How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with...

51
How to achieve good results with the GORE® VIABAHN® Endoprosthesis for treating complex femoropopliteal occlusive disease: Results from the Japanese IDE trial National Principal Investigator: Takao Ohki, MD, PhD Jikei University, Tokyo, Japan

Transcript of How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with...

Page 1: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

How to achieve good results with the GORE® VIABAHN® Endoprosthesis for treating complex

femoropopliteal occlusive disease: Results from the Japanese IDE trial

National Principal Investigator:

Takao Ohki, MD, PhD

Jikei University, Tokyo, Japan

Page 2: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Disclosure

Speaker name: Takao Ohki

.................................................................................

I have the following potential conflicts of interest to report:

Consulting

Employment in industry

Stockholder of a healthcare company

Owner of a healthcare company

Other(s)

I do not have any potential conflict of interest

Page 3: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface

Ultra-thin wall ePTFE tube

Unique, durable bonding film

Polished nitinol support

Contoured proximal edge

Lengths: 2.5, 5, 7.5, 10, 15, 25 cm

Diameters: 5–13 mm

CBAS Heparin Surface*

* PROPATEN Bioactive Surface is synonymous with the CBAS Heparin

Surface.

Page 4: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Achilles' heel of FP bypass (PTFE): Thrombosis

CBAS technology

3 years post implant

Propaten

Control PTFE

Page 5: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Compliant with the Mechanical Forces of the SFA

• More than 700,000 GORE® VIABAHN®

Endoprosthesis sold worldwide

• Very low incidence of reported fractures(< 0.015%)

• Capable of longitudinal compression with little residual force

• Superb flexibility

Page 6: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Mechanical Forces in the SFAFlexion

Flexion“The curvature of the femoral vessels was studied and quantified in stretched and flexed positions…Three or more small curves were seen proximal to the knee joint in all volunteers”1

1 Wensing PJ, Scholten FG, Buijs PC, Hartkamp MJ, Mali WP, Hillen B. Arterial tortuosity in the femoropopliteal region during knee

flexion: a magnetic resonance angiographic study. Journal of Anatomy 1995;187(Pt 1):133-139.

Page 7: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Mechanical Forces in the SFALongitudinal Compression

Longitudinal Compression“From the supine position to the fetal position, the SFA shortened 13% ± 11% (P < .001)”1

The GORE® VIABAHN®

Endoprosthesis is capable of longitudinal compression with little residual force

1 Cheng CP, Wilson NW, Hallett RL, Herfkens RJ, Taylor CA. In Vivo MR Angiographic Quantification of

axial and twisting deformations of the superficial femoral artery resulting from maximum hip and knee

flexion. Journal of Vascular & Interventional Radiology 2006;17(6):979-987.

Page 8: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

#1 Achilles' heel of SFA stenting: ISR

Page 9: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Blocking Neointimal Hyperplasia

• Achilles' heel of SFA stenting Mechanical Barrier

Stent Graft(VIABAHN)

• Original stimulus for stenosisremoved from the equation

• Pore size provides a barrier to tissue ingrowth

Page 10: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Difference between BMS ISR (diffuse, usually symptomatic) and covered stent edge stenosis (focal and often asymptomatic)

If stenosis does recur, have changed an initial TASC II C or D lesion to TASC A

Tough restenosis

Page 11: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Ideal SFA stent

- Mechanical barrier against intimal hyperplasia

- Fracture resistant

- Thrombosis resistant

Page 12: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Viabahn: 5.0-250mm

Edge stenosis

Typical Viabahn restenosis

Only at the edge

PSVR:2.4

Re-intervention easily performed with

Viabahn 5.0mm x 2.5cm

Page 13: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

The evolution of a device

Heparin-coated, contoured edge, low profile (0.018” compatible) device used in the Japan IDE Trial.

