High Technology Medical Devices 1 guide.pdf · High Technology Medical Devices 1 | ... is resistant...

21
High Technology Medical Devices 1| Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu

Transcript of High Technology Medical Devices 1 guide.pdf · High Technology Medical Devices 1 | ... is resistant...

High Technology Medical Devices 1 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

 

   

High Technology Medical Devices 2 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

INSTRUCTIONS FOR USE (USER’S GUIDE)  GENERAL GUIDELINES 

 

This Operating manual, including specifications, is a User’s manual for the PLASON Air‐Plasma Scalpel‐Coagulator‐ 

Stimulator  

 

Read this User’s manual before using the PLASON Air‐Plasma Scalpel‐Coagulator‐Stimulator 

 

The present User’s manual contains the description of device, operating principle, technical characteristics and operating 

instructions for the PLASON Air‐Plasma Scalpel‐Coagulator‐Stimulator 

 

 Safety Signs 

 

                  ‐ Refer to user manual/booklet 

 

                 ‐ General safety sign 

 

                 ‐ General prohibiting sign 

 

                   ‐  General warning sign 

   

High Technology Medical Devices 3 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

 1.  GENERAL INDICATIONS 

 1.1.   As the basis of the action of the Air‐plasma scalpel‐coagulator‐stimulator “PLASON” (further as “device”) is assumed the action of the flow of air plasma containing the molecules of the mono‐oxide of nitrogen on the tissue of the human organism, for therapeutic effect.  1.2.   The User´s Guide is intended for the medical staff that operates the device, and engineering personnel, that conducts the maintenance and repair of the device. 

1.3.   Device is a restorable article and in the case of malfunction can undergo routine repair.  1.4.   Only medical staff that has studied this user´s guide and received training of the device use, can operate the device.  1.5.  The use of the device during the application of combustible anesthetics is inadmissible.   2. DESIGNATION 

 2.1.   Device  is  intended  for  coagulation  and  sterilization  of wound  surfaces,  evaporation  and  destruction  of  nonviable cloths and pathologic formations, dissection of biological tissues by plasma flow with the temperature up to 4000 °C, and also  for  the  stimulation  of  reparative  processes  during  the  treatment  of wounds,  trophic  ulcers,  bedsores,  sharp  and chronic inflammatory processes, scar and sclerous changes, other defeats of the external covers of soft tissues, mucous membranes and internal organs by gas flow with the temperature up to 40°C, which contains the mono‐oxide of nitrogen (NO), in the conditions of surgical departments. 

  3. TECHNICAL SPECIFICATIONS 

 3.1.   The overall dimensions of device correspond to the values, indicated in the Figure 2.  3.2.   The mass of device with the equipment does not exceed 7.5 kg. The electric power supply of device is achieved from the electric network with alternating current with a voltage of 220‐230 V with a frequency of 50 Hz  3.4.   The power, consumed by device, does not exceed 500 V∙A.  3.5.  Device  works  with  interchangeable  manipulators,  which  ensure  the  regimes  of  coagulation,  destruction  and therapeutic action (NOx‐therapy).  3.6. The overall dimensions of manipulators correspond to those indicated in Table 1.   Таble 1  

MANIPULATOR  LENGTH Mm 

DIAMETER mm 

DIAMETER OF OUTLET mm 

MARKING

Coagulator  130±2  14±0.1  4.0+0.05  dark‐blue 

Coagulator  130±2  14±0.1  1.2+0.05  red 

Destructor  130±2  14±0.1  0.7+0.05  yellow 

Stimulator‐Coagulator  130±2  14±0.1  1.8+0.05  green 

    

High Technology Medical Devices 4 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

A     B     C     D  

Fig. 1.1: Various types of manipulators: A ‐ blue coagulator, B ‐ red coagulator, C ‐ green stimulator, D – yellow destructor 

 

 

Fig. 1.2: Various types of manipulators 

   

High Technology Medical Devices 5 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

 

Fig. 2: Dimensions of PLASON device 

High Technology Medical Devices 6 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

3.7.   Nominal operating current of the manipulator 2.30 A ± 0.23 A. 

 

3.8.   Time of the start of manipulator operation after pressing the “START” button is not more than 5s. 

 

3.9.   The device ensures work for 4 hours in the intermittent service. Time of operation ‐ 5 min., pause ‐ 1 min. 

 

3.10.  Upon the switching of the “on/off” switch from “0” to “I” the “on/off” switch and the “STOP” button is illuminated. 

In  this  setting  the  desired  flow  rate  of  air  (MIN,  NORM, MAX)  can  be  selected  by  pressing  one  of  respective  control 

buttons, which correspond to the selected flow rate. 

 

3.11.  On the safety the device corresponds to requirements GOST R IEC 60601‐1‐2010 (IEC 60601‐1:2005‐12) class I with 

the working part of the type B. 

 

3.12.  The device is resistant to the mechanical actions by operation acc. to GOST R IEC 60601‐1‐2010 (national standard) 

for group 2. 

 

3.13.   The device by the operation is resistant to the action of climatic factors by GOST R 15150‐69 (national standard) 

for the the climatic performance UKhL of category 4.2              

 

3.14.  The device  in the packing of manufacturing concern is resistant to the climatic actions during the transportation 

and the storage by GOST 15150‐69 (national standard) for the conditions for storage nr. 5. 

 

3.15.  The device  in the transport packing  is resistant to the mechanical actions with the transportation and possesses 

vibration strength and shockproof quality by GOST R IEC 60601‐1‐2010 (national standard) for group 2. 

 

3.16.  The external surfaces of device and manipulators are resistant to disinfection in accordance with the requirements 

of MU‐287‐113 (national standard) 

 

3.17.  The device on  the possible consequences of  failure  relates  to  the class B by GOST R  IEC 60601‐1‐2010  (national 

standard). 

