IN VITRO DEGRADATION BEHAVIOUR AND FRACTURE MECHANISM OF Mg-0.8Ca UNDER A CONSTANT STRAIN

download IN VITRO DEGRADATION BEHAVIOUR AND FRACTURE MECHANISM OF Mg-0.8Ca UNDER A CONSTANT STRAIN

of 16

description

Master degree proposal.

Transcript of IN VITRO DEGRADATION BEHAVIOUR AND FRACTURE MECHANISM OF Mg-0.8Ca UNDER A CONSTANT STRAIN

  • In vitro degradation behaviour and fracture mechanism of Mg-0.8Ca under a constant strain

    Nur Amirah binti Shaharom Supervised by: i) Dr. Hendra Hermawan ii) Assoc. Prof. Ir. Dr. Mohammed Rafiq bin Abdul Kadir

    Abstract

    Magnesium-based alloys are currently attracted much interest owing to its

    potential for biodegradable implants application. However, despite many of its

    advantages which are superior to those of current implant materials, tackling the

    rate of degradation of magnesium alloys is one of the grand challenges

    highlighted. Mg-0.8Ca is believed to have optimum qualities to serve as implant

    material, thus have been chosen as the testing materials in present work. Failure

    mechanism of tested material is proposed to be evaluated to further understand

    its degradation behaviour under synergistic presence of mechanical loading

    along with corrosive physiological environment. It is believed that under such

    condition the materials are highly susceptible to stress corrosion cracking (SCC)

    which might lead to catastrophic complications and sudden failure. This study

    aims to evaluate the susceptibility Mg-0.8Ca towards SCC in physiological

    environment, determine the mode of SCC crack propagation and recognise the

    causes of failure which will be depicted from the propose testing methods. The

    understanding of failure mechanism will give contributions in the improvement of

    material properties in order to be successfully applied as biodegradable implants.

  • Introduction

    Metallic biomaterials are predominately applied for impaired bone tissue fixture and

    replacement [1] due to comparable monotonic strength they exhibit for such purposes.

    However, evaluation on current metallic materials used [1-5], i.e. stainless steel,

    titanium and cobalt-chromium-based alloys, revealed a deleterious complications due to

    debris produced from loading-related scenarios, in combination with corrosive

    environment of the implantation site. Stiffness ratio of current metallic implants toward

    the adjacent bones is relatively huge leading to various clinical issues due to stress

    shielding effects [5, 6]. Table 1 below shows typical properties of current metallic

    materials in comparison to those of human bone.

    Table 1 Mechanical properties of commonly used metallic biomaterials and human bone [4]

    Stainless

    steel Cobalt-chromium

    alloy Titanium

    alloy Magnesium

    Human bone

    Density (g cm-3

    ) 7.9 8.1 8.3 9.2 4.4 4.5 1.74 2 1.8 2.1

    Youngs modulus (GPa)

    189 205 230 110 117 41 45 3 20

    Compressive yield strength (MPa)

    170 310 450 1000 758 115 65 100 130 180

    Development of orthopedic implants had recently shifted from classic standard of

    material selection to a rather socially and economically beneficial. The selection is now

    demanded the materials to be biologically active in enhancing physiological response

    and implant-tissue interactions while progressively degrade during the healing period of

    damaged tissue [5]. Designated to avoid secondary surgical procedure for implant

    removal, biodegradable magnesium-based implants which generate mainly soluble,

    non-toxic degradation products have been generally approved to serve as bone fixation

    devices [2, 5, 7]. With sufficient mechanical strength and load-bearing capacity,

    magnesium-based implants exceeded cotemporary polymer based implants for

    orthopaedic applications. Furthermore, Youngs modulus of magnesium-based implants

    is a lot closer to human bone compared to permanent metallic implants currently used

  • [4] which may contribute to the reduction of stress shielding effects [5]. However, the

    rate of degradation of magnesium-based alloys is in biggest concern [1, 2, 4, 5, 8, 9] as

    the degradation process evolves intolerant hydrogen gas accumulation and the implant

    might lose its mechanical integrity before the fractured bones completely heal [4, 10].

    Generally, poor corrosion resistance in magnesium alloys is believed to be influenced

    by a few factors which include [11] (a) corrosion product films are too weak to inhibit the

    high intrinsic dissolution tendency of magnesium, and (b) speedup of micro-galvanic

    localized corrosion due to the presence of second phases serving as local cathodes. It

    is believed that understanding degradation mechanism of magnesium alloys is the best

    perspective in establishing a material with optimised properties most suited the implant

    applications.

