40
General Specifications <<Contents>> <<Index>> InfraSpec NR800 Fourier Transform Near-Infrared Analyzers Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 GS 12Y03A03-01E GS 12Y03A03-01E ©Copyright Aug. 2001 5th Edition Mar. 18, 2014 n Overview The NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to- noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes. n Features (including options) Newly developed interferometer and detector Used for a variety of applications, ranging from over to combination tone. • High resolution: Up to 4 cm –1 , user selectable setting • High S/N ratio: 2250:1 (RMS, 4 cm -1 resolution, 4100 to 4200 cm -1 , 1 sec.) • Wide wavelengths scanning range 900 to 2500 nm (11,000 to 4000 cm -1 ) • Wavelength reproducibility: 0.007 cm -1 • Wavelength accuracy: 0.04 cm -1 Enhanced environmental resistance, durability, and reliability • Provides high vibration resistance by a unique design free of sliding parts • Features a multi-channel measurement optical system free of moving parts • Eliminates the need for a PC for continuous operation. A PC is now only necessary for generating the calibration model and loading data. Outstanding wavelength accuracy allows calibration model transport between NR800s Outlier detection and self-diagnostic features come as standard function Various standard features and available options for optimal system configuration • Dust-proof and drip-proof: IP53 equivalent or NEMA TYPE 3R • Optional Explosion-proof Enclosure: FM, CSA, ATEX, TIIS (*1), KOSHA • Line up with CE marking • Non-moving multi-channel measurement: Expandable to 4 channels • Up to 12 items can be measured per stream: limited to 64 items / unit • The optional I/O unit offers a variety of inputs/outputs: Analogue output (up to 40 points), analogue input, and contact input/output • Communication output: RS-422 • Fast Ethernet communication between Engineering PC and the analyzer (*1) TIIS: The Technology Institution of Industrial Safety Explosion proof Model

Transcript of General InfraSpec Specifications Near-Infrared Analyzers · General Specifications InfraSpec...

Page 1: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

GeneralSpecifications

<<Contents>> <<Index>>

InfraSpecNR800 Fourier TransformNear-Infrared Analyzers

Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JapanTel.: 81-422-52-5617 Fax.: 81-422-52-6792

GS 12Y03A03-01E

GS 12Y03A03-01E©Copyright Aug. 2001

5th Edition Mar. 18, 2014

n OverviewThe NR800 Fourier Transform Near-Infrared Analyzers offers high-resolution, high S/N (signal-to-noise) ratio, and wide wavelength scanning range measurement with its newly developed interferometer and detector. The NR800 also offers high stability, vibration resistance, and durability, inheriting features from NR500 earlier successful models. The NR800 allows online, real-time, continuous, multiple, and simultaneous measurement for properties and component concentration of various processes.

nFeatures(includingoptions) Newlydevelopedinterferometeranddetector

Usedforavarietyofapplications,rangingfromovertocombinationtone.• Highresolution:Upto4cm–1, user selectable

setting• HighS/Nratio:2250:1(RMS,4cm-1 resolution,

4100 to 4200 cm-1 , 1 sec.)• Widewavelengthsscanningrange900to2500nm(11,000to4000cm-1)

• Wavelengthreproducibility:0.007cm-1

• Wavelengthaccuracy:0.04cm-1

Enhancedenvironmentalresistance,durability,andreliability• Provideshighvibrationresistancebyaunique

design free of sliding parts• Featuresamulti-channelmeasurementopticalsystemfreeofmovingparts

• EliminatestheneedforaPCforcontinuousoperation.

APCisnowonlynecessaryforgeneratingthecalibration model and loading data.

OutstandingwavelengthaccuracyallowscalibrationmodeltransportbetweenNR800s

Outlierdetectionandself-diagnosticfeaturescomeasstandardfunction

Variousstandardfeaturesandavailableoptionsforoptimalsystemconfiguration• Dust-proofanddrip-proof:IP53equivalentorNEMATYPE3R

• OptionalExplosion-proofEnclosure:FM,CSA,ATEX,TIIS(*1),KOSHA

• LineupwithCEmarking• Non-movingmulti-channelmeasurement:

Expandable to 4 channels• Upto12itemscanbemeasuredperstream:limitedto64items/unit

• TheoptionalI/Ounitoffersavarietyofinputs/outputs:

Analogueoutput(upto40points),analogueinput,andcontactinput/output

• Communicationoutput:RS-422• FastEthernetcommunicationbetweenEngineeringPCandtheanalyzer

(*1) TIIS:TheTechnologyInstitutionofIndustrialSafety

ExplosionproofModel

Page 2: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

2

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E Mar. 18, 2014-00

1.Configuration1.1 Configurationexamples

• Modbusoutputwithsamplingunit(explosion-proof,1measuringchannel)

RS-422/RS-232Cconverter *2

RS-232C(MODBUS)

(15m max.)(1km max.)

Sampleejection

Process

Fiber-optic cable *2

Clean, dry air

Hazardous area Non-hazardous area

Alarm contact output

Optical/electricalsignal converter

Monitor/ControlSystem

Engineering PC

Samplingunit

Ethernet(Fiber-optic cable *1)

RS-422(MODBUS)

Ethernet(10 Base-T)

InfraSpecNR800

explosion-proofmodel

Measurement

cell

• Analogueoutputwithsamplingunit(explosion-proof,2measuringchannels)

I/O unit*2

RS-422Process 2

Process 1

Sampleejection

Sampleejection

Clean, dry air

Fiber-optic cable *2

Hazardous area Non-hazardous area

Optical/electricalsignal

Alarm contact output

Engineering PC

Ethernet(10 Base-T)

Analog signal

Contact signal

Contact signal

Monitor/ControlSystem

Sampling cell 1

Sampling cell 2

Ethernet(Fiber-optic cable *1)

InfraSpecNR800

explosion-proofmodel

Measurementcell 1

Measurementcell 2 (1km max.)

*1: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel,anelectriccablecannotbeused. For a general purpose model, the type of cable depends on the length. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused.*2: OnlyusetheRS-422outputofanexplosion-proofmodelincombinationwithanI/Ounit(seesection2.2,selectSuffixcode

toExplosion-proofmodel)orRS-422/RS-232Cconverter(seesection2.5).TheseunitswillblockthecommunicationsignaluponreceivingaPurgefailuresignalfromtheAnalyzer.

Andusethededicatedfiber-opticcableformeasurement(seesection2.4)andEthernetcommunication(seesection2.6). Thus, ensuring the explosion-proof integrity of the analyzer.

Page 3: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

3<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

1.2 Component,softwareandcalibrationmodelgeneration

Item Require-ment(*1) Model Description Reference

1 InfraSpec NR800 analyzer A

NR801AG General purpose model, CSA standard

1.32.1

NR805AG Explosion-proofmodel,FMcertified(excludingNR805AG/CSA),CSAcertified(onlyforNR805AG/CSA)

NR801EG General purpose model with CE markingNR805EG Explosion-proofmodel,ATEXcertifiedandCEmarkingNR801JG Generalpurposemodel,Japaneseversion

NR805JG Explosion-proofmodel,TIIScertified(excludingNR805JG/K),KOSHAcertified(onlyforNR805JG/K)

2 I/Ounit BNR893AG To be used with NR801,805AG, CSA standard

1.32.1NR893EG To be used with NR801,805EG, and CE marking

NR893JG To be used with NR801,805JG

3 Measurement cell A

NR510 Flow through cell 2.3.1NR512 Flow through cell with constant temperature water tube 2.3.2NR511 Special measurement cell 2.3.3FIR200 In-situ probe 2.3.4

4 Fiber-optic cable for measurement A

NR821 Applicablewavelengthrange:900to2100nm,Silica,single2.4.1

NR822 Applicablewavelengthrange:900to2100nm,Silica,dualNR823 Applicablewavelengthrange:900to2500nm,Fluoride,single

2.4.2NR824 Applicablewavelengthrange:900to2500nm,Fluoride,dual

5 RS-422/RS-232Cconverter B

K9404LAor

K9404LDConvertstheRS-422signalfromtheanalyzerintoRS-232C. 2.5

6 Ethernetcable(*2) BNR895 Dedicatedfiber-opticcable(*3) 2.6

- Electriccable,providedbyuser 6.2

7 SoftwareA NR831 SPECTLAND2managementandmaintenancesoftware 2.7B NR530 Chemometrics software 2.8

8 Sampling unit B J439 Yokogawa will propose an optimum unit based on sample pressure, temperature, properties, and measurement items 2.9

9On-siteguidanceof calibration model generation

B J964 Hands-onpracticeandguidanceformodelgenerationonsite. 5.1

10 Calibration model generation B J965 Calibration model generation by Yokogawa based on user-

providedsamplewithlaboratoryanalysisresults 5.2

11 EngineeringPC A - Providedbyuser.Seerecommendedspecifications. 6.1

12Optical/electricalsignalconverterforEthernet

B -ConvertsopticalsignalsforanEthernetoutputintoelectricalsignalsforengineeringPCInterface.Providedbyuser.Seerecommendedspecifications

6.2

13 Customer inspection

B J963 For Instrument except following J443 -B J443 For J439 Sampling unit and FIR200 In-situ probe -

14 Equipmentstart-up B - Start-up work for analyzers and sampling units -

Notes*1: A:Required,B:optional*2: AnEthernetcableisrequired.Chooseeitheraspecialfiber-opticcableorelectriccabledependingonthefollowing

conditions: Alwaysuseaspecialfiber-opticcable(seesection2.6)foranexplosion-proofmodel,anelectriccablecannotbeused. For a general purpose model, the type of cable depends on the length. •Cablelengthlessthan40m:Eitheranelectricorspecialfiber-opticcable(seesection2.6and6.2)canbeused. •Cablelengthof40morlonger:Useaspecialfiber-opticcable(seesection2.6and6.2).Anelectriccablecannotbeused.*3: Foratotalcabledistanceoflongerthan20m,anadditionalfiber-opticcable(seesection2.6)shallbeprovidedbytheuser.