 

3.18.  Mean operating time of device to the failure ‐ is not less than 1000 hours. 

 

3.19.  Average working  life  to  the writing off of device  ‐  is  not  less  than 5 years.  The  criterion of  limiting  condition  is 

impossibility or the economic inexpediency of the restoration of device. 

 

3.20.  The time required for changing of the manipulator is not more than 3 minutes. 

 

3.21.  The electrical installation of device corresponds to requirements RDT 25.106‐88 (national standard). 

 

3.22.  The device corresponds to requirements of electromagnetic compatibility IEC 60601‐1:2005‐12. 

 

3.23.  Metallic  and  nonmetallic  mineral  coatings  correspond  to  GOST  9.303‐84  (national  standard)  for  the  group  of 

operating conditions 1 GOST 15150‐69 (national standard). 

 

3.24.  Paint  and varnish  coats  ‐ GOST 9.032‐74  (national  standard)  for  the operating  conditions  correspond  to UKhL 4 

GOST 9.104‐79 (national standard). Coatings of external surfaces ‐ not lower than the III class by GOST 9.032‐74 (national 

standard). 

 

3.25.  The level of the acoustic power at a distance of 1.0 m of the outline of device does not exceed 65 dBA. 

 

3.26.  The  temperature of  the external parts of  the device, which do not have contact with  the patient, accessible  for 

touch when in normal operation, does not exceed 55ºC. 

 

3.27.  Class of device concerning the potential risk of application ‐ 2a by Annex IX of the Directive 93/42/EEC.  

High Technology Medical Devices 7 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

4. DELIVERY SET  

4.1.   The delivery set of device corresponds to the indications in table 2.   Таble 2  

№  

Designation  Designation of the document  Quantity, pieces

1.    2.  3.  4.  5.  6.  7.  8.  9.  

10.    

11.  

12. 

Air‐plasma scalpel‐coagulator‐stimulator “PLASON” 

Set of the equipment  Coagulator  Destructor  Generator‐ stimulator  Distancer   Gas‐tube  Cold nitric oxide tip 130 mm  Key 

 Pedal  Transporting bag  

Operational documentation  

Passport  Description of medical technology on CD 

КРЛД 38642.001.    

КРЛД 38642.001.07.  

КРЛД 38642.001.08.  

КРЛД 38642.001.09.  

КРЛД 38642.001.11.  

КРЛД 38642.001.12.  

КРЛД 38642.001.14.  

КРЛД 38642.001.15.  

 КРЛД 38642.001.16 .  

КРЛД 38642.001 ПС 

1   2  1  2  1  1  1  1  1  1    1  1 

  5. DEVICE, OPERATING PRINCIPLE AND THE REGIMES OF WORK  

 5.1.   Construction of device.    The  device  consists  of  the  service  block  (SB),  electro‐hydro  pneumatic  (EHP)  supply,  interchangeable  manipulators, silicone tube with a metallic tip and footswitch. SB is structurally executed in a rectangular metal housing, which consists of two parts ‐ lower and upper. The lower part of the housing consists of horizontal base, front and rear panels and serves for the installation of all systems, which ensure the work of manipulator.    Front panel includes:  ‐ illuminating “ON/OFF” switch; ‐ button “START” and “STOP” with illumination for starting and turning off the manipulator; ‐ button “MIN”, “NORM”, “MAX” with illumination for the control of the air flow rate; ‐ the hinged connection of the EHP supply; ‐ thread carbine for the connection of silicone tube with a metallic tip; ‐ jack opening for the connection of the footswitch (pedal).  Two brackets  for  the  coil  of  electric  cable  are  located on  the  rear panel, where also a  special  key  for  changing of  the manipulators is fixed. The upper part of the housing is the removable cover of SB, on its top two containers with threaded 

High Technology Medical Devices 8 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

caps  are  located  at  both  ends,  used  for  storage  of  manipulators.  The  central  container  also  serves  as  a  handle  for transporting of  the device. On  the upper panel  there  is  also  the  tube  for  inserting of  the manipulators  for  the built‐in cooler of air‐plasma flow, closed with threaded cap with inscription “GAS COOLER”, and also the filler neck of reservoir of the cooling liquid, closed with threaded cap with inscription “COOLING LIQUID”  1.3 m long EHP supply is attached to the front panel of SB and is executed in the form of plastic bellows, on end of which the  connection  joint  for manipulators  is  located.  Inside  the  bellows  there  are  flexible  tubes,  intended  for  supply  and outlet of the manipulator cooling  liquid, supply of working gas ‐ atmospheric air to the manipulator and electric power supply wires of the manipulator.  The threaded carbine located on the front panel is used for connecting of the 1.3 m long silicone tube with the detachable metallic tip for the output of the cooled NOx‐ containing gas flow (NO ‐CGF).  5.2.  Description of device and modes of its operation.    Service block contains all  the  systems ensuring  the work of  the device:  the  supply of atmospheric air,  cooling, electric power supply, automation, control and sound communication.   The  basic  element  of  device  is  the  medical  manipulator,  connected  to  the  service  block  with  the  flexible  electro‐hydropneumatic (EHP) supply tube.   Manipulator is fixed on EHP supply by an adapter nut and can be easily changed for a new manipulator of the same type or a manipulator with other functionality.   The device is supplied with the manipulators of three types ‐ coagulator, destructor and generator – stimulator.    The manipulators of all three types are the generators of the air plasma flow, made according to the linear three‐electrode diagram and which differ from each other in terms of the construction of outlet.    Coagulator  (dark‐blue,  red marking)  has  outlet  channel with  a  diameter  of  1.2  (blue)  and 4.0  (red) mm.  The work  of coagulator with any selected air flow rate it is formed the vividly luminous plasma flow with the outlet temperature from the channel of 3000‐4000°C and small gas‐dynamic pressure.  Destructor (yellow marking) has outlet channel with a diameter of 0.7 mm. The destructor forms more localized plasma stream (compared to the coagulator) with the temperature of 2500‐3000 °C and an increased gas‐dynamic pressure.  Generator‐ stimulator (green marking) has outlet channel with a diameter of 1.8 mm. The generator‐ stimulator by any selected air flow rate forms a weakly  luminous plasma flow with the outlet temperature of 700‐800°C and a small gas‐dynamic pressure. All manipulators are not only the sources of air plasma, but also sources of the nitric oxide NO, which is formed in the air plasma as a result of the plasma‐chemical reactions.   Available  regimes  of  the  work  of  device  are  determined  by  the  characteristics  of  gas  flow,  which  escapes  from  the manipulator, basic parameters of which are its temperature and content of the nitric oxide in it.  For the scalpel regime of action on the cloth the determining parameter is the gas flow temperature, values of which on its axis for different types of manipulators are represented on Fig. 3.  For  the  therapeutic action  (NOx‐therapy)  the content of  the nitric oxide  is  the determining parameter of gas  flow. The nature of a change in the NO gas flow content on its axis is represented in Fig. 3.    Depending on the arrangement of manipulators, fixed on the EHP supply tube, device allows to realize two groups of the operation modes (Fig. 4).  The first group of regimes ‐ regimes with the free expiration from the manipulators of air‐ plasma flow, in this case device is in state A on Fig. 4, and manipulator is kept in the hand of user.  The second group of regimes are the regimes of action on the biological  tissue by the cooled (room temperature) NO‐ containing gas flow. To achieve this, any manipulator needs to be put into the tube of the built‐in cooler. The supply of 