    Alloying is one of the methods currently developed to enhance magnesium mechanical

    properties and corrosion resistance [2, 4, 5, 7, 11-16]. Addition of alloying elements has

    been observed analytically producing magnesium alloys with a lower corrosion rate than

    that of high purity magnesium in 3% NaCl [11]. Among various alloying elements

    available, calcium has been favourably chosen and considered as a unique alloying

    addition owing to Mg2Ca phase features exhibited in low alloyed Mg-Ca systems which

    further affect its microstructure, mechanical properties, electrochemical behaviour and

    degradation kinetics [5, 7, 14, 15]. Addition of calcium in Mg-Ca alloys reduces grain

    size and improves general and pitting corrosion resistance as well [5, 14]. Li et al. [14]

    highlighted a few considerations in choosing calcium as alloying element in magnesium

    for biodegradable material which include the composition of calcium in human body, its

    close density to those of human bones, and the need of magnesium in calcium

    incorporation into the bone have been reported [17] with potential benefit for bone

    healing. Corrosion in Mg-Ca alloys with calcium content up to 0.8 wt% is well distributed

    and homogenous in texture while those with higher calcium content demonstrate

    uncertain, dissipated corrosion. Magnesium with 0.6 0.8 wt% calcium has been

    proven exhibited slowest degradation rate [9, 15] and good biocompatibility [7].

  • Figure 1 Hydrogen evolution of Mg-Y in 0.1M NaCl. The rate of degradation can be represented by instantaneous

    value from hydrogen evolution curve or by average value over a particular time period [11].

    Most assessments done in vitro considered only on general practice of magnesium

    alloys for orthopaedic implants, instead of simulating the implantation site condition

    specifically. Improvement in corrosion resistance and mechanical properties focused on

    the effect of alloying element and mechanical processes only, and despite of failure

    reported experienced by current metallic biomaterial devices, no further assessment on

    the mechanism of failure in corrosive environment for biodegradable implants has been

    made despite of the fact that SCC in magnesium alloys would give severe effects to the

    materials when in service. Thus, the present work utilizes every aspects of Mg-0.8Ca

    which have been evaluated as the absolute choice of material to be applied specifically

    for craniofacial fixation devices. Since the magnesium alloys corrode very fast in

    chloride environment, the occurrence of SCC needs to be ascertained whether the loss

    of mechanical properties is resulted from the SCC or just the reduction in specimen

    cross-section due to anodic dissolution [18]. Understanding fracture mechanism

    endures by Mg-0.8Ca under constant strain in simulated physiological media is

    expected to contribute further improvement on its corrosion resistance and service life.

  • Literature Reviews

    Current developments of metallic materials encompass groups of materials with fine

    interaction towards biological environment and capable to progressively degrade during

    the healing process of damaged tissue [5]. Instead of relying upon polymers to be

    applied as implant materials, biodegradable metallic substitutes are worthier to be

    established on account of better load-bearing capacity and mechanical strength they

    demonstrate [5]. Magnesium and magnesium-based alloys are currently drawing much

    interest as promising candidates to be applicable for such purposes. Current research

    development on magnesium-based biodegradable implants is focusing mostly on Mg-Al

    and Mg-RE alloy systems [14]. In vitro assessment [19] of degradation rate and

    mechanism for magnesium has been summarized by Kirkland et al. detailing their

    advantages and limitations.

    Witte et al. [20] have conducted an investigation to compare the corrosion rates of

    AZ91D and LAE442 and concluded that corrosion rates measured in vivo were orders

    of magnitude lower than those measured in substitute ocean water prepared according

    to ASTM-D1141-98 [21]. In vitro test has been accomplished using various solutions

    which include; (i) simulated body fluid (SBF), (ii) artificial plasma (AP), (iii) phosphate

    buffer saline (PBS), and (iv) minimum essential medium (MEM, Invitrogen). In vitro test

    [16] conducted to evaluate the corrosion rate of various magnesium alloys reported that

    high purity magnesium had the lowest corrosion rate which increased with increasing

    amount and effectiveness of second phase. Song [16] concluded that Zn-containing

    magnesium alloy is the most promising material to be developed into a biodegradable

    material with some modifications such as anodised coating to delay the degradation and

    purification to reduce the rate.

