Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of...

6
ORIGINAL ARTICLE Insertion torque, resonance frequency, and removal torque analysis of microimplants Yu-Chuan Tseng a,b , Chun-Chan Ting a , Je-Kang Du a , Chun-Ming Chen a,c , Ju-Hui Wu d , Hong-Sen Chen d, * a School of Dentistry, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan b Department of Orthodontics, Dental Clinics, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan c Department of Oral and Maxillofacial Surgery, Dental Clinics, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan d Faculty of Oral Hygiene, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung, Taiwan Received 6 April 2016; accepted 19 July 2016 Available online 24 August 2016 KEYWORDS Insertion torque; Orthodontic microimplant; Removal torque; Resonance frequency Abstract This study aimed to compare the insertion torque (IT), resonance frequency (RF), and removal torque (RT) among three microimplant brands. Thirty microimplants of the three brands were used as follows: Type A (titanium alloy, 1.5-mm 8-mm), Type B (stainless steel, 1.5-mm 8-mm), and Type C (titanium alloy, 1.5-mm 9-mm). A synthetic bone with a 2-mm cortical bone and bone marrow was used. Each microimplant was inserted into the synthetic bone, without predrilling, to a 7 mm depth. The IT, RF, and RTwere measured in both vertical and horizontal directions. One-way analysis of variance and Spearman’s rank correlation coef- ficient tests were used for intergroup and intragroup comparisons, respectively. In the vertical test, the ITs of Type C (7.8 Ncm) and Type B (7.5 Ncm) were significantly higher than that of Type A (4.4 Ncm). The RFs of Type C (11.5 kHz) and Type A (10.2 kHz) were significantly higher than that of Type B (7.5 kHz). Type C (7.4 Ncm) and Type B (7.3 Ncm) had significantly higher RTs than did Type A (4.1 Ncm). In the horizontal test, both the ITs and RTs were significantly higher for Type C, compared with Type A. No significant differences were found among the Conflicts of interest: All authors declare no conflicts of interest. * Corresponding author. Faculty of Oral Hygiene, School of Dental Medicine, Kaohsiung Medical University, Kaohsiung, Number 100, Shih-Chuan 1st Road, Kaohsiung 807, Taiwan. E-mail address: [email protected] (H.-S. Chen). http://dx.doi.org/10.1016/j.kjms.2016.07.007 1607-551X/Copyright ª 2016, Kaohsiung Medical University. Published by Elsevier Taiwan LLC. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/). Available online at www.sciencedirect.com ScienceDirect journal homepage: http://www.kjms-online.com Kaohsiung Journal of Medical Sciences (2016) 32, 469e474

Transcript of Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of...

Page 1: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

Kaohsiung Journal of Medical Sciences (2016) 32, 469e474

Available online at www.sciencedirect.com

ScienceDirect

journal homepage: http: / /www.kjms-onl ine.com

ORIGINAL ARTICLE

Insertion torque, resonance frequency, andremoval torque analysis of microimplants

Yu-Chuan Tseng a,b, Chun-Chan Ting a, Je-Kang Du a, Chun-Ming Chen a,c,Ju-Hui Wu d, Hong-Sen Chen d,*

a School of Dentistry, College of Dental Medicine, Kaohsiung Medical University, Kaohsiung,Taiwanb Department of Orthodontics, Dental Clinics, Kaohsiung Medical University Hospital,Kaohsiung, Taiwanc Department of Oral and Maxillofacial Surgery, Dental Clinics, Kaohsiung Medical UniversityHospital, Kaohsiung, Taiwand Faculty of Oral Hygiene, College of Dental Medicine, Kaohsiung Medical University,Kaohsiung, Taiwan

Received 6 April 2016; accepted 19 July 2016Available online 24 August 2016

KEYWORDSInsertion torque;Orthodonticmicroimplant;Removal torque;Resonance frequency

Conflicts of interest: All authors d* Corresponding author. Faculty of

Shih-Chuan 1st Road, Kaohsiung 807,E-mail address: [email protected]

http://dx.doi.org/10.1016/j.kjms.2011607-551X/Copyright ª 2016, KaohsiuCC BY-NC-ND license (http://creativ

