Gsp 900 Manual Final

download Gsp 900 Manual Final

of 255

Transcript of Gsp 900 Manual Final

  • 7/27/2019 Gsp 900 Manual Final

    1/255

    Spectrum Analyzer

    GSP-930

    USER MANUALREVISION 1.1 APRIL 2012

    ISO-9001 CERTIFIED MANUFACTURER

  • 7/27/2019 Gsp 900 Manual Final

    2/255

    This manual contains proprietary information, which is protected bycopyright. All rights are reserved. No part of this manual may bephotocopied, reproduced or translated to another language withoutprior written consent of Good Will company.

    The information in this manual was correct at the time of printing.However, Good Will continues to improve products and reserves therights to change specification, equipment, and maintenanceprocedures at any time without notice.Good Will Instrument Co., Ltd.No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.

  • 7/27/2019 Gsp 900 Manual Final

    3/255

    SAFETY INSTRUCTIONS

    1

    Table of ContentsSAFETY INSTRUCTIONS .................................................. 3 GETTING STARTED .......................................................... 8

    GSP-930 Introduction ............................ 9Accessories .......................................... 11Appearance .......................................... 13First Use Instructions .......................... 24

    BASIC OPERATION ........................................................ 36Frequency Settings ............................... 39Span Settings ....................................... 43Amplitude Settings .............................. 46Autoset ................................................ 59Bandwidth/Average Settings ................ 61Sweep .................................................. 66

    Trace .................................................... 72Trigger ................................................. 79Marker ................................................. 84Display................................................. 98System Settings ................................. 105Preset ................................................ 109

    ADVANCED OPERATION ...... ...... ...... ...... ...... ...... ...... ..... 111Measurement ..................................... 112Limit Line Testing .............................. 159Sequence ........................................... 165Tracking Generator ............................ 170Power Meter ...................................... 173

    FILE ............................................................................. 178File Overview ..................................... 178

  • 7/27/2019 Gsp 900 Manual Final

    4/255

    GSP-930 User Manual

    2

    REMOTE CONTROL ...................................................... 194Interface Configuration ...................... 195

    FAQ .............................................................................. 205APPENDIX .................................................................... 206

    Replace the Clock Battery ................... 206Glossary of Acronyms ........................ 207GSP-930 Default Settings ................... 209Menu Tree .......................................... 211 GSP-930 Specifications ...................... 240GSP-930 Dimensions ......................... 249Declaration of Conformity .................. 250

    INDEX .......................................................................... 251

  • 7/27/2019 Gsp 900 Manual Final

    5/255

    SAFETY INSTRUCTIONS

    3

    SAFETY INSTRUCTIONSThis chapter contains important safetyinstructions that you must follow duringoperation and storage. Read the following beforeany operation to insure your safety and to keepthe instrument in the best possible condition.

    Safety Symbols

    These safety symbols may appear in this manual or on theinstrument.

    WARNINGWarning: Identifies conditions or practices thatcould result in injury or loss of life.

    CAUTIONCaution: Identifies conditions or practices thatcould result in damage to the instrument or toother properties.

    DANGER High Voltage

    Attention Refer to the Manual

    Earth (ground) Terminal

    Frame or Chassis Terminal

    Do not dispose electronic equipment as unsortedmunicipal waste. Please use a separate collection

    facility or contact the supplier from which thisinstrument was purchased.

  • 7/27/2019 Gsp 900 Manual Final

    6/255

    GSP-930 User Manual

    4

    Safety Guidelines

    GeneralGuideline

    CAUTION

    Do not place any heavy object on theinstrument.

    Avoid severe impact or rough handling thatleads to damaging the instrument.

    Do not discharge static electricity to theinstrument.

    Use only mating connectors, not bare wires, forthe terminals.

    Ensure signals to the RF input do not exceed

    +30dBm. Ensure reverse power to the TG output terminal

    does not exceed +30dBm.

    Do not supply any input signals to the TGoutput.

    Do not block the cooling fan opening.

    Do not disassemble the instrument unless you

    are qualified.(Measurement categories) EN 61010-1:2010 specifies themeasurement categories and their requirements as follows. Theinstrument falls under category II.

    Measurement category IV is for measurement performed at thesource of low-voltage installation.

    Measurement category III is for measurement performed in thebuilding installation.

    Measurement category II is for measurement performed on the

    circuits directly connected to the low voltage installation.

    Measurement category I is for measurements performed oncircuits not directly connected to Mains.

    Power Supply

    WARNING

    AC Input voltage range: 100V~240V

    Frequency: 50/60Hz

    To avoid electrical shock connect the protectivegrounding conductor of the AC power cord toan earth ground.

  • 7/27/2019 Gsp 900 Manual Final

    7/255

    SAFETY INSTRUCTIONS

    5

    Battery

    CAUTION

    Rating: 10.8V, 6 cell Li-ion battery

    Turn off the power and remove the power cordbefore installing or removing the battery.

    Cleaning Disconnect the power cord before cleaning.

    Use a soft cloth dampened in a solution of milddetergent and water. Do not spray any liquid.

    Do not use chemicals containing harsh materialsuch as benzene, toluene, xylene, and acetone.

    Operation

    Environment

    Location: Indoor, no direct sunlight, dust free,

    almost non-conductive pollution (Note below) Temperature: 5C to 45C

    Humidity:

  • 7/27/2019 Gsp 900 Manual Final

    8/255

    GSP-930 User Manual

    6

    Disposal Do not dispose this instrument as unsortedmunicipal waste. Please use a separate collectionfacility or contact the supplier from which thisinstrument was purchased. Please make sure

    discarded electrical waste is properly recycled toreduce environmental impact.

  • 7/27/2019 Gsp 900 Manual Final

    9/255

    SAFETY INSTRUCTIONS

    7

    Power cord for the United Kingdom

    When using the instrument in the United Kingdom, make sure thepower cord meets the following safety instructions.

    NOTE: This lead/appliance must only be wired by competent persons

    WARNING: THIS APPLIANCE MUST BE EARTHEDIMPORTANT: The wires in this lead are coloured in accordance with thefollowing code:Green/ Yellow: EarthBlue: NeutralBrown: Live (Phase)

    As the colours of the wires in main leads may not correspond withthe coloured marking identified in your plug/appliance, proceedas follows:

    The wire which is coloured Green & Yellow must be connected tothe Earth terminal marked with either the letter E, the earth symbol

    or coloured Green/Green & Yellow.

    The wire which is coloured Blue must be connected to the terminalwhich is marked with the letter N or coloured Blue or Black.

    The wire which is coloured Brown must be connected to theterminal marked with the letter L or P or coloured Brown or Red.

    If in doubt, consult the instructions provided with the equipmentor contact the supplier.

    This cable/appliance should be protected by a suitably rated andapproved HBC mains fuse: refer to the rating information on theequipment and/or user instructions for details. As a guide, a cableof 0.75mm2 should be protected by a 3A or 5A fuse. Larger

    conductors would normally require 13A types, depending on theconnection method used.

    Any exposed wiring from a cable, plug or connection that isengaged in a live socket is extremely hazardous. If a cable or plug isdeemed hazardous, turn off the mains power and remove the cable,any fuses and fuse assemblies. All hazardous wiring must beimmediately destroyed and replaced in accordance to the abovestandard.

  • 7/27/2019 Gsp 900 Manual Final

    10/255

    GSP-930 User Manual

    8

    GETTING STARTEDThis chapter provides a brief overview of theGSP-930, the package contents, instructions forfirst time use and an introduction to the frontpanel, rear panel and GUI.

    GSP-930 Introduction ...................................................... 9Main Features .................................................................................................................9Accessories..................................................................... 11Package Contents ....................................................................................................... 12

  • 7/27/2019 Gsp 900 Manual Final

    11/255

    GETTING STARTED

    9

    GSP-930 IntroductionThe GSP-930 is the most advanced spectrum analyzer GW Instek

    has produced to date. The GSP-930 features a split window displayto view data in spectrum, topographic or spectrographic views.

    Main Features

    Performance 9kHz~3GHz bandwidth

    1Hz resolution

    Nominal RBW accuracy of 5% 750kHz

    Video bandwidth 1Hz~1MHz (10 steps)

    Amplitude measurement range: DANL~30dBm(frequency dependent)

    Input attenuation: 0 ~ 50dB

    Phase noise: < -88dBc/Hz@1GHz, 10kHz

    Features 10%-step increments for RBW bandwidth

    Three display modes: Spectrum, Topographicand Spectrographic

    Split window display

    Built-in EMI filter

    Auto Wake-up

    Built-in preamplifier

    Gate sweep

    Marker Frequency counter

    Two operating modes: Spectrum and PowerMeter mode

    SEM measurement

    ACPR measurement

    OCBW measurement

  • 7/27/2019 Gsp 900 Manual Final

    12/255

    GSP-930 User Manual

    10

    Channel power measurement

    Demodulation analyzer

    Diverse marker functions and features with

    Peak Table Sequence function to automatically perform pre-

    programmed sequential operations

    Optional battery operation

    Interface 8.4 color LCD (800600)

    On-screen menu icons

    DVI-I video output RS-232 with RTS/CTS hardware flow control

    USB 2.0 with support for USB TMC

    LAN TCP/IP with LXI support

    Optional GPIB/IEEE488 interface

    IF output @ 886MHz

    Headphone output

    REF (reference clock) input/output BNC ports

    Alarm/Open collector output BNC port

    Trigger/Gate input BNC ports

    RF N-type input port

    Tracking generator output

    DC +7V/500mA output SMB port

  • 7/27/2019 Gsp 900 Manual Final

    13/255

    GETTING STARTED

    11

    Accessories

    Standard

    Accessories

    Part number Description

    Region dependant User manual

    Region dependant Power cord

    Options Option number Description

    Opt1. Tracking generator

    Opt2. Battery (11.1V/5200mAH

    Li-ion battery)Opt3. GPIB interface (IEEE 488

    bus)

    OptionalAccessories

    Part number Description

    PWS-06 USB Average Power Sensor(up to 6200 MHz;-32 to 20 dBm)

    GRA-415 6U Rack mount kit

  • 7/27/2019 Gsp 900 Manual Final

    14/255

    GSP-930 User Manual

    12

    Package Contents

    Check the contents before using the GSP-930.

