Forced Oscillations Technique, the reference method … · The Quark i2m is a forced oscillation...
Transcript of Forced Oscillations Technique, the reference method … · The Quark i2m is a forced oscillation...
The Quark i2m is a forced oscillation technique (FOT) system for the measurement of the me-chanical properties of the respiratory system, total respiratory input impedance (Zrs), under tidal breathing conditions. FOT employs small-amplitude pressure oscillations superimposed on the normal breathing and, therefore, has the advantage of not requiring the perfor-mance of complex respiratory manoeuvres.
The FOT test is simple and fast (8 seconds) and it is performed with the patient breathing nor-mally into the measuring device., making the test ideal in uncooperative patients or those unable to perform forced expiratory manoeu-vres, such as children or the elderly. As the FOT is noninvasive, it may be used for routine testing, epidemiological studies, research and monitoring the efficacy of treatment in COPD, asthma and muscoviscidosis. Novel FOT appli-cations include tests during mechanical venti-lation and sleep.
Quark i2m superimposes a random signal to the patient’s breath and measure the respira-tory apparatus response. This signal is gen-erated by an acoustic source connected to a mouthpiece with a pressure transducer, where the patient breaths at rest. The acoustic pulses are sent to the mouth at high frequency in ca-sual sequence (Pseudo Random Noise - PRN).
Quark i2m measures pressure and flow at the subject’s mouth while an imperceptible, low intense and high frequency (4-48 Hz) pressure signal is applied. Elaborating flow and pres-sure signals, Quark i2m provides respiratory impedance measurement (Zrs) and its two components: resistance (R) and reactance (X).
Quark i2m can be used either as a standalone system or as a module fully integrated in a COSMED Quark PFT system. A special rack can be added to the Quark PFT medical graded cart, allowing thus the operator to manage, with one single PC program, tests of respira-tory mechanics, diffusing lung capacity, static and dynamic lung volumes.
®® Total Respiratory Impedance measurement by Pseudo Random Noise Signal
®® No cooperation required from the patient
®® Ideal for assessing lung function in pre-school children
®® Quick and easy assessment (8 seconds tidal breathing)
®® Accurate, reliable and reproducible technique
Forced Oscillations Technique, the reference method for respiratory mechanics in children
Forced Oscillations
Respiratory care, when breath does matter
© 20
14/0
9 COS
MED
E
& OE
. Sub
ject t
o alte
ratio
ns w
ithou
t prio
r not
ice.
REF
C090
10-0
2-93
COSMED Srl
Via dei Piani di Monte Savello 37 Albano Laziale - Rome 00041, Italy
+39 (06) 931-5492 Phone +39 (06) 931-4580 Fax
[email protected] | cosmed.com
Validation articles• Beyon et al. “An Official American Thoracic Society/Euro-
pean Respiratory Society Statement: Pulmonary Function Testing in Preschool Children” Am J Respir Crit Care Med Vol 175. pp 1304–1345, 2007
• E. Oostveen, et al. “The forced oscillation technique in clinical practice: methodology, recommendations and future developments ERS Task Force on Respiratory Im-pedance Measurements” Eur Respir J 2003; 22: 1026-1041 (Ref1)
• J. Hellinckx, et al “Evaluation of impulse oscillation sys-tem: comparison with forced oscillation technique and body plethysmography” Eur Respir J 2001 ; 18 : 564-570.
• More scientific studies on www.cosmed.com/bibliography
Technical Specifications
To know more:
Product Description REFQuark i2m Forced Oscillation Technique (FOT) Laboratory C09010-01-99Standard packaging Quark i2m unit, Quark i2m head, arm support for Quark trolley, calibration unit, antibacterial
filters (20 pcs), nose clips (2 pcs), Irubber mouthpieces (20 pcs), RS232 serial communication cable (2 pcs), PC software, user manual.
Standard TestsForced Oscillation Technique Total Respiratory System Impedance (Zrs), Resistance (Rrs) in hPa/(l/s) & Reactance (Xrs) in
hPa/(l/s), Resonance frequency, frequency dependency, average Rrs and Xrs, discriminant function
FOT TechnologyType Optimized pseudorandom noise (PRN) between 4Hz and 48HzMeasurement range 0.5 - 40 hPa/(L/s)Frequencies between 4 and 48Hz (frequency step 2 Hz)Accuracy ± 2%Coherence 0-1, (Index = 0,95)Processing time 8 secPressure TransducerType Piezo-resistiveRange 0-1 psiHardwareDimensions & Weight 33 x 20 x 16 cm / 6.5 KgInterface ports RS232Electrical requirements 100-240V ± 10% 50/60 HzEnvironmental conditions Temperature 0-50 °C (32 - 122 °F); Barometer 400-800 mmHg; Humidity 0-100%Software PFT SuiteAvailable languages Italian, English, German, Spanish, French, Portuguese.PC Configuration Pentium or faster, Windows XP, VISTA (32 bit), Windows 7 (32 bit), 128 Mb RAM or more, USB
or RS 232, CD-Rom reader, 80 Mb on HD space available.Optional Module Description REFPulse Oximetry Oximeter (Xpod) requires probe
Oximeter ipod (w/ finger probe)C02600-01-05 C02390-01-05
Accessories Description REFLarge Medical Cart 3 cylinder holder (230 or 120 VAC) C02900-01-04 (230)
C02900-02-04 (120)Table arm support flexible arm for holding Quark i2m head C02870-01-05Bracket extension Special rack to be added to the medical graded cart, allowing
thus the operator to fully integrate FOT tests with other Quark PFT tests
C03660-01-11
Safety & Quality StandardsMDD (93/42 EEC); EN 60601-1 (safety) / EN 60601-1-2 (EMC)