Abstract This study aimed to compare the insertion torque (IT), resonance frequency (RF),and removal torque (RT) among three microimplant brands. Thirty microimplants of the threebrands were used as follows: Type A (titanium alloy, 1.5-mm � 8-mm), Type B (stainless steel,1.5-mm � 8-mm), and Type C (titanium alloy, 1.5-mm � 9-mm). A synthetic bone with a 2-mmcortical bone and bone marrow was used. Each microimplant was inserted into the syntheticbone, without predrilling, to a 7 mm depth. The IT, RF, and RT were measured in both verticaland horizontal directions. One-way analysis of variance and Spearman’s rank correlation coef-ficient tests were used for intergroup and intragroup comparisons, respectively. In the verticaltest, the ITs of Type C (7.8 Ncm) and Type B (7.5 Ncm) were significantly higher than that ofType A (4.4 Ncm). The RFs of Type C (11.5 kHz) and Type A (10.2 kHz) were significantly higherthan that of Type B (7.5 kHz). Type C (7.4 Ncm) and Type B (7.3 Ncm) had significantly higherRTs than did Type A (4.1 Ncm). In the horizontal test, both the ITs and RTs were significantlyhigher for Type C, compared with Type A. No significant differences were found among the

eclare no conflicts of interest.Oral Hygiene, School of Dental Medicine, Kaohsiung Medical University, Kaohsiung, Number 100,Taiwan.u.tw (H.-S. Chen).

6.07.007ng Medical University. Published by Elsevier Taiwan LLC. This is an open access article under theecommons.org/licenses/by-nc-nd/4.0/).

Page 2: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

470 Y.-C. Tseng et al.

groups, and the study hypothesis was accepted. Type A had the lowest inner/outer diameterratio and widest apical facing angle, engendering the lowest IT and highest RF values. Howev-er, no significant correlations in the IT, RF, and RT were observed among the three groups.Copyright ª 2016, Kaohsiung Medical University. Published by Elsevier Taiwan LLC. This is anopen access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).

Figure 1. The microimplants manufactured with threedesigned types, from left to right: Type A (1.5-mm � 8-mm),Type B (1.5-mm � 8-mm), and Type C (1.5-mm � 9-mm).

Introduction

Favorable anchorage design is a critical factor for suc-cessful orthodontic treatment. Orthodontic microimplantshave been verified as highly stable anchorage devicesexhibiting diverse applications for effectively overcomingthe difficulties encountered in orthodontic treatment. Thestability and reliability of microimplants enable the suc-cessfully controlling orthodontic forces, limiting undesiredtooth movements and correcting severe malocclusion. Or-thodontic microimplants have a success rate of 60e90%[1e3]; therefore, they can be used as an effective tool fororthodontic treatment.

The stability of orthodontic microimplants that areinserted into bones can be categorized into two types;primary and secondary. Primary stability is the initialstrength of the mechanical interlock between a microim-plant and bone, whereas secondary stability is a biologicalosseointegration between an orthodontic microimplant andbone during healing. However, orthodontic microimplantsare typically loaded with the orthodontic force immedi-ately or after a period of 2e3 weeks, unlike dental implantsthat require at least 4 months for bone integration (sec-ondary stability). Therefore, primary stability is the mostcritical concern in the application of orthodontic implants.

Different technologies have been employed to evaluatethe stability of orthodontic microimplants, and such tech-nologies include insertion torque (IT) [4,5], removal torque(RT) [6,7], and resonance frequency (RF) analysis [8e10].RF analysis is a noninvasive, harmless, repeatable, andreliable method that has been successfully and widely usedto measure the stability of dental implants. However, thismethod has seldom been used to study the stability of or-thodontic implants. Therefore, the objective of the currentstudy was to use the IT, RF, and RT analyses to investigateand compare the mechanical forces among three differentbrands of orthodontic microimplants.

Methods

As illustrated in Figure 1, 30 commercial orthodonticmicroimplants exhibiting three distinct features andbelonging to three different brands were used in this study,and they can be categorized as follows: Type A (titaniumalloy, 1.5-mm � 8-mm), Type B (stainless steel, 1.5-mm � 8-mm), and Type C (titanium alloy, 1.5-mm � 9-mm). From each of the three brands, five microimplantswere used for vertical tests (90�) and five for horizontaltests (0�). Both the vertical and horizontal tests couldinclude and interpret the degree of insertion of the clinicalcondition. Each test included IT, RF, and RT analyses.

