Harmonic and Flicker

56
Regulatory standards for ac mains phenomena are critical to maintaining the quality of ac power distribution systems. The goal of this application note is to provide an introduction to these standards as well as insight into their scope and intent. An effort has also been made to openly explore the interpretive as well as technical issues of the stan- dards. The discussion of these issues is included to stimulate end-users who test products for com- pliance to consider both the interpretive and tech- nical merit of test solutions. This application note discusses the following topics: the origins of harmonic current, voltage fluctuation, and flicker phenomena IEC 1000-3-2 (1995), EN 61000-3-2 (1995), IEC 1000-3-3 (1994), EN 61000-3-3 (1995) and related standards Agilent Technologies’ interpretation of the standards with regard to test solution implementation the Agilent 6840 Series Harmonic/Flicker Test System Although this document offers a comprehensive discussion of the above standards and Agilent’s test solution, it is most beneficial when read with reference to the relevant standards. Agilent AN 1273 Compliance Testing to the IEC 1000-3-2 (EN 61000-3-2) and IEC 1000-3-3 (EN 61000-3-3) Standards Application Note

Transcript of Harmonic and Flicker

Page 1: Harmonic and Flicker

Regulatory standards for ac mains phenomena are critical to maintaining the quality of ac powerdistribution systems. The goal of this applicationnote is to provide an introduction to these standardsas well as insight into their scope and intent. Aneffort has also been made to openly explore theinterpretive as well as technical issues of the stan-dards. The discussion of these issues is included to stimulate end-users who test products for com-pliance to consider both the interpretive and tech-nical merit of test solutions. This application notediscusses the following topics:

• the origins of harmonic current, voltage fluctuation, and flicker phenomena

• IEC 1000-3-2 (1995), EN 61000-3-2 (1995), IEC 1000-3-3 (1994), EN 61000-3-3 (1995) and related standards

• Agilent Technologies’ interpretation of the standards with regard to test solution implementation

• the Agilent 6840 Series Harmonic/Flicker Test System

Although this document offers a comprehensivediscussion of the above standards and Agilent’stest solution, it is most beneficial when read withreference to the relevant standards.

Agilent AN 1273

Compliance Testing to the IEC 1000-3-2 (EN 61000-3-2) and IEC 1000-3-3 (EN 61000-3-3)StandardsApplication Note

Page 2: Harmonic and Flicker

Table of Contents

Introduction 3IEC 1000-3-2 and 1000-3-3 Standards and the

Implications for Electronics Manufacturers 3

Understanding the Regulatory Testing Environment 4The Relationship between the IEC, CENELEC,

and the European Union 4EMI/RFI Standards in General 4Low Frequency Emissions 5The Standards Process 5

The Necessity of AC Mains Regulations 6Harmonic Currents 6Why Switch-mode Power Supplies Generate

Harmonic Currents 7Effects of Harmonic Currents 8Quasi-stationary and Fluctuating Harmonics 8Voltage Fluctuations and Flicker 9

An Overview of IEC 1000-3-2 (EN 61000-3-2) and IEC 1000-3-3 (EN 61000-3-3) Requirements 11The Difference Between Precompliance and

Compliance Testing 12Reference Methodologies 12

IEC 1000-3-2 and EN 61000-3-2 13Class Definition 13Class D Equipment 14Test Limits 15AC Power Source Requirements 16Harmonic Current Measurement Requirements 16

IEC 1000-3-3 and EN 61000-3-3 18Test Limits 19AC Power Source Requirements 19Voltage Fluctuations/Flicker Measurement

Requirements 20

The Agilent 6840 Series Harmonic/Flicker Test System Solution 21Basic Measurement Definitions 22

Interpretive Issues for Test Solutions Vendors and Equipment Manufacturers 23Normative References and Their Scope 23Stated, Implied, and Inferred Requirements 24Appropriateness of Requirements 25Technical Errors 26

Implementation of IEC 1000-3-2 and EN 61000-3-2 Standards 29Harmonic Current Measurements 29Agilent 6840 Series Measurement Accuracy 30Measurement Synchronization Using

Rectangular Windows 31Shunt Burden 331.5 Second Smoothing Filter 33Source Impedance 33The Pre-Test Window for Class Determination 34Class D Active Power Tests and the Special

Waveshape Requirement 35Class C Power Factor and Fundamental

Current Measurement 37Class B Harmonic Current Limits 37Voltage Distortion Measurement 38Delay to Start of Test 38Programmable Current Limits 39

Implementation of IEC 1000-3-3 and EN 61000-3-3 Standards 40Flicker Measurement—UIE Flickermeter 40RMS Measurements 40Pst Classifier 41Agilent 6840 Series Flickermeter 41Use of a Simulated Impedance 42Squarewave Response Tests per IEC 868 42Unclear Definitions for Dmax, Dc, and Dt 44Delay to Start of Test 47

Appendix A 48Flickermeter Verification Tests 48Quantization Errors 50

References 55

Reading This Application NoteThis application note is written with two audiencesin mind. The first part of the application note, frompages 3 to 22, is for readers who wish to familiarizethemselves with the IEC 1000-3-2 and 1000-3-3 (EN 61000-3-2 and 61000-3-3) standards. This gen-eral discussion concludes with a brief description ofthe Agilent 6840 Series Harmonic/Flicker Test System.

The second part of the application note, from pages23 to 55, is for readers who are familiar with IEC1000-3-2 and 1000-3-3 (EN 61000-3-2 and 61000-3-3)and are aware of some of the technical and inter-pretive issues surrounding the standards. Thesepages identify the issues and describe in detail howAgilent has resolved them with the 6840 SeriesHarmonic/Flicker Test System.

Page 3: Harmonic and Flicker

3

IEC 1000-3-2 and 1000-3-3 Standards and theImplications for Electronics ManufacturersThe quality of distributed electrical power hasbecome a critical concern in recent years. This concern has grown with the advent of positiveindustry trends to improve electrical product effi-ciency and with the proliferation of electronicequipment in households and commercial environ-ments. Interestingly enough, the very technologythat allows for more efficient use of electrical energy, switch-mode power conversion technology,is also a culprit that can negatively impact powerquality. Switch-mode power supplies can drawhigh harmonic currents from the ac mains, whichcan cause a variety of undesirable effects in the ac power distribution system. The end result forelectronic products connected to the ac mains canbe improper operation, over-stressed ac input components, and product failure. The power distri-bution system must bear the burden of excessiveneutral currents that over-stress neutral wiringand inefficient power transmission to the loadsconnected to the system.

The power quality of the ac mains can also deterio-rate due to voltage fluctuations caused by devicessuch as electronic ballasts and light dimmer switches.The phase-controlled ac input of these devices maycause large rms current changes on the ac mains,which result in substantial rms voltage deviations.In addition, the amplitude and frequency of thesedeviations can cause incandescent lamps to flicker,which can be irritating to humans.

Concern has grown regarding the consequences of these ac mains phenomena. This has resulted inaction from worldwide regulatory commissions toupdate requirements of standards and implementregulations for electrical and electronic products.In fact, the European Union considers these conse-quences sufficiently severe as to require complianceto harmonic emissions standards under the require-ments of the EMC Directive. These standards weremost recently known by their older voluntary ver-sions, the IEC 555-2 and 555-3 (EN 60555 parts 2and 3) standards. Very recently, these standardshave been revised and published as the IEC 1000-3-2and 1000-3-3 (EN 61000-3-2 and 61000-3-3) stan-dards, which set the limits for harmonic currentemissions and voltage fluctuations and flicker.

The publication of these standards in the OfficialJournal of the European Union (OJEC) and theirimpact on the ability of electronic and electricalequipment manufacturers to participate in theworldwide marketplace has greatly increased indus-try anxiety levels. The intent of this applicationnote is to clarify the requirements of the standards,identify some of the technical and interpretiveissues raised by the standards, and explain howthe capabilities of Agilent’s test solution greatlysimplify compliance testing to these standards.

Introduction

Page 4: Harmonic and Flicker

4

The Relationship between the IEC, CENELEC,and the European UnionAlthough the necessity of regulations such as IEC1000-3-2 and 1000-3-3 may be well understood, thesame may not be said of the process that drives theirdevelopment. The worldwide organization respon-sible for developing most of the electrical standardsrelated to the commercial and consumer industryis the International Electrotechnical Commission,or IEC. Through liaisons with international, national,and other standards-writing organizations as wellas its own IEC Technical Committees, the IEC devel-ops and publishes electrical standards and guide-lines for voluntary use for entire industries at large.The IEC standards and guidelines can then be har-monized into national and regional standards with-out any alteration; any differences are clearly doc-umented. The national and regional standards havedifferent designations for the same IEC standard.For example, EN 61000-3-2 is the European Unionversion of IEC 1000-3-2 and is a regional standardpublished and ratified by CENELEC. BSEN 61000-3-2refers to the national United Kingdom version ofthe EN 61000-3-2 standard.

The European Union and the European Free Tradecountries have the need to development new EMI/RFI immunity and emission standards. Their goalis to establish a comprehensive and uniform set ofrequirements for compliance to the EMC Directive(89/336/EEC). All electrical or electronic productsunder the scope of the EMC Directive intended forsale and distribution in the participating countrieswill be required to comply, and must carry the “CEmark” or label as proof of compliance by January1, 1996. The specific requirements and associateddetail are covered in the EMC Directive publicationand its addendums. IEC and CENELEC are there-fore working in close collaboration to prepare thesestandards related to the EMC Directive, which willbe published under the IEC and EN designations.

EMI/RFI Standards in GeneralThe IEC standards are developed by committees andsub-committees comprised of national committeemembers. The IEC TC77 committee has the respon-sibility for creating standards that address mostaspects of electromagnetic disturbances. They haveclassified these disturbances in terms of low andhigh frequency conducted, magnetic, and radiatedphenomena; electrostatic discharge; and nuclearelectromagnetic pulse phenomena. Once published,these standards will be designated by the followingformats:

• IEC 1000-1-x General considerations, definitionsand terminology

• IEC 1000-2-x Description and classification of theenvironment levels and compatibility

• IEC 1000-3-x Emission and immunity limits

• IEC 1000-4-x Testing and measurement method-ologies and techniques

• IEC 1000-5-x Installation and mitigation guidelines

• IEC 1000-9-x Miscellaneous

Understanding the Regulatory Testing Environment

Page 5: Harmonic and Flicker

5

Low Frequency EmissionsSub-committee A of IEC TC77 is responsible forconducted low frequency disturbances includingthose that pertain to ac power distribution systems.One of the committee’s efforts has been the revisionof the IEC 555-2 and 555-3 emissions standards,which led to the development of the new IEC 1000-3-2 and 1000-3-3 standards and the correspondingCENELEC versions (EN 61000-3-2 and 61000-3-3).

The Standards ProcessElectrical and electronic equipment manufacturersusually have a critical need to understand whencompliance to a standard is required. The processby which a document becomes an IEC standardinvolves several drafts or revisions, and typicallytakes anywhere from 4 to 12 years. While in theform of a revision, the “pre-standard” document isidentified by an IEC designation and number thatreflects the committee preparing the standard andthe particular revision cycle. For example, the IEC 1000-3-3 standard prepared by sub-committeeTC77A was designated as IEC 77A(CO)38 duringits Draft International Standard stage just prior to approval. Compliance to a “pre-standard” docu-ment during any stage of the revision process isnot required because the potential for change stillexists and the committees have not officiallyapproved the document for release.

These documents are released to National Committeessolely for the purpose of technical commentary,and are not intended to be used as testing guide-lines. It is not until the representative IEC NationalCommittees vote on and approve the document for publication that it becomes an official IEC stan-dard. The IEC 1000-3-2 and 1000-3-3 standards will not become regulations (i.e., a mandatory testrequirement) for the European Union until CEN-ELEC harmonizes the documents and the EuropeanCommission publishes them in the Official Journal,scheduled for January, 1996. They are now listedwith reference to the EMC Directive as the EN61000-3-2 and 61000-3-3 standards.

Page 6: Harmonic and Flicker

6

Harmonic CurrentsUp until the last few decades, ac mains harmoniccurrents were mainly a concern in heavy industrialenvironments because of the type of loading thisequipment placed on the electrical distribution sys-tem. Facilities of this type were the main users ofelectrical equipment with phase-angle controlledand rectified power devices. Most households andcommercial environments still used fairly simpleinductive and resistive electronic devices or loadssuch as incandescent lights, motors, and heaters.When connected to the sinusoidal voltage of the ac mains, these loads draw a sinusoidal currentwaveform which has little or no harmonic content(see Figures 1 and 2).

The proliferation of electronic equipment such ascomputers, printers, televisions, and audio equip-ment has expanded the harmonic current problemto households and commercial environments. Thisequipment has ac input circuitry that draws non-sinusoidal currents (see Figure 3). Typically, thisinput circuitry consists of a switch-mode powersupply with a capacitor-diode rectifier input stage.Devices such as lamp dimmers and electronic bal-lasts also draw non-sinusoidal currents because ofthe phase-angle control circuits used to regulatethe input power.

The Necessity of AC Mains Regulations

Figure 1. Resistive load

Figure 2. Inductive load

Figure 3. Non-sinusoidal currents

Page 7: Harmonic and Flicker

7

Why Switch-mode Power Supplies GenerateHarmonic CurrentsAlthough switch-mode power supply technologyoffers high efficiency in a small size, power sup-plies using this technology are one example of a generator of ac mains harmonics. Switch-modepower supplies typically have a capacitor-dioderectifier input stage, which creates a high voltagedc source from the ac input voltage (see Figure 4).This dc voltage is then modulated or chopped toprovide a regulated voltage to the load. When theac voltage is first applied, the charging of thecapacitor is accompanied by a large inrush of accurrent until the capacitor is charged to the peakvalue of the rectified voltage. Once the inrush con-ditions subside, the diodes conduct ac current andcharge the capacitor only when the ac voltage isgreater than the capacitor voltage. AC current flowis blocked by the diodes when the ac voltage islower than the capacitor voltage (see Figure 5).

The charging of the capacitor at the peak of the acvoltage waveform causes ac current to flow throughthe capacitor in successive narrow current pulses.This creates a nonlinear ac mains current that isharmonically rich. This current waveform is madeup of odd-order harmonics which are integer mul-tiples of the fundamental frequency (see Figure 6).