May 15, 2014-01

Page 4: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

4

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

1.3Safetystandard,EMCstandardModel SafetyStandard EMCstandard

FT-NIRanalyzer I/OunitNR801AG, NR805AG, NR805AG/CSA*2

NR893AG*2

CSAC22.2No.61010-1*1FMClass3810(excludeNR805AG/CSA)ANSI/ISA61010-1(excludeNR805AG/CSA)ANSI/ISA-82.02.02(IEC61010-2-031)(excludeNR805AG/CSA)

-

NR801EG, NR805EG*2,*3

NR893EG*2,*3

EN61010-1*1 EN61326-1:ClassA,Table2(Foruseinindustrial location)EN 61326-2-3EN 61000-3-2 +A1 +A2EN 61000-3-3KoreaElectromagneticConformityStandardRCM(C-tick)mark

NR801JG, NR805JG(excludingNR805JG/K)

NR893JG - -

NR805JG/K - - KoreaElectromagneticConformityStandard

*1: •Installationcategory(Overvoltagecategory)II Describesanumberwhichdefinesatransientovervoltagecondition.Itimpliestheregulationforimpulsewithstand

voltage. “II”appliestoelectricalequipmentwhichissuppliedfromfixedinstallationslikedistributionboards. •PollutionDegree2 Describesthedegreetoasolid,liquid,orgaswhichdeterioratesdielectricstrengthorsurfaceresistivityisadhering. “2”appliestonormalindooratmosphere.Normally,onlynon-conductivepollutionoccurs.*2: Usea5Abreakerinthepowersupplycircuit.*3: ThisinstrumentisaClassAproduct,anditisdesignedforuseintheindustrialenvironment.Pleaseusethisinstrumentin

theindustrialenvironmentonly.

2.Componentspecifications(includingoptions)

2.1 NR800FourierTransformNear-InfraredAnalyzers

2.1.1 HardwarespecificationsPrinciple: Fourier-transformremotemeasurement

viafiber-opticcableMeasurementmethod: Opticaltransmission

absorptionMeasuredsample: LiquidBeamsource: Halogenlamp(recommended

replacementintervalforcontinuousoperation: 5000 hours)

Detector: InGaAs(indiumgalliumarsenide)photodiode,effectivewavelengthrange:900 to 2500 nm

Numberofmeasuringchannels: 1to4(non-moving)Housingstructures: Fieldsuitablewithfull

hingedfrontcoverFiber-optic cable entry: Cable glandFiber-optic cable connectors:

For measurement; FCFor Ethernet; ST

Display: LEDKeyboard: Coveredwithwater-proofsheetingOperatinglocationrequirements: Seechapter4.Grounding type: Independent, 100ΩorlowerInsulationresistance: 10MΩorgreater(500VDC)Withstandingvoltage:

PowerSupply WithstandingVoltage100VAC,115VAC 1000VACfor1min.200VAC,230VAC 1500VACfor1min.

Coating: Epoxy resin coatingPaintcolor:

ModelNR801AG,NR801EG,NR801JG

NR805AG,NR805AG/CSA,

NR805EG,NR805JG

Flame proof enclosure - Lampblack

(equivalenttoMunsell0.8Y2.5/0.4)Airregulatorcover -

Case Frostywhite(equivalenttoMunsell2.5Y8.4/1.2)

Displaypanel Lampblack(equivalenttoMunsell0.8Y2.5/0.4)

Wallmountbracket,Free standing rack

Deepseamossgreen(equivalenttoMunsell0.6GY3.1/2.0)

Mar. 18, 2014-00

Page 5: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

5<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

Model NR801AG NR801EG NR801JGNR805AG(excluding

NR805AG/CSA)

NR805AG/CSA NR805EG

NR805JG(excludingNR805JG/K)

NR805JG/K

Explosion proof General purpose type(nonexplosion-proof) FM*1,*2 CSA*3 ATEX*2,*4 TIIS*5 KOSHA

*8CE marking - A*6 - - - A*6 - -Housingstructure IP53,NEMA3RorCSAType3RAir(purgegas)connection - 1/4NPT(F) Rc1/4or1/4NPT(F) Rc1/4or1/4NPT(F)

Cable connection Packinggland 3/4NPT(F) G3/4(F),3/4NPT(F),M25x1.5(F),orPG21(F)

Flame-proof packing glandorsealingfitting

Weight(kg)*7

Withoutbracket Approx. 50 Approx. 65

Withwallmount bracket Approx. 55 Approx. 70

Free stand rack Approx. 85 Approx. 100

*1: TypeXpurgingandexplosionproofforClassI,Division1,GroupsB,CandD.TemperatureClassT5. Allwiringshallcomplywiththemanufacture’sinstructions,NationalElectricalCodeANSI/NFPA70andLocalElectricalCodes.*2: Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*3: TypeXpurgingandexplosionproofforClassI,Division1,GroupsB,CandD.TemperatureClassT5. Allwiringshallcomplywiththemanufacture’sinstructions,CanadianElectricalCodePartIandLocalElectricalCodes.*4: KEMA02ATEX2175,II2G,EExpdIIB+H2T5 Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*5: ExpdIIB+H2T5X Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.*6: A: applicable*7: Notincludingweightofthepackagingmaterials.*8: KOSHA,ExdpIIB+H2T5 Allwiringshallcomplywiththemanufacture’sinstructionsandLocalElectricalCodes.

2.1.2 PerformanceWavelengthscanningrange: 900to2500nm

(11000to4000cm–1)Wavelengthresolution: 4,8,16,32,and64cm-1

(userselectable)Wavelengthreproducibility: 0.007cm-1Wavelengthaccuracy: 0.04cm-1S/Nratio 2250:1(RMS,resolution4cm -1, 11000 to

4000 cm -1, 1 sec.)

2.1.3 Inputs/Outputs(seealsosection2.2I/OUnit)

(1)CommunicationInterface:EngineeringPC: 1channel(Ethernet)DCS/I/Ounit: 2channels(RS-422) 1channelforDCS(Modbus)and

anotherforI/Ounit;or2channelsforI/Ounit

(2)Contactoutputs:Specification: NC/NOselectable,2channelsRating: 0.5A,30VDCorlessAction:

General purpose model:NR800analyzerpower”ON/OFF”;Powerfailsignal

StatusDO1TerminalNumber1to2 2to3

Power“off” Short OpenPower“on” Open Short

Mar. 18, 2014-00

NR800Systemalarm

StatusDO2TerminalNumber1to2 2to3

Power“off” Short OpenAlarm occurred Short OpenAlarm not occurred Open Short

Explosion proof model:NR800Systemalarm

StatusTerminalNumber

D1toD2 D2toD3Power“off” Open ShortAlarm occurred Open ShortAlarm not occurred Short Open

Annunciatorsignaloutput

StatusTerminalNumber

A1toA2 A2toA3Power“off” Open ShortDuringpurging Open ShortOverrideswitchpressed* Open ShortAfter purging Short Open

*: Overriding Whenopeningthedoorofthepressurized

explosion-protected enclosure under power on, the inner-pressure protection system will automatically turn off the power. The function that disables this automatic power off is called “overriding.”

Theoverrideswitchislocatedinsidetheanalyzer(flameproof).Youcanuseoverridingtoopenthedoor of the explosion-protected section.

Whenopeningthedoor,alwaysmakessuretheambient atmosphere is non-hazardous using, for example, a gas sensor.

Page 6: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

6

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.1.4 Operatingmodes*1

Basicoperatingmodeanddescription

Channeloperatingmode

No. Auto/Manual.*3

Maintenance *2

Spectrum analysis of a reference sample,equipmentmaintenance

- -

Run

On-linemeasurement(allowsspectrum analysis on selected channels)

1AUTO

MANUAL

2AUTO

MANUAL

3AUTO

MANUAL

4AUTO

MANUAL

Notes*1: Whenthepoweristurnedon,theanalyzerstarts

accordingtoapredefinedoperatingmode.*2: Canperformspectrumanalysis(notcontinuous

measurement).*3: Auto: Performscontinuousmeasurement Manual: Canperformspectrumanalysis(not

continuous measurement).

2.1.5 ChangingandSettingOperatingModeSome operations are prohibited depending on the user level.

UserlevelDescription Changing/

settingmodeAbbreviationUserA UA For operator Unauthorized

UserB UB Forequipmentsupervisor Authorized

UserC UC For maintenance Authorized

•AuserlevelcanbeswitchedonthefrontconsolepaneloftheanalyzerorfromSPECTLAND2screenofanengineeringPC.

•ApasswordisrequiredtoswitchlevelsfromUAtoUBandswitchfromUAorUBtoUC.

2.1.6 SampleMeasurement

(1)ContinuousmeasurementNumber of measuring channels: 1to4(Specificationisrequired.)Dataupdatingperiod: 4secondsorlonger(dependingonthe

numberofaveragingandmeasuringchannels, as well as measurement items)

Number of measurement items: Upto12perchannel(48max.per

analyzer)Number of outlier detection items: Upto12perchannel(48max.per

analyzer) Separatemaintenanceisavailableforeachchannel(exceptforcommonhardware).

(2)Streamswitchingbyinputsignal(intermittentmeasurement)

Calibration model set for the each channel can be changed by external input signal. This function is used for multi-stream application or multi sample application.

