IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head,...
-
Upload
kelley-conley -
Category
Documents
-
view
242 -
download
0
Transcript of IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head,...
![Page 1: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/1.jpg)
IAEAInternational Atomic Energy Agency
IAEA quality audits in radiotherapy
Joanna Izewska
Head, Dosimetry Laboratory
Dosimetry & Medical Radiation Physics Section, Division of Human Health
![Page 2: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/2.jpg)
IAEA
IAEA audits in radiotherapy
• SINCE 1969: IAEA/WHO TLD postal dose audits of radiotherapy beam calibration (45 years of operation):
130 countries, >2000 hospitals, >10000 beam checks• SINCE 2005: Quality Assurance Team for Radiation Oncology (QUATRO)
• SINCE 2010:
TPS end-to-end audits
![Page 3: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/3.jpg)
IAEA 3
Currently >90% results are within 5% acceptance limit
2012-2013: N = 1271 , m = 1.005, SD = 0.068, 92% res. within 5%
1969-2013: N = 10226 , m = 1.008, SD = 0.08, 83% res. within 5% limit
±5%
![Page 4: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/4.jpg)
IAEA
TLD results within the 5% limit
4
![Page 5: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/5.jpg)
IAEA
TLD results: use of CoPs (2000-2013)
126 deviations beyond 20% excluded
TLD results 2000-2013
5
0
10
20
30
40
50
60
70
80
90
100
2002 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012
Per
cen
tag
e u
sin
g p
roto
col
Year of irradiation
ND,w based
Nk based
Nx based
![Page 6: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/6.jpg)
IAEA 6
IAEA TPS audit procedures
Set of practical tests for dosimetry calculations reflecting basic treatment techniques in a typical radiotherapy hospital
IAEA scientific secretary: S. Vatnitsky
![Page 7: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/7.jpg)
IAEA
IAEA supported TPS audit in Europe
• Eight European countries conducted TPS audits with the IAEA support
• Approx. 60 hospitals participated
• Approx. 200 datasets have been collected for beam/TPS algorithm combination
• Several dosimetry and TPS-related problems were identified and fixed
7
![Page 8: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/8.jpg)
IAEA 8
Comprehensive RT audit: QUATRO
![Page 9: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/9.jpg)
IAEA
QUATRO audit procedures
• Auditors: RO, MP, RTT
• Typically 5 days per RT centre
• Entrance briefing
• Assessment: tour of facility, staff interviews, review & evaluation of procedures and documentation, measurements, tests, observation of practical work
• Exit briefing: feedback to the department, preliminary recommendations, questions, discussion.
9
![Page 10: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/10.jpg)
IAEA
QUATRO missions
lower middle incomeupper middle incomehigh income
• Training of auditors and regional QUATRO workshops in all regions • 70 QUATRO missions to date: Africa – 6; Asia – 18; Europe – 30 + 3 re-
audits; Latin America – 13
![Page 11: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/11.jpg)
IAEA 11
Europe: recommendations by QUATRO
• Support staff continuous development
• Strengthen QA programmes and procedures
• Equipment: get more, use better
• Increase staffing levels: MP, RO, RTT
• Staff requires training: RTT, RO, MP
• Improve treatment protocols• Improve structure and
organization of cancer centres
![Page 12: IAEA International Atomic Energy Agency IAEA quality audits in radiotherapy Joanna Izewska Head, Dosimetry Laboratory Dosimetry & Medical Radiation Physics.](https://reader036.fdocuments.net/reader036/viewer/2022081503/56649d755503460f94a55a1b/html5/thumbnails/12.jpg)
IAEA
THANK YOU
12