    Opening the box

    Contents(single unit)

    Main unit(may include optionalGPIB, TG output)

    Quick Start manual

    User Manual CD

    Power cord x1 (regiondependent)

    Optional battery pack

    Calibration certificate

  • 7/27/2019 Gsp 900 Manual Final

    15/255

    GETTING STARTED

    13

    Appearance

    GSP-930 Front Panel

    Frequency

    Autoset

    Amplitude

    Span

    BW/Avg

    Display

    Trace

    Sweep

    Meas

    Trigger

    Sequence

    File

    Recall

    Save

    Peak

    Marker->

    Marker

    System

    Mode

    Preset

    Search

    QuickSave

    LimitLine

    7

    4

    1

    0

    8

    5

    2

    9

    6

    3

    / Enter

    BK SP

    F4

    F3

    F2

    F1

    F6

    F5

    F7

    GSP-930 9 kHz 3 GHzSpectrum Analyzer

    LOCAL

    RF INPUT50

    REV PWR +30dBm

    TG OUTPUT50 W W500mA MAX.

    DC 7VOUTPUT +30dBm MAX.

    MAX.DC 50V MAX.DC 50V

    CONTROL FILE

    M AR KER S TA TE

    LCDdisplay

    Mainkeys

    Functionkeys

    Controlkeys Power key

    File keys

    State keys

    Scroll wheel

    Arrow keys

    RF inputterminal

    DC powersupply

    Trackinggeneratoroutput

    Numeric,Enter andBK SP keys

    USB A,Micro SDport

    Marker keys

    LCD display 800600 color LCD display. The display shows thesoft keys for the current function, frequency,amplitude and marker information.

    Function keysF 1

    ~ F 7The F1 to F7 function keys directlycorrespond to the soft keys on theright-hand side of display.

    Main keysFrequency

    Sets the center frequency, startfrequency, stop frequency, centerfrequency step and frequency

    offset values.

  • 7/27/2019 Gsp 900 Manual Final

    16/255

    GSP-930 User Manual

    14

    Span

    Sets the span, with options for fullspan, zero span and last span.

    Amplitude

    Sets the amplitude reference level,attenuation, pre-amplifiercontrols, scale and other optionsfor attenuation and scale.

    Autoset

    Automatically searches the peaksignal with maximum amplitudeand displays it with appropriate

    horizontal and vertical scales.

    Control keysBW/Avg

    Sets the resolution bandwidth,video bandwidth, average typeand turns the EMI filter on/off.

    Sweep

    Sets the sweep time and gate time.

    Trace

    Sets traces and trace relatedfunctions.

    Display

    The Display key configures thewindowing mode and basicdisplay properties.

    Meas

    Accesses measurement options

    such as ACPR, OCBW,demodulation measurements,SEM, TOI and other advancedmeasurements.

    LimitLine

    Sets and tests Pass/Fail limit lines.

    Sequence

    Access, set and edit program

    sequences.

  • 7/27/2019 Gsp 900 Manual Final

    17/255

    GETTING STARTED

    15

    Trigger

    Sets the triggering modes.

    File File

    File utilities options

    QuickSave

    The Quick Save utility allows youto save either the state, trace,screen limit line, correction orsequence with only a single press.

    Save

    Save the trace, state etc., and saveoptions.

    Recall

    Recall the trace, state etc., andrecall options.

    MarkerMarker

    Turns the Markers on/off andconfigures the markers.

    Marker->

    TheMarker-> key positions themarkers on the trace.

    PeakSearch

    Finds each maximum andminimum peak. Used with theMarker function.

    StatePreset

    LOCAL

    The Preset key will restore the

    spectrum analyzer to the Factoryor User-defined settings.

    The Preset key will also return theinstrument back to local controlafter it has been in remote controlmode.

  • 7/27/2019 Gsp 900 Manual Final

    18/255

    GSP-930 User Manual

    16

    Mode

    TheMode key sets the spectrumanalyzer to either Spectrum orPower Meter mode.

    System

    The System key shows systeminformation, settings and othersystem related functions.

    Power key Turns the instrument on/off.

    Scroll wheel Edit values, select listed items.

    Arrow keys Increment/decrement values (insteps), select listed items.

    RF input terminal

    RF INPUT 50W

    +30dBm MAX.MAX.DC 50V

    RF input port. Accepts RF inputs.

    Maximum input: +33dBm

    Input impedance: 50

    Maximum DC voltage: 50V

    N-type: female

    DC power supply

    500mA MAX.

    DC 7VOUTPUT

    SMB port supplies power foroptional accessories.

    DC +7V

    500mA Max.

  • 7/27/2019 Gsp 900 Manual Final

    19/255

    GETTING STARTED

    17

    Numeric keypad7 8 9

    654

    1 2 3

    /0

    BK SP

    Enter

    The numerickeypad is used toenter vales and

    parameters. It isoften used inconjunction withthe arrow keys andscroll wheel.

    TG output port

    REV PWR +30dBm

    TG OUTPUT 50 W

    MAX.DC 50V

    The Tracking Generator (TG)

    output source. N-type: female

    Input impedance: 50

    Output power: -50dBm to 0dBm

    Maximum reversed power:+30dBm

    USB A, Micro SD USB A port, Micro SD port forsaving/recalling settings/files.

  • 7/27/2019 Gsp 900 Manual Final

    20/255

  • 7/27/2019 Gsp 900 Manual Final

    21/255

    GETTING STARTED

    19

    Battery pack Voltage: 10.8VCapacity: 5200mAH

    REF IN BNC female reference input.

    REF OUT BNC female reference output:10MHz, 50 impedance

    Security Lock

    ALARM OUT BNC female open collector Alarmoutput.

    TRIG IN/GATE IN BNC female 3.3V CMOS triggerinput/gated sweep input.

    Phone 3.5mm stereo headphone jack(wired for mono operation)

    USB B USB B Device port. USB 1.1/2.0

    LAN RJ-45 10Base-T/100Base-Tx

  • 7/27/2019 Gsp 900 Manual Final

    22/255

    GSP-930 User Manual

    20

    Display

    Scale andReferencelevel

    Attenuationlevel

    Dateand time Function

    menu

    Markerinformation

    Triggersettings

    Pre-ampsettings

    Trace andDetectionsettings

    Frequency/Bandwidthsettings

    Status icons

    Sweepsettings

    Traces and

    waveforms

    LXI icon

    Entry /Messagearea

    Reference level Displays the reference level. For details, see page46.

    Attenuation Displays the vertical scale (attenuation) of theinput signal. For details, see page 47.

    Date/Time Displays the date and time. See page 106 for

    details.

    Markerinformation

    Displays marker information. For details see page83.

    LXI icon This icon indicates the status of the LXI connection.For details, see page 196.

    Function menu Soft menu keys associated with the F1 to F7function keys to the right of the display.

  • 7/27/2019 Gsp 900 Manual Final

    23/255

    GETTING STARTED

    21

    Sweep settings Sweep icon that shows the sweepstatus. See page 66 for details.

    Trigger settings Trigger icon that shows the triggerstatus. See page 79 details.

    Pre-amp settings Pre-amplifier icon that shows thePre-amplifier status. See from page48 for details.

    Trace anddetection settings

    Trace icon that shows the trace typeand the detection mode used foreach trace. See from page 72 fordetails.

    Status Icons Displays the interface status, power source status,and alarm status, etc. See the Status Icon Overviewon page 22 for a list of the status icons.

    Frequency/Bandwidthsettings

    Displays the Start, Center and Stop frequencies,RBW, VBW, Span and Sweep settings.

    Entry/Messagearea

    This area is used to show system messages, errorsand input values/parameters.

    Trace and

    waveforms

    Main display showing the input signals, traces

    (page 72), limit lines (159) and marker positions(83).

  • 7/27/2019 Gsp 900 Manual Final

    24/255

    GSP-930 User Manual

    22

    Status Icon Overview

    PreAmp Indicates that the pre amplifier ison.

    AC Shown when running on ACpower.

    AC Charge Shown when the AC power ischarging the battery.

    Alarm Off Alarm buzzer output is currentlyoff.

    Alarm On Alarm buzzer output is currentlyon.

    Amplitude Offset Indicates that the amplitude-shift is

    active. This icon appears whenamplitude-related functions areused:Reference level offsetAmplitude CorrectionInput Z = 75Input Z cal >0

    Battery indicator

    ~

    Indicates the battery charge.

    BandwidthIndicator

    Indicates that the RBW or VBWsettings are in manual mode.

    Average Indicates that the Average functionis active.

  • 7/27/2019 Gsp 900 Manual Final

    25/255

    GETTING STARTED

    23

    External Lock Indicates that the system is nowlocked and refers to the externalreference input signal

    External Trigger External trigger signal is beingused.

    Math Trace math is being used.

    SequenceIndicator

    Shown when a sequence is running.

    Sweep Indicator Indicates that the sweep time ismanually set.

    Trackinggenerator

    Indicates the tracking generator isturned on.

    TG Normalization Indicates that the tracking generator

    has been normalized.

    Wake-up clock Indicates that the wake-up clock isturned on.

