Infant Flow® LP system headgear
![Page 2: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/2.jpg)
![Page 3: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/3.jpg)
Prepare Infant Flow® SiPAP driver ....................1Infant Flow LP components ............................2Generator assembly components ...................3Low pressure generator .................................4Nasal prongs ..................................................5Nasal mask ....................................................6Nasal prong and mask sizing ..........................7Headgear sizing .............................................8Generator assembly application .....................9 Headgear application ........................... 9–11 Check for proper headgear placement ......12 Incorrect headgear placement ............ 13–14 Attach interface to generator ...................15 Remove support cradle ............................16 Apply support cradle to headgear ............17 Align generator assembly .........................18
Secure locking strap .................................19 Attach side straps to generator ................20Final check of generator assembly ................21 Check for correct headgear placement ......21 Check placement of exhalation tube ........22 Check for proper nasal prong position .......23 Incorrect nasal prong position ..................24 Verify proper mask placement ..................25 Incorrect mask placement ........................26 Incorrect mask size—too small .................27 Incorrect mask size—too large .................28Warnings and cautions ................................29Application tips ............................................30Inspection ....................................................31Notes ...........................................................32
Table of contents
![Page 4: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/4.jpg)
1
100
90
80
706050
40
30
21
2
1
2
3
4
5
L/minL/min
NCPAP/Pres Low
Pres High
Infant FlowSiPAP
4
6
8
10
12
14
XDCR PPROX
1. Attach the infant circuit to the humidifier
and nCPAP driver.
2. Attach the generator to the circuit.
3. Perform oxygen sensor calibration and
user verification.
4. Enter the desired settings.
Prepare Infant Flow SiPAP driver
3 4
1
2
![Page 5: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/5.jpg)
2
A
B
E
FD
C
A. Mask and prongs
B. Generator
C. Bonnet sizing guide
D. Mask/Prong sizing guide
E. Headgear
F. Bonnet
Infant Flow LP components
![Page 6: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/6.jpg)
3
Generator assembly components
A. Exhalation tube
B. Securing strap
C. Generator receiver
D. Fixation tab
E. Mask/Prong
attachment base
F. Support cradle
G. Pressure and drive lines
C
B
F
E
A
G
D
![Page 7: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/7.jpg)
4
Low pressure generator
• Lower drive pressure
– Prevents pressure backup into the
water auto-feed system
• Pressure relief valve
Pressure relief valve
![Page 8: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/8.jpg)
5
Nasal prongs
A. Flared tip
B. Flexible bellow
C. Base
D. Size indicator
E. Key
A
A
BC
E
B
C
D
D
Anatomically designed
![Page 9: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/9.jpg)
6
Nasal mask
Anatomically designed
A. Nasal bridge cushion
B. Contour shape
C. Key
D. Flexible bellow
E. Assessment window
F. Size indicator
A
B
C
E
E
D
F
D
![Page 10: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/10.jpg)
7
• Use the sizing guide to determine the
appropriate mask or prong size.
• Nasal prongs: Choose the dots that fill the
nare space.
• If between sizes, choose the larger size.
• Nasal masks: Choose the triangle that fits
over the nose.
Nasal prong and mask sizing
![Page 11: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/11.jpg)
8
Headgear sizing
Measure the circumference of the head; from the crown at back of the head to
the brow line (in centimeters).
![Page 12: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/12.jpg)
9
1
2
Headgear application
1. Align the headgear with the midline
of the face.
2. Align the bottom of the headgear with
the nape of the neck.
Generator assembly application
![Page 13: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/13.jpg)
10
3
4
Headgear application (continued)
3. Place strap 1 just above the brow line.
Note: Straps are labeled 1, 2 and 3.
4. Place strap 2 midline over strap 1.
![Page 14: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/14.jpg)
11
5
6
Headgear application (continued)
5. Place strap 3 on top of straps 1 and 2.
Note: The strap should be snug—do
not overtighten.
6. Fold strap 2 up and over strap 3,
and secure it to the top strap.
![Page 15: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/15.jpg)
12
Check for proper headgear placement
• The back of the headgear rests on the
nape of the neck.
• The forehead straps rest on the brow line.
![Page 16: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/16.jpg)
13
Incorrect headgear placement
• The horizontal strap rests close to the eye.
• The forehead strap rests too far above
the brow line.
![Page 17: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/17.jpg)
14
Incorrect headgear placement (continued)
• A headgear fold/bulge appears above
the nape of the neck.
• The top strap is too tight, and forehead
strap 3 is pulled back.
