OPERATING INSTRUCTIONS
[GROWTH CHAMBER] Model : GC-300/1000
Manual No : 00HAA0002010 (Rev. 3)
Page 2
◆ NOTICE.
Thank you for purchasing Jeio Tech’ s product. This operation manual forms a definition of warning marks according to the level of importance and danger in order to use the product safely and correctly and prevent the users from accidents or injuries. Hence, please use the product in accordance with the instructions.
JEIOTECH CO., LTD
Page 3
■ Notice for the manual
1. Always keep this manual near to the instrument.
2. Check the manual in detail for correct operation
3. This manual is only for instruction of how to operate the instrument. Therefore,
anything out of the contents of the manual is not guaranteed.
4. It is prohibited to print and distribute the manual partly or entirely without
leave.
5. Please contact where you purchased the instrument if faults are found in the
manual.
■ Alert mark for safety operation
1. The alert marks are for safety operation and protect user and instrument from
damage.
2. Each mark is identified following to the degree of risks or importance.
User can damage seriously, or lose his or her life
if they operate the instrument wrongly.
User can damage or the instrument can be damaged if
they operate the instrument wrongly.
This is indicated as “ground connection"
The instrument should be set on a plat part.
This is for reference
3. Pay attention to the contents of alert marks.
4. Specially important “alert marks" are attached on the main body.(On the door)
5. Please change the alert marks if they are worn out
☞ Ask alert marks to where the instrument is purchased from if they are needed to
replace.
Page 4
■ Notice for safety and alteration
1. Read the manual carefully for safety before starting to operate the instrument.
2. Any damage that is happened due to carelessness is out of the manufacturer's
responsibility.
3. It's prohibited to alter the instrument.
4. Departing, fixing, and altering the instrument at your will can cause electric
shock, fire, and dysfunction.
5. Please contact where you purchased the instrument when some necessary parts should
be changed.
6. External shock to this instrument can cause damage and dysfunction.
■ Exemption
1. Any claim which is out the quality guaranteed by the manufacturer is our of the
manufacturer's responsibility.
2. Any damage that is from unexpected fault or damage of user by Acts of God is out
of Manufacturer's responsibility.
Page 5
◈ Unpacking your growth chamber
Please follow the orders as below.
1. Check the instrument.
1. Check the package condition before unpacking
2. Unpack carefully
3. Check the instrument carefully if damage is happened while transportation.
4. Check the accessories and related parts.
2. How to deal with damaged items
1. Please contact where you purchased the instrument if there is damage to the
instrument.
2. Keep the instrument as you received for the first time and check with the
transporter.
3. Issue claims as soon as you can to where you purchased the instrument.
3. Warranty for transportation
1. Responsibilities for damages happened while transportation is up to transporters.
2. The manufacturer will deal with any damage except for periods of transportation.
3. The instrument that is not transported by manufacturer's entrusted transporter
is out of manufacturer's responsibility.
Page 6
4. Composition of the unit 1) GC-300
<Image> <Description / Quantity>
Unit
1EA
Operating instructions
1EA
Shelves
3 EA
Shelf bracket
12EA
Industrial plug
/socket
20A / 1EA
Silicon hose
0.5M caliber 8mm / 2EA
1.5M caliber 10mm / 1EA
Page 7
Water tank
10 L / 1EA
Water tank prop
1EA
2) GC-1000
<Image> <Description / Quantity>
Unit
1EA
Operating instructions
1EA
Selves
6 EA
Shelf bracket
24EA
Page 8
Industrial plug/ socket
30A / 1EA
Silicon hose
0.5M caliber 10mm / 3EA
1.5M caliber 10mm / 1EA
Upper light
1 SET
Metal lamp
400W / 3EA
Recorder instructions
Optional / 1EA
Page 9
CONTENTS ◆ NOTICE …………………………………………………………………………………………… 3
◆ Transportation …………………………………………………………………………………… 5
◆ Composition of the unit ……………………………………………………………………… 6
OPERATING MANUAL
1 Alert and notice for Safety ………………………………………………………………… 10
2 Installation of the composition …………………………………………………………… 14
3.3-1 General purposes ……………………………………………………………………… 16
3-2 Features ……………………………………………………… ………………………… 17
4. Name of parts
4-1. GC-300 ……………………………………………………………………………………… 19
4-2. GC-1000 ……………………………………………………………………………………… 20
4-2. Lamp ………………………………………………………………………………………… 21
5. How to use Controller
5-1. Name & Operation ………………………………………………………………………… 22
5-2. How to operate Controller …………………………………………………………… 24
6. How to solve problems ……………………………………………………………………… 40
7. Daily Checking ………………………………………………………………………………… 41
8. How to Clean …………………………………………………………………………………… 42
9. Water tank usage …………………………………………………………………………… 44
10. Specification ………………………………………………………………………………… 46
11. Warranty ………………………………………………………………………………………… 47
12. How to waste ………………………………………………………………………………… 47
Page 10
Alert for safety
♣ Read the notice carefully
♣ This is for keeping user and their property from damage
Warning means that failure to follow this
safety statement could result in serious
personal injury, or death.