Page 14: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Japan VIABAHN® IDE Trial:Basic strategy of the trial

• No interventional treatment was approved for lesions greater than 15cm in length in Japan

• Significant number of FP bypass was done for TASC II C/D lesions at the start of the trial

• A device that was proven in long, complex lesions was needed• Invasiveness measures devised to prove lower invasiveness of

VIABAHN compared to FP bypass: if equal performance percutaneous wins

• Repeat percutaneous procedures under local anesthesia for ISR not considered as a defeat

• Avoidance of bypass surgery and maintenance of flow important• Led to Primary Assisted Patency =avoidance of surgery as primary

endpoint

Page 15: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

For the purpose of IDE approval the downside or FP bypass was exaggerated

Wound infection following FP bypass

Page 16: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Gore Japan IDE Clinical Study:Performance in Patients that may Otherwise Require Bypass

Page 17: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Gore Japan IDE Clinical Study: Performance in patients who may otherwise require bypass

Follow-up:• 1, 3, 6, 12 and 24 month CDUS

• Annual visits through 5 years to assess

fracture, AE’s

• Primary patency data available through 24

months

• fTLR data will be collected through 60

months

Page 18: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Outcome depends on what you treat:Where Viabahn Japan data fits

* Prospective Randomized or Prospective Multicenter (> 2 sites) SFA studies included. Studies that did not report primary patency at 12-months are not included. Patency

definitions may vary: where Kaplan-Meier estimates with a PSVR of ≥ 2.5 are available, these were used for comparison. For the GORE® VIABAHN® Endoprosthesis, only

studies that utilized the GORE® VIABAHN® Endoprosthesis with Heparin Bioactive Surface were included. Drug-coated balloon studies were only included for those studies

that report data for devices currently approved in the US. Atherectomy: non-SFA lesions were included in the reported patency rates in the CELLO (N = 5 out of 65 total)

and Pathway PVD trials (N = 76 out of 210 total). Non-SFA lesions were included in the reported patency rates in the Micari, et al. study (N = 12 out of 114 lesions total),

LEVANT I Trial (N = 4 out of 49 total), LEVANT II Trial (N = 31 out of 316 total), and IN.PACT SFA I and II Trial (15 out of 221 total).1-29

** Gore RELINE clinical study was for treatment of in-stent restenosis in the SFA.27

Page 19: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Proper Sizing Correlates with Optimal OutcomesSize was not measured but eyeballed in the Viper Trial

IVUS seldom used

* Gore VIPER Clinical Study: A prospective, multi-center study of 119 limbs in the U.S. (19 cm average lesion length). Overall results 73% Primary Patency. Retrospective

analysis was done on device sizing by Core Lab on 95 limbs. For devices oversized < 20% at the proximal edge Primary Patency was 88%.1

** GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 12-month primary patency of 92% as defined by evidence of flow with no target lesion

revascularization (TLR). The same study demonstrated 88% 12-month primary patency when defined by PSVR of < 2.5 without TLR.

Page 20: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Inclusion / Exclusion Criteria Designed to Target Complex SFA Disease

• Inclusion Criteria

—Rutherford 2–5

—ABI ≤ 0.9 or TBI ≤ 0.5

—Surgical bypass candidate

—Angiographic:

• Lesion length ≥ 10 cm (No upper limit)

• SFA lesion

– 1 cm below the SFA origin and ending 1

cm above the intercondylar notch

• Patent distal popliteal artery

• At least one patent tibial artery

• Lesion can be pre-dilated

• Reference vessel diameters between 4.0

and 7.5 mm. Must be measured and not

estimated

• IVUS encouraged

• Exclusion Criteria

—Untreated flow-limiting aortoiliac

disease (could be treated during index

procedure)

—Any previous stenting or surgery

in target vessel

—Femoropopliteal aneurysm

—Rutherford 5 with active infection

—Known coagulation disorder

—Current dialysis

—Contraindication to

anticoagulation or antiplatelet

Page 21: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Anatomical inclusion criteria

