Surgical technique and Ordering information This surgical technique applies only for the US. IDSF

Transcript of IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2)...

Page 1: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

Surgical technique and Ordering informationThis surgical technique applies only for the US.

IDSF

Page 2: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)2

Page 3: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

3

OsteoBridge™ IDSF System

Table of ContentsDescription ���������������������������������������������������������������������������������������������������������������������������������������������������������������� 4

Indications ���������������������������������������������������������������������������������������������������������������������������������������������������������������� 5

Contraindications �������������������������������������������������������������������������������������������������������������������������������������������������� 5

System Compatibility �������������������������������������������������������������������������������������������������������������������������������������������� 6

Surgical Technique ������������������������������������������������������������������������������������������������������������������������������������������������ 9

Pre-operative planning ������������������������������������������������������������������������������������������������������������������������������������������� 9

Preparing the defect to be treated ���������������������������������������������������������������������������������������������������������������������11

Non-cemented implantation ������������������������������������������������������������������������������������������������������������������������������12

Cemented implantation ��������������������������������������������������������������������������������������������������������������������������������������12

Implant selection ���������������������������������������������������������������������������������������������������������������������������������������������������13

Non-cemented insertion and locking of intramedullary nails ����������������������������������������������������������������������14

Cemented insertion of intramedullary nails �����������������������������������������������������������������������������������������������������18

Mounting the spacer ��������������������������������������������������������������������������������������������������������������������������������������������19

Wound closure and post-operative care ����������������������������������������������������������������������������������������������������������24

Additional surgical technique �����������������������������������������������������������������������������������������������������������������������������25

OsteoBridge™ IDSF revision �������������������������������������������������������������������������������������������������������������������������������26

Ordering Information ����������������������������������������������������������������������������������������������������������������������������������������28

OsteoBridge™ IDSF Implants ������������������������������������������������������������������������������������������������������������������������������28

OsteoBridge™ IDSF Trial nails ���������������������������������������������������������������������������������������������������������������������������29

OsteoBridge™ IDSF Instrument Tray �����������������������������������������������������������������������������������������������������������������30

Page 4: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)4

Description

Description

Clamping screws

SpacerUpper half-shell

SpacerLower half-shell

Interlocking screws

All components are made of: TiAl6V4Titanium

Alloy

The OsteoBridge™ IDSF – Intramedullary Diaphyseal Segmental Defect Fixation system is intended to be used in the management of segmental diaphyseal bone loss secondary to radical bone loss and/or resection due to tumors in either humerus, tibia or femur in oncology patients� The device is divided into three categories/product size ranges that are suitable for each boney application (see tables on pages 6-8)�

The system is designed for non-cemented and cemented fixation in bone� The modular system includes a spacer made of two cylindrical half-shells with eight screws that clamp over the end of two intramedullary nails� Non-cemented implants are additionally fixated with two interlocking screws per nail� Spacers and intramedullary nails of different sizes can be selected to accommodate varying intramedullary canal diameters and different bone defect sizes� If the defect size is larger than 70 mm, two spacers may be connected with a spacer connector� All components of the OsteoBridge™ IDSF system are manufactured from Ti6Al4V ELI titanium alloy according to ASTM F136 and ISO 5832-3�

Page 5: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

5

OsteoBridge™ IDSF SystemDescription

X Indications

• Long-term stabilization of major bone defects resulting from

o Radical bone loss due to tumors and/or metastases

o Bone resections following tumors and/or metastases

• For use only in the diaphyseal region of humerus, tibia and femur

Nail

X Contraindications • Ongoing local or systemic infections

• Severe muscular, nervous, or vascular conditions that endanger the affected extremities

• Insufficient bone structure, preventing good fixation of the implant

• Any accompanying conditions which could potentially endanger the functionality or success of the implant

• Patients with mental or neurological illnesses, or patients who are not capable of following instructions with

regards to necessary post-operative treatment

Page 6: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)6

System Compatibility

CAUTION: The 20 mm, 25 mm and 34 mm diameter spacers have different inner diameters�

The appropriate nails have different clamping diameters and therefore cannot combined with each other� Each bone has its own nail portfolio�

System Compatibility

Humerus:

Connection DIA. 10

DIA. 7 DIA. 8 DIA. 9 DIA. 10

60 GB80706S GB80806S GB80906S GB81006S

70 GB80707S GB80807S GB80907S GB81007S

90 GB80709S GB80809S GB80909S GB81009S

110 - - - GB81011S

130 - - - GB81013S

Length

Diameter

DIA

� 10D

IA� 10

DIA� 20

20 mm diameter spacerIncluding clamping screws / sterile (Inner diameter 10 mm)

Spacer

Length Ref.