![Page 18: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/18.jpg)
15
1
3 4
2
Attach interface to generator
Align the notch on the prongs/mask to straddle the pressure lines.
![Page 19: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/19.jpg)
16
1
2
Remove support cradle
1. Expand the collapsible corrugated tubing.
2. Remove the support cradle from the
generator assembly.
![Page 20: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/20.jpg)
17
1
2
Apply support cradle to headgear
1. Center the support cradle properly
on the straps.
2. Secure the center tab.
![Page 21: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/21.jpg)
18
1 2 3
Align generator assembly
1. Center the generator and interface over
the patient’s nose and support cradle.
2. Insert the nasal prongs using a gentle
rocking motion.
3. Place the drive line and pressure line in the
support cradle with the exhalation tube
resting on top.
![Page 22: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/22.jpg)
19
1
2
Secure locking strap
1. Wrap the locking strap over the
exhalation tube, and attach it to the
side of the support cradle.
2. Position the small slit over the ridge of
the exhalation tube.
![Page 23: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/23.jpg)
20
1 2
3
Attach side straps to generator
1. Wrap the side straps around the
generator tabs.
2. Use the Velcro® tip to secure the
strap—loose is better.
3. Use the side straps to adjust the
tension on the interface—do
not overtighten.
![Page 24: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/24.jpg)
21
Check for correct headgear placement
A. The generator aligns and centers with
the nose.
B. The side straps are properly placed.
C. The support cradle tabs are secured
to the headgear.
Note: The straps should be
snug—not tight. B
A
C
Final check of generator assembly
![Page 25: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/25.jpg)
22
Incorrect Correct
Check placement of exhalation tube
![Page 26: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/26.jpg)
23
Check for proper nasal prong position
• The bellows are visible.
• The septum is visible.
• The generator assembly does not
contact skin.
![Page 27: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/27.jpg)
24
• The mask is too tight, and the bellows
are compressed.
• The prong is not in the nare.
Incorrect nasal prong position
![Page 28: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/28.jpg)
25
Verify proper mask placement
• The mask covers the nasal profile.
• The mask rests below the eyes.
• The bellows are not compressed.
• The side strap rests low on the cheek.
Note: Do not overtighten.
![Page 29: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/29.jpg)
26
Incorrect mask placement
• Too much tension applies to one side.
• The mask is collapsed.
• The nares are not visible.
![Page 30: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/30.jpg)
27
Incorrect mask size—too small
• The mask blocks the nares.
• The mask cannot cover the nose.
![Page 31: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/31.jpg)
28
Incorrect mask size—too large
• The mask rests on the bridge of
the nose.
• The mask rests against the eyes.
• The mask rests on the upper lip.
![Page 32: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/32.jpg)
29
Warnings and cautions
• Use this product only as directed in the product literature to reduce risk of
nasal irritation, septal distortion, skin irritation and pressure necrosis.
• Use this product only as a trained practitioner, under a qualified physician’s
direct supervision.
• Use the Infant Flow LP generator only with variable flow nCPAP drivers.
• Do not overtighten the fixation straps.
![Page 33: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/33.jpg)
30
Application tips
• Select the appropriate nasal mask size to minimize leaks and dead space.
• Select the appropriate nasal prong size; if between sizes, select the larger size.
• To avoid affecting the generator’s stability, only use a correctly sized prong, mask, bonnet
or headgear.
• Adjust the straps to stabilize the generator, and maintain a seal at the nose using the least
tension possible.
• Consider alternating the use of prong and mask interfaces at set intervals to change the
pressure points on the face.
• Continuously monitor the respiratory status.
• Cover both ears evenly; ensure the ears are not folded.
• Use humidification with nCPAP systems.
![Page 34: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/34.jpg)
31
Inspect the system at least every three to four hours to:
• Ensure clearly visible eyes and unblocked nares if using a mask.
• Check the fixation device and straps for proper tension, adjusting them as needed to
maintain a proper fit.
• Monitor gastric insufflation and abdominal distension.
• Monitor for excessive condensation in the circuit and generator.
• Alternate the use of prong and mask interfaces to change the pressure points on the face.
Inspection
![Page 35: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/35.jpg)
32
Notes
![Page 36: Infant Flow LP system headgear - CareFusion...• Use the Infant Flow LP generator only with variable flow nCPAP drivers. • Do not overtighten the fixation straps. 30 Application](https://reader033.fdocuments.net/reader033/viewer/2022041720/5e4e4a107c20e576a851e143/html5/thumbnails/36.jpg)