
Select type of installation
Typical should be used in most cases
Code P010 Date: 20.2.20076
Software Installation
Enter the User code corresponding to the HW
key used
Leave blank for30 day Trial mode
Chose acquisition device which would you
like to set as default
Code P010 Date: 20.2.20077
Software Installation
Wait for next screen, the installation is in progress
Code P010 Date: 20.2.20078
Software Installation
Start the Hardware installation
Wait for next screen, the installation is in progress
Code P010 Date: 20.2.20079
Software Installation
Finish the Hardware installation
Wait for the installation to finish
Code P010 Date: 20.2.200710
Software Installation
Finish the software installation
Show readme.txt file
Run Clarity(do not check with hardware not yet
installed!)
Run IQ utility
Code P010 Date: 20.2.200711
Hardware Installation
Switch off the computer
Insert the INT7 A/D board to a free PCI slot
Switch on the computer
Follow the New Hardware Found dialogs
Code P010 Date: 20.2.200712
Hardware Installation
Select „No, not this time“ for updates search
Select Install Automatically
(first time installation on a new computer)
Code P010 Date: 20.2.200713
Hardware Installation
Wait, Installation is in progress
This warning may appear under some
circumstances.It is safe to ignore it, use
„Continue Anyway“ to proceed
Code P010 Date: 20.2.200714
Hardware Installation
Insert the HW key New hardware found notice will shortly appear,
the installation is automatic
Finish the hardware installation
Code P010 Date: 20.2.200715
Inputs IN1-IN4 Symmetric inputs – the shield cable must be connected
Starting cable
S ignal cable
START (R EM O TE, … )
R ed
Shie ld ing
SYM M ETR IC AL C O N N EC TIO N
+ (+1V, H i, … )
- (-1V, Lo, … )
CO M (0, G N D )
R ed
Shie ld ing
W hite
Digita l output (Relay)
START (R EM O TE, … )
1V (100m V, H i, + , … )
0 (CO M , AG ND , … )
N O N SYM M ETR IC AL C O N N EC TIO N
R ed
S hie ld ing
R ed
Shie ld ing
W hite
Starting cable
S ignal cable
D igita l output (Relay)
Electric connections
Code P010 Date: 20.2.200716
Start Clarity using the desktop icon
Registration DialogNeeded for on-line updates
Login DialogEnter your name
Click here to open instrument
Clarity First Start
Code P010 Date: 20.2.200717
User Accounts user access privileges
Configuration configuration of the datastation
Directories working directories for individual instruments
Digital Outputs setting and assignment of digital outputs
System/Configuration
Configuration
Code P010 Date: 20.2.200718
To define users When several
operators use the station for different analyses
The desktop file (user profile) stores information about last used project, table and graph settings, last opened files, etc.
User Accounts window
Code P010 Date: 20.2.200719
Fill in accordance with your company rules for GLP and 21CFRpart11 compliance
Leave zeros for empty passwords or not restricted password life time.Leave blank for unprotected mode.
(All users will share common desktop file)
User can change his/her password in the User
Account Details dialog
User Accounts
Code P010 Date: 20.2.200720
Available Control modules
Selected instrument configuration (add devices)
Setting of displayed instruments (data acquisition allowed only on instruments with purchased licence)
Installed devices
Select the digital input and output for synchronisation
Instrument type selection
System Configuration window
Code P010 Date: 20.2.200721
Available Control
Modules list
Device configuration (doubleclick the device
icon)
System Configuration window
Code P010 Date: 20.2.200722
Directories window
Working directories enables working with data over network
Directory = where the available projects are searched forUseful for connecting of Eval version from other computer
Blank = Clarity instalation directory
Blank = LOG subdirectory of
Clarity instalation directory
Code P010 Date: 20.2.200723
Chromatography data storage according to user analysis type sample type time
Projects
Code P010 Date: 20.2.200724
Main Clarity directoryprogram files, configuration, user settings
COMMON Subdirectory report styles (*.sty), default method
templates, other files shared between projects
Projects Subdirectory project files info (*.prj)
Project Directorymethods (met.*), sequences (seq.*)
Project CALIB subdirectory calibration files (cal.*) and standard
chromatograms (*.prm)
Project DATA subdirectory sample chromatograms (*.prm)
File and Folders structure
Code P010 Date: 20.2.200725
RAW raw data file data as received from A/D converter
created at run start and incremented during analysis
actual and last processed stored in TMP subdirectory
PRM primary data integrated data
created when analysis is finished
contains RAW data, history and copy of calibration file
CAL calibration file contains data for calibrated results calculations
referenced from method and chromatograms
actual copies stored in saved chromatogram
Chromatogram files
Code P010 Date: 20.2.200726
Functionality Check
1) Open Data Acquisition window. The signal will drift if the signal cable wires are free and should be within a few µV from 0 mV if the are interconnected. You can also compare the detector display reading and reaction to autozero.
2) Upon the injection (closing the marker contact) the analysis will start. Data Acquisition window will change in appearance.
Note: The optional Validator can be used to check the system correct function