High Technology Medical Devices 9 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

NO‐CGF to the biological tissues is achieved through the silicone tube with the metallic tip with a diameter of the outlet of 0.7 mm fixed at its end (state of device B in Fig. 4).  For checking of the correct functioning the device is equipped with a system of sound communication. It starts the first sound  signal  during  the  establishment  of  the  work  regime  and  shorter  sound  signals  after  every  60  seconds  of  the manipulator work.   For maintaining  the stable  thermal condition of manipulators  the device  is  supplied with a closed  liquid system, which ensures their effective cooling and rejection of the heat into the surrounding space through two radiators and fans on the rear panel of the device. 

Temperature  and  content  of  the nitrogen oxide  are  the basic  physical chemistry  parameters  of  the  NO‐containing gas flow. The temperature of  expiration  is  ~4000  °C  for  the destructor  and  ~1000  °C  for  the coagulator (Fig. 3).  It should be noted that  the molecule of the nitrogen oxide possesses active oxidation  potential  and  in  the  air medium  enters  into  reaction  with molecular  oxygen,  forming  a chemically  resistant  compound  ‐  the nitrogen  dioxide  (NO2),  which  will always be present as a result of this in the NO‐containing gas flow.   The  application  of  the  standardized generator  part  of  all  manipulators makes it possible to reach flows with identical  axial  content  of  nitrogen oxide,  what  is  also  shown  in  Fig.  3. Range of the obtained concentrations of NO  is  very wide  ‐  from 2500 ppm to  0.  This makes  it  possible  to  select those  parameters,  which  the  doctor considers  to  be  most  acceptable  for specific purposes.   The  analysis  of  dependence  of temperature  and  content  of  the nitrogen oxide on the axis of gas flow shows  that  the  device  allows  an enormous field of possibilities on the combining  of  the  thermo‐physical and  biochemical  factors  of  action  on the biological tissues and objects.    

 Fig.3: The temperature of flow is rapidly reduced with the distancing from  the outlet of manipulator and at a distance 100 mm does not exceed 100°C.  The content of nitrogen oxides on the axis of gas flow is represented in the  second graph.  

High Technology Medical Devices 10 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

  5.2.1 Coagulation of the wound surfaces  Coagulation  is  achieved  by  bringing the high‐temperature air‐plasma flow (HPF)  directly  to  the wound  surface. The temperature of flow in the region of contact must be not less than 2000 °C,  therefore,  the  distance  from  the outlet  of  coagulator  to  the  region of action  must  be  not  more  than  25 mm.  In  the  first  step  (Fig. 5A) during application  of HPF with  temperature higher  than  2000°C  to  the  biological tissue, its heating to the boiling point of  woven  liquid  occurs.  Compact layer  of  necrosis  (CLN)  is  formed, which  consists  of  the  degradation products  of  protein  connections, partially damaged and nonviable, but with  their  structure  cells  preserved (55°C), and the zone of paranecrosis ‐ with  partially  reversible  changes  in the  cells.  Subsequently  the  zone  of paranecrosis  becomes  the  source  of regeneration.    

 Fig. 5: Coagulation of the wound surfaces  The further increase of the time of action (Fig. 5B) leads to strengthening of the process of boiling and evaporating the woven liquid. The spongy layer of necrosis (SLN) is formed ‐ the porous, elastic and airtight structure, which consists of the dehydrated protein and fatty connections. Clearly expressed physical boundary of the evaporation of liquid appears in the cloth, which divides the spongy (SLN) and compact (CLN) layers of necrosis (100°C). Upper temperature boundary of the spongy  layer of necrosis coincides with  the  temperature of  the beginning of  the  thermal decomposition of protein 

Work regime A

Work regime B

Fig. 4: Two operational regimes of PLASON device work 

High Technology Medical Devices 11 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

and fatty compounds into the simplest chemical elements (550 °C, Fig. 5C). Above SLN the carbonized (carbon) layer of necrosis (CaLN) is formed, which is the burnt and charred cloth. Qualitative hem‐aero‐lymph‐cholestasis is determined by the positive dynamics of shaping of the compact and spongy layers of necrosis or by the low rate of shaping of the carbon layer of necrosis. The thermo‐physical and geometric parameters of the plasma flow of coagulator are optimized so that reliable  hem‐aero‐lymph‐cholestasis  is  reached on  any  biological  cloths  and  at  any  speeds  of  counter  blood  flow.  The significant (about 200 m/s) discharge velocity of hot gas from the outlet of coagulator creates condition on the surface of the coagulated section of  the cloth, with which coagulation  layers are  formed directly  from cloth  itself, but not of  the counter‐flows of the liquid, for example, of the blood or lymph entering.  A change  in the time of action and distance from the outlet of coagulator  to the region of coagulation  is  the universal method of  the dosage of  the energy necessary  for  the coagulation. For obtaining  the  reliable coagulation  the distance from the outlet of manipulator lies within the limits from 2 to 25 mm.   

