DU-DR-0016-ME0003 Rev B 4/02/01 Page 1 of 3 GCX Mounting Assembly Operation/Installation Manual Dräger Model Cato ® and Cato ® Edition with GEMS Information Technologies Eagle ® 4000 Patient Monitor The purpose of these instructions is to: 1. Describe attachment of GCX mounting hardware to the Cato/Cato Edition anaesthesia machine. 2. Describe attachment of the Eagle 4000 monitor to the GCX mounting hardware. WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE ANAESTHESIA MACHINE. ONLY THE CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH EXTENDED WHEEL ADAPTER UPGRADE SET FROM DRÄGER (PART NO.: 86 018 68). Installer: When installation is completed, provide these instructions to the end user for future reference. INSTALLATION - Left/Right with Pivot Arm Tools Required: Philips head screw driver, Hex key (incl.). Mounting the Channel to Cato/Cato Edition Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2). Brass Plate with inserts from Dräger (M29110) 1. To attach the Channel to the Cato left side accessory track, locate the brass plate in the accessory track, align the mounting holes in the Channel with the threaded holes on the brass plate then run the M6 flat head screws (2) partly in. (Note: Fixed Stop is at the bottom of the Channel) 2. Set the height of the Channel in the accessory track and tighten the M6 screws to hold in place. Mounting the Adapter to Eagle 4000 Parts Required: Adapter, (4) #8-32 x 11/4" screws, (4) .125" washers. 1. Position the handle at the back of the monitor. 2. Attach the Mounting Adapter/Handle to the rear of the Eagle 4000 using the #8-32 x 1¼” machine screws and .125” spacers (between plate and handle). (Note: If the Handle is not used, used the supplied #8-32 x 3/8" screws) Mount the Pivot Arm into the Channel 1. Align the Rear Slide on the Pivot Arm with the Channel and insert. Mounting the Eagle 4000 to the Pivot Arm 1. Slide the Eagle 4000 horizontally into the Mounting Plate until the Spring Plunger locates in the clearance hole in the Mounting Adapter. Tighten the (2) nylon thumb screws at the rear of the Mounting Plate. Eagle 4000 Patient Monitor M6 x 16 mm (2) FHMS Accessory Track Pivot Arm Mounting Plate Spring Plunger Mounting Adapter #8-32 x 1 ¼” Machine Screws Handle Fixed Stop Mounting Positions: - Left/Right Side w/Pivot Arm - Top Shelf (Not with Cato Edition)

Transcript of GCX Mounting Assembly Operation/Installation Manual · Eagle 4000 using the #8-32 x 1¼” machine...

Page 1: GCX Mounting Assembly Operation/Installation Manual · Eagle 4000 using the #8-32 x 1¼” machine screws and .125” spacers (between plate and handle). (Note: If the Handle is not

DU-DR-0016-ME0003 Rev B 4/02/01 Page 1 of 3

GCX Mounting Assembly Operation/Installation Manual Dräger Model Cato® and Cato® Edition with GEMS Information Technologies Eagle® 4000 Patient Monitor

The purpose of these instructions is to:

1. Describe attachment of GCX mounting hardware to the Cato/Cato Edition anaesthesia machine. 2. Describe attachment of the Eagle 4000 monitor to the GCX mounting hardware.

WARNING: USE OF MOUNTING HARDWARE AND MONITORING COMPONENTS OTHER THAN THOSE DESCRIBED IN THIS DOCUMENT MAY RESULT IN SERIOUS INJURY DUE TO TIPPING OF THE ANAESTHESIA MACHINE.

ONLY THE CATO MOUNTING CONFIGURATION REQUIRES THE 40KG COUNTERWEIGHT WITH EXTENDED WHEEL ADAPTER UPGRADE SET FROM DRÄGER (PART NO.: 86 018 68).

Installer: When installation is completed, provide these instructions to the end user for future reference. INSTALLATION - Left/Right with Pivot Arm

Tools Required: Philips head screw driver, Hex key (incl.).

Mounting the Channel to Cato/Cato Edition

Parts Required: 4" Channel, M6 x 16 mm FHMS screws (2). Brass Plate with inserts from Dräger (M29110)

1. To attach the Channel to the Cato left side accessory track, locate the brass plate in the accessory track, align the mounting holes in the Channel with the threaded holes on the brass plate then run the M6 flat head screws (2) partly in. (Note: Fixed Stop is at the bottom of the Channel) 2. Set the height of the Channel in the accessory track and tighten the M6 screws to hold in place.

Mounting the Adapter to Eagle 4000

Parts Required: Adapter, (4) #8-32 x 11/4" screws, (4) .125" washers.

1. Position the handle at the back of the monitor. 2. Attach the Mounting Adapter/Handle to the rear of the Eagle 4000 using the #8-32 x 1¼” machine screws and .125” spacers (between plate and handle). (Note: If the Handle is not used, used the supplied #8-32 x 3/8" screws)

Mount the Pivot Arm into the Channel

1. Align the Rear Slide on the Pivot Arm with the Channel and insert.

Mounting the Eagle 4000 to the Pivot Arm

1. Slide the Eagle 4000 horizontally into the Mounting Plate until the Spring Plunger locates in the clearance hole in the Mounting Adapter. Tighten the (2) nylon thumb screws at the rear of the Mounting Plate.