Check Voltage, Phase, and Capacity for
correct connection.
☞ Wrong connection can cause fire or
electric shock.
☞ Power cord should be connected to consent
with ground terminal
Don't insert a lot of plug into the
outlet at once!
☞ It causes fire and electric shock
Power with ground connection should be
used.
☞ Power without ground connection
causes serious damage to users or
equipment . Do not ground with gas or
water pipe.
Please check the power supply and
frequency on the ID plate before
installation.
☞ The unit requires a grounded and
polarized 220 VAC, single phase.
GC-300 : 16A, 2.0 SQ,
GC-1000 : 31A, 5.5SQ
Do not settle the equipment near to
places where gas can be leaked.
Do not use the equipment in places
where industrial oil, smoke, metal
dusts are existed
☞ That causes fire or electric shock
Off the power if the equipment has
strange sound and smell, smoke etc.
☞ Ask After Service.
It causes fire and electric shock.
v
◈ Setting
1
Page 11
Alert for safety
♣ Check the list as below during setting and operation.
♣ This is for keeping user and their property from damage.
Warning means that failure to follow this
safety statement could result in serious
personal injury, or death.
Do not put explosive or inflammable
materials into the chamber
☞ The equipment is not explosion-
proof structure. Sample can be
exploded in the chamber
Do not keep inflammable materials near
to the instrument.
☞ It can cause fire or explosion.
Do not depart, fix, and alter the
instrument by yourself.
☞ It can cause fire or electric shock
☞ It can cause dysfunction of the
instrument.
☞ Use only for the original purpose.
Do not put heavy things on the power
line. Do not put the equipment on the
line.
☞ It causes fire and electric shock.
Do not use the equipment near to places
where vibration occurs.
☞ It causes damage to equipment and is
possible that the equipment does not
works properly.
Do not let the product take any strong
shock or vibration.
☞ It causes abnormal operation or
trouble.
◈ Operating
1
Page 12
Notice for safety
♣ Check the list as below during setting and operation.
♣ This is for keeping user and their property from damage
Warning means that failure to follow this
safety statement could result in serious
personal injury, or death.
Do not clean the instrument with
solvent abstergent. Use smooth cloths.
☞ It can cause fire and deformity
Do not spread insecticide or
inflammable spray on the equipment.
☞ It causes damage to equipment ,
fire, electric shock as well as harms
to user.
Keep out of the direct sunlight.
☞ It may cause fire and abnormal
operation.
☞ When use in the direct sunlight,
please set up a proper interception
to make shade
Set the instrument apart from wet
materials.
☞ It can cause dysfunction of the
instrument and electric shock.
☞ Please ask After Service if the
instrument is wet.
Do not spread insecticide or
inflammable spray on the equipment.
☞ It causes damage to equipment ,
fire, electric shock as well as harms
to user.
Do not ask After Service to anyone who
is not from the manufacturer or its
entrusted party.
☞ It can cause dysfunction of the
instrument if the control is different
from Manufacturer's.
◈ Operating
1
Page 13
Notice for safety
♣ Check the list as below during setting and operation
♣ This is for keeping user and their property from damage
Warning means that failure to follow this
safety statement could result in serious
personal injury, or death.
▪ Setting place
Keep the equipment more than 1.5m from lighting and 20cm from the back. Make wide space for opening left and right side of the equipment. ☞ It can cause damage while changing fluorescent lamp.
Set the equipment on a plat place and do not heat or push down the equipment.
The equipment needs proper tool for
transportation.
■ Checking after settle down.
☞ Check whether the electric lines are at correct locations.
☞ Check whether the wire ground is correct
☞ Check whether there is spare place near the instrument. This is for setting
and after service of the instrument.
☞ Check the water level in the water tank and connect the tank to where it
should be located.
☞ Please operate the instrument after checking all the necessary part for
safety.
◈ Setting & operating
GC-300 : 250kg GC-1000 : 550kg
1
Page 14
Installation of the composition
2-1. Installing the GC-300 Water Tank
Following the bellow direction to install the water tank.