    USB Indicates that a USB flash drive isinserted into the front panel and isrecognized.

    Micro SD Indicates that a micro SD card isinserted into the front panel and isrecognized.

  • 7/27/2019 Gsp 900 Manual Final

    26/255

    GSP-930 User Manual

    24

    First Use InstructionsUse the procedures below when first using the GSP-930 to tilt the

    stand, insert the battery pack, power up the instrument, setting theinternal clock, the wake-up clock, updating the firmware andrestoring the default settings. Lastly, the Conventions sections willintroduce you to the basic operating conventions used throughoutthe user manual.

    Tilting the Stand

    Description The GSP-930 has two adjustable rubber feetthat can used to position the instrument intotwo preset orientations.

    Upright Position Tuck the feet underthe bottom of theinstrument to stand

    the instrumentupright.

    Leaning Position Pull the feet back to

    have the instrumentleaning back.

  • 7/27/2019 Gsp 900 Manual Final

    27/255

    GETTING STARTED

    25

    Inserting the Battery Pack

    Description The GSP-930 has an optional battery pack. Thebattery should be inserted before power isconnected to the AC power socket and beforethe unit is turned on.

    Steps 1. Ensure the power is off and the AC power isdisconnected.

    2. Remove the battery cover.

    3. Insert the battery as shown in the diagrambelow.

    4. Replace the battery cover.

    Display Icon The battery icon is displayed when GSP-930 is running on battery power.

    Insertion Diagram

  • 7/27/2019 Gsp 900 Manual Final

    28/255

    GSP-930 User Manual

    26

    Power UP

    Steps 1. Insert the AC power cord into the powersocket.

    2. The power button exterior will be lit blue toindicate that the GSP-930 is in standby mode.

    3.

    Press the power button to turn the GSP-930 on.

    4. The power button will turn orange and theGSP-930 will start to boot up.

    Note

    It takes approximately 1.5 minutes for the GSP-930

    to fully startup.

  • 7/27/2019 Gsp 900 Manual Final

    29/255

    GETTING STARTED

    27

    Power Down

    Description The GSP-930 has two methods to power down:Normal and Forced Power Down.

    The normal power down method will save thesystem state and end any running processes.The state is saved for the next time theinstrument is turned back on.

    The forced power down method only does aminimum state save.

    Normal PowerDown

    Press the power button. The system willautomatically handle the power downprocedure in the following order:

    The system state is saved.

    Outstanding processes are closed in

    sequence.

    The LCD backlight is turned off.

    The system enters standby mode (thepower key changes from orange toblue).

    NoteThe process takes ~10 seconds.

    Forced PowerDown

    Press and hold the power button for ~4seconds until the system turns off and thepower button turns blue.

    NoteThe forced down mode might cause the GSP-930 to perform a longer system check the nexttime it is powered up.

  • 7/27/2019 Gsp 900 Manual Final

    30/255

    GSP-930 User Manual

    28

    Setting the Date, Time and Wake-Up Clock

    Description The GSP-930 can be setup to power-upautomatically using the Wakeup Clockfunction. This feature is useful to wake-up theinstrument early and eliminate settling time.

    System Date Example: Set the System Date to March 1, 2012

    1. Press System >Date/Time[F4]>Set

    Date[F1]>Year[F1].

    2. Press 2012>Enter[F1].

    3. PressMonth[F2]>3>Enter[F1].

    4. Press Day[F3]>1>Enter[F1].

    5. Press Return[F7].

    NoteThe System Date will be shown at the top of thedisplay.

    System Time Example: Set the System Time to 9.00 AM

    1. Press System >Date/Time[F4]>SetTime[F2]>Hour[F1].

    2. Press 9>Enter[F1].

    3. PressMinute[F2]>0>Enter[F1].

    4. Press Second[F3]>0>Enter[F1].

    5. Press Return[F7].

  • 7/27/2019 Gsp 900 Manual Final

    31/255

    GETTING STARTED

    29

    NoteThe System Time will be shown at the top of thedisplay.

    System Wake-UpClock

    Example: Set the GSP-930 to wake up at 9.00AM

    1. Press System >Date/Time[F4]>Wake-UpClock[F3]>Clock[F1].

    2. Press Clock[F1] to choose a clock (1 ~ 7).

    3. Press State[F2] to turn the clock on/off.

    4. Press Hour[F3]>9>Enter[F1].

    5. PressMinute[F4]>0>Enter[F1].

    6. Press [F5] and choose Rept. (Repeat) or Single.

    7. Press Select Date[F6] and select a day.

    8. Press Return[F7] to save the Wake-Up Clocksettings.

    NoteThe system time is kept with the CR2032 clockbattery. If the system time/ wake up clock can nolonger be set, please replace the clock battery. Seepage 206.

  • 7/27/2019 Gsp 900 Manual Final

    32/255

    GSP-930 User Manual

    30

    Firmware Update

    Description The GSP-930 allows the firmware to beupdated by end-users. Before using the GSP-930, please check the GW Instek website or askyour local distributor for the latest firmware.

    System version Before updating the firmware, please check thefirmware version.

    1. Press System >System Information[F1].

    2. The firmware will be listed on the display.

    Firmware: T. 1. 0. 0

    3. Press any other key to exit out of the SystemInformation screen.

    4. To upgrade the firmware, insert the newfirmware onto a USB flash drive or Micro SDcard and put the drive/ card into theappropriate front panel port. The firmwarefiles should be located in a directory namedgsp930.

    5. Press System >More 1/2[F7]>Upgrade[F3].

  • 7/27/2019 Gsp 900 Manual Final

    33/255

    GETTING STARTED

    31

    6. The spectrum analyzer will automatically findthe firmware on the USB flash drive and startto update the firmware. When finished, the

    message Upgrade is finished will be shownat the bottom of the screen followed byRebooting.

    Rebooting

    7.

    The system will automatically restart after therebooting message.

    NoteThe upgrade process may take a few minutes.

    Restoring Default Settings

    Description The factory default settings or user-definedpresets can be easily restored using the Presetkey on the front panel. By default, the factorydefault settings are restored with the Presetkey.

    For details on how to configure the presetsettings, please see page 109.

    Steps 1. Press Preset .

    2. The spectrum analyzer will load the presetsettings.

  • 7/27/2019 Gsp 900 Manual Final

    34/255

    GSP-930 User Manual

    32

    Conventions

    The following conventions are used throughout the user manual.

    Read the conventions below for a basic grasp of how to operate theGSP-930 menu system and front panel keys.

    Soft Menu keys The F1 to F7 function keys on the right side ofthe display correspond directly to the soft-menu keys on their left.

    Frequency

    Autoset

    Amplitude

    Span

    BW/Avg

    Display

    Trace

    Sweep

    Meas

    Trigger

    Sequence

    File

    Recall

    Save

    Peak

    Marker->

    Marker

    System

    Mode

    Preset

    Search

    QuickSave

    LimitLine

    7

    4

    1

    0

    8

    5

    2

    9

    6

    3

    / Enter

    BKSP

    F4

    F3

    F2

    F1

    F6

    F5

    F7

    GSP-930 9 kHz 3 GHzSpectrum Analyzer

    LOCAL

    R FIN P U T50

    REVPWR +30dBm

    TGOU TP UT 50W W500mAMAX.

    D C 7VO UT PU T + 30 dB m M AX.

    MAX.DC 50V MAX.DC50V

    CONTROL FILE

    M AR KE R S TA TE

    Soft-menu keys

    F1 ~ F7 function keys

    Input ParameterValues

    Selecting this type of menu key will allow youto enter a new value with the numeric keypador increment/decrement the value using thescroll wheel.

    Toggle State

    Pressing this menu key will toggle the state.

  • 7/27/2019 Gsp 900 Manual Final

    35/255

    GETTING STARTED

    33

    Toggle State &Input Parameter

    Pressing this menu key will allow you to togglethe state of the function between Auto andMan(ual) state. When in the Man state, theparameter value can be manually edited. Usethe numeric keypad to enter the new value oruse the scroll wheel to increment/decrementthe current value.

    Sub Menu

    Pressing this menu key will enter a submenu.

    Sub Menu to

    select parameter

    Pressing this menu key will enter a submenu toselect a parameter.

    Active Function

    Pressing this type of menu key will activatethat function. The menu key will behighlighted to show it is the active function.

  • 7/27/2019 Gsp 900 Manual Final

    36/255

    GSP-930 User Manual

    34

    Parameter input Numerical keypadScroll wheel

    7

    4

    1

    0

    8

    5

    2

    9

    6

    3

    / Enter

    BK SP

    Directional arrow keysBackspace, Enter keys

    Parameter values can be entered using thenumeric keypad, the scroll wheel andoccasionally with the arrow keys.

    Using the numeric

    keypad

    When prompted to enter a parameter, use the

    number keys (0~9), the decimal key (.) and thesign key (+/-) to enter a value. After a valuehas been entered, the soft-menu keys can beused to select the units.

    The value of the parameter is shown at thebottom of the screen as it is edited.

    Edited parameter

    Span: 1.5

    Back Space Use the backspace key to delete the lastcharacter or number entered.

  • 7/27/2019 Gsp 900 Manual Final

    37/255

    GETTING STARTED

    35

    Using the scrollwheel

    Use the scroll wheel to alter the current value.Clockwise increases the value, anti-clockwisedecreases the value.

    Directional arrows Use the directional arrows to select discreteparameters or to alter values by a coarserresolution than the scroll wheel. Left decreasesthe value, right increases the value.

