Hettich Rotina 38-Brochure

14
ROTINA 38 Benchtop Centrifuges, non - refrigerated and refrigerated ! """ ! #$%#&’$ ()%&

description

Furniture delailing

Transcript of Hettich Rotina 38-Brochure

Page 1: Hettich Rotina 38-Brochure

ROTINA 38

BenchtopCentrifuges, non-refrigeratedand refrigerated

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 2: Hettich Rotina 38-Brochure

Large capacitye.g. 4 x 250 ml, 48 x 15 ml, 10 x 50 ml (Falcon® tubes).Extremely short run-up and run-down times.

High RCFup to 16,233 with standard tubes, 24,400 with microlitre tubes.Efficient coolingcontrollable from -20 OC to +40 OC (ROTINA 38 R).On option, the ROTINA 38 R is alsoavailable● with connexion to an external refrig-

erated/heating circulator (to be provided by the customer) – Cat. No. 1707-50 (without internalcooling unit). Dependent on the performance of the circulator, thecentrifuge can heat or cool within a range of 0 OC + 90 OC.

● with connexion for flushing withnitrogen – Cat. No. 1707-20 (withoutmonitoring and controlling system)

3 complete running programmes can be stored and recalled.

ROTINA 38

® Registered trademark of Becton Dickinson

Developed for high efficiency ROTINA 38/38 R are mainly em-ployed in diagnostics, cell culture,genetics research and microbiology.

The high-performance benchtop centrifuges ROTINA 38 and ROTINA 38 R impress by state-of- the -art technology, largecapacity, extremely easy handlingand exceptionally smooth and quietrunning.

Accommodating a multitude oftubes used in today s laboratoryroutine the ROTINA 38 type of cen-trifuges enable a wide range ofapplications.

Smooth and quiet running

Extraordinary speed stability

Brushless maintenance- free induction drive

Ergonomically arranged controls and displays

Viewing port in the lid

A special lid locking device enableseffortless closing and opening of the lid with one hand.

Housing and lid of metal for stability

Housing finished in scratch andimpact resistant powder coating

Centrifuging chamber and innerlid cover of stainless steel

Additionally steel -armoured centrifuging chamber

Easily exchangeable rotors

Automatic rotor recognition

Imbalance switch - off

Lid locking and holdingduring rotor run

Lid dropping protection

Emergency lid lock release

Top quality is Hettich standard

Performance

Technology

Design

Hettich manufactures high quality centrifuges.

All stages from design to end product are continuously subjected to the strictest controls.

Evidence of conformity withnational and international regu-lations such as IEC 61010 is provided by correspondingcertificates.

Of course, Hettich centrifugesare CE marked.

The ISO 9001:2000 certificateaccredited to the entire com-pany bears witness to the extreme care and responsibilityHettich puts into the manufactureof centrifuges and their acces-sories.

This is why Hettich centrifugesare known for safety and reliabi-lity all over the world.

2

Safety

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 3: Hettich Rotina 38-Brochure

The fully developed up- to -date N control panel meets the requirements of theusers and is extremely easy to operate.

Entry of the parameters is fast, precise and convenient with selector keys on the water- protected foil keypad.

The large display is well readable. During centrifugation the actual values of the parameters are indicated.

The parameters remain in the memory even after the centrifuge has been switched off.

Speed RPM: Input in increments of 10.

Relative centrifugal force RCF: Input in increments of 10.

Change-over RPM/RCF: By entering the rotor radius r/mm the centrifuge calculates and indicates speed or RCF.

Running time t: Input in minutes and seconds, max. 99 min., 59 sec.

Run-up/run-down times: Input of acceleration and braking ramps in levels1– 9 or unbraked run-down.

Temperature T: With the refrigerated model input in increments of 1 OC within a range of -20 OC to +40 OC.

Memory: 3 complete running programmes can be stored and recalled.

Impulse key: For short centrifuging operations. The rotor spins at the preselected speed or RCF as long as the key remains pressed.

ROTINA 38benchtop centrifugedimensions (HxW x D):379 x 468 x 563 mm1707Cat.-No.