The current fundamental and odd harmonic ampli-tudes sum at 90 degrees and 270 degrees of the fundamental, which creates the current pulse wave-form. If the fundamental frequency is 50 Hz, the3rd harmonic is at 150 Hz, the 5th harmonic is at250 Hz, and so on. This type of current waveformis representative of the Class D equipment with a“special wave shape” as described by IEC 1000-3-2and EN 61000-3-2.

Figure 4. Switch-mode power supply circuit

Figure 5. Capacitor voltage with rectifier voltage andcurrent waveforms

Figure 6. Summing odd harmonic currents

Page 8: Harmonic and Flicker

8

Effects of Harmonic CurrentsElectronic equipment and devices that cause non-linear current waveforms produce many detrimen-tal effects on the ac power distribution. As more of these devices load the ac power distribution system, the following problems can occur:

• Significant current can flow that does not deliverreal power and must be supported by the ac dis-tribution system.

• Elevated neutral currents can flow in three-phasewye distribution systems which cause ampacityratings for wiring and connectors to be exceeded.

• Equipment connected to the same branch circuitcan operate improperly due to severe voltagedistortion caused by harmonic currents interact-ing with impedances in the distribution system.For example, suppressed or clipped peak voltagelimits the ability of computer power supplies toride through momentary ac mains sags.

Equipment manufacturers generally use power factor correction (PFC) circuitry to minimize theeffects of nonlinear loads and the power inefficien-cies they cause. Testing to the harmonic currentregulations helps to ensure that these power factorcorrection circuits are designed and operatingproperly.

Quasi-stationary and Fluctuating HarmonicsThe harmonic current distortion caused by a non-linear load on the ac mains can be classified in one of two general categories: quasi-stationary orfluctuating. Quasi-stationary or steady-state har-monic currents are produced by electronic equip-ment that generates non-varying levels of currentdistortion, where the amplitude of each harmonicremains constant over time. This equipment appearsas an unchanging load on the ac mains (see Figure 7).Examples include test and measurement instru-mentation such as oscilloscopes and multimeters,electronic ballasts, and video display equipment.

Figure 7. Steady-state harmonic currents

Page 9: Harmonic and Flicker

9

In contrast, fluctuating harmonic currents arecaused by electronic equipment that represent time-varying loading on the ac mains. This equipmentgenerates varying levels of current consumption ordrain where the amplitude of individual harmonicschange over time (see Figure 8). Examples includemicrowave ovens, dishwashers, laser printers, andphotocopiers. The IEC 1000-3-2 and EN 61000-3-2standards address both classes of harmonic currents.

Different measurement considerations and method-ologies are used for determining compliance forthe two classes, with more stringent testing meth-ods applied to the fluctuating harmonic class. Inparticular, compliance testing for fluctuating har-monics requires non-stop harmonic analysis overextended periods of time.

Voltage Fluctuations and FlickerIn addition to loads that cause harmonic currentson the ac mains, there are also loads that haveautomatic turn on/turn off controls such as ther-mostats and timers. Kitchen appliances, spaceheaters, air conditioners, copiers, and other equip-ment include these type of controls. These loadsare frequently resistive in nature, which meansthat they draw sinusoidal currents and generate no harmonics except during transient events.

When automatic controls cycle on and off, theycause frequent changes of the load to the supply.When the fluctuating load is in a branch circuitwith other loads, these changes cause rms voltagefluctuations that affect all of the loads in the branch(see Figure 9). In particular, variations in voltageamplitude cause changes in the light output of any filament lamps in the branch circuit. Becausethe output of a filament lamp is proportional to the square of the applied voltage, changes in lightintensities can be significant even for small changesin voltage.

Figure 8. Fluctuating harmonic currents

Figure 9. Branch circuit loads

Page 10: Harmonic and Flicker

10

This effect of variation in light output as perceived bythe human observer is referred to as flicker. Becauseflicker is annoying, and for some individuals pres-ents a health hazard (persons that have epilepsyfor example), the standard seeks to regulate flickergeneration to an acceptable level. Figure 10 illus-trates an rms voltage variation on the ac mains.Figure 11 is from the IEC standard that definessuch a voltage change characteristic. The voltagechange characteristic defines changes in rms volt-age levels, U(t), in terms of percentage of nominalline voltage, Un.

The IEC 1000-3-3 and EN 61000-3-3 standardsaddress voltage fluctuations and flicker. Compliancewith these standards ensures that voltage distur-bances in the electrical distribution system do notinterfere with other equipment connected to the ac mains or cause incandescent lights to visiblyflicker in a way that causes an annoyance or health risk to a human observer.

Figure 10. Variations in voltage amplitude

Figure 11. Voltage change characteristic

Page 11: Harmonic and Flicker

11

Developed as part of a comprehensive strategy toset guidelines for EMC compliance under the EMCDirective, the recently published standards includethe following:

• IEC 1000-3-2 (1995)/EN 61000-3-2 (1995): Specifiesthe limits for harmonic currents created byequipment connected to public low-voltage supply systems. These standards cover bothquasi-stationary harmonics and fluctuating harmonics.

• IEC 1000-3-3 (1994)/EN 61000-3-3 (1995): Specifiesthe limits for voltage fluctuations and flickerproduced by equipment connected to public low-voltage supply systems.

The specified limits are applicable to distributionsystems with nominal voltages of 230 V (single-phase) and 400 V (three-phase) at 50 Hz, and forall electrical and electronic equipment with ratedcurrents up to 16 Amps. The harmonics of interestare the 2nd through the 40th harmonic. Note thatIEC 1000-3-2/EN 61000-3-2 also refers to 60 Hz systems, but IEC 1000-3-3/EN 61000-3-3 does not.

There currently appears to be a wealth of confusionregarding the effective dates of these standardsand when they will become mandatory for all prod-ucts shipped into the European Community. Muchdiscussion is being focused on the criteria thatconstitutes whether or not an electronic productmust comply when the standards are published

as part of the EMC Directive guidelines in January1996, or whether mandatory compliance should beenforced after the official date of withdrawal of theolder versions of the standards on January 1997(for EN 60555 part 2) or June 1998 (for EN 60555part 3). In addition, there may be temporary certi-fication clauses exempting certain products depend-ing on the date of introduction into the marketplaceand whether they conform to the older versions of the standards.

IEC 1000-3-2 and EN 61000-3-2 and Related Standards

Published Version Older Version Normative Reference

IEC 1000-3-2 (1995) IEC 555-2 (1982) IEC 1000-4-7 (1991)EN 61000-3-2 (1995) EN 60555-2 (1987)

IEC 1000-3-3 and EN 61000-3-3 and Related Standards

Published Version Older Version Normative References

IEC 1000-3-3 (1994) IEC 555-3 (1982) IEC 725 (1981)EN 61000-3-3 (1995) EN 60555-3 (1987) IEC 868 (1986)

IEC 868 Amendment 1(1990)

An Overview of IEC 1000-3-2 (EN 61000-3-2) and IEC 1000-3-3 (EN 61000-3-3) Requirements

Page 12: Harmonic and Flicker

12

The Difference Between Precompliance andCompliance TestingTesting requirements can differ based on thedesired end result. Precompliance testing is tradi-tionally used during product development to pro-vide confidence that the equipment under develop-ment can meet the standard prior to full compli-ance testing at an outside regulatory laboratory.The instrumentation used is typically general pur-pose equipment found in any well-equipped laband is often less expensive than that used in com-pliance testing. Precompliance testing provides acost-effective means of ensuring that money investedin external testing services will result in the equip-ment being certified for compliance.

Compliance testing requires instrumentation thathas traceable specifications which guarantee per-formance levels sufficient to certify that equipmentmeets the required standards. More importantly,this equipment must meet any special performancecharacteristics defined by the associated normativereference standards. Compliance-level test instru-mentation is generally more expensive than thatused for precompliance testing. The Agilent 6840Series Harmonic/Flicker Test System, however,effectively utilizes new technology to provide com-pliance-level testing at costs significantly belowprecompliance test instrumentation.

Reference MethodologiesCompliance-level testing often encompasses severalpossible testing methodologies as defined by thestandards. The rating of these methodologies rangefrom “acceptable,” which represents the lowestlevel of test instrumentation performance neces-sary to meet the compliance-level testing require-ments, to the “most preferred” methodologies,which represent the optimum level of test instru-mentation performance endorsed by the standardscommittee. The “most preferred” testing method-ologies are called reference methodologies. Equip-ment utilizing these methodologies meets the moststringent performance requirements and the mostpreferred test and measurement techniques. Forexample, IEC 1000-3-3 uses this categorization forinstruments that can assess all types of voltagefluctuations by direct measurement using a UIE/IECflickermeter as defined by IEC 868. The Agilent6840 Series Harmonic/Flicker Test System utilizesreference-level methodologies both for measuringvoltage fluctuations and flicker for IEC 1000-3-3, andfor measuring harmonic currents for IEC 1000-3-2.

Page 13: Harmonic and Flicker

13

The IEC 1000-3-2 and EN 61000-3-2 standardsdefine measurement requirements, ac power sourcerequirements, and limits for testing the harmoniccurrent emissions of electronic and electrical equip-ment. Additional harmonic current testing andmeasurement techniques and instrumentationguidelines for these standards are covered in IEC1000-4-7. IEC 1000-3-2 and EN 61000-3-2 also pro-vide definitions for commonly used terms usedthroughout the standards and describe test condi-tions for particular types of equipment in Annex C.

Compliance to these standards ensures that testedequipment will not generate harmonic currents at levels which cause unacceptable degradation ofthe mains environment. This directly contributesto meeting compatibility levels established in otherEMC standards such as IEC 1000-2-2, which definescompatibility levels for low-frequency conducteddisturbances in low-voltage supply systems.

The block diagram in Figure 12 illustrates theequipment functions required to implement testsper the IEC 1000-3-2 and EN 61000-3-2 standards.

Class DefinitionTo establish limits for similar types of harmoniccurrent distortion, equipment under test must becategorized in one of four defined classes. Use theflowchart in Figure 13 to determine equipmentclassification.

Class A: Balanced three-phase equipment, and allother equipment except that stated in one of theremaining three classes.

Class B: Portable electrical tools, which are handheld during normal operation and used for a shorttime (a few minutes) only.

Class C: Lighting equipment, including dimmingdevices.

Class D: Equipment having an input current with a “special wave shape” (e.g., equipment with off-linecapacitor-rectifier ac input circuitry and switch-mode power supplies) and an active input power≤600 W. The active power is defined as Watts. Forthe Class D mA/W limits to apply, the active powermust also be greater than 75 W. This lower limitwill be reduced to 50 W within 4 years of the imple-mentation of IEC 1000-3-2. Note that the standardis ambiguous as to whether or not the maximumpermissible harmonic current limits of the stan-dard (see Table 3, page 15) apply to devices withactive power below these thresholds.

IEC 1000-3-2 and EN 61000-3-2

Figure 12. Equipment diagram for testing harmonic current emissions

Page 14: Harmonic and Flicker

14

Class D EquipmentFor electronic equipment to be classified as Class D,two requirements must be met, assuming the equip-ment does not meet the definition of the other threeclasses. The first requirement, as previously dis-cussed, is that the active input power of the equip-ment must be ≤600 W. The second requirementaccording to IEC 1000-3-2 states that:

“equipment shall be deemed to be Class D if underthe test conditions given in Annex C, the inputcurrent waveshape of each half-period—referredto its peak value ipk —is within the envelope shownin the following figure for at least 95% of the dura-tion of each half period; this implies that wave-forms having small peaks outside the envelopeare considered to fall within the envelope. Thecenter line M, coincides with the peak value of the input current.”

Figure 13. Flowchart for class determination

Figure 14. Class D waveshape envelope

Page 15: Harmonic and Flicker

15

Test LimitsThe following tables describe the test limits forharmonic currents according to class. Harmoniccurrent limits are noted for Class A, C, and Dequipment in Tables 1, 2, and 3, respectively. Thelimits for Class B equipment are 1.5 times those for Class A.

Table 1. Limits for Class A Equipment

Harmonic order Maximum permissible harmonic current(n) (A)

Odd Harmonics

3 2,305 1,447 0,779 0,4011 0,3313 0,2115 ≤ n ≤39 0,15 x 15/n

Even Harmonics

2 1,084 0,436 0,308 ≤ n ≤40 0,23 x 8/n

Table 2. Limits for Class C Equipment

Maximum permissible harmonic currentexpressed as a percent of the input current

Harmonic order at the fundamental frequency(n) (%)

2 23 30 • l* 5 107 79 511 ≤ n ≤39 3(odd harmonics only)

* l is the circuit power factor.

Table 3. Limits for Class D Equipment

Maximum permissible Maximum permissible Harmonic order harmonic current per watt harmonic current(n) (mA/W) (A)

3 3,4 2,305 1,9 1,147 1,0 0,779 0,5 0,4011 0,35 0,3313 ≤ n ≤39 3,85/n See Table 1(odd harmonics only)

The general harmonic test limits are as follows:

• Harmonic currents <0.6% of the input current or less than 5 mA are disregarded.

• If the envelope of the spectrum of harmonics 20 through 40 decrease monotonically, only harmonic 2 through 19 need to be measured.

• For fluctuating (or transitory) even harmonicsfrom 2 through 10 and odd harmonics from 3 through 19, harmonic amplitudes can be 1.5 times the limits for 15 seconds of any 2.5minute period.

Page 16: Harmonic and Flicker

16

The IEC 1000-3-2 and EN 61000-3-2 standardsdescribe harmonic measurement test conditionsonly for some types of equipment. The objective isto test all applicable equipment under conditionsthat produce the maximum harmonic amplitudesunder normal operating conditions for each har-monic component. A critical element for successfulcompliance testing is the quality of the ac powersource and harmonic measurement instrumentation.