40 mm GB60001S

50 mm GB60002S

60 mm GB60003S

70 mm GB60004S

DIA� 20

DIA� 10

Page 7: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

7

OsteoBridge™ IDSF SystemSystem Compatibility

Tibia:

Connecion DIA. 14

DIA. 9 DIA. 10 DIA. 12 DIA. 14

60 GB40906S GB41006S - -

70 GB40907S GB41007S GB41207S -

90 GB40909S GB41009S GB41209S GB41409S

110 - GB41011S GB41211S GB41411S

130 - GB41013S GB41213S GB41413S

150 - - GB41215S GB41415S

200 - - - GB41420S

Length

Diameter

DIA

� 14D

IA� 14

DIA� 25

Spacer

Length Ref.

40 mm GB60011S

50 mm GB60012S

60 mm GB60013S

70 mm GB60014S

25 mm diameter spacerIncluding clamping screws / sterile (Inner diameter 14 mm)

DIA� 14

DIA� 25

Page 8: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)8

System Compatibility

34 mm diameter spacerIncluding clamping screws / sterile (inner diameter 16 mm)

Spacer

Length Ref.

40 mm GB60021S

50 mm GB60022S

60 mm GB60023S

70 mm GB60024S

Femur:

DIA

� 16

DIA� 16

DIA� 34

DIA� 16

DIA� 34

Nail with connection DIA. 16

DIA. 10 DIA. 12 DIA. 14 DIA. 16 DIA. 18 DIA. 20

60 GB70001S - - - - -

70 GB70002S GB70006S - - - -

90 GB70003S GB70007S GB70011S - GB70020S GB70022S

110 GB70004S GB70008S GB70012S GB70016S GB70021S GB70023S

130 GB70005S GB70009S GB70013S GB70017S - -

150 - GB70010S GB70014S GB70018S - -

200 - - GB70015S GB70019S - -

Length

Diameter

Page 9: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

9

OsteoBridge™ IDSF SystemSurgical Technique

Surgical Technique

Please note: Use only the OsteoBridge™ instruments included in delivery when implanting OsteoBridge™ IDSF components.

OsteoBridge™ IDSF is a disposable product and must not be re-used. Do not use parts that are damaged, have damaged packaging, or whose intended use is unclear.

X Pre-operative planning

The surgical technique shown here serves as an example to help illustrate the basic procedure for implanting the OsteoBridge™ system. Merete GmbH, manufacturer of this medical product, does not stipulate that this or any other treatment method is to be used for any specific patient. The responsibility for selecting a suitable method of treating a patient lies with that patient’s operating physician. Patient information is to be provided in accordance with the product information sheet.

Pre-operative planning has a significant influence on the success of the operation� Correct implant selection depends upon it� Merete GmbH provides support in this regard� X-ray templates are available for pre-operative planning purposes� They can be used to determine correct nail and spacer lengths and diameters prior to surgery�

When selecting nails, remember to subtract the 18 mm clamping length from the selected nail length (Fig� 1)� This shortens the nail’s implantation length (see Table 1)�

Total length Implantation length

60 mm 42 mm

70 mm 52 mm

90 mm 72 mm

110 mm 92 mm

130 mm 112 mm

150 mm 132 mm

200 mm 182 mm

Table 1

Imp

lan

tati

on

len

gth

18 mmClamping length

Tota

l le

ng

th

Figure 1 Nail length

Page 10: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)10

Surgical Technique

Please note the following when using the instruments:

Use the ratchet screwdriver with 3.8 mm and 5.0 mm diameter interlocking screws�

Spacer: 20 and 25 mm diameter Spacer: 34 mm diameter

Clamping screws Clamping screws

3.5 mm 5.0 mm

Ref. GB90204 Ref. GA90024

Ref. GB90213

Ref. GB90210

Ref. GB90205 Ref. GA90026

When using 3.5 mm diameter clamping screws (20 mm diameter humerus spacer + 25 mm diameter tibia spacer), use the instruments with the blue handles�

When using 5.0 mm diameter clamping screws (34 mm diameter femur spacer), use the instruments with the yellow handles�

Nails: 7 mm and 8 mm diameter Nails: 9, 10, 12, 14, 16, 18 and 20 mm diameter

Interlocking screws Interlocking screws

3.8 mm diameter(Ref� GB33818S - GB33832S)

5.0 mm diameter(Ref� GB35020S - GB35065S)

Ref. GB90209 Ref. GB90204

Ref. GB90213

Page 11: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

11

OsteoBridge™ IDSF SystemSurgical Technique

X Preparing the defect to be treated

Figure 2 Marked bone

Figure 3 Preparing the medullary cavity

Mark beside the bone segment to be resected (e�g�, with a SteriPen™ or two K-wires), so that the bone can be aligned along its original axis following resection (Fig� 2)�

Only resect after the markings are in place!

The bone defect can now be resected and the defect zone prepared for the spacer� Be sure to saw away the bone section being replaced in a straight line, in order to maintain a good surface for the spacer� Measure the resected bone using the steel ruler (Ref� AI90300) and select a spacer whose length approximately matches that of the resected bone� Selecting a slightly shorter spacer will make it easier to fit� The diameter of the spacer depends on the limb being operated upon�

1

For the operation to succeed, it is important to prepare the medullary cavity carefully, in accordance with the following instructions� Prepare the medullary cavity by reaming it proximally and distally to the desired diameter (Fig� 3), ideally using flexible intermedullary reamer (not included)� The medullary cavity should be reamed in 0.5 mm steps until contact with the corticalis has been made. Make sure there is at least 4 cm in length of cortical bone contact, in order to ensure stable anchorage of the nail within the bone� Correct nail length can be determined based on the length of intramedullary reamer guide wire inserted�

2

Page 12: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)12

Surgical Technique

X Non-cemented implantation

Ream the medullary cavity in accordance with the following table (Table 2)� Exact preparation of the medullary cavity is necessary to ensure proper PressFit fixation�

Nail diameter Reamer diameter

7 mm 7 mm

8 mm 8 mm

9 mm 9 mm

10 mm 10 mm

12 mm 12 mm

14 mm 14 mm

16 mm 16 mm

18 mm 18 mm

20 mm 20 mm

X Cemented implantation

For cemented implantation, the medullary canal should have a diameter at least 2 mm greater than the desired nail diameter� This will permit a cement coating thickness of 1 mm� Make sure that the thickness of the cement coating does not exceed 2 mm (Table 3)�

Nail diameter Reamer diameter

7 mm 9 mm

8 mm 10 mm

9 mm 11mm

10 mm 12 mm

12 mm 14 mm

14 mm 16 mm

16 mm 18 mm

18 mm 20 mm

20 mm 22 mm

Table 3

Table 2

Caution: Do not use interlocking screws for cemented nail fixation� Pilot drill oscillations in the interlocking holes of the nail will generate a clearance between implant and cement which deteriorates the rotational stability of the construct� Also, hardened cement should not be perforated since cracks may be introduced when using the drill� Postoperatively cyclic loading may enlarge the clearance or the cracks until breakage, dislocation or migration of a component� Consequently, the risk for the following iatrogenic complications increases when the cemented construct is weakended through interlocking screws insertion• Loosening of the locking screw• Locking screw breakage• Loosening of the nail• Breakage of the nail• Breakage of the cement mantle• Heat generation• Breakage of the drill (drill is not designed for use in bone cement)

Page 13: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

13

OsteoBridge™ IDSF SystemSurgical Technique

Trial nails (Ref� GB92201 - GB92423) can be used to check the appropriate nail diameters and lengths (Fig� 4)� Different nail lengths and diameters may be used in the proximal and distal sections of the bone�

Use the spacer gauge (Ref� GB90212) to determine the spacer length required (Fig� 5)� With bone defects larger than 70 mm, the steel ruler (Ref� AI90300) may be used as well�

X Implant selection

4010 10 10

Figure 4 Using the trial nails

Figure 5 Spacer gauge

Use the extractor for trial nails (Ref� GB90211) and the slotted hammer (Ref� AI00048) to remove the trial nails� To do this, screw the extractor into the proximal bore of the nail (Fig� 6)�

Figure 6 Extracting the trial nail

NOTE: Make sure that the extractor is always completely screwed into the trial nail.

