Illustrative Flow Visualization from 3D Doppler Ultrasound

10
Illustrative Flow Visualization from 3D Doppler Ultrasound Åsmund Birkeland 1 , Linn Heljesen 1,2 , Kim Nylund 2 , Dag Magne Ulvang 3 , Odd Helge Gilja 1,2 and Ivan Viola 1,3 University of Bergen (1) Haukeland University Hospital (2) Christian Michelsen Research (3)

description

Illustrative Flow Visualization from 3D Doppler Ultrasound. Åsmund Birkeland 1 , Linn Heljesen 1,2 , Kim Nylund 2 , Dag Magne Ulvang 3 , Odd Helge Gilja 1,2 and Ivan Viola 1,3. University of Bergen (1) Haukeland University Hospital (2) Christian Michelsen Research (3) . Motivation. - PowerPoint PPT Presentation

Transcript of Illustrative Flow Visualization from 3D Doppler Ultrasound

Page 1: Illustrative Flow Visualization from 3D Doppler Ultrasound

Illustrative Flow Visualization from 3D Doppler Ultrasound

Åsmund Birkeland1, Linn Heljesen1,2, Kim Nylund2, Dag Magne Ulvang3, Odd Helge Gilja1,2 and Ivan Viola1,3

University of Bergen (1) Haukeland University Hospital (2) Christian Michelsen Research (3)

Page 2: Illustrative Flow Visualization from 3D Doppler Ultrasound

MotivationBlood flow imagingWhat we want:

3D Blood flow measurementsLiveInexpensiveNon-invasive

What we have:4D MRI – expensive, not liveB-flow – only in 2DCEUS – invasive Doppler – ...

2/10

Page 3: Illustrative Flow Visualization from 3D Doppler Ultrasound

MotivationPros:

InexpensiveLiveNon-invasive

Cons:NoisySingle component of the blood flowDifficult to interpret for laymen

3/10

Page 4: Illustrative Flow Visualization from 3D Doppler Ultrasound

Methodology

4/10

Page 5: Illustrative Flow Visualization from 3D Doppler Ultrasound

Geometry ExtractionSegmentation

Extract vessel segments during an ultrasound scanImage based registration for fusing segments together

Vessel geometryExtract center-line of segmented vesselsGenerate a vessel-direction map based on nearest center-line point

5/10

Page 6: Illustrative Flow Visualization from 3D Doppler Ultrasound

3D Flow ReconstructionProbe position tracked using image based registrationReconstruction:Flux through a cross-section: Error dependent on

6/10

Page 7: Illustrative Flow Visualization from 3D Doppler Ultrasound

Visual Mapping

7/10

Animated particles depicting flow direction and velocity magnitudeVelocity legends for identifying velocity magnitudeColor according to flux through cross section

Page 8: Illustrative Flow Visualization from 3D Doppler Ultrasound

ResultsVerification based on 4D MRI data Accuracy: 69% for angles below 55 degrees

8/10

Page 9: Illustrative Flow Visualization from 3D Doppler Ultrasound

Future DirectionsIn vivo evaluationImprove visual representationMore advanced flow model

9/10

Page 10: Illustrative Flow Visualization from 3D Doppler Ultrasound

Thank you for listening!Questions?

10/10