Improved Speed and Confidence in Toxicology …...- Scan = best confirmation of targets and...
Transcript of Improved Speed and Confidence in Toxicology …...- Scan = best confirmation of targets and...
Bruce QuimbyApplication ScientistApplication ScientistAgilent TechnologiesOctober 28, 2009
Page 1 For Forensic Use.
Tools For Improving GC/MS Analysis
• Retention Time Locking (RTL)
• Method Translation• Method Translation
• Synchronous SIM/Scan
• Column effluent split to an NPD
• Column backflushing
• Deconvolution Reporting Software (DRS)
Page 2 For Forensic Use.
Retention Time LockingGet precisely the same retention times:• After cutting or replacing column• On all instruments in a lab and other labs (method sharing)
Get precisely the same retention times:
( g)• Greatly reduces method maintenance for updating:
– peak recognition (RT) windows– SIM group start times– integration events
• Reduces data review timeReduces data review time• Makes management of large screening databases much easier• Simple procedure restores all RTs to original calibrated values• Process moves RTs back to calibrated values, does not change calibrated
value to current (different) RT• Look for new targets in archived data, since RTs always the same
Page 3
g , y
For Forensic Use.
How Does RTL Work?
For each method:• A set of five runs of retention time vs inlet pressure for a• A set of five runs of retention time vs inlet pressure for a
single locking compound is collected (only once per method)• The calibration is stored with the method• When locking a new instrument, the locking compound is run and its RT is
measured• RTL software calculates new inlet pressure that makes the RT of the
locking compound and all analytes precisely match that of the originalmethod
• U
•
pdating RTs of individual cal compounds and timed events unnecessaryp g p y
Page 4
If a table of hundreds of RTs is collected under RTL conditions, anyoneanywhere can lock to the same conditions and get the same RTs. This isbasis of Agilent screening databases for Forensic Tox compounds, pesticides, etc.
For Forensic Use.
RTL Example
RTL keeps the retention times of all analytes typically within 0.030 min absolute. Locking is a simple procedure using 1 compound. After pressure adjustment, all compounds fall within their RT recognition window.
Initial run 4.72 psi4.296 min.
Original locked method
Column trimmed lockingTrim 1 meter 4.72 psi4.064 min. Column trimmed, locking
compound run at original pressure (RT is too short)
Relock 4.42 psi4.297 min.
RTL calculates new inlet pressure to restore RTs
Page 5 For Forensic Use.
Agilent’s Method Translation SoftwareA technique invented by Agilent that calculates the inletA technique invented by Agilent that calculates the inletpressure and oven temperature program for changing:
– Analysis speed (example: 4 x faster)Analysis speed (example: 4 x faster)– Column dimensions (with same phase ratio)– Carrier gas type (example: Changing from He to H2)– Column outlet pressure (MSD vs NPD vs CFT Splitter)
Determines the required conditions with a calculator instead of trial and error GC runs in the laboratorytrial and error GC runs in the laboratory
Key benefits:Relative elution order is maintained• Relative elution order is maintained.
• Retention times with new method are accurately predicted• Substantial saving in method development time, especially for screening
Page 6
methods with large numbers of analytes.
For Forensic Use.
Speed Scaling Constant Pressure RTL MethodsAllows method to be adjusted to optimal speed on a
Use MTL software to calculate changes in pressure and
Allows method to be adjusted to optimal speed on aspecific system.
Use MTL software to calculate changes in pressure andtemperature ramps corresponding to desired speed gain
Generate new RTL calibration for the new methodGenerate new RTL calibration for the new method
Peaks will elute in same relative order as before
L k t i t t d tLock system using target compound to:(original target time) (speed gain)
N ll t ti ti i th d ill b l tNow all retention times in new method will be equal to:original RT speed gain
Page 7 For Forensic Use.
