Guide Wire Failure Analysis Brad James, Ph.D., P.E.

10
Guide Wire Failure Guide Wire Failure Analysis Analysis Brad James, Ph.D., P.E. Brad James, Ph.D., P.E.

Transcript of Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Page 1: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Guide Wire Failure AnalysisGuide Wire Failure Analysis

Brad James, Ph.D., P.E.Brad James, Ph.D., P.E.

Page 2: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Background

• 304 stainless steel guidewire fractured after percutaneous balloon dilation of the pulmonary artery

• After successfully crossing lesion and dilation, the Doctor reported the wire became caught

• Attempts to remove the wire were unsuccessful• Ultimately, guide wire fractured approximately 5

cm from the distal tip• Remaining wire end remained in patient• Patient remained “asymptomatic”

Page 3: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Background

• The wire was sent to a metallurgical laboratory for examination

• The laboratory found an inclusion on the fracture surface and indicated the inclusion acted as a “stress riser and reduced the cross-sectional area”

• Tensile tests of other sections of the subject guidewire indicated a “significant range of ductility from fairly ductile to brittle”

Page 4: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Fracture Surface Analysis

• SEM analysis of subject break indicated ductile overload with microvoid coalescence

• Inclusion with shear lips toward top edge at 12:00

• Inclusion was roughly round, approximately 1/15th of wire diameter

• Rub observed on outside edge of fracture surface

Page 5: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Inclusion Analysis

• Inclusion contained primarily silicon, carbon, and oxygen

• Note shear lips around inclusion

Page 6: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

• Three tensile test samples were also supplied

• Visual inspection indicated each sample broke in the grips

• SEM analysis indicated ductile fracture (but not classic “cup and cone”

Tensile Sample AnalysisTensile Sample Analysis

Page 7: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Failure Analysis

• Inclusion equal to approximately 1/15 the diameter of the wire (area fraction of approximately 0.4%)

• Inclusions tend to have low interfacial bonding strength with matrix, therefore little difference between inclusions and pores

• In ductile materials, plastic deformation reduces stress concentration at notches

• Ductile materials also exhibit notch strengthening: notch strength = max tensile load/area at notch

• Conservative assumption for this example is no notch strengthening, wire strength reduced by reduction in area

Page 8: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Failure Analysis

• With no notch strengthening, wire strength reduced by area reduction (0.4%)

• Specified wire tensile strength was 405 to 435 ksi

• Assuming subject wire strength was 420 ksi (middle of range), inclusion reduced tensile strength by approximately 1.7 ksi.

Page 9: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Failure Analysis

Inclusion was near outside edge of fracture, inside shear lip area, away from initiation location

Page 10: Guide Wire Failure Analysis Brad James, Ph.D., P.E.

Conclusions

• The subject guidewire fractured in a ductile manner when service loads exceeded it’s tensile strength

• Inclusion reduced strength of wire, but only by a miniscule amount (well within specifications)

• Variations in apparent ductility from tensile testing (measured by elongation) only occurred because of the test methods employed

• SEM examination indicated each of the fractures were ductile in nature, no evidence of “brittle fracture” was observed