Insmeth Lecture 8 - Gas Chromatography

download Insmeth Lecture 8 - Gas Chromatography

of 33

Transcript of Insmeth Lecture 8 - Gas Chromatography

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    1/33

    Gas Chromatography

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    2/33

    General Design of a GC

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    3/33

    Some of the designdetailsGas supplies usually have either in-line or instrument

    mounted traps to remove any water, oxygen,hydrocarbons or other contaminants fromcompressed gases

    Instruments can have multiple injectors, detectors orcolumns

    Injectors and detectors usually have their owntemperature controlled zones (small heaters)

    The GC oven has a large fan and a vent door to helpwith rapid cooling of the oven

    Data collection (and integration) can be done usinga chart recorder, integrator or a computerized datasystem

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    4/33

    Separation Processes in GC The analyte is in the gas phase in the GC and partitions between

    the mobile phase (carrier gas) and the liquid stationary phase thatis coated on the inside of an open-tubular capillary column or onparticles inside a packed column

    Some packed-column GC uses non-coated solid stationaryphases, in which case one is performing gas-solid adsorptionchromatography

    Capillary, open-tubular (WCOT specifically) column GC is theprimary type of GC used in quantitative analysis: higher resolution = greater ability to discriminate between

    components smaller capacity of the column is not important as long as sufficient

    analyte is available for detection

    pg/mL (ppt) to g/mL (ppm) concentration range for liquid analytes

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    5/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    6/33

    The Objective in

    Chromatography (all types) Separate your analytes in the shortest amount of

    time possible and detect them.

    How can we do this in GC? Use different columns for different analyte types

    stationary phase

    diameter of column, stationary phase thickness

    column length

    Use different injection types/temperatures to optimizethe process of loading the sample on the column

    Use different temperature (or pressure) programs forthe column

    Select and use a detector that is suitable for theanalyte(s) of interest

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    7/33

    GC Columns (concentrating on open-tubular

    capillary columns)

    Column frame constructed of fused silica tubing Polyamide coating on the outside gives it strength

    Liquid stationary phases coated or bonded to the inside ofthe tubing

    0.1 - 0.53 mm + ID, 5-100 meters in length, stationary

    phases usually 0.10 to 1.5 m in thickness Mounted on a wire cage to make them easier to handle

    5-150 meters long.

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    8/33

    Capillary

    ColumnStationary

    Phases

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    9/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    10/33

    Choosing a GC Column Is the column compatible with your analytes

    polar analytes require polar stationary phases so theywill spend some of their time in the stationary phase

    non-polar analytes require non-polar stationary phases

    You usually have to compromise on the stationaryphase to get a good column for your analytes (whichare probably a mix of polar and non-polar)

    DB-5, HP-5, EC-5, RTX-5 (5% dimethyl, 95% diphenylpolysiloxane) most common general use column.

    Temperature range, solvent and carrier gascompatibility

    Sample capacity versus resolution usually determines packed vs.. capillary

    GCs usually setup for either packed or capillary

    Lets say you choose a capillary column, theresmore to think about!

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    11/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    12/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    13/33

    For capillary GC columns. Increased length = greater N, therefore a

    greater R

    expense is possible band broadening if analytes are onthe column too long!

    Increased length leads to longer separations. Do youhave the time?

    Increased stationary phase thickness and

    column diameter provides increased samplecapacity and can provide increased resolution tradeoffs are a longer analysis time and more column

    bleed with thicker stationary phases

    Is the column compatible with the detector? Thick stationary phases bleed more and will

    contaminate a mass spectrometer.

    For most analytical work, a best compromisecolumn is chosen and other variables (temp,etc.) are altered to optimize the separation.

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    14/33

    Capillary vs. Packed ColumnsCapillary Columns:

    Higher resolution (R)Greater HETP and N

    Shorter analysis time

    Greater sensitivity

    Most common in

    analytical laboratory GCinstruments

    Smaller sample capacity

    Higher cost/column

    Columns more

    susceptible to damage

    Packed ColumnsGreater sample capacity

    Lower cost (can make yourown)

    More rugged

    Most common in process labsor separating/determining

    major components in asample (prep GC)

    Limited lengths reduces R andN

    Not compatible with some

    GC detectors

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    15/33

    Temperature Programming in GC The simplest way to alter the separation in GC

    is to alter the temperature program in the oven.

    You can also alter the pressure of the carrier gas,but this is less common (much).

    Isothermal = constant temperature

    Gradient = varied temperature

    By altering the temperature, you vary the rate of

    the reaction for any analyte: they spend more or less time in the stationary phase the greater the difference in the times between

    analytes, the better the separation!