1-5 ng Test Mix: DB-5MS, 1x MSD Direct10C/min and 1.8 mL/min initial flow1 Amphetamine2 Phentermine3 Methamphetamine4 Nicotine Tetrahydrocannabinol
15
18
Lorazepam16 Diazepam17 Hydrocodone
10C/min and 1.8 mL/min initial flow
4 Nicotine5 Methylenedioxyamphetamine(MDA)6 Methylenedioxymethamphetamine(MDMA)7 Methylenedioxyethylamphetamine8 Meperidine9 Phencyclidine
Tetrahydrocannabinol1819 Oxycodone20 Temazepam
22 Diacetylmorphine23 Nitrazepam17
21 Flunitrazepam
y10 Methadone11 Cocaine12 SKF-525a (RTL Compound)13 Oxazepam14 Codeine
p24 Clonazepam25 Alprazolam26 Lysergide (LSD)27 Strychnine28 Trazodone
9
1012
17
1816
23
4
6
7
811
1415
1921
22
25 27
1
5
6
13
20 23 24 2628
Page 8
TIC: 25mix_340C_2.D\data.ms
4 8 12 16 20 24 28
For Forensic Use.
DB-5MS: Comparison of Different Speeds on MSD4X method RTs are precisely the same as 1X divided by 4
1 x30 m, 10C/min and 1.7 mL/min: 120V oven and Standard Turbo
2 x15 m, 20C/min and 1.2 mL/min: 120V oven and Standard Turbo
4 8 12 16 20 24 28
2 x
15 m, 30C/min and 2.7 mL/min: 240V oven and Perf Turbo2 4 6 8 10 12 14
3 x 15 m, 30C/min and 2.7 mL/min: 240V oven and Perf Turbo
1 2 3 4 5 6 7 8 9
4 x15 m, 40C/min and 5.9 mL/min: 240V oven and Perf Turbo
1 2 3 4 5 6 7
Page 9
1 2 3 4 5 6 7
For Forensic Use.
Why Use Agilent’s Synchronous SIM/Scan?
Get SIM and Scan Data in a single run which saves time
SIM maximum sensitivity for target compounds- SIM = maximum sensitivity for target compounds
- Scan = best confirmation of targets and identification of unknownsunknowns
- For most methods, changing to SIM/Scan results in little, if any, degadation in signal-to-noise compared to SIMany, degadation in signal to noise compared to SIMonly or Scan only modes
Page 10 For Forensic Use.
How Does Synchronous SIM/Scan Work?
SIM data pointsChromatographic Peak SIM and Scan data
points are alternatelyScan data pointsSIM data points points are alternately
collected
At end of run, two separate data signals are constructed
Each signal can be d lik SIM
Scan timeSIM
Scan
SIM
Scan
SIM
Scan
SIM
Scan
SIM
Scan
processed like SIM or Scan collected separately
Page 11 For Forensic Use.
Why Collect NPD and MSD Simultaneously?
NPD gives sensitive nitrogen selective detection:
Identity confirmation (if it is a nitrogen drug it should- Identity confirmation (if it is a nitrogen drug, it shouldrespond on NPD)
- Faster data review alternate to MSD for quantitation- Faster data review, alternate to MSD for quantitation
- Highlights nitrogen drugs that are not in target list
S litt d l ll b kfl hi d h i l- Splitter used also allows backflushing and changing columnswithout venting MSD
Page 12 For Forensic Use.
Example of Synchronous SIM/Scan/NPDFentan l
Ion 245 Scan S/Npk-pk = 8
Fentanyl Real whole blood tox sample:
Screen for 278 drugs in Scan mode
Ion 146 Scan
Ion 189 Scan
mode.
Screen for low level targets in SIM mode
S/Npk-pk = 77Ion 245 SIMFentanyl found by SIM at trace level because of 10 fold better S/N
Ion 146 SIM
Ion 189 SIM
NPD trace shows nitrogen peak at same RT as fentanyl in MS dataMS data
NPDS/Npk-pk 42
Page 13
5.52 5.60 5.68 5.76
For Forensic Use.
How Does the Splitter Work?Liquidq
Injector
XEPC NPDVent
New splitter technology
7890A GC
Column 5975CMSD
7890A GC
Column effluent split between MSD and NPD
Reliable, inert, easy to use two way splitter with solvent venting
Auxillary EPC control allows no vent column changing and post run backflushing
Page 14
backflushing
For Forensic Use.
Why Use Agilent’s Column Backflushing?