AC Power Source RequirementsIn general, the ac power source used for testingequipment to the harmonic current emissions lim-its must have low output impedance to minimizeoutput voltage distortion, sufficient output voltageratings to supply the appropriate test voltages,peak current ratings sufficient to power the equip-ment under test without clipping, and the ability tosufficiently regulate the output voltage, frequency,and phase relationship (for three phase testing).The specific requirements are:

Output voltage ratings: 230 Vrms (single-phase)400 Vrms (three-phase)

Voltage and frequencyaccuracy: 230 Vrms ±2%, 50 Hz ±0.5%

Phase angle accuracy: 120° ±1.5° between the funda-mental voltage of each phase

Harmonic content of 0.9% (3rd harmonic)the output voltage 0.4% (5th harmonic)(% of Vfundamental): 0.3% (7th harmonic)

0.2% (9th harmonic)0.1% (even harmonics from 2 to 10)0.1% (all harmonics from 11 to 40)

Peak output voltage 1.40 to 1.42 times the rms value value: reached within 87° to 93° after

the zero crossing

Harmonic Current Measurement RequirementsThe scope of the harmonic emissions standardsinclude measurement requirements for both quasi-stationary (steady-state) and fluctuating harmon-ics. The methodology and limits used to measureharmonic levels for compliance is dependent uponwhich type of harmonics are generated by theequipment under test. Annex B of IEC 1000-3-2and IEC 1000-4-7 jointly describe a number ofrequirements for instrumentation used to measureharmonic currents in ac mains circuits. SectionsB.1, B.2, and B.4 are applicable to the FFT imple-mentation used in Agilent’s 6840 Series Harmonic/Flicker Test System. Figures 15 and 16 illustratethe measurement circuits for single-phase andthree-phase equipment according to IEC 1000-3-2and EN 61000-3-2 (Annex A).

The IEC 1000-4-7 standard defines harmonic cur-rent measurement instrumentation in terms ofaccuracy class (A or B) and signal characteristicsof the measured parameter (quasi-stationary orfluctuating). Class A accuracy levels are requiredfor compliance-level harmonic current measurements.

Page 17: Harmonic and Flicker

17

General measurement guidelines are specified for the test conditions defined in Annex C of thestandards or user-defined test conditions that generate worst case harmonic levels (either quasi-stationary or fluctuating harmonics). The objectiveof IEC 1000-3-2 and EN 61000-3-2 is to test equip-ment under normal operation, and therefore cur-rent inrush conditions that typically occur whenequipment is first powered up or powered downare disregarded. A wait period of ≤10 seconds isspecifically called out in the fluctuating (transitory)harmonic measurement requirements. General har-monic measurement requirements are as follows:

• The total measurement error for steady-stateharmonics shall not exceed 5% of the permissiblelimits or 0.2% or the rated current of the EUT.

• The input impedance of the measuring instru-ment (e.g., a current shunt) shall be such thatthe voltage drop across the input circuit shallnot exceed 0.15 Volts peak.

• In the case of fluctuating harmonics, a smooth-ing filter having a response equivalent to a firstorder lowpass filter with a time constant of 1.5seconds ±10% shall be applied to the DFT outputs.

Figure 15. Single-phase measurement circuit

Figure 16. Three-phase measurement circuit

Page 18: Harmonic and Flicker

18

The IEC 1000-3-3 and EN 61000-3-3 standardsdefine the measurement requirements, ac powersource requirements, line impedance requirements,and voltage fluctuation and flicker limits for assess-ing electronic and electrical equipment’s propensityto cause voltage disturbances on the ac mains.Compliance with these standards ensures that volt-age fluctuations do not interfere with other equip-ment connected to the ac mains or cause incandes-cent lights to visibly flicker in a way that causesan annoyance or health risk to a human observer.

The amplitude and frequency of occurrence ofthese voltage fluctuations directly determine thelevel of flicker perceived by the human observer.Figure 17 is a representative curve showing the

relationship between relative amplitude changes,U(t)/Un, versus frequency of occurrence for a con-stant level of flicker irritability, or Pst. The flickermeasurement requirement of these standards, there-fore, quantifies the annoyance a typical humanwould experience due to voltage fluctuations interms of amplitude and frequency. Common sourcesof flicker are equipment that produce time-varyingloading conditions. Some specific examples are hot plates, microwave ovens, and laser printers.

The block diagram in Figure 18 illustrates theequipment functions required to implement theIEC 1000-3-3 and EN 61000-3-3 standards.

IEC 1000-3-3 and EN 61000-3-3

Figure 17. Representative Pst = 1 curve

Figure 18. Equipment diagram for testing voltage fluctuations and flicker

Page 19: Harmonic and Flicker

19

Test LimitsCompliance is determined if the following testparameters are within the following defined limits:

Short-term Flicker (Pst): The flicker severity evaluatedover a short period of time (10 minutes). Pst = 1 is the conventional threshold of irritability, andtherefore the limit.

Long-term Flicker (Plt): The flicker severity evaluatedover a long period (typically 2 hours) using succes-sive Pst values. Plt = 0.65 is the conventional thresh-old of irritability, and therefore the limit.

For voltage changes that are caused by manualswitching of equipment or that occur less frequentlythan once per hour, Pst and Plt are not applicable.However, the following voltage change (“D”) parame-ters are applicable, with the limits multiplied by 1.33.

Relative Steady-state Voltage Change (Dc): The differ-ence between two adjacent steady-state voltagesrelative to the nominal voltage. Dc must be ≤3%.

Relative Voltage Change Characteristic (D(t)): Thechange in rms voltage, relative to the nominal volt-age, as a function of time and between periodswhen the voltage is in a steady-state condition forat least 1 second. D(t) must not be >3% for morethan 200 milliseconds continuously during a volt-age change event.

Maximum Relative Voltage Change (Dmax): The differ-ence between maximum and minimum rms valuesof the voltage change characteristic relative to thenominal voltage. Dmax must be ≤4%.

AC Power Source RequirementsSince flicker is determined by monitoring voltagechanges, requirements for these standards includemaking measurements when the equipment undertest is powered through a specified line or referenceimpedance. This reference impedance is definedaccording to IEC 725. For 230 V/50 Hz power distri-bution systems, the reference value is 0.4 + j 0.25 V(phase to neutral).

Traditionally, this reference impedance (seriesResistor and Inductor) is provided as a separatebox that is connected between the ac source andthe equipment under test as shown in Figure 19.Some ac source equipment has built-in sourceimpedance control, but only the Agilent 6840Series Harmonic/Flicker Test System has the accuracy to provide a true compliance-level reference impedance.

Figure 19. Source with additional R and L (LIN) and load

Page 20: Harmonic and Flicker

20

Other ac power source requirements are:

Output voltage ratings: 230 Vrms (single-phase)400 Vrms (three-phase)

Voltage and frequency accuracy: 230 Vrms ±2%, 50 Hz ±0.5%

Voltage THD: 3%

Maximum source short-term flicker (Pst): 0.4

Voltage Fluctuations/Flicker MeasurementRequirementsThe measurement of the rms voltage fluctuationson the ac mains caused by the equipment undertest is the basis for flicker measurements. Thereare different methods of evaluating flicker severity(Pst) that range from direct measurement via adevice called a UIE/IEC flickermeter to the use ofmathematical analysis or a Pst graph provided bythe standard. The direct method using the UIE/IECflickermeter is the reference method for true com-pliance-level testing. This instrument must comply

with the specifications given in the IEC 868 refer-ence standard. This is the method used in Agilent’simplementation. A detailed discussion of Agilent’simplementation of the UIE/IEC flickermeter is covered under “Implementation of IEC 1000-3-3and EN 61000-3-3 Standards.”

Measurement of voltage fluctuations is also a critical part of determining if electrical equipmentcauses excessive voltage disturbances on the acmains. While flicker measurements provide anaccurate assessment of the effect of continuousvoltage changes, the voltage fluctuation measure-ments provide a better indication of the effect of sudden large voltage changes. Both can be visi-bly annoying and detrimental to other loads con-nected to the same branch circuit. The relativevoltage changes must be measured with a totalaccuracy better than ±8%.

Page 21: Harmonic and Flicker

21

Finding a testing solution that truly meets therequirements of the IEC 10003-2 and 1000-3-3standards can be a time-consuming and costlyprocess. This process requires full understandingof the standards, the evaluation of multiple testinstruments for compliance to the performanceand implementation criteria, hardware systemintegration, and software development. An addi-tional complicating factor can be unclear specifica-tions and poor or misleading documentation fromtest equipment vendors. The alternative turnkeytest systems are often very costly. They typicallyinclude the instruments from multiple vendors,some of which would fail to withstand technicalscrutiny if evaluated as individual components.These problems have inspired Agilent Technologiesto introduce a revolutionary testing solution specif-ically designed for testing to the new ac mains emis-sion standards called the 6840 Series Harmonic/Flicker Test System.

Procuring a solution is now simple and affordablewith the 6840 Series Harmonic/Flicker Test System.Three single phase models are available, the 6841A,6842A, and 6843A. Power levels range from 750 VAup to 4800 VA, the maximum required by the stan-dards (230 Vrms, 16 Arms). These complete solu-tions are backed by proven Agilent expertise inpower generation, precision measurement, andhigh quality products.

The Agilent 6840 Series Harmonic/Flicker TestSystem provides compliance-level testing using ref-erence methodologies for the following ac mainsharmonic current, voltage fluctuation, and flickerregulatory standards.

• IEC 1000-3-2 (1995) and IEC 555-2 (1982) • IEC 1000-3-3 (1994) and IEC 555-3 (1982) • EN 61000-3-2 (1995) and EN 60555-2 (1987) • EN 61000-3-3 (1995) and EN 60555-3 (1987)

Test equipment evaluation, system integrationtime, and equipment cost is greatly reduced withthese “One-Box” test systems. Each model is a complete solution with hardware and software,eliminating the need for a separate ac source, lineimpedance network (LIN), power analyzer, andflickermeter. This saves you the time it takes tospecify, rack mount, cable, and develop softwarefor separate instruments. System costs are greatlyreduced since you don’t have to purchase expen-sive instruments, upgrades to the required per-formance levels, or costly preconfigured multiplebox test systems.

Low distortion power generation, low and pro-grammable output impedance, and an accuratemeasurement system assure compliance-level performance. The 6841A, 6842A, and 6843A were designed according to the reference method-ologies outlined in the standards for voltage andharmonic current measurement techniques (IEC1000-3-2 and IEC 1000-4-7), flicker measurements(IEC 868), and reference impedance requirements(IEC 725). This eliminates the need to spend timeand effort comprehending the complexities of the standards and allows you to spend more timetesting your equipment.

The Agilent 6840 Series Harmonic/Flicker Test System Solution

Page 22: Harmonic and Flicker

22

Each 6840 Series Harmonic/Flicker Test System isshipped with Harmonic/Flicker Test System (HFTS)software for Windows, providing a fast and easyway to access the IEC testing capabilities. TheHFTS software provides the following capabilities:

• Test set-up and execution• Pre-test for EUT (equipment under test) class

determination (IEC 1000-3-2)• Real-time test data display (graphical and

tabular)• On-line/off-line test data review with user-

specified search criteria• Test termination under user-defined conditions• Pass/Fail indication• Diagnosis of test results via advanced features• Report generation

Basic Measurement DefinitionsThe Agilent 6840 Series Harmonic/Flicker TestSystem relies on simultaneous digitization of out-put voltage and current waveforms with subsequentdigital processing as the basis for all measurementfunctions. Voltage and current waveforms are digi-tized at a 38.4 kHz rate with the digitized samplesaccumulated into 4K data buffers.

Sampling A/D converters with 16-bit resolution areused to provide superior measurement resolutionand tight synchronization between the voltage andcurrent data records. Synchronization is maintainedto within a few nanoseconds. Synchronization ofmeasurement windows to the input waveforms isalso carefully controlled, as described later in thisdocument. In IEC measurement mode, processingof voltage and current buffers involves the use ofFIR (finite impulse response) implemented multi-rate filters to lower sampling rates to frequenciescompatible with specific IEC requirements for har-monic and flicker measurements. Options for select-ing Rectangular or Hanning windows are provided.

The computational algorithms used for basic volt-age, current, and power measurement functionsand for the windowing functions are as follows:

Page 23: Harmonic and Flicker

23

Regulatory testing presents unique challenges that are equally troublesome for vendors of testsolutions as they are for equipment manufacturersseeking to certify their products for sale into markets where demonstration of compliance isrequired. Not only are standards such as IEC 1000-3-2 and IEC 1000-3-3 difficult to understand,but careful reading reveals that what is not said is frequently as important as what is said. In addi-tion, the standards sometimes impose apparentlycontradictory requirements, and in some casescontain errors, despite the best efforts of compe-tent, dedicated people serving on the standardscommittees.

Fortunately for equipment manufacturers, most of the interpretive burden falls upon the solutionsvendor. Equipment manufacturers need to beaware of the issues and the tradeoffs that havebeen made, but generally do not need to troublethemselves with discovering and proving that aparticular interpretation results in an implementa-tion that cannot be made to work. This sectiondescribes the issues in a general way, with exam-ples that apply to both standards. More detaileddiscussions of issues specific to each standard andAgilent’s solutions for these issues may be foundunder the appropriate standard as discussed laterin this document.

Normative References and Their ScopeAs previously stated, IEC 1000-4-7 is one of thenormative references for IEC 1000-3-2, while IEC725 and IEC 868 are normative references for IEC 1000-3-3. A normative reference may refer toanother standard or it may be a self-containedannex. The “normative” designation means that the reference includes requirements that must bemet for compliance. The other possible designationfor a reference is “informative,” which means thatthe reference contains useful information relatedto the standard, but that the material is not a compliance requirement.

When interpreting the impact and scope of norma-tive references, a general rule that may be appliedis that the most specific standard carries the great-est weight when evaluating the applicability ofrequirements in normative references, particularlyif any of the requirements appear to conflict withthe more specific standard.

As an example of how this rule applies, considerthe following: IEC 1000-4-7 is a general referencethat documents requirements for instrumentationand measurement techniques to be used for assess-ing mains circuit harmonic and interharmonic signals. Interharmonics are signals occurring atfrequencies other than integer multiples of themains frequency. Since voltage and current meas-urements are both described, a plausible interpre-tation is to conclude that instrumentation used to provide compliance testing for a standard thatreferences IEC 1000-4-7 must comply with all ofthe requirements documented in IEC 1000-4-7.However, Section 4 of JEC 1000-4-7 clearly statesotherwise:

“The instrumentation may be constructed eitherto cover a particular need and application (e.g.,steady-state harmonics on the supply voltage) or to comply with the requirements for severalcases (e.g., voltage and current measurements,harmonics and interharmonics).”

