GDS10+ · 2020. 7. 22. · Sensor TP3 Volts TP2 TP4 L A 3 Wire Sink 3 Wire Source JPI Jumpers DB...
Transcript of GDS10+ · 2020. 7. 22. · Sensor TP3 Volts TP2 TP4 L A 3 Wire Sink 3 Wire Source JPI Jumpers DB...
E: [email protected] T: +44 (0)113 286 0166
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
PRODUCT DESCRIPTION: The GDS10+ control unit boasts a compact design, initially for monitoring gases within cabins, enclosures or cabinets. However, with the optional feature of a mains supply input the system can be expanded for use in conventional areas of gas detection such as commercial and other light industrial applications.
GDS10+GAS MONITOR
E: [email protected] T: +44 (0)113 286 0166
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
controller configuration
E: [email protected] T: 0113 286 0166 F: 0113 287 8178 GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
T E C H N O L O G I E SGDS10+ CONTROLLER CONFIGURATION
Data Sheet ref C1606 Issue A v.2
This document is not contractual and the equipment specification may be modified at any time without prior notice.
15J
Internal orRemote Sensor
XDI Sensor
3 co
re 1
.5m
m
Remote Sensor OptionsCatalytic mV only
Status Indicatorx unlimited
Aspirated Internal Sensor Option
Aspirated Sensor Option
Defender Sensor
145
147.0
Cover Lid Removed
Gas Sen
sor
10
145
.2
147.0
99.
3
Cover Lid Removed
A
B B
A
A A
CC
CC
85.0 (D) - 30J60.0 (S) - 15J
104
.5
Cover Lid Removed / Mounting Points
A
B B
A
A A
CC
CC
98
DETECTOR HEADType – P1 Safe area useIngress ProtectionIP63, + Universal fitting IP65MaterialABS - blackSensor Inserts - 4 seriesElectrochemicalInfra RedSemiconductorPellistor - Catalytic PID
Accessories - OptionalCollector Cone + Universal FittingUniversal Fitting (test gas applicator/spray deflector)Flow BlockDuct Mount KitDetector Head Weather Shield
JUNCTION BOXType - 15/30J safe area use only 15J - shallow box (S) 30J - deep box (D)Ingress ProtectionIP64MaterialABS Flame retardent FR40Lid Screws M4SSFinishSignal White RAL 9003Weight380gmsEntriesRear 5-20mm knock outsSides, Top, Bottom - not specifiedMountingStandoffs - M4 or No 8 screws Drill at (C) when standoffs removedConduit box - drill at (A) 4.5mmSurface mount box - drill at (B) 4.5mmEU Design No.01359723-0001
E: [email protected] T: 0113 286 0166 F: 0113 287 8178
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
This document is not contractual and the equipment specification may be modified at any time without prior notice. C1268
Specification Sheet ref C1274A V5
T E C H N O L O G I E S
E: [email protected] T: 0113 286 0166 F: 0113 287 8178
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
This document is not contractual and the equipment specification may be modified at any time without prior notice. C1015
Specification Sheet ref C1546C v.2
Oxygen 0~25%
Carbon Monoxide 0~250ppm
Hydrogen Sulphide 0~50ppm
Sulphur Dioxide 0~10ppm
Nitrogen Dioxide 0~10ppm
Hydrogen 0~2000ppm
Chlorine 0~10ppm
Hydrogen Cyanide 0~25ppm
Nitrogen Monoxide 0~100ppm
Hydrogen Chloride 0~10ppm
Ammonia 0~100ppm 0~1000ppm 0~5000ppm
Ozone 0~2ppm
Ethylene Oxide 0~25ppm
Flammable 0~100LEL
Refrigerant 0~1000ppm
CO2 0~2% Vol. & 0~100% Vol.