ATTENTION! 

In practice hem‐aero‐Iymph‐cholestasis can be achieved also without completeshaping of the entire region of thermal changes. Coagulation effect begins, for example, and without the formation of the carbon layer of necrosis. 

  5.2.2 Sparing coagulation of the wound surfaces  In a number of cases, when the intensive entering of liquid (for example, the blood or of lymph) from the deep layers of cloth to the surface is absent, it is necessary  to  conduct  partial  drying  surface  in  the  regime  of  the  sparing coagulation for the purpose of the final evaporation of pathologic liquids (for example, exudate) and achievement of the effect of sterilization.  For  achievement of  the  effect  of  partial  drying  and  tasks  formulated  above there is no need for using the entire temperature potential of air‐plasma flow (HPF). Sufficient to ensure in the place of the contact of plasma jet with the cloth temperature  in the range from 1000 °C to 2000 °C. The flow charts of the sparing coagulation are shown in the figure.   The sparing coagulation can be carried out by the coagulator (Fig. 6A), as well as  by  stimulator‐coagulator  (Fig.  6B).  With  the  use  of  the  coagulator  the distance from the outlet to the region of coagulation must be not less than 25 mm. Visually this approximately corresponds to the end of the luminous part of the plasma torch of coagulator. The decrease of distance from the outlet to the region of coagulation will lead to excess temperature limit to 2000 °C.  The  thermo‐physical  processes,  proceeding  during  the  sparing  coagulation, are  in  many  respects  similar  to  the  processes,  proceeding  during  the traditional coagulation, but in this case low temperature potential flow does not make possible to be formed for the complete region of thermal changes. Process is limited to the formation of the small zone of spongy necrosis (SLN), zone of compact necrosis (CLN) and zone of paranecrosis. As a result of this action an elastic film of  lusterless nuance with a thickness not of more than 200 mkm is formed on the surface of cloth, which also is the purpose of the sparing coagulation.  

Fig. 6: Sparing coagulation  When conducting the sparing coagulation by stimulator‐coagulator the distance from the manipulator outlet to the cloth is minimal (Fig. 6B), since the temperature of the escaping flow does not exceed 1000 °C. This way,  if  it  is necessary to realize the sparing coagulations on large surface, it is possible to use a coagulator, and if it is necessary to coagulate the bound spaces ‐ stimulator‐coagulator.  

ATTENTION! For obtaining an uniform coagulative membrane it is necessary to constantly change the angle of the tilting of plasma flow about the wound surface according to its relief.

High Technology Medical Devices 12 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

5.2.3 Technology of conducting of the sessions of NOx‐therapy by coagulator and by destructor  Any manipulator for the action of exogenous nitrogen oxide on the biological  subjects  allows  using  of  various  generator  parts. Temperature factor is a fundamental physical limitation for applying of  the  NOx‐therapy  technology  as  a  whole.  The  optimal temperature of action is determined by the following conditions:   1) the absence of possible thermal damages of cellular structures;   2) by the absence of the sensations of thermal discomfort in patient.   Practice has shown that the optimal temperature of the flow is 40 ± 10 °C.   This  is  less  than  the  temperature  of  the  beginning  of  protein connections  denaturing,  and  also  creates  a  pleasant  sensation  of blowing warm air at the patient.  Minimum  distance  from  the  outlet  to  the  object,  with which  it  is possible to carry out the sessions of the NOx‐therapy by coagulator or  by  destructor,  is  150  mm.  The  flow  chart  of  conducting  the sessions of NOx‐therapy   is shown in the Fig. 7. The NO‐containing gas flow (NO‐CGF) is directed to surface desirably at right angle.   

Fig. 7: NOx‐therapy  by coagulator and by destructor  Bright  luminescent  spot  on  the  surface  (optical  and geometric  projection of  the outlet  of manipulator  on  the  surface) helps the user to determine the center of flow and to direct it in the necessary direction. In this case, a stationary circular region with the increased content of the nitrogen oxide is formed in the region of the contact of flow with the surface. Maximum concentration of NO is observed on the axis of gas flow and it falls to the periphery smoothly. Concentration of NO and NO2 on the centerline of gas flow is determined by the physical chemistry flow parameters (temperature and the content of oxides of nitrogen ‐ fundamental characteristics of the air‐plasma flow).  The effective scope diameter of the workable surface NO ‐ CGF can be estimated according to the simple formula:  

4

mm object, ther tomanipulato ofoutlet thefrom Distance diameter Scope  

 If the geometric dimensions of the workable region or object are greater than the scope diameter, working is conducted by scanning, linear or circular motions with a movement speed of approximately 0.5‐1.5 cm/s. 

   

ATTENTION! The diameter of luminescent spot is lower than the diameter of the seizure of the workable surface. Luminescent spot only allows to correctly direct the center of the air flow, which contains the nitrogen oxide. 

ATTENTION! When conducting NOx‐therapy  sessions by coagulator or by destructor it is necessary to maximally attentively check the distance from the outlet of manipulator to the region of working to avoid the possibility of causing thermal damages. 