Interpretive Issues for Test Solutions Vendors and EquipmentManufacturers

Page 24: Harmonic and Flicker

24

IEC 1000-3-2 describes limits for harmonic cur-rents only; voltage harmonics are not regulated by thestandard nor is there any discussion or regulation ofinterharmonics. Using the rule that the most specificstandard is given greatest weight, an instrumentthat meets just those IEC 1000-4-7 requirementsthat apply to harmonic current measurement pro-vides a fully compliant test solution for IEC 1000-3-2,assuming that it also meets all of the requirementsof JEC 1000-3-2 itself. This latter point is impor-tant. Normative Annexes A and B to IEC 1000-3-2impose, respectively, requirements for the measure-ment circuit and supply source and for the meas-urement equipment. These additional requirementsmust be met, and in cases where differences existwith respect to requirements described in IEC1000-4-7, the requirements in the annexes takeprecedence.

Stated, Implied, and Inferred RequirementsA related set of issues involves recognition of thedistinctions between stated, implied, and inferredrequirements. Consider the example of voltage distortion monitoring for IEC 1000-3-2.

Section A.2 from Annex A of IEC 1000-3-2 docu-ments a set of requirements for the quality of the mains source to be used during tests whichinclude: accuracy specifications for mains voltageamplitude and frequency, permissible harmoniccontent of the mains source voltage on a harmonic-by-harmonic basis, and the crest-factor of the volt-age waveform. The standard also states that theseconditions must be met “with the EUT connectedas in normal operation.” (EUT stands for equip-ment under test.) These are the stated requirements.

Even though it is not explicitly stated, the languagedescribed above clearly implies that the specifiedvoltage distortion should be maintained through-out the test, and this is indeed the widely acceptedinterpretation. In other words, the accepted inter-pretation is that the standard implies a requirementthat the specified voltage distortion be maintainedthroughout the test and that some evidence thatthis condition is met should be provided.

The next interpretive step is to infer that the stan-dard requires continuous monitoring of voltageharmonics throughout the test. This is absolutelynot stated, and the inference is not valid. While at first glance it may seem desirable to have thiscapability, this view does not necessarily withstandscrutiny, particularly if the function adds signifi-cant cost or complexity to the test equipment orprocess without adding any real value. Such aninterpretation on the part of the solutions vendormay also entail design tradeoffs that compromiseperformance for the task immediately at hand bydiverting instrument resources towards activitiesthat are not required. If a test scenario can bedevised that can guarantee worst-case equipment-under-test behavior, such a scenario would berepeatable, in which case simultaneous monitoring ofvoltage and current throughout the test is clearly notneeded to prove that the voltage distortion specifi-cation is met. Separate test runs will also providethe necessary proof, and the resulting efficienciesmay help optimize the value of the solution.

Page 25: Harmonic and Flicker

25

It is worth noting at this point that the perspectivesof IEC solutions vendors and that of equipmentmanufacturers seeking to demonstrate compliancemay differ considerably on inferred requirements.From the perspective of the solutions vendor, somerequirement overkill may be desirable, particularlyif there is any serious doubt about the nature of a valid interpretation. Less positively, overkill mayalso be desirable by letting a solutions vendor intro-duce a “lockout” specification that confers a com-petitive advantage or justifies a high price for thesolution.

From the perspective of the equipment manufac-turer, the entire compliance process is a cost, withthe minimization of cost being the goal. This is notto say that the manufacturer does not recognizethe appropriateness of the standard and its valuein promoting a quality mains environment. Compe-tent manufacturers seek to improve efficienciesand minimize costs wherever possible as a normalpart of conducting business wisely. Opportunitiesto minimize the cost of compliance testing includecontrolling the cost of capital expenditures for testequipment, the cost associated with maintainingtest labs and staff, and the hourly cost of conduct-ing the tests themselves.

As both a manufacturer of a broad range of equip-ment that must comply with EMI/RFI standardssuch as IEC 1000-3-2 and IEC 1000-3-3, and as avendor of test solutions for demonstrating compli-ance, Agilent is uniquely positioned to appreciatethe perspective of the equipment manufacturer aswell as that of the solutions vendor. By keeping theequipment manufacturers perspective foremost inmind, it is Agilent’s belief that the interests of thebroadest range of actual and potential customersfor EMI/RFI solutions will be best met. This meanscarefully controlling the impulse to infer require-ments that do not actually exist.

Appropriateness of RequirementsIssues of appropriateness constitute another generalconcern that frequently arises when interpretingstandards such as IEC 1000-3-2 and IEC 1000-3-3.Although not limited to this area alone, many appro-priateness issues are related to performance of themains source. This situation is in part due to thehistory of mains testing. It is only a recent develop-ment that power amplifiers have been substitutedfor use of the local utility (perhaps conditioned by impedance networks and/or crude regulators).The standards still reflect the earlier environmentwhere external disturbances to utility power hadto be monitored during a test to ensure that theresults were valid. Typically there was little or noability to correct improper source behavior and thetest was simply repeated if the source was shownto be out of specification during the test.

The language in Section 6.3 of IEC 1000-3-3 providesa good example of appropriateness related to theperformance of the mains source:

“Fluctuations of the test supply voltage during atest may be neglected if the Pst value is less than0.4. This condition shall be verified before andafter each test.” (Pst is short-term flicker severity.)

This requirement is interesting in several respects.First, it does not provide for continuous monitoringof the source during the test, a fact that has alreadybeen discussed earlier under “Stated, Implied, andInferred Requirements.” Secondly, the statementimplies that source voltage fluctuations should beexpected. This is an artifact of the history of usingutility power as the source. Third, since other lan-guage in the same section makes reference to the“open circuit voltage,” some solutions vendorshave inferred that source flicker must be verifiedon the source side of the reference impedance. It isnot clear that this is the only valid interpretation.

Page 26: Harmonic and Flicker

26

When the test solution includes a high perform-ance source such as is provided in Agilent’s 6840ASeries Harmonic/Flicker Test System, all of theconditions described in Section 6.3 may easily beguaranteed by design. In this situation, the require-ment for verifying Pst <0.4 before and after eachtest is unnecessary. In fact, not only is verificationunnecessary, but the performance of the sourceguarantees that the conditions will always be metthroughout the entire test, something that is notguaranteed by pre- and post-test verification.

Verification may still be performed, however. Since the reference impedance of the 6840 SeriesHarmonic/Flicker Test System is loop implemented(i.e., there is no physical inductor or resistor), veri-fication may be performed by disconnecting theequipment under test, which provides verificationon the load side of the reference impedance. As analternative, the reference impedance may be pro-grammed to its minimum value (0.0 + j 0.0063 ohms)with the equipment under test still connected, thusproviding verification on the source side of the ref-erence impedance. Note that this latter techniqueis very conservative since there is no impedance to limit the current drain of the equipment undertest and thus the source is potentially subjected tohigher currents than otherwise would be the case.

In either case, tests conducted by Agilent show thattypical source Pst values of 0.07 are obtained withequipment under test exhibiting failing Pst valueswell above the Pst limit of 1.0. The results are simi-lar regardless of which verification method is used.The equipment under test used for this test was a 5 V, 100 A switch-mode power supply without powerfactor correction that was subjected to a 0 to 100

ampere DC load chopped at 9 Hz. Under these con-ditions, the test unit draws 1640 VA and 800 Wattswith a mains peak current of 27 amperes at 230 VAC.The 1750 VA model 6842A Harmonic/Flicker TestSystem was used to conduct the test. The amplifierwithin the 6842A maintained source voltage THDat 0.41 % under these loading conditions.

In the immediate future, verification may be rou-tinely performed in order to meet the letter of thestandard, but the requirement is clearly inappro-priate when a properly designed high performancesource is used in place of utility power. The usermay wish to consider appropriateness when plan-ning verification tests and may reasonably expectthat this requirement will be eliminated at sometime in the future as the use of high performancesources becomes commonplace.

Technical ErrorsOne of the most difficult aspects of correctly inter-preting IEC 1000-3-2 and IEC 1000-3-3 involves thepresence of technical errors. The entire burden ofidentifying errors and documenting them for thebenefit of users properly falls upon the solutionsvendor. This is a key part of the value that a com-petent solutions vendor provides. A number of theseissues as they specifically relate to the IEC 1000-3-2and IEC 1000-3-3 standards are documented inthis application note under the appropriate stan-dard. Again, there is a need to point out that theexistence of errors is natural given the complexityof the standards and the manner in which they haveevolved; there is no implied criticism of the IECcommittees. The following example of sliding meanfilters should help illustrate the subtle nature oftechnical errors in the IEC standards.

Page 27: Harmonic and Flicker

27

Sliding mean filters are mentioned in both IEC 868,which describes requirements for flickermeters,and in working documents from SC 77A/WG01/TF02,which is an IEC committee developing future revi-sions to IEC 1000-4-7. In both cases, the languageused tends to treat “first order lowpass,” “slidingmean,” and “moving average” as interchangeableterms. The language in Section 3.9 of IEC 868 isparticularly explicit about this interchangeabilityin specifying the smoothing filter located at theinput to the statistical classifier block (Block 5) of the IEC/UIE flickermeter.

“The sliding mean operator shall have the trans-fer function of a first order low-pass resistance/capacitance filter with a time constant of 300 ms.”

Figure 20, which displays the frequency responsecharacteristics of an IIR (Infinite Impulse Response)implementation of a first order RC lowpass and asliding mean filter tuned to the same 3 dB frequency,shows why this language is especially troublesome.

For frequencies at or below the 3 dB point, the twofilters exhibit very similar although not identicalbehavior. At higher frequencies, however, the slid-ing mean filter shows a series of null responses at frequencies having periods equal to 1/N timesthe filter’s period of 838 ms where N is an integer.Note that a period of 838 ms “tunes” the slidingmean filter for a 3 dB corner frequency equivalentto that provided by the 0.3 second time constantspecified for the RC filter. The null responses pro-vide essentially “perfect” filtering of AC signals at these frequencies.

The statistical processing that follows the smooth-ing filter in a UIE flickermeter makes it difficult to achieve an intuitive understanding of how nullresponses exhibited by the sliding mean filterimpact instrument performance. Tests conductedduring the development of the 6840 Series Harmonic/Flicker Test System showed that test frequenciesclose to the null frequencies yielded improperresponses with a sliding mean filter implementa-tion, but not with an equivalent IIR lowpass design.

Figure 20. Frequency response of IIR filters versus sliding mean filters

Page 28: Harmonic and Flicker

28

For harmonic testing, the relative behavior of thetwo filter implementations is as follows: The filterin question here is the 1.5 second smoothing filterused to lowpass filter DFT outputs for fluctuatingharmonics prior to comparison with the test limits.Consider a load that exhibits one or more fluctuat-ing harmonics with a 50% duty cycle (i.e., a loadwith squarewave modulation). Assume also thatthe load fluctuates between 20% and 160% of thespecified limit for a given harmonic at a frequencyexactly equal to one of the null frequencies exhibitedby the sliding mean filter. The null response willaverage these fluctuations to a non-fluctuating levelcorresponding to 90% of the specified limit. Underthese conditions the equipment under test willpass, but with a slightly different modulating fre-quency it will fail the 2.5 minute window criteriafor fluctuating harmonics, both because individualmeasurements will exceed 150% of specificationand because greater than 10% of the measurementswill exceed 100% of the limit. The test could also bemisinterpreted as showing quasi-stationary behav-ior and a false “pass” would still occur. Other sce-narios may be devised to produce similar results.

Agilent believes that the IEC committee somehowoverlooked this aspect of sliding mean filter per-formance. For the solutions vendor, the choice is to deviate from a literal interpretation of the stan-dard in favor of a focus on “intent” which clearlydictates use of an IIR implementation. In the caseof IEC 868, this decision is made easier by the con-fusing language quoted above that would appear to make either implementation compliant. If futurerevisions to IEC 1000-4-7 or IEC 868 specify a“moving average” filter, a more difficult interpre-tive decision may be required.

Page 29: Harmonic and Flicker

29

Harmonic Current MeasurementsAnnex B of IEC 1000-3-2 and IEC 1000-4-7 jointlydescribe a number of requirements for instrumen-tation used to measure harmonic currents in acmains circuits. Sections B.1, B.2, and B.4 are appli-cable to the FFT implementation used in Agilent’s6840A Series Harmonic/Flicker test systems.

There is language in IEC 1000-4-7 that implies that certain parts of Section B.3, which describesrequirements for frequency domain instruments(i.e., wave analyzers), may also be applicable sinceit is the intent of the standards that time and fre-quency domain techniques produce equivalentresults. In particular, measurement bandwidth andattenuation of adjacent signals in the frequencydomain are defined by language in Section B.3. It turns out that using the reference grade method-ologies described in Section B.4 for fluctuating harmonics (i.e., rectangular windows without over-lap and 16 cycles of the fundamental within themeasurement window) meets the bandwidth andattenuation requirements of Section B.3. The keyrequirements for harmonic current measurementare summarized as follows:

• The total measurement error for steady-stateharmonics shall not exceed 5% of the permissiblelimits or 0.2% of the rated current of the EUT.

• For steady-state (i.e., quasi-stationary) harmon-ics the width of the measuring window shall bebetween 4 and 30 cycles of the fundamental.

• For steady-state harmonics, either Hanning orRectangular windows may be used. There is nospecification for gaps or overlaps.

• For fluctuating harmonics, either Hanning orRectangular windows may be used. The use ofRectangular windows requires no gaps or over-laps, while the use of Hanning windows requires50% overlap.

• In either case, use of Rectangular windowsrequires synchronization of the measuring win-dow to the mains frequency to 0.03% or better.

• When test limits are exceeded (i.e., when the test results are close to the limits), either aRectangular window or a Hanning window may be used with the additional requirementthat 16 (Rectangular) or 20-25 (Hanning) cycles of the fundamental shall appear in themeasuring window.

• The input impedance of the measuring instru-ment (e.g., a current shunt) shall be such thatthe voltage drop across the input circuit shallnot exceed 0.15 Volts peak.

• In the case of fluctuating harmonics, a smooth-ing filter having a response equivalent to a firstorder lowpass filter with a time constant of 1.5seconds ±10% shall be applied to the DFT outputs.

Additionally from Section B.3:

• When test limits are exceeded (i.e., when the testresults are close to the limits), an instrumentshall be used with a bandwidth of 3 Hz ±0.5 Hz.

Implementation of IEC 1000-3-2 and EN 61000-3-2 Standards

Page 30: Harmonic and Flicker

30

The 6840 Series Harmonic/Flicker Test Systemuses a base digitization rate of 38.4 kHz for allcombinations of mains frequency and windowtype. FIR (Finite Impulse Response) implementedmultirate digital filtering techniques are used tolower the sampling rates to frequencies compatiblewith acquisition of 4K data buffers having meas-urement windows (i.e., acquisition windows) con-sistent with stated requirements for the number of fundamental cycles to be contained within themeasurement window.