Scanning electron microscope analysis (Hitachi SU8010,Tokyo, Japan) was applied to evaluate the surface featureof a thread (Figure 2). Under a clinical condition, a micro-implant is placed in the interdental alveolar bone, whichpossesses a 2-mm thick cortical plate. A synthetic bone(Sawbone, Pacific Research Laboratories Inc., Vashon Is-land, WA, USA) with a 2-mm thick cortical plate (40 pcf)was developed from rigid polyurethane foam. The densityof the cortical plate represented the relative densities ofthe maxillary and mandibular cortices, whereas the densityof the cancellous bone (20 pcf) represented that of thebone marrow.

Each microimplant was inserted into the synthetic bone,without predrilling, to a depth of 7 mm, leaving at least 1-mm gingival thickness for IT and RT measurements using adigital torque meter (Lutron, Taipei, Taiwan). The analyzer(Implomates, BioTech One, Inc., Taipei, Taiwan) was basedon the impulse force method and was used to measureresonance frequencies (Figure 3).

Page 3: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

Figure 2. Scanning electron microscope analysis (15 kV � 30, Hitachi SU8010, Japan): diameters and thread depths of the threetypes of mini-implants (from left to right: Type A, Type B, and Type C).

Figure 3. Resonance frequency analysis (Implomate;BioTech One, Inc., Taipei, Taiwan).

Table 1 The parameters (mm) of microimplants.

Microimplants Type A Type B Type C

Inner diameter 0.92 1.08 1.01Outer diameter 1.50 1.52 1.52Inner diameter/outerdiameter

0.61 0.71 0.66

Thread depth 0.29 0.22 0.25Thread pitch 0.61 0.67 0.66Apical facing angle(degree)

35.00 34.00 32.00

Coronal facing angle(degree)

17.00 15.00 14.00

Figure 4. The dimensions of microimplant.

Mechanical strengths of microimplants 471

Statistical analyses in this study were conducted usingSPSS (IBM SPSS 20, New York, USA). One way analysis ofvariance with Tukey Honestly Significant Difference (HSD)post comparison was applied to test the significant differ-ences among the three mechanical forces (IT, RF, and RT).Spearman’s rank correlation coefficient test was used toinvestigate correlations among the three experimentalvalues (IT, RF, and RT) during intragroup comparisons.Absence of correlations among the mechanical forcesamong the three brands used was considered as the nullhypothesis. Statistical significance was tested at p < 0.05.

Results

The dimensions of the three microimplants are presented inTable 1 and Figure 4. The inner diameter of Type B (1.08-mm) was greater than those of Type C (1.01-mm) andType A (0.92-mm). Similarly, the inner/outer diameter ratioof Type B (0.71) was higher than those of Type C (0.66) and

Type A (0.61). Type A (0.29-mm) had the highest threaddepth, compared with Type C (0.25-mm) and Type B (0.22-mm). Moreover, Type A demonstrated the greatest apicalfacing angle (35�) and coronal facing angle (17�), comparedwith the other two types.

Page 4: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

Figure 5. The summary of insertion torque (IT), resonancefrequency (RF), and removal torque (RT) in the vertical test.

472 Y.-C. Tseng et al.

Table 2 shows a summary of the IT, RF, and RT values,and Table 3 presents the results of the Tukey HSD postcomparison tests for the intergroup comparison. In thevertical test (Figure 5), the ITs of Type C (7.8 Ncm) andType B (7.5 Ncm) were significantly higher (p < 0.0001) thanthat of Type A (4.4 Ncm). Moreover, the RFs of Type C(11.5 kHz) and Type A (10.2 kHz) were significantly higher(p Z 0.005) than that of Type B (7.5 kHz). In addition, theRTs of Type C (7.4 Ncm) and Type B (7.3 Ncm) weresignificantly higher (p < 0.0001) than that of Type A (4.1Ncm).