Stream switchingNumberofstreamstoswitch: Upto16Switching patterns: See table 1. “Stream numbers

assignable to measuring channels corresponding to switching patterns”

Contact inputs: Seesection“2.2I/OUnit”RS-422 communication: Seesection“2.1.8MODBUS”

Dataupdatingperiod: 8secondsorlonger(dependingonthe

switchingpattern,numbersofaveragingand measuring channels, and stream configuration)

Number of measurement items: Upto12perstream(64max.per

analyzer)Number of outlier detection items: Upto12perstream(64max.per

analyzer)Separatemaintenanceisavailableforeachchannel

(exceptforcommonhardware).InputsignalsandmeasurementcomputationsequenceStreamidentification: Usedtoidentifytheselectedsample.

The analyzer will choose a calibration modelsettosuitefortherelevantstream based on this signal.

Measurementenabled(validsample): Whenclosed,theanalyzerperforms

measurement using the calibration modelspecifiedbythesampleidentificationsignalabove.

Schematictimingchartofmeasurementandcomputationsequence(e.g.2streams)

Stream 1 Stream 2 Stream 1

Computationperformed

Computationperformed

Computationperformed

OpenClosed

Closed

Open

Stream identification*

Measurement enabled

Notes(*) Astreamnumberisdefinedbyacombinationof

open/closedstatesofanidentificationcontactsignal.

Mar. 18, 2014-00

Page 7: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

7<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

Table1 Streamnumbersassignabletomeasuringchannelscorrespondingtoswitchingpatterns

Streamswitch MeasuringchannelNo.*3Totalstreams Applicable

channelNo.Case PatternNo. 1 2 3 4None 0 1 2 3 4 4 max 1 to 4

Stream switching per channel*2

1*1 1to(17-N) 18-N 19-N 20-N 16 max 1 to 42 1 to 8 9 to 16 None None 16 max 1 to 23 1 to 4 5 to 8 9 to 12 13 to 16 16 max 1 to 4

Notes*1: N:Maximumnumberofmeasuringchannelsincludedwithintheanalyzer.*2: Thestreamnumberforameasuringchannelthatdoesnotswitchpathsmustbethesmallestnumberintherelevant

column.*3: Measuringchannelnumbersthatequalthenumberofmeasuringchannelsincludedwithintheanalyzerorsmallerarevalid.

2.1.7 ConsolePanel

(1)Displayitems

OperatingmodeLEDs(MODE) ThefollowingLEDsindicatethecurrent

basic operating mode:RUN: LitwhenintheRunmode.MAINT: LitwhenintheMaintenancemode. The operating modes of each channel

aredisplayedbyLEDswhenintheRunmode.

TwoLEDsareprovidedforeachchannel,amountingtoatotalof8LEDs.

AUTO: LitwhenintheAutomode.MAN: LitwhenintheManualmode.

Self-diagnosisLEDs(DIAGNOSIS)ALARM: Litwhenanalarmoccurs.LAMP: Litwhenalamphasburnedoutorafter

atimeperioddefinedbytheservicelifesetting elapses.

LASER: Litwhenthelaserhasburnedoutorafteratimeperioddefinedbytheservicelifesettingelapses.

LEDdisplay(16digits,STREAM/CONSTITUENT/VALUE)

The content depends on the operating status or operation.

Operationkeys Thefollowingsixkeysareprovided:ITEM: Usedtochangeitems.ENT: Usedtoconfirmtheentry.Arrowkeys: Usedtomovethecursororchange

display.

(2)Analyzerbehaviorforeachbasicoperatingmode

Maintenance modeBasicoperatingmodeLEDs: OnlyMAINTlightsup.ChanneloperatingmodeLEDs: Litinaccordancewitheachsetting.

Self-diagnosisLEDs: Litinaccordancewiththeresultsof

self-diagnosis.LEDdisplay: The display depends on the operating

status.Run modeBasicoperatingmodeLEDs: OnlyRUNlightsup.ChanneloperatingmodeLEDs: Litinaccordancewitheachsetting.Self-diagnosisLEDs: Litinaccordancewiththeresultsof

self-diagnosis.LEDdisplay: The display depends on the operating

status.

2.1.8 MODBUS

(1)Communicationspecifications

Item SpecificationCommunication standard*

RS-422orRS-232C(aconverterisrequired)

Start bit 1Number of data bits 7(ASCIImode)or8(binarymode)Paritybit 1Stop bit 1Communication speed 4800,9600or19200(selectable)

Error detection Oddnumberparity,EvennumberParityornone(selectable)

*: OnlyRS-232CcanbeusedinNR805*Gexplosion-proof type.

See section 1.2 and 2.5

(2)Serialcommunication

Item ASCIImode RTUmodeNumber of data bits 7(ASCII) 8(binary)

Message starting character Colon “:” None

Message ending character

Carriagereturn/Linefeed“<CR><LF>” None

Error detection LRC(LogicalRedundancy Check)

CRC-16(CyclicRedundancy Check)

Inter-character time out None 100msec

Mar. 18, 2014-00

Page 8: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

8

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

(3)SupportfunctionintheMODBUSprotocol

FunctionNo. Description

01 ReadstheON/OFFstatusofcoils.02 ReadstheON/OFFstatusofinputrelays.03 Readsthecurrentvalueofholdingregisters.04 Readsthecurrentvalueofinputregisters.05 Forcibly changes the status of a coil.06 Writeavaluetoaholdingregister.

08Loop-backtestSends back the same message as the command message.

(4)Recommendedvalueofmasterside

Item RecommendedvalueTransmissionspeed(bps) 4800/9600/19200Time monitor of Inter-character gap 1000msNo reply time 2 sec.Number of transmission retries 5 timesCommunicationRecoveryPeriod 30 sec.

Mar. 18, 2014-00

2.1.10ModelandSuffixCodes

(1)NR801AG,Generalpurposemodel

Model SuffixCode OptionCode DescriptionNR801AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodelLanguage -E ••••••••••••••••• EnglishPowersupply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1

2

•••••••••••••••••

•••••••••••••••••

Electric cable, only for general purpose model and less than 40 mFiber-optic cable

Mounting ABC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction

2.1.9 OtherFunctions

(1)Baselinecompensation:Upto10points

(2)Outputoftheresultcompensatedusinganalogueinputs

The following operation result can be output.Compensation operation processing:

P2=C1*P1+C0+D D=An2 *D2+An*D1+D0

An: Analogueinput(n:4orless)P1: Measurementvaluebeforethe

compensationP2: Measurementvalueafterthe

compensationD0,D1,D2,C0,C1: Coefficient

(3)OutputsmoothingSimplemovingaverage(Upto10points)Exponentialmovingaverage

(4)On-linemeasurementspectrasaving

Page 9: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

9<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(2)NR805AG,TypeXpurgedandFMapproved,CSAapprovedexplosion-proofmodel

Model SuffixCode OptionCode DescriptionNR805AG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion proof model, FM

approvedversion(*1)Language -E ••••••••••••••••• EnglishPowersupply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N"Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A

BC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N"- -N ••••••••••••••••• Always “-N"Cable entrance -2 ••••••••••••••••• Female3/4NPTPurgeairconnection 2 ••••••••••••••••• Female1/4NPTOption /SS

/CSAWithstreamswitchinputfunctionCSAapprovedversion(*1)

(*1) Optioncode"/CSA"isspecifiedCSAapproved Optioncode"/CSA"isnotspecifiedFMapproved

Mar. 18, 2014-00

Page 10: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

10

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

(3)NR801EG,Generalpurposemodel,withCEmarking

Model SuffixCode OptionCode DescriptionNR801EG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodel,

CE markingLanguage -E ••••••••••••••••• EnglishPowersupply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1

2

•••••••••••••••••

•••••••••••••••••

Electric cable, only for general purpose model and less than 40 mFiber-optic cable

Mounting ABC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction

Mar. 18, 2014-00

Page 11: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

11<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(4)NR805EG,ATEXapprovedexplosion-proofmodel,withCEmarking

Model SuffixCode OptionCode DescriptionNR805EG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model,

ATEXapproved,CEmarkingLanguage -E ••••••••••••••••• EnglishPowersupply 1

346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A

BC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -3

-4-5-6

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

FemaleG3/4Female3/4NPTFemale M25x1.5FemalePG21

Purgeairconnection 12

••••••••••••••••••••••••••••••••••

FemaleRc1/4Female1/4NPT

Option /SS Withstreamswitchinputfunction

Mar. 18, 2014-00

Page 12: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

12

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

(5)NR801JG,Generalpurposemodel

Model SuffixCode OptionCode DescriptionNR801JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800FT-NIRAnalyzer,GeneralPurposemodelLanguage -E

-J••••••••••••••••••••••••••••••••••

EnglishJapanese

Powersupply 1346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 1

2

•••••••••••••••••

•••••••••••••••••

Electric cable, only for general purpose model and less than 40 mFiber-optic cable

Mounting ABC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”- -0 ••••••••••••••••• Always “-0”- 0 ••••••••••••••••• Always “0”Option /SS Withstreamswitchinputfunction

Mar. 18, 2014-00

Page 13: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

13<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(6)NR805JG,TIISapprovedexplosion-proofmodel*1

Model SuffixCode OptionCode DescriptionNR805JG •••••••••••••••••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• NR800 FT-NIR Analyzer, Explosion-proof model,

TIISapproved*1Language -E

-J••••••••••••••••••••••••••••••••••

EnglishJapanese*1

Powersupply 1346

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

100VAC±10%,50/60Hz115VAC±10%,50/60Hz200VAC±10%,50/60Hz230VAC±10%,50/60Hz

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Number of measuring channels

-S1-M1-M2-M3-M4

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

1 channelExpandable to 4 channels, comes with 1 channelExpandable to 4 channels, comes with 2 channelsExpandable to 4 channels, comes with 3 channels4 channels

Wavelengthscanningrange W1W2

••••••••••••••••••••••••••••••••••

900 to 2100 nm900 to 2500 nm

Fiber-optic cable -1-2

••••••••••••••••••••••••••••••••••

Single cableDualcable

- -00 ••••••••••••••••• Always “-00”Ethernet output cable 2 ••••••••••••••••• Fiber-optic cableMounting A

BC

•••••••••••••••••••••••••••••••••••••••••••••••••••

WithoutbracketsWithwall-mountingbracketsWithfreestandingrack

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Cable entrance -1

-2-3-4

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Metalconduit(G3/4)Metalconduit(3/4NPT)Flameproofpacking(G3/4)Flameproofpacking(3/4NPT)

Purgeairconnection 12

••••••••••••••••••••••••••••••••••

FemaleRc1/4Female1/4NPT

Option /SS/K

WithstreamswitchinputfunctionKOSHAapprovedexplosionproofmodel*1

*1: “TIIS”approvalisnotapplicablewhenoptioncode“/K”isselected.“Japanese”cannotbeselectedwithoptioncode“/K.”

