fMRI Quality Assurance (QA)(v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre for Scotland, University of Edinburgh

Transcript of FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical...

Page 1: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

fMRI Quality Assurance (QA)(v2)

Data acquisition, processing and analysis

Katherine LymerMedical Physicist in MR Support

SFC Brain Imaging Research Centre for Scotland, University of Edinburgh

Page 2: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

fMRI QA: Objectives

Using measures of SNR, SFNR, fluctuation and drift

• Describe the quality of the fMRI data

• Monitor the long-term stability of scanner performance in fMRI

• Allow between site-assessments of fMRI data

Page 3: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Acquisition

Coil: As per usual fMRI examination (e.g. quadrature / 8 channel-head coil)Test object: Manufacturer provided homogenous sphere

Equipment set-up

Test-object set-up

Use test object holder (if applicable);Place test object in centre of coil;

Align lasers with centre of coil;Mark as “zero” positionCentre coil / test object in scanner

Page 4: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Acquisition

Scanning parametersNote magnet room temperatureSequence: Two approaches:(1) As per typical fMRI examination(2) “Pushing” the scanner as hard as possible (measure stability “under stress”)In Edinburgh, we use the following protocol on our 8 channel head coil:

TE TR NEX FOV #slicesSlice

thicknessspacing nVols Matrix

Flip angle

Tscan

40ms 2044ms 1 24cm 28 5mm 0 202* 128*64 90º6min 53sec

* Includes dummy scans (processing software expects 2 dummy vols)

Frequency of acquisition:Site dependant: May be useful to acquire data on a weekly / fortnightly basis to establish a baseline; The results from our QA suggest that a monthly test is acceptable on our scanner.

Note the RF-receiver and transmit gain and the resonant frequency

Page 5: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Processing

NB This processing protocol is based upon that provided by the BIRN (http://www.nbirn.net/research/function/fbirn_tools_primer.shtm)and is similar to that used in “Calibrain”.

See also:Weisskoff et al, MRM 36: 643-645, 1996Friedman et al, JMRI 23: 827–839, 2006

Requirements:(1) PC / workstation with matlab;(2) SPM5 (as supplied by the FIL: http://www.fil.ion.ucl.ac.uk/spm/)(3) fMRI_qa.m script (as supplied by fBIRN: http://www-calit2.nbirn.net/tools/fbirn_stability_phantom/index.shtm)This script needs to be modified to reflect local / in-house needs(4) mricro

Transfer data to “off-line” to image analysis work station

Page 6: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Data Management

(1) Ensure that data is stored in central fMRI data store, in appropriate folder denoting the current QA session

(E.g. In Edinburgh, the data is stored in Y:\MONTHLY_QA using the format: QAddmmyyyyTEMPtt_yyyymmdd_studyNumber)

(2) In the relevant monthly QA folder, make two new sub-folders to differenentiate between the raw and reconstructed data e.g. “raw_data” and “SPM5_output”.

(3) Copy the fMRI dicom files into the raw_data folder(E.g. in Edinburgh, this is: QAddmmyyyyTEMPtt_yyyymmdd_studyNumber\

raw_data)

Page 7: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Pre-processing

(1) Start Matlab (ensure all paths are correct)(2) Start SPM5(3) Convert the DICOM files using the “DICOM Import” function in SPM5:

Select the files to be reconstructed (all in the “raw data” folder)Select “Done”;Select output dir: SPM5_output;Select “Done

(NB if performing reconstruction on a windows machine: In pop-up window: Output image format: Two file (img +hdr) NIFTIUse ICEDims in filename: No)

This may take a few minutes

Page 8: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Pre-processing

(4) Start mricro

Check the central slice in volumeLoad the first “real” fMRI volume (i.e. the third listed volume) into mri croReview entire volume and note central sliceAmend fMRI_qa.m accordingly

Page 9: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Processing: fMRI_qa_calcEdits required to fMRI_qa (original version obtained from

http://www-calit2.nbirn.net/tools/fbirn_stability_phantom/index.shtm)

(1) Ensure TR is correct;

(2) Ensure that the date is in the correct format;

(3) Define dimensions of ROI for ghosting measurement by changing:X1G, X2G, Y1G, Y2G

(4) Ensure that the central ROI is placed appropriately within the central image

(5) Check that the central slice number is correct (i.e. it matches that practically acquired)

Page 10: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Processing: fMRI_qa_calcRun your version of fMRI_qa

Import values into excel spreadsheet;Plot results to help identify any trends / outliers.