  

    

High Technology Medical Devices 13 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

5.2.4 Technology of conducting the sessions of NOx‐therapy  by the stimulator‐coagulator   Stimulator‐coagulator  as  a  device with  a  cooler  part  connected  to  the  generator generates  gas  flow,  which  contains  nitrogen  oxide,  with  temperature  potential smaller  than  coagulator  or  destructor.  When  conducting  the  sessions  of  NOx‐therapy  ,  this makes  it possible  to bring  the manipulator nearer  to  the biological subject  and  as  a  result,  create  higher  concentration  of  NO  in  the  near‐surface layers.  The  flow  chart  of  conducting  the  sessions  of NOx‐therapy    by  stimulator‐coagulator is shown in the Fig. 8A. According to the dependence of the content of the nitrogen oxide on  the axis of  flow, minimum distance  from  the outlet  to  the surface,  where  the  temperature  will  not  exceed  40  ±  10  °C,  is  about  80  mm (temperature  and  the  content  of  oxides  of  nitrogen  are  the  fundamental characteristics of air‐plasma flows).  Directed  desirably  in  the  right  angle  to  the  workable  object,  flow  will  smoothly spread  in  parallel  to  surface,  creating  axisymmetrical  region  with  an  increased content of nitric oxide.   The center of luminescent spot indicates the user, where the flow is directed at the given moment. The effective scope diameter of the workable surface is defined as ¼  of  distance  from  the  outlet  of  manipulator  to  the  surface.  For  averting  the possibility of causing thermal damages the manipulators are finished with a thread for  installation  of  the  detachable  thermal  spacer,  whose  application  makes  it possible to fix the physical chemistry parameters of flow (Fig. 8B).    Thermal  spacer  is  either  manufactured  as  a  hollow  cylinder  with  wide  lateral windows or a cylindrical base with a narrow distancing needle.  Its  length assures that the temperature of the NO‐containing gas flow (NO‐CGF) on the end of spacer reaches 40 ± 10 °C, in this case the concentration of the nitrogen oxide will have a value of ~750 ppm. The installation of thermal spacer does not introduce changes in  geometric  parameters  of NO‐CGF.  The effective  scope  diameter  is ¼  from  the distance of  the maximal possible approximation of stimulator‐coagulator which  is ~20 mm. 

 Fig. 8: NOx‐therapy  by stimulator‐coagulator  

ATTENTION! During continuous operation of manipulator with thermal spacer attached, the spacer may become hot. Avoid the contact of the spacer with biological tissues. 

 

 5.2.5 Technology of conducting NOx‐therapy sessions by narrowly‐directed cooled NO‐containing gas flow.  The application of additional cooler of  labyrinth  type allows  to completely cool  the air‐plasma flow (APF) to ambient temperature. To obtain a narrowly‐directed NO‐containing gas  flow (NO‐CGF) of ambient  temperature  it  is necessary  to establish any manipulator into the cooler tube and start the device work.  The  cooled  flow will  be  delivered  to  the  object  through  a  flexible  silicone  tube  and  a metallic tip, a thin‐walled tube with a diameter of 4 mm from biologically inert materials. Localization  and  focusing  is  achieved by  the  calibration of  the  tip  outlet.  The optimum diameter of the outlet for NOx‐therapy is less than 1 mm.   The  application  of  narrowly‐directed  cooled  NO‐CGF  allows  acting  on  the  region  of pathology  from any distance  to 250 mm. The cooling of NO‐containing gas  flow allows using it when working in cavities in different variants. 

Fig. 9: Cooled NO usage 

High Technology Medical Devices 14 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

Variant 1.   If  cavity  has  a  form  of  pocket  and  the  opening,  which  connects  it  with  the surrounding space, is bigger than the diameter of the metallic tip, then the tip is directly  introduced  into  the  opening.  Air  in  the  cavity  (pocket)  is  displaced  by forced  NO‐CGF.  The  atmosphere  with  the  increased  content  of  the  nitrogen oxide is created after a certain time in the cavity.  

ATTENTION! 

When using narrowly‐directed NO‐CGF, choose procedure duration with attention since at small distance of the tip to the surface of processed object, high concentration of NO is formed. 

 Variant 2.   Connection of metallic tip to the already existing or specially created vent  lines allows  processing  of  almost  any  internal  cavities  by  nitrogen  oxide.  Here  the metallic tip  is air tightly connected to the vent  line (tube of supply) fixed in the cavity.  NO‐CGF  is  then  introduced  into  the  cavity,  and  the  gas  from  the  cavity flows out through another tube, which connects the cavity with the surrounding space (tube of discharge).     

Fig. 10: Variants of cooled NO usage  

ATTENTION! When working in the cavities through the opening it is necessary that the metallic tip would enter into the opening with a clearance for drainage of gas. Otherwise appearance of excess pressure in the cavity is possible. 

 The version of working  in cavities with  the aid of  the  two‐clearance drainages  (Fig. 10.1) or of  the coaxial  (tube  in  the tube) version of NO‐CGF introduction into the cavity (Fig. 10.2), and also the supply of the nitrogen oxide to the region of pathology through the puncture needles and the biopsy channels of endoscopic instruments are possible.  

  

Fig. 11: Various forms of cooled NO‐CGF introduction tips  

ATTENTION! During the supplying of nitrogen oxide into any cavity it is necessary to ensure its drainage for exiting of the "waste" gas into the surrounding space. 