    AnalyteAnalyte phasestationaryphasemobile

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    16/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    17/33

    The traps of temperature

    If your temperature at a given time is too high,you can cause the peaks to co-elutepoor resolution vs but a faster separation

    If your temperature at a given time is too low,

    you can get still get a good separationadequate resolution, but a separation that takes

    very long

    You have to choose a compromisetemperature program

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    18/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    19/33

    GC Carrier Gases (the mobile phase) Usually inert gases (dont react with analytes except

    sometimes in the detector)

    Purpose: sweep sample through the column

    protect column from oxygen exposure at temperature

    assist with function of the detector

    Most common: Helium (available relatively pure without extensive purification

    after it leaves a compressed gas cylinder)

    Nitrogen (usually requires an oxygen and water trap)

    Hydrogen

    normally used only with flame ionization detectors (FID) since theFID needs it as fuel for the flame

    still rarely used due to safety concerns (and chromatographicones)

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    20/33

    GC Injection. Samples are injected through a septum:

    keeps oxygen out of the column

    provides a seal to keep the carrier gas pressure up atthe head of the columncarrier gas flow rate is determined by the pressure or the

    gas at the opening of the column

    Many different (mostly proprietary) materials

    red rubber (bleeds at about 250 C)Thermogreen (good up to about 300 C)

    High-temperature blue (good a little over 300 C)

    The injector is usually lined with a de-activatedglass liner prevents metal injector-sample reactions that would

    alter analytes or damage the metal of the injector

    Can be cleaned/replaced regularly

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    21/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    22/33

    Injection types

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    23/33

    On-Column Injection: used widely in packed-column GC, less in capillary GC

    sample is deposited directly on the column

    Good for thermally unstable compoundsGood for quantitative analysis at low concentrations

    all sample is available to travel to the detector

    BUT, you can inject only a relatively small amount ofsample in capillary GC anyhow.

    Splitless Injection: Sample is vaporized in the injector itself and ALL of the

    sample is swept onto the column by the carrier gas

    Again, relatively small samples are injected (10 L orless in capillary GC)

    Sample spends a large amount of time in the injector Best for trace (1 -100 ppm range) concentrations of

    high boiling point analytes in low boiling point solvents

    extra time in the injector helps volatilize the analytes.

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    24/33

    Split Injection:the injection is split, with only a portion of the

    sample (usually 1% - 20%) actually making it

    to the columnthe most common method of injecting

    samples onto small diameter, open-tubularcolumns.Even if you inject 20 L, only a fraction

    (adjustable) makes it on to the columnNot good for analytes with a wide range of

    boiling pointssome may be swept out the split vent before they

    are volatilized

    Modern capillary GCs come with aSplit/Splitless injectors standardyou switch between modes by changing

    the split vent gas flow and using a different

    injection liner.

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    25/33

    Dont Forget SPME (Solid Phase Microextraction)

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    26/33

    GC Detectors A dozen or more varieties (some obscure)

    Must be: sensitive to the analytes of interest compatible with the column, carrier gas, solvent, etc.

    rugged enough to withstand general unattended used

    Ive run our new GC for 36 hours straight without touching it!

    Should have a known linear range if the detector response is very linear, you can use a

    response factor instead of a calibration curve forquantitation!

    Usually require separate gas supplies (other than the

    carrier gas), have their own temperature control.Measure nothing more than a voltage or a current.

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    27/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    28/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    29/33

    ThermalConductivity (TCD)

    The carrier gas has a knownthermal conductivity.

    As the thermal conductivityof the column eluent (gasflow in) changes, the

    resistance of the filamentchanges.

    The presence of analytemolecules in the carrier gasalter the thermal conductivityof the gas (usually He)

    There is normally a secondfilament to act as areference (the carrier gas issplit)

    Increased sensitivity withdecreasing temperature

    (detector), flow rate andapplied current.

    Filaments will burn out(oxidized) in the presence ofoxygen if hot!

    Non-destructive

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    30/33

    FID D t ti l l t

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    31/33

    FID Destructive, sample lost. Analytes containing C

    burn in a hydrogen-oxygen flame andproduce ions

    CHO+ ions are collectedon a cathode and thecurrent they produceresults in the signal

    WILL NOT detect non-C

    containing compounds! Requires H2 supply (tank or

    generator) and O2 supply(compressed air)

    H2 carrier gas can be

    used, eliminating theneed for a supply for thedetector

    A makeup gas can alsobe required!

    -FlameO,H

    eCHOOCH22

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    32/33

  • 7/28/2019 Insmeth Lecture 8 - Gas Chromatography

    33/33

    ECD Particularly sensitive to halogens

    nitriles, carbonyls, nitro compounds

    Analytes pass through a cell, in whichelectrons are traveling between a 63Ni

    electrode and a collector electrode As analytes with electron capturing

    ability pass through the cell, the flowof electrons is interrupted.

    The change in current, due toreduced flow of electrons, isrecorded.

    EXTREMELY SENSITIVE TO HALOGENS

    could ruin detector with 1 ppm

    hexachlorocyclohexane by

    contaminating it with excess analyte

    Widely used for the determination ofpesticides, herbicides and PCBs inenvironmental samples.

    Non-destructive