In the horizontal test (Figure 6), the IT of Type C (6.4Ncm) was significantly higher (p Z 0.028) than that of TypeA (5.0 Ncm). The RFs of the three types were as follows:Type A, 9.8 kHz; Type C, 8.4 kHz; and Type B, 7.5 kHz. Nosignificant differences in RF were observed in the inter-group comparison (p Z 0.160). The RT of Type C (6.6 Ncm)was significantly higher (pZ 0.036) than that of Type A (4.7Ncm).

In the intragroup comparison, Spearman’s rank correla-tion test (Table 4) revealed no significant correlationsamong the IT, RF, and RT values in both the vertical andhorizontal tests. Therefore, the proposed null hypothesiswas accepted.

Table 2 Resonance frequency (kHz), insertion torque(Ncm), and pullout strength (Ncm) analysis ofmicroimplants.

Microimplant IT RF RT

Mean SD Mean SD Mean SD

VerticalType A 4.4 0.56 10.2 1.65 4.1 0.43Type B 7.5 0.79 7.5 0.49 7.3 1.17Type C 7.8 0.80 11.5 2.05 7.4 0.80

HorizontalType A 5.0 0.27 9.8 2.66 4.7 1.40Type B 6.1 1.17 7.5 0.91 6.0 1.00Type C 6.4 0.55 8.4 1.11 6.6 0.55

IT Z insertion torque; RF Z resonance frequency;RT Z removal torque; SD Z standard deviation.

Table 3 Statistical significance (p < 0.05) of intergroup inthe Tukey HSD post comparison.

ITVertical p < 0.0001; C > A, B > AHorizontal p Z 0.028; C > A

RFVertical p Z 0.005; C > B, A > BHorizontal p Z 0.160

RTVertical p < 0.0001; C > A, B > AHorizontal p Z 0.036; C > A

A Z Type A; B Z Type B; C Z Type C;IT Z insertion torque; RF Z resonance frequency;RT Z removal torque.

Discussion

In this study, the interradicular areas between the secondpremolar and first molar in the maxilla and mandible, themost common positions for the insertion of microimplants,were selected. Cha et al. [11] used an ultrasonic instrumentto detect the thickness of interradicular keratinizedgingiva, between the second premolar and first molar, inthe maxilla and mandible of men and women; they re-ported the thickness to be approximately 1-mm to 2-mm.Therefore, in the present study, we set the gingival thick-ness to 1-mm to 2-mm. However, because the Type C im-plants were 9-mm in length, the gingival thickness waseventually set to 2-mm so that each microimplant could beinserted at the same depth of 7-mm in the artificial bone (2-mm cortical bone).

Park and Cho [12] conducted a three-dimensional evalu-ation of interradicular spaces and cortical bone thickness,where microimplants are commonly inserted in clinicalpractice. They reported that the maxillary interradiculardistances ranged from 1.6-mm to 3.46-mm and tended toincrease from the cementoenamel junction to the apex.They were the greatest between the second premolar andthe first molar. Moreover, the mandibular interradiculardistances ranged from1.99-mm to 4.25-mmand tended to begreater than those of the maxilla. The thickness of themaxillary and mandibular buccal cortical bone ranged from1.12-mm to 1.33-mmand 1.25-mmto 2.98-mm, respectively,

Figure 6. The summary of insertion torque (IT), resonancefrequency (RF), and removal torque (RT) in the horizontal test.

Page 5: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

Table 4 Intragroup comparisons by Spearman’s rho rank correlation coefficient for hypothesis test (statistical significance,p < 0.05).

Microimplants Type A Type B Type C

Vertical Horizontal Vertical Horizontal Vertical Horizontal

IT vs. RF 0.300 �0.400 0.000 �0.410 �0.205 �0.740RF vs. RT 0.821 �0.700 0.460 0.158 �0.300 0.000IT vs. RT 0.051 �0.100 �0.410 �0.649 �0.564 �0.167

No statistical significance was found.IT Z insertion torque; RF Z resonance frequency; RT Z removal torque.

Mechanical strengths of microimplants 473

and demonstrated a tendency to increase from the cemen-toenamel junction to the apex in both jaws. To prevent or-thodontic microimplants from contacting the teeth ordamaging periodontal membrane tissue, each microimplantwas inserted at least 0.5-mm away from the adjacent roots.Therefore, in the present study, we selected 1.5-mm diam-eter microimplants.