Eagle 4000 Patient Monitor

M6 x 16 mm (2) FHMS

Accessory Track

Pivot ArmMounting Plate

Spring Plunger

MountingAdapter

#8-32 x 1 ¼” Machine Screws

Handle

Fixed Stop

Mounting Positions:

- Left/Right Side w/Pivot Arm

- Top Shelf (Not with Cato Edition)

Page 2: GCX Mounting Assembly Operation/Installation Manual · Eagle 4000 using the #8-32 x 1¼” machine screws and .125” spacers (between plate and handle). (Note: If the Handle is not

DU-DR-0016-ME0003 Rev B 4/02/01 Page 2 of 3

INSTALLATION - Top Shelf

FOR CATO ONLY, NOT CATO EDITION

Tools Required: Philips head screw driver Install The Top Shelf RS Housing Pan Parts Required: RS Top Shelf Pan 1) Insert the front guides into the RS housing. 2) Lower the back edge an insert the rear guides into the RS housing. 3. Thread the thumbscrews onto the threaded post straddled by the rear guides of the tray and tighten to secure tray. Install the Eagle 4000 Bottom Attachment Pan/Horizontal Channel Adapter Parts Required: Bottom Attachment Pan with pre-installed Horizontal Channel Adapter 1. Install the Bottom Attachment Pan on the Eagle 4000 using the supplied #8-32 x 3/8" BHMS (5) screws. Install the Eagle 4000 (into Channel) 1. Align the Horizontal Channel Adapter on the Bottom Attachment Pan with the Horizontal Channel on the RS Top Shelf Pan and slide to desired position. 2. Once in position, tighten the #8-32 x 7/16" PHMS (2) set screws located at the rear of the Horizontal Channel Adapter.

Install and tighten thumbscrews (2)

#8-32 x 3/8" BHMS (5)

Horizontal Mount not Shown

Page 3: GCX Mounting Assembly Operation/Installation Manual · Eagle 4000 using the #8-32 x 1¼” machine screws and .125” spacers (between plate and handle). (Note: If the Handle is not

DU-DR-0016-ME0003 Rev B 4/02/01 Page 3 of 3

INSTRUCTIONS FOR USE - Pivot Arm Pivot

1. To pivot or move the Flat Panel Display, simply push on the side of the Arm or the Display. 2. To adjust the tension of the pivot action, tighten or loosen the pivot adjustment screw (at the top of pivot block, above the slot, just below the arm) with the hex key provided. Swivel

1. To rotate the Flat Panel Display, simply push or pull the corners of the display. Tilt

1. Adjust the overall tilting tension by evenly tightening or loosening the (2) #8-32 x 1/2” SHCS - (Tension Adjustment Screws) with the hex key provided. Use the Adjustment Lever to fine tune and lock the tilt position. 2. Turn Adjustment Lever counter clockwise to loosen. Grasp the top and/or bottom of the monitor/display and tilt to the desired angle. 3. Turn Adjustment Lever clockwise to tighten and lock position. Cleaning

1. The mounting assembly may be cleaned with most mild, non-abrasive solutions commonly used in the hospital environment (e.g. diluted bleach, ammonia, or alcohol solutions). 2. The surface finish will be permanently damaged by strong chemicals and solvents such as acetone and trichloroethylene. 3. Steel wool or other abrasive material should never be used. 4. Damage caused by the use of unapproved substances or processes will not be warranted. To verify compatibility, It is recommended that any cleaning solution be tested on a small area of the arm that is not visible. 5. Never submerge the mounting arm. Wipe any cleaning agents off of the arm immediately, using a water-dampened cloth. Dry the arm thoroughly after cleaning.

CAUTION: GCX makes no claims regarding the efficacy of the listed chemicals or processes as a means for controlling infection. Consult your hospital’s infection control officer or epidemiologist. To clean or sterilize mounted devices or accessory equipment, refer to the specific instructions delivered with those products.

PARTS LIST - Left/Right with Pivot Arm

GCX P/N Dräger P/N GE-MMS P/N DESCRIPTION DR-0016-82 MM15719 4” Cato Channel WM-0017-13 MM15669 12” Pivot Arm with Manual Tilt and Riser ME-0003-86 MM15792 Eagle 4000 Rear Adapter Plate/Handle

M29110 Accessory Track Brass Plate (DRÄGER)

Optional: ME-0003-17 MM15692 Eagle 4000 Rear Adapter Plate w/o Handle

PARTS LIST - Top Shelf (FOR CATO ONLY, NOT CATO EDITION)

GCX P/N Dräger P/N GE-MMS P/N DESCRIPTION DR-0021-04 MM15690 Top Shelf RS Housing Pan w/o Channel DR-0021-02 MM15708 15” Channel for P/N MM15690 ME-0003-88 MM15678 Eagle 4000 Bottom Attachment Horizontal Mount

Mounting Plate

Tilt Tension Adjustment Lever

Pivot Block

Pivot TensionAdjustment

Screw

#8-32 x ½” SHCS(2) Tension Adjustment Screws

Slide