Page 14: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)14

Surgical Technique

X Non-cemented insertion and locking of intramedullary nails

Figure 7 NGI with nail

Figure 8 Maximum penetration depth

Caution: Do not drive the nail any further into the bone, as the nail’s clamping area must be 18 mm in length for it to be clamped securely with the spacer!

Use the trial nails (Ref� GB92201 - GB92423) to determine the appropriate nail sizes to use based on the location of the defect and the size of the bone in question� Two different sized nails may be used as well (see system compatibility on page 6-8)�

To insert a nail into the bone, place it onto the nail guiding impacting instrument (NGI, Ref� GA90100), and then tighten the connection screw so that the nail is seated firmly on the NGI (Fig� 7)�

3

Tap the connection screw on the NGI with a hammer to drive the nail into the bone until the spacer (the wedge on the NGI, Fig� 8) is flush with the bone�4

Page 15: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

15

OsteoBridge™ IDSF SystemSurgical Technique

Figure 9 Distal and proximal holes for 110 mm nail

Protection sleeve

Trocar

Proximal hole

Distal hole

Place the static interlocking screw first, followed by the dynamic interlocking screw� Place the tissue protection sleeve (Ref� GB90101) into the NGI drill hole corresponding to the nail length (note labelling on NGI); use the tip of the trocar (Ref� GB90102) to center punch a hole for the interlocking screws (Fig� 9)�

5

Page 16: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)16

Surgical Technique

Figure 10 Drilling the locking holes with the help of the drill sleeve

Drill

Drill sleeve

Tissue protection sleeve

Drill a bicortical locking hole with the help of the appropriate spiral drill� Use the corresponding drill sleeve (Ref� GB90132/GB90145) as a guide (Fig� 10)�

• 3�2 mm diameter drill (Ref� GB90232) for 3�8 mm diameter interlocking screws• 4�5 mm diameter drill (Ref� GB90245) for 5�0 mm diameter interlocking screws

6

Page 17: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

17

OsteoBridge™ IDSF SystemSurgical Technique

Guide the correct interlocking screw through the tissue protection sleeve and tighten it using the included screwdriver� To do this, connect the correct screwdriver to its ratcheted hand grip (see page 10)�

Figure 11 Depth gauge for determining screw length

Figure 12 Distal bone with locked nail

Note:Dynamisation of the nails is a decision left to the discretion of the operating physician.

Use the scale on the drill (Fig� 10) or the depth gauge (Ref� AI00200, Fig� 11) to determine the necessary screw length� Start by removing the drill sleeve� Then guide the depth gauge through the protection sleeve and place it on the bone� Push the narrow rod of the depth gauge through the bone and hook it to the opposite corticalis� Always round odd numbers to the next size up (e�g�: 43 mm measured 44 mm interlocking screw)�

Insert the second interlocking screw using the same procedure described above� Once the holes have been placed and the screws have been inserted, the NGI can be removed (Fig� 12)� Do not begin locking the second nail until the first nail is completely locked into the bone�

All nails have two holes so that they can be double-locked� On nails 110 mm or longer, the second hole is elongated for optional dynamisation, which entails removing the distal screw several weeks after the operation�

7

Page 18: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)18

Surgical Technique

X Cemented insertion of intramedullary nails

Figure 13 NGI with nail

Figure 14 Maximum insertion depth

Caution: Do not additionally lock the nails after cementation.

Caution: Make sure that the clamping surface of the nail within the spacer remains free of bone cement!