1703Cat.-No

ROTINA 38 Rbenchtop centrifuge, refrigerateddimensions (H x W x D): 379 x 468 x 721 mm

3��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 4: Hettich Rotina 38-Brochure

illustrated with buckets No.1430 and lids No.1382,tested for biocontainment 5)

1724 without bucketsCat.-No

Swing -out rotor, 4 -place,n= 4,600 min -1, RCF 3,832

90°

50

29 x 115

16

– 4

162 157 156 156 157 156 156 157 155

3,832

4 20 28 20 20 16 20 20 16 16

6318 1383

16 16

1348 1329 1457 1383 1383 1459 1383 1383 1459 1329

19 232

16

– 4

19 232

158 160 156 156 156 155 155 149 148 156 156 150 160

155

108 36 20 20 20 16 16 4 4 4 16 4 4

4

– – – 0501 0578 0500 0518 0519 0521 0526 – 0545 0546

0534

1339 1343 1383 1383 1383 1329 1329 1330 1331 0761 1348 4417 4416

0765

159

4

0547

1396

150

20

2078/0536

1351

*

capacity ml

run-up in sec

radius in mm

RCF

tubes per rotor

dimensions ØxLmm

run-down in sec

* A rubber stopper for closing the tube is available with Cat.No. 0535

1 3 5 6 7 9 15 25 50 100 10 30 50

30

3,738 3,785 3,690 3,690 3,690 3,667 3,667 3,525 3,501 3,690 3,690 3,549 3,785

3,667

6 x 45 Rhesus 10 x 60 13 x 75 12 x 82 12 x 100 14 x 100 17 x 100 24 x 100 34 x 100 44 x 100 16 x 80 26 x 95 29 x 107

44 x 105

85

3,761

38 x 106

rubber insert

1.5/2.0

3,549

11 x 38

chromebath tube

rubberinsert

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

lid No.1382biocontainment 5)

1430Cat.-No

Accessories

1.6 - 7 1.1 - 1.4 2.7 - 3 2.6 - 2.9 4 - 5.5 4.5 - 5 4.9 7.5 - 8.2 9 - 10

156 162

3,690 3,832 3,714 3,690 3,690 3,714 3,690 3,690 3,714 3,667

13 x 75/100

4 - 10 10

16 x 75/100 15 x 102 11 x 66 13 x 65 15 x 75 11 x 92 13 x 90 15 x 92 16 x 928 x 66

blood collection tubes / urine tubes

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

lid No. 1382biocontainment 5)

1430Cat.-No

braked run-down at level 9 Adapters for additional tubes on request.

unbraked run-down

1) Lowest attainable temperature at max. speed in the refrigerated ROTINA 38 R.

4) Withoutrubber insert

5) Tested by the TÜV in conformity with DIN EN 61010, part 2-020

4

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

2)

162 162

4 4

0509 0513

1347 1384

50

29 x 11517 x 120

15

3,832 3,832

12 25 30

17 x 100 25 x 90 25 x 110

144

3,407

162

3,832

162

3,832

4 4 4

-- -- --

6311 1363 1365

4)

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 5: Hettich Rotina 38-Brochure

braked run-down at level 9

unbraked run-down

1) Lowest attainable temperature at max. speed in the refrigerated ROTINA 35 R.

15 50 1.6 - 5 4 - 7 8.5 - 10

16

–9

158 158 152 152 156 152 152 152 152 152 152 152

4,416 4,416 4,248 4,248 4,248 4,248

16 4 36 28 28 28 28 28 28 16

1736 1443 1458 1438 1438 1441 1438 1438 1431 1439

17 x 120 29 x 115 13 x 75 13 x 100 16 x 100

0509 0513

19 232

capacity ml

run-up in sec

radius in mm

RCF

tubes per rotor

dimensions ØxLmm

run-down in sec

50

29 x 115

158

3,738

4 28 28

1737 1438

--

braked run-down at level 9

Adapters for additional tubes on request.unbraked run-down

1) Lowest attainable temperature at max. speedin the refrigerated ROTINA 38 R.

Hettich rotors are designed for continuous duty and longservice life.

Customers requirements and demands on safety as wellas easy handling have beenconsidered.

The automatic rotor recognitionsystem defines the max. speedof each rotor so that it cannotbe stressed beyond its designlimits.