  • 7/27/2019 Gsp 900 Manual Final

    38/255

    GSP-930 User Manual

    36

    BASIC OPERATIONFrequency Settings .......................................................... 39Center Frequency ....................................................................................................... 39Start and Stop Frequency .......................................................................................... 40Center Frequency Step .............................................................................................. 41Frequency Offset ........................................................................................................ 42Span Settings ................................................................. 43Span .............................................................................................................................. 43Full Span ...................................................................................................................... 44Zero Span .................................................................................................................... 44Last Span ...................................................................................................................... 45Amplitude Settings ......................................................... 46Reference Level .......................................................................................................... 46

    Attenuation .................................................................................................................. 47Scale/Div ..................................................................................................................... 48Scale Type .................................................................................................................... 48

    View Scale .................................................................................................................... 49Vertical Scale Units .................................................................................................... 50Reference Level Offset .............................................................................................. 50

    Amplitude Correction ................................................................................................ 51Create a Correction Set ............................................................................... 52

    Amplitude Correction On/Off .................................................................. 54Delete Correction Set .................................................................................. 55Save Correction Set To Memory ............................................................... 55Recall Correction Set From Memory ........................................................ 56

    Input Impedance ........................................................................................................ 56Input Impedance Calibration ................................................................................... 57Using the Built-in Pre-Amplifier ............................................................................. 57Autoset .......................................................................... 59Using Autoset ............................................................................................................. 59Limiting the Autoset Vertical Search Range .......................................................... 60

  • 7/27/2019 Gsp 900 Manual Final

    39/255

    BASIC OPERATION

    37

    Limiting the Autoset Horizontal Search Range .....................................................60Bandwidth/Average Settings ........................................... 61Resolution Bandwidth Setting (RBW) .....................................................................61

    Video Bandwidth Settings (VBW) ...........................................................................62

    VBW/RBW Ratio .......................................................................................................62Average Trace ..............................................................................................................63Average Type ...............................................................................................................65EMI Filter .....................................................................................................................66Sweep ............................................................................ 66Sweep Time ..................................................................................................................66Single Sweep .................................................................................................................67Continuous Sweep ......................................................................................................68

    Gated Sweep Overview ..............................................................................................68Using the Gated Sweep Mode ..................................................................................70Trace ............................................................................. 72Selecting a Trace ..........................................................................................................72

    Trace Math ...................................................................................................................74Trace Detection Mode ...............................................................................................75Trigger ........................................................................... 79Selecting a Trigger Type .............................................................................................79

    Free Run Mode ..............................................................................................79Activate Video Trigger .................................................................................79Activate External Trigger .............................................................................81

    Selecting the Trigger Mode .......................................................................................82Set the Trigger Delay Time .......................................................................................83Marker ........................................................................... 84

    Activating a Marker.....................................................................................................85

    Activate a Normal Marker ...........................................................................85Move Marker Manually ................................................................................86Move Marker to Preset Locations ..............................................................86

    Activate Delta Marker ..................................................................................87Move Delta Marker(s)Manually ..................................................................88

    Marker Functions ........................................................................................................89Marker Noise .................................................................................................89Frequency Counter .......................................................................................90

    Move Marker to Trace ...............................................................................................91Show Markers in Table ..............................................................................................92Peak Search ..................................................................................................................93

  • 7/27/2019 Gsp 900 Manual Final

    40/255

    GSP-930 User Manual

    38

    Move Marker to Peak .................................................................................. 93Move Marker and Peak to Center ............................................................. 93Search for Peaks ........................................................................................... 94Peak Configuration ...................................................................................... 95

    Peak Table ................................................................................................................... 96Display ........................................................................... 98

    Adjusting the LCD Brightness ................................................................................. 98Turning the LCD Backlight Off .............................................................................. 98Setting a Display Line (Reference Level Line) ...................................................... 99Using the Video Out Port ......................................................................................... 99Setting the Display Mode ........................................................................................ 100Split Spectrum View ................................................................................................ 103System Settings ............................................................ 105System Information ................................................................................................. 105Error Messages ......................................................................................................... 105Set the System Language ......................................................................................... 106Set the Date and Time ............................................................................................. 106Using the Wake-Up Clock ...................................................................................... 107

    Alarm Output ............................................................................................................ 108Preset ........................................................................... 109Using the Preset Key ............................................................................................... 109Save the User Preset Settings ................................................................................. 109Preset Type Settings ................................................................................................. 110Power on Preset Settings ........................................................................................ 110

  • 7/27/2019 Gsp 900 Manual Final

    41/255

    BASIC OPERATION

    39

    Frequency Settings

    Center Frequency

    Description The center frequency function sets the centerfrequency and centers the display to the centerfrequency.

    Operation 1. Press Frequency >Center[F1] and enter the

    frequency and unit.

    Range:Resolution:Default

    0kHz~3GHz1Hz1.5GHz

    Display Center frequency

    Set Center Frequency

  • 7/27/2019 Gsp 900 Manual Final

    42/255

    GSP-930 User Manual

    40

    Start and Stop Frequency

    Description The start/stop frequency function will set thestart and stop frequency of the span.

    Operation 1. To set the start frequency, press Frequency > Start

    Freq[F2] and enter the frequency and unit.

    2. To set the stop frequency, press Frequency > Stop

    Freq[F3] and enter the frequency and unit.

    Range:Resolution:Default Start frequency:Default Stop frequency:

    0kHz~3GHz1Hz0Hz3GHz

    Display Start Frequency

    Start Frequency Stop Frequency

    Stop Frequency

    NoteThe start and stop frequency can change when thespan settings are used.The stop frequency must be set higher than the

    start frequency (for spans 0), otherwise the span

    will be automatically set to 100Hz.

  • 7/27/2019 Gsp 900 Manual Final

    43/255

    BASIC OPERATION

    41

    Center Frequency Step

    Description The CF Step function sets the step size of thecenter frequency when using the arrow keys.

    When the arrow keys are used to alter thecenter frequency, each press will move thecenter frequency by the step size specified bythe CF Step function.

    In auto mode, the center frequency step size isequal to 10% (1 division) of the span.

    Operation 1. Press Frequency > CF Step[F4] and set the CF Step to

    Auto or Man.

    2. If Man was selected, set the frequency and unitof the center frequency step size.

    Manual Range:Auto range:

    100Hz~3GHz1/10 of span frequency

    Display

    CF Step size

  • 7/27/2019 Gsp 900 Manual Final

    44/255

    GSP-930 User Manual

    42

    Frequency Offset

    Description The Freq Offset function allows you to add anoffset to the Center, Start and Stop frequenciesas well as the marker frequencies. The offsetvalue does not affect displaying the trace on thedisplay.

    Operation 1. Press Frequency >Freq Offset[F5] and set the offset

    value.

    The Center, Start, Stop and Marker frequenciesare updated accordingly.

    Offset Range: 0Hz~100GHz

    Display

    Frequency

    Offset

  • 7/27/2019 Gsp 900 Manual Final

    45/255

    BASIC OPERATION

    43

    Span Settings

    Span

    Description The Span function will set the frequency rangeof the sweep. The sweep will be centeredaround the center frequency.

    Setting the span will alter the start and stopfrequencies.

    Operation 1. Press Span > Span[F1] and enter the span

    frequency range and unit.

    Range:Resolution:Default Span:

    0kHz~3GHz1Hz3GHz

    Display Span

    Set Span

  • 7/27/2019 Gsp 900 Manual Final

    46/255

    GSP-930 User Manual

    44

    Full Span

    Description The Full Span function will set the span to thefull frequency range.

    This function will set the start and stopfrequencies to 0Hz and 3GHz respectively.

    Operation 1. Press Span > Full Span[F2].

    Zero Span

    Description The Zero Span function will set the frequencyrange of the sweep to 0Hz and fixes the startand stop frequencies to the center frequency.The Zero Span function measures the timedomain characteristics of the input signal at the

    center frequency. The horizontal axis isdisplayed in the time domain.

    Operation 1. Press Span > Zero Span[F3].

    The span changes accordingly.

  • 7/27/2019 Gsp 900 Manual Final

    47/255

    BASIC OPERATION

    45

    Display Time domain

    0Hz Span

    Example: Amplitude modulation

    NoteThe measurement functions such as TOI, SEM,CNR, CTB, CSO, ACPR, OCBW, phase, Jitter andNdB are not available with the zero span setting:

    Last Span

    Description The last span function returns the spectrumanalyzer to the previous span settings.

    Operation 1. Press Span > Last Span[F4].

  • 7/27/2019 Gsp 900 Manual Final

    48/255

    GSP-930 User Manual

    46

    Amplitude SettingsThe vertical display scale is defined by the reference level

    amplitude, attenuation, scale and external gain/loss.

    Reference Level

    Description The reference level defines the absolute level ofthe amplitude on the top graticule in voltage orpower.

    Operation 1. Press Amplitude > Ref Level[F1] and enter the

    reference level amplitude and unit.

    Range:Units:Resolution:

    -120dBm ~ 30dBmdBm, W, V, dBmV, dBuV1dBm

    DisplayReference LevelRef Level reading

  • 7/27/2019 Gsp 900 Manual Final

    49/255

    BASIC OPERATION

    47

    Attenuation

    Description The attenuation of the input signal level can setto automatic (Auto) or manual (Man). When theattenuation is set to Man, the input attenuatorcan be changed manually in 1dB steps.

    Operation 1. Press Amplitude >Attenuation[F2] and select Auto

    or Man.

    2. If Man was selected, enter the attenuation leveland unit.

    Range:Units:Resolution:

    0dBm ~ 50dBmdBm1dB

    Display Attenuation level

  • 7/27/2019 Gsp 900 Manual Final

    50/255

    GSP-930 User Manual

    48

    Scale/Div

    Description Sets the logarithmic units for the verticaldivisions when the scale is set to Log.

    Operation 1. Press Amplitude > Scale/Div[F3] repeatedly to select

    the vertical division units.