High Technology Medical Devices 15 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

5.2.6 Destruction (dissection) 

 Basic difference in the manipulator‐destructor from the coagulator is the diameter of  outlet  (coagulator  1.2  or  4 mm,  destructor  0.7 mm).  This  allows  to  aggravate and concentrate the air‐plasma flow (APF) energy and to increase the speed of its expiration.  In  working  of  living  biological  tissues  by  the  concentrated  and aggravated  APF,  the  carbon  layer  of  necrosis  is  heated  to  the  temperature  of ~800°C  and  it  begins  to  be  decomposed  into  small  particles  ‐  to  sublimate,  and high‐speed flow exhausts the decay products of CoLN from the zone of action.   With  this  dynamics  the  carbon  layer  of  necrosis  ceases  to  be  the  natural  heat shield as  in  the case of  coagulator, and entire  structure of  the  region of  thermal changes is moved into the depths of the cloth. As a result, a zone of destructured cloths  is  formed,  surrounded  by  the  coagulation  layer,  which  ensures  reliable hemostasis.  

Fig. 12: Destruction (dissection)  Effective distance from the outlet of manipulator must be the smallest possible. This allows obtain qualitative section with causing smallest possible damages to the adjacent cloths and organs.  

ATTENTION! During destruction or dissection of the hollow organ walls the danger of damage of their internal surfaces as a result of the penetration through incision of hot gas can arise. 

  5.2.7 Technology of obtaining antiseptic solution on the basis of hydrogen peroxide and nitrogen oxide (solution of Aleksandrov M.T.) 

To produce an antiseptic  solution on  the basis of hydrogen peroxide and nitrogen oxide  it  is  necessary  to  dissolve  the  specific  quantity  of  NO  in  3%  solution  of hydrogen peroxide. The low solubility of the nitrogen oxide in the aqueous solutions does  not  make  it  possible  to  prepare  this  solution  without  the  application  of technology  of  bubbling.  Flow  chart  is  shown  in  the  Fig.  13.  A  transparent  tank  is filled with fixed quantity of 3% solution of hydrogen peroxide. The metallic tip of the supply  of  the  NO‐containing  gas  flow  (NO‐CGF)  is  introduced  through  the  tightly closed cover in such way, that the outlet of the tip is located below the level of liquid in the tank. Additionally, the cover has a vent tube, connecting the empty space of the tank with the ambient space. With the starting of the device "PLASON", NO‐CGF enters  directly  into  the  3%  solution  of  hydrogen peroxide  in  the  form of  bubbles, saturating peroxide by the nitrogen oxide. The optimum time for the preparation of solution depends on the volume of the prepared solution and is determined by the formula: 

t = 0,06∙ V 

where the V is volume of the solution in ml, t ‐ time of bubbling in min.  

Fig. 13: Technology of preparing                of antiseptic solution  The obtained solution can be used in the therapeutic process immediately or in the course of 20 days if stored in a cool, place protected from the light.  

ATTENTION! Obtaining solution with the described technology must be conducted in the exhaust hood or in a room equipped with exhaust ventilation. 

     

High Technology Medical Devices 16 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

5.3 Control systems of PLASON  For  the  control  of  work  of  manipulators  and  the  device  the  device  is  supplied  with  the  system  of  the  sound communication, which forms the appropriate sound signals with the presence of disturbances or malfunctions.   Table 2: Control systems of PLASON  

Name of the system 

Functions  Effect  Audible sidnal 

Visual signal

Temperature control system 

Coolant Temperature Control (CTC) 

1. When coolant temperature reaches 32 ± 2 °C: ‐ second (intensive) cooling stage is initiated 

 

2. When the coolant temperature reaches 50 ± 2 °C: a) the manipulator turns off; b) blocking of start until the coolant 

temperature is lowered to 45 ± 2 °C.

 

Coolant consumption control system  

Quality control of flow rate and amount of coolant 

1. When the flow of coolant is reducedbelow 0.8 ± 0.1 liters / minute: a) the manipulator turns off; b) blocking of start until the coolant 

flow rate is restored 

Continuous beep signal 

Flashing red LED on the control panel (inside the device, only for technicians). 

2. Violation the continuity of coolant flow (bubbles): a) the manipulator turns off; b) blocking of start until the coolant 

flow rate is restored 

Continuous beep signal 

Flashing red LED on the control panel (inside the device, only for technicians). 

Level control system of coolant  

Control the coolant level 

1. The volume of coolant decreases by 0.25 liters: a) the message by beep signal. 

Intermittent beep signal 

Flashing green LED on the control panel (inside the device, only for technicians). 

  5.4 Remote Control of the Device  For the remote control of the starting and stopping of work with manipulators the device is equipped with a pedal. In order to use the pedal for operating the device remotely, connect it to the pedal connection on the front panel.   When  the  pedal  is  pressed  down,  the manipulator  starts  working.  Once  the pedal is released, the manipulator stops working immediately.  

Fig. 14: Pedal for PLASON remote control  

 6. INDICATIONS ON SECURITY MEASURES  

6.1.   Use the device only after thorough study and with observance of requirements, presented in this User´s guide. 

6.2.  Work  only  with  original  manipulators.  When  a  failure  is  detected  it  is  necessary  to  send  the  device  to  the  EU representative (ONKOCET Ltd.) for repair and testing. 

6.3.   The operation of device in a room, where combustible anesthetics have been applied, is forbidden. 

6.4.   Connection to the electrical network must be through the protection from short circuit (automatic switches, fuses, safety devices).  

6.5.   It is forbidden to use the device with the cover of service block removed. 

6.6.   Before conducting of surgical interventions with the aid of the device the surgeon is obliged to become acquainted with the action of plasma flow on soft tissues by trial actions on the damp meat. 

High Technology Medical Devices 17 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

6.7.   Special attention is required when switching manipulators to set them up on the manipulator tube correctly. 

6.8.  When working with the device it is strictly forbidden to subject any objects under the plasma flow, other than the treated surface of the patient’s body. Don’t screw any objects on the tip of the manipulators. 

6.9.   Continuous operation of the device is possible only in the premises, supplied with exhaust ventilation. 

6.10. Use only prescribed cooling fluid to fill into the cooling system of the device. Using other substances can damage the device and will result in void warranty. 