Note that the need to perform this sample ratereduction in order to meet the requirements of thestandard completely negates claims made by somesolutions vendors that 100K+ sample/second digiti-zation rates are required to perform IEC 1000-3-2tests. What is required is careful design of anti-aliasing filters to ensure that adequate rejection of signals past the Nyquist limit is obtained with-out compromising measurement accuracy in thedc-to-2 kHz passband.

A 4K FFT (Fast Fourier Transform) is performedregardless of the selection of mains frequency andwindow type. Note that an FFT technique is usedto obtain harmonic data and that this technique is fully compatible with requirements for use of a DFT (Discrete Fourier Transform) despite someconfusing of this issue by other solutions vendors.An FFT is simply an algorithm for rapidly comput-ing a DFT that meets certain constraints; the cal-culation is still a DFT. The relationships betweenselection of mains frequency and window type andthe dependent variables of sampling rate, decima-tion rate, measurement (acquisition) window peri-od, overlap, record (data) rate, fundamental binnumber, bin spacing, and 3 dB bin bandwidth aresummarized in the tables below.

6841A 6842A 6843A

Fundamental (Low Range) 0.03% + 1.5 mA 0.03% + 1.5 mA 0.03% + 3 mAHarmonics 2-40 (Low Range) 0.03% + 1 mA + 0.2%/kHz 0.03% + 1 mA + 0.2%/kHz 0.03% + 2 mA + 0.2%/kHzFundamental (High Range) 0.05% + 5 mA 0.05% + 5 mA 0.05% + 6 mAHarmonics 2-40 (High Range) 0.05% + 3 mA + 0.2%/kHz 0.05% + 3 mA + 0.2%/kHz 0.05% + 3 mA + 0.2%/kHz

Agilent 6840 Series Measurement AccuracyThe harmonic current measurement accuracy forthe 6840 Series Harmonic/Flicker Test System issummarized in the following table:

Mains Window Sample Decimation AcquisitionFrequency Type Rate Rate Window Overlap

50 Hz RECTangular 12.800 kHz 3X 320 ms None50 Hz HANNing 8.533 kHz 4.5X 480 ms 50%60 Hz RECTangular 15.360 kHz 2.5X 266.7 ms None60 Hz HANNing 7.680 kHz 5X 533.3 ms 50%

Mains Window Record Bin Number Bin 3 dBFrequency Type Rate (F1) Spacing Bandwidth

50 Hz RECTangular 320 ms 16 at 50 Hz 3.125 Hz 2.8 Hz50 Hz HANNing 240 ms 24 at 50 Hz 2.083 Hz 3.0 Hz60 Hz RECTangular 266.7 ms 16 at 60 Hz 3.750 Hz 3.3 Hz60 Hz HANNing 266.7 ms 32 at 60 Hz 1.875 Hz 2.7 Hz

Page 31: Harmonic and Flicker

31

The bin number for the fundamental is the same asthe number of cycles of the fundamental frequencycontained within the measuring (acquisition) win-dow. Note that for 60 Hz with a Hanning window,the number of fundamental cycles within the win-dow is 32, which would appear to be a violation ofthe requirements of the standard to have 20 to 25cycles within the acquisition window when using a Hanning window. In Agilent’s view, this is eithera minor technical error within the standard or areflection of the IEC committee’s focus on 50 Hzsystems. The possible integer range for fundamen-tal cycles consistent with a 3 Hz ±0.5 Hz bandwidth(as described in Section B.3 of IEC 1000-3-2) at amains frequency of 60 Hz is 25 to 34. A very difficultimplementation decision would have been requiredto provide a 25 cycle window and this would haveproduced a bandwidth of 3.46 Hz, which wouldhave been at the upper limit of the stated range forbandwidth. Given this situation, a decision wasmade to select 32 cycles with a resulting bandwidthof 2.7 Hz. Agilent’s design objective was to use ref-erence level methodologies in all cases. The Rectan-gular window with 16 cycles is an option if literalcompliance is demanded. This would in any casebe the proper and recommended choice for thosesituations where measured values are close to thespecified limits.

Measurement Synchronization UsingRectangular WindowsOne of the unique performance advantages pro-vided by Agilent’s fully integrated solution is elim-ination of concern about synchronization whenusing Rectangular windows. Because the outputwaveform generation and the measurement sys-tems share a common clock, synchronization tobetter than 0.25 ppm, or 0.000025%, is maintainedcontinuously. In addition, the transient eventscaused by the response time of the phase- or fre-quency-locked loops used by alternative solutionscan never occur.

Although the technique just described completelysolves the synchronization issue for quasi-stationaryharmonic tests, neither it nor any other synchro-nization method completely solves the problemsbrought about by fluctuating harmonic signals.

Synchronization is a concern when Rectangularwindows are used because the extremely poor side-lobe suppression provided by the window permitsspectral leakage from nearby bins that can severelycorrupt magnitude measurements. This effect iscompletely eliminated if perfect synchronization ismaintained, which is another way of saying that allfrequencies present must be harmonics of a basefrequency that has an exact integer number ofcycles within the measurement window. Normallythis base frequency has exactly one cycle withinthe acquisition window and is equal to 1/TW whereTW is the width of the measurement window. Thesefrequencies correspond to the bin spacing and areshown in the previous tables. The requirement for16 cycles of the fundamental within the measure-ment window for IEC 1000-3-2 measurements mod-ifies, but does not undo, the logic of this constraint.Unfortunately, fluctuating harmonics are virtuallyguaranteed to produce many nonharmonicallyrelated frequency components which do not meetthis criterion.

The requirement for use of either Rectangular orHanning windows, with an apparent preference forRectangular windows, may reflect the IEC commit-tee’s awareness that both of these windows, whenused with the overlap specified for reference grademeasurements, provide unity weighting for eachinput sample. Despite this unique characteristic(the Triangular window is the only other commonlydescribed window with similar properties), it isgenerally accepted that other windows offer superiorperformance for spectral estimation. Requirementsfor use of Rectangular windows in the presence offluctuating harmonics may thus be seen as a fun-damental technical issue that impacts the perform-ance of any IEC 1000-32 test solution.

Page 32: Harmonic and Flicker

32

The following three figures illustrate the differencesin sidelobe performance between Rectangular,Hanning, and KBessel windows.

The IEC committees have been advised of theseconcerns, and recommendations have been madeto the committees for use of other windows thatexhibit superior sidelobe performance. Unless thecommittee concurs and elects to modify the windowrequirements, solutions vendors have little choiceother than meeting the letter of the standard.

Figure 21. Sidelobe performance for Rectangular windows

Figure 23. Sidelobe performance for KBessel windows

Figure 22. Sidelobe performance for Hanning windows

Page 33: Harmonic and Flicker

33

Shunt BurdenThe voltage drop across the measuring instru-ment input is 0.0 volts for the Agilent 6840 SeriesHarmonic/Flicker Test System because the regulat-ing loop for the source amplifier senses voltage at the load and thus corrects for any drop acrossthe internal measuring shunt.

1.5 Second Smoothing FilterAn IIR type lowpass filter with a time constant of 1.5 seconds is used to smooth outputs wheneverfluctuating harmonic measurements are selected.The time constant for this filter is controlled tobetter than 0.1%. It may be noted at this point thatsliding mean filters, in addition to exhibiting nullresponses at certain frequencies as previously dis-cussed under “Technical Errors,” also suffer thedrawback of having coarse tuning resolution rela-tive to IIR implementations in digital systemswhen the filter corner frequencies and samplingfrequencies are close to one another. This is thecase for filtering DFT outputs from IEC 1000-3-2tests. The effect occurs because the length of thefilter becomes shorter as the comer frequencyapproaches the sample rate. This in turn meansthat integer steps in filter length have a relativelylarge impact on tuning. Please note that “close” in this case means ratios of 1:30, not close in thesense that the Nyquist limit is approached.

Source ImpedanceUsers should be aware that source impedance candramatically affect the magnitudes of harmoniccurrents. This situation is sometimes used to anadvantage where a marginally passing EUT can be made to pass solidly by slightly increasing thesource impedance. More commonly, source imped-ance variations from one source to another causemeasurement correlation problems when differenttest systems are used. This variability can be par-ticularly frustrating when a marginal EUT unex-pectedly fails a test it was thought to pass.

Several things should be kept in mind regardingsource impedance. First, users frequently questionthe accuracy of the measuring equipment whensource impedance variability is the real culprit.Secondly, it is unreasonable to expect exact corre-lation between test systems, particularly if differ-ent mains sources are used; lower source imped-ance will produce higher harmonic currents. Third,source impedance is a function of frequency. Thismeans that identical results may be obtained forsome harmonics, even with different sources, butother harmonics may vary considerably. This willbe especially true for higher harmonics. Fourth,high source impedance will cause voltage distortion.A harmonic current working against the sourceimpedance will produce a corresponding harmonicvoltage. IEC 1000-3-2 indirectly controls sourceimpedance by limiting the permissible levels of out-put voltage harmonics as documented in Annex Aof the standard.

Lastly, IEC 1000-3 -2 contains in Figure A.1, thevague statement that source impedance ZS is notspecified, but must be “sufficiently low to suit thetest requirements.” There is growing recognitionthat the near-ideal source impedances provided by high performance sources do not accuratelyreplicate typical mains environments. There hasbeen some discussion of adding a requirement fora LIN (Line Impedance Network) that would inserta known source impedance into the measurementcircuit. To date no official action has been taken on this issue in Europe although some countries,notably Japan, have added LIN requirements forharmonic current testing.

Presently, Agilent’s design decision is to set theoutput impedance to the lowest possible level dur-ing harmonic tests, but because output impedanceis a programmable loop-based implementation in the 6840 Series Harmonic/Flicker Test System,future changes may be easily accommodated viaupgrades to the software package included withthe system.

Readers interested in this issue may wish to con-sult IEC 725, which contains the results of a studyof mains impedances in most European countries.This study formed the basis for selection of the values used for the reference impedance that isrequired for flicker testing and may be used as well for justifying inclusion of a LIN for harmoniccurrent testing.

Page 34: Harmonic and Flicker

34

The Pre-Test Window for Class DeterminationAgilent’s HFTS software application, which is partof the 6840 Series Harmonic/Flicker Test Systemsolution, provides a Pre-Test window that containscontrols and displays to facilitate determination of EUT class as well as providing other usefulinformation about EUT and system performance.Frequent references will be made to functions pro-vided in this window. The following figure showsthe Pre-Test window, included here to help thereader visualize how these functions may be usedas part of a compliance test. As its name implies,the Pre-Test window centralizes functions that, following test setup, need to be performed prior to running an actual test. This includes runningpreliminary tests to measure quantities that aresubsequently used to set test limits for Class C and Class D devices.

Test limits for Class C devices are a function of thepower factor (PF) and fundamental current drawnby the EUT, while for Class D devices limits are afunction of active power (i.e., Watts) drawn by theEUT. Agilent’s general design philosophy has beento provide flexible tools in the Pre-Test and Set-up

windows that simplify measurement of the neces-sary parameters, confirm user setup decisions, andpermit overriding of pre-test results if required.The basic assumption is that users will performthese tasks prior to running the actual compliancetest in order to establish uniform test limits for use throughout the subsequent test.

Agilent has taken this position rather than aninferred requirement that for Class D testing inparticular, the current waveform must be evalu-ated continuously for each measurement windowduring a harmonic current test, with class deter-mination and test limits calculated simultaneouslyand uniquely for each measurement window. Whileit may indeed be possible to assemble instrumen-tation and analysis software to support such aninterpretation of the standard, it is difficult to knowexactly what value a mass of test data with con-stantly varying pass/fail limits would be to users.This is particularly true for those attempting todevelop a diagnostic perspective on a failed test so that corrective modifications to the EUT mightbe made.

Figure 24. Pre-test window

Page 35: Harmonic and Flicker

35

It is appropriate to ask, what benefit does an extremeand costly interpretation of the standards provideto equipment manufacturers seeking product certi-fication, to solutions vendors seeking to profit byproviding test solutions for the equipment manufac-turers, or to governmental and quasi-governmentalbodies seeking to regulate the mains environment?Equipment manufacturers clearly do not benefitfrom extreme interpretations that complicate thejob of getting products over the regulatory hurdles.Solutions vendors may in some cases find moreextreme interpretations to be in their interest, butas the resulting requirements become ever moreonerous and the pursuit of “lockout” specificationsbecomes more obvious, the motives of these solu-tions vendors become suspect. Regulatory agenciesand quasi-governmental bodies such as the powerutilities, the drivers behind the standards, seek topromote a “clean” mains environment that mini-mizes interactions between their customers’ loads.They also wish to promote public safety and trans-mission efficiencies by discouraging low power factor loads. It is not at all in the interest of theseagencies to promote standards that are so strin-gent that industry either ignores the standards oris aroused to lobby against them.

Class D Active Power Tests and the SpecialWaveshape RequirementCertain devices, particularly switch-mode powersupplies using off-line conversion technologies,may draw high peak current through a small con-duction angle near the peak of the mains voltagewaveform. If this characteristic is sufficiently pro-nounced, IEC 1000-3-2 requires that the EUT betested as Class D device. Harmonic current limitsfor Class D devices are a function of the activepower drawn from the mains circuit (i.e., Watts),and generally will be lower than those provided forClass A devices. The more stringent requirements

for Class D reflects the higher harmonic content of the current waveform typical for these devicesand a corresponding desire on the part of regulatorybodies to control these emissions. Note that allEUT’s that are not portable tools (Class B) or light-ing devices (Class C) must be either Class A orClass D devices.

The following criteria must be met before the EUTis classified as a Class D device:

1. The input current waveform must have the “special shape” defined in Figure 1 of IEC1000-3-2.

2. The EUT must not be “... motor driven equipmentwith phase angle control ...”

3. The active input power must ≤600 W.

4. For the Class D mA/W limits to apply, the activepower must also be greater than 75 W. Thislower limit will be reduced to 50 W within fouryears of the implementation of IEC1000-3-2. As noted earlier, the standard is ambiguous asto whether or not the “maximum permissibleharmonic current” limits given in Table 3 of the standard apply to devices with active powerbelow these thresholds. Agilent takes the posi-tion that these maximum limits do apply, butthe user is free to interpret the standard as notrequiring a test if this condition is met.