Backflushing removes heavy material from column after elution time of last analyte:y
- Prevents heavy matrix compounds from eluting in later runs
- Reduces column trimming and detector maintenance- Reduces column trimming and detector maintenance
- Significantly reduces postrun bakeout times (example: 2 min backflush vs 10 min bakeout)backflush vs 10 min bakeout)
Heavies past end of Whole blood tox sampleacquisition
TIC
p
Page 15
1 2 3 4 5 6 7 8
For Forensic Use.
How Does Backflushing Work?
Auxillary EPC provides constant pressure makeup
t t l d i
During GC Run
A EPC
SplitVent
gas to post column device.During run, makeup flow is low (~ 1 mL/min)S/S Inlet
Splitter or
4 psi
MSD
Aux EPCTrap
Flow
Splitter orQuickSwapColumn
25 psi
After GC Run A EPC i d
45 psiAux EPC
SplitVent Trap
After GC Run Aux EPC is programmedto high pressure and inlet to low pressure. Reversed flow quickly sweeps
S/S Inlet
Column1 psi
45 psi
MSDFlow
Splitter or QuickSwap
flow quickly sweepsheavies out to split vent trap
Page 16
Column p
For Forensic Use.
Why Use Deconvolution Reporting Software?
Deconvolution of mass spectra removes/reduces interferences from chromatographically overlapped peaks:g p y pp p• Better identification and confirmation of analytes in high matrix samples • Reduces data review time, especially for large screening methods• Reduces false positives and false negatives in dirty samples• Identification based on matching entire spectrum cleaned of interferences
against library. Much more reliable than target/qualifier ratio method.against library. Much more reliable than target/qualifier ratio method.
Page 17 For Forensic Use.
How Does Deconvolution Work?Ions with the same abundance vs time profile are groupedIons with the same abundance vs time profile are groupedtogether to create spectra “cleaned” of interferences from overlapped peaks
Three Overlapped Chromatographic Peaks
DeconvolutionInterference 1
TIC
Interference 2
m/zInterference 1Interference 2 Time
Time
Interference 2Target
Time Time
Target
Page 18
m/z
For Forensic Use.
AMDIS Looks at Apex Retention Time and Peak Shape
28p
50170
280
3118516
075160
Extracted Ion
160 – wrong shape
Chromatograms(EIC)
Ions whose EICs have the:
50170
g p1. same apex RT
2. same shapeThese have same shape and same apex RT280
75 – wrong rt185 – wrong rtare grouped to form
cleaned (extracted) spectra
apex RT
Page 19
g310 – wrong rt
spect a
For Forensic Use.
Ions With Same RT and Shape Are Grouped and Called Components (Spectra)p ( p )
50170
280
This spectrum isExtractedspectrum This spectrum is
compared to msl for target hits.
spectrumcleaned of interferences
ComponentExtracted Ion
ChromatogramsPeak heights of EICs give
l ti
50170
Chromatograms(EIC)relative
abundancevalues for
t 170280
spectra
Page 20 For Forensic Use.
Example:Spectral Confirmation of Carisoprodol in Blood Extract With High Level of Fatty AcidsBlood Extract With High Level of Fatty Acids
Apex spectrum 180000
200000
Abundance
Average of 3.564 to 3.601 min.: 120C_1MIN_GB20_VIAL2.D\DATA.MS54.9
#1 NIST Hit is Oleic acid.
Carisoprodol is 80000
100000
120000
140000
160000
96.9 Spectrum at Carisoprodol RT, not deconvoluted
Carisoprodol isnot in top 100 hits
40 60 80 100 120 140 160 180 200 220 240 260 280 300 3200
20000
40000
60000
m/z-->
158.0128.9
185.0
284.1213.0 241.1
75.8 261.9 340.0311.9
58
DeconvolutedSpectrum
0
50
100
44
58
71 86
97104
114132
149
158
184205 223
245257 288 327
Deconvoluted Spectrum at Carisoprodol RT Match = 80
Carisoprodol is #1 Hit in NIST05a
0
50
100
43
55
83
97
114127 144
158
184202 217
245260
Library Reference Spectrum of Carisoprodol
Page 21
40 70 100 130 160 190 220 250 280 310 340
55
For Forensic Use.