    Unit Range: 10, 5, 2, 1

    Display Scale

    Note The Scale/Div function is only selectable when thescale is set to Log (logarithmic).

    Scale Type

    Description Sets the vertical scale in linear or logarithmicunits.

    By default the linear scale is set to volts and thelogarithmic scale is set to dBm.

    Operation 1. Press Amplitude > Scale Type[F4] and set the vertical

    scale to Log or Lin.

    NoteIf the unit scale is changed (i.e. dBm volts), thedisplayedvertical scale type will remain in the set

    linear or logarithmic setting.

  • 7/27/2019 Gsp 900 Manual Final

    51/255

    BASIC OPERATION

    49

    View Scale

    Description The Scale function turns the vertical scale

    on/off. The value of each graticule division isdisplayed with same units used for the RefLevel settings.

    Operation 1. Press Amplitude > Scale[F5] to toggle the Scale on or

    off.

    2. Press Scale Pos.[F6] to toggle the position of thescale when on.

    Scale position: Left, Center, Right

    Display

    ScaleLeft

    The vertical scale is displayed on the left hand sideby default.

  • 7/27/2019 Gsp 900 Manual Final

    52/255

    GSP-930 User Manual

    50

    Vertical Scale Units

    Description Change the vertical units for both linear orlogarithmic scales.

    Operation 1. Press Amplitude >More[F7]>Y Axis[F1] and then

    choose the desired units.

    2. The units are changed accordingly.

    Units: dBm, dBmV, dBuV,Watts, Volts

    Reference Level Offset

    Description The Reference Level Offset function sets anoffset value to the reference level to compensate

    for any loss or gain from an external network ordevice.The offset value does not affect the inputattenuation or the on-screen trace.This setting will change the reference levelreadout, the scale readout and the markerreadout.

    Operation 1.

    PressAmplitude

    >More[F7]>RefLvlOffset[F2] and setthe offset level and unit.

    2. To remove the offset level, set the referenceoffset to 0 dB.

    Range: 0dB ~ 50dB

    Display Icon The AMP icon is displayed at the bottom

    of the screen.

  • 7/27/2019 Gsp 900 Manual Final

    53/255

    BASIC OPERATION

    51

    Example: Ref: 0dBm

    Before reference level offset(offset: 0dB)

    Ref: 10dBmReference level

    offset: 10dB

    After reference level offset (offset: 10dB)

    Amplitude Correction

    Description Amplitude correction adjusts the frequencyresponse of the spectrum analyzer by alteringthe amplitudes at specified frequencies. Thisallows the spectrum analyzer to compensate forloss or gain from an external network or deviceat certain frequencies.

    Range Correction Sets: 5 sets of 30 points

    Amplitude: -40dB to +40dBAmplitude Resolution: 0.1dBFrequency: 9kHz to 3GHzFrequency Resolution: 1Hz

  • 7/27/2019 Gsp 900 Manual Final

    54/255

    GSP-930 User Manual

    52

    Display Correction points

    Original waveform

    Correctedoutput

    Example: The diagram above shows how amplitude

    correction is used to compensate for any losses orgains at specific frequencies.

    Create a Correction Set

    Description The GSP-930 can create and edit up to 5 sets ofcorrection points. The correction points andassociated values are all tabulated for ease ofuse.

    Operation 1. Press Amplitude >More[F7]>Correction[F3]>

    Correction Set[F1] and choose a correction set toedit/create.

    Correction set: 1~5

  • 7/27/2019 Gsp 900 Manual Final

    55/255

    BASIC OPERATION

    53

    2. Press Edit[F3].

    The GSP-930 will split into two screens. The top

    screen will show the waveform and the bottomscreen will provide an overview of thecorrection points.

    Spectrum display

    Correction points

    Frequency ofselected point

    3. Press Point Num[F1] and choose a point numberto edit.

    Point Num: 1~30

    4. Press Frequency[F2] and choose the frequency of

    the selected point.Press Gain Offset[F3] and choose the amplitudeof the select point. The units will be the same asthose used for the vertical scale.

    The frequency of the point values displayed inthe correction table on the bottom display.

  • 7/27/2019 Gsp 900 Manual Final

    56/255

    GSP-930 User Manual

    54

    Correction Table

    5. Repeat steps 2 to 4 for any other correctionpoints.

    6. To delete the selected point, press DeletePoint[F6].

    7. Press Return[F7]>Save Correction[F5] to save thecorrection set.

    NoteNote that the correction points are automaticallysorted by frequency (low high).The correction set must be saved before it can beturned on.

    The frequency values displayedin the correctiontable are rounded down for display purposes only.The actual frequency for each point can be seen inthe Frequency soft-key.

    Amplitude Correction On/Off

    Description Any one of the 5 correction sets can be turnedon.

    ActivateCorrection

    1. Press Amplitude >More[F7]>Correction[F3]>

    Correction Set[F1] and choose a correction set.

    Correction Set: 1~5

    2.

    Press Correction[F2] and toggle correction on.

  • 7/27/2019 Gsp 900 Manual Final

    57/255

    BASIC OPERATION

    55

    DeactivateCorrection

    1. Press Amplitude >More[F7]>Correction[F3]>

    Correction[F2] to turn correction back off.

    Delete Correction Set

    Operation 1. Press Amplitude >More[F7]>Correction[F3]>

    Correction Set[F1] and choose the correction setto delete.

    Correction Set: 1~5

    2. Press DeleteCorrection[F6].The selected correction set will be deleted.

    Save Correction Set To Memory

    Operation 1. Press Save > Save To[F1] and choose the save

    location.

    Location: Register, Local, USB, SD

    2. Press Type[F2]> Correction[F5].

    3. Press Data Source[F3] and choose a correction.

    Correction Set: Correction 1~5

    4. Press Save Now[F7].

    5. The correction set will be saved to the selectedlocation. For more information on Save andRecall, please see page 170.

  • 7/27/2019 Gsp 900 Manual Final

    58/255

    GSP-930 User Manual

    56

    Recall Correction Set From Memory

    Operation 1. PressRecall

    > Recall[F1] and choose the recalllocation:

    Location: Register, Local, USB, SD

    2. Press Type[F2]> Correction[F5].

    3. Press Destination[F3] and choose a correctionset.

    Correction Set: Correction 1~5

    4. Press Recall Now[F4].

    5. The current correction set will be recalled fromthe selected location. For more information onSave and Recall, please see page 170.

    Input Impedance

    Description Sets the input impedance to 75 or 50.

    Operation 1. Press Amplitude >More[F7]>Input Z[F4] to toggle

    the input impedance.

    Range: 75, 50

  • 7/27/2019 Gsp 900 Manual Final

    59/255

    BASIC OPERATION

    57

    Input Impedance Calibration

    Description When an external impedance converter module(optional accessory ADP-101) is used to convertthe impedance of a device from 50 to 75,some external loss can be induced. The Input ZCal function can be used to compensate forthese losses with an offset value.

    Note

    The Input Z Cal function is only available when the

    input impedance is set to 75.

    Operation 1. Press Amplitude >More[F7]>Input Z Cal[F5] and set

    the impedance offset.

    Range:Resolution:

    0dB to +10dB1dB

    Display Icon The AMP icon is displayed at the bottomof the screen when Input Z Cal0dB.

    Using the Built-in Pre-Amplifier

    Description The built-in pre-amplifier boosts weak inputsignals, such as EMI testing signals, to levels

    that are easy to handle, over the entirefrequency range. The built-in pre-amplifier onthe GSP-930 has a nominal gain of 20dB.

    In the Auto setting, the pre-amplifier will beautomatically turned on when the referencelevel is less than -30dBm. When the referencelevel is greater than -30dBm, the pre-amplifieris turned off.

    The bypass setting turns the pre-amplifier off.

  • 7/27/2019 Gsp 900 Manual Final

    60/255

    GSP-930 User Manual

    58

    Operation 1. Press Amplitude >More[F7]>Preamp[F6] to toggle

    the Preamp state.

    Range: Auto, Bypass

    Display Icon The Pr-amp icon indicates that the preamplifier is on.

    Example: Pr-amp icon

    NoteWhen the pre-amplifier is on, the attenuatorbecomes fixed at 0dB (i.e. Attenuation = 0dB).

  • 7/27/2019 Gsp 900 Manual Final

    61/255

    BASIC OPERATION

    59

    AutosetThe Autoset function searches the peak signals in two stages (full

    span & 0Hz - 100MHz limited span), picks the signal peak with themaximum amplitude, and then shows it in the display.

    Using Autoset

    Operation 1. Press Autoset >Autoset[F1].

    Autoset Range Amplitude:Span:

    -80dBm ~ +20dBm100Hz ~ 3GHz

    Example:

    Before Autoset, preset state

    After Autoset

  • 7/27/2019 Gsp 900 Manual Final

    62/255

    GSP-930 User Manual

    60

    NoteRBW, VBW and sweep settings are reset to Autowhen the Autoset function is used.

    Limiting the Autoset Vertical Search Range

    Description You can set the amplitude floor so that thesignals lower than the setting will be ignoredby the Autoset search.

    Operation 1. Press Autoset >Amp.Floor[F2] and switch the

    range from Auto to Man.

    2. Enter the amplitude limit and unit for theAutoset search.

    Range: 80 to +20dBm

    NoteSee page 48 for setting the amplitude units.

    Limiting the Autoset Horizontal Search Range

    Description You can change the frequency span limit in thedisplay to get a better view of the Autosetresult. By default, the frequency span afterAutoset is set at 3MHz.

    Operation 1. Press Autoset > Span[F3] and switch the range

    from Auto to Man.