    A study [22] conducted to understand failure mechanisms of commercial Kntscher

    nails of type 316L stainless steel under the combined action of corrosion and stress has

    been done by G. Bombara and M. Cavallini. Evaluation on damage morphology of the

    device showed transgranular, branched and unblunted cracks which were first identified

    as SCC behaviour rather than corrosion fatigue (CF). This was corroborated with

  • appearances of minor and indeterminate fretting fractographs using scanning electron

    microscope (SEM). The results obtained agreed with investigation done by M.

    Sivakumar and S. Rajeswari [23] which concluded SCC of failed intramedullary nail has

    been worsened by high inclusion content and large grain size of the implant.

    Fractography of failed commercial AISI stainless-steel 316L presented by S.

    Bandyopadhyay and P. Brockhurst [24] has shown evidence of SCC failure with crack

    propagation both in intergranular and transgranular modes which determined the

    susceptibility of austenitic stainless steel at a rather low temperature than mostly

    reported.

    Kannan and Raman [25] had claimed that SCC susceptibility of Mg-Al-Zn alloy is not

    substantial and should not be a concern for implant applications despite of stress-

    assisted failures that have been reported [2, 12, 26, 27] for current orthopaedic

    implants. However, a critical review done by Atrens et al. [11] highlighted a number of

    grand challenges in development of magnesium-based implants, which include

    consideration of SCC due to corrosion-assisted stresses that is known existed in

    implant-bone systems. This review is in agreement with investigation by Choudhary and

    Raman [18] which had confirmed the SCC susceptibility of both smooth and notched

    specimens of a magnesium alloy in physiological environment. In a different

    investigation that has been done by Choudhary et al. [28] on AZ91D in modified SBF

    and air at different strain rates revealed that the alloy is susceptible to SCC.

  • Materials and Methods

    Sample preparation

    Samples will be used are cast Mg-0.8Ca samples whose geometry and dimensions

    presented in Figure 2. The sample will be ground successively to 1200 grit using SiC

    paper, ultrasonically cleaned in acetone, and dried to room temperature for 24 hours

    prior to testing.

    Figure 1 Dimension of the proposed sample.

    Microstructural characterisation

    The samples which will be used for microstructure characterisation will be mechanically

    ground with SiC paper up to 2500 grit and polished, follow by chemical etching in 2%

    Nital solution. Surface morphology of the samples before and after the introduction of

    strain will be observed using optical microscope. Furthermore, microstructure of

    samples surface will be examined using SEM before and after the samples failure.

    Failed samples will be first cleaned chemically using 20% CrO3 and 10% AgNO3

    solution. The changes of microstructure morphology will be observed to further

    understand the influence of strain towards failure of samples and is believed to be

    closely related to the fracture mechanism. The characterisation of failed samples is

    expected to have features of brittle and/or ductile fracture, and show the mode of

    cracking either transgranular or intergranular.

  • Immersion test

    The samples will be assembled to a custom-made apparatus for immersion test in

    simulated body fluid (SBF) with pH value will be adjusted to 7.4 and strain variation of

    0.25% and 0.5% elongation. The test will be done for 1, 2, 4, and 6 hours to evaluate

    the rate of degradation under such condition from the weight loss percentage, and then

    prolonged until failure. Each test will be performed in triplets in order to examine the

    reproducibility. Another set of immersion test will be conducted as control experiment.

    The samples for the control experiment will not be put under strain, but evaluate the

    same measurement of weight loss for the same time intervals, which will then be

    prolonged until failure. All samples will be taken out, left to dry at room temperature for

    24 hours and be measured of the weight loss, Wa before chemical cleaning. Then, the

    samples will be cleaned with chromic acid solution to remove corrosion products and let

    dry to room temperature before the final weight loss, Wb will be recorded.

    Wb from failed samples from immersion test with introduction of strain will be compared

    to Wb of the samples from immersion test without strain at the same failure time. The

    microstructure of both samples will be examined and is expected to verify the effects of

    strain on the samples microstructure and their influences on degradation rate.

    In situ observation

    A sample from the immersion test with strain value of 0.5% will be randomly picked for

    in situ observation of failure using stereo microscope. Fractography and failure

    mechanism are expected to be better visualised using this method.