6.11. Use only the supplied footswitch to operate the device remotely. 

6.12.  Avoid  long  or  very  close‐distance  exposure  of  human  tissues  to  protect  from  burns.  Follow  feedback  from  the patient in order to state the comfortable application distance or use the provided distance.  

6.13. It is recommended to use Plason in rooms with temperature up to 25 °C and humidity up to 80 percent.  

  

7. ORDER OF INSTALLATION AND THE PUTTING INTO OPERATION  

7.1.   Extract the device from the transport package. 

7.2.   Verify the completeness of device. 

7.3.   Place  the device on a horizontal  surface. Arrangement must ensure  free access  to  the control elements and their convenient observation. 

7.4.   Unscrew and remove the red transport plug from the bottom of the device. 

7.5. Connect the foot pedal to the device. 

  

8. PREPARATION OF DEVICE FOR OPERATION  

8.1.   Device is delivered to the user in completely assembled form with a manipulator set up on the end of the flexible EHP supply tube. Remove the red screw plug from the bottom of the device prior to its operation. The red screw is used  for  locking of  the overflow discharge  tube  for  transport.  The preparation of  device  for  the work  consists  in checking of its fitness for work with different manipulators with different operating modes on the air flow rate.  

8.2.  Make sure that the network supply rosette has a grounding contact. In the case of its absence ensure the installation of network supply rosette with a grounded contact. 

8.3.   Set the “ON/OFF” switch into the initial state ‐ position “0”. 

8.4.   Plug the network cable of the device in a socket with the supply voltage 230 V of 50 Hz with the permissible current not less than 6 A. 

8.5.   Start the device via the “on/off” switch from the position “0” to the position “I”. When done, the “on/off” switch, the “STOP” button and one of the control buttons of airflow “MIN”, “NORM” or “MAX” are illuminated; pump and cooling‐system fans of device begin to work. 

8.6.   Establish the maximum air flow rate by pressing of the “MAX” button. 

8.7.   Start  the work  of  the  coagulator  by  pressing  the  button  “START”.  Normal  operation  of  device  and  coagulator  is shown by glowing “START” button, by a sound signal and by outlet of steady plasma torch from the coagulator. 

8.8.   Establish the minimum air  flow rate by pressing the “MIN” button. The air  flow rate through the coagulator must decrease. 

8.9.   Turn off the manipulator by pressing of the “STOP” button. 

8.10. Turn off the device via the “on/off” switch from the position “I” to the position “0”. 

High Technology Medical Devices 18 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

8.11. Remove the coagulator  from the  tip of  the EHP supply, after unscrewing of  the adapter nut, which connects  the manipulator and the EHP supply tube. If necessary, use the special tool, fixed on the rear wall of device. 

8.12. Establish a destructor (yellow marking) on the EHP supply tube by screwing the adapter nut thoroughly. Before the installation of manipulator to the EHP tube check for moisture tracks presence and wipe them with a gauze napkin. 

8.13. For functional test of the device and destructor repeat the actions in sequence 8.5‐8.10. 

8.14. Establish a stimulator (green marking) on the EHP supply tube, after carrying out actions in sequence 8.11, 8.12. 

8.15. For the functional test of the device and stimulator repeat the actions in sequence 8.5‐8.10. 

8.16. For the functional test of the device in the regime of remote control start with any manipulator mounted on the EHP supply  tube by pressing of  the pedal;  “START” button glowing,  the presence of  sound signal and outlet of  steady plasma torch from the manipulator testify about the normal operation of the device. 

8.17. Release the pedal ‐ the flow from the manipulator must be stopped and device should go into the state as in 8.5. 

8.18. Disinfection of the external surfaces of device and manipulators can be done by five‐step working, each of whom consisting of two rubbings of external surfaces with a gauze napkin, moistened by 3% solution of hydrogen peroxide by  GOST  177‐88  (national  standard)  with  the  addition  of  the  cleaning  agent  according  to  the  GOST  25644‐96 (national standard) or by 1% chloramine solution, according to technical specifications TU 9392‐031‐00203306‐2003. 

8.19. Device is ready for operation. 

8.20. During the intra‐operating application of the device a sterile sleeve should be put on the EHP supply from cotton cloth or the EHP supply should be wrapped into a sterile sheet (towel).

 

 The sterilization of manipulators should be carried out by chemical gas methods by ethylene oxide or by formaldehyde. The  sterilization of manipulators  in UV or  gas‐formalin  chamber,  and also by  rubbing with different  solutions with  the subsequent covering up of manipulator into the sterile napkin is possible.  

ATTENTION!  Moistening of manipulators is not allowed!

  

 

   9. OPERATIONAL PROCEDURE  

9.1.   Set “on/off” switch into the initial state ‐ position “0”. 

9.2.   Plug the network cable of the device in a socket with the supply voltage 230 V of 50 Hz with the permissible current not less than 6 A. 

9.3.   Select and connect the required manipulator to the EHP supply tube. 

9.4.   Start the device via the “on/off” switch from the position “0” to the position “I”. 

9.5.   Select the requested air flow rate by pressing the button “MIN”, “NORM” or “MAX”. 

9.6.   Start the manipulator by pressing the “START” button or by pressing of the foot pedal. 

9.7.   Carry out the necessary surgical or therapeutic procedures. 

9.8.   Turn off the manipulator by pressing the “STOP” button or by releasing of the foot pedal. 

9.9.   Turn off the device by setting the “on/off” switch to position “0”. 

9.10. Take out the fork of the electric supply cable from the electric plug.  

9.11. During the interruptions between therapeutic actions (procedures) it is recommended only to turn off the manipulator by pressing the “STOP” button and not to turn off the whole device by the “on/off” switch until the end of the outlined cycle of procedures. 