Choose the appropriate nail sizes to use based on the location of the defect and the size of the bone in question� Two different sized nails may be used as well (consult the allocation tables, which are included with the instruments as laminated sheets)�

To insert a nail into the bone, place it onto the nail guiding / impacting instrument (NGI) (Ref� GA90100), and then tighten the connection screw so that the nail is seated firmly on the NGI (Fig� 13)�

8

Once the prepared medullary canal has been cleaned and dried, insert the bone cement� Tap the connection screw on the NGI with a hammer to drive the nail into the bone until the spacer (the wedge on the NGI, Fig� 14) is in contact with the bone� Do not drive the nail any further into the bone, as the nail’s clamping area must be at least 18 mm in length for it to be clamped securely with the spacer! Wait until the bone cement has completely hardened before proceeding with the remaining steps�

9

Page 19: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

19

OsteoBridge™ IDSF SystemSurgical Technique

X Mounting the spacer

Figure 15 Preparing to mount the spacer

Figure 17 Gap between bone and spacer

Figure 16 Nail with clamping area completely inside spacer

Caution:When using 18 mm and 20 mm diameter nails, be sure to leave a gap between the nail and the edge of the spacer (Fig. 17).

Gap

Once the nails have been inserted, the spacer can be mounted onto the two nails� Start by placing the thread-guiding half-shell under the nails dorsally (Fig� 15)� At this point, it will no longer be possible to move the nails to adjust their length within the medullary cavity� The spacer is flush with the clamping areas of the nails (Fig 16)�

10

Page 20: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)20

Surgical Technique

Use the Guiding pins to place the upper spacer half-shell on the nails (Fig� 19)�

Figure 18 Guiding pins for centering the spacer half-shells

Figure 19 Guided spacer half-shells

Guiding pins

TibiaHumerus M 3�5 Ref� GB90200

(Instrument tray GB90010)

Femur M 5�0 Ref� GA90003(Instrument tray GB90011)

Table 4

Four guiding pins are included with the instruments for use in centering the spacer half-shells� Screw all four guiding pins into the outer set of threaded holes on the dorsal half-shell (Fig� 18)� This is easiest when the thread-guiding half-shell has already been placed on the nails dorsally�

11

Page 21: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

21

OsteoBridge™ IDSF SystemSurgical Technique

Figure 21 Using the markings to align the bone

Figure 20 Secured spacer

Caution:To position the spacer correctly using the the spacer clamp, make sure the clamp is centered on the spacer.

The spacer can be clamped into the correct position using the included spacer clamp (Ref� GB90208)� Insert the inner four clamping screws and tighten them slightly with the corresponding screwdriver (see page 10)� Then remove the guiding pins and screw in the four remaining clamping screws (see page 23)� Each spacer includes the eight required clamping screws as well as two reserve clamping screws�

Make sure that the markings made on the bone at the beginning are opposite one another again�

12

Page 22: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)22

Surgical Technique

Figure 22 Correctly positioned spacer

Note:Accumulation of bone material on the spacer is advisable to facilitate bone bridge growth�

Figure 23 Correctly positioned spacer

The spacer half-shells must be parallel to one another� Make sure they are exactly parallel (uniform gap on both sides) before tightening the screws to the same slight extent (Figs� 22 and 23)�13

Page 23: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

23

OsteoBridge™ IDSF SystemSurgical Technique

Figure 24 Using the torque limiter

Figure 25 Sequence for tightening clamping screws

The torque limiter consists of two parts, a T-handle and a hex screwdriver (see page 10) connected to one another with an AO-connection� Starting on the inside with Clamping Screw 1, tighten the screws in order crosswise (see Fig� 25)� When the correct torque is reached, the torque limiter will overtwist with a “pop”�

Tighten all eight screws three times in the order shown above until the correct torque has been reached�

X IMPORTANT INFORMATION: TIGHTENING THE CLAMPING SCREWS

1. 4.

3. 2.

7. 6.

5. 8.

