FGM-1 MPDR, 2/12/2015 FLUXGATE MAGNETOMETER Fluxgate Magnetometer (FGM) Kathryn Rowe

Transcript of Fluxgate Magnetometer (FGM) Kathryn Rowe...Fluxgate Magnetometer ‐ 3 Axis data stream ‐ 24 bit,...

  • FGM-1 MPDR, 2/12/2015

    FLUXGATE MAGNETOMETER

    Fluxgate Magnetometer (FGM)

    Kathryn Rowe

  • FGM-2 MPDR, 2/12/2015

    TEAM ORGANIZATION

  • FGM-3 MPDR, 2/12/2015

    Instrument Overview Instrument Description Development Overview

    AGENDA

  • FGM-4 MPDR, 2/12/2015

    INSTRUMENT OVERVIEW

    4

    ELFINFluxgate Magnetometer (FGM)Mission PDR Meeting

  • FGM-5 MPDR, 2/12/2015

    EXPLODED VIEW

  • FGM-6 MPDR, 2/12/2015

    Magnetic sectoring information for the EPD Potential to detect EMIC waves

    WHY DOES ELFIN NEED AMAGNETOMETER?

  • FGM-7 MPDR, 2/12/2015

    1. FGM Electronics

    2. FGM Sensor/Cable

    WHAT DOES THE FGM CONSISTOF?

  • FGM-8 MPDR, 2/12/2015

    BLOCK DIAGRAM

  • FGM-9 MPDR, 2/12/2015

    RFAS FROM PDR

    1

    FGM Resolution Requirement -- Waiting Board ApprovalThe FGM Resolution of 10pT was presented as a requirement whereas the noise level is 0.5nT/sqrt(Hz). The PRM may generate data with

  • FGM-10 MPDR, 2/12/2015

    INSTRUMENT DESCRIPTION

    ELFINFluxgate Magnetometer (FGM)Mission PDR Meeting

  • FGM-11 MPDR, 2/12/2015

    Fluxgate Magnetometer‐ 3 Axis data stream‐ 24 bit, 6pT resolution‐ Dynamic Range: ±55,000nT‐ Noise: 0.4nT/√Hz @ 1Hz‐ Sample rate: 200Hz

    Instrument DescriptionMagnetometer Electronics

  • FGM-12 MPDR, 2/12/2015

    POWER REQUIREMENT - FGM

    REQ ID Requirement Parent(s) Verification Method Verification Procedure

    PLD‐35

    The Payload subsystem shall not exceed the power allocated in the ELFIN system power budget

    SYS‐22

    T/A: Measured power consumption data during component tests;  Analysis of the complete set of Instrument states.

    Power BudgetPowerBudget‐##‐B.xlsx

    Power allocated for FGM: 0.5 W (during science data collection)Current best estimate:  0.423 WMargin: 15 % (0.063 W)CBE + Margin: 0.486 WContingency: ~0.014 W

    Supply voltages: Currents:+8 V+3.3 V+1.5 V

    30 mA35 mA45 mA

    Power:0.24 W0.1155 W0.0675 W

    0.423 W+

  • FGM-13 MPDR, 2/12/2015

    PWA mounts to 4 Delrin® brackets with 2-56 socket head brass bolts at each corner

    ELECTRONICSMECHANICAL INTERFACE

    The Delrin® brackets connect to cubesat chassis with #4-40 countersunk brass screws, and the PWA lies perpendicular to the instrument stack

  • FGM-14 MPDR, 2/12/2015

    INSTRUMENT DESCRIPTIONMAGNETOMETER SENSOR

    3 Axis Fluxgate 48mm x 48mm x 25mm 58 grams without cable 106 grams with 1 m cable

  • FGM-15 MPDR, 2/12/2015

    1. Conductor – 3 Twisted Pairs Material: Silver-Coated

    Strength Alloy Thickness of Silver Coating:

    2um min Size: AWG 28

    2. Binder Material: Gore-Tex®

    3. Severed Wire Shield Material: Solver-Coated,

    oxygen free Cu Thickness of silver coating:

    2.5um min4. Shcatch

    Material: Gore-Tex®

    FGM CABLE

    Characteristics:Operating Voltage: 100VrmsOperating Temp Range: -150C to 200C

  • FGM-16 MPDR, 2/12/2015

    Stacer is 75cm long Sensor connects to

    stacer with #4-40 non-magnetic screws

    Final harness length TBD. CBE is 100cm

    MLI installed over sensor

    SENSOR MECHANICAL INTERFACE

  • FGM-17 MPDR, 2/12/2015

    STACER Keleva Design Inc A 75cm BeCu stacer boom with a

    SMAR actuator and a flywheel break

  • FGM-18 MPDR, 2/12/2015

    MASS REQUIREMENT - FGMREQ ID Requirement Parent(s) Verification Verification Document

