High-Level Dosimetry systems used at CERN

Markus Fuerstner1, Doris Forkel-Wirth1, Helmut Vincke1, Sabine Mayer1, Isabel Brunner1, Idelette Floret2

1 CERN, CH-1211 Geneva 23, Switzerland

2 EIG, Ecole d’Ingénieurs de Genève

Page 2: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Contents

Contents

• Overview of HLD systems

• HLD systems used at CERN

• Further development of the systems

• HLD measurement campaigns in mixed radiation fields

• Conclusion

Page 3: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Types of dosimeters

• Hydrogen Pressure Dosimeter (HPD)• LiF Crystal• Thermo Luminescent Dosimeter (TLD) • Dye Film• Radio Photo Luminescent (RPL)• Polymer-Alanine Dosimeter (PAD)

Commercial available passive dosimeters:

Page 4: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Dose range of various dosimeters

Page 5: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

RPL and alanine dosimeter systems

• For both systems a gamma calibration exists up to a dose of 1MGy.

• The dosimeter response in a gamma irradiation field is well understood

Two main goals need to be achieved within the next years:

1. Better understanding of detector response to mixed radiation fields

2. Improvement of the current detector readout systems

Page 6: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Principle of physics of RPL

• Very small (~ 6 mm length) silver doped glas cylinder

• Radiation creates – RPL centres– Colour centres

100 Gy 105 Gy 106 Gy

Page 7: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Readout of dosimeter

•spectrometry–UV light–Luminescence –Signal RPL

transparentlight brown

dark brown

•Relation between dose and RPL signal

Project new RPLreader

Page 8: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Dose [Gy]

1000

RP

L R

eadi

ng V

alue

90

100

110

120

130

140

150

160

Pot

entia

l diff

eren

ce o

n th

e re

sist

or [

V]

0.0

0.2

0.4

0.6

0.8

RPL Calibration CurveMeasured Optical DensityOptical Density Fit

Project new RPL-Reader

1. Feasibility Study: Measurement of the optical density of the Dosimeter

Photo-Diode+ OP-Amp.

UV

RPLUV

RPLSignal

Page 9: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Project new RPL-Reader

12

3

1. UV pulse2. Dirt3. RPL Signal4. Predose with long decay

t [µs]

4

rel.

Lig

ht I

nte

nsi

ty

ns t1 t2

Dose A(t1) – f(A(t2))

2. UV-Pulse Excitation [1]

UV-Pulse

Page 10: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Project new RPL-Reader

Advantages:

- measurable dose range 1µGy (*) – 1MGy… 12 decades (before 7 decades)- well defined reading in the plateau- no washing (for low dose)

(*) According to the brochure of the FGD-1000 Readerfrom Chioda Technol Coporationhttp://www.c-technol.co.jp/

Page 11: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Principle of physics of alanine

• Alanine =amino-acid molecule

• Radiation causes formation of free radicals

NH CHCOOH

CH

CHCOOHCH

radiation

100 Gy 105 Gy 106 Gy

Cylinder (~ 3 cm length) consisting of alanine/polymer mixture

Page 12: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Experiments in mixed radiation fields

• CERF 2003

• CERF August 2004

• CERF October 2004

• p+ Calibration Curve (IRRAD1)

• Radiation Source TT40 Beam dump

Aim of the irradiation campaigns: Collecting data in high-energy

mixed radiation fields before the PS and the SPS are switched off.