1. Setting the water tank prop on ②or ③ wherever you want with 6mm bolts.
2. Setting the water tank ① on where you want.
3. Connecting the water tank ① with the inlet ④ by using the silicon hose ⑤. (1.5M)
4. Pouring distilled water into the water tank ① and opening the valve.
(Using distilled water improves durability of the unit.)
5. Connecting two silicon hoses (0.5M) with part ⑥.
(On part ⑥, the right with a valve is a draining valve to drain water from the
inside of chamber. The left without a valve is a pipe for over flow.)
2
1
2
3
4
5
6
Page 15
2-2. Installing of the GC-1000 upper light
Following the below direction to install the upper light.
(Must take off the main plug in safety.)
1. Ready for the upper light first as the image <1>.
2. Put the upper light up on the upper side of unit as the image <2>, set the light
with bolts (4X10) that be set on the unit.
(Use a ladder with at least two people in safety.)
3. On the image <3>, open the upper cover (A) and put the lighting line in through the
hole (B).
4. As the image <4>, connecting the lighting line with (L1+, L1-, L2+, L2-, L3+, L3-,
FAN+, FAN-) from the left one in order. (blue= +, black= - )
5. Close the upper cover and open the lighting part <1> then set three metal lamps up.
<1>
<3> <4>
B A
C
<2>
Page 16
-1. General purposes
* Agriculture researches by using plant growth
* Plant tissue culturing
* Seeds germination researches
* Photosynthesis and light safety tests
* Growth tests for various animals and plants
* Preservation tests for food
* Genes and life science researches
* Increase, breeding, acclimation and photosynthesis tests of plants
* Marine plant, water microorganism, cultivation of insects and light safety tests
* Minute ocean currents cultivation
* Cultivation of sensitive microorganisms with heat and light
* Transmutation tests of plants
The equipment can make the optimal condition (temperature, humidity, intensity of
illumination) for the above various tests and researches like photosynthesis of plants
test.
3
Page 17
-2. Features
1. Accurate Temp. Control
▶ Perfect Temp. and heating control by PID Controlled Microprocessor Touch
Pad Type
2. Proper airflow for growing plants
▶ Air flowing from bottom to top by 2 fans (Top FAN(GC-300), bottom FAN(GC-
1000)
3. Controlling heat rising
▶ Minimizing heat rising by light and easy setting the condition of
illumination of day and night.
▶ Heat can be taken out through top part. (GC-1000)
4. Protect inner air from leaking out
▶ Outer part of chamber and door have been produced from Adiabatic material
and specially made silicon packing protects inner air from leaking out by
making close adherence between inner glass door and chamber.
5. Magnetic Packing
▶ Door with Magnetic Packing functions shock absorber and duplicate air
tightness.
6. Programming operation
▶ Ideal test condition by programming setting value by 10 grades for temp.
illumination, humid (GC-1000) . Repeatable test is possible till 999 Cycles.
7. Safety Circuit
▶ Over Temp. LED is on and power becomes off when the temp. is reached on Over
Temp. Limit.
3
Page 18
-2. Features
8. Corrosion-proof material
▶ Chamber material is made of Stainless Steel (SUS 304) and outer body is made
of Steel Plate with coating for protecting from corrosion.
9. Glass door for observation inside the chamber
▶ Tempered glass helps the door safe and durable. User can observe inside the
chamber without temp. change by opening only outer door.
10. Making perfect environmental condition
▶ Designed for maintaining stable condition inside the chamber
▶ Temp. illumination, humid. can be programmed by grades and illuminating
condition of day and night is controllable.
- GC-300 : Temp. illuminating program (Humid, CO2 OPTION)
- GC-1000 : Temp. illumination, humid program (CO2 OPTION)
3
Page 19
Name of each part
10 7
9 8
1 3
6
4
1. CONTROL PANEL
For inputting and checking setting
condition
2. Over temp. limit
For stopping heater when temp. is rising
over the set value unusually
3. Main Power Switch
On and off the main power switch.
4. Water Tank
For supplying water to control humidity.
5 Lamps
For supplying light into the chamber
6 Casters
Wheels For transporting and fixing the
equipment
7 Pair glass
Double glass for maintaining temp inside
the chamber
8 Filters
Separable filter can be easily cleaned.
(Clean filters once a week for maintaining
good condition of Compressor)
9 Shelf level adjustor
Shelves can be adjusted following to
sample size
10.Inner Door
User can observe inside the chamber
Tempered glass
11. Door
4-1. GC-300
4
11
2
5
Page 20
Name of each part
2
4
5
7 9
10
11
12
1. Recorder (OPTIONAL)
For recording Temp. and Humid. data
2. Control Panel
For inputting and checking setting condition
3. Over temp. limit
For stopping heater when temp. is rising over
the set value unusually.