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

15 50 1.1 - 1.4 2.7 - 3 2.6 - 2.9 4 - 5.5 4.5 - 5 4.9 7.5 - 8.2 9 -10

3,738 3,738 3,596 3,596 3,690 3,596 3,596 3,596 3,596 3,596 3,596 3,596

17 x 120 29 x 115 8 x 66 11 x 66 13 x 65 15 x 75 11 x 92 13 x 90 15 x 92 16 x 92

blood collection tubes / urine tubes

1425Cat.-No

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

5

1441

1.6 - 7

13 x 75/100

4 - 10

16 x 75/100

illustrated with buckets No.1425,adapters No. 1438 for blood collectiontubes and No. 1443 for conical 50 mltubes with screw cap

1724 without bucketsCat.-No

Swing -out rotor, 4 -place,n= 4,600 min -1, RCF 3,832

16

– 9

154 152 152 152 152 152 152 152 152 148

144 56 28 48 48 28 28 8 4 4

– – 0501 0578 0500 0518 0519 0521 0526

1432 1433 1438 1434 1434 1431 1431 1435 1436 1437

19 232

151

36

2078/0536

1444

capacity ml

dimensions ØxLmm

1 3 6 7 9 15 25 50 100

3,643 3,596 3,596 3,596 3,596 3,596 3,596 3,596 3,596 3,501

6 x 45 Rhesus 10 x 60

5

13 x 75 12 x 82 12 x 100 14 x 100 17 x 100 24 x 100 34 x 100 44 x 100

1.5/2.0

3,572

11 x 38

1425

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

Cat.-No

90°

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 6: Hettich Rotina 38-Brochure

1724 without bucketsCat.-No

illustrated with buckets No.1427 and lids No.1421

Swing -out rotor, 4 -place,n= 4,600 min -1,RCF 3,809

90°

Accessories

unbraked run-down

1) Lowest attainable temperature at max. speed in the refrig. ROTINA 38 R.

braked run-down at level 9

3) When using these tubes bucket No.1427 cannot be closed with lid No.1421. Adapters for additional tubes on request.

6

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

19 232

1.6 - 5 4 - 7 4 - 7 8.5 - 10 10 1.1 - 1.4 2.6 - 2.92.7 - 3 4.5 - 5 4.9 7.5 - 8.2 9 -10

16

– 2

156 156 154 153 159 159 157 157 156 159 154

3,690 3,690 3,643 3,620 3,761 3,761 3,714 3,714 3,690 3,761 3,643

32 32 20 24 20 48 48 48 32 20 20

1732 1732 5271 5231 5279 5278 5229

156

3,690

32

1732 5229 1732 5279 5271

13 x 75 13 x 100 16 x 75 16 x 100 15 x 102 8 x 66 13 x 6511 x 66 11 x 92 13 x 90

4 - 5.5

159

3,761

20

5279

15 x 75 15 x 92 16 x 92

3) 3) 3) 3)

blood collection tubes / urine tubes

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

lidCat.-No. 1421

1427Cat.-No.

1.5/2.0

16

19 232

– 2

159 158 157 156 157 153 153 153 153 153

3,761 3,738 3,714 3,690 3,714 3,620 3,620 3,620 3,620 3,620

36 120 48 32 48 48 24 24 8 4

2078/0536 – – – 0501 0578 0500 0518 0519 0521

5277 1357 1327 1732 5229 5230 5237 5231 5232 5233

11 x 38

1 3 5 6 7 9 15 25 50

6 x45 Rhesus 10 x 60 13 x 75 12 x 82 12 x 100 14 x 100 17 x 100 24 x 100 34 x 100

1.5/2.0

157

3,714

48

2078/0536

5273

11 x 38

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

lidCat.-No. 1421

1427Cat.-No.

4)

4) Without rubber insert

19 232

15

16

– 2

161 154 154

3,809 3,643 3,643

12 8 8

1733 1731 1731

17 x 120

3)

0509 – –

25 30

25 x 90 25 x 110

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

lidCat.-No. 1421

1427Cat.-No.

50

157

3,714

4

5276

29 x 115

3)

0513

153

3,620

4

5233

50

29 x 115

+1 x 6316

3)

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 7: Hettich Rotina 38-Brochure

7

With Cat. No. 5247-91 or 5248-91 these adapterscan be ordered with a decanting aid.