The clamping screws must be tightened into their final position with the corresponding torque limiter (see page 10) in accordance with the schema below, in order to ensure even spacer clamping (Fig� 24):14

Note:To avoid tilting or misalignment of the clamping screws, ensure that there is no tension in the system from soft tissue�

Page 24: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)24

Surgical Technique

Before closing the wound, perform and document an intra-operative X-ray check to ascertain that the implant is correctly seated (Fig� 26)�

Post-operatively, it is important to ensure load reduction on the affected limb in accordance with its healing progression�

Figure 26 X-ray image of a spacer with bone overgrowth

Sour

ce: P

rof�

C� L

ohm

ann,

Uni

vers

ity C

linic

of M

agde

burg

X Wound closure and post-operative care

Page 25: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

25

OsteoBridge™ IDSF SystemSurgical Technique

When bridging a bone defect larger than 70 mm, two spacers are to be used in combination with a spacer connector� Place the spacer half-shells on the connector and tighten the screws as described above (Fig� 27)� The connectors are adapted to the spacers’ respective inner diameters� It is advisable to screw one spacer into place completely, and then begin fitting the second spacer�

Figure 27 Two spacers connected by a spacer connector

X Additional surgical technique

Page 26: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)26

Surgical Technique

In the event that the OsteoBridge™ IDSF System needs to be revised, please order the instrument tray GB90006�Start by removing all clamping screws from the spacer (Fig� 28), and then remove the upper and lower spacer half-shells (Fig� 29)�

Figure 28 Removing the clamping screws

Figure 29 Removing the spacer half-shell

X OsteoBridge™ IDSF revision

Page 27: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

27

OsteoBridge™ IDSF SystemSurgical Technique

Figure 30 Removing interlocking screws

Figure 31 Screwing in the extractor Figure 32 Extracting the nails

Before extracting the implanted nails, it is important to remove all of the interlocking screws (Fig� 30)� After that, screw the extractor (Ref� GB90203) into each nail (Fig� 31) and tap it out with the slotted hammer (Ref� AI00048) to remove it (Fig� 32)�

Caution:Use only the implant extractor (Ref. GB90203) to extract implanted nails.

Note: Make sure that the extractor is always completely screwed into the nail.

Page 28: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)28

Ordering Information

Ordering Information X OsteoBridge™ IDSF Implants

Spacer (incl. 8+2 Screws), sterile

Length20 mm

diameter25 mm

diameter34 mm

diameter

40 mm GB60001S GB60011S GB60021S

50 mm GB60002S GB60012S GB60022S

60 mm GB60003S GB60013S GB60023S

70 mm GB60004S GB60014S GB60024S

Spacer Connector for Spacer, sterile

Diameter Ref.20 mm GB61001S

25 mm GB61002S

34 mm GB61003S

Interlocking screws3.8 mm diameter, sterile

Length Ref.18 mm GB33818S

20 mm GB33820S

22 mm GB33822S

24 mm GB33824S

26 mm GB33826S

28 mm GB33828S

30 mm GB33830S

32 mm GB33832S

Interlocking screws5.0 mm diameter, sterile

Length Ref.

20 mm GB35020S

22 mm GB35022S

24 mm GB35024S

26 mm GB35026S

28 mm GB35028S

30 mm GB35030S

32 mm GB35032S

34 mm GB35034S

36 mm GB35036S

38 mm GB35038S

40 mm GB35040S

42 mm GB35042S

44 mm GB35044S

46 mm GB35046S

48 mm GB35048S

50 mm GB35050S

52 mm GB35052S

54 mm GB35054S

56 mm GB35056S

60 mm GB35060S

65 mm GB35065S

Page 29: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

29

OsteoBridge™ IDSF SystemOrdering Information

X OsteoBridge™ IDSF Trial nails

OsteoBridge™ IDSF Trial Nail Tray - Humerus

Ref.: GB92200

Diameter/Length

7 mm dia.

8 mm dia.

9 mm dia.

10 mm dia.

60 mm GB92201 GB92204 GB92207 GB92210

70 mm GB92202 GB92205 GB92208 GB92211

90 mm GB92203 GB92206 GB92209 GB92212

110 mm - - - GB92213

130 mm - - - GB92214

Hu

meru

s

OsteoBridge™ IDSF Trial Nail Tray - Tibia

Ref.: GB92300

Diameter/Length

9 mm dia.

10 mm dia.

12 mm dia.

14 mm dia.