4. Main Power Switch
On and off the main power switch
5. Lamps
For supplying light into the chamber
6. Casters
Wheels For transporting and fixing the
equipment
7. Pair glass
Double glasses for maintaining temp. inside
the chamber.
8. Door Limit Switch
To stop operating except for lamp when the
inner door is open.
9. Filters
Separable filter can be easily cleaned.
(Clean filters once a week for maintaining
good condition of Compressor)
10. Shelf level adjustor
Shelves can be adjusted following to sample
size.
11. Inner Door
User can observe inside the chamber Tempered
glass
12. Door
4-2. GC-1000
4
3
6
1
8
Page 21
Structure and feature of Lamp
1. Metal Halide Lamp
① Consuming power : 400W(per 1EA)
② Long life span, excellent light
efficiency.
③ Small heat loading capacity does
not affect samples. The instrument
is easy to be installed with small
maintenance cost.
④ Life span: 6000 hrs.
2. Fluorescent lamp
① High-grade light efficiency with
using economic-type electronic
safety device.
② Each lamp has it’ s own ON/Off
switch. - Illumination capacity can
be controlled step by step.
3. How to replace ramps.
① Open the top part by keeping the
handle.
② The lamp is very hot. Please wait
at least 10 min. after power-off.
③ When replacing fluorescent lamps,
turn it to clockwise rotation about
90°, and pull it to take off from
the socket.
④ Be careful when you replace the
lamps.
☞ High temp. “Burns" (Over 100℃)
Be careful when you replace the lamps. You can be
dead or have serious damage.
4-3. Lamp
(Showing Lamp part)
2
1
Page 22
Operating Controller
1. LCD display
For checking operating condition of Plant
Growth Chamber.
2. Condition display
1. LCD display
3. Controller buttons 4. Set value input
button
2. Condition display
5-1 Name and Operation
Showing temp. controlling heater operation.
Showing humid. Controlling heater operation
Showing CO2 supplying condition
5
Page 23
LAMP condition display
3. Controller operation button
4. Set value inputting button
Showing PROG/MANU operation mode
Showing Auto-tuning operation condition
Showing on/off condition of compressor
Showing whether or not Timer is used
Showing water level (Light is on and humi. Control
stops when water is lacked,)
START/STOP : START/STOP button
PROG/MANU : PROG/MANU MODE change button
MODE : Parameter setting button
ENTER/A-T: Set value changing button
☞ Auto-tuning starts when the button is put for 5 sec.
(In case of performing mode)
1~0 : Set value no. inputting button
◀ : Set value unit changing button
- : (-)mark inputting button
Page 24
Each set value is shown on SvT orderly.
SvT : Set temp. value
SvH : Set humid. value
SvC : Set CO2 density value
At First
You can see the message when
putting the instrument on.
Starting
Manual
Mode
Starting
Manual
Mode
.
Display shows the present condition of the
chamber when the instrument is on.
T : Present temp. in the chamber.
Co2 : CO2 Density
H : Present humid. In the chamber
Lux : Intensity of illumination in the
chamber
Tm : Remained time SvT : Set temp.
<Ref>
Manual Mode : 1 Step can be controlled
Program Mode : 10 Steps can be controlled
5-2 Controller operation
Page 25
1. Operation (start to control)
1. Put on the Main Power Switch on the right top part of the instrument.
2. Put on the Main Power Switch after putting on circuit breaker (This is for GC-1000)
1-1. Changing control mode
1-2. Performing screens for each of control mode.
Manual mode control screen
(Manual Mode) Screens
for
each
Control
Mode
Program mode control screen
The instrument works when you put the Start button
on in condition of Control stand-by
<Exchanging Program/manual mode>
PROG/MANU modes are exchanged by putting the button
for 5 sec.
ON
OFF
GC-300 GC-1000
Page 26
2. Operating manual mode
2-1. Putting set values in manual modes.
Exchanging
and
putting
Set values
Put the mode button on.
2-2. Setting control modes
<Ref>
T : Temp. input Co2 : CO2 input
H : Humid input
Tm : Time input F : Lamp input
(1 : ON, 0: OFF)
☞ Time, Temp. Humid. And CO2(GC-1000) values can be
set with and button or with Number
button.
☞ button is for moving of each unit.
☞ button is for moving of each item.
<Input set range>
Temp. : 0℃ ~ 50℃ Time : 00 : 00 ~ 99hr. 59min.
Humid. :50 ~ 90RH% CO2 density : 0 ~ 5000ppm(OPTION)
Values are returned to the present ones if they are not set within 1 min.