  • Expected results and discussion

    The degradation rate will be computed from experiments conducted in different

    conditions; (i) sample will be put under strain in SBF, (ii) sample will be put under strain

    in air, and (iii) sample will be immersed in SBF for 24 hours then be put under strain in

    air. Each of the experiments however, will be repeated thrice. The variation in testing

    condition is to recognise the causes of failure [18]; either it is indeed a result of the

    synergistic effect of stress and corrosive environment (i.e. SCC) or simply attributed to

    the continuously reducing cross-sectional area of the sample due to high corrosion rate

    of magnesium alloys.

    Table 2 Design of experiment that is expected to be achieved with some data obtained from current conducted

    experiments.

    Media Displacement Time to failure Weight Loss pH Degradation Rate

    SBF

    EXPERIMENT 1 - Degradation Rate of MgCa Under Tension Stress

    0.50%

    Time (min) W W W - W Wb W - Wb pH pHi (W - Wb)/t

    60 1.87 1.869 0.001 1.838 0.032 7.2 7.9 0.000533333

    120 1.89 1.886 0.004 1.852 0.038 7.2 7.9 0.000316667

    240 1.929 1.927 0.002 1.835 0.094 7.2 8.2 0.000391667

    360 1.803 1.804 -0.001 1.656 0.147 7.4 8.4 0.000408333

    tf - 960 1.94 1.86 0.08 1.654 0.286 7.3 9.2 0.000297917

    EXPERIMENT 2 - Degradation Rate of MgCa Under Tension Stress

    0.25%

    Time (s) W W W - W Wb W - Wb pH pHi (W - Wb)/t

    60 1.75 1.78 -0.03 1.655 0.095 7.1 7.6 0.001583333

    120 1.78 1.82 -0.04 1.727 0.053 7.1 7.6 0.000441667

    240 1.88 1.9 -0.02 1.788 0.092 7.1 7.8 0.000383333

    360 1.75 1.73 0.02 1.64 0.11 7.1 7.9 0.000305556

    T2 W20

    EXPERIMENT 3 - Degradation Rate of MgCa Under Tension Stress

    0%

    W W W - W Wb W - Wb pH pHi

    1h W21

    2h W22

    4h W23

    6h W24

    T5 W25

  • Results obtained in Table 3 are from the first round of experiments. From the data,

    some graphical evaluation could be made as shows in Table 3. Computing stress vs.

    time curves will show changes in mechanical properties of the sample tested for each

    assessed conditions. It is expected that maximum reduction in mechanical properties

    which might fail the fastest among the samples tested is the one under strain in SBF. It

    is expected that stress-time curve for the samples that will be pulled in air indicate some

    similar behaviours in comparison to those pre-immersed samples.

    Table 3 Degradation rate evaluated for current conducted experiments.

    Fractograph of the sample that will be tested under strain in SBF is expected to show

    common mode for SCC crack propagation of magnesium alloys, i.e. localised attack at

    the sample circumference which in higher magnification, will reveal transgranular

    cracking and localized cracks (Figure 3).

    60 120 240 360 tf - 960

    0.50% 0.00053333 0.00031667 0.00039167 0.00040833 0.00029792

    0.25% 0.00158333 0.00044167 0.00038333 0.00030556

    0

    0.0002

    0.0004

    0.0006

    0.0008

    0.001

    0.0012

    0.0014

    0.0016

    0.0018

    Weig

    ht

    Lo

    ss (

    g)

    Rate of Degradation (gm-1)

    Degradation rates of Mg-0.8Ca under various strain values in SBF

  • Figure 2 Factograph of failed sample with common mode of SCC cracks propagation for magnesium alloys: (a)

    overall fracture surface, and (b) evidence of trransgranular cracking [18].

    The experiment will use weight loss method in evaluation of degradation rate of Mg-

    0.8Ca sample. At least two from three types of measurement, i.e. weight loss rate,

    hydrogen evolution rate and rate of Mg2+ leaving the metal surface, are recommended

    by Atrens et al. [11] to be implemented to better clarify the rate of corrosion on

    magnesium alloy in service. However, since this study is not focus on the rate of

    corrosion, only weight loss rate will be measured to give a framework of service and

    learn the impact of such simulated condition on the specimen towards fracture. It is

    believed that the rate of corrosion is closely related to the failure mechanism of

    implants.

