Page 1: FUNDUS MODULE 300 Quick guide – Image capture step-by ......FUNDUS MODULE 300 Quick guide – Image capture step-by-step Image capture step-by-step Preparing slit lamp Move microscope

Tradition and Innovation – Since 1858, visionary thinking and a fascination with technology have guided us to develop innovative products of outstanding reliability: Anticipating trends to improve the quality of life.

FUNDUS MODULE 300Quick guide – Image capture step-by-step

Page 2: FUNDUS MODULE 300 Quick guide – Image capture step-by ......FUNDUS MODULE 300 Quick guide – Image capture step-by-step Image capture step-by-step Preparing slit lamp Move microscope

Image capture step-by-stepPreparing slit lamp

■ Move microscope and illumination tower to the right until stop position (for BM 900 set to 60°)

Installing Fundus Module 300 ■ Lift Fundus Module 300 from cradle (if Fundus Module 300 has been turned off, press ON button for starting)

■ Latch Fundus Module 300 to base plate on microscope

Preparing Fundus Module 300 for capture

■ Move Fundus Module 300 forward until stop position ■ Select ‘manual mode’ and enter patient’s refractive error

Start capture ■ Start examination AFTER Windows recognizes Fundus Module 300 in EyeSuite (wait for Windows connection sound or pop-up)

■ Move Fundus Module 300 towards patients eye ■ Centre retina and move in until the picture detail of IR image shows section of interest

Capturing ■ Press release bar on control panel or capture button on Fundus Module 300

■ Captured image is shown briefly then transmitted to EyeSuite

Page 3: FUNDUS MODULE 300 Quick guide – Image capture step-by ......FUNDUS MODULE 300 Quick guide – Image capture step-by-step Image capture step-by-step Preparing slit lamp Move microscope

How to avoid image artefacts Disturbing re-flexes in lower part

Reason: Camera is not close enough

Action: Move camera slightly closer to patient’s eye

Image is out of focus

Reason: Camera focus was not on retina

Action: Enter patient’s refractive error to bring focus on retina

Disturbing re-flexes in upper part

Reason: Camera is too close

Action: Move camera slightly fur-ther from patient’s eye

Picture detail incomplete

Reason: Patient’s pupil was too small

Action: Darken room further or if necessary dilate patient’s eye

Image too dark

Reason: Flash intensity is too low

Action: Increase flash intensity

Poor detail despite sharp image

Reason: Image was captured through hazy media

Action: Best im-ages can only be achieved through relatively clear media

Image too bright

Reason: Flash intensity is too high

Action: Reduce flash intensity

Ideal image

Page 4: FUNDUS MODULE 300 Quick guide – Image capture step-by ......FUNDUS MODULE 300 Quick guide – Image capture step-by-step Image capture step-by-step Preparing slit lamp Move microscope

HS-Art.No. 1504.7020040.02020

©HAAG-STREIT AG, 3098 Koeniz, Switzerland2. Edition / 2019 – 01

HAAG-STREIT Holding AGwww.haag-streit-holding.com

HAAG-STREIT AG, Diagnosticswww.haag-streit.com

HAAG-STREIT AG, Verkauf Schweizwww.haag-streit.ch

HAAG-STREIT Deutschland GmbHwww.haag-streit.de

HAAG-STREIT Far Eastwww.haag-streit-fareast.com

HAAG-STREIT Medtech AGwww.haag-streit-medtech.com

HAAG-STREIT Surgical GmbHwww.haag-streit-surgical.com

HAAG-STREIT UKwww.haag-streit-uk.com

HAAG-STREIT USAwww.haag-streit-usa.com

CLEMENT CLARKE Ltd.www.clement-clarke.com

HS DOMS GmbHwww.hs-doms.com

IPRO GmbHwww.ipro.de

John Weiss Ltd.www.johnweiss.com

Möller-Wedel GmbH & Co KGwww.haag-streit-surgical.com

Möller-Wedel Optical GmbHwww.moeller-wedel-optical.com

OptoMedical Technologies GmbHwww.haag-streit-surgical.com

Reliance Medical Inc.www.haag-streit-usa.com

SPECTROS AGwww.spectros.ch

Members of HAAG-STREIT Group

HAAG-STREIT AGGartenstadtstrasse 103098 KoenizSwitzerlandPhone +41 31 978 01 11Fax +41 31 978 02 [email protected]

Distributed by:

Manufactured by:Optomed OyYrttipellontie 190230 Oulu, Finland

0598