Page 27
2-3. Lamp setting
2-2. Set values are inputted following to Lamp diagram.
GC-300 LAMP diagram GC-1000 LAMP diagram
☞ Examples of Lamp setting(GC-300)
No. Lux(lx) Set Values Lamp Diagram
1 1,500 0000 010 2
2 3,500 0000 101 1,3
3 5,500 0000 111 1,2,3
4 6,000 0101 010 2,4,6
5 8,000 0101 101 1,3,4,6
6 10,000 1010 010 2,5,7
7 12,000 1010 101 1,3,5,7
8 14,000 1111 010 2,4,5,6,7
9 17,000 1111 101 1,3,4,5,6,7
10 20,000 1111 111 1,2,3,4,5,6,7
* Illuminating measurement- Values are measured from the middle of the chamber toward
topside.
<Lamp diagram>
1(No.8) 1(No.7) 1(No.6) 1(No.5) 1(No.4) 1(No.3) 1(No.2) 1(No.1)
☞ Location of ramps are shown orderly.
1(No.8) is not used with GC-300 and GC-1000
<Ref> 1=> Lamp On, 0 => Lamp Off.
Page 28
☞ Examples of Lamp setting(GC-1000)
No. Lux (lx) Set Values Lamp Diagram
1 5,000 0101 000 4, 6 or 5, 7
2 7,500 0000 010 2
3 9,500 1111 000 4, 5, 6, 7
4 12,500 0101 010 2/4, 6 or 5, 7
5 15,000 0000 101 1, 3
6 18,500 1111 010 2 / 4, 5, 6, 7
7 21,000 0101 101 1, 3 / 4, 6, or 5, 7
8 24,000 0000 111 1, 2, 3
9 28,000 1111 101 1, 3 / 4, 5, 6, 7
10 31,500 0101 111 1, 2, 3 / 4, 6 or 5,
7
11 35,000 1111 111 1, 2, 3 / 4, 5, 6, 7
* Illuminating measurement- Values are measured from the middle of the chamber toward
topside.
2-4. Operating Manual mode
2-3. Operate the instrument by putting “MODE"and“START"buttons after inputting set
values
1. Button “Mode” after inputting each control item
2. The instrument is operated by manual mode when “START"
key is buttoned.
Page 29
3. Example of performing manual mode
▶ The discription shows how to perform in Manual modes.
Please follow the orders as below
Main Power Switch should be ON.
1. Put MODE button
2. Input tmep, CO2, Humid. And time.
ex)T: 30.0℃, Co2: 600ppm, H: 60.5%, Tm : 01:00
3. F : Input No.1,3,5,7
ex) F: 1010101
(Using ◀, ENTER, and No. button )
4. Get out from the screen by putting MODE
button
Put START/STOP button.
.
The instrument works following to the inputted set
values.
T, Co2, H, and Lux on the display show Temp. Co2
density, Humid. and Illumination in the chamber.
Set values SvT, SvH, SvC are shown orderly.
● Put PROG/MANU button for 5 sec. if
you want to start with Program mode.
<Example of inputting set value>
This is how to operate the instrument by inputting fixed temp.
humid. Illumination, CO2 , and time as below.
ex) Temp.: 30.0℃, CO2: 600ppm, H: 60.5%, Time: 1hr
Illumination: 12000lx(Lamp No. 1,3,5,7)
Page 30
4. Operating Program Mode
4-1. Inputting set values in Program Mode
The mode is changed to Program Mode with beep sounds by putting
PROG/MANU button for 5 sec.
☞ Set values of program mode can be inputted while operation and
stop of the instrument.
(Referring to 4-4 parameter input)
Put MODE button
.
4-2. How to set program mode
<Ref>
T : Temp.input Co2 : CO2 input
H : Humid. input
Tm : Time input F : Lamp input
(1 : ON, 0: OFF)
Next STEP input screen is shown by putting button
Input values at Each STEP by the same way.
Inputting
Set value
☞ Time, Temp. Humid. And CO2(GC-1000) values can be set with
and button or with Number button.
☞ button is for moving of each unit.
☞ button is for moving of each item.
<Input set range>
Temp. : 0℃ ~ 50℃ Time : 00 : 00 ~ 99hr. 59min.
Humid. :50 ~ 90RH% CO2 density : 0 ~ 5000ppm(OPTION)
Values are returned to the present ones if they are not set within 1 min.
Lamp Input is the same as Manual mode (Please refer to 2-3 Lamp)
Page 31
4-3. Input and close of STEP
Maximum STEP nos. of the instrument is 10 ea.
If time is inputted as 00:00 at STEP 10, the steps before STEP 10 are inputted.
