From Theory to Reality - MDRAO · 2018. 5. 26. · • Place endoscope in an automated re-processor...
Transcript of From Theory to Reality - MDRAO · 2018. 5. 26. · • Place endoscope in an automated re-processor...
-
From Theory to Reality
-
Objectives
To understand the mechanics of an endoscope
To learn proper reprocessing methods for flexible endoscopes
To understand the day to day challenges with reprocessing endoscopes
-
An instrument/medical device designed to observe, diagnose and perform therapeutic procedures.
-
Anatomy of an Endoscope
-
Body Mechanics Control Section
• Angulation control knobs
• Free/engage lever
• Suction and air/water valves
• Switch unit (buttons)
• Biopsy port
• Stiffness control mechanism (colonoscopes)
• Auxiliary water port
• Special channel control (duodenoscopes)
-
Body Mechanics
Insertion Tube
• Includes the light guide tube boot
• Bending rubber
• Distal Tip
Umbilical Cable/Control Head
• Includes the light guide fibers
• Suction port
• Electrical connector
Insertion Tube
Umbilical Cable
Distal Tip
-
Suction Channel Flow
-
Air/Water Channel Flow
Air
Water
-
Auxiliary Channel Flow
-
• Pre-clean at the bedside
• Dry/Wet leakage testing and visual inspection
• Manual Cleaning
• Sterilization/High level Disinfection (Automated or Manual)
• Drying
• Alcohol Flush
• Storage
-
Bedside Pre-Cleaning
• Performed immediately after the procedure • The first step in removing organic and microbial bioburden
• Mandatory prior to automated or manual re-processing
-
Wipe the outside of the insertion tube with a clean, lint free cloth soaked in enzymatic
solution.
-
Cover Air/Water button hole to check for air flow. No air flow indicates a blockage. Inform your scope re-processor upon
delivery/pick up.
-
Press Air/Water button to check for water flow. No water flow indicates a blockage.
Inform your scope re-processor upon delivery/pickup.
-
Press suction button to test suction.
-
Alternate between enzymatic solution and air aspirating until flow is clear.
-
Day to Day Challenges
-
No Bedside Cleaning Bedside Cleaning
-
Dried Bioburden
-
• Endoscopes are to be transported to the re-processing room immediately following bed side cleaning
• One endoscope per bin
• Accessories are to be separated from the endoscope during transportation to reduce the risk of damage
-
Day to Day Challenges
Sharp Instruments such as Forceps could potentially damage the
scope during transport
-
Leakage Test
• Must be performed before each cleaning
• Turn on pump and check that air is emitting from it
• Attach tester outside of water
• Remove all detachable parts
• Perform dry leak test
• If leak is detected, proceed to clean the endoscope under pressure adhering to the regular steps in reprocessing if possible.
-
Visually check the insertion tube. Note any dents, bulges, or other irregularities.
Crushed Insertion Tube
-
Test the brakes to ensure they are not engaged.
-
Turn the angulation controls and release. If the brakes are not engaged,
the bending section will return to normal position.
-
• Fill sink with water
• Confirm leak tester is connected
• Completely immerse endoscope
• Perform wet leakage test
• Add enzymatic solution
• Brush all valves/buttons with valve brush
• Brush all channels (ie: Biopsy channel, Suction channel)
• With each passage, check the brush for visible bioburden before withdrawing through the scope
-
Wet Leakage Test
Valves/Buttons Removed
Leakage Test Connector
-
Brush the buttons in enzymatic solution paying close attention to all areas. Compress the spring loaded top
of both buttons for further cleaning.
Bioburden in Biopsy valve cap
-
Brush Suction Channel at 45 degree until brush appears clean.
-
Brush the Suction Channel at 90 degree until brush appears clean
-
Brush through the Biopsy Channel until brush appears clean.
-
• Attach all cleaning adapters including auxiliary water port adapter on certain scopes (ie: colonoscopes, duodenoscopes)
• Irrigate all channels with enzymatic solution and leave inside the scope for the manufacturer’s recommended time (usually 2-3 min.)
• Rinse all channels and valves/buttons with water
-
Cleaning Adapters (Manual)
Channel Plugs
Enzymatic Solution Connector
-
Cleaning Adapters (Automated)
Enzymatic Solution Connector
Channel Plugs
-
• The brush tip must be smooth and inspected before each use
• Reusable brushes must be disinfected with each endoscope
• Only use brushes recommended for channel cleaning
-
• Place endoscope in an automated re-processor or basin for manual disinfection
• Connect proper adapters to all channels
• For manual disinfection, the endoscope must be in contact with the disinfectant for the manufacturer’s recommended time
• After soaking, the endoscope and channels are then rinsed with bacteria-free water
-
Manual Process for High Level Disinfection
• High Level Disinfectant • Flush all channels
-
Rinsing after High Level Disinfection
• Bacteria free water • Rinse all channels
-
Endoscope placed in an Automated Re-processor
-
Day to Day Challenges
Crushed Umbilical Cable
Umbilical Cable damaged due to
Improper Loading
-
Missing Connector
-
Missing Sterilant
Missing Chemical Indicator
-
• Endoscopes must be dried after re-processing and before storing at the end on the day
• Purge endoscope channels with medical air
• Flush all channels with alcohol to assist in drying
• Purge endoscope channels with medical air
• Dry exterior of the endoscope with a clean, lint free cloth
• Dry buttons, do not put back into the endoscope during storage
-
STORAGE • Store in a well ventilated cabinet, free
of dust and in close proximity to the procedure rooms
• Remove all detachable parts including water proof cap and buttons
• Attach EtO/venting cap to open all internal channels
• Scopes must be hung vertically with
the control head supported on a smooth bracket
• Distal tips hanging freely
-
Day to Day Challenges
* Arrows indicate missing EtO/venting caps*
-
*Arrows indicate buttons and biopsy valve cap left
in during storage*
-
Bending Rubber 43%
Biopsy Channel 16%
Angulation 11%
Air/Water Repair 8%
CCD Replacement 7%
Bending Channel 7%
Suction Cylinder 2%
Light Guide Lens 2%
Adjust Forcep Elevator 2%
Repair Elevator Channel 1% Replace Light Guide
Bundle 1%
Damage Analysis Olympus Canada 2008