Mar. 18, 2014-00

Page 14: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

14

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.1.11Externaldimensions

(1)Generalpurposemodel(NR801*Gtype)•Withoutbrackets(Suffixcode:A)

6282 60 60 60 60 62 62

68±5

30±5

480

92

60430

277

665

355550

Mounting screws4-M8

2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)

Earth terminal(M4 screw)

Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Unit: mm

For NR801AG, NR801EG

Part A

Part A

Note: The wall construction should be withstood a force of four times the weight of the NR800.

Mar. 18, 2014-00

Page 15: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

15<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

•Wall-mountingmodel(Suffixcode:B)

30±5

665

550

60 430

50 50

603086

0

355 20

920

6282 60 60 60 62 6260

277

68±5

Mounting holes4-ø10.5 holes

Earth terminal(M4 screw)

Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)

2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Wall mounting bracket

Unit: mm

For NR801AG, NR801EG

Part A

Part A

Note: The wall construction should be withstood a force of four times the weight of the NR800.

Mar. 18, 2014-00

Page 16: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

16

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

•Free standing rack model(Suffixcode:C)66

5

1500

70

350 8080

30±5

68±5

550 355 50

630

550

490

56827

7

6282 60 60 60 62 6260

Mounting holes 4-ø17 (for anchor bolts)

4-special fiber-optic cable outlets(CH1, CH2, CH3, CH4 from the left side)

Earth terminal(M4 screw)

Unit: mm

Power supply cable connection inlet (For NR801AG, NR801EG)(Packing gland, acceptable cable: ø10 to 15.9)

2-cable outlets (For NR801AG, NR801EG)(Packing glands, acceptable cables: ø10 to 13)

3-cable outlets (For NR801JG)(Packing glands, acceptable cables: ø10 to 18.5)

Special Ethernet fiber-optic cable outlet orEthernet communications cable outlet (10Base-T, acceptable cable ø5 to 7.4)

For NR801AG, NR801EG

Part A

Part A

Mar. 18, 2014-00

Page 17: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

17<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(2)Explosion-proofmodel(NR805*Gtype)•Withoutbrackets(Suffixcode:A)

62

82180

93

36

6262332

665

355550

App

rox.

300

200

430 60

9248

0

4-M8Mounting screws

95

200

32±5

Connection: M25x1.5

Connector *2

For NR805EG

Internal-pressure protection unit(flameproof enclosure)Cable connectors(with sealing fitting orflameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT(only 1/4NPT for NR805AG)

Connector *4

Connection:3/4NPT

For NR805AG

Earth terminal(M4 screw)

86

277

216

65

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

Special Ethernet fiber-optic cable outlet

Unit: mm

For NR805AG, NR805EG

Part B

Part A

Part A Part A

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.Note: The wall construction should be withstood a force of four times the weight of the NR800.

Mar. 18, 2014-00

Page 18: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

18

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

•Withwall-mountingbrackets(Suffixcode:B)

62

App

rox.

300

36

82180

6093

6262332

665

550

60 430

50 50

3086

0

200

200

32±5

For NR805AG

Connection: 3/4NPT

Connector *2

Connection: M25x1.5

Mounting holes 4-ø10.5 holes

For NR805EG

Internal-pressure protection unit(flameproof enclosure)

Cable connectors(with sealing fitting or flameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

95

Earth terminal(M4 screw)

86

277

21665

Special Ethernet fiber-optic cable outlet

2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

920

355 20

Wall mounting bracket *5

Connector *4

Unit: mm

For NR805AG, NR805EG

Part A

Part A

Part A

Part B

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.*5: ThewallconstructionshouldbewithstoodaforceoffourtimestheweightoftheNR800.

Mar. 18, 2014-00

Page 19: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

19<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

•Withfreestandingrack(Suffixcode:C)

568

490

550

630

70

1500

665

App

rox.

300

80

200

550

Mounting holes 4-ø17 (for anchor bolts)

3693

Connection: 3/4NPT

For NR805AG

For NR805AG, NR805EG

32±5

200

Connector *2

Connection: M25x1.5

Internal-pressure protection unit(flameproof enclosure)

Cable connectors(with sealing fitting or flameproof packing gland *1)

Air (purge gas) connection: Rc1/4 or 1/4NPT (only 1/4NPT for NR805AG)

For NR805EG

82180

350 80

95

Connector *4

50355

Bottom of analyzer

277

216

8665

Earth terminal (M4 screw)

62332 62 62

4-special fiber-optic cable outlets (CH1, CH2, CH3, CH4 from the left side)

2-cable outlets *3 *4(Packing glands, acceptable cables: ø10 to 15.9 or sealing fittings)

Special Ethernet fiber-optic cable outlet

Unit: mm

Part B

Part A

Part A

Part A

Part B

*1: ThecableODforflameproofpackingglandisø8toø15.9mm.*2: Selecttheconnectorfromthefollowing. 3/4NPT,PG21orG3/4*3: InthecaseofNR805AGorNR805EG,Sinceattachedferritecoreareused,thepracticalcableODisuptoø13mm.*4: InthecaseofNR805EG,selecttheconnectorfromthefollowing. 3/4NPT,PG21,G3/4orM25x1.5 TheconnectorofNR805AGis3/4NPT.

Mar. 18, 2014-00

Page 20: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

20

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.2 I/OUnit

2.2.1 OverviewTheI/Ounitisaninput/outputinterfacebetweentheanalyzerandexternalmonitor/controlsystemsuchasDCS.TheI/OunitconnectsNR800analyzertoserialcommunications(RS-422).UptotwoI/Ounitscanbe connected to the Analyzer, and one of which can incorporate the contact input for stream switching and analogue input for compensation options.

2.2.2 SpecificationsPowersupply: Seesection“3.UtilitySpecifications”.Insulationresistance: 5MΩormore,500VDCWithstandvoltage: 1500VACfor1min.Analogue output: Outputdata: Measurementresults(Propertiesand

concentration)Number of outputs: 0 to 40Outputspecifications:

Item Description

Outputrange 4to20mADC(3.0to21.0mADC,Floating-common type)

Isolation method

Between output terminals and internal circuit:PhotocouplerisolationBetween output terminals: non-isolated, Commonnegative

Withstandvoltage 500VDCfor1min.Allowable load resistance 600Ωorless

12-bitD/Aconverterresolution 5.7 µA

System accuracy ±0.5%offullscale

Separately,24VDCpowersupplyisrequired(seesection“2.2.3.ModelandSuffixcodes”and“3.UtilitySpecifications”).

Contact output:

Alarm Quantityofoutputs Alarm Quantity

ofoutputsGeneral 1 Outlier 4Communication failure 1 I/Ounitfailure 1

Operatingmode 5 - -

Contactoutputspecifications:Item Description

I/Ounitfailure

Rating 24VDC.0.3A

Action and number1 normally open and 1 normallyclosed(sharedcommon)

Alarm, operating mode, and outlier detection

Insulation method Mechanical isolationRated load voltage

DC 24VAC 100to240V

Maximum load current 2Amps/point,8Amps/common

Servicinglife

Mechanical At least 20 million actions

Electrical At least 100 thousand actions

Surge killer NoneNumber per common 8points/commonExternal power supply Notrequired.

Contact input:Description:Streamidentificationformulti-streamsampling

unit : 8 points, status signals to identify

sample streams that pass through measurement cells.

Measurementenabled(streamvalid): 4points,statussignalstoconfirmthat

samples inside the measurement cells are ready for measurement.

Inputspecifications:Item Description

Input type VoltagefreecontactCommon terminal Common to 8 pointsIsolation method Transformer isolation

WithstandvoltageBetween external connectors collectivelyandinternalcircuit:500VDCfor1min.

Off-stateopen-circuitvoltage 5to7V

On-stateloadcurrent 1 to 3 mA

On-stateloadresistance 200Ωorsmaller

Off-stateloadresistance 100Ωorlarger

Provideexternalcontact-inputsignalsmeetingtheaboverequirements.

Analogue input:Input data and number: Analogue output

compensation signal, 4 pointsInputspecifications

Item DescriptionInputrange(actual) 1to5VDC(-0.25to5.25VDC)

Isolation method

Between input terminals and internal circuit: photo coupler isolationBetween input terminals: non-isolated, negativecommon

Withstandvoltage 500VDCfor1min.Input resistance 1MΩ12-bitA/Dconverterresolution 1to5VDC:1.4mV

System accuracy ±0.5%offullscaleat0to40°C

Installationrequirements:Seechapter4.Mounting: WallmountingExternal connection terminal: M3.5 screwDrip-proofanddust-proofconstruction: IP4X(withenclosure)

Mar. 18, 2014-00

Page 21: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

21<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.2.3 ModelandSuffixcodes

(1)NR893AG:forNR801AGandNR805AGanalyzer

Model SuffixCode OptionCode DescriptionNR893AG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit(nonexplosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1

2••••••••••••••••••••••••••••••••••

NR805AG(Explosion-proofmodel) (*1)NR801AG(GeneralPurposemodel)

Contact input for stream switching

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Analogue input for compensation

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

None481216202428323640

Enclosure -A-B-C

•••••••••••••••••••••••••••••••••••••••••••••••••••

Number of analog outputs: 0 to 12Number of analog outputs: 0 to 28Number of analog outputs: 0 to 40

- -N ••••••••••••••••• Always “-N”

*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.