☞ Total 9 steps are inputted if time is inputted as 00:00 at STEP 10
Temp. set graph for each STEPs.
Page 32
4-4. Input parameter
Put MODE button after inputting “STEP"
The screen shows as below in case of inputting “STEP" value while operation
<Run Cycle No> Input Step repeating times.
Maximum: 999 times. Input “000" to repeat
infinitely.
<Lamp On Delay>
Lamp is on or off as indicated at the term
between steps. The unit is sec.
(**:This is for protecting plant growth from fast
change of illumination)
The screen shows as below when inputting STEP value after stopping operation.
<Pgm-On Delay>
This is shown when setting program on “Controller
Stop" condition.
The instrument works after the set time(Min.)
(Inputting“0000"doesn't works anything)
4-5. Performing Program Mode
1. In case of inputting “STEP" value while operation
1. Put “MODE" button after finishing each control item.
2. “Program mode" starts if you put “START" button.
Page 33
2. In case of inputting STEP value after stopping operation
4-6. Example of setting STEP input of Program Mode
If you want to repeat the step 60 times from 1 to 5
Example of setting to repeat 60 times
1. Input values with and or Nos. from step1 to Step 5.
2. Set values(Tm:) as 00:00 on Step 6.
3. “Run Cycle Number" mode is on if you put button.
Set “Run Cycle Number" as 60
4. Put button again
☞ Total Cycle(130) = STEP 5 ⅹRun Cycle Number(60)
4-7. Changing “Operation STEP" on Program Mode.
▶ This function is for changing present STEP while operating on program mode.
Select a STEP you want with putting ◀(SHIFT) button when you
want to change from program mode to present mode.
STEP 1 -> STEP 2 -> STEP 3 is selected repeatedly
when you put the SHIFT button
The instrument works automatically after the set time(min.) for
“Pgm-On Delay"if you put “MODE" button after inputting each
control item.
Step 1
Step 2
Step 3
Page 34
5. Example of operating Program Mode. 6.
▶ The discription shows how to perform in Program mode.
Please follow the orders as below
Put on Main Power Switch
1. Put “MODE" button
2. Input Temp. CO2, Humid. and Time.
ex) T: 60.0℃, Co2: 500ppm, H: 60.5%, Tm : 01:00
3. F : Input nos. 1,2,3,4,6
ex) F: 00101111
(Use ◀, ENTER, or No. button)
Next 5 step input screen is shown if you put
“MODE" button.
1. input Temp. CO2, Humid. and time.
ex) T: 30.0℃, Co2: 400ppm, H: 80.5%, Tm : 01:00
2. F : Input No.1,2,3
ex) F: 00000111
3. If second STEP input is done, input values of
next STEP by buttoning “MODE'
4. Input set values by 10 steps with the sam way.
● Put PROG/MANU button for 5 sec. if the
screen starts with manual mode
<Example of inputting set value>
This is how to operate the instrument by inputting fixed
temp. humid. Illumination, CO2 , and time as below.
<STEP 1>
Temp.: 30.0℃, CO2: 600ppm, H: 60.5%, Time: 1hr
Illumination: 14000lx(Lamp No. 1,2,3,4,6)
<STEP 2>
Temp: 30.0℃, CO2: 400ppm, H: 80.5%, Time: 1hr.
Illumination: 12000lx(Lamp No. 1,2,3)
Page 35
Parameter input screen is shown if STEP input is done
till 10 steps or input the time of STEP as 00:00
1. This is for setting repeating time of each
programmed STEP. Maximum 999 times can be
inputted. STEPs are repeated infinitely if you
input as “000"
ex) To set 10 STEP , input Run Cycle No as 005
10 STEP ⅹ 005 = 50 (Total Cycle)
2. Input Lamp-On delay time
ex) 60s : 60sec
3. Input Pgm-On delay time
ex) 0005 : 5min(The instrument works 5min. later)
4. Move to stand-by condition by putting
Button.
Put button.
This is operated automatically if you input Pgm-On
delay
T, Co2, H, Lux on the display shows present temp.
CO2 density, humid. and illumination of the
chamber.
The set values SvT, SvH, SvC are shown by turns.
The present STEP and cycle are shown by turns on
the display
Page 36
6. Operation and setting of timer. 6-1. Operating theory of timer
Timer works automatically by the set value when you put “START"
button.
1. Timer works when it satisfies some cases as below.
* If present temp. is under set value. (Set values(Tm:) as 01:00)
* If present temp is over set temp. (Set values(Tm:) as 01:00)
Page 37
6-2. Setting timer
<Setting range>
▶ 00:00 – 99:59(Hr:Min.)