    2. Enter the span frequency for the Autosetsearch.

    Manual Range: 100Hz to 3GHz

  • 7/27/2019 Gsp 900 Manual Final

    63/255

    BASIC OPERATION

    61

    Bandwidth/Average SettingsBW/Avg key sets the resolution bandwidth (RBW), video

    bandwidth (VBW) and averaging functions. The resolution, sweeptime, and averaging are in a trade-off relationship, so configurationshould be done with care.

    Resolution Bandwidth Setting (RBW)

    Description RBW (Resolution Bandwidth) defines the width

    of the IF (intermediate frequency) filter that isused to separate signal peaks from one another.The narrower the RBW, the greater thecapability to separate signals at closefrequencies. But it also makes the sweep timelonger under specific frequency spans (thedisplay is updated less frequently).

    Operation 1. Press BW/Avg > RBW[F1] and set the RBW to

    Auto or Man.

    2. Set the resolution bandwidth and unit for Manmode.

    Mode:Frequency Range(3dB):

    Frequency Range(6dB):

    Auto, Man10Hz~3kHz (1-3-10 step)10kHz~1MHz (10% step)

    200Hz, 9kHz, 120kHz

    Display Icon The BW icon is displayed at the bottom ofthe screen when the RBW is in Man mode.

    NoteIf the RBW settings have an asterisk (*), itindicates that the -6dB filters are used.

  • 7/27/2019 Gsp 900 Manual Final

    64/255

    GSP-930 User Manual

    62

    Video Bandwidth Settings (VBW)

    Description VBW (Video Bandwidth) defines thesmoothness of the trace on the display.Combined with RBW, VBW defines the abilityto sort out the target signal from surroundingnoise or adjacent peaks.

    Operation 1. Press BW/Avg > VBW[F2] and set the VBW to

    Auto or Man.

    2. Set the video bandwidth and unit for Manmode.

    Mode:Frequency Range(3dB):

    Auto, Man1Hz~1MHz (1-3-10 step)

    Display Icon The BW icon is displayed at the bottom of

    the screen when the VBW is in Man mode.

    VBW/RBW Ratio

    Description The VBW/RBW function is used to view theratio between the video bandwidth and theresolution bandwidth.

    The VBW/RBW ratio is altered by setting theRBW and or VBW settings, see page 61 & 62respectively.

    View VBW/RBWratio

    1. Press BW/Avg .

    2. The ratio is displayed on the VBW/RBW[F3] softkey.

  • 7/27/2019 Gsp 900 Manual Final

    65/255

    BASIC OPERATION

    63

    Display

    VBW/RBWRatio

    Tip Signals that are masked by the noise floor levelshould have a ratio of less than 1 to smooth thenoise out.

    Signals with strong frequency components shoulduse a ratio equal to or greater than 1.

    Average Trace

    Description The Average function averages the trace for auser-defined number of times before it is

    displayed. This feature smoothes the noiselevel, but has the drawback of slowing downthe display update rate.

    Operation 1. Press BW/Avg >Average[F4] and toggle Average

    on or off.

    2. Set the number of averages.

    Range:Default:

    4 ~ 20020

    Display Icon The AVG icon is displayed at the bottom ofthe screen when the Average function ison.

  • 7/27/2019 Gsp 900 Manual Final

    66/255

    GSP-930 User Manual

    64

    Example:

    Average:Off

    Average: On (20)

  • 7/27/2019 Gsp 900 Manual Final

    67/255

    BASIC OPERATION

    65

    Average Type

    Description The Average Type function determines how theGSP-930 determines the average value.

    LOG Average: Averages the trace points on alogarithmic scale.

    Volt Average: Averages the amplitudes of thetrace points on a linear voltage scale.

    Power Average: Averages the trace points on alogarithmic scale in watts.

    Operation 1. Press BW/Avg >Average Type[F5] and choose the

    average type.

    Range:

    Default:

    LOG Power, VoltAverage, Power AverageLOG Power

  • 7/27/2019 Gsp 900 Manual Final

    68/255

    GSP-930 User Manual

    66

    EMI Filter

    Description The built-in EMI filter is used for specificmeasurement situations such as EMI averagedetection, where a higher level of sensitivity isrequired than the standard configuration.When turned on, the RBW is set to -6dB,indicated by an asterisk (*).

    When any measurement functions are turned

    on (see page 111 for details), the EMI filter isautomatically disabled. Conversely if the EMIfilter is turned on, any measurement functionsare turned off.

    Operation 1. Press BW/Avg > EMI Filter[F6] and toggle EMI

    filter on or off.

    SweepThe GSP-930 has a number of sweep options including setting thesweep time and sweep mode(continuous, single). The GSP-930 alsohas gated sweep modes.

    Sweep Time

    Description Sweep time defines the length of time thesystem takes to "sweep the current frequencyspan. Note, however, that sweep time andRBW/VBW are in a trade-off. Faster sweeptimes update the display more frequently butmake RBW and VBW wider, reducing thecapability to separate signals at closefrequencies.

  • 7/27/2019 Gsp 900 Manual Final

    69/255

    BASIC OPERATION

    67

    Operation 1. Press Sweep > Sweep Time[F1] and toggle the

    Sweep time to Auto or Man.

    2. Set the sweep time for the Man mode.

    Mode:Range:

    Resolution:

    Auto, Man22ms ~ 1000s (span>0Hz)50us ~ 1000s (span=0Hz)10us

    Display Icon The SWT icon will be displayed at the

    bottom of the screen when in the sweep isin manual mode.

    Single Sweep

    Description The single sweep function is used to perform asingle sweep. When Sweep Single is pressed theGSP-930 will perform a single sweep and thenstop.

    Operation 1. Press Sweep > Sweep Single[F2] to put the

    spectrum analyzer into single sweep mode.

    2. Press Sweep Single[F2] again to perform a singlesweep.

    Display Icon The Sweep Single icon is displayed on theright-hand side of the screen when thesweep is in single mode.

    NoteYou must wait for the single sweep to finish beforepressing the Single Sweep key again.

    If a setting is changed whilst the spectrum analyzer

    is still sweeping, the single sweep will immediatelystart over.

  • 7/27/2019 Gsp 900 Manual Final

    70/255

    GSP-930 User Manual

    68

    Continuous Sweep

    Description The GSP-930 has two main sweeping modes:single and continuous. Use the continuousmode to have the sweep constantly updated.

    Operation 1. Press Sweep > Sweep Cont[F3] to put the

    spectrum analyzer into continuous sweepmode.

    Display Icon The Sweep Cont icon is displayed on theright-hand side of the screen when thesweep is in continuous mode.

    NoteThe GSP-930 will now continuously sweep unlessthe mode is changed to single sweep mode or ifthe system is waiting for a trigger condition.

    Gated Sweep Overview

    Description The Gated Sweep mode allows a trigger signalto dictate when the spectrum analyzer cansweep. This mode is useful for characterizingsignals that are pulsed on and off, such as RFburst transmissions or for measuring spuriousnoise levels between transmission bursts.

    Overview 1. The trigger signal must be synchronized to theperiod of the input signal (shown as RF burstbelow).

    2. The start of the gate time is produced from thepositive or negative edge of the trigger signal +the delay time.

    3. The end of the gate time is determined by the

  • 7/27/2019 Gsp 900 Manual Final

    71/255

    BASIC OPERATION

    69

    set gate length.

    4. The gated sweep should not be positioned ateither end of the transmission.

    Delay Length

    Gate

    Trigger

    RF Burst

    period

    Example: The diagram above demonstrates therelationship between the input trigger, the inputsignal and the position of the gated sweep relativeto the input signal.

    NotePlease take into consideration RBW settling time.Setting the delay time too short may not leaveenough time for the RBW filter to resolve.

  • 7/27/2019 Gsp 900 Manual Final

    72/255

    GSP-930 User Manual

    70

    Using the Gated Sweep Mode

    Connection 1. Connect a trigger signal (3.3v CMOS) to theGATE IN port on the rear panel.

    Trigger

    Operation 1. Press Sweep > GateDelay[F5] and set the gate

    delay time.

    2. Press Sweep > Gated Length[F6] and set the gate

    time length.

    3. Press Sweep > Gated Sweep[F4] and turn the

    mode on.

    Gate Delay:Gate Length:

    0s ~ 1000s10us ~ 1000s

    Display Icon The Sweep Gated icon is displayed whenGated Sweep is turned on.

  • 7/27/2019 Gsp 900 Manual Final

    73/255

    BASIC OPERATION

    71

    Example: The example below shows the spectrum of an FSKmodulated signal when gated sweep mode is off.

    The example below shows the same signal with thegated sweep timed to sweep when only the desiredfrequency is output.

    Gated sweep icon

    NoteGate Delay and Gate Length must first be setbefore Gated Sweep is turned on.

  • 7/27/2019 Gsp 900 Manual Final

    74/255

    GSP-930 User Manual

    72

    TraceThe GSP-930 is able to set the parameters of up to 4 different

    trances on the display at once. Each trace is represented by adifferent color and is updated with each sweep.

    Selecting a Trace

    Description Each trace (1, 2, 3, 4) is represented by adifferent color. When activated, an icon for each

    trace color and function is shown to the left ofthe display. When a trace is selected,parameters can be set/edited from the tracemenu.

    Trace Color: 1: Yellow2: Pink3: Blue4: Red

    Trace Type The type of trace used determines how the tracedata is stored or manipulated before beingdisplayed. The analyzer updates each traceaccording to the type of trace used.

    Clear andWrite

    The GSP-930 continuouslyupdates the display with each

    sweep.

    Hold Max/Hold Min

    The maximum or minimumpoints are maintained for theselected trace. The trace pointsare updated each sweep if newmaximum or minimum pointsare found.