Toxic/Oxygen PCB Flammable/Refrigerant PCB
24v DCOutput
W Y P
No C Nc
NCCNO
- ++ +
SensorTerminal
Sounder
RELAYBOARD
SensorTerminal
EOL
Output SignalAdjust
Test Points
Alarm LevelAdjust
AlarmLED
JP2
20m
A
4mA
Spa
n
Zero
DP
M
E
A
JP1
-+
W Y R
- +J4
JP3
JP4
X Y
BCD
W+
Y SIG
P -Field Terminalsto Alarm Unit
U5
LK1
A L
Y SIG
Field Terminals toalarm unit
Output SignalAdjust
W Y P
No C Nc
NCCNO
SensorTerminal
Sounder
RELAYBOARD
REMOTEOPTION
T F
EOL
1ARelayRL2Sensor
volts adjustSensor
Ala
rmS
et
AlarmLED
Gas
3ARelayRL1
1ARelayRL2
3ARelayRL1
ON/Zero
DP
MZ
20m
A
4mA
Zero
DP
MS
E
W Y PJ4
JP1
A B
C
LK1
JP2
W+
P -++-
TP1
TP3SensorVolts
TP4TP2L A
3 WireSink
3 WireSource
JPI Jumpers DB
JPI Jumpers AC
250 ohms maximum load
-
+
-
+E
SIG
-
+
-
+E
SIG
-
24v DCOutput
-
24v DCOutput
W Y P
No C Nc
NCCNO
- ++ +
SensorTerminal
Sounder
RELAYBOARD
SensorTerminal
EOL
Output SignalAdjust
Test Points
Alarm LevelAdjust
AlarmLED
JP2
20m
A
4mA
Spa
n
Zero
DP
M
E
A
JP1
-+
W Y R
- +J4
JP3
JP4
X Y
BCD
W+
Y SIG
P -Field Terminalsto Alarm Unit
U5
LK1
A L
Y SIG
Field Terminals toalarm unit
Output SignalAdjust
W Y P
No C Nc
NCCNO
SensorTerminal
Sounder
RELAYBOARD
REMOTEOPTION
T F
EOL
1ARelayRL2Sensor
volts adjustSensor
Ala
rmS
etAlarmLED
Gas
3ARelayRL1
1ARelayRL2
3ARelayRL1
ON/Zero
DP
MZ
20m
A
4mA
Zero
DP
MS
E
W Y PJ4
JP1
A B
C
LK1
JP2
W+
P -++-
TP1
TP3SensorVolts
TP4TP2L A
3 WireSink
3 WireSource
JPI Jumpers DB
JPI Jumpers AC
250 ohms maximum load
-
+
-
+E
SIG
-
+
-
+E
SIG
-
24v DCOutput
-
24v DCOutput
W Y P
No C Nc
NCCNO
- ++ +
SensorTerminal
Sounder
RELAYBOARD
SensorTerminal
EOL
Output SignalAdjust
Test Points
Alarm LevelAdjust
AlarmLED
JP2
20m
A
4mA
Spa
n
Zero
DP
M
E
A
JP1
-+
W Y R
- +J4
JP3
JP4
X Y
BCD
W+
Y SIG
P -Field Terminalsto GDS Alarm Unit
U5
LK1
A L
Y SIG
Field Terminals toGDS alarm unit
Output SignalAdjust
W Y P
No C Nc
NCCNO
SensorTerminal
Sounder
RELAYBOARD
T F
EOL
1ARelayRL2Sensor
volts adjustSensor
Ala
rmS
et
AlarmLED
Gas
3ARelayRL1
1ARelayRL2
3ARelayRL1
ON/Zero
DP
MZ
20m
A
4mA
Zero
DP
MS
E
W Y PJ4
JP1
A B
C
LK1
JP2
W+
P -++-
TP1
TP3SensorVolts
TP4TP2L A
3 WireSink
3 WireSource
JPI Jumpers DB
JPI Jumpers AC
250 ohms maximum load
-
+
-
+E
SIG
-
+
-
+E
SIG
-
24v DCOutput
-
2 Wire
24vDC
145
.0
Cover Lid Removed
Cover Lid Removed
A
B B
A
A A
90.0 - D10
CC
CC
145
.2
147.0
99.