Proximal medial condyle

Proximal landing zone >1cmProfunda

SFA

Popliteal

1cm

1cm

Inclusion criteria•Lesion length >10cm•No upper limit•Rutherford category 2-5•Reference vessel diameter 4 to 7.5 mm

Page 22: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Majority of Patients Were Older Men with a History of Smoking and Diabetes

Page 23: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Evaluated in Long, Complex Lesions Generally indicated for Bypass

Population enrolled generally suitable for bypass: Mean lesion length 22cm and 84.5% TASC II C or D

Page 24: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Device Evaluated in a Challenging Biomechanical Environment

With over half the patients reporting sitting seiza-style or cross-legged at least 25% of the time, this population presents a challenging biomechanical environment for long lesions that were allowed to extend through Hunter’s canal to 1 cm above the intercondylar notch.

Seiza style

Page 25: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Stent-Grafts Were Less Invasive than Bypass

Treatment with stent-grafts shown to be less invasive than bypass by all measures

* Retrospective series of 68 bypass patients collected from study sites (6 / 15 sites). Patient selection

was done according to a similar set of inclusion / exclusion criteria as those treated by the study

device.

Page 26: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Primary effectiveness endpoint met

91 / 100 patients at the end of one year follow-up were occlusion free

94.1% Primary Assisted Patency K-M at day 365

Statistically non-inferior to 80% PP of surgical bypass from literature

Page 27: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

88% Primary Patency in 21.8 cm Average Length Lesions

• In the Gore Japan IDE Clinical Study, the GORE® VIABAHN® Endoprosthesis demonstrated 88% 12-month primary patency*

• Average Lesion Length 21.8 cm

* GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 12-month primary patency of 92% as

defined by evidence of flow with no target lesion revascularization (TLR). The same study demonstrated 88% 12-

month primary patency when defined by PSVR of < 2.5 without TLR.

Page 28: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Subgroup analyses93% Primary Patency in Lesions Between 10–20 cm

• In the Gore Japan IDE Clinical Study, the GORE® VIABAHN® Endoprosthesis demonstrated 92.7% 12-month primary patency in lesions between 10 and 20 cm*

• Average Lesion Length 16.2 cm

* GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 100% 12-month primary patency as defined by evidence of flow with no target lesion

revascularization (TLR) in lesions ≤ 20 cm and ≥ 10 cm (shortest length allowed for enrollment). Average lesion length 16.2 cm. N = 43 of 103. The same study

demonstrated primary patency of 92.7% when calculated using PSVR of < 2.5 with no TLR in lesions ≤ 20 cm and ≥ 10 cm (shortest length allowed for enrollment).

Page 29: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Subgroup analyses85% Primary Patency in Lesions > 20 cm

• In the Gore Japan IDE Clinical Study, the GORE® VIABAHN® Endoprosthesis demonstrated 84.8% 12-month primary patency in lesions greater than 20 cm*

• Average lesion length: 25.7 cm

* GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 86.5% 12-month primary patency as defined by evidence of flow with no target lesion

revascularization (TLR) in lesions > 20 cm. Average lesion length 25.7 cm. N = 60 of 103. The same study demonstrated primary patency of 84.8% when calculated using

PSVR of < 2.5 with no TLR in lesions > 20 cm.

Page 30: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Several Key Patient Factors Showed No Impact to Primary Patency

No significant differences in patency

• Diabetes

• Lesion calcification

• Subject age

• Number of runoff vessels

• Smoking status

• Hypertension

Page 31: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

PTA dilation

6 x 15cm

VIABAHN®

Merit and demerit of Covering Collaterals

Page 32: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Covering Collaterals

Complete lesion coverage / Increased flow through device

PreserveCollaterals

Minimize failureLess risk of limb threat?