(2)NR893EG:forNR801EGandNR805EGanalyzer

Model SuffixCode OptionCode DescriptionNR893EG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit,CEmarking(nonexplosion-proof)Language -E ••••••••••••••••• EnglishAnalyzer 1

2••••••••••••••••••••••••••••••••••

NR805EG(Explosion-proofmodel) (*1)NR801EG(GeneralPurposemodel)

Contact input for stream switching

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Analogue input for compensation

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Number of analogue outputs -00-04-08-12-16-20-24-28-32-36-40

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

None481216202428323640

Enclosure -C ••••••••••••••••• Always “-C”- -N ••••••••••••••••• Always “-N”

*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.

Mar. 18, 2014-00

Page 22: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

22

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

(3)NR893JG:forNR801JGandNR805JGanalyzer

Model SuffixCode OptionCode DescriptionNR893JG ••••••••••••••••••••••••••••••••••••••••• ••••••••••••••••• IOunit(nonexplosion-proof)Language -E

-J••••••••••••••••••••••••••••••••••

EnglishJapanese

Analyzer 12

••••••••••••••••••••••••••••••••••

NR805JG(Explosion-proofmodel) (*1)NR801JG(GeneralPurposemodel)

Contact input for stream switching

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Analogue input for compensation

AB

••••••••••••••••••••••••••••••••••

AvailableNone

Number of analogue outputs (*2)

-00-04-08-12-16-20-24-28-32-36-40

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

None481216202428323640

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”Option /C Withenclosureincluded24VDCpowersupplyunit

foranalogoutput.(*2)

*1: Alwaysspecify1forusewithanexplosion-proofmodel.Thisblocksthecommunicationsignaluponreceivingapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity of the analyzer.

*2: Separate24VDCpowersupplyrequiredwhenusingananalogueoutputandoptioncode“/C”arenotspecified.(seechapter 3).

Mar. 18, 2014-00

Page 23: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

23<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.2.4 Externaldimensionsandweight:

(1)NR893AG,NR893EGPaintcolor:Lightbeige(Munsell5Y7/1orequivalent)

B D F

CA

E G

H

75 17

16118°

1743

5028

0

400

68.5

200

6100 505050505050

L + 50

L50 50

4-ø14 holesfor mounting *1

ø20 holefor protective ground wiring

Number ‘N’ of wiring holesSuitable cable OD *2 A: ø10 to 15.9 (to the power supply) B to H: ø10 to 18.5

Unit: mm

Enclosurecode

Numberofanalogoutputs L(mm) N Approximate

WeightApplicablemodelname

-A 0 to 12 400 6(AtoF) 15.5 kg NR893AG-B 0 to 28 500 7(AtoG) 17.5 kg NR893AG

-C*3 0 to 40 600 8(AtoH) 20.0 kg NR893AGNR893EG

*1: The wall construction should be withstood a force of four times the weight.*2: Sinceattachedferritecoresareused,thepracticalcableODareuptoø13mm.*3: InthecaseofNR893EG,theenclosurecodeisonly-Cirrespectiveofthenumberofanalogoutputs.

Mar. 18, 2014-00

Page 24: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

24

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

(2)I/OunitWithoutenclosure(onlymodelNR893JG)•ForExplosionproofanalyzerunit•ThesizeofIOunitforgeneralpurposeanalyzerunitisthesameasabovefigurebutRS-422/232Cconverterpowersupplycaseremoved.Paintcolor:Baseplate,RS-422/232Cconverterpowersupplycase;Modulecolors: Light cobalt blue (Munsell 6.2PB4.6/ 8.8 or equivalent), lamp black (Munsell 0.8Y2.5/0.4 or

equivalent)

215

199

L

303 151510

(101)Approx. 120

N-ø7.2 holesfor mounting *1

Safety barrier power supply caseused only with NR805JG explosion-proof analyzer

Unit: mm

Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 333 4 6.5 kg-16, -20, -24, -28 16, 20, 24, 28 450 6 7.0 kg

-32, -36, -40 32, 36, 40 537 6 7.5 kg

*1: Thewallconstructionshouldbewithstoodaforceoffourtimestheweight.

(3)Withenclosure(Optioncode:/C)Paintcolor:Lightbeige(Munsell5Y7/1orequivalent)

C E

D

A

F H

G

5050

B

5050 50505075 17

10118°

1743

5028

0

400

App

rox.

30

200

6

L + 50

L50 50

4-ø14 holesfor mounting *1

ø20 holefor protective ground wiring

Number ‘N’ of wiring holesSuitable cable OD: ø10 to 18.5

Unit: mm

I/O Unit (inside)

Numberofanalogoutputscode Numberofanalogoutputpoints L(mm) N ApproximateWeight-00, -04, -08, -12 None, 4, 8, 12 400 6(AtoF) 15.5 kg-16, -20, -24, -28 16, 20, 24, 28 500 7(AtoG) 17.5 kg

-32, -36, -40 32, 36, 40 600 8(AtoH) 20.0 kg

*1: Thewallconstructionshouldbewithstoodaforceoffourtimestheweight.

Mar. 18, 2014-00

Page 25: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

25<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.3MeasurementCells

2.3.1 FlowthroughCell

(1)SpecificationsOpticalpathlength: 1,2,5,10,or20mmFiber-optic connector: FC typeWettedpartmaterial: Fluorinerubber,Kalrez4079,316SS,PTFE,Borosilicatecrownglass,orSapphireSamplepressurerange: Atmosphericpressureupto1.9MPaGSampletemperaturerange: –5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther

information.)Sampleconnection: FemaleRc1/4Installationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.3kg

(2)ModelandSuffixCodes

Model SuffixCode OptionCode DescriptionNR510 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cellWindowmaterial and optical path length

-B00

-B01-B02-B05-B10-B20-S00

-S01-S02-S05-S10-S20

•••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Borosilicatecrownglass,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm

Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ringmaterial -B

-K••••••••••••••••••••••••••••••••••

Fluorine rubberKalrez4079

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”

(3)ExternalDimensions

Unit: mm(96)55 29

102

92

L4-ø5.2 holes

2-Rc1/4

Optical path length (mm)

1251020

L (mm)

64±2

65±2

68±2

73±2

83±2

Fiber-optic connection(in the other side)

Mar. 18, 2014-00

Page 26: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

26

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.3.2 FlowthroughCellwithConstantTemperatureWaterTube

(1)SpecificationsOpticalpathlength: 1, 2, 5, 10, or 20 mmFiber-optic connector: FC typeWettedpartmaterial: Fluorinerubber,Kalrez4079,316SS,PTFE,Borosilicatecrownglass,orSapphireSamplepressurerange: Atmosphericpressureupto1.9MPaSampletemperaturerange: +5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther

information.)Constantwatertemperaturerange:+5°Cto+80°C(ItvarybythematerialsofO-ring.ContactYokogawaforfurther

information.)Sampleconnection: FemaleRc1/4Connection for water with constant temperature: 6-mm outside diameter, copper tubeInstallationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.3kg

(2)ModelandSuffixCodes

Model SuffixCode OptionCode DescriptionNR512 •••••••••••••••••••••••••••• ••••••••••••••••• Flow through cell with constant temperature water

tubeWindowmaterial and optical path length

-B00

-B01-B02-B05-B10-B20-S00

-S01-S02-S05-S10-S20

•••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Borosilicatecrownglass,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Borosilicate crown glass, 1 mmBorosilicate crown glass, 2 mmBorosilicate crown glass, 5 mmBorosilicate crown glass, 10 mmBorosilicate crown glass, 20 mmSapphire,10mm,withvariableopticalpathadapter(1,2,5,or20mm)Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mmSapphire, 20 mm

Body material SUS ••••••••••••••••• Stainless steel, 316SSO-ringmaterial -B

-K••••••••••••••••••••••••••••••••••

Fluorine rubberKalrez4079

- -N ••••••••••••••••• Always “-N”- -N ••••••••••••••••• Always “-N”

(3)ExternalDimensions

Unit: mm

4-ø5.2 holes

2-Rc1/4

L29

Approx. 160

App

rox.

160

102

92

B-SIDEB-SIDE A-SIDE A-SIDE

Thermal insulation jacket

Approx. 145

70111

Optical path length (mm)

1251020

L (mm)

64±2

65±2

68±2

73±2

83±2Fiber-optic connection (in the other side)

Mar. 18, 2014-00

Page 27: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

27<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.3.3 SpecialmeasurementcellAPFAflowthroughtypespecialcellisdescribedhere.

(1)SpecificationsOpticalpathlength: 1,2,5,or10mmFiber-optic connector: FC typeWettedpartmaterial: PFA,Kalrez4079orMoriseiperflioro(MP4275B),orSapphire (Stainlesssteelispartlyusedasnon-wettedparts.)Samplepressurerange: Atmosphericpressureupto0.3MPaGSampletemperaturerange: 5°Cto80°CSampleconnection: 1/4inchs(Ø6.35xØ3.95mm)PFAtube,orØ6xØ4mmPFAtube (length:3000mm)Installationangle: VerticalInstallationlocationrequirements:Seesection4.Weight: Approx.1.5kgOthers: O-ringsandPFAbodymustbeexchangedsimultaneously.