The requirements, while appearing straightforward,actually require some interpretation and consider-able care on the part of the test engineer duringtest setup.

Page 36: Harmonic and Flicker

36

Note that Figure 1 from IEC 1000-3-2, which isduplicated in Figure 14, contains an ambiguity that requires interpretation. The vertical axis is in units of i/ipk, which normalizes the waveform to the peak value and takes the absolute value ofthe waveform as long as i and ipk are equal in sign.So far, so good! If, however, a small DC componentdue either to EUT behavior or to offsets in themeasurement circuit causes a significant portion of the waveform, typically in the first or last thirdof the half-cycle, to be opposite in sign to the peakvalue in the corresponding half-cycle, the require-ment that at least 95% of waveform fall within thewindow will not be met even though the waveformhas characteristics that otherwise qualifies the EUTfor Class D treatment. This does not seem to be avalid interpretation and Agilent’s implementationinstead takes the absolute value of the ratio of i to ipk.

The current waveform peak is required to be cen-tered in the Class D envelope for proper evaluationof the 95% criterion. Fitting the current waveformto the Class D envelope may be problematic if anoisy and relatively flat-topped current waveformbecomes sufficiently offset in time relative to theenvelope to cause misinterpretation. Agilent’simplementation makes use of several filtering algo-rithms to assist in proper location of the peakwithin the waveform record which is then analyzedwithout filtering. As a final check, the waveform is actually displayed at high resolution in the Pre-Test window along with the Class D envelope. Thisenables the operator to visually detect and reject a false test result should one occur.

A second issue, once a Class D determination hasbeen made and confirmed, is the requirement tomeasure active power in order to set the test lim-its. The definitions section of IEC1000-3-2 definesactive power as: “The mean value, taken over oneperiod, of the instantaneous power.”

Agilent takes this definition as a statement describ-ing a mathematical relationship between instanta-neous values and the range of integration for ameaningful integral of those values, not as a literalstatement that ranges of integration must only beone mains period.

Agilent’s view is that a tool that integrates powerover a user-selected interval of time provides themost appropriate means for measuring active power.This permits the user to adjust the test based onknowledge of the EUT’s behavior in manner thatbest captures its performance under typical usagescenarios. The Pre-Test window provides just sucha tool. A “Pre-Test Duration” control permits theuser to select pre-test durations ranging from 100 msto 7 days. Power is averaged for an integer numberof mains cycles equal to the duration of the pre-test.This technique provides the user with an accuratemeasure of the behavior of the load integrated overa period of his choosing. Rms voltage and currentare also integrated over this same period, again foran exact integer number of mains cycles. Variouscalculations for derived parameters such as VA,power factor, etc. are then made and all of thisinformation is displayed for the user.

Page 37: Harmonic and Flicker

37

At the end of the pre-test interval, the currentwaveform is digitized, filtered, analyzed, and pre-sented against the Class D envelope. A calculationfor percent within the envelope is made and theresults displayed. The user’s tentative class selec-tion made previously in the test setup screens ischecked against the results of this calculation andmeasured active power. The user’s selection is thenverified within a window that displays the actualclass with a green background indicating a con-firmed choice and a red background indicating an invalid choice.

Finally, the user is given one additional control inthe advanced setup options screen where, at theuser’s discretion, a choice can be made to overridethe default mode of automatic Class D limit settingbased on the measured active power and insteadmanually enter the active power to be used in set-ting limits. Use of this control gives an indicationthat the test is potentially invalid, but gives the useran additional level of control that may aid diagnos-tic activities or ease testing of an otherwise prob-lematic EUT.

Laser printers are frequently mentioned as “problem”EUT’s. This occurs because these devices typicallyexhibit both Class A and Class D behavior depend-ing upon their operating mode. When the fuserheater is on, or in some instances when the deviceis in print mode, Class A waveshapes are present,but in standby mode Class D behavior is usuallyobserved. A case has been made by some observersthat this behavior means that laser printers cannotbe tested unless instantaneous class changes arepermitted during the test. Agilent believes that care-ful use of the long-term integration tool providedin the Pre-Test window in the context of a Class Dtest will provide a test that is both reasonable andcompliant. Note that the objection that momentaryClass A behavior will cause certain failures giventhe tighter limits imposed by Class D testing doesnot hold up; the currents drawn during these peri-ods are primarily fundamental currents and there isno limit specified for fundamental current in any class.

Class C Power Factor and Fundamental CurrentMeasurementThe controls and displays provided in the Pre-Testwindow may also be used to set limits for Class Cdevices. As mentioned previously during the dis-cussion of Class D devices, rms voltage, rms cur-rent, and active power (Watts) are integrated overthe entire duration of the pre-test. Since VA and PF are derived parameters that are calculated fromthese three basic parameters, the same benefits of user-controlled, long-term integration of devicebehavior are provided. In addition, a control per-mitting the test engineer to override automatic useof measured power factor in setting test limits isprovided in the advanced options setup window.

Class B Harmonic Current LimitsAs noted above, Tables 1, 2, and 3 in IEC 1000-3-2provide limits for permissible harmonic currentsrespectively for Class A, Class C, and Class Ddevices. Section 7.2 of IEC 1000-3-2 contains thefollowing language:

“For Class B equipment, the harmonics of theinput current shall not exceed the maximum permissible values given in Table 1 multiplied by a factor of 1.5.”

In addition, Section 6.2.2 of IEC 1000-3-2 containsthe following language regarding transitory har-monic currents:

“a) harmonic currents lasting for not more than10 s when a piece of equipment is brought intooperation or is taken out of operation, manuallyor automatically, are disregarded.

b) the limits in Tables 1 to 3 are applicable to allother transitory harmonic currents occurringduring the testing of equipment or parts of equip-ment, in accordance with Annex C.

However, for transitory even harmonic currentsof order from 2 to 10 and transitory odd harmoniccurrents of order from 3 to 19, values up to 1.5times the limits in Tables 1 to 3 are allowed foreach harmonic during a maximum of 10% of anyobservation period of 2.5 min.”

Page 38: Harmonic and Flicker

38

One solutions vendor has interpreted the standardto exclude use of the 1.5 multiplier for fluctuatingharmonics when testing Class B equipment. Agilentdoes not find justification for this interpretation inthe language, and has therefore elected to providethe multiplier for Class B equipment.

Voltage Distortion MeasurementA discussion about the issues surrounding meas-urement of voltage distortion during harmonic current tests was provided under “Stated, Implied,and Inferred Requirements.” As previously noted,neither IEC 1000-3-2 nor IEC 1000-4-7 contains anyexplicit statement that provisions for monitoringvoltage distortion must be included in the instru-mentation intended for IEC 1000-3-2 testing. Ifinferences about measurement of source perform-ance are to be made, the most logical inference isto conclude that the use of a mains source withperformance guaranteed by design, as is the casewith Agilent’s 6840 Series Harmonic/Flicker TestSystem, is the preferred approach and would elimi-nate the necessity for source distortion monitoring.This point notwithstanding, Agilent’s implementa-tion does include the means for monitoring voltagedistortion, but in the interests of cost-effectivenessdoes not conduct these tests simultaneously withharmonic current measurements.

During the pre-test, voltage distortion is continu-ously monitored on a harmonic-by-harmonic basiswith the highest measured values subsequentlycompared to the limits given in Annex A of IEC1000-3-2 for voltage harmonics. An indicator con-trol labeled “Voltage THD” directly beneath the current waveform display shows either “IN SPEC”or “OUT OF SPEC” depending on the outcome ofthis comparison. Note that there is also a “VoltageTHD” meter display near the bottom right comer of the Pre-Test window. This control displays meas-ured Total Harmonic Distortion, but does not reflectthe harmonic-by-harmonic comparison that drivesthe other display.

Agilent’s recommended procedure is to conduct apre-test of sufficient duration to include all antici-pated EUT behavior. This method ensures that volt-age distortion performance is continuously verifiedunder actual test conditions, and provides at thesame time the integrated measurements used toestablish test limits as described above for Class Cand Class D devices. Note also that pre-test resultsmay be copied to new filenames. This permits asingle pre-test to serve an entire sequence of maintests, a procedure that is both valid and efficientwhen running a series of tests on a single EUT.

Delay to Start of TestSection 6.2.2 of IEC 1000-3-2 contains languagethat states:

“... harmonic currents lasting for not more than10 s (seconds) when a piece of equipment is broughtinto operation or is taken out of operation, man-ually or automatically, are disregarded; ...”

This language is advantageous to most equipmentmanufacturers because it permits the “inrush” cur-rents that typically occur when a piece of equipmentis first powered-up to be excluded from the test.Agilent’s solution includes a control in the advancedsetup screen that sets the delay from application ofmains power to the EUT to the beginning of meas-urement. The default value for this delay is 10 sec-onds, but it may be set anywhere within of range of 0 to 10,000 seconds. Users should be aware thatsettings greater than 10 seconds make the test non-compliant, but this additional flexibility may bedesirable in diagnostic situations.

A similar control is provided in the Pre-Test windowto set equivalent delays for pre-tests.

Page 39: Harmonic and Flicker

39

Programmable Current LimitsThe “inrush” current phenomenon previouslydescribed can create “protection faults” if the peakcurrent levels are high enough to trigger protectioncircuits contained within the source portions ofthe 6840 Series Harmonic/Flicker Test System.Some of these circuits provide “latching” operationthat must be explicitly “cleared” and which conse-quently will abort test runs.

All high power solid state sources must includeprotection circuitry to prevent self-destruction inthe presence of uncontrolled load behavior. Properuse of additional controls provided in the Agilentsolution can guarantee orderly start-up of EUT’sand avoid protection induced test aborts. Note thatthis type of load behavior will normally not be seenwith compliant EUT devices because inrush currentis regulated directly by the Dmax specification inIEC 1000-33, as well as indirectly by IEC 1000-3-2,and thus should not occur at levels capable of generating protection faults.

The “delay to start of test” or measurement delaycontrols described in the section above may beused in conjunction with current limit controlsprovided in the advanced setup options screen to control inrush current to levels below those atwhich triggering of latching protection circuits will occur. Current limiting normally generatesdetectable errors that invalidate compliance testssince voltage regulation is overridden when cur-rent limiting occurs. During the “delay to start of

test” portion of both pre-tests and main tests, cur-rent limiting events are ignored. However, once the actual acquisition of measurement data begins,current limit events are again treated as errorsthat invalidate the test. Since high inrush currentsusually occur only during the first few mains cyclesfollowing application of mains voltage, a properlyset current limit threshold will not be crossed oncethe “delay to start of test” period has ended. Thisconstructive interaction between the test systemand EUT permits the test engineer to limit loadcurrents without invalidating test results.

Both “rms” and “peak” current limits are providedfor the Agilent 6841A and 6842A models. The 4800VA 6843A model provides “rms” current limitingonly, in part due to use of a different circuit topol-ogy that inherently limits susceptibility to damagefrom excessive peak currents. The “peak” currentlimit provides the most useful control for limitinginrush currents since this loop has wide bandwidth.The “rms” limit controls may be used to provideprotection to the EUT in the event that it experi-ences a circuit failure during a test, but care shouldbe taken to pick a setting that provides sufficientmargin to avoid unintended limiting during tests.These events will be detected should they occur,but as previously noted, they will also invalidatetests. Default values for all current limit controlsare automatically set to the system’s maximum ratings unless overridden by the test engineer.

Page 40: Harmonic and Flicker

40

Flicker Measurement—UIE FlickermeterIEC 1000-3-3 states that the reference methodologyfor measuring flicker requires use of a UIE flicker-meter. The UIE (International Union for Electro-Heat) is a quasi-governmental body based in Europethat was involved in early efforts to coordinateactivities in individual European countries withthe intent of developing internationally accepteddefinitions for flicker and for instrumentationintended for flicker measurement. The portion ofthis work directed towards definition of measuringinstrumentation eventually was published as IEC868. This document defines flickermeter implemen-tations and is a normative reference in IEC 1000-3-3.

A flickermeter is fundamentally a specialized AM modulation analyzer designed to operate withmains frequency carrier inputs. The instrument’soutput is calibrated in terms of human perceptionof the flicker, or light intensity variations, inducedin a 60 W coiled filament incandescent lamp oper-ated at 230 VAC in a 50 Hz mains system. A func-tional block diagram of a flickermeter as defined in IEC 868 is shown below. The heart of a flicker-meter consists of Blocks 2, 3, and 4 which takentogether comprise a simulation of the lamp-eye-brain system’s response to rms voltage variationsthat are induced in mains circuits by varying cur-rent drains operating against branch circuit imped-ances. Block 1 consists of scaling circuitry and anautomatic gain control function that normalizesinput voltages to Blocks 2, 3, and 4, while Block 5consists of a statistical classifier the output of

which is a probability density function used tocompute short-term flicker severity, or Pst. Theoverall instrument is calibrated so that an outputof 1.00 from Block 4 corresponds to the “referencehuman flicker perceptibility threshold” while aPst value of 1.00 derived from the outputs of Block5 corresponds to the “the conventional thresholdof irritability” per IEC 1000-3-3.

RMS MeasurementsIn 50 Hz mains systems, the output of Block 1 of the flickermeter is a 100 sample/second datastream consisting of rms voltage measurementsmade continuously (without gaps) on a half-cycleby half-cycle basis. In a 60 Hz mains system, thecorresponding data rate is 120 samples/second;however, it should be noted that at present there is no flicker standard for 60 Hz systems. Note that Agilent’s implementation does include func-tionality for 60 Hz systems. When standards for 60 Hz systems are written, it may require modi-fication to the transfer functions in Blocks 2, 3,and 4.

The rms voltage data stream within Block 1 may be taken before the automatic gain control func-tion and processed to extract the “relative voltagechange” characteristic that is the basis for evaluat-ing EUT performance against limits for Dmax, Dc,and Dt per IEC 1000-3-3.