Competing interests

-Complete coverage prevents progression

-Prevents competing flowCollaterals provide a flow path in case of device failure

Page 33: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Safety endpointNo Acute Limb Ischemia, No Fractures, No Bypass

•No cases of Acute Limb Ischemia (ALI) were

observed through end of 12 month follow-up

*Occlusions : 6

• No fractures were identified by core laboratory

• No patient required bypass

• Limb salvage 100%

• No patient death, target vessel revascularization, or major

amputation within 30 days

Page 34: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Proper Sizing Correlates with Optimal OutcomesSize was not measured but eyeballed in the Viper Trial

IVUS seldom used

* Gore VIPER Clinical Study: A prospective, multi-center study of 119 limbs in the U.S. (19 cm average lesion length). Overall results 73% Primary Patency. Retrospective

analysis was done on device sizing by Core Lab on 95 limbs. For devices oversized < 20% at the proximal edge Primary Patency was 88%.1

** GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 12-month primary patency of 92% as defined by evidence of flow with no target lesion

revascularization (TLR). The same study demonstrated 88% 12-month primary patency when defined by PSVR of < 2.5 without TLR.

Page 35: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Unlike the Viper trial size was measured in the Japan Trial

* Gore VIPER Clinical Study: A prospective, multi-center study of 119 limbs in the U.S. (19 cm average lesion length). Overall results 73% Primary Patency. Retrospective

analysis was done on device sizing by Core Lab on 95 limbs. For devices oversized < 20% at the proximal edge Primary Patency was 88%.1

** GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 12-month primary patency of 92% as defined by evidence of flow with no target lesion

revascularization (TLR). The same study demonstrated 88% 12-month primary patency when defined by PSVR of < 2.5 without TLR.

Proper sizing is critical to achieving successful outcomes.1, 5

Page 36: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Unlike the Viper trial size was measured in the Japan Trial

* Gore VIPER Clinical Study: A prospective, multi-center study of 119 limbs in the U.S. (19 cm average lesion length). Overall results 73% Primary Patency. Retrospective

analysis was done on device sizing by Core Lab on 95 limbs. For devices oversized < 20% at the proximal edge Primary Patency was 88%.1

** GORE® VIABAHN® Endoprosthesis Japan IDE Clinical Study demonstrated 12-month primary patency of 92% as defined by evidence of flow with no target lesion

revascularization (TLR). The same study demonstrated 88% 12-month primary patency when defined by PSVR of < 2.5 without TLR.

Proper sizing is critical to achieving successful outcomes.1, 5

Page 37: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

How to achieve good results with the GORE® VIABAHN® Endoprosthesis: Lessons from the Japan trial

Best practices followed throughout the study

•Vessel size at landing zones was required to be measured, not estimated

—Quantitative angiography was used in all cases

—IVUS was used in 70%

—Device sizes were chosen per IFU (5 – 20% oversizing)

•Landing zones were carefully selected so that all disease was covered

•Pre and post-dilation performed in all cases (high pressure balloons in most)

•Dual antiplatelets were required for six months and recommended for 12

months

—Aspirin and clopidogrel (or other theopyridine class)

—Cilastozol additionally prescribed in some cases

•Consistent follow-up was performed

—1, 3, 6, and 12 month duplex ultrasound

Page 38: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

A BFigure 1

* *

VIABAHN Japan IDE Example

Page 39: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

3 yrs post stent

A

B

C

ABI unchanged at 0.9

Page 40: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

pre post pre post

Viabahn

Rt SFA treated with Viabahn Lt SFA with BMS

BMS

2Yrs 6Month

Japan Viabahn Case with internal controlCourtesy of Dr Yamaoka

Page 41: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

• Despite including mostly TASC II C/D lesions (84.5%), the GORE® VIABAHN® Endoprosthesis showed 88% Primary Patency and 94% Primary Assisted Patency

• These are lesions normally recommended for treatment by above knee bypass

• The GORE® VIABAHN® Endoprosthesis has the potential to replace bypass for treatment of long/complex lesions