Page 13: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Experiment at CERF in 2003

RPL 4RPL 3 RPL 2 RPL 1

Al1

Al2

beam

Mixed hadron beam Momentum Total intensity

60% +, 35% p, 5% K+ 120GeV/c 1.55×1012

Page 14: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Preliminary resultsComparison: simulation vs. experiment

S1

S2

S3

S4

Al1

Al2

 

Simulat-ion

Stat. measured

Measure-ment

Sim/ Meas

RPL 1 30.3 4.4 29 1 1.04

RPL 2 168 5 182 6 0.92

RPL 3 169 4 183 6 0.92

RPL 4 75 7 76 2 0.99

Al1 130 10 127 7 1.02

Al2_a 724 15 854 40 0.85

Al2_b 862 17 879 40 0.98

Al2_c 958 21 879 40 1.09

Al2_d 861 17 789 40 1.09

Simulation results give Gy in air

Gy/pp

Page 15: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

CERF August

Motivation: Investigation of the response functionof Alanine and RPL Dosimeter (e.g. according

to the direction of the dosimeters)

Beam

Cu Target

PMMA Platewith Dosimeters

Cu shielding

Page 16: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

CERF August

BE

AM

Preliminary results: Comparison between simulation andRPL/Alanine dosimeter measurements

Page 17: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Conclusion

• Various measurement campaigns in mixed high-energy radiation fields were performed in order to improve the knowledge of the HLD systems.

• Measurements at IRRAD1 were performed to build up a proton calibration curve.

• RPL and Alanine dosimeter are well suited for High Level Dosimetry at CERN .

• First simulations were performed allowing a better understanding of the behavior of the HL dosimeters in mixed radiation fields.

• An upgrade of both systems is under way.

• Further analysis of the measurements and simulations is under way.

• Preliminary results look very promising.

Page 18: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Acknowledgment

• Glaser Maurice1 PH/TA1, Moll Michael1 PH/TA1• Ravotti Federico1 TS/LEA• My colleagues at SC/RP

1 CERN, CH-1211 Geneva 23, Switzerland

Page 19: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

References

[1] E. Piesch, B.Burgkhardt and M. Viligis, Photoluminescence dosimetry: progress and present state of the art, Radiation Protection Dosimetry, Vol. 33 No. ¼ pp. 215-226 (1990)

Page 20: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Readout of dosimeter

• Amount of radiacals can be detected with EPR

• Resonance condition

h = g µ B• Relation between EPR signal and

absorbed dose

• New dosimeters are under development

• New dosimeters are under development • These new dosimeters allow the use of the newer EPR machine•Further investigation concerning the use of most recent EPR machines for the means of HLD are ongoing

Page 21: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

CERF October

Motivation:

Comparison of the Results from August with:

Positions of the Dosimeter:- TLD600/TLD700- Mini Ionizationchamber- RPL-Dosimeter

• Analysis is currently ongoing

Page 22: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

Fluence [ Protons cm-2 ]

1010 1011 1012 1013 1014 1015 1016 1017

Rea

ding

Val

ue

0.01

0.1

1

10

100

1000

IRRAD1 proton calibration

•Irradiation facility IRRAD1 provides PS proton beam with a momentum of 24 GeV/c

• RPL and alanine dosimeters were exposed up to 1016 protons/cm2

• Measurement of Fluence detector readout curve as a function of dose

Simulations: Fluence to dose in material conversion

Proton dose calibration curve

Fluence [ Protons cm-2 ]

1010 1011 1012 1013 1014 1015 1016

Rea

ding

Val

ue

0

20

40

60

80

100

120

140

160

Page 23: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

HLD systems used at CERN Further development of the systems

Page 24: High-Level Dosimetry systems used at CERN Markus Fuerstner 1, Doris Forkel-Wirth 1, Helmut Vincke 1, Sabine Mayer 1, Isabel Brunner 1, Idelette Floret.

SPS LHC extraction tests (450 GeV/c Protons on beam dump)

Gy/pp

• More than two hundred dosimeters (RPL and alanine) were installed on various positions on the dump (TED). • Analysis of the read out results is ongoing.• Preliminary simulations were performed.• Accurate simulations will be performed in order to evaluate the various field components at the given detector positions.• Measurements will be compared with simulations in order to improve the knowledge of the detector response in mixed fields.

450 GeV/cprotons