Implementation of IEC 1000-3-3 and EN 61000-3-3 Standards

Figure 25. Flickermeter block diagram

Page 41: Harmonic and Flicker

41

Pst ClassifierVarious implementations for the classifier withinBlock 5 are described in IEC documents. IEC 868Amendment 1 describes a logarithmic classifierthat eliminates the need for the ranging functional-ity shown at the output of Block 3, and which alsomitigates the need for “zero extrapolation” whenrelatively low flicker levels are measured. Agilent’simplementation uses this technique with a 1024 binclassifier that covers 6 decades of flicker percepti-bility ranging from 0.01 to 10,000 perceptibilityunits. The classifier “bins” individual flicker meas-urements by level which results in a 1024 pointarray. This array contains the accumulated totalcounts of flicker levels corresponding to each bin’s“weight” in perceptibility units that have occurredduring the Pst integration period. IEC 868 statesthat this “observation” period may be 1, 5, 10, or15 minutes. IEC 1000-3-3 specifies that the integra-tion period will be 10 minutes.

Agilent’s HFTS software application, included with the 6840 Series Harmonic/Flicker Test System, further processes the 1024 point array to producea cumulative probability function that gives theflicker levels (or percentiles) associated with prob-abilities of occurrence corresponding to 0.1, 1.0, 3,10, and 50% of the time during the observationperiod. These values are then used to calculate Pstby means of the following equations specified byIEC 868 Amendment 1:

The “s” subscript indicates use of “smoothed” valueswhich are calculated from the cumulative probabil-ity function as follows:

P50s = (P30 + P50 + P80) /3

P10s = (P6 + P8 + P10 + P13 + P17)/5

P3s = (P2.2 + P3 + P4)/3

P1s = (P0.7 + P1 + P1.5)/3

IEC 1000-3-3 specifies that a flicker test shall con-sist of 12 Pst integration periods. The Pst valuesfrom each integration period are used to calculatelong-term flicker severity (Plt) according to an addi-tional equation specified by IEC 868 Amendment 1:

Agilent 6840 Series FlickermeterThe fully compliant flickermeter embedded in the6840 Series Harmonic/Flicker Test System makesIEC 1000-3-3 testing for flicker a very straightfor-ward process. Agilent’s implementation provides a Pre-Test window similar to the one used for IEC1000-3-2 tests. A pre-test must be run prior to running main tests, but the information gatheredduring the pre-test is presented mainly as a confi-dence check for proper EUT operation. None of the acquired data is used to set test limits nor isthere any possibility that the specified source dis-tortion limit of 3% THD will be exceeded while testing compliant EUT’s. As with IEC 1000-3-2 testing however, there are several technical andinterpretive issues surrounding IEC 1000-3-3 testing and IEC 868 that merit discussion.

Pst = Ï0.0314P0.01 + 0.0525P1s + 0.0657P3s + 0.28P10s + 0.08P50s

Page 42: Harmonic and Flicker

42

Use of a Simulated ImpedanceIEC 1000-3-3 requires use of a “reference impedance”that is placed between an ideal source and the EUT.For single phase 230 V, 50 Hz systems, line andneutral impedances may be lumped and are speci-fied to have a value in total of 0.4 + j 0.25 ohms.This value represents an internationally agreedupon “typical” branch circuit impedance acrosswhich flicker voltages are developed in response to EUT current drains. Voltage is sensed at themains input terminals of the EUT and flicker volt-ages are thus properly observed.

Section 6.4 of IEC 1000-3-3 states:

“For equipment under test the reference imped-ance, Zref, according to IEC 725, is a conventionalimpedance used in the calculation and measure-ment of the relative voltage change “d” and thePst and Plt values.”

Agilent’s implementation uses multiple feedbackloops to provide a loop implemented source imped-ance having programmable resistive and inductivecomponents. The nominal range of values for theresistive component is 0 to 1.0 ohms and for theinductive component is 20 to 1000 mH. The j 0.25inductive component required by IEC 1000-3-3 corresponds to 796 mH. Remote sensing at the EUTinput terminals causes the total mains circuit imped-ance to equal the programmed values.

Typical IEC test solutions make use of discretepassive components to implement the referenceimpedance. There are no statements in the stan-dard that specify implementation details for thecomponent parts of the reference impedance. It isAgilent’s belief that the word “conventional” citedabove merely refers to the existence of a mutualagreement among the parties involved in the dis-cussions that are extensively documented in IEC

725. A loop implementation has the decided advan-tages of being dissipationless and programmable,with the latter capability opening the possibility ofefficient accommodation of differing national stan-dards for reference impedances. Agilent believesthese advantages will lead to widespread adoptionof loop-based implementations as use of high per-formance sources becomes more common.

Section 6.2 of IEC 1000-3-3, which states measure-ment accuracy requirements, contains a require-ment for an “overall accuracy” of ±8%. Agilent’sdesign philosophy was to allocate 3% to the refer-ence impedance while reserving the remainder ofthe permissible error, in accordance with statedrequirements, for other sources. This approach isconsistent with general allowances for total meas-urement error given in IEC 868, and provides aconservative overall error budget especially consid-ering that errors of <1% for both the resistive andinductive components measured independentlywere documented during design evaluations ofAgilent’s loop implemented impedance. Worst caseperformance of ±3% performance is guaranteed.

Squarewave Response Tests per IEC 868IEC 868 includes tables containing provisions for“overall analog response” for sinusoidal and rec-tangular (i.e., squarewave) voltage fluctuations(i.e., modulations). Tables 1 and 2, respectively,prescribe modulation levels for modulating fre-quencies ranging from 0.5 Hz to 25 Hz that corre-spond to outputs of 1.00 from Block 4. The actualmodulation levels producing outputs of 1.00 arerequired to be within ±5% of the values shown inthe tables.

Page 43: Harmonic and Flicker

43

An extensive discussion of the performance ofAgilent’s flickermeter implementation is given inAppendix A. However, there is one issue related to performance with squarewave modulation thatis more fully discussed here.

Complying with values given for sinusoidal modu-lations is straightforward given a proper imple-mentation of Blocks 2, 3, and 4. Requirements forrectangular modulations are another matter. Thetest requirements given in Table 2 are impossibleto meet unless the modulating signal is lowpass filtered, prior to modulation, by a filter having theequivalent of a single pole RC characteristic withthe corner frequency set to approximately 100 Hz.Filtering is necessary because the higher orderharmonics present in squarewave signals producemultiple responses within the passband of theweighting filter in Block 3, and are not properlyaccounted for in Table 2’s prescriptions when modulating frequencies are above 15 Hz.

The fifth harmonic of a 22 Hz squarewave modulat-ing signal is particularly troublesome in this respect.Considering for a moment this 110 Hz frequencycomponent alone, modulation of a 50 Hz carrierproduces upper and lower sidebands at Fc ±Fmod

(i.e., at 60 Hz and 160 Hz). The component at 60 Hzis the lower sideband “folded back” around DC sothat it appears reversed in phase at 60 Hz. This 60 Hz component, when detected by the squaringdemodulator, appears at 10 Hz, which is nearly atthe response peak of the weighting filter containedwithin Block 3. Despite the fact that this detectedcomponent is considerably lower in magnitudethan the detected 22 Hz fundamental component,the relative response of the weighting filter at thetwo frequencies produces sufficient output fromthe fifth harmonic to exceed the limits specified in Table 2.

Figure 26. Carrier and modulating components (50 Hz carrier with 1%, 22 Hzsquarewave modulation)

Figure 27. Detected modulating components from Figure 26, prior to weightingfilter in Block 3

Page 44: Harmonic and Flicker

44

It is always risky to speculate on why an oversightof this type found its way into a document such asIEC 868, particularly since squarewave response is correctly tabulated at lower frequencies. Despitethis, one guess may be that at the time this workwas done, the relatively limited bandwidths avail-able in mains sources may have suppressed anyobvious manifestations of higher harmonic responsesfor modulating frequencies above 15 Hz. A recentIEC publication, 77A(SEC)113, which discussesproposed revisions to IEC 868, includes modifica-tions to Table 2 which appear to account for theeffects described above.

Unclear Definitions for Dmax, Dc, and DtIn Agilent’s view, the most difficult interpretiveissue in IEC 1000-3-3 involves Definitions 3.1through 3.4, Figures 2 and 3, and requirements for evaluating Dmax, Dc, and Dt. An edited versionof Figure 3, which shows the generally acceptedinterpretations for these parameters, is shownbelow as is the complete text of Definitions 3.1through 3.4.

“3.1 R.M.S. voltage shape, U(t): The time functionof the r.m.s. voltage evaluated stepwise over suc-cessive half-periods of the fundamental voltage(see Figure 2).

3.2 voltage change characteristic, ∆(t): The timefunction of the change in the r.m.s. voltage betweenperiods when the voltage is in a steady-state condition for at least 1 s (second) (see 4.2.3 andFigure 2).

3.3 maximum voltage change, ∆Umax: The differ-ence between maximum and minimum r.m.s. values of the voltage change characteristic (see Figure 2).

3.4 steady-state voltage change, ∆Uc: The differ-ence between two adjacent steady-state voltagesseparated by at least one voltage change charac-teristic (see Figure 2).

NOTE – Definitions 3.2 to 3.4 relate to absolutephase-to-neutral voltages. The ratios of these magnitudes to the phase-to-neutral value of thenominal voltage (Un) of the reference network in Figure 1 are called: – relative voltage change characteristic:

d(t) (definition 3.2): – maximum relative voltage change:

dmax (definition 3.3): – relative steady-state voltage change:

dc (definition 3.4): These definitions are explained by the example in Figure 3.”

Figure 28. Detected modulating components after weighting filter in Block 3

Page 45: Harmonic and Flicker

45

The greatest interpretive difficulty arises from thelack of an adequate definition of “steady-state.” Itseems clear, given the text reproduced above, thatsuccessive half-cycle rms voltage measurementsmust meet some criteria for at least one second toqualify as steady-state. What is not given anywherewithin IEC 1000-3-3 is a definition of the ampli-tude characteristic of steady-state. Since it is veryunlikely that successive measurements will pro-duce exactly identical readings even with perfectlyconstant load behavior, it is left up to the reader toconclude what might be a reasonable band withinwhich one second’s worth of successive readingsmust fall in order to qualify as steady-state.

Agilent’s inferred definition is ±0.15% of the nomi-nal line voltage (Un). While the complete absence of a definition makes one interpretation as good as any other, there are solid reasons for makingthe choice just described. The error allocation formeasuring instrumentation is 5% while the statedlimit for Dc is 3%. Taking both specifications intoaccount (i.e., 5% of 3%) suggests that a tolerance of±0.15% for a steady-state condition is a reasonableguess at the standard’s intent. This band is alsoconsistent with the output noise performance ofAgilent’s sources and readily permits detection of steady-state EUT behavior.

Since setting this tolerance band is critical todetection of voltage change events, and to avoidimposing a specific interpretation on others whomay see it differently, Agilent’s HFTS softwareapplication uses an entry in the hfts.ini filethat contains initialization information for theapplication to set the tolerance band. The entry is factory set to 0.003 (i.e., 0.3% for the total tolerance band), but may be easily edited by those subscribing to a different interpretation.

There are other interpretive issues with Dmax, Dc, and Dt. Agilent takes the language in the noteincluded in Definition 3.4, particularly the inclu-sion of the word “relative” and the reference toFigure 3 of IEC 1000-3-3, to mean two things:

1. The term “relative” means that voltage changesare evaluated relative to the value of the steady-state voltage immediately preceding the voltagechange characteristic under evaluation.

2. Relative voltage changes are expressed as a percentage of the nominal mains voltage Un.

Figure 29. Edited version of Figure 3

Page 46: Harmonic and Flicker

46

Establishing a tolerance band to define steady-state together with the interpretations justdescribed permits a reasonably straightforwardevaluation for Dc.

The voltage change characteristic shown in Figure 3of IEC 1000-3-3 lends itself to a simple evaluationfor Dmax as well. Consider, however, a situationwhere a sudden increase in EUT current causes adip in rms voltage (in excess of 2% for example)followed by an extended period of time during whichsteady-state conditions are not re-established, butduring which the rms voltage approaches the pre-vious steady-state value. Assume now that a suddendecrease in EUT current drain causes a surge inrms voltage that just exceeds a 2% increase relativeto the previous steady-state voltage. This is followedby a decrease in rms voltage so that eventually asteady-state condition is re-established at a levelnot too different from the previous steady-statecondition.

A literal interpretation of Definition 3.3 wouldseem to require that Dmax be reported at a valuegreater than 4% which would constitute a failure.Using what is known about how human perceptionof flicker works, this would seem to be an exces-sively conservative assessment, particularly sincethe period of time between the minimum and

maximum values is unspecified and could rangefrom milliseconds to minutes. Pending clarificationon this issue, Agilent has adopted a definition morein keeping with Figure 3 of IEC 1000-3-3 that recordsthe greatest excursion in absolute terms from theprevious steady-state value as Dmax.

Because of the potentially unlimited storage require-ments for on-line analysis of Dmax, Dc, and Dtevents, Agilent has made two additional interpre-tive inferences that are directed towards meetingthe intent of the standard while still permitting a practical implementation:

1. Steady-state conditions may never be achievedduring a test, in which case values for Dmax, Dc, and Dt will not be reported.

2. In the situation where steady-state conditionsare established, Agilent’s solution reports themaximum Dmax, Dc, and Dt events occurringwithin each Pst integration period. These valuesare then compared to the specifications statedin IEC 1000-3-3 to determine pass/fail.

Figure 30. Dmax evaluation example

Page 47: Harmonic and Flicker

47

Delay to Start of TestUnlike IEC 1000-3-2, IEC 1000-3-3 contains no language excluding inrush current events in thefirst 10 seconds following power-up. Events of thistype will have little or no impact on flicker testresults in any case, but potentially will have majorimpact on Dmax, Dt, and Dc test results. For theserequirements, the language included in Section 5 of IEC 1000-3-3 is explicit:

“If voltage changes are caused by manual switch-ing or occur less frequently than once per hour,the Pst and Plt requirements shall not be applica-ble. The three requirements related to voltagechanges shall be applicable with the previouslymentioned voltage values, multiplied by a factorof 1.33.”

Since this language implies that the EUT, if it issubject to manual switching, will in fact be operatedmanually during the test, the measurement delaycontrols have no consequential impact on testresults. For this reason, Agilent’s solution makesthese controls available in the interest of providingconsistent functionality across the application, butwith the expectation that the test engineer willmake use of them in an informed manner.

Similarly, voltage change test results (i.e., Dmax,Dc, and Dt) are reported as measured and com-pared against the normal limits. It is expected that the test engineer will determine if the 1.33multiplier should properly be applied, and if so,that the multiplier will be manually applied as part of the analysis of test results as reported inthe “long form” report. Alternatively, the limitoverride control provided for rms parameters inthe “Advanced Options” setup screen may be set to 133% to provide automatic scaling of the rmstest limits.