• In medium length lesions ≤ 20cm, Primary Patency was 93%

• Other interventional devices have mainly been studied in shorter lesions

• Stent-grafts may be preferred to other interventional treatments in medium length lesions

• With no cases of ALI, and no patients requiring bypass or amputation, safety has been shown

• The Japanese IDE Trial results are a result of careful technique

• Quantitative sizing

• Complete disease coverage

• Diligent post PTA

• Dual antiplatelet administration

• Good follow-up

Japan Viabahn Trial: Summary

Page 42: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Evolution of a device, evolution of techniques,

accumulation of data

Heparin-coated, contoured edge, low profile (0.018” compatible) device

used in the Japan IDE Trial.

Japan IDE approval

Page 43: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

2 Year Results

Will be presented in Main Arena Room 1 at 17:14 in the session titled “New concepts for

complex femoropopliteal disease”.

Eluvia stent (Majestic trial)

R 2-4 n=57

Mean Lesion length 7.1 cm

Primary patency 1 yr 96% 2 yr 78.2%

Hulsbeck SM LINC 2017

1yr success does not guarantee 2yr so stay tuned

Page 44: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film
Page 45: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film
Page 46: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Japan Viabahn conducted with Ideal SFA stent, Ideal practice

- Mechanical barrier against intimal hyperplasia

- Fracture resistant

- Thrombosis resistant

- Contour edge

- Viabahn implant tricks

- <20% oversizing

- Full lesion coverage

- DAPT, Diligent FU

Page 47: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

Proven Patency for Complex SFA Lesions

• 88% 12-month primary patency

• Good technique drives great outcomes

— Proper sizing is critical

• 93% Primary Patency in lesions between 10–20 cm (average 16.2 cm)

• Zero cases of ALI, bypass, or amputation

Page 48: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

References

1. Saxon RR, Chervu A, Jones PA, et al. Heparin-bonded, expanded polytetrafluoroethylene-lined stent graft in the treatment of

femoropopliteal artery disease: 1-year results of the VIPER (Viabahn Endoprosthesis with Heparin Bioactive Surface in the

Treatment of Superficial Femoral Artery Obstructive Disease) Trial. Journal of Vascular & Interventional Radiology

2013;24(2):165-173.

2. Lammer J, Zeller T, Hausegger KA, et al. Heparin-bonded covered stents versus bare-metal stents for complex femoropopliteal artery

lesions: the randomized VIASTAR trial (Viabahn endoprosthesis with PROPATEN bioactive surface [VIA] versus bare nitinol stent in

the treatment of long lesions in superficial femoral artery occlusive disease). Journal of the American College of Cardiology

2013;62(15):1320-1327.

3. Zeller T, Peeters P, Bosiers M, et al. Heparin-bonded stent-graft for the treatment of TASC II C and D femoropopliteal lesions: the

Viabahn-25 cm Trial. Journal of Endovascular Therapy 2014;21(6):765-774.

4. Schillinger M, Sabeti S, Loewe C, et al. Balloon angioplasty versus implantation of nitinol stents in the superficial femoral artery.

New England Journal of Medicine 2006;354(18):1879-1888.

5. W. L. Gore & Associates. Multi-center study for stent graft system for peripheral artery (VJH 11-01). Bethesda, MD: National

Library of Medicine; 2012. Available from: https://clinicaltrials.gov/ct2/show/NCT01575808. NLM Identifier: NCT01575808.

Published April 9, 2012. Updated June 9, 2015. Accessed May 10, 2016.

6. Laird JR, Katzen BT, Scheinert D, et al; RESILIENT Investigators. Nitinol stent implantation versus balloon angioplasty for

lesions in the superficial femoral artery and proximal popliteal artery: twelve-month results from the RESILIENT randomized

trial. Circulation: Cardiovascular Interventions 2010;3(3):267-276.