When a screw is inserted into the bone, the continuousresistance force created between the bone and the screw isreferred to as the IT. Brown et al. [13] found that stainlesssteel miniscrews had significantly higher IT levels,compared with those made of titanium alloy. In the presentstudy, the IT of Type B was significantly higher than that ofType A, but not significantly different from that of Type C.Yoo et al. [14] studied the shape of miniimplants anddetermined that the IT of tapered miniimplants wassignificantly higher than that of cylindrical miniimplants.However, they reported that success rates of these twotypes of miniimplants were similar. Similarly, in the presentstudy, the IT of Type B (tapered shape) was significantlyhigher than that of Type A (cylindrical shape), but notsignificantly different from that of Type C (cylindricalshape).

In the present study, the inner/outer diameter ratios ofthe microimplants are outlined as follows: Type B, 0.71;Type C, 0.66; and Type A, 0.61. We observed that a higherinner/outer diameter ratio could lead to a higher IT.Regarding the application of both vertical and horizontalforces, the IT of Type A was significantly lower than those ofType B and Type C. In terms of thread design, Type Ademonstrated the highest thread depth, lowest threadinner diameter, and lowest inner/outer diameter ratio,compared with the other two types used in this study. Theapical facing angle of the Type A screw was the widestamong the three types, implying that the resistance expe-rienced during implant insertion was relatively low. Thethread depth and thread pitch diameter of the Type B andType C screws were similar, whereas the apical facing angleof the Type C screw was narrower than that of Type B.Hence, Type C exhibited a higher IT than did Type B. Tosummarize the preceding results, the IT is correlated withthe thread inner diameter, inner/outer diameter ratio, andapical face angle.

Since 1998, several reports have revealed that RFanalysis is a reliable and noninvasive method of detectingthe stability of dental implants [8e10]. Therefore, RFanalysis is also an efficient and safe method of assessingthe primary stability of orthodontic microimplants. Inter-group comparisons reveal that the RFs of Type C (11.5 kHz)

and Type A (10.2 kHz) were significantly higher than thatof Type B (7.5 kHz) in the vertical test. In the horizontaltest, despite the lack of significant differences, the RFs ofType A (9.8 kHz) and Type C (8.7 kHz) were higher thanthat of Type B (7.5 kHz). Furthermore, our results suggestthat the inner/diameter ratio could affect the RF level.Type B had the highest inner/outer diameter ratio andlowest thread depth. Hence, Type B demonstrated theleast anchorage on the artificial bone, thereby affectingits RF level.

Regarding the RT in both the vertical and horizontal di-rections, all the microimplants demonstrated a perfor-mance level similar to that observed for the IT. In otherwords, the RT of Type A was lower than those of Type B andType C in both the vertical and horizontal directions. Alower thread inner diameter and inner/outer diameterratio may imply lower resistance during RT. Notably, thecoronal facing angle of Type A was also wider than those ofType B and Type C. Because the RT involves anticlockwiserotation, a wider coronal facing angle necessitates a lowerstrength for RT. In our study, the Type B microimplantswere made of stainless steel and were tapered in shape,whereas the Type A and Type C microimplants were madeof titanium alloy and had a cylindrical shape. Therefore,there is still controversy that mechanical forces (IT, RF, andRT) depend upon the material composition and shape oforthodontic microimplants. Nienkemper et al. [15] reportedthat no significant correlation existed between the IT andRF. In the correlation analysis of the present study, the IT,RF, and RT were not significantly correlated in both thevertical and horizontal directions; therefore, our resultsare consistent with those of Nienkemper et al. [15]. Thisimplies that the IT cannot be used to infer the RF (primarystability), and similarly, the RF cannot be used to infer theRT. The lack of a significant correlation between the IT andRT indicates that the application of a higher force duringinsertion does not necessitate a larger force during micro-implant removal. The findings of our study reveal that alower inner/outer diameter ratio accompanied by a widerapical facing angle may reduce the IT and serve an energy-saving method for the placement of microimplants.Furthermore, a lower inner/outer diameter ratio accom-panied by a wider coronal facing angle may reduce the RTand serve as an energy-saving method for the removal ofmicroimplants. However, the RF was still unaffected by thedesign of threads.