    PLD‐34 The Payload subsystem shall not exceed the mass allocated by Systems SYS‐14

    System mass documented and tracked through assembly, integration and test

    Mass BudgetMassBudget‐##‐E.xlsx

    Mass allocated for FGM: 237 gCurrent best estimate (CBE): 199 gMargin: 15 % (29.9 g)CBE + Margin: 228.9 gContingency: 8.1 g

    FGMComponent Maturity Qty Unit mass CBE Margin CBE+Margin

    FGE (FGM Electronics) P 1 90.00 90.0 15% 103.5FGS (FGM Sensor with 1m cable) P 1 109.00 109.0 15% 125.4

  • FGM-19 MPDR, 2/12/2015

    DEVELOPMENT OVERVIEW

    ELFINFluxgate Magnetometer (FGM)PDR Meeting

  • FGM-20 MPDR, 2/12/2015

    FGM REQUIREMENTS

    REQ ID Requirement Parent(s) Comply?

    FGM

    PLD‐21 The FGM shall have a range of 55,000 nT SCI‐01, SCI‐04, SCI‐05Yes.  Current dynamic range is ±58,400nT

    PLD‐22 The FGM shall have a resolution of 0.1nT SCI‐01, SCI‐04, SCI‐05 Yes. Current resolution is 7pT.

    PLD‐23 The FGM shall have an absolute stability of 100 nT SCI‐01, SCI‐04, SCI‐05 Need to test

    PLD‐24 The FGM shall have a relative stability of 1 nT/2 hr SCI‐01, SCI‐04, SCI‐05 Need to test

    PLD‐25 The FGM shall have a noise level of 0.5 nT/sqrt(Hz) SCI‐01, SCI‐04, SCI‐05Yes. Current resolution is 0.4nT/sqrt(Hz)

    PLD‐27 The FGM shall have a digitization of > 20 bits SCI‐01, SCI‐04, SCI‐05Yes. Current digitization is 24 bits

    PLD‐28 The orthogonality of the FGM sensor shall be known to 

  • FGM-21 MPDR, 2/12/2015

    Prototype Complete (Board and Sensor)

    CURRENT STATUS OF INSTRUMENT

  • FGM-22 MPDR, 2/12/2015

    Elfin Status (2/26/12) PRM with ST5 Sensor Feedback Resistor: 2.7K to 3.3K

    Problems: Frequency Response Incorrect: LabView needs to be

    updated Need New Drive Circuit

    X axis Offset: 1867nT Dynamic Range: +/-64,000nT Noise Level: 402pT at 1Hz

    Y axis Offset: -49nT Dynamic Range: +/-64,000nT Noise Level: 379nT at 1Hz

    Z axis Offset: 3129nT Dynamic Range: +/-58,400nT Noise Level: 373pT at 1Hz

    Power Consumption 3.3V: 32mA 1.5V: 42mA 8V: 28mA

    TEST DATA

  • FGM-23 MPDR, 2/12/2015

    STABILITY TESTING(2 HOURS)

  • FGM-24 MPDR, 2/12/2015

    FGM IMPLEMENTATION PLAN

  • FGM-25 MPDR, 2/12/2015

    1. Reduce dynamic range on X and Y axis From ±64,000nT to ±55,000nT Feedback resistors needs to be changed Increases margin on noise level

    HARDWARE CHANGES NEEDED ONPROTO

  • FGM-26 MPDR, 2/12/2015

    1. Change input clock from 32MHz to 16.384MHz GSE oscillator needs change

    2. Change sample rate from 122Hz to 200Hz Need to implement anti-aliasing filter IDPU to average down to 10Hz

    3. Change FPGA code from Libero Schematics to Verilog

    FPGA CHANGES NEEDED ONPROTO

  • FGM-27 MPDR, 2/12/2015

    1. Electronics Thermal Cycling (Expected temperature range needed) Need test plan Need to determine electronics temperature

    dependence2. Sensor Temperature Cycling (Expected

    temperature range needed) Need test plan Need to determine sensor temperature

    dependence3. Alignment Test Need orthogonality of sensor

    PROTOTYPE TESTING

  • FGM-28 MPDR, 2/12/2015

    1. Minor board outline change Need to accommodate S/C harnessing

    2. Remove JTAG connections on IDPU connector Lose ability to program in-flight (low risk)

    3. Change connector from 26pin to 20pin Smaller connector to increase board space

    4. Add ±15V supplies to IDPU connector WCA revealed supply rails for analog switch was

    undervoltage5. Bring the drive circuit outside PRM Unable to change supply rails for analog switch

    inside PRM

    CHANGES FROM PROTO → EM