    There are two types of apparatus for transgranular SCC evaluation, the Linearly

    Increasing Stress Test (LIST) and the Constant Extension Rate Test (CERT). Review

    done by Atrens et al. [11] described that both of these methods are useful to identify the

    occurrences of SCC and have capabilities to measure the threshold stress and crack

    velocity from the final crack size divided by the time of cracking when coupled in

    identification of crack initiation. The increased crack propagation time under CERT

    conditions may be important in determining the mechanism for SCC in magnesium

    alloys. However, the experimental setup presented in this study will be custom-made,

    an apparatus that simulate the condition which initiate conventional SCC. There may be

  • a room for improvement by applying either LIST or CERT, or both to better evaluate

    failure mechanism of proposed material.

    Choudhary and Raman [18] had done SSRT under different testing conditions which

    included continuous cathodic charging and simultaneous strain in SBF for indication of

    possible combined effect of hydrogen-induced cracking and anodic dissolution.

    Furthermore, circumferential notch tensile (CNT) testing was used in their work to

    determine the threshold stress intensity factor for SCC (KISCC) of the tested samples in

    SBF. SEM fractographs (Figure 4) of the failed samples from CNT test had provided

    better description on the failure mechanism.

    Figure 3 SEM fractographs obtained from CNT: (a) the overall fracture surface showing machined notch, fatigue pre-

    crack, SCC and mechanical failure zones, (b) fatigue pre-cracked and SCC zones (arrow indicate crack propagation

    direction), and (c) mechanical failure zone showing dimples [18].

  • Intergranular in magnesium alloys occurs typically by micro-galvanic corrosion, when

    there is a presence of continuous or nearly continuous second phase along the grain

    boundaries, a microstructure typical of many cast creep-resistant magnesium alloys. In

    contrast, transgranular SCC occurs through the -phase magnesium matrix. From the

    review done by Atrens et al. [11], they concluded that SCC is associated with

    environmental conditions leading to the local breakdown of a partially protective surface

    film which is usually caused by localized corrosion by chloride ions. However, load may

    be influenced the breakdown as well since the occurrence of SCC is observed on pure

    magnesium and magnesium alloys (AZ91, AM60, AS41, ZK60A-T5) in distilled water.

    The influenced of hydrogen to the fracture behaviour and the effect of alloying elements

    on SCC propagation are not well understood currently. In advance, the understanding of

    microstructure influences on SCC is required to better understand magnesium

    transgranular SCC. Furthermore, the understanding of effective hydrogen diffusion rate

    to the fracture process zone needs to be included which requires an understanding of

    hydrogen-trap interactions in magnesium alloys. Although Kannan and Raman [13]

    found insignificant decreases of strength and ductility for AZ91 in simulated body fluid

    using CERT which indicate SCC, such failure is still largely unexplored.

  • Conclusion

    Current studies on the possibilities to implement magnesium-based alloys as

    degradable implants demanding a much deeper understanding of the degradation

    behaviour, particularly involving microstructural analysis, to tackle rapid corrosion of the

    material in physiological environment. SCC is one of many failures recognised among

    metallic materials, thus it is demanded to have absolute evaluation on such issues to

    provide crucial information on whether the materials will be tolerable for the given

    service time of the device in practical. Such studies have been done on industrial

    magnesium alloys, which are less likely applicable as implant materials. Thus, the

    present work emphasised the SCC susceptibility assessment and the mechanism of

    SCC involved, if applicable, leading to failure of the proposed implants material, Mg-

    0.8Ca.

  • References

    1. Niinomi, M., Recent metallic materials for biomedical applications. Metallurgical and Materials Transactions A, 2002. 33(3): p. 477-486.

    2. Antunes, R.A. and M.C.L. de Oliveira, Corrosion fatigue of biomedical metallic alloys: Mechanisms and mitigation. Acta Biomaterialia, 2012. 8(3): p. 937-962.

    3. Blackwood, D.J., 2.27 - Corrosion in Body Fluids, in Shreir's Corrosion, J.A.R. Editor-in-Chief: Tony, Editor. 2010, Elsevier: Oxford. p. 1308-1322.

    4. Staiger, M.P., et al., Magnesium and its alloys as orthopedic biomaterials: A review. Biomaterials, 2006. 27(9): p. 1728-1734.

    5. Salahshoor, M. and Y. Guo, Biodegradable Orthopedic Magnesium-Calcium (MgCa) Alloys, Processing, and Corrosion Performance. Materials, 2012. 5(1): p. 135-155.

    6. Nagels, J., M. Stokdijk, and P.M. Rozing, Stress shielding and bone resorption in shoulder arthroplasty. Journal of Shoulder and Elbow Surgery, 2003. 12(1): p. 35-39.

    7. Erdmann, N., et al., Evaluation of the soft tissue biocompatibility of MgCa0.8 and surgical steel 316L in vivo: a comparative study in rabbits. Biomed Eng Online, 2010. 9: p. 63.

    8. Singh Raman, R.K., N. Birbilis, and J. Efthimiadis, Corrosion of Mg alloy AZ91 the role of microstructure. Corrosion Engineering, Science and Technology, 2004. 39(4): p. 346-350.

    9. Drynda, A., et al., Development and biocompatibility of a novel corrodible fluoride-coated magnesium-calcium alloy with improved degradation kinetics and adequate mechanical properties for cardiovascular applications. Journal of Biomedical Materials Research Part A, 2010. 93A(2): p. 763-775.

    10. Witte, F., et al., In vivo corrosion of four magnesium alloys and the associated bone response. Biomaterials, 2005. 26(17): p. 3557-3563.

    11. Atrens, A., M. Liu, and N.I. Zainal Abidin, Corrosion mechanism applicable to biodegradable magnesium implants. Materials Science and Engineering: B, 2011. 176(20): p. 1609-1636.

    12. Gu, X.N., et al., Corrosion fatigue behaviors of two biomedical Mg alloys AZ91D and WE43 In simulated body fluid. Acta Biomaterialia, 2010. 6(12): p. 4605-4613.

    13. Kannan, M.B. and R.K.S. Raman, In vitro degradation and mechanical integrity of calcium-containing magnesium alloys in modified-simulated body fluid. Biomaterials, 2008. 29(15): p. 2306-2314.

    14. Li, Z., et al., The development of binary MgCa alloys for use as biodegradable materials within bone. Biomaterials, 2008. 29(10): p. 1329-1344.

    15. Rad, H.R.B., et al., Microstructure analysis and corrosion behavior of biodegradable MgCa implant alloys. Materials & Design, 2012. 33(0): p. 88-97.

    16. Song, G., Control of biodegradation of biocompatable magnesium alloys. Corrosion Science, 2007. 49(4): p. 1696-1701.

    17. Serre, C.M., et al., Influence of magnesium substitution on a collagenapatite biomaterial on the production of a calcifying matrix by human osteoblasts. Journal of Biomedical Materials Research, 1998. 42(4): p. 626-633.

    18. Choudhary, L. and R.K. Singh Raman, Magnesium alloys as body implants: Fracture mechanism under dynamic and static loadings in a physiological environment. Acta Biomaterialia, 2012. 8(2): p. 916-923.

    19. Kirkland, N.T., N. Birbilis, and M.P. Staiger, Assessing the corrosion of biodegradable magnesium implants: A critical review of current methodologies and their limitations. Acta Biomaterialia, 2012. 8(3): p. 925-936.

  • 20. Witte, F., et al., In vitro and in vivo corrosion measurements of magnesium alloys. Biomaterials, 2006. 27(7): p. 1013-1018.

    21. ASTM, Standard Practice for the Preparation of Substitute Ocean Water. 2008. 22. Bombara, G. and M. Cavallini, Stress corrosion cracking of bone implants. Corrosion

    Science, 1977. 17(2): p. 77-85. 23. Sivakumar, M. and S. Rajeswari, Investigation of failures in stainless steel orthopaedic

    implant devices: pit-induced stress corrosion cracking. Journal of Materials Science Letters, 1992. 11(15): p. 1039-1042.

    24. Bandyopadhyay, S. and P. Brockhurst, Fractographic evidence of mixed mode stress corrosion cracking in stainless steel surgical implants. Journal of Materials Science, 1979. 14(12): p. 3002-3003.

    25. Bobby Kannan, M. and R.K. Singh Raman, Evaluating the stress corrosion cracking susceptibility of MgAlZn alloy in modified-simulated body fluid for orthopaedic implant application. Scripta Materialia, 2008. 59(2): p. 175-178.

    26. Teoh, S.H., Fatigue of biomaterials: a review. International Journal of Fatigue, 2000. 22(10): p. 825-837.

    27. Fleck, C. and D. Eifler, Corrosion, fatigue and corrosion fatigue behaviour of metal implant materials, especially titanium alloys. International Journal of Fatigue, 2010. 32(6): p. 929-935.

    28. Choudhary, L., et al., Investigations into stress corrosion cracking behaviour of AZ91D magnesium alloy in physiological environment. Procedia Engineering, 2011. 10(0): p. 518-523.