In conclusion, Type A exhibited the lowest inner/outerdiameter ratio and the widest apical facing angle, leadingto the lowest IT and a higher RF values, compared with the

Page 6: Insertion torque, resonance frequency, and removal torque … · Table 1 The parameters (mm) of microimplants. Microimplants Type A Type B Type C Inner diameter 0.92 1.08 1.01 Outer

474 Y.-C. Tseng et al.

two other microimplants used in this study. The detaileddimensions of microimplants, including the inner diameter,inner/outer diameter ratio, thread pitch, thread depth,and apical as well as coronal face angles, are critical fac-tors affecting their mechanical strength.

Acknowledgments

This work was partially supported by a grant (KMUH99-9M30) from Kaohsiung Medical University Hospital, Taiwan.

References

[1] Tseng YC, Hsieh CH, Chen CH, Shen YS, Huang IY, Chen CM.The application of mini-implants for orthodontic anchorage.Int J Oral Maxillofac Surg 2006;35:704e7.

[2] Wu TY, Kuang SH, Wu CH. Factors associated with the stabilityof mini-implants for orthodontic anchorage: a study of 414samples in Taiwan. J Oral Maxillofac Surg 2009;67:1595e9.

[3] Rodriguez JC, Suarez F, Chan HL, Padial-Molina M, Wang HL.Implants for orthodontic anchorage: success rates and reasonsof failures. Implant Dent 2014;23:155e61.

[4] Meursinge Reynders RA, Ronchi L, Ladu L, van Etten-Jamaludin F, Bipat S. Insertion torque and success of ortho-dontic mini-implants: a systematic review. Am J Orthod Den-tofacial Orthop 2012;142:596e614.

[5] Wilmes B, Drescher D. Impact of bone quality, implant type,and implantation site preparation on insertion torques ofmini-implants used for orthodontic anchorage. Int J OralMaxillofac Surg 2011;40:697e703.

[6] Pithon MM, Nojima MG, Nojima LI. In vitro evaluation ofinsertion and removal torques of orthodontic mini-implants.Int J Oral Maxillofac Surg 2011;40:80e5.

[7] Okazaki J, Komasa Y, Sakai D, Kamada A, Ikeo T, Toda I, et al.A torque removal study on the primary stability of orthodontictitanium screw mini-implants in the cortical bone of dog fe-murs. Int J Oral Maxillofac Surg 2008;37:647e50.

[8] Kahraman S, Bal BT, Asar NV, Turkyilmaz I, Tozum TF. Clinicalstudy on the insertion torque and wireless resonance fre-quency analysis in the assessment of torque capacity andstability of self-tapping dental implants. J Oral Rehabil 2009;36:755e61.

[9] ChangWJ, Lee SY,WuCC, Lin CT, Abiko Y, Yamamichi N, et al. Anewly designed resonance frequency analysis device for dentalimplant stability detection. Dent Mater J 2007;26:665e71.

[10] Ito Y, Sato D, Yoneda S, Ito D, Kondo H, Kasugai S. Relevanceof resonance frequency analysis to evaluate dental implantstability: simulation and histomorphometrical animal experi-ments. Clin Oral Implants Res 2008;19:9e14.

[11] Cha BK, Lee YH, Lee NK, Choi DS, Baek SH. Soft tissue thick-ness for placement of an orthodontic miniscrew using an ul-trasonic device. Angle Orthod 2008;78:403e8.

[12] Park J, Cho HJ. Three-dimensional evaluation of inter-radicular spaces and cortical bone thickness for the place-ment and initial stability of microimplants in adults. Am JOrthod Dentofacial Orthop 2009;136:314.e1e314.e12.

[13] Brown RN, Sexton BE, Gabriel Chu TM, Katona TR, Stewart KT,Kyung HM, et al. Comparison of stainless steel and titaniumalloy orthodontic miniscrew implants: a mechanical and his-tologic analysis. Am J Orthod Dentofacial Orthop 2014;145:496e504.

[14] Yoo SH, Park YC, Hwang CJ, Kim JY, Choi EH, Cha JY. Acomparison of tapered and cylindrical miniscrew stability. EurJ Orthod 2014;36:557e62.

[15] Nienkemper M, Wilmes B, Pauls A, Drescher D. Mini-implantstability at the initial healing period: a clinical pilot study.Angle Orthod 2014;84:127e33.