Figure 1
-
Upload
garry54 -
Category
Technology
-
view
282 -
download
0
Transcript of Figure 1
Implantable Cardiac Devices and Electromagnetic Tracking Systems
Are Patients at Risk?
2004
Authors:Chris Holmes – Southern Illinois University at Edwardsville
Amy McKinney – University of Nebraska-Lincoln
Faculty Sponsor:Dr. Hank Grant, EMC Center Director
Glenn Kuriger, EMC Center Assistant Director
Abstract
By definition, an active implantable medical device (AIMD) is a device that is inserted, partially or fully, into a human body or body cavity for permanent use by medical intervention. These devices include equipment such as pacemakers, implantable cardioverter defibrillators (ICDs), pain stimulators, respiration-stimulators, insulin or drug pumps, cochlea implants, and electrocardiogram monitors. Of particular concern are pacemakers and ICDs, which are cardiac implantable devices that treat different heart conditions. According to Faraday’s law, for any time varying magnetic field, a voltage may be induced at the input of an implanted cardiac device, thus creating a surge of excess electricity. These fields may emanate from any electronic device, such as electronic article surveillance system, weapon detectors, or high voltage overhead lines. Hospitals are now utilizing electromagnetic tracking systems during surgery as a more accurate way to view the area being operated. A wireless probe is inserted into the patient and communicates with a base station to relay information through pulsed electromagnetic waves. Hence, these waves could possibly interfere with patients who have implantable cardiac devices if the base station is located too close to the device.
The Center for the Study of Wireless Electromagnetic Compatibility at the University of Oklahoma is conducting in vitro testing to find if there is any possible interaction between implantable cardiac devices and electromagnetic tracking systems. Four questions are under consideration:
1) What degree of interaction, if any, is present between a wireless tracking system base station and various implantable cardiac devises?
2) What is a safe distance between a patient wearing an implantable cardiac device and the base station during transmission?
3) Is there a relationship in the placement of the implantable cardiac device within the body and the corresponding interaction that could occur from exposure?
4) Do some implantable cardiac devices have greater interactions with an electromagnetic tracking system than others?
By using a torso simulator, systematic testing of various degrees of interaction is possible. Due to limited research in the field of medical electromagnetic device interference, a study is warranted to investigate these issues. However, as of the writing of this report, cardiac equipment has yet to be shipped, therefore causing the delay of measurement. As a consequence, it is unwarranted to attempt any speculation on possible results. Hopefully, future research will continue with the aforementioned goals presented in consideration.
Table of Contents
Table of Contents..............................................................................................................................i
Table of Illustrations........................................................................................................................ii
Introduction......................................................................................................................................1
Literature Review............................................................................................................................2
Problem Definition..........................................................................................................................5
Analysis of Problem........................................................................................................................6
Solution Methods Employed...........................................................................................................8
Results............................................................................................................................................10
Recommendations..........................................................................................................................10
Limitations.....................................................................................................................................11
References......................................................................................................................................12
Appendix..........................................................................................................................................a
Holmes McKinneyi
Table of Illustrations
Figure 1: 1950's Era Pacemaker................................................................................3
Figure 2: Modern Pacemaker..........................................................................................................3
Figure 3: Left vs. Right Side Orientation.......................................................................................4
Figure 4: Electromagnetic Spectrum..............................................................................................7
Figure 5: Pulsed EM Field..............................................................................................................8
Figure 6: Torso Simulator...............................................................................................................9
Table 1: Test Procedures...............................................................................................................a
Holmes McKinneyii
Introduction
One research topic that has received much attention by the academic and commercial
world is that of electromagnetic interference with active implantable medical devices (AIMDs).
Specifically, cardiac devices such as implantable pacemakers and cardioverter defibrillators have
been shown to have potential hazards associated with electromagnetic interference from various
sources such as wireless phones or security systems. The Center for the Study of Wireless
Electromagnetic Compatibility (EMC Center) at the University of Oklahoma has conducted
several studies examining such interactions. In their research, as well as other academic
investigations, interference has proven minimal, if not nonexistent. However, even with these
results, further examination must continue to determine compatibility between new wireless
technology and AIMDs. Medical technology in particular is experiencing rapid technological
advances, which may have benefits for some patients, but hazards for others.
In the mid 1990’s, the need arose for an independent center to investigate the use of
wireless devices and their potential interaction with AIMDs. In response, Dr. Hank Grant at the
University of Oklahoma created the Center for the Study of Electromagnetic Compatibility
(EMC Center), shortly after his arrival at the university in 1994. Since then, the EMC Center has
investigated wireless interference with devices such as hearing aids, pacemakers, and
implantable defibrillators. In recent years, cardiac devices have received much attention from
the EMC Center, with investigations revealing that while several select models of pacemakers or
ICDs have shown interaction with older style analog wireless phones, newer digital phone
formats such as PCS or GSM are essentially interaction free. However, wireless technology is
always changing, proliferating in both format and use.