(2)ModelandSuffixCodes

Model SuffixCode OptionCode DescriptionNR511 ••••••••••••••••••••••• ••••••••••••••••• Special measurement cellWindowmaterial and optical path length

-S01-S02-S05-S10

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Sapphire, 1 mmSapphire, 2 mmSapphire, 5 mmSapphire, 10 mm

Body material P ••••••••••••••••• PFAO-ringmaterial -K

-P••••••••••••••••••••••••••••••••••

Kalrez4079Moriseiperflioro(MP4275B)

Tube -A-B

••••••••••••••••••••••••••••••••••

1/4inchs(Ø6.35xØ3.95mm),3000mm,PFAØ6xØ4mm,3000mm,PFA

Option /Z Alwaysadd“/Z”

(3)ExternalDimensions

Unit: mm(80)(50)

90

(300

0)(3

000)

74

77(94)

4-ø5.2 holes

Fiber-optic connection(in the other side)

Mar. 18, 2014-00

Page 28: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

28

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.3.4 In-situprobe

(1)SpecificationsFiber-optic connector: FC typeWetpartmaterial:

Body; HastelloyC-276orUNSS31603(Stainlesssteel316L)Others; Sapphire,HastelloyC-276,Gold

Samplepressurerange: Atmosphericpressureto0.98MPa*Sampletemperaturerange: Normaltemperatureto80°C*Purgegasconnection: Female1/8NPTx2Fiber-opticentrance: Female1NPTInstallationangle: Vertical(Insertportionisdownward)orHorizontalInstallationlocationrequirements:Seesection4Weight: Approx.5kg(excludingflange)*: Whenthesamplepressureistheoutsideoftherange,pleaseconsultthesalesperson.

(2)ModelandSuffixCodesFIR200/Z

(3)ExternalDimensions

L *3

( 560 ) *4 ( 80 )

Approx. 35 A *1

Flange *2

Fiber-optic connection box

Fiber-optic entrance(Exclusive packing)

Purge-gas connections2-1/8NPT

Approx. ø27

Unit: mm

*1:Opticalpathlength“A”:2/5/10mm,selectable(Dependentonsampleconditions,measurementitems,measurementrange, etc. Consult Yokogawa.)

*2:Mountingflange(HastelloyC-276/UNSS31603comparable)shouldbeprovidedbycustomer.*3:Specifylength“L”withintherange150to250mm.*4:Theshapeandsizeofconnectionboxmaychange.

Note1:Keepthetemperatureattheconnectionofthefiber-opticcableat80°Corbelow.Ifthetemperatureattheconnectionexceeds80°C,theconnectionboxshouldbepurgedwithclean,dryair,orotherappropriatemeasuresshouldbetaken.Inthiscase,theconnectionboxmaybedifferentinshapeandsizefromtheoneshownabove.Ifafiber-opticcableformeasurement is exposed to radiant heat, it should be shielded.

Note2:Dimensionsofeachpartmaychangewithoutpriornotice.

2.4 Fiber-opticCablesformeasurementFiber-optic cables connect an analyzer to measurement cell. Select appropriate cables to meet your application requirement,e.g.,wavelengthsscanningrange.

2.4.1 SilicaFiber-opticCable

(1)SpecificationsApplicablewavelengthrange: 900to2100nmConnector: Double-endFCorFC-SMAtypeStructure: Dual(forsampleandreference)orsingle(sample),two-core,protectedby

stainlessflexibletubeMinimum bending radius: 100 mm. To reduce optical attenuation, make the radiuses along the cable as

large as possible when laying cables.Installationlocationrequirements:Seesection4.Cabling: Conduit protected cabling is recommended.

Mar. 18, 2014-00

Page 29: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

29<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(2)ModelandSuffixCodes

a.Singlefiber-opticcableModel SuffixCode OptionCode Description

NR821 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2100nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side

Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

5 m10 m20 m30 m50 m100 m150 m200 m300 m

- -000 ••••••••••••••••• Always “-000”

b.Dualfiber-opticcableModel SuffixCode OptionCode Description

NR822 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2100nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side

Cable length -L005-L010-L020-L030-L050-L100-L150-L200-L300

•••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

5 m10 m20 m30 m50 m100 m150 m200 m300 m

- -000 ••••••••••••••••• Always “-000”

(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.

• Double-endFCconnector

BA

A B

Unit: mmFC connector FC connector

(Cable length)600±50

(57) 3010

1510

• FC(analyzer)-SMA(measurementcell)connector

BA

A B

Unit: mmFC connector (analyzer) SMA connector

(measurement cell)

(Cable length)600±50

(57) 3010

600±50

(53) 8

1510

Mar. 18, 2014-00

Page 30: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

30

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.4.2 FluorideFiber-opticCable(1)Specifications

Applicablewavelengthrange: 900to2500nmConnector: Double-endFCtypeorFC-SMAtypeStructure: Dual(forsampleandreference)orsingle(sample),two-core,protectedby

stainlessflexibletubeMinimum bending radius: Analyzer side; 100 mm, Measurement side;120 mm To reduce optical attenuation, make the radiuses along the cable as large as

possible when laying cables.Installationlocationrequirements:Seesection4.Note: Fluoride,amaterialusedintheNR823andNR824fiber-opticcables,hasthedeliquescenceproperties.Whenstoring

thesecables,takeappropriatemeasurestopreventmoisture,e.g.,acableshouldbeplacedwithdesiccantinasealed,moisture-barrier plastic bag and the desiccant replaced regularly.

(2)ModelandSuffixCodes(ModelcodetoincludefibersforAnalyzer–Cell(probe)–Analyzer)a.Singlefiber-opticcable

Model SuffixCode OptionCode DescriptionNR823 •••••••••••••••••••• ••••••••••••••••• Singlefiber-opticcableforwavelengthof900to2500nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side

Cable length -L003-L004-L005

•••••••••••••••••••••••••••••••••••••••••••••••••••

3 m4 m5 m

- -000 ••••••••••••••••• Always “-000”

b.Dualfiber-opticcableModel SuffixCode OptionCode Description

NR824 •••••••••••••••••••• ••••••••••••••••• Dualfiber-opticcableforwavelengthof900to2500nmConnector -FF

-FS••••••••••••••••••••••••••••••••••

Double-endFCtypeFC type on analyzer side and SMA type on measurement cell side

Cable length -L003-L004-L005

•••••••••••••••••••••••••••••••••••••••••••••••••••

3 m4 m5 m

- -000 ••••••••••••••••• Always “-000”

(3)ExternalDimensionsAdualfiber-opticcableisasetoftwosinglefiber-opticcableofthesamespecification.(Connectthelongerendtothe measuremrnt cell side.)• Double-endFCconnector

BA

A B

Unit: mmFC connector (analyzer) FC connector

(measurement cell)

(Cable length)600±50 800±50

(57) 3010

1510

• FC(analyzer)-SMA(measurementcell)connector

BA

A B

Unit: mmFC connector (analyzer) SMA connector

(measurement cell)

(Cable length)600±50

(57) 3010

800±50

(53) 8

1510

Mar. 18, 2014-00

Page 31: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

31<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.5 RS-422-to-RS-232CConverter(PartNumber:K9404LAorK9404LD)ThisunitconvertsoutputsignalsoftheanalyzermainunitfromtheRS-422formattotheRS-232CformatforpersonalcomputerorDCScommunicationandalsoblockscommunicationsignalswhenitreceivesapower-offsignal from the Analyzer, thus ensuring the explosion-proof integrity.

2.5.1 SpecificationsPowersupply: Seesection3.Signal terminals: Analyzermainunitside(RS-422):M4,outputside(RS-232C):FemaleDsub25-pinGroundingtype: 100ΩorlowerInstallationlocationrequirements:Seesection4.Housingstructure: Desktop(K9404LA),Screwcramp(K9404LD)Weight: Approx.2kgNote: TheconverterdoesnotcarrytheCEMarking.Anappropriatehousingshouldbeprovidedbycustomer,ifnecessary.

2.5.2 TerminalDiagram

CN3 CN2

LN-+RDBRDASDBSDA

RS-422

From analyzer

D/I 12V AC power supply

Female D-SUB 25-pin connectorCN1

RS-232C

2.5.3 RS-232CCommunicationStraightCable

GND

123456789

CDRDSDERSGDRRSCSCI

FGCDRDSDERSGDRRSCSCI

GND183220764522

Internal connection

To converter To host

2.5.4 ExternalDimensions

4-ø4.5 holes203

126

126

39.5

15 15

9611

.5 218

203

37

K9404LA Desktop type K9404LD Screw cramp type

Unit: mm

May 15, 2014-01

Page 32: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

32

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.6 EthernetFiber-opticCableCableforEthernetcommunicationbetweentheanalyzerandtheengineeringPC.

2.6.1 SpecificationsLength: Upto20mConnector: ST typeStructure: Two-core,protectedbystainlessflexibletubeInstallationlocationrequirements:Seesection4.Minimum bending radius: 50 mm. To reduce optical attenuation, make the radiuses along the cable as

large as possible when laying cables.

2.6.2 ModelandSuffixCodesModel SuffixCode OptionCode Description

NR895 ••••••••••••••••• ••••••••••••••••• Ethernetfiber-opticcableLength*1 -L003

-L005-L010-L020

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

3 m5 m10 m20 m

- -000 ••••••••••••••••• Always “-000”Option /JB Withjunctionbox*2

*1: Foratotalcabledistanceof20morlonger,anadditionalfiber-opticcable(fittedwithSTconnectors)mustbeprovidedbythe user.

Seechapter6.2forthecablespecificationofuserprovisions.*2: TheJunctionboxisforconnectionbetweenanNR895fiber-opticcableandtheadditionalcable.

