Gravimètre Quantique Absolu : une utilisation...

30
Gravimètre Quantique Absolu : une utilisation opérationnelle des atomes froids pour la mesure de gravité Dr. Jean Lautier-Gaud Journée CNFGG

Transcript of Gravimètre Quantique Absolu : une utilisation...

GravimètreQuantiqueAbsolu:uneutilisationopérationnelledesatomes

froidspourlamesuredegravité

Dr.JeanLautier-GaudJournéeCNFGG

Outline

• Brief overview ofMuquans

• Absolutegravimetry with coldatoms

• TheAbsoluteQuantumGravimeter(AQG)

– Instrumentdescription

– Operation

• Firstresults &status

2

OurcompanyFounded in 2011, Muquans delivers turnkey high-precision sensors based on quantum manipulation oflaser-cooled atoms

Muquansprovides acomplete lineofhigh-performancescientific instrumentsdedicated tometrology :

100millionofatomstemperature ofafewµK

AQG MuClock Lasersystems

Frequencytransfer links

MUQUANS: COMMERCIALLY PROVIDEOPERATIONAL AND TURNKEY HIGH-PRECISION SENSORSAND RESEARCH-GRADE LASER SYSTEMS

Technology transfers from SYRTE laboratory(Observatoire de Paris) & Institut d’Optique

Active in more than 20 countries

Team of 24 engineers, scientists, and techniciansstate-of-the-art laboratories & production lines

Core technology developed for over 15 years

Muquans is located in the Bordeaux area

Ourteamandtechnical capacities

• Muquansgathers adeep expertiseoverallthekeycompetencies :– Quantumphysics– Optics & laser– Systemengineering, opto-mechanics andultra-highvacuum– Electronics,RFandmicro-wave– Real-timesoftware,dataacquisition,andsignalprocessing

• Fully equiped laboratories andproductionlines tocontroleach step ofour activity

High-precision gravimetrywith cold-atoms

g = 9.8142627….. m/s2

Flattening

Mountains & trenches

Internal density anomalies

Large reservoirs Time variations due to tides

Big buildings

1000 µGals

Figure:OlivierdeVironIPGP

100 µGals 10 µGals 1 µGalà

Themeasurementofg• Absolutemeasurement ofg:freefall ofatestmassinvacuum.Here,testmass=ensembleoflasercooled atoms

We accurately monitorthebalistictrajectory ofthefreefalling atoms withlasers.

Measurement*sequence*

100millionofatomstemperature ofafewµK

8

AQGfollowstheleadoftheCAG• ColdAtom Gravimeterdeveloped atLNE-SYRTElaboratory• State-of-the-artperformances:

• 20+peer-reviewed publications• Only atomic gravimeter toparticipatetointernationalkeycomparisons(since 2009)

Accuracy 4,3μGal

Recordsensitivity 6μGal/√Hz

Ultimate stability 0,2μGal

P.Gillot,Metrologia 51,L15-L17(2014)

Themeasurementofg• Free-fall oftheatoms is compared toastanding-wavereferenced toafixed retroreflecting mirror.

- Verticalaccelerometer- Mach-Zehnder interferometer- Stimulated Ramantransitions

Measurement*sequence*

𝜑 ∝ 𝑎. 𝑇&

11

Take-homemessage1:Keyadvantages ofquantumgravimetry

• Absolute gravity measurements atthe μGal level withexcellentsensitivity, long-term stability,andaccuracy

• Continuous measurements formonths,fewHzrep.Rate

• Nomoving partsunder vacuum(reduced maintenance)

• Nomechanical nor optical alignment,noprimary pumping

• Robustness toground vibration

• Different measurement principle (metrology)

AbsoluteQuantumGravimeterofMuquans

à Instrumentdescriptionà Scientificresultsà AQGoperation

Absolute QuantumGravimeterFirstcommercialquantumgravimeter intheworld

Actual versiondedicated tocontrolled-environment13

Absolute QuantumGravimeterDegree ofmaturity ofthetechnology andlevel ofintegration