7. Zeller T, Tiefenbacher C, Steinkamp HJ. Nitinol stent implantation in TASC A and B superficial femoral artery lesions: the

Femoral Artery Conformexx Trial (FACT). Journal of Endovascular Therapy 2008;15(4):390-398.

8. Matsumura J; DURABILITY II Investigators. Durability II. Full data set at 12 months. Presented at the International

Symposium on Endovascular Therapy (ISET); January 15-19, 2012; Miami Beach, FL.

9. Krankenberg H, Schlüter M, Steinkamp HJ, et al. Nitinol stent implantation versus percutaneous transluminal angioplasty in

superficial femoral artery lesions up to 10 cm in length: the femoral artery stenting trial (FAST). Circulation 2007;116(3):285-292.

10. Dake MD, Ansel GM, Jaff MR, et al; Zilver PTX Investigators. Paclitaxel-eluting stents show superiority to balloon angioplasty

and bare metal stents in femoropopliteal disease: twelve-month Zilver PTX randomized study results. Circulation:

Cardiovascular Interventions 2011;4(5):495-504.

11. Dick P, Wallner H, Sabeti S. Balloon angioplasty versus stenting with nitinol stents in intermediate length superficial femoral artery

lesions.

Catheterization & Cardiovascular Interventions 2009;74(7):1090-1095.

12. Medtronic Complete™ SE Vascular Stent System [Instructions for Use]. Minneapolis, MN: Medtronic; 2013. M729425B001 Rev. 1B.

13. Bosiers M, Deloose K, Callaert J. et al. Less invasive interventions: how feasible is the low-profile approach? 4EVER study-final

results. Presented at the Leipzig Interventional Course (LINC); January 23-26, 2013; Leipzig, Germany.

14. Cordis S.M.A.R.T.® CONTROL® Stent [Instructions for Use]. Miami Lakes, FL: Cordis; 2013. 100000000922.2.

Page 49: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

12. Soukas PA. The Supera Vasculomimetic Stent in SFA disease. Presented at the New Cardiovascular Horizons 14th Annual

Conference (NCVH); June 5-7, 2013; New Orleans, LA.

13. Dake MD, Scheinert D, Tepe G, et al; Zilver PTX Single-Arm Study Investigators. Nitinol stents with polymer-free paclitaxel coating

for lesions in the superficial femoral and popliteal arteries above the knee: twelve-month safety and effectiveness results from the

Zilver PTX single-arm clinical study. Journal of Endovascular Therapy 2011;18(5):613-623.

14. Yokoi H; On behalf of the Investigators. Zilver PTX® post-market surveillance study of Paclitaxel-eluting stents for treating

femoropopliteal artery disease in Japan: 12-month results. Presented at the LINC Asia-Pacific Meeting; March 9-11, 2014; Lantau

Island, Hong Kong.

15. Ohki T, Angle JF, Yokoi H, Jaff M, Popma J. Effectiveness of the Misago stent for the treatment of superficial femoral artery disease:

12-month results of the First Japan and the United States Collaboration Trial (OSPREY). Presented at the Transcatheter

Cardiovascular Therapeutics Twenty-Sixth Annual Symposium Transcatheter Cardiovascular Therapeutics (TCT); September 13-17,

2014; Washington, DC. Journal of the American College of Cardiology 2014;64(11):B155. TCT-530.

19. Tepe G, Laird J, Schneider P, et al; IN.PACT SFA Trial Investigators. Drug-coated balloon versus standard percutaneous

transluminal angioplasty for the treatment of superficial femoral and popliteal peripheral artery disease: 12-month results

from the IN.PACT SFA Randomized Trial. Circulation 2015;131(5):495-502.