#

34 265

6 7 15 25 50

32

155 154

2,773 2,755

80 80 48 20 8 4

0501 0578 0518 0519 0521 0523

5227 5247

155 154 154

2,773 2,755 2,755

48

5264 5248 5242 5243 5249

12 x 82 12 x 100

9

0500

14 x 100 17 x 100 24 x 100 34 x 100

100

40 x 115

capacity ml

run-up in sec

run-down in sec

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

temperature °C 1) – 2

Deckel5095

##

34 265

32

4

capacity ml

run-up in sec

run-down in sec

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

temperature °C 1) – 2

Deckel5095

0526

5262

100

44 x 100

93/149

1,664/2,665

8

two-tierframe

for cytoprepara-

tions

Pleaseask for our

detailedcyto

brochure !

5280

153 155 157 155

2,737 2,773 2,808 2,773

80 80 48 48

5267 5227 5268 5264

1.1 - 1.4 2.7 - 3 2.6 - 2.9 4 - 5.5

8 x 66 11 x 66 13 x 65 15 x 75

blood collection tubes

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

5091Cat.-No.

5091Cat.-No.

unbraked run-down

1) Lowest attainable temperature at max. speedin the refrigerated ROTINA 38 R.

braked run-down at level 9

155 157 155

2,773 2,808 2,773

80 48 48

5227 5268 5264

4.5 - 5 4.9

11 x 92 13 x 90

7.5 - 8.2

15 x 92

154

2,755

44

5258

9 - 10

16 x 92

1798 without bucketsCat.-No

illustrated with buckets No. 5091 and lid No. 5095

Swing -out rotor, 4 -place,n= 4,000 min -1,RCF 3,095

90°

158 153109/158

2,826 2,7371,950/2,826

64160

5281

80

155

2,773

80

5267 52275257

1.5/2.0 1.5/2.0

11 x 38 11 x 38

2078/0536

3

10 x 60

5

13 x 75

–��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 8: Hettich Rotina 38-Brochure

8

34 265

32

157 157 155 154 154 162 158

2,808 2,808 2,773 2,755 2,755 2,898 2,826

48 48 48 48 44 28 8

5268

blood collection tubes / urine tubes

5268 5264 5248 5258 6306 5259

0509 0513

1.6 - 5 4 - 7 4 - 7 8.5 - 10 10

15 x 10213 x 75 13 x 100 16 x 75 16 x 100

15 50

17 x 120 29 x 115

4) 4) 4) 4)

– 2

--

12

162

2,898

28

6306

17 x 100

154

2,755

--

50

8

5243

29 x 115

+2 x 6316

braked run-down at level 9

unbraked run-down

--

30

154

2,755

20

5266

25 x 110

1) Lowest attainable temperature at max. speed in the refrigerated ROTINA 38 R.4) When using these tubes, bucket No. 5091 cannot be closed with lid No. 5095.

run-up in sec

run-down in sec

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

temperature °C 1)

5091Cat.-No.

lidCat.-No. 5095

run -down in sec

0501 0518 0519 0509

34 265

6

32

48

5128

12 x 82

5092

lidNo. 5093

biocon-tainment 5)

28

5121

15

17 x 100

capacity ml

run-up in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

Cat.-No.

25 1550 100 100 250250

173165

2,898 3,095 3,0052,952

16 284

0521 0526 0523 51270530

5122 51295125 5126 631910575124

24 x 100 17 x 12034 x 100 44 x 100 40 x 115 62 x 12265 x 115

0578

7

48

5120

12 x 100

3,005

168 162 165 173

-3

2,952 3,095

4 4

0513

50

8

5123

29 x 115

--

12

168 158

2,826

28

5121

17x 100

--

25

12

5134

25 x 90

--

30

162 169

2,898 3,023

16

5122

25 x 110

--

50

8

5135

29 x 115

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

Accessories

6)

5) Tested by the TÜV in conformity with DIN EN 61010, part 2-0206) Please remove the insert in adapter No. 5129, before closing bucket No. 5092 wih lid No. 5093

capacity ml

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 9: Hettich Rotina 38-Brochure

1726 without suspensionsCat.-No.

illustrated with suspensions No. 1742

Accessories

90°

Swing-out rotor, 6 -place, n= 4,000 min -1, RCF 2,808

16

20 210

- 10

129 128 128

180 72 72

– – –

1357 1327 1326

130

54

5277

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions ØxLmm

1 3 4

2,308 2,290 2,290

6 x45 rhesus 10 x 60 12 x 60

2,325

2078/0536

1.5/2.0

11 x 38

1366Cat.-No.