Note any variation in RF-receiver / -transmit gain and resonant frequency (since the phantom and the protocol remain the same, these should remain constant)

Page 11: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:All image analysis is performed on the central slice of the volume. (NB It is

important to check that the central slice as designated in KL_fMRI_qa.m matches that practically acquired.)

Application of the BIRN / Calibrain protocol yields six measurements:

(1) Mean signalA mean image is produced by calculating the mean, on a voxel by voxel

basis, across all the central slices.A signal summary value is obtained from an ROI placed in the centre of this

mean image.

Example mean image:

Page 12: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(2) Temporal Fluctuation Noise ImageA fluctuation noise image is produced by subtracting the calculated trend

line from the data (the BIRN function uses a second order polynomial). (It’s effectively a standard deviation image.)

By removing the trend from the data, we can estimate the fluctuations in the data about that trend: A linear trend may indicate a systematic increase or decrease in the data (caused by e.g. sensor drift).

Example standard deviation image:

Page 13: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(3) Signal-to-Fluctuation Noise Ratio (SFNR) Image and Summary SFNR

ValueA SFNR image is created by dividing the mean signal image by the temporal

fluctuation (SD) image on a voxel-by-voxel basis. The summary SFNR value is obtained from the mean SFNR from the same

ROI placed in the centre of the SFNR image.

Example SFNR image:

Page 14: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(4) Static Spatial NoiseIn order to measure the spatial noise, the sum of the odd and even

numbered slices are calculated separately (so there are two results); the difference between these two sums is then calculated. This approximates the static spatial noise.

Is there is no drift in either amplitude or geometry across the time series, then this difference image will show no structure of the phantom and provides a measure of intrinsic noise.

Example difference image:

Page 15: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(5) SNR SummaryThe same ROI is placed in the centre of the static spatial noise (difference)

image.Using the signal summary value (see (1)), the

)sintpotime/#)valuesummaryiance((var

valuesummarysignalSNR

Page 16: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(6) SNR Percent Fluctuation and DriftThe average signal intensity is calculated from each of the central slices

across the time series using the same ROI, placed in the centre of each image.

A time series of volume number vs average intensity is plotted and a second order polynomial is fitted to the data to establish a trend line.

The mean signal intensity (prior to detrending) and the SD of the residuals after detrending are calculated.

signalensityintmean

residualsofSD100nfluctuatio%

signalensityintsignalmean

)valuefit(min)valuefit(max100drift%

Page 17: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(7) Summary graphsIn addition, a series of summary graphs showing the change in raw signal,

SGR, magnitude spectrum and relative standard deviation will be displayed

Page 18: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

Image Analysis:(8) Signal-to-Ghost Ratio (SGR)

In addition to the BIRN tests, a signal to ghost (SGR) measurement was added to the protocol by Dr Gordon Waiter, Aberdeen Biomedical Imaging Centre, University of Aberdeen.

The image may have to be windowed to observe the ghost.

In order to make the SGR measurement, a second (ghosting) ROI is defined within the ghosting region of the image. The SGR summary value is calculated from the mean signal (from the central ROI) / mean signal (from the ghosting ROI).

Page 19: FMRI Quality Assurance (QA) (v2) Data acquisition, processing and analysis Katherine Lymer Medical Physicist in MR Support SFC Brain Imaging Research Centre.

SINAPSE Reporting Mechanism:Short-termMaintain results locally;Distribute summary results with colleagues e.g. for multi-centre studies,

comparisons of scanner performance (KL happy to hold central data base and distribute as required).

Medium - Long termWe aim to produce a SINAPSE QA reporting website to which these results

can be uploaded.Watch this space…………..!