Food Analysis Lecture 24 (04/23/2013)
GC (2)
Qingrong Huang
Department of Food Science
![Page 2: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/2.jpg)
Sample Injection
• Direct Injection•Sample Derivatization
![Page 3: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/3.jpg)
Column
• Either packed or capillary column• Packaging Materials: - Packed column: silane-treated diatomaceous earth - Capillary column: hollow fused silica glass• Stationary phase: - polysiloxane-based (-Si-O-Si-) -General rules: choose polar phase to separate polar compounds and phenyl-based column to separate aromatic compounds.
![Page 4: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/4.jpg)
Detector
• Thermal Conductivity Detector (TCD)• Flame Ionization Detector (FID)• Electron Capture (ECD)• Flame Photometric Detector (FPD)• Photoionization Detector (PID)
![Page 5: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/5.jpg)
Carrier Gas
• In capillary chromatography, A term is small;• N2 is the most efficient carrier gas, but its minimum occur in low u;• Hydrogen is a better choice than He and N2
![Page 6: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/6.jpg)
SPME
• Microextraction technique which employs a thin film of sorptive polymer on a glass fiber
• Fiber
Having varying film thickness of polymer
Coatings are based on both absoption and adsorption mechanisms
• Convenient, solventless extraction technique• Used to extract analytes from both liquid and solid
matrices
![Page 7: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/7.jpg)
SPME Advantages
• less solvent required
• speed
• less glassware is needed
• better precision and accuracy
• minimal solvent evaporation
• readily automated
![Page 8: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/8.jpg)
SPME Device
![Page 9: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/9.jpg)
SPME Procedure
Extraction
Desorption
![Page 10: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/10.jpg)
Degradation of 2,4-Decadienal
• 2,4-Decadienal is an Important Aldehyde Flavor
• Contributes Deep Fat Characteristics at 10 ppm
• Key Component in Flavors Such as:
– Chicken
– Lamb
– Beef
– French Fries
![Page 11: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/11.jpg)
Degradation of 2,4-Decadienal
Red Peaks: Neo FD 60°C 30 Days Blue Peaks: Neo SD 60°C 30 Days
Hex
anal
1-H
epta
nol
Hex
anoi
c A
cid
2-O
cten
al
Lim
onen
e
Met
hylo
ctan
oate
Tra
ns-2
-Non
enal
Oct
anoi
c A
cid
Dod
ecan
e
E2,
Z4-
Dec
adie
nal
E2,
E4-
Dec
adie
nal
![Page 12: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/12.jpg)
Neral Geranial
What is citral?
![Page 13: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/13.jpg)
Citral DegradationCitral
Cyclization (induced by protons)
OH OH
and/or
Monoterpene alcohols
and more…OH O
OHOH
.
,p-dimethylstyrene p-methylacetophenone
O
p-cresol
OH
p-cymene
![Page 14: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/14.jpg)
No. Compounds No. Compounds
1 2,3-dehydro-1,8-cineole 8 p-mentha-1(7),2-dien-8-ol
2 p-cymene 9 a-terpineol
3 p-cresol 10 (2R, 5R)-tetrahydrofuran isomer
4 a,p-dimethylstyrene 11 (2S, 5R)-tetrahydrofuran isomer
5 p-mentha-1,5-dien-8-ol 12 Trans-p-menth-2-ene-1,8-diol
6 p-methylacetophenone 13 Cis-p-menth-2-ene-1,8-diol
7 p-cymen-8-ol 14 8-hydroperoxy-p-cymene
Citral Degradation
Ueno, T., et al., J. Agric. Food Chem., 2006. 54(8): p. 3055.
Disappears in TWO weeks!
![Page 15: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/15.jpg)
Mass Spectrometry
![Page 16: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/16.jpg)
Mass Spectrometry Basics
• Ionization of molecules;• Separation of ions based on mass-to-charge ratio (m/z);• Detected by mass analyzer under electrostatic field;• Routine in analytical labs;• GC-MS: interface of mass spectrometer with GC;• LC-MS: interface of mass spectrometer with HPLC.
- An analytical tool used for measuring the molecular mass of a molecule
![Page 17: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/17.jpg)
Sample Introduction
• Direct Injection: for a gas or a volatile liquid;• Direct Insertion: for somewhat volatile solid, the sample is placed in a small cup at the end of a stainless steel rod or probe;• For mixtures, we can use GC-MS or LC-MS through an interface which removes excess GC carrier gas or HPLC solvent.
![Page 18: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/18.jpg)
Ionization
• In ion source, the compound is exposed to a beam of electrons emitted from a filament composed of rhenium or tungsten metal.• When a direct current is applied to the filament (~70 eV), it heats and emits electrons that move across the ion chamber toward a positive electrode;• As the electron pass through the source region, they come in close proximity to the sample molecule and extract an electron, forming an ionized molecule;• The ionized molecules can further fragment into smaller molecular fragments.
![Page 19: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/19.jpg)
Ionization Process
![Page 20: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/20.jpg)
Example of Ionization Process
![Page 21: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/21.jpg)
Chemical Ionization (1)
- Chemical Ionization (CI) is especially useful technique when no molecular ion is observed in EI mass spectrum, and also in the case of confirming the mass to charge ratio of the molecular ion. -Chemical ionization technique uses virtually the same ion source device as in electron impact, except, CI uses tight ion source, and reagent gas. Reagent gas (e.g. ammonia) is first subjected to electron impact. -Sample ions are formed by the interaction of reagent gas ions and sample molecules. This phenomenon is called ion-molecule reactions. -Reagent gas molecules are present in the ratio of about 100:1 with respect to sample molecules. -Positive ions and negative ions are formed in the CI process. -Depending on the setup of the instrument (source voltages, detector, etc...) only positive ions or only negative ions are recorded.
![Page 22: Food Analysis Lecture 24 (04/23/2013)](https://reader036.fdocuments.net/reader036/viewer/2022062309/56813bca550346895da4f4e8/html5/thumbnails/22.jpg)
Chemical Ionization (2)
- In CI, ion molecule reactions occur between ionized reagent gas molecules (G) and volatile analyte neutral molecules (M) to produce analyte ions. Pseudo-molecular ion MH+ (positive ion mode) or [M-H]- (negative ion mode) are often observed. Unlike molecular ions obtained in EI method, MH+ and [M-H]- detection occurs in high yield and less fragment ions are observed.
Positive ion mode: (example) CH4 + e -----> CH4
+. + 2e ------> CH3+ + H.
CH4+. + CH4 -----> CH5
+ +CH3.
Negative ion mode: (example)NH3 + e -----> NH3
+. + 2eNH3
+. + NH3 ------> NH4+ + NH2
.