45

152

2,719

6

31 x 100

15

149

2,665

24

17 x 100

5

141

–13 –10 –13 –9 –13 –9 –9

149

2,522

102

13 x 75

5

13 x 75

9

–10

2,665

30

0500– – 0578 – –0518 0519 0521

14 x 100

50

157

2,808

6

34 x 100

20

16

20 210

152

2,719

12

21 x 100

25

157

2,808

12

24 x 100

1345 1746

0521

50

148

2,647

6

34 x 100

13081346 174513691372

7

149

2,665

24

12 x 100

–10

1369-92

149

2,576

24

–10

1369-91 1370

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions ØxLmm

1726Cat.-No.with

decantingaid

with decanting

aid

16

20 210

–9

155 157

60 36 60

0500

0701 – –0701

# #

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions ØxLmm

1741Cat.-No.

9 1.1 - 1.4 4.5 - 5 4.9

2,773 2,808

blood collection tubes

14 x 100 8 x 66 11 x 92 13 x 90

rubberinsert

rubberinsert

–4

16

20 210

157 144 157 157157 150 150 157

42 42 42 1818 42 42 42

0507 0509

– – 1744 – – 0716 0716 –

# #

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions ØxLmm

1742Cat.-No.

15 4 - 5.5 9 - 10 1015 1.6 - 5 4 - 7 8.5 - 10

2,808 2,576 2,808 2,8082,808 2,683 2,683 2,808

blood collection tubes / urine tubes

17 x 100 15 x 75 16 x 92 15 x 10217 x 120 13 x 75 16 x 75 16 x 100

rubberinsert

rubberinsert

PPinsert

Adapters for additional tubes on requestbraked run-down at level 9unbraked run-down

1) Lowest attainable temperature at max. speed in the refrig. ROTINA 38 R.

# # May only be loadedaccording to pattern

9

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 10: Hettich Rotina 38-Brochure

Adapters for additional tubes on requestbraked run-down at level 9

unbraked run-down

1) Lowest attainable temperature at max. speed in the refrig. ROTINA 38 R.

90°

illustrated with buckets No.1790

1794 without bucketsCat.-No.

Swing-out rotor, 4-place, n= 4,000 min -1, RCF 2,808

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

10

Ausschwingrotor, 4 - fach, n= 4.000 min-1, RCF 2.719

65 7 15 25 50 100100

16

– 3

20 246

152 151 151 151 145 149 148 148 157 157

2,7192,701 2,701 2,701 2,594 2,665 2,647 2,647 2,808 2,808

4848 48 28 16 4 4 4 16 8

05780501– 0518 0519 0521 0526 0523

12 x 8213 x 75 12 x 100 17 x 100 24 x 100 34 x 100 40 x 11544 x 100

15 50

0509 0513

17 x 120 29 x 115

157

2,808

4

250 250 10

0530 5127 urinetube

65 x 115 62 x 122 15 x 102

Cat.-No. 1790

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

142 145157 151

2,540 2,5942,808 2,701

-- --

25 30

12 164 28

51285128 5120 5121 5122 5124 5125 5126 5133 17931791 6319 51215134 5122

25 x 90 25 x 110

* The screw cap for tube No. 0538 isavailable with Cat. No. 0539 (dimensions with cap: 38 x 106 mm).

98/115

105/120 1,110/1,103

1796

dimensions Ø x L mm

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF ROTINA 38

RCF ROTINA 38 R

tubes per rotor

Cat.-No.

Cat.-No.

capacity ml

Cat.-No.