Page 48: Harmonic and Flicker

48

Flickermeter Verification TestsTo the best of Agilent Technologies’ knowledge,there is no official certification for UIE flicker-meter implementations. The IEC commission doesnot endorse products offered by instrumentationmanufacturers, nor does it provide services or recommend test laboratories to verify compliancewith its published standards, either for the prod-ucts regulated by the standards or for instrumen-tation designed to test these products.

The design team for the 6840A Series Harmonic/Flicker Test System conducted their own assess-ment of measurement accuracy for Agilent’s flick-ermeter implementation. The results are attachedand a discussion of the test methodologies follows.This discussion includes a description of alternatemethods, the relative merits of these methods, andthe reasons for selecting the particular method used.

A UIE Flickermeter is a specialized AM receiverthat is designed to detect amplitude modulations of the ac mains and present the detected signals in terms of human perceptibility of incandescentlamp flicker. The attached tables show the resultsof three sets of tests, defined by IEC 868 and IEC 868Amendment 1, that are intended to test complianceof flickermeter implementations with measurementaccuracy requirements described in the standards.One set of the tests checks accuracy in terms ofpeak flicker response with sinewave modulations(Table A-1), while the other two check peak flickerresponse (Table A-2) and Pst measurement accuracywith squarewave modulations (Table A-3).

A common feature of all three tests is a require-ment for relatively small modulation depths andfor considerable precision in controlling modula-tion depth. This may be seen by examining thecolumns labeled “Modulation dV/V (%); nominal” in each table. The dV/V parameter describes thedifference between the maximum rms value of themodulated carrier (nominally 230 Vac @ 50 Hz)and the minimum rms value of the carrier and isexpressed as a percentage of the rms value of theac mains frequency carrier. In the more conven-tional measure of modulation as a percentage ofpeak-to-peak carrier amplitude, the equivalentvalue is one half the value shown in the tables.Therefore, the requirement for precision is actuallymore severe than it appears to be at first glance. In addition, IEC 868 specifies that the modulationlevel producing a peak flicker response of 1.00must be within ±5% of the nominal value. Thesedelta values are shown in the tables for peak flickerresponse in the columns labeled “Modulation dV/V (%); low and high.” Examination of the mag-nitude of the differences from the nominal valuemakes it clear that performance verification requirescontrol over modulation depth to resolutions ofapproximately 1 ppm and accuracies of 10 ppm.

Appendix A

Page 49: Harmonic and Flicker

49

The required test signals might be obtained by pre-cisely modulating a load, by precisely modulatingthe generated output without a load, or by measur-ing the effect produced by an uncalibrated load or modulation using a 230 Vac 50 Hz input modu-lation meter having measurement resolutions to0.0001% and corresponding accuracy. Off-the-shelfequipment with these capabilities is simply notavailable. Another technique described in some of the literature is to select a very low frequencysquarewave modulation and use a precision DVMto measure the two rms levels. This technique works,but becomes an act of faith when the assumption is made that once calibrated in this manner, thetest source may be switched to sinewave modula-tion and/or to higher modulation frequencies without loss of calibration.

The problems documented above led Agilent’s designteam to adopt a different approach. The block dia-gram in Figure A-1 shows the technique used.

Agilent’s 6840 Series Harmonic/Flicker Test Systemuses a firmware implemented DDS (direct digitalsynthesis) technique to generate the output wave-form. DDS generation produces programmable out-put frequencies with fine frequency resolution andcrystal accuracies. The 6840 Series products alsoinclude a fully digital implementation of the UIE

Flickermeter. This implementation includes digiti-zation of the mains voltage input to the EquipmentUnder Test (EUT) with a 16-bit A/D converteroperating at approximately 40K samples/second.All digital signal processor operations are per-formed on 32-bit floating point values and the A/Dconverter readings are immediately converted tofloating point representations following digitiza-tion. Since these numbers have approximately 7decimal digits of resolution, despite having only 16 bits of quantization and dynamic range, it ispossible to implement a highly precise digital AMmodulator that acts directly on the stream of digitized samples of the voltage input to the EUT.Addition of a second DDS generator as the modu-lation frequency source permits generation ofextremely accurate and repeatable test signals thusproviding a means for characterizing all systemerrors except those introduced by the A/D converterand the differential amplifier connecting it to theEUT. These additional errors are discussed sepa-rately later. Note that a lowpass filter is shown inthe modulation signal path. This attenuates thehigher frequency components of squarewave modu-lating waveforms to prevent problems with responsesto these signals that are not accounted for in Table 2 from IEC 868.

Figure A-1. Flickermeter block diagram

Page 50: Harmonic and Flicker

50

One outstanding quality of the Agilent 6840 SeriesHarmonic/Flicker Test System, both from the stand-point of making actual flicker measurements andfrom the standpoint of verifying performance, isthat given identical inputs, its fully digital architec-ture provides identical results every time.

All tests were conducted with the output pro-grammed to 230 Vac and 50 Hz with no load. Thissets the A/D outputs to the level corresponding to the midrange value encountered during an actual test. Since the filters in a UIE flickermeterhave very long time constants, the test proceduredescribed in IEC 868 was modified from a proce-dure of sweeping the modulation level to produce a reading of 1.00 to one of testing at the specifica-tion limits of 95% and 105% of nominal modulationand then using a simple algebraic manipulation tocalculate the modulation level that would producea Pst reading of 1.00. The extremely wide dynamicrange and fine resolution provided by the 32-bitfloating point implementation makes this approachcompletely appropriate for demonstrating compli-ance with IEC 868 requirements.

The results for peak flicker response with sine-wave and squarewave modulations at the modulat-ing frequencies and depths described in IEC 868are shown in the columns labeled “dV/V (%) forreading of 1.00” and “Deviation from nominal, % ofnominal.” The second column documents the errorsintroduced by the flickermeter implementation inaccordance with the test limits as described in thestandard and must show values between –5% and+5% for the implementation to meet specification.These errors are predominantly caused by thebandpass filter at the heart of the flickermeter andarise largely because the transfer function speci-fied in IEC 868 and the specified test modulationlevels are all empirically derived, but unfortunatelynot from the same data sets. This brings up the

interesting question of whether it is possible to“improve” the transfer function. Our conclusion is that the answer is “yes,” but the design teamdecided that it was better to implement the trans-fer function exactly as specified and live with theerrors rather than attempt to second guess thestandard’s authors. In any case, the 6840 SeriesHarmonic/Flicker Test System meets specified test accuracy across the board.

Quantization ErrorsThe errors introduced by the A/D converter andthe differential amplifier must also be taken intoaccount. The differential amplifier acts as a voltagedivider and is implemented with precision compo-nents that cause it to introduce total errors of lessthan 0.1% over the entire operating range of theinstrument. These errors therefore make a negligi-ble contribution to the total error budget. Thespecified voltage measurement accuracy for the6840 Series Harmonic/Flicker Test System is betterthan 0.08% at 230 Vac and thus this error sourcealso makes a negligible contribution.

Another more subtle issue with the A/D conver-sion process exists, however. Quantization error, orthe limited ability of the A/D converter to resolvedifferent input magnitudes, contributes the majorerror associated with the A/D conversion process.A number of different techniques may be used toanalyze this effect and depending upon the partic-ular technique selected, and upon the detail of theanalysis, various factors including frequencydomain noise floor, quantization error, averagingeffects, linearity errors, modulation waveshape,etc. may be taken into account. A detailed account-ing of these issues is beyond the scope of this dis-cussion, but a simple examination of quantizationerror (or resolving power) is illustrative.

Page 51: Harmonic and Flicker

51

The 16-bit A/D converter used in the 6840 SeriesHarmonic/Flicker Test System, when taking intoaccount the 500 V peak range of the voltage meas-urement circuit, provides a bit weight of approxi-mately 15.3 mV. Assuming monotonic and linearbehavior, the maximum deviation between an actualinput and a quantized value is 1/2 LSB or approxi-mately 7.6 mV. Dividing this value by the peak volt-age change at 230 Vac that is equivalent to thespecified modulation level and multiplying by 100yields the percentage of nominal modulation equiv-alent to the resolving power of the A/D converter.This value is shown for each test case in the tablesfor peak flicker response in the column labeled“A/D resolution @ 230 V.” These values are thensummed with the errors attributable to the flicker-meter implementation to yield a “worst case” error.Once again, errors for all specified test conditionsare within the specified limits. Note that factorssuch as averaging effects have not been considered,but generally these factors will act to improve reso-lution rather than degrading it, and resolution isthe issue here rather than absolute accuracy. It isassumed that the benefits of averaging togetherwith the fact that A/D conversions extend over awide range of codes will overcome errors intro-duced by nonlinearities in the conversion process.

The third table is for Pst measurement verification.This test is described in IEC 868 Amendment 1.The Pst test is specified somewhat differently inthat an exact modulation is prescribed and theimplementation must simply produce an outputthat is between 0.95 and 1.05 for each prescribedmodulation. The results of tests conducted usingthe methodologies described above are shown inthe table for Pst measurement accuracy. The 6840Series Harmonic/Flicker Test System again meetsrequirements across the board. Because Pst isderived from a statistical processing of 60,000 indi-vidual measurements, the A/D resolution issuesdescribed above are largely removed by the averag-ing effects implicit in the statistical calculation.

The final contributor to total measurement error is the reference impedance, 0.4 + j 0.25 ohms forsingle phase 50 Hz systems. Agilent’s programma-ble impedance is specified at <±1.0% error at roomtemperature and <±3% error over the full operat-ing range of the instrument. This error, when alge-braically summed with the errors described above,maintains the instrument within the ±8% total errorbudget permitted by the IEC standards.

Page 52: Harmonic and Flicker

52

Table A-1. Squarewave Modulation Test Results

Page 53: Harmonic and Flicker

53

Table A-2. Sinewave Modulation Test Results

Page 54: Harmonic and Flicker

54

Table A-3. Pst Accuracy with Squarewave Modulation

Modulation Frequency Modulation Pst Reading Pst Reading(Voltage changes/min.) Hertz (dV/V in %) (Flicker Perceptibility) (% of nominal)

1 0.00833 2.724 1.044 4.42 0.01667 2.211 1.025 2.57 0.05833 1.459 1.022 2.239 0.32500 0.906 1.027 2.7110 0.91667 0.725 1.018 1.81620 13.50000 0.402 0.990 –1.0

Figure A-2. Agilent 6840 Series squarewave and sinewave modulation test summary

Figure A-3. Agilent 6840 Series Pst accuracy with squarewave modulation summary

Page 55: Harmonic and Flicker

55

Ford, John, The VOLTECH Handbook of Testing to JEC555, 1993

McCarthy, Darren, “Immunity and Low FrequencyEmissions Standards,” IEEE SCV EMC Symposium,March 1994

Satoh, Akira, Japan Powerline HarmonicsRequirements Overview, 1995

IEC 555-1 (1982), “Part 1: Definitions”IEC 555-2 (1982), “Part 2: Harmonics”

Amendment No. 1 (1984)Amendment No. 2 (1988)Amendment No. 3 (1991)

IEC 77A (SEC) 113 (1994), “Part 4: Testing andmeasuring techniques Section 15: Flickermeter—Functional and design specifications.”

IEC 725 (1981), “Considerations on referenceimpedances for use in determining the disturbancecharacteristics of household appliances and similarelectrical equipment.”

IEC 816 (1984), “Guide on methods of measure-ment of short duration transients on low voltagepower and signal lines.”

IEC 827 (1985), “Guide to voltage fluctuation limits for household appliances (relating to IECPublication 555-3).”

IEC 868 (1986), “Flickermeter, Functional anddesign specifications.”

IEC 868 Amendment 1 (1990), “Flickermeter,Functional and design specifications.”

IEC 868-0 (1991), “Part 0: Evaluation of flickerseverity.”

IEC 1000-1-1 (1992), “Part 1: General—Section 1:Application and interpretation of fundamental definitions and terms.”

IEC 1000-3-2 (1995), “Part 3: Limits—Section 2:Limits for harmonic current emissions (equipmentinput current ≤16 A per phase).”

IEC 1000-3-3 (1994), “Part 3: Limits—Section 3:Limitation of voltage fluctuations and flicker inlow-voltage supply systems for equipment withrated current ≤16 A.”

IEC 1000-4-1 (1992), “Part 4: Testing and measure-ment techniques. Section 1: Overview of immunitytests. Basic EMC Publication.”

IEC 1000-4-7 (1991), “Part 4: Testing and measure-ment techniques. Section 7: General guide on har-monics and inter-harmonics measurements andinstrumentation for power supply systems andequipment connected thereto.”

IEC SC 77A/WGOI/TF02 (1995), “Draft, Minutes of the 3rd meeting held in Paris on January 26,1995.”

U.I.E. “Flicker Measurement and Evaluation,” 1986

Report to IEC 77A WG6 Committee members, from XITRON Technologies, September 15, 1995

Parts of IEC 1000-3-2, 1000-3-3, 1000-4-7, 868 are reproduced by kind permission of theInternational Electrotechnical Commission (IEC).Copies of the complete standard may be obtainedfrom IEC, P.O. Box 131, 1211 Geneva 20, Switzerland,Phone: 41 22 919 0228, FAX: 41 22 919 0300 or, in the United States, from the AmericanNational Standards Institute, Sales Department, 11 West 42nd Street, New York, NY 10036, Phone: 212 642 4900, FAX: 212 302 1286.

References

Page 56: Harmonic and Flicker

By internet, phone, or fax, get assistance with all your test and measurement needs.

Online Assistancewww.agilent.com/find/assistPhone or FaxUnited States:(tel) 1 800 452 4844

Canada:(tel) 1 877 894 4414(fax) (905) 282 6495

Europe:(tel) (31 20) 547 2323(fax) (31 20) 547 2390

Japan:(tel) (81) 426 56 7832(fax) (81) 426 56 7840

Latin America:(tel) (305) 269 7500(fax) (305) 269 7599

Australia:(tel) 1 800 629 485 (fax) (61 3) 9210 5947

New Zealand:(tel) 0 800 738 378 (fax) (64 4) 495 8950

Asia Pacific:(tel) (852) 3197 7777(fax) (852) 2506 9284

Product specifications and descriptions in this document subject to change without notice.

Copyright © 1995, 2000 Agilent TechnologiesPrinted in U.S.A. 10/005964-1917E