9.12. To transfer the device into the working regime B (cooled NO‐CGF regime), the threaded “GAS COOLER” cap, which closes the tube of the built‐in cooler on the upper panel of device, should be removed. Blue manipulator should be inserted into the tube to the maximum possible depth (don´t use excess force). It is recommended to fix the 

High Technology Medical Devices 19 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

threaded cover to its holder, located on the rear bracket of the knob of the device, opposite the built‐in gas cooler. 

9.13. Transfer from work regime A into regime B and vice versa must be done only when the manipulator is turned off. 

 Table of possible malfunctions  

Designation of malfunction, external manifestation and additional signs 

Possible reasons  Method of the elimination 

1. When switching on the device by the “on/off” button the cooling liquid leaks from the manipulator outlet. 

Non‐hermetic state of the manipulator cooling system. 

Replace the manipulator. 

 

2. The manipulator does not start when the “START” button is pressed. The “START” button is not illuminated. 

1. Manipulator reached the end of its lifetime. 

2. Malfunction of electric power supply system. 

1. Replace the manipulator2. Contact the producer or 

official representative. 

3. When switching on the device by the “on/off” button or when the manipulator is started, an intermittent sound signal is occurring. 

Insufficient amount of liquid in the cooling system. 

Add the cooling liquid. 

 

4. When switching on the device by the “on/off” button a continuous sound signal is present; it is not possible to start the manipulator by the “START” button. 

The malfunction of cooling system (there is an unsufficient flow rate of the cooling fluid). 

Contact the producer or the official representative. 

 

10. THE MAINTENANCE  

10.1.    For guaranteeing the reliable work of device to conduct timely maintenance.  

10.2.    To follow the cleanliness of device and manipulators. External surfaces of device and manipulators to      disinfect on p. 8.18. present passport.  

10.3.    To conduct daily maintenance with the regular work with the device.  

10.4.    Monthly maintenance by regular operation of device is carried out by medical staff that exploits the device.  

10.5.    Annual servicing of device with the regular work with it to carry out not less than one time per year.  

10.5.1.  Plug the device out from the electric network, remove the cover of service block.  

10.5.2.  Clean the dust from internal spaces of device, using vacuum cleaner.  

10.5.3.  Look around thoroughly through the installation, paying attention to the presence of clearances between the    components, the exterior view of the elements of installation, the integrity of wire insulation, fastening units.   

10.5.4.  Periodic annual maintenance is carried out by the specially prepared operating personnel.  

  11. WARNINGS AND SAFETY NOTICES 

11.1. Modification of this equipment is not allowed. Hazards can result from unauthorized modification! 

11.2. Putting fingers into openings of the device intended for holding of manipulator is prohibited. 11.3  Using of PLASON in an oxygen rich environment is forbidden. An unacceptable risk of fire is considered to exist in an 

oxygen rich environment when a source of ignition (plasma) is in contact with ignitable material and there is no means that would limit the spread of fire. 

11.4 It is forbidden to use PLASON with flammable anesthetics. 11.5 It is forbidden to use PLASON in conjunction with flammable agents. 

11.6 It is forbidden to use PLASON when the operating staff is under the influence of alcohol, and medicines negatively affecting performance and attention. 

High Technology Medical Devices 20 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

12. LABELING  

Front panel 

 

Back panel 

 

 

 

 

   REMOVE THE RED PLUG ON THE BOTTOM OF THE DEVICE BEFORE FIRST USE! 

 

 

 

The instructions for use describe the nature of the HAZARD and the precautions for avoiding it or minimizing the associated RISK.

 

 

Warning ! Keep the end of manipulator in

minimal distance 20 cm away from the body of patient or operator

before pressing the START button.

Attention! Screw in the red plug to the outlet

channel at the bottom of unit before transportation.

Attention! Please remove the red plug on the

bottom of the unit before use.

Warning ! Don’t put fingers into any of the

openings of the device.

High Technology Medical Devices 21 | Sreznevského 17, 831 03 Bratislava, Slovakia Tel: +421 2 446 409 77; Mobile: +421 948 610 228, +421 903 114 944; Mail: [email protected]; Web: www.onkocet.eu 

13. DISPOSAL OF PRODUCT AND PACKAGING 

DISPOSAL OF PACKAGING - packaging components (cardboard, plastics etc.) are all classified as solid waste and can be therefore easily disposed of with the use of recycling processes.

Before disposal is always advisable that you check the validity of the relevant provisions in the place of installation. PLEASE MAKE proper disposal!

PRODUCT DISPOSAL – PLASON is made of various materials. Most of them (plastic, metal, electric conductors, etc.) may be handed over in recycling sites and can be recycled. However, other components (electric plates, etc.) may contain hazardous substances. Therefore the following components should be submitted in the respective centers or the European representant of the producer, where the qualified ensure their disposal. Before disposal is always advisable that you check the validity of the relevant provisions of the disposal site. PLEASE MAKE proper disposal!

14. STORAGE AND TRANSPORTATION OF PLASON The PLASON device should be transported and stored in the following conditions: ambient temperature from +5C till + 40C; relative humidity less than 80%; avoid aggressive vapor in the air; avoid dusty spaces; avoid direct sunshine; avoid strong electric and magnetic fields; avoid moist environment. Lifetime of the PLASON device, if the conditions of usage, storage and transportation meet the requirements of present User’s manual, is 7 years. Warranty repairs are made by: Manufacturer:

CJSC “Center BMSTU” 2nd Baumanskaya, 5, bld. 1, 105005 Moscow, Russia

or the authorized service centres: Authorized representative in European Union:

Onkocet Ltd. 4 Kutuzova str., 90201 Pezinok, Slovakia, tel.: 421 244 64 09 77

e-mail: [email protected]

All valid permissions and certificates are placed on the homepage of European Representative www.onkocet.eu

 Last revision: November 1st 2014