Holmes McKinney1
Literature Review
By definition, an active implantable medical device (AIMD) is a device that is inserted,
partially or fully, into a human body or body cavity for permanent use by medical intervention.
These devices include equipment such as pacemakers, implantable cardioverter defibrillators
(ICDs), pain stimulators, respiration-stimulators, insulin or drug pumps, cochlea implants, and
electrocardiogram monitors (Irnich, 2002). Of particular concern are pacemakers and ICDs,
which are cardiac implantable devices that treat different heart conditions. A pacemaker is
utilized in patients experiencing heart failure, which is when heart muscle has become diseased
to the point that it lacks the ability to beat on its own (Ashley, et al, 1998). The device supplies
electric shocks to “pace” the heart either continuously or as needed. A heart “attack” is another
condition altogether. During this time, the heart beats faster than blood can flow, causing the
heart muscle to become starved for oxygen, thus damaging the muscle. An ICD monitors for this
type of activity, then provides a shock of electricity to correct the heart’s rhythm (some cutting-
edge models also function as pacemakers) (Irnich, 2002). ICDs are relatively new compared to
pacemakers, first implanted in 1980 (Cannom and Prystowsky, 2004), as opposed to pacemakers
in 1952 (Hayes and Furman, 2004). However, since the devices use electric signals for both
programming and monitoring of a patient, it can be dangerous if a patient is exposed to any
electric equipment emitting harmful levels of electromagnetic radiation. Figure 1 and Figure 2
demonstrate the extreme change that pacemakers have undergone in the last 50 years.
Pacemakers have now become smaller and more versatile in their use.
Holmes McKinney2
Figure 1: 1950's Era Pacemaker Figure 2: Modern Pacemaker
The causes that attribute to an AIMD to misinterpret a signal can easily be explained.
According to Faraday’s law, for any time varying magnetic field, a voltage may be induced at the
input of an implanted cardiac device, thus creating a surge of excess electricity (Scholten and
Silny, 2001). These fields may emanate from any electronic device, such as electronic article
surveillance system, weapon detectors, or high voltage overhead lines (Witters, et al, 2001).
Some studies have constructed specific “Faraday Cages” to examine this phenomenon (Hedjiedj,
2002). Assuming that the magnetic field is sinusoidal and located in the AIMD range, acting
perpendicularly to the frontal plane of the thorax and is homogeneous, there exists the possibility
that a voltage could be produced. Also, if pacemakers or ICDs are under consideration, the
placement of the device can also greatly influence any induced voltage. Left side implantation
creates an induction loop, versus a right side that typically has an “s-shape” wire lead orientation
Holmes McKinney3
(Scholten and Silny, 2001). Figure 3 pictorially demonstrates the different placement options of
pacemakers within a patient’s chest, and the resulting reference loops that are incurred. Different
types of wireless technologies, primarily found in wireless phones, also attribute to AIMD
interaction.
Figure 3: Left vs. Right Side Orientation
The FDA has found that many different electrically powered medical devices such as
pacemakers and implantable defibrillators can be affected by electromagnetic interference
(Witters, et al, 2001). Both in vitro and in vivo testing has been conducted dealing with this
issue. Studies have focused on GSM (Global System for Mobile Communication) wireless
technology and any interactions that may occur with a cardiac device’s functionality (Barbaro, et
al, 2003b). In vitro results have concluded that when the worst-case scenario is tested (phone at
maximum output and implantable cardiac device at its highest sensitivity) there is some
interaction between certain types of implantable cardiac devices and certain types of wireless
phones. The interaction found could be that the device perceived the electromagnetic waves as a
heart beat, or that the device “fired” when it should not have. However, in all cases, when the
wireless phone was removed and turned off the implantable cardiac device showed no signs of
Holmes McKinney4
being re-programmed (Grant, 1998) (Kuriger, 1998). Based on prior in vitro results, in vivo
testing showed that there were no threatening interactions between specific devices and phones
when in close contact (Barbaro, et al, 2003a). However, it should be noted that in the Barbaro
study, only two phone models and one pacemaker model was examined, so these results may not
hold true for all combinations of pacemakers and wireless technologies.
Hospitals are now utilizing electromagnetic tracking systems during surgery as a more
accurate way to view the area being operated. This new technology is an upgrade from the
optical systems that have been widely used for many years (Poulin, 2002). This technology
provides the surgeon with a 3D “roadmap” of the patient as well as the exact location of the
probe within the body. The system uses an emitter to produce an electromagnetic field that is
received by a base station. Some stations are capable of converting this information into six
degrees of freedom (Poulin, 2002), allowing the surgeon to view the patient in ways never before
capable with optical systems. The possibility of interference with AIMD patients requires
examination.
Even though there is no major threat to the general public, there are a substantial number
of AIMD patients who could be affected by an electromagnetic interference during surgery.