If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

12
RapidVet ® -H Blood Group Determination System for Identifying DEA 1.1 Positive and DEA 1.1 Negative Dogs dm s laboratories,inc.

description

RapidVet ® -H Blood Group Determination System for Identifying DEA 1.1 Positive and DEA 1.1 Negative Dogs. - PowerPoint PPT Presentation

Transcript of If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Page 1: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

RapidVet®-HBlood Group Determination Systemfor Identifying DEA 1.1 Positive and

DEA 1.1 Negative Dogs

dmslaboratories,inc.

Page 2: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Kit contains:auto-agglutination saline screen cards, patient test cards, pipettes, stirrers, diluent, instructions, interpretation guideand report card.

Shelf-life: 24 months at room temperature

dmslaboratories,inc.

Page 3: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

If agglutination is observed, STOP the test and perform

cell washing procedures before proceeding.

If auto-agglutination does not occur, proceed to the next step.

dmslaboratories,inc.

Prior to typing, check for auto-agglutination using the card provided in the kit. This card is used for three patients and, after use, should be returned to the kit box.

Page 4: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Remove the Test Card from its plastic sleeve. Write the name/number of the dog and the

testing date on the card.

dmslaboratories,inc.

Page 5: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Dispense 1 drop of diluent from the dropping bottle into each well.

dmslaboratories,inc.

Page 6: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Aspirate a small amount of patient sample into a pipette and release 1 drop (50µl) into each well, taking care not to cross-contaminate the wells.

dmslaboratories,inc.

Page 7: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Using a NEW stirrer and pressing downward with the flat portion of the stirrer, spread and mix the materials

within the entirety of the first well for about 10 seconds.

dmslaboratories,inc.

Page 8: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Using a NEW stirrer, spread and mix the materials within the entirety of the second well

for about 10 seconds.

dmslaboratories,inc.

Page 9: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Using a NEW stirrer, spread and mix the materials within the entirety of the third well

for about 10 seconds.dmslaboratories,inc.

Page 10: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Rock the card, using a transaxial motion, for 1 minute.

Be sure materials are mixing and

“rotating” within each individual well.

Be careful not to cross-contaminate.

dmslaboratories,inc.

Page 11: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

Read the results, noting whether agglutination has or has not occurred in the “Patient Test” well.

dmslaboratories,inc.DEA 1.1+ DEA 1.1

Page 12: If agglutination is observed, STOP the test and perform cell washing procedures before proceeding.

After the card has been read, take a digital photograph for a permanent record, or allow to dry and replace in its

plastic sleeve as a permanent record.