  • 7/27/2019 Gsp 900 Manual Final

    75/255

    BASIC OPERATION

    73

    View View will hold the selectedtrace and stop updating thetrace data for the selected trace.

    Pressing View[F4] will displaythe trace data that was clearedusing the Blank[F5] key.

    Blank Clears the selected trace fromthe display and stores tracedata. The trace data can berestored by pressing View[F4].

    Display Icon

    Clear & Write

    Hold Max

    View

    Blank

    Trace 1

    Trace 2

    Trace 3

    Trace 4

    Operation 1. Press Trace > Trace[F1] and choose the trace

    number.

    Trace: 1, 2, 3, 4

    2. Select the trace type:

    Clear & Write[F2]Max Hold[F3]Min Hold[F4]View[F5]Blank[F6]

    Note Traces, 2, 3 and 4 are set to Blankby default.

  • 7/27/2019 Gsp 900 Manual Final

    76/255

    GSP-930 User Manual

    74

    Trace Math

    Description Performs trace math from two traces (TR1, TR2)and stores the result in the currently selectedtrace. It also performs trace shift.

    Math functions Power Diff Subtracts the TR1 amplitudedata from the TR2 amplitudedata. The TR1 data TR2 dataare converted to watts. The

    result is converted back todBm.

    Log Diff Subtracts the TR1 amplitudedata from the TR2 amplitudedata and then adds alogarithmic reference. Both theTR1 and TR2 data is in dBm.The resultant trace of thesubtraction is in dB. When theresult is added to a logarithmicreference the resulting data isin dBm.

    LOG Offset Adds a reference to the TR1trance

    Operation 1. Press Trace >More[F1]>Trace Math[F1].

    2. Press TR1[F1] and select the first trace source:

    TR1: Trace 2, Trace 3, Trace 4

    3. Press TR2[F2] and select the second tracesource:

    TR2: Trace 2, Trace 3, Trace 4

  • 7/27/2019 Gsp 900 Manual Final

    77/255

    BASIC OPERATION

    75

    4. Select the trace math function:

    PowerDiff[F3]

    LogDiff[F4]LogOffset[F5]

    5. If LogDiff was selected, set the reference leveland unit.

    LogDiff ref range:LogDiff ref units:

    -120dBm ~ 30dBmdBm, W

    6. If LogOffset was selected, set the offset leveland unit.

    LogOffset range: -50dB~+50dB

    7. To turn trace math off, press the OFF[F6].

    Display Icon The Math icon is displayed when tracemath is turned on.

    Trace Detection Mode

    Description Each time the spectrum analyzer samples datafor each point on the trace, a number of samples

    are usually taken for each point, known as asample bucket. The actual value of each point isdetermined by the detector from the samples ineach bucket.

    Each selected trace, (1, 2, 3, 4), can use adifferent detection mode.

  • 7/27/2019 Gsp 900 Manual Final

    78/255

    GSP-930 User Manual

    76

    Detection modes Auto Automatically chooses anappropriate mode based on thevalues of all the samples.

    Normal While the signal level is constantlyincreasing or decreasing, thepositive peaks are detected.Otherwise, detecting modeswitches between positive peakand negative peaks. Useful forpicking up burst phenomenon

    while avoiding excessive noise.

    Peak+ Detects positive peak signals byselecting the maximum peak valuefor each point from each bucket.This mode is useful for sinusoidalsignals.

    Peak- Detects negative peak signals by

    selecting the lowest peak value foreach point from each bucket. Thismode is not recommended foramplitude measurement.

    Sample Randomly selects a value from thebucket sample. Useful for noisesignals.

    Average Calculates the average of all thesamples in the sample bucket.

  • 7/27/2019 Gsp 900 Manual Final

    79/255

    BASIC OPERATION

    77

    Auto DetectorSelection Method

    Below is a flow chart diagram showing theDetector selection for the Auto mode.

    Is (trace) Average

    on?

    Are any traces set

    to Max Hold?

    Are any traces setto Min Hold?

    Is Average Type

    set to Power?

    AverageSample

    Peak+

    Peak-

    Normal

    Y

    N

    Y

    N

    Y

    N

    N Y

    Is Phase Jitter

    measurement on?Sample

    Y

    N

    Are any windows

    set to Topographic?Peak+

    Y

    N

    Detector(Auto mode)

    Operation 1. Press Trace >More[F7]>Detection[F2].

    2. Select the trace detection mode for the selected

    trace:

    Auto[F1]Normal[F2]Peak+[F3]Peak-[F4]Sample[F5]Average[F6]

  • 7/27/2019 Gsp 900 Manual Final

    80/255

    GSP-930 User Manual

    78

    3. The display will return the Trace menu.

    Display Icon Normal Average icon

    Peak+ icon Peak - icon

    Sample icon Average icon

  • 7/27/2019 Gsp 900 Manual Final

    81/255

  • 7/27/2019 Gsp 900 Manual Final

    82/255

    GSP-930 User Manual

    80

    Parameters Video Edge: Determines the polarity of thevideo trigger.

    Positive: The signal voltageexceeds the video level at thetrigger frequency.

    Negative: The signal voltage islower than the video level at thetrigger frequency.

    Video Level:The trigger voltage level.

    TriggerFrequency:

    Sets the frequency to starttriggering

    Operation 1. Press Trigger >Trigger Condition[F2]>Video[F1]

    2. Press Video Edge[F1] and choose the edge.

    Range: Positive, Negative

    3. Press Video Level[F2] and set the video voltagetrigger level.

    Trigger level: (-120dBm to +30dBm)+Ref Level Offset

    4. Press Trigger Freq[F3] and choose the frequencyat which the spectrum analyzer will check thetriggering conditions.

    Frequency: 0-3GHz+frequency offset

    NoteSet the trigger back to Free Run to disable thevideo trigger.

  • 7/27/2019 Gsp 900 Manual Final

    83/255

    BASIC OPERATION

    81

    Activate External Trigger

    Description The external trigger is used when an externaltrigger signal is input into the rear panel TRIGIN port. The external trigger signal can beconfigured as positive or negative edge.

    Trigger: 3.3V, CMOS

    Operation 1. Press Trigger >Trigger Condition[F2]>Ext.Edge[F2]

    and select the trigger edge:

    Pos:Neg:

    Positive edgeNegative edge

    2. Connect the external trigger signal to the rearpanel TRIG IN port.

    Trigger

    3. PressAction Now[F5] to activate the externaltrigger.

    4. The system will now wait the trigger conditionsto be matched before starting a sweep.

    Display Icon The EXT Trigger icon is displayed when theexternal trigger is activated.

    NoteThe trigger will revert back to the Free Run mode ifany parameter settings are changed, such as thespan or amplitude settings.

  • 7/27/2019 Gsp 900 Manual Final

    84/255

    GSP-930 User Manual

    82

    Selecting the Trigger Mode

    Description In free run mode all signals are captured andthe trigger conditions are not used.

    Modes Normal: The spectrum analyzer capturesevery signal that meets the triggerconditions.

    Single: The spectrum analyzer captures

    the first signal that meets thetrigger conditions.

    Continuous: The spectrum analyzer capturesthe first signal that meets thetrigger conditions then switches tofree run mode thereafter.

    Operation 1.

    PressTrigger

    > Trigger Mode[F3] to toggle thetrigger mode:

    Nor.:Sgl.:Cont.:

    NormalSingleContinuous

    2. PressAction Now[F5] to manually starttriggering.

  • 7/27/2019 Gsp 900 Manual Final

    85/255

    BASIC OPERATION

    83

    Set the Trigger Delay Time

    Description Sets the delay time between when the analyzertriggers and when the analyzer begins tocapture the signal.

    Delay time range: 1ns to 1ks

    Operation 1. Press Trigger >Trigger Delay[F4] and set the

    trigger delay time.

    Delay range: 0~1000s

  • 7/27/2019 Gsp 900 Manual Final

    86/255

    GSP-930 User Manual

    84

    MarkerA Marker shows the frequency and amplitude of a waveform point.

    The GSP-930 can activate up to 6 markers or marker pairssimultaneously as well as up to 10 peak markers in the marker table.

    The marker table helps editing and viewing multiple markers in asingle display.

    A delta marker shows the frequency and amplitude differencesbetween the reference marker.

    The GSP-930 can automatically move a marker to various locationsincluding the peak signal, center frequency, and start/stop

    frequency. Other marker operations regarding signal peaks areavailable in the Peak Search function.

    Activating a Marker from page85

    Move Marker Manually from page86

    Move Marker to Preset Locations from page86

    Activate Delta Marker from page87

    Move Delta Marker(s)Manually from page 88

    Marker Functions from page89

    Move Marker to Trace from page91

    Show Markers in Table from page92

    Peak Search from page93

    Peak Configuration from page95

    Peak Table from page96

  • 7/27/2019 Gsp 900 Manual Final

    87/255

    BASIC OPERATION

    85

    Activating a Marker

    There are two basic marker types, normal markers and delta

    markers. Normal markers are used to measure the frequency/timeor amplitude of a point on the trace. Delta markers are used tomeasure the difference between a reference point and a selectedpoint on the trace.

    Activate a Normal Marker

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    Marker: 1~6

    2. Press [F2] to turn the selected marker on.

    3. Press Normal[F3] to set the selected marker tothe Normal type.

    4. The display will show the marker on the trace(centered by default) with the markermeasurement at the top of the display.

    Marker

    Maker ID, Frequency, Amplitude

  • 7/27/2019 Gsp 900 Manual Final

    88/255

    GSP-930 User Manual

    86

    Move Marker Manually

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. Use the left/right arrow keys tomove the marker one grid division.