20. DeFord J. LutonixTM Drug Coated Balloon Device for the Treatment of Femoropopliteal Artery Disease. CR Bard, Inc. FDA

Executive Summary Prepared for the June 12, 2014 meeting of the Circulatory System Devices Advisory Panel.

http://www.fda.gov/downloads/

AdvisoryCommittees/CommitteesMeetingMaterials/MedicalDevices/MedicalDevicesAdvisoryCommittee/CirculatorySystemDevicesPa

nel/ UCM401269.pdf. Published June 12, 2014. Accessed April 6, 2015.

21. McKinsey JF, Zeller T, Rocha-Singh KJ, Jaff MR, Garcia LA; DEFINITIVE LE Investigators. Lower extremity revascularization using

directional atherectomy: 12-month prospective results of the DEFINITIVE LE study. JACC: Cardiovascular Interventions

2014;7(8):923-933.

22. Dave RM, Patlola R, Kollmeyer K, et al; CELLO Investigators. Excimer laser recanalization of femoropopliteal lesions and 1-year

patency: results of the CELLO registry. Journal of Endovascular Therapy 2009;16(6):665-675.

23. Zeller T, Krankenberg H, Steinkamp H, et al. One-year outcome of percutaneous rotational atherectomy with aspiration in

infrainguinal peripheral arterial occlusive disease: the multicenter pathway PVD trial. Journal of Endovascular Therapy

2009;16(6):653-662.

24. Scheinert D, Duda S, Zeller T, et al. The LEVANT I (Lutonix paclitaxel-coated balloon for the prevention of femoropopliteal restenosis)

trial for femoropopliteal revascularization: first-in-human randomized trial of low-dose drug-coated balloon versus uncoated balloon

angioplasty. JACC: Cardiovascular Interventions 2014;7(1):10-19.

25. Micari A, Cioppa A, Vadalà G, et al. Clinical evaluation of a paclitaxel-eluting balloon for treatment of femoropopliteal arterial

disease: 12-month results from a multicenter Italian registry. JACC: Cardiovascular Interventions 2012;5(3):331-338.

26. Micari A, Vadalà G, Castriota F, et al. 1-year results of paclitaxel-coated balloons for long femoropopliteal artery disease: evidence

from the SFA-long study. JACC: Cardiovascular Interventions 2016;9(9):950-956.

27. Bosiers M, Deloose K, Callaert J, et al. Superiority of stent-grafts for in-stent restenosis in the superficial femoral artery: twelve-

month results from a multicenter randomized trial. Journal of Endovascular Therapy 2015;22(1):1-10.

References

Page 50: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

28. Krankenberg H; on behalf of the FAIR Trial Investigators. DCB and ISR: Insights from the FAIR Study. Presented at the

Leipzig Interventional Course (LINC); January 27-29, 2015; Leipzig, Germany.

29. Dippel EJ; EXCITE ISR Investigators. EXCITE ISR: initial results. Presented at the Transcatheter Cardiovascular

Therapeutics Twenty-Sixth Annual Symposium Transcatheter Cardiovascular Therapeutics (TCT); September 13-17,

2014; Washington, DC.

30. Medtronic Vascular, Inc. IN.PACT® Admiral® Paclitaxel-coated PTA Balloon Catheter. [Product Brochure]. Santa

Rosa, CA: Medtronic Vascular, Inc; 2014. UC201502574EN.

31. Lutonix® 035 Drug Coated Balloon PTA Catheter [Instructions for Use]. New Hope, MN: Lutonix, Inc; 2015. DW5200-01

Rev 01.

References

Page 51: How to achieve good results with the GORE® VIABAHN® … · GORE® VIABAHN® Endoprosthesis with PROPATEN BioactiveSurface Ultra-thin wall ePTFE tube Unique, durable bonding film

How to achieve good results with the GORE® VIABAHN® Endoprosthesis for treating complex

femoropopliteal occlusive disease: Results from the Japanese IDE trial

National Principal Investigator:

Takao Ohki, MD, PhD

Jikei University, Tokyo, Japan