InfusO.R. Syringe PumpSPECIFICATIONS
Product Code2L3100
AccuracyInfusion: -t- 3% linear rateBolus: +/- 3% linear displacementBattery life150 hours typical use and 4 hours at low battery
BatteriesFour C Size alkaline batteries (ANSI/NEDA 14A)
SyringesKendall Monoject" or B-D® 20 cc plasticKendall Monoject" or B-D® 60 cc plastic
Flow Rates0-600 mL/hr, depending on Smart Label
Flow ProfileInfusion: pulsed intermittentlyBolus: continuous
Occlusion Force (Pressure) ~---'8 ± 1 Ibs
Maximum Occlusion Pressure11 psi with 60 cc syringe20 psi with 20 cc syringe
Occlusion Detection Time*(Inversely related to flow rate)120 seconds at 36 mL/hr
Volume Stored on Occlusion*1.1 mL (approximate)
Back Pressure Effect on AccuracyNone to occlusion pressure
Size9.2 x 4.5 x 2.0 inches (238 x 113 x 50 mm)
Weight2 Ibs (1 kg) with batteries
* Using a microbore anesthesia set and a 60 cc syringe
ADMINISTRATION SETS
~n Ane~th(esi~Set with blue ~f?;~~ijU_M2~~N~~L~2C~
stripe - Sterile Package
~. ..... ..._*"'" ~ ~~;r;;;;~:JJ.Anesthesia Set with green stripe NO.2C9216
~. ~""""",,, ..k¥ .. u."..,,,", .b ... &c.~~~~:rr:=:IUniversal Anesthesia Set with yellowstripe - Higher viscosity drugs
SMART LABELSMedical Description
Cisatracurium
Dopamine
General mUhr
General imUhr-8oJus
Mivacurium
Propofol
Propofol-Pediatric Dose
Remifentanil-General Anesthesia
Remifentanil-MonitoredAnesthesia Care
. Sufernanil
NO.2C9218
MedicalConcentration I III.. ..
2 mg/mL 213260
4000 mcg/mL 2L3230,
50 mcg/mL 2L3243
Not Applicable 2L3241
Not Applicable 2L3225
2 mg/mL 2L3248
10 mg/mL 2L3242
10 mg/mL 2L3259
50 mcg/mL 2L3269
./
./
./
./
./
./
25 mcg/mL 2L3268 ./
50 mcg/mL
Sufentanil ./
ntan il~f1ediatric Dose
Vecuronium
5 mcg/mL
10 mcg/mL
1 mg/mL
2L3249
2L3240 ./
./2L3227
For more information contact yourBaxter Sales Representative,
or call 1-800-933-0303.Rx Only. For safe and proper use of this device, please refer to the completeinstructions in the user's manual.
COMMITTE~fsafer~HEALTHCARE ENVIRONMENT
IVIE.~IC::.A."TICJ~ DELIVERY
Baxter, InfusorO.R and Committed to a Safer Healthcare Environment are trademarks of Baxter International Inc.B~D and Monoject are registered trademarks of Becton-Dickinson Co. and Tyco Healthcare Group Lp, Inc.Baxter Healthcare Corporation, Medication Delivery. Route 120 and Wilson Road, Round Lake, IL 60073 www.baxter.com 420255 5K 10/08