3
Cover Lid Removed
A
B B
A
A A
CC
CC
147.0
GDS10+Gas Sensor
E: [email protected] T: +44 (0)113 286 0166
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
Gases and Ranges
Over 700 gas types are available – Contact our sales department with your requests - [email protected]
Gas Relative Density Air =1 Units Standard Range Alarm Level 1 Alarm Level 2
Oxygen (O2) 1.105 % Vol 0-25% 19% 17%
Carbon Monoxide (CO) 0.96 ppm 0~200ppm 20ppm 100ppm
Sulphur Dioxide (SO2) 2.26 ppm 0~5ppm 0.5ppm 1ppm
Nitric Oxide (NO) 1.03 ppm 0~10ppm 2ppm 3ppm
Nitrogen Dioxide (NO2) 2.62 ppm 0~5ppm 0.5ppm 1ppm
Chlorine (Cl) 2.41 ppm 0~5ppm 0.3ppm 0.5ppm
Chlorine Dioxide (ClO2) 2.33 ppm 0~1ppm 0.1ppm 0.3ppm
Hydrogen (H2) 0.06 ppm 0~2000ppm 250ppm 500ppm
Hydrogen Cyanide (HCN) 0.93 ppm 0~10ppm 0.9ppm 4.5ppm
Hydrogen Sulphide (H2S) 1.18 ppm 0~50ppm 5ppm 10ppm
Hydrogen Chloride (HCl) 1.26 ppm 0~10ppm 1ppm 5ppm
Ammonia (NH3) 0.58 ppm 0~100ppm 25ppm 35ppm
Ozone (O3) 1.66 ppm 0~1ppm 0.1ppm 0.2ppm
Hydrogen Fluoride (HF) 0.7 ppm 0~10ppm 1.8ppm 3ppm
Methane (CH4) 0.55 % LEL 0~100% 20% 40%
Propane (C3H8) 1.55 % LEL 0~100% 20% 40%
Butane (C4H10) 2.11 % LEL 0~100% 20% 40%
Pentane (C5H12) 2.48 % LEL 0~100% 20% 40%
LPG 1.55 % LEL 0~100% 20% 40%
Hydrogen (H2) 0.06 % LEL 0~100% 20% 40%
Carbon Dioxide (CO2) 1.52 % Vol 0-2% 0.50% 1.50%
Methane (CH4) 0.55 % LEL 0~100% 20% 40%
Propane (C3H8) 1.55 % LEL 0~100% 20% 40%
Butane (C4H10) 2.11 % LEL 0~100% 20% 40%
Pentane (C5H12) 2.48 % LEL 0~100% 20% 40%
LPG 1.55 % LEL 0~100% 20% 40%
Ethanol (C2H60) 1.51 % LEL 0~100% 20% 40%
Iso-butylene (C4H8) 1.94 ppm 200ppm 20ppm 40ppm
E: [email protected] T: +44 (0)113 286 0166
GDS TECHNOLOGIES LTD | FUSION POINT | ASH LANE | GARFORTH | LEEDS | LS25 2GA | UK www.gds-technologies.co.uk
GDS10+Channels Single Channel
Measurements Combustible gas – LEL, % Vol.Toxic Gas – ppm, % Vol.
Oxygen - % Vol. – Depletion / Enrichment Refrigerant – ppm
Sensor Types Remote - Bridge type 3 wire mv (Catalytic) Internal - Infrared / Catalytic / PID / Semiconductor / Electrochemical
Single Stage Alarm 5~100% FSD
Channel Outputs Analogue output 4-20mA – option 1~5 volts. 250 ohms max load
Relays S.P.C.O. per relay – auto resetRL1 – 3A/230v AC control relay
RL2 – 1A/24v DC signal relay*Single Alarm*
Sensor Cable 3 Core 1.5mm² screened 1.5km
Power 12 to 30v DC – 24v nominal – 160mA230-115v AC optional
Panel Indication Alarm - Red LED
Audible Alarm Auto reset 75dBs @ 1m
Operating Temperature -10 to + 50° C
Storage Temperature +5° C to + 55° C
Humidity 0-95 non condensing
Ingress Protection IP64 Weatherproof
Dimensions H-147 W-145 D-65mm (24vDC Version)H-147 W-145 D-100mm (AC Version)
Weight 0.32KG (24vDC Version)0.76KG (AC Version)
Enclosure Material ABS Flame Retardant FR40 – Lid Screws M4SSFinish – Signal White RAL 9003
Standards EMC BS EN 50270Low Voltage BS EN 50270
Specification
GDS10+GAS MONITOR
Specification Sheet Ref C1822A
GDS Technologies Ltd operates under an integrated quality, health and safety and environmental management system certified to ISO 9001, ISO 45001 and ISO 14001 by NQA Certification Ltd.
This document is not contractual and the equipment specification may be modified at any time without prior notice.