    3. Use the scroll wheel to move themarker in fine increments.

    4. Alternatively, the numeric keypadcan be used to directly enter thefrequency of the marker position.

    7 8 9

    654

    1 2 3

    /0

    Move Marker to Preset Locations

    Description The Marker-> key is used to move the selected

    marker to a number of preset positions.

    Functions Mkr>Center:Mkr>Start:Mkr>Stop:Mkr>CF Step:Mkr>Ref Lvl:

    Move to center frequency.Move to start frequency.Move to stop frequency.Move to step frequency.Move to reference level

    amplitude.

    NoteWhen the Marker-> key is used, the span and other

    settings may be automatically changed.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. Press Marker-> and select a marker position:

  • 7/27/2019 Gsp 900 Manual Final

    89/255

  • 7/27/2019 Gsp 900 Manual Final

    90/255

    GSP-930 User Manual

    88

    Move Delta Marker(s)Manually

    Move Delta orReference Marker 1. PressMarker

    > Delta[F4]>Move[F2]>Move Ref[F2]to move the reference marker.

    2. Press Marker > Delta[F4]>Move[F2]>Move

    Delta[F3] to move the Delta marker.

    3. Move the selected marker in the same fashionas a normal marker, see page 86

    Move Bothreference anddelta marker

    1. Press eitherMove Pair Span[F4] orMove PairCenter[F5] to move both markers at the sametime.

    Move Pair Span: Sets the frequency spanbetween both markers.The span can be positiveor negative:

    1+span1

    1-span1

    Move Pair Center: Moves both markers atthe same time, keepingthe span between both

    markers eventhroughout.

    2. Move both markers in the same fashion as anormal marker, see page 86

  • 7/27/2019 Gsp 900 Manual Final

    91/255

    BASIC OPERATION

    89

    Marker Functions

    Marker Noise

    Description The noise marker function calculates theaverage noise level over a bandwidth of 1Hz,referenced from the marker position.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. Press [F2] to turn the selected marker on.

    3. Press Normal[F3] and then position the markerto the desired location.

    4. Press Function[F5]>Marker Noise and turnMarker Noise on.

    5. The display will show the noise levelmeasurement at the top of the screen indBm/Hz.

    Marker ID, Frequency, dBm/Hz

    Marker

  • 7/27/2019 Gsp 900 Manual Final

    92/255

    GSP-930 User Manual

    90

    Frequency Counter

    Description The frequency counter function is used to makeaccurate frequency measurements.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. Press [F2] to turn the selected marker on.

    3. Press Normal[F3] and then position the markerto the desired location.

    4. Press Function[F5]>Frequency Counter[F1] andturn the counter function on.

    5. Press Resolution[F2] and set the resolution:

    Auto: Automatically chooses

    the best resolution.Man: Allows the resolution to

    be manually set.

    Man Range: 1Hz, 10Hz, 100Hz, 1kHz

    6. The display will show the frequencymeasurement at the top of the screen at the

    selected resolution.

    Frequency

    Marker

  • 7/27/2019 Gsp 900 Manual Final

    93/255

    BASIC OPERATION

    91

    Move Marker to Trace

    Description The Marker Trace function moves the selectedmarker to any of the currently active traces.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. Press [F2] to turn the selected marker on.

    3. PressMore 1/2[F7]>Marker Trace[F1] and choosea trace to move the current marker to. Onlyactive traces can be selected.

    Auto[F1]Trace1[F2]Trace2[F3]Trace3[F4]Trace4[F5]

    4. In the example below, marker 1 is set to Trace1and marker 2 is set to Trace2.

    Marker 1, Trace 1

    Marker 2, Trace 2

  • 7/27/2019 Gsp 900 Manual Final

    94/255

    GSP-930 User Manual

    92

    Show Markers in Table

    Description The GSP-930 has a Marker Table function to

    show all the active markers and measurementsat once.

    Operation 1. Press Marker >More 1/2[F7]>Marker Table[F2] and

    turn the marker table on.

    2. The display will split into two screens. Thebottom half will show the Marker Table withthe marker ID(normal, reference or delta), trace,x-axis position (frequency/time) and theamplitude of the marker.

    Marker Table

  • 7/27/2019 Gsp 900 Manual Final

    95/255

    BASIC OPERATION

    93

    Peak Search

    Move Marker to Peak

    Description ThePeak

    Search key is used to find trace peaks.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. PressPeak

    Search >Peak Search[F1]. The marker will

    move to the highest signal peak.

    3. To continually search for the peak each sweep,

    press,Peak

    Search >More 1/2[F7]>Peak Track[F1] and

    set Peak Track to on.

    Move Marker and Peak to Center

    Description The Center function moves the marker to thehighest signal peak and moves the centerfrequency to that peak.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. PressPeak

    Search >Mkr>Center[F2].

    NoteThe span will not be changed.

  • 7/27/2019 Gsp 900 Manual Final

    96/255

    GSP-930 User Manual

    94

    Search for Peaks

    Description ThePeak

    Search key can be used to search for anumber of different peaks.

    Peak Search Next Peak: Searches for next highestpeak visible on the display.

    Next Peak Right: Searches for the next peak tothe right of the marker.

    Next Peak Left: Searches for the next peak tothe left of the marker.

    Min Search: Searches for the lowest peak.

    Operation 1. Press Marker > Select Marker[F1] and select a

    marker number.

    2. PressPeak

    Search and select the type of peak you

    wish to find.

    Example:Next Peak

  • 7/27/2019 Gsp 900 Manual Final

    97/255

    BASIC OPERATION

    95

    Example:Next Peak Right

    Example:Next Peak Left

    Peak Configuration

    Description There are two peak search configurationoptions: Peak Excursion and Peak Threshold.

    Peak Excursion: Peak Excursion sets theminimum value above thepeak threshold for whichpeaks will be detected.

  • 7/27/2019 Gsp 900 Manual Final

    98/255

    GSP-930 User Manual

    96

    Peak Threshold: Peak threshold sets theminimum threshold level forthe analyzer to detect peaks.

    Any value above the PeakThreshold + Peak Excursionwill be detected as a peak.

    Peak ThresholdPeak Excursion

    1 Peakdetection area

    Operation 1. PressPeak

    Search >More 1/2[F7].

    2. Press Peak Excursion[F2] to set the excursionlevel.

    3. Press Peak Threshold[F3] to set the peakthreshold.

    Peak Excursion: 0~100dB

    Peak Threshold: -120dB~+30dB

    Peak Table

    Description The Peak Table function will display all peaks(up to 10) that meet the peak configurationsettings. The amplitude and frequency for eachpeak is listed.

    Operation 1. PressPeak

    Search >More 1/2[F7]>Peak Table[F5].

  • 7/27/2019 Gsp 900 Manual Final

    99/255

    BASIC OPERATION

    97

    2. Press Peak Sort[F2] and set the sorting type:

    Freq:

    Amp:

    Sort by frequency in

    ascending order.Sort by amplitude inascending order.

    3. Press Peak Table[F1] to turn the peak table on.

    4. The display splits in two. The bottom screenshows the peak table with the peak marker ID,

    X-axis position and amplitude.

    Peak Table

    Peak markers

  • 7/27/2019 Gsp 900 Manual Final

    100/255

    GSP-930 User Manual

    98

    DisplayThe Display key configures the basic display settings as well as

    setting up the display mode (spectrum, spectrographic,topographic) and the split screen modes.

    Adjusting the LCD Brightness

    Description The LCD brightness levels can be adjusted tothree pre-set levels.

    Operation 1. Press Display > LCD Brightness[F2] to toggle the

    display brightness:

    Hi:Mid:Lo:

    High brightnessMedium brightnessLow brightness

    Turning the LCD Backlight Off

    Description The LCD backlight can be turned off topreserve power or to prolong the lifetime of theLCD display when not in use.

    Operation 1. Press Display > LCD Backlight[F3] and turn the

    LCD backlight off.

    2. When the backlight is off, press any functionkey to turn the LCD backlight back on.

  • 7/27/2019 Gsp 900 Manual Final

    101/255

    BASIC OPERATION

    99

    Setting a Display Line (Reference Level Line)

    Description The Display Line function is used to super-impose a reference level line over the traces.

    Operation 1. Press Display > Display Line[F4] to turn the

    display line on.

    2. Set the display line level and unit.

    Example:

    Display line

    Display line set at -50dBm

    Using the Video Out Port

    Description The GSP-930 has a dedicated DVI terminal to

    output the display to an external monitor. Thevideo output is always on.

    Output resolution 800 x 600 (fixed)

    Operation 1. Connect an external monitor to the rear panelDVI terminal.

  • 7/27/2019 Gsp 900 Manual Final

    102/255

    GSP-930 User Manual

    100

    Setting the Display Mode

    Description The GSP-930 has three different display modesfor viewing: spectrum, spectrograph andtopographic. It is also possible to view thespectrum with the spectrographic ortopographic views using a split screen.

    Spectrum Default display mode.

    Spectrogram Useful for viewingfrequency or power inthe time domain.

    Topographic Useful for observing thefrequency of events witha trace.

    Operation 1.

    PressDisplay

    >Window Setup[F1] and select thedisplay mode:

    Spectrum[F1]:Spectrogram[F3]:Topographic[F4]:Spectrogram+Spectrum[F5]:Topographic+Spectrum[F6]:

    NoteThe same trace is used on the top and bottom forthe Spectrogram+Spectrum andTopographic+Spectrum modes.

  • 7/27/2019 Gsp 900 Manual Final

    103/255

    BASIC OPERATION

    101

    Example:Spectrogram

    The Spectrogram view shows signals in boththe frequency and time domain. The X-axisrepresents frequency, the Y-axis represents timeand the color of each point represents theamplitude at a particul