302515 10 50 85 94 250

3228 12 12

--0518 0519 05460545 51270538*0547

56415646 5643 5644 5644 --

16 x 8017 x 100 24 x 100 26 x 95 38 x 10629 x 107 62 x12238 x102

+ 4

4 4 4 4

5642 5642

112 108 113 108 100 102 119

11,301 10,897 11,402 10,897 10,090 10,292 12,007

10.142 9.780 10.233 9.780 9.056 9.237 10.776

4)

Cat.-No. 1796

25°

Angle rotor, 4-placeROTINA 38 R: n= 9,500 min -1, RCF 12,007ROTINA 38: n= 9,000 min-1, RCF 10,776

4) Withoutrubber insert

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 11: Hettich Rotina 38-Brochure

braked run-down at level 9

unbraked run-down

Cat.-No. 1792

Adapters for additional tubes on request

1) Lowest attainable temperature at max. speed in the refrig. ROTINA 38 R.

45°

56 / 53

+ 7

53 93056 1040

11 x 38

117 115 115 112 117 122 115 115

24 12 6 6 6 6 6 6 6 6 6 6

1449 1448 1451 1447 1446 –

26 x 95 29 x 107 38 x 10217 x 100

2078/0536 – 0518

112

1447

24 x 100

0519 0545 0546 0538*

16 x 80

117 119

0509 0513

1466 1454

29 x 11517 x 120

122

0547

--

38 x 106

1451 1451

58 / 56

15 x 92

Angle rotor, 6 -place, with lidROTINA 38 R: n= 11,000 min -1, RCF 16,504ROTINA 38: n= 10,000 min-1, RCF 13,640

RCF ROTINA 38 R

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF ROTINA 38

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

1792Cat.-No.

10 15 30 50 9425 15 5085

15,828 15,557 15,828 15,557 15,557

1.5/2.0 7.5 - 8.2

15,151 15,82816,504

13,081 12,857

15,557

12,857 13,081 12,857 12,85712,522

15,151

12,522 13,081

16,098

13,30413,640

16,504

13,640

blood collec-tion tube

15 x 102

10

urine tube

*The screw cap for tube No. 0538 is available with Cat.-No. 0539 (dimensions with cap: 38 x 106 mm).

Angle rotor, 6 -place, ROTINA 38 R: n= 11,000 min-1, RCF 16,504ROTINA 38: n= 10,000 min-1, RCF 13,640

1720Cat.-No.

58 / 56

– 1

57 780 58 960

113 113 112 117 122 113

12 6 6 6 6 6

1448 1451 1447 1446 – 1451

10 15

26 x 95 29 x 107 38 x 10617 x 100

30 50 85 9 - 10

– 0518 0545 0546 0547

16 x 80

1454

115 119

15

17 x 120

50

29 x 115

6 6

1466

0509 0513

115

24

1449

11 x 38

2078/0536

12,633 12,633 12,522 13,081 13,640 12,633

16 x 92blood collec-

tion tube

12,857 13,30412,857

1.5/2.0

15,557 15,286 15,286 15,151 15,828 16,504 15,286 16,09815,557

capacity ml

run-up in sec

run-down in sec

temperature °C1)

radius in mm

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

1720Cat.-No.

RCF ROTINA 38

RCF ROTINA 38 R

45°

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

11

2)

2) 2)

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 12: Hettich Rotina 38-Brochure

1) Lowest attainable temperature at max. speed in the refrig. ROTINA 38 R.

5) Tested by the TÜV in conformity with DIN EN 61010, part 2 -020.

1711 without framesCat.-No.

illustrated with frames No. 1377

15 241

6 x 18/45 8 x 45

---- 2078--

20232023 --2024

8 x 30 11 x 38

15

+ 4

97

30

0536

11 x 38

25 / 16

8 x 30 8 x 45

+ 6

75

60192 126 126

2078– – --

13771378 1379 1379

11 x 386 x 18/45

60

0536

1377

11 x 38

355 402 28 18

45°

Drum rotor, 6 -place, with lidn= 15,000 min-1, RCF 18,866

Adapters for additional tubes on request.

RCF

Cyto rotor, 8 -place

More information is available in a detailed cyto brochure.Please ask for it!

Accessories

Cat.-No.

Angel rotor, 30-place, with lidn= 15,000 min-1, RCF 24,400

1789tested for biocontainment 5)

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

tubes per rotor

Cat.-No.

Cat.-No.

1789Cat.-No.

0.50.2 / 0.4 0.8 1.5

24,400

2.0

For centrifu-gation at

high speedswe recom-

mend to use adapter

No. 2031

1711Cat.-No.