▶ Timer does not work if “Tm:00:00" is shown.
“Tm:**:**"is indicated on the screen.
▶ Timer signal lamp is on when the time is over.
“Time-out"is shown on the screen.
▶ Every control output and relay is off when time is
over.
▶ Put “START/STOP" button to get out from “Time-
out"
● Timer does not work if a temp. is not reached to set temp.
In case of this, the Timer lamp is flickering
● Timer works when the present temp. is over the set temp.
In case of this, the Timer lamp is on and “:"part of time
signal is flickering.
Page 38
7. Operating “AUTO TUNING"
7-1. Start to “AUTO TUNING"
Operating mode Waiting mode
“Auto-tuning" selected
Screen. Auto-
tuning
LED
screen
Example of Temp. auto-tuning.
Please put “AT" button for 5 sec in START condition, then you
can see selectable screens for Auto-tuning
Auto-tuning is only done in Operating mode
1. Select a control item for “Auto-tuning" with Enter button.
2. “Auto-tuning" is operated by putting “MODE" button and
“Auto-tune" signal lamp is flickering.
Condition of AUTO-TUNE signal lamp
Page 39
7-2. Example of Temp. AUTO TUNING
7-3. Example of Humid. AUTO TUNING
If you move cursor to Humidity
by putting Enter button, the
screen shows Humidity (Y) and
others (N)
“Auto-tuning" is operated if
you put MODE button and AT-TUNE
lamp is on.
If you move cursor to TEMP by
putting Enter button, the screen
shows Temp(Y) and others (N).
“Auto-tuning" is operated if
you put MODE button and AT-TUNE
lamp is on.
Page 40
How to solve problems(check before ask)
Symptoms Confirmation Arrangement
Check the main power
Operate after checking
whether it's rated
voltage.
Check whether the power plug
is disconnected from an
outlet
put a power plug in an
outlet after turn off
the main power switch.
Check whether Run LED on
Display is blowing out.
Operate by pushing
START/STOP button.
If equipment does not
operate
Check weather it is power
failure.
Operate the equipment
after powering it on.
Please check if an obstacle
is in an intake ventilation
Remove obstacles from
the intake.
If temperature does not
go down
Please confirm whether there
is too much dirt in Filter.
Clean the Condenser Fin
And Filter.
The machine becomes off
even though turning on
the power switch.
Leakage
Stop operating and
contact with purchase
store
Check a power plug is
disconnected from an outlet
correctly
Turn off the power
switch and connect the
plug into the outlet. Display does not show
anything even though the
power switch gets on
Check whether the power
switch is ok.
Ask a purchase store
for After Service.
6
Page 41
Daily Checking
Section Daily
check
Weekly
check
Monthly
check
Yearly
check Note
Cleanliness
state of CASE ○ Condition Check
State of a
fluorescent
Lamp ○ Condition Check
State of an
illuminometer ○
CO2
Regulator ○
Air Filter
Cleaning ○
Connecting
condition ○
Noise and
Vibration ○
Controller
Check ○ Condition Check
① Selects a manager, and please check very often and
don't make other person to operate except for the
person in charge of operating.
② Don't repair product in the state that connected the
power. This makes the equipment to be broken earlier
7
Page 42
How to clean
◆ To clean appearance.
1. Turn off main power switch of the
equipment.
2. Pulls out a power plug of a
equipment and turn off the power.
3. Wipe the equipment with sponge or
the soft cloth that wetted a
neutral detergent.
4. Wipe the equipment with cloth in
order to remove a neutral
detergent.
5. Wipe with dry cloth.
◆ To clean air filter
1. If Filter is polluted, pneumatic
circulation decreases and obstructs
normal operation of the equipment.
2. Wash Air Filter according to a degree
of pollution with a vacuum cleaner or
water, and dry it at the shade where
ventilation is fluent in after washing
* Don’ t use insecticide or an
organic solvent when you clean the
equipment
GC-300
GC-1000
8
Page 43
How to clean
◆ To clean internal chamber
1. Turn off main power switch of the
equipment.
2. Take off inner Shelves and a
Shelves hanger of the equipment.
3. Wipe the inside with soft cloth
wetted with a neutral detergent.
4. Wipe with wet cloth in order to
remove a neutral detergent
5. Wipe with dry cloth.
◆ To replace lamp
1. Check the condition of a door
fluorescent lamp periodically.
2. Replace it after turn off the power
of the equipment to avoid electric
shock.
◆ To store equipment for a long
time without operation
1. Off the power by unplugging the
equipment.
2. Wipe with soft cloth cleanly.
3. Please keep it after packing.
8
Page 44
Water Tank Usage
1. How to use GC-300 Water Tank
Fill water into the water tank following to the
orders as below.
