From Theory to Reality - MDRAO · 2018. 5. 26. · • Place endoscope in an automated re-processor...

52
From Theory to Reality

Transcript of From Theory to Reality - MDRAO · 2018. 5. 26. · • Place endoscope in an automated re-processor...

  • From Theory to Reality

  • Objectives

    To understand the mechanics of an endoscope

    To learn proper reprocessing methods for flexible endoscopes

    To understand the day to day challenges with reprocessing endoscopes

  • An instrument/medical device designed to observe, diagnose and perform therapeutic procedures.

  • Anatomy of an Endoscope

  • Body Mechanics Control Section

    • Angulation control knobs

    • Free/engage lever

    • Suction and air/water valves

    • Switch unit (buttons)

    • Biopsy port

    • Stiffness control mechanism (colonoscopes)

    • Auxiliary water port

    • Special channel control (duodenoscopes)

  • Body Mechanics

    Insertion Tube

    • Includes the light guide tube boot

    • Bending rubber

    • Distal Tip

    Umbilical Cable/Control Head

    • Includes the light guide fibers

    • Suction port

    • Electrical connector

    Insertion Tube

    Umbilical Cable

    Distal Tip

  • Suction Channel Flow

  • Air/Water Channel Flow

    Air

    Water

  • Auxiliary Channel Flow

  • • Pre-clean at the bedside

    • Dry/Wet leakage testing and visual inspection

    • Manual Cleaning

    • Sterilization/High level Disinfection (Automated or Manual)

    • Drying

    • Alcohol Flush

    • Storage

  • Bedside Pre-Cleaning

    • Performed immediately after the procedure • The first step in removing organic and microbial bioburden

    • Mandatory prior to automated or manual re-processing

  • Wipe the outside of the insertion tube with a clean, lint free cloth soaked in enzymatic

    solution.

  • Cover Air/Water button hole to check for air flow. No air flow indicates a blockage. Inform your scope re-processor upon

    delivery/pick up.

  • Press Air/Water button to check for water flow. No water flow indicates a blockage.

    Inform your scope re-processor upon delivery/pickup.

  • Press suction button to test suction.

  • Alternate between enzymatic solution and air aspirating until flow is clear.

  • Day to Day Challenges

  • No Bedside Cleaning Bedside Cleaning

  • Dried Bioburden

  • • Endoscopes are to be transported to the re-processing room immediately following bed side cleaning

    • One endoscope per bin

    • Accessories are to be separated from the endoscope during transportation to reduce the risk of damage

  • Day to Day Challenges

    Sharp Instruments such as Forceps could potentially damage the

    scope during transport

  • Leakage Test

    • Must be performed before each cleaning

    • Turn on pump and check that air is emitting from it

    • Attach tester outside of water

    • Remove all detachable parts

    • Perform dry leak test

    • If leak is detected, proceed to clean the endoscope under pressure adhering to the regular steps in reprocessing if possible.

  • Visually check the insertion tube. Note any dents, bulges, or other irregularities.

    Crushed Insertion Tube

  • Test the brakes to ensure they are not engaged.

  • Turn the angulation controls and release. If the brakes are not engaged,

    the bending section will return to normal position.

  • • Fill sink with water

    • Confirm leak tester is connected

    • Completely immerse endoscope

    • Perform wet leakage test

    • Add enzymatic solution

    • Brush all valves/buttons with valve brush

    • Brush all channels (ie: Biopsy channel, Suction channel)

    • With each passage, check the brush for visible bioburden before withdrawing through the scope

  • Wet Leakage Test

    Valves/Buttons Removed

    Leakage Test Connector

  • Brush the buttons in enzymatic solution paying close attention to all areas. Compress the spring loaded top

    of both buttons for further cleaning.

    Bioburden in Biopsy valve cap

  • Brush Suction Channel at 45 degree until brush appears clean.

  • Brush the Suction Channel at 90 degree until brush appears clean

  • Brush through the Biopsy Channel until brush appears clean.

  • • Attach all cleaning adapters including auxiliary water port adapter on certain scopes (ie: colonoscopes, duodenoscopes)

    • Irrigate all channels with enzymatic solution and leave inside the scope for the manufacturer’s recommended time (usually 2-3 min.)

    • Rinse all channels and valves/buttons with water

  • Cleaning Adapters (Manual)

    Channel Plugs

    Enzymatic Solution Connector

  • Cleaning Adapters (Automated)

    Enzymatic Solution Connector

    Channel Plugs

  • • The brush tip must be smooth and inspected before each use

    • Reusable brushes must be disinfected with each endoscope

    • Only use brushes recommended for channel cleaning

  • • Place endoscope in an automated re-processor or basin for manual disinfection

    • Connect proper adapters to all channels

    • For manual disinfection, the endoscope must be in contact with the disinfectant for the manufacturer’s recommended time

    • After soaking, the endoscope and channels are then rinsed with bacteria-free water

  • Manual Process for High Level Disinfection

    • High Level Disinfectant • Flush all channels

  • Rinsing after High Level Disinfection

    • Bacteria free water • Rinse all channels

  • Endoscope placed in an Automated Re-processor

  • Day to Day Challenges

    Crushed Umbilical Cable

    Umbilical Cable damaged due to

    Improper Loading

  • Missing Connector

  • Missing Sterilant

    Missing Chemical Indicator

  • • Endoscopes must be dried after re-processing and before storing at the end on the day

    • Purge endoscope channels with medical air

    • Flush all channels with alcohol to assist in drying

    • Purge endoscope channels with medical air

    • Dry exterior of the endoscope with a clean, lint free cloth

    • Dry buttons, do not put back into the endoscope during storage

  • STORAGE • Store in a well ventilated cabinet, free

    of dust and in close proximity to the procedure rooms

    • Remove all detachable parts including water proof cap and buttons

    • Attach EtO/venting cap to open all internal channels

    • Scopes must be hung vertically with

    the control head supported on a smooth bracket

    • Distal tips hanging freely

  • Day to Day Challenges

    * Arrows indicate missing EtO/venting caps*

  • *Arrows indicate buttons and biopsy valve cap left

    in during storage*

  • Bending Rubber 43%

    Biopsy Channel 16%

    Angulation 11%

    Air/Water Repair 8%

    CCD Replacement 7%

    Bending Channel 7%

    Suction Cylinder 2%

    Light Guide Lens 2%

    Adjust Forcep Elevator 2%

    Repair Elevator Channel 1% Replace Light Guide

    Bundle 1%

    Damage Analysis Olympus Canada 2008