2.6.3 ExternalDimensions(1)Fiber-opticCable

15

9.310

Blue

Yellow

Unit: mm

1000 0+100

43 30

(Cable length)

Mar. 18, 2014-00

Page 33: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

33<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

(2)JunctionBoxforEthernetfiber-opticcables(Optioncode:/JB)

400 430

305

300

142

24120

280

60

20260

4-ø10

Handle to open

100 30Hole

2-Cable gland

Unit : mm

NR895_JB.ai

Housingstructure: IP44Coating: Epoxy resin coatingPaintcolor: Lightbeige(equivalenttoMunsell5Y7/1)Material: IronWeight: Approx.10kg

Mar. 18, 2014-00

Page 34: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

34

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.7 SPECTLAND2MeasurementandMaintenanceSoftware

2.7.1 OverviewSPECTLAND2isanapplicationsoftwarethatcontrolsNR800operationandmonitoringonaPC.Itdisplaysthe analyzer status, and instructs measurement and sets parameters for the analyzer. To use SPECTLAND2,firstinstallitintheEngineeringPC.ThecontinuousoperationofSPECTLAND2forextended periods of time is not guaranteed.

MainFeatures•Instructstheanalyzertomeasurespectra,savedata

and display. The measured spectra can be used to generate calibration models.

•Showstrendgraphsofthemeasurementsandsavesthemtofilesduringacontinuousmeasurement.

•Setsparametersforcontinuousmeasurement.•Displaysvariousdataoftheanalyzer,suchas

operating modes, alarm status, and maintenance data.

•Instructstheanalyzertoperformsuchtasksasoperation mode change, spectra measurement, and setting property information, calibration models, or measurement conditions.

2.7.2 MainWindows(1)ManualSpectrumWindow

Enables the analyzer to measure spectra, which are tobeprocessedbyChemometrics(calibrationmodelgenerationsoftware).Alsoallowsdatatobesavedtofilesanddisplayed.

(2)AutoSpectrumWindowAllows the user to upload spectra data to the EngineeringPCduringcontinuousmeasurementuponreceivingasignalatperiodicintervals,outlierdetection,orapropertyvaluevariationfailure.Thisdata is displayed for each measuring channel.

(3)PowerSpectrumWindowIt displays power spectra. ThiswindowisavailableforUBlevel(equipmentsupervisor)users.

(4)InterferogramWindowIt displays interferogram data collected. This window isavailableforUBlevelusers.

(5)Real-timeTrendWindowsDisplaymeasurementvaluesofNos.1to6andNos.7 to 12 components in two separate trend graph windowsforeachstream.Upto10windowscanbeopen at the same time.

(6)HistoricalTrendWindowsDisplayhistoricaltrenddatasaved.Trenddataof24hoursforeachstreamissavedtoafile.Upto4windows can be open at the same time.

(7)ParameterWindowDisplaysthecurrentparametersettingsfortheanalyzer.Inaddition,UBleveluserscanchangethesettings.

(8)AlarmStatus/HistoryWindowsTheAlarmStatuswindowdisplaystheactivealarmsfortheanalyzer,whiletheAlarmHistorywindowdisplays all the past alarms. The alarm history can be deleted with commands.

(9)Tab-controlledMaintenanceWindowDisplaystheA/Dreferencevalueandservo-relateddataoftheanalyzer.ThiswindowisavailableforUClevelusers.

(10) Tab-controlledCommunicationStatusWindow

Displaysthecommunicationstatusbetweenthepersonal computer and the analyzer. This window is availableforUClevel(maintenance)users.

2.7.3 ModelandSuffixCodes

Model SuffixCode

OptionCode Description

NR831 ••••••••••• ••••••••••• SPECTLAND2measurement and maintenance software

Language -E-J

••••••••••••••••••••••

EnglishJapanese

- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”Option /UP Versionup

Packagecontents: OneCD-ROM Oneuser’smanual

2.8 ChemometricsSoftware

2.8.1 Specifications•Calibrationmodelgeneratingtechnique: Partialleastsquare(PLS)andothers

2.8.2 ModelandSuffixCodes

Model SuffixCode

OptionCode Description

NR530 ••••••••••• ••••••••••• Chemometrics softwareLanguage -E ••••••••••• English- -N ••••••••••• Always “-N”- -N ••••••••••• Always “-N”

Packagecontents: TwoCD-ROMs(Installdisc/User'smanual)

Onesetofuserregistrationdocument

Mar. 18, 2014-00

Page 35: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

35<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

2.9 SamplingUnitUseofasamplingunitishighlyrecommendedtoensurecompatibilityoftheuser’sprocesssamplewiththemeasurementcell.Inaddition,itallowsseparationofthemeasurementcell(probe)apartfromtheanalyzerwithafiber-opticcableupto300minlength.Thisenablesselectionofameasurementlocationindependentoftheanalyzer.Theoptimumsamplingunitispreparedforindividualapplicationrequirements.ContactYokogawaforfurther information.

l ExampleofSamplingSystemThefollowingintroducesaconceptualsamplingunitformeasuringtheproperties(RON,RVP,Distillationetc.)ofgasolineblendinginapetroleumrefinery.Notethateachsamplingunitshouldbedesignedforeachapplication;designswillvary.

MeasurementSampleConditionsItem Condition

Fluid to be measured Gasoline

Sample inlet/outletcondition

Inlet pressure 0.3to0.9MPaGOutletpressure 0.3to0.9MPaGInlet temperature 0°Cto+40°C

Notes1: Whensamplesdonotcontainfreewatercontent,acoalescer(CF1)isnotrequired.2: Whensamplesdonothavebubbles,adegasser(DG1)isnotrequired.3: Ifthepressuredifferencebetweenthesampleinletandoutletisgreaterthan0.3MPa,asamplingpump(PU1)isnot

required.

com comcom com

PU1CHV1, 2, 3BV1, 2, 3, 4BV5, 6, 7, 8, 9V1, 2, 3, 4, 5, 6NV1, 2, 3, 4, 5FM1, 2, 3, 4PR1, 2CF1DG1P1, 2DT1TK1, 2, 3, 4CON1RN1

PUMPCHECK VALVE3-WAY BALL VALVE2-WAY BALL VALVESTOP VALVENEEDLE VALVEFLOW METERPRESSURE REGULATORCOALESCER FILTERDEGASSERPRESSURE GAUGEDETECTORTANKInfraSpec NR800RESISTANCE TEMP. SENSOR

: JIS 10K-15A-RF: Rc1/2: Rc1/4

<connection>

: SUS316TP 15A sch40: ø12/ø10 SUS316: ø8/ø6 SUS316: ø4/ø3 Nylon tube: ø6/ø4 Nylon tube: ø22/ø6 Braid tube

<piping>

outlet

Cooling waterinlet

*2PR2

Air inlet

*3

*4

*1PR1

PU1Sample inlet

P1

V1

NV1

FM1NV3NV4

BV1

DG1 BV2

CF1

BV9

Sample outlet

P2

V2

FM4FM3

Cooling water

*3 *4

*1

TK1

V3

V4

CH

V2

DT1 BV3C

HV

1

*2NV5 RN1

CON1

TK2 V5V6

a'

NV2FM2

BV4

Drain

TK4

BV

7

TK3

CH

V3

BV

8

a'

BV

5

BV

6SAMPLING UNIT

N2 gas inlet

N2 gas

Mar. 18, 2014-00

Page 36: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

36

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

2.10 AuxiliaryParts*Therecommendedreplacementintervalsforauxiliaryparts,listedinSection2.10.1,ModelandSuffixCodes,aresetforpreventivemaintenanceanddonotguaranteetheservicelifeofparts.Actualreplacementintervalsforauxiliarypartsvarydependingonoperatingconditions,sampleconditions,andotherfactors.Besuretoperformpreventivemaintenancebasedonoperationalstatusofinstrument.Theauxiliarypartsareclassifiedintotwotypes:thosethatcanbereplacedbycustomerandthosethatmustbereplacedbyaYokogawaserviceengineer.Priortoreplacement,refertotheuser’smanualandfollowtheinstructiontherein.* Auxiliarypartsmeanlimitedlifecomponentsthatareexceptedtobeintheirwear-outfailureperiodwithin5yearsunder

normal operating or storage conditions.

2.10.1ModelandSuffixCodes

Model SuffixCode

OptionCode Description

RecommendedReplacement

IntervalNR8SP01 •••••••• ••••••••• Auxiliary parts set for NR800

-01-02-03-04-05-06

-07-08-10-11-12-13-14-15-16-17

••••••••••••••••••••••••••••••••••••••••••••••••••••••

••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••••

Sealingpackingforanalyzerdoor(forNR801AG/NR805AG/NR801JG/NR805JG)Analyzercoolingfan(5pcs)Spectrometersupportinsulator(4pcs)Lamp(adjusted)LaserheadDesiccantforspectrometer(9pcs)

Desiccantforflow-throughcell(10pcs)*1Sealingpackingforanalyzerdoor(forNR801EG/NR805EG)O-ringforflow-throughcell(JISP38,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP21,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP16,FPM(fluorinerubber),2pcs)O-ringforflow-throughcell(JISP38,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP21,Kalrez4079,2pcs)O-ringforflow-throughcell(JISP16,Kalrez4079,2pcs)Windowpaneofflow-throughcell(borosilicatecrownglass,2pcs)*3Windowpaneofflow-throughcell(sapphire,2pcs)*3

1 year3 years3 years5,000 hours12,000 hours1year(Shelflife:1year after shipment)3 months1 year*1,*2*1,*2*1,*2*1,*2*1,*2*1,*2

*1: Whenameasurementcellisdisassembledforservice/cleaning,itisrecommendedthatO-ringsshouldbereplaced.*2: Replacementintervalsvarydependingonsamplecomposition,humidity,pressure,etc.Determinereplacementintervals

based on actual operational status.*3: Ifawindowisscratchedorfogged,replacethewholesetofwindows(2pieces/set).

3.UtilitySpecifications3.1PowerSupplies

(1)NR800analyzerPowersupply: 100,115,200,230VAC,singlephase,

50/60Hz(*1)Voltagefluctuation: Rating±10%,50/60±2HzPowerconsumption: Approx.250VA(*1): Tobespecified.Seethecorrespondingmodel

andsuffixcodesfordetails.