1. Take out the Water Tank from the supporter.
2. Open the cover and fill water into the
Water Tank.
3. Keep the water tank back to the supporter.
1. Water Tank : for controlling Humidity
by supplying water in the chamber.
(Capacity: 10L.)
2. Supporter: To keep the water tank on the body.
1
2
(Water Tanks are indicated by
orange colored line)
(Water Tanks can be set by 2 kinds shown)
User can select the position of water tank
as the pictures.
If the water tank lacks of water, lamp is
getting on and the humidity controlling is
stopped operating.
Use a ladder to move the water tank from
the supporter.
- Name and function of water tank
9
Page 45
2. GC-1000 Water Tank Usage
If the water tank lacks of water, lamp is getting on and the humidity controlling is stopped operating.
Please fill water in Water Tank according to the
following procedures.
1. Open water inlet of the bottom in front of the
equipment. (Pull the handle of Water inlet
forward)
2. Fill water by using funnel or kettle
3. Push water tank inlet back to the original
position.
1. Waterspout: Condensation steam is supplied to
Water Tank again.
2. Water Inlet : Please fill water in Water Tank
according to the following procedures..
(Total Capacity of Water Tank is 20L.)
3. Water level gauge : It shows water level in
the Water Tank.
2
(Water Tank part.)
1
- Water Tank names and function
3
If the water tank lacks of water, lamp is
getting on and the humidity controlling is
Page 46
Specification
Model GC-300TL GC-300TLH GC-1000TLH
Chamber Volume 300L 1000L
Permissible
Environmental Condition
Temperature 5℃ to 40℃
Maximum Relative Humidity 80%
Altitude up to 2,000m
Range +5℃ ~ 50℃(Lamp off), +10℃ ~ 50℃(Lamp on),
+20℃ ~ 50℃(with humidity)
Fluctuation ±0.1℃ at 25℃(without humidity) Temperature
Variation ±0.8℃ at 25℃(without humidity)
Range 0~20,000Lux 0~35,000Lux
Controls 10 Step 11 Step Illumination
Lamp FL 32W×18EA FL 32W×16EA
METAL 400W×3EA
40~80RH%(at20~35℃) 50 ~ 90RH% (at 20~35℃)Range -
70~90RH%(at36~50℃) 60 ~ 90RH% (at 36~50℃)Humidity
Fluctuation - ±3% at 60%RH
Chamber Stainless steel, 0.8t Stainless steel, 1.0t
Body Steel 1.2t, powder coating Steel 1.6t, powder
coating
Shelves Stainless steel wire, electro polished
Material
Inner door Tempered safety glass
Range 300 to 5,000ppm CO2(Optional)
Sensor NDIR CO2 sensor
Controller Microprocessor PID controller
Safety device Power Breaker, Level monitor, Door limit
Over temp. limit Hydraulic over temp. limit
Refrigerator 1/2HP 3/4HP
Temp. Heater Fin type 750W ⅹ 2EA Fin type 1500Wⅹ 2EA
Humid. Heater - 1300W 1500W
Internal(mm) 510ⅹ540ⅹ1100 1200ⅹ800ⅹ1080 Dimensions
(WⅹDⅹH) External(mm) 700ⅹ805ⅹ1900 1410ⅹ1070ⅹ2150
Electric Requirements AC 230V, 50/60Hz
Power Consumption 12.5A 16.0A 30.0A
Weight(net) 250Kg 550kg
* The above specification can be changed by quality of a product and a performance
elevation.
10
Page 47
Warranty
1. Service under Warranty period
If trouble occurs during product use, user can get free service for 2 year
limited warranty from the date of purchase.
2. Exceptional cases even during under warranty period
User can not be credited by warranty in case of as below.
① If trouble occurs by an act of God.
② If the equipment breaks down due to misusing of available voltage.
③ If damage occurs by dropping a product, or impact.
④ If damage occurs in an appearance by organic solvents such as thinner, benzene.
⑤ If damage occurs without following to notice in the manual
⑥ If damage occurs by fixing the equipment by any person who is not related
with Jeio tech.
⑦ If damage occurs by a mistake of a customer
How to waste products
Check the notice as below when you are going to waste the equipment
How to waste each model
Element Model Weight (kg) Dimension (mm) How to waste
Main Body GC-300TL 250 kg 700ⅹ805 19ⅹ 00
Main Body GC-300TLH 250 kg 700 805 19ⅹ ⅹ 00
Main Body GC-1000TLH 550 kg 1410 1ⅹ 070 2150ⅹ
Connect with a company that
is dealing with waste
matter.
11