60 mm GB92304 GB92307 - -

70 mm GB92305 GB92308 GB92312 -

90 mm GB92306 GB92309 GB92313 GB92317

110 mm - GB92310 GB92314 GB92318

130 mm - GB92311 GB92315 GB92319

150 mm - - GB92316 GB92320

200 mm - - - GB92321

Tib

ia

OsteoBridge™ IDSF Trial Nail Tray - Femur

Ref.: GB92400

Diameter/Length

10 mm dia.

12 mm dia.

14 mm dia.

16 mm dia.

18 mm dia.

20 mm dia.

60 mm GB92401 - - - - -

70 mm GB92402 GB92406 - - - -

90 mm GB92403 GB92407 GB92411 - GB92420 GB92422

110 mm GB92404 GB92408 GB92412 GB92416 GB92421 GB92423

130 mm GB92405 GB92409 GB92413 GB92417 - -

150 mm - GB92410 GB92414 GB92418 - -

200 mm - - GB92415 GB92419 - -

Fem

ur

Page 30: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)30

Ordering Information

X OsteoBridge™ IDSF Instrument Tray

Nr. Ref. Description Qty.

1 GB90200 Guiding Pin M 3�5 4

2 GB90232 3�2 mm dia� Drill with AO-connector for 3�8 mm dia� Interlocking screws 5�0 mm dia� 2

3 GB90245 4�5 mm dia� Drill with AO-connector for 5�0 mm dia� Interlocking screws and Clamping screws 2

4 GB90204 Hex 3�5 mm Screwdriver with AO-connector for 3�5 mm dia� Clamping screws 1

5 GB90209 Hex 2�5 mm Screwdriver with AO-connector for 3�8mm dia� Interlocking screws 1

6 GB90208 Spacer clamp 1

7 GB90205 Torque Limiter with T-handle and AO-connector 4�5 Nm for Clamping screws 3�5 mm dia� 1

8 AI00048 Slotted hammer 1

9 AI00200 Depth Gauge 1

10 GB90213 Handle with ratchet and AO-connector small for Hex� 2�5 mm and Hex� 3�5 mm Screwdriver 1

11 AI90300 300 mm Steel ruler 1

Ref. Description

GB90010 OsteoBridge™ IDSF Instrument Tray 1/2

Please have intramedullary reamer ready at hand, if needed

Page 31: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

31

OsteoBridge™ IDSF SystemOrdering Information

1

2

3

4

5

6

7

8

9

10

11

Page 32: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)32

Ordering Information

Nr. Ref. Description Anzahl

1 GA90100 Nail Guiding / Impacting Instrument 2

2 GB90101 Protection Sleeve 1

3 GB90212 Spacer Gauge 1

4 GB90211 Extractor short for Trial Nails 1

5 GB90132 Drill Sleeve for 3�2 mm dia� drill 1

6 GB90145 Drill Sleeve for 4�5 mm dia� drill 1

7 GB90102 Trocar 1

8 GB90210 Handle with ratchet and 1/4"-connector for Hex� 5�0 mm Screwdriver 1

9 GA90003 Guiding Pin M 5�0 4

10 GA90024 Hex 5�0 mm Screwdriver with 1/4"-connector for 5�0 mm dia� Clamping Screws 1

11 GA90026 Torque Limiter with T-handle and 1/4"-connector 9�5 Nm for 5�0 Clamping Screws 1

Ref. Description

GB90011 OsteoBridge™ IDSF Instrument Tray 2/2

Page 33: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

33

OsteoBridge™ IDSF SystemOrdering Information

1

2

3

4

5

6

7

8

9

10

11

Page 34: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

OsteoBridge™ IDSF System

Merete Technologies Inc. (MTI)34

Surgical Technique

Page 35: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

35

OsteoBridge™ IDSF SystemSurgical Technique

Page 36: IDSF - mereteusa.com · Ream the medullary cavity in accordance with the following table (Table 2) Exact preparation of the medullary cavity is necessary to ... Do not use interlocking

Merete Technologies Inc� (MTI)

U.S. Distributor:

Merete Technologies Inc. (MTI)One Lincoln Centre18W140 Butterfield RoadOakbrook Terrace, IL 60181

Customer Service Phone: 630-869-1091Fax: 630-445-1752

E-Mail: service@merete-medical�comwww�mereteUSA�com