(2)RS-422/RS-232CconverterPowersupply: 100to120VACor200to240VAC,

singlephase,50/60HzVoltagefluctuation: Rating±10%,50/60±2Hz(Don’texceed

250VAC)Powerconsumption: Approx.15VA

(3)I/OunitACPowersupply: 100to240VAC±10%,singlephase,

50/60Hz±5%Powerconsumption:

NR893AG, NR893JG with enclosure (AOpoints0to12):Approx.150VA (AOpoints16to28):Approx.190VA (AOpoints32to40):Approx.220VANR893EG (AOpoints0to40):Approx.220VANR893JG without enclosure (AOpoints0to40):Approx.100VA

DCPowersupply: 24VDC±10%Powerconsumption: Approx.180mVA/AO4points*: RequiredwhenintheNR893JGI/Ounit,option

code“/C”isnotspecified.

May 15, 2014-01

Page 37: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

37<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

3.2 Others

(1)Clean,dryairforAnalyzerpurge(explosion-proofmodel)

Item DescriptionTemperature -10°Cto+40°CPressure 0.3to0.9MPa

Dewpoint-20°Corlower(atpressure)Compressed air should not dew at the installation ambient temperature.

Cleanliness

Mustbefreefromdust,corrosive,andtoxicelements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.

Volume; Approx.75l/min(stp.)

(2)Clean,dryairforin-situprobeItem Description

Temperature -10°Cto+40°CPressure Atmospheric pressure

Dewpoint-20°Corlower(atpressure)Compressed air should not dew at the installation ambient temperature.

Cleanliness

Mustbefreefromdust,corrosive,andtoxicelements.Impurities, such as oil, cause damage to the optical components. Eliminate them with an activatedcarbonfilter.

Volume;Approx.0.3to0.5l/min(stp.).Dependonsampletemperatureandambient temperature.

(3)Waterforflowthroughcellwithwatertube;Dependonmeasurementspecifications.Temperature,response speed, etc.,

(4)Utilityforsamplingunit(whenused)Dependonsamplingsystemconfiguration,sampletemperature, response speed, etc.

4.InstallationLocationRequirementsAvoidphysicalshockasitmaycausedamagetotheequipment.

4.1 NR800analyzer,measurementcell,andfiber-opticcable

Item Description

Location

Hazardousarea/non-hazardousarea.Outdoor/indoor.Avoiddirectexposuretowind,rain,sunlight and radiation heat.Altitude 2,000 m max.

Ambient temperature -10°Cto+40°CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2 or less)

AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.

4.2 RS-422/RS-232CconverterandI/OunitItem Description

Location

Non-hazardous area, indoor.Avoiddirectexposuretowindandrain, sunlight, or radiation heat.Altitude 2,000 m max.

Ambient temperature +5°Cto+35°CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2orless)

AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.

4.3MaintenanceSpaceforNR800AnalyzerWheninstallingtheanalyzer,tofacilitateoperationandmaintenance,providespacearounditasfollowingillustration.

561

300 577 200305

1200

1077

(Maintenance space)

50

Unit: mm

*NR800

Analyzer

* JunctionboxforEthernetfiber-opticcables.(WhenNR895/JBisselected.)Seesection2.6.3for external dimensions.

4.4 StorageRequirementsAvoidphysicalshockasitmaycausedamagetotheequipment.

Item Description

LocationIndoor.Avoiddirectexposuretowindandrain, sunlight, or radiation heat

Ambient temperature -10°Cto+50°CAmbient humidity 5%to95%RH(nocondensation)

VibrationInstalltheequipmentinaplacewithminimumvibration(vibrationaccelerationof2m/s2 or less)

AtmosphereMustnotcontaincorrosiveortoxicsubstances.Must be free from dust.

Note: Fluoride, a material used in the NR823 and NR824 fiber-opticcables,hasthedeliquescenceproperties.Whenstrongthesecables,takeappropriatemeasurestorpeventmoisture,e.g.,acableshouldbeplacedwithdesccant in a sealed, moisture-barrier plastic bag and the desiccant replaced regularly.

Mar. 18, 2014-00

Page 38: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

38

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

5.SupportforCalibrationModelGeneration

5.1 On-siteGuidanceofCalibrationModelGeneration

AYokogawaengineerwilltrainanuser'ssitepersonnelin the procedure to generate a calibration model for one measureditemusingauser-providedsamplewithitslaboratory analysis results.

5.2 CalibrationModelGenerationYokogawa generates a calibration model using the necessaryquantityofuser-providedsampleswithlaboratoryanalysisresults.ApredefinedSEP(standarderrorofprediction)valueof1σwillbeusedasthemeasurementtargetvalue.Thetargetvalue,samplequantity,andotherdetailsaredeterminedseparatelyfor each application.

5.3 OthersOthersupportoptionsforcalibrationmodelgenerationand maintenance include:•Samplingtestforpotentialusers•Maintenancecontracts•Sampling/modelgeneration/maintenanceconsultingservice.

ContactaYokogawasalesrepresentativeforfurtherinformation,andadviceonthebestsolutionforyourneeds.

6.RecommendedSpecificationsforEngineeringPCandrelatedEquipmentprovidedbyuserthatsatisfiesthefollowingrequirements:

6.1 PC

Computer IBMPC/ATcompatible

Operatingsystem(OS)

MicrosoftWindows7Professional32bitSP1 orMicrosoftWindowsVistaBusinessEdition

CPU 1GHzorhigherRAM 1 GB or more

Harddiskspace 10GB(forprogram)and25GB(fordatastorage) or more

Ethernet adapter 10 BASE-TDisplay SVGAmode(1024x768pixelsormore)Other CD-Rdriveisrecommended

l ColorPrinterPrepareifnecessary.

l Connectioncablesandotherdevicesandconsumables

Prepareifnecessary.

6.2 EthernetCommunicationWhenfiber-opticcablesareusedforEthernetcommunication(seesection2.6)Anoptical-to-electricalsignalconverter(wavelength:850nm)foramultimodefiber-opticcablethatcomplieswithEthernetversion2.0andIEEE802.310Base-Tand10Base-FLstandardsandisfittedwithSTconnectors,fiber-opticextensioncables, and electric cables for Ethernet communication shouldbeprovidedbycustomer.WhenelectriccablesareusedforEthernetcommunication Electric cables for Ethernet communicationshouldbeprovidedbycustomer.

l AdditionalEthernetfiber-opticcableMaterialandconnector: SilicaglassfiberwithST

connectorsMode: Multi-mode GINumber of cores: 2Core/Claddiameters: 50/125µmApplicablewavelength: 850nmLength: 1000morlessintotalincludingan

NR895fiber-opticcable

l SpecificationsofelectriccableforEthernetcommunication

Type: 10Base-T, 8-core shielded, straight connection, RJ-45 connectors

NotethatwhenanNR895fiber-opticcable for Ethernet communication is not used,useanEthernetcrossovercable.Ifyou use a hub, consult Yokogawa.

Length: 3to40mFinished outside diameter: 5.0 to 7.4 mm

May 15, 2014-01

Page 39: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

39<<Contents>> <<Index>>

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation GS 12Y03A03-01E

SpecificationSheetsforinquiryofNR800series(1/2)Thankyouforyourinquiry.Pleasefillitcustomer’srequestsinordertomakeaquotation.Tick all which apply.

1. ScopeofQuotationEssential: AnalyzerMeasurementcell/probe,Fiber-optic,Software,User’smanual(1copy)Optional: Document(Specificationforapproval copy Testcertificate copy) Factoryacceptancetest CalibrationModelimplementationbyYokogawa On-sitehands-ontrainingofCalibrationmodelimplementation

2. GeneralUser’sname: Purpose : Control Monitor/alarm OthersPlantname: Document: English Japanese

3. SpecificationsNumberofmeasuringchannel: 1ch 2ch 3ch 4ch PossibilityofthechannelexpansionAnalogueoutput: Need( points) NotneedCommunication: Need NotneedMeasurementcell/probe: No.1ch(cellprobe) No.2ch(cellprobe) No.3ch(cellprobe) No.4ch(cellprobe)Lengthoffiber-optics: No.1ch(Layingdistance: m) No.2ch(Layingdistance: m) No.3ch(Layingdistance: m) No.4ch(Layingdistance: m)

4. InstallationConditionAnalyzer: Indoors Outdoors explosion-proofnon-ex-proof Toxic/Corrosiveambient Restrictedinstallationarea( m x m)Measurementcell/probe: Indoors Outdoors Toxic/Corrosiveambient

5. UtilitiesPowersupply: VAC, HzInstrumentair: (Pressure MPa) NoneNote:Clearanddryairisrequiredforpurgingexplosion-prooftypeanalyzer.Seesection3.2.

6. SampleconditionPleasefillinnextpage.

7. SpecialnotePleasefilloutothernotes.

Mar. 18, 2014-00

Page 40: General InfraSpec Specifications Near-Infrared Analyzers · General Specifications   InfraSpec NR800 Fourier Transform Near-Infrared Analyzers

40

AllRightsReserved.Copyright©2001,YokogawaElectricCorporation

<<Contents>> <<Index>>

GS 12Y03A03-01E

SpecificationSheetsforinquiryofNR800series(2/2)Samplecomposition/MeasuringConstituentList

Pleasefilloutallsampleconstituentincludingnon-measuringconstituent.No.1chSamplename:

Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Changing/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.2chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Changing/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.3chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Changing/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

No.4chSamplename: Sample temperature: to ºC, normal ºCSample pressure: to MPa,normal MPaBabble/slurry: Bubble Slurry Others( )

Tick for measuring constituent

ConstituentFill out all constituent Changing/measuringrange Normal Unit

1 to2 to3 to4 to5 to6 to7 to8 to9 to10 to

Mar. 18, 2014-00Subjecttochangewithoutnotice.