0.5 0.80.2 / 0.4 1.5

18,866

2.0

braked run-down at level 9unbraked run-down

1748 without bucketsCat.-No.

illustrated with buckets No. 16801717 with bucketsCat.-No.

Swing-out rotor, 10-place, n= 4,600 min-1, RCF 3,856

45°

163

19 206

15 50

15

– 3

3,856

10

0509 0513

1462 –

17 x 120 29 x 115

capacity ml

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

tubes per rotor

Cat.-No.

Cat.-No.

dimensions Ø x L mm

Cat.-No. 1717

12

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

dimensions Ø x L mm

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 13: Hettich Rotina 38-Brochure

illustrated with carriers No.1735

Swing-out rotor, 2 - place, n= 4,000 min-1, RCF 2,075

Cat.-No. 1713 without carriers

90°

90°

Cat.-No.1717

116

17

31 430

- 15

8 6 4 2

1735

118

16

18 182

- 16

4

4345

Adapters for additional tubes on request.

2,111

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

plates per rotor

Cat.-No.

run-up in sec

run-down in sec

temperature °C 1)

radius in mm

RCF

plates per rotor

Cat.-No.

1) Lowest attainable temperature at max. speed in the refrigerated ROTINA 38 R.

braked run-down at level 9unbraked run-down

illustrated with carriers No. 4345

1725 without carriers

Swing-out rotor, 4 - place, n= 4,000 min-1, RCF 2,111

Cat.-No

1725Cat.-No.

1713Cat.-No.

Microtitreplate

Cultureplate

2,075

Cultureplate

Deep wellplate

Microtitreplate

Microtitreplate

with lid

All models are of sturdydesign and feature largecastors and two locks.

Any vibrations of the centrifuge are safely compensated.

765

260

550 550

765

260550

550

low and narrow,with one drawer

depth 650 mm

high and narrow,with two drawers

depth 650 mm

low and wide,with one drawer

depth 650 mm

high and wide,with two drawers

depth 650 mm

4612 Cat.-No. 4613 Cat.-No.

4614 Cat.-No. 4615 Cat.-No.

...with rolling cabinets for Hettich centrifugesSave space on the laboratory bench and gain more flexibility

Please take care not to exceed the tubes’ max. permissible RCF. 2) For glass tubes the max. RCF is 4,000.

13

Microtitreplate

with lid

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�

Page 14: Hettich Rotina 38-Brochure

ROTINA 38 RBenchtop centrifuge, refrigerated, without rotor

Cat.-No.1707 Cat.-No.1707-01

Andreas Hettich GmbH & Co. KGFöhrenstr. 12 · D-78532 Tuttlingen / GermanyPhone +49(0)7461 705-0 · Fax +49 (0) 7461 705-122National Sales (07461) 705-200 · National Service (07461) 705-202Internat. Sales +49 (0)7461 705-201 · Internat. Service +49 (0)7461 705-203

www.hettichlab.com · [email protected] · [email protected] me/ga � © by Hettich Tuttlingen · Printed in Germany · Design, form and colour modifications reserved.

Housing and lid of metal for safety. Panelling finished in a scratch and impact resistant coating. Centrifuging chamberand inner lid of stainless steel. Lid locking, lid holding, and imbalance switch -off are standard.

Design

220-240 V 1~50-60 Hz 880 VAEN 55011 ISM class B

110-127 V 1~50-60 Hz 760 VAFCC class A

EN 50082-1 -- --EN 50082-1

4 x 250 mln= 15,000 min-1 / RCF 24,400

1-99 min, ∞ (continuous run), impulse key (short cycle mode)

*other voltages on request

ROTINA 38Benchtop centrifuge, without rotor

Cat.-No.1703 Cat.-No.1703-01

*Power supply:Frequency:

Consumption: Radio interference:

Immunity:

Dimensions (H x W x D): Weight:

379 x 468 x 563 mmapprox. 47 kg

Capacity:Speed/RCF:

Centrifuging chambertemperature:

Running time:

208-240 V 1~50 Hz 1300 VAEN 55011 ISM class B

110-127 V 1~60 Hz 1500 VAFCC class A

379 x 468 x 721 mmapprox. 73 kg

controllable from -20°C to +40°C

Technical Details

��������������� �������������������������������

������������� �� �!���""" �������� �� �!�

���

���

���

���#$%#&�'$()%&