14

Absolute QuantumGravimeter• Compactatomic absolute gravimeter

Performances(demonstrated):

Physicalpackage:

Sensitivity 50μGal/√HzLongtermstability 1µGal

Accuracy fewµGal

DimensionsSensorhead:h=70cm,diam=38cmLaser:100x50x70cm3

Mass 100kg(instrumentonly)

Powerconsumption 300W

Nomechanical isolationplatformNosuperspring 15

PrincipleofoperationMeasurement*sequence*

Freefall=0.12sTemperature≈1µK

Measurementcyclingfrequency:afewHzQ.Bodart,etAl.« Acoldatompyramidalgravimeterwithasinglelaserbeam » ,AppliedPhysicsLetters96,134101(2010)

KeytechnologicalaspectsofAQG

• Hollow pyramid retroreflectorà Singlelaserbeam atom interferometer(instead of6usually)

• Fiber lasertelecom technologiesà Fibered,reliable androbust lasersystem

• Activecompensationofground vibrationà Excellentimmunity tovibrationnoise(nomechanical isolation)

Immunity toground vibrations• AQGmeasures therelativeacceleration between thefree-falling atoms andthesensor head,laying onthefloor:

ameasured =g +acceleration noise(vibration)FewHzcycling frequencyà aliasingofthevibrationnoise

• Vibrationrejection:correlation with aclassicalaccelerometer

Classical accelerationsignal

AtomicsignalAQGsensor

head

Classicalaccelerometer

g

Data• TheAQG-A01hasbeenrunningforalmost 2years now

àAsitecan be properly measured within 30minutes

Demonstratedsensitivity:50μGal/√Hz

Resolution:1.5μGalwithin 30min.

19

1,5months continuous data(1μGal=10-8 m/s2)

ComparisonAQGvsiGravJune2015(Larzac)

60hlongcont.meas.at2Hz

σres =2μGal

AQG

Noapparentdrift

Sensitivity inurban environment

Demonstrated sensitivity:50μGal/√Hz,1.5μGal reached in30min

• Measurements at2nd floor ofinner citybuilding(Talence,France)

4.5longcontinuous meas.at2Hz

σres =0.6μGal

Cf.Gmob

Sensitivity inurban environment• Measurements at2nd floor ofinner citybuilding(Talence,France)

Allanstandarddeviation

Sensitivity inurban environment• Measurements at2nd floor ofinner city

Allanstandarddeviation

Measurement campaign inLarzacComparison w/absolute grav.

FG5-228(N.LeMoigne,CNRS,France)

Larzac,France2015/06/12

FG5FG5

AQG

AQG

Measurement campaign inTrappesComparison w/gravimeters (S.Merlet,LNE-SYRTE,France)

Trappes,France2015/09/29

Deliveryoftheworld’s 1stcommercialatom gravimeter

DeliveryoftheAQG-01toRESIF(France) onMarch,29th2016

(N.LeMoigne &S.Bonvalot)

FG5FG5

AQG

AQG

27

Take homemessage2:Keyadvantages oftheAQG

• Advantages ofquantumgravimetry

+• Compacity andmobility• Easy andquicktooperate (technician level)• Low operator dependent measurements• Greatrobustness wrt toground vibrations

àNosuperspring

Actualstatus- 01• About10tripsoutside ofMuquansfacility (even toHolland)

• Transportability,robustness,easeandquickness toset-upà validated

• AQGready tomeasure within 1h1μGal level reached within 30min(quietplace)

• Excellentrepeatability

• Continuous measurements overmonths

à AQGis operationalAQGinits boxesinacar

Actualstatus- 02• Urban sensitivity :2μGal level reached within 15min

• Improvement ofweight &volume

• 2units under production

• Metrological assessment ofbiases

à development ofthefield version

AQGinits boxesinacar

Thankyouforyourattention!