Steps in DRS Process
Software goes through entire GC/MS scan file and extracts all components (spectra).p ( p )
Each component is automatically searched against the target spectral library.
If a component has both:• a spectral match factor > than user defined minimump• a retention time matching the table RT within a user defined limit• It is reported as present.
Data reviewer inspects the results and confirms compounds present. Those compounds are then quanted and report generated.
Page 22
g
For Forensic Use.
Example: Design 725 Compound Fast Tox Screening SystemScreening SystemRetention Time Locking (RTL)-
use Agilent’s New G1674AA Forensic Toxicology RTL Library– use Agilent s New G1674AA Forensic Toxicology RTL Library
Method Translation– Run 4x faster version (40C/min) of method– Run 4x faster version (40C/min) of method
Synchronous SIM/Scan and Split to NPD– Collect 3 signals Scan SIM and NPD in one short runCollect 3 signals Scan,SIM, and NPD in one short run
Column backflushing– Saves bakeout time, gives cleaner baselines, less maintenanceSaves bakeout time, gives cleaner baselines, less maintenance
Deconvolution Reporting Software (DRS)– Biggest time saver. Data reviewed in ~10 min instead of ~1
Page 23
gghour/sample
For Forensic Use.
Fast Tox Screening System Post column device provides:
LiquidI j t
pA) 1.4:1 MSD:NPD effluent splitB) Solvent ventingC) BackflushingD) No vent column changingInjector
XEPC NPDSolvent
D) No vent column changing
XEPC3.8 psig
NPD
S/S Inlet
14 9 i
SolventVent
Column
14.9 psig
(nom)
7890A GC
Column
10 m X 0 25 mm id X 0 25 um DB 5ms
5975CMSD
Page 24
10 m X 0.25 mm id X 0.25 um DB-5ms
For Forensic Use.
Method Parameters for DB-5ms, 4X, NPD/MSD Split on 7890on 7890
Ramp 'C/min 'C Hold minInitial 100 0.25
MSD Agilent 5975CSolvent Delay 0.5 min
Ramp 1 40 325 1.25Runtime 7.13 min
Inlet Split/Splitless
Scan Range 40 to 570Threshold 0Sampling 1Quad Temp 180 'C
Postrun 325 'C/min for 0.5 min
Acquisition Mode SIM/Scan
p pTemp 280 'CMode Splitless, Constant PressurePressure 14.90 (adj to lock)
Quad Temp 180 CSource Temp 300 'C
Purge Flow 50 mL/min
Transfer Line 300 'C
P ti 0 4 i S litt 2 litt / l t t
Tune Gain Normalized 1X
Purge time 0.4 min SplitterPressure 3.8 psig
Column DB-5ms part # (custom) MSD Restictor 0.694 m x 0.15 mm id
Split ratio 1.4:1 MSD:NPDInitial Flow 2 52 mL/min
2-way splitter w/ solvent vent
NPD Restictor 0.361 m x 0.15 mm id10m x 0.25 mm id x 0.25 um film
Injection volume 1 uL
Split ratio 1.4:1 MSD:NPDInitial Flow 2.52 mL/minVent time range 0-0.75 minOutlet SplitterBackflush Time 0.5 minOutlet Pressure 3.8 psig
Backflush Temp 325Backflush Press 76RT locked to Proadifen at 4.285 min
Page 25 For Forensic Use.
SIM Groups Acquired With SIM/ScanSIM Group Peak Name RT min Tgt Q1 Q2p g
1 Amphetamine 0.600 44 91 652 Methamphetamine 0.700 58 91 653 Methylenedioxyamphetamine(MDA) 1.319 136 135 514 Methylenedioxymethamphetamine(MDMA) 1.431 58 135 774 Ecgonine Methyl Ester 1 481 94 82 964 Ecgonine Methyl Ester 1.481 94 82 964 Ethylecgonine 1.482 94 82 965 Meperidine 1.884 246 218 2476 Ketamine 2.092 180 182 2096 Phencyclidine 2.166 243 242 2006 Tramadol 2.259 58 263 596 Tramadol 2.259 58 263 597 Methadone 2.578 72 57 1657 Dextromethorphan 2.597 271 212 2708 Cocaine 2.695 182 82 948 Cocaethylene 2.783 196 82 949 Diazepam 3.065 258 286 2579 Tetrahydrocannabinol 3.111 299 300 2319 6-Acetyl-Morphine 3.182 268 327 32810 Oxycodone 3.201 315 230 11510 Temazepam 3.281 271 273 27210 Diacetylmorphine 3.328 310 268 32710 Fentan l 3 451 245 146 18910 Fentanyl 3.451 245 146 18911 Zolpidem 3.555 235 236 21911 Clonazepam-M (amino-) 3.622 285 258 28612 Alprazolam 3.753 308 279 28012 Zaleplon 3.797 305 263 24813 Zopiclone 3 937 112 99 139
Page 26
13 Zopiclone 3.937 112 99 13913 Lysergide (LSD) 4.000 323 324 222
For Forensic Use.
Screening A Real Blood ExtractBlood extracts are very complex. There are more than 400 detectable peaks in y p pthis sample. The problem is to find the compounds of toxicological interest.
Fast method: 9.75 min injection to injection (including backflush)
Scan TIC
SIM TIC
NPD
Page 27
1 2 3 4 5 6 7
For Forensic Use.
Target Compounds Identified in Sample By DRS1 Nicotine2 Nicotinamide
7 Carisoprodol8 Methadone
7 9 11
2 Nicotinamide3 Carisoprodol artifact4 Cotinine5 Meprobamate6 Caffeine
8 Methadone9 Cyheptamide (ISTD)
10 Oxycodone11 Cholesterol
Scan TIC3 8
7 11
3
215
4
8
610
3 4 7
10 NPD
21
56 8 9
11
Page 28
1 2 3 4 5 6 7
For Forensic Use.
Deconvolution of Spectra Eliminates Review Bottleneck For Large Screening MethodsBottleneck For Large Screening MethodsConventional quant approach (~ 1hour / sample review time):
Must examine every peak with response at target ion (367 here)– Must examine every peak with response at target ion (367 here)– Takes ~ 1 hour to review, delete false positives, quant real ones
DRS approach (~ 10 min / sample review time)– Gives list of compounds present in 1-2 minutesGives list of compounds present in 1 2 minutes– List has many fewer compounds to review. (12 instead of 367)– Very few false positives and false negatives– In the current example, all compounds identifiable by QEDIT or Screener
approach were found immediately with deconvolution– Can find compounds not identifiable by conventional approach when
Page 29
p y ppsevere overlap with matrix exists
For Forensic Use.
DRS Report From Scan Data of Sample
Page 30For Forensic Use.
Caffeine: Is It There?
All 3 qualifiers have problems. It would be good to have more info lik t l t h >
Interference
Caffeine?
TIC Scan
like a spectral match >60 to confirm ID, since qualifiers are questionable.q
194 Scan
Q1 and Q2 have interference overlap
67 S
109 Scan
82 Scan
Q3 has very low S/N
NPD confirms nitrogen at caffeine RT
67 Scan
NPD
Q3 has very low S/N
Page 31
3.00 3.05 3.10
For Forensic Use.
Caffeine Spectrum Has Interference Problem With Matrix CompoundMatrix Compound
Match = 51Caffeine Apex Spectrum194 Caffeine Apex Spectrum(no deconvolution)
50
100
4355
67 109
194
Target Library
0
42
71 82 91
94 122
124
136
137 154 165
165
207 221 235 250260 273 284
Target LibraryReference Spectrum of Caffeine
50
100
5567
82
109
194
38 58 78 98 118 138 158 178 198 218 238 258 278
194
Page 32 For Forensic Use.
Caffeine Component Searched Against Target LibraryLibrary
Deconvolution improves spectral match quality by removing interfering compound spectrum
Match = 70
Component Found at RT of Caffeine
100194
0
50
42
55 6772 82
94
97
109
122 136 165
Target Library Reference Spectrum of Caffeine
50
100
42
5567
82
109
194
38 58 78 98 118 138 158 178 198 218 238 258 278
194
Page 33 For Forensic Use.
Methadone: Is It There? Scan and NPD Data
SIM i hi h S/N 72 Scan
57 Scan
SIM gives high S/Nconfirmation of ratios, but 57 still doubtful
57 Scan165 Scan
NPD confirms N
NPDcontaining peak at Methadone RT
Page 34
3.8 3.9 4.0
For Forensic Use.
Methadone Spectrum Has Interference Problem With Octadecanoic Acid in MatrixWith Octadecanoic Acid in Matrix
Match = 42100
72
Methadone Apex Spectrum
(no deconvolution)50
100
43 55
Target Library Reference
0
50
42
43 55
57 77
83
91
97 111
115 128
129 143151
165
165 178
185
191199 223
236
241 253 264 284
294
Target Library ReferenceSpectrum of Methadone
38 58 78 98 118 138 158 178 198 218 238 258 278 298 318
50
100 72
38 58 78 98 118 138 158 178 198 218 238 258 278 298 318
Page 35 For Forensic Use.
Methadone Deconvoluted Spectrum Searched Against Target LibraryAgainst Target Library
Deconvolution significantly improves spectral match quality by removing interfering octadecanoic acid spectrum
Component found at RT of Methadone
Match = 80
50
100 72
Methadone
042
42
56
57 77
85
91
115
115 128130 143
151165
165
178
178 189195 223
236294
294
Target Library Reference Spectrum of Methadone
50
100 Spectrum of Methadone38 58 78 98 118 138 158 178 198 218 238 258 278 298 318
100 72
Page 36 For Forensic Use.
Alprazolam: Is It There?Deconvolution and conventional quant are both limited byDeconvolution and conventional quant are both limited bysignal-to-noise ratio of scan data. As S/N drops, ions disappear from deconvoluted spectrum
Deconvoluted Spectrump
Forward Match 57.5
Reverse 57.5
Second Hit in NIST 05a50
100 204
273
279
308
Second Hit in NIST 05a
0
50
3951
63 89102
116137
151 163177
190 219245
253293
38 58 78 98 118 138 158 178 198 218 238 258 278 298 318 338
50
100
77
204
273
279308
Page 37 For Forensic Use.
Alprazolam at Limit of Detectability in Scan ModeDeconvolution and conventional quant are both limited by
279 Scan
signal-to-noise ratio of scan data
204 Scan
308 Scan
9 ScaBoth deconvolution and standard quant are at the detection li it h
203 Scan
308 Scanlimit here
279 SIM
308 SIMSIM and NPD data aid in confirming very
280 SIM
NPD
in confirming verylow level of Alprazolam
Page 38
5.55 5.60 5.70 5.80NPD
For Forensic Use.
Time Savings
Conventional Method:Scan run: 35 min– Scan run: 35 min
– SIM run: 35 min– NPD run: 35 min– Data Review: ~1 hour– Total: 165 min/sample
F t T M th dFast Tox Method:– Scan, SIM, and NPD: 9.75 min
Data Review: ~10 min– Data Review: ~10 min– Total: 19.75 min/sample
Page 39For Forensic Use.
Summary
Retention Time Locking (RTL)-the same RTs from now onthe same RTs from now on
Method Translationspeed scale to new column sizes carrier gases analysis speedsspeed scale to new column sizes, carrier gases, analysis speeds
Synchronous SIM/Scan and Split to NPDCollect 3 signals Scan SIM and NPD in one runCollect 3 signals Scan,SIM, and NPD in one run
Column backflushingSaves bakeout time, gives cleaner baselines, less maintenanceSaves bakeout time, gives cleaner baselines, less maintenance
Deconvolution Reporting Software (DRS)Data review time reduced while quality is improved
Page 40
q y p
For Forensic Use.
Get started quicklyGet started quickly
Application kitColumns, consumables & checkout mixApplication Note and Quick start guide
Application kit
Application Note and Quick start guideDVD with method parameters729 compound DRS database for fast screening729 compound DRS database for fast screeningVideo tutorials with step-by-step instructionsPre- configured and pretested for your applicationOn-site application checkout
Page 41
Page 41
FY09 Solution PlanAgilent Restricted
Feb. 2009
For Forensic Use.
Thanks For Attendingg
Wow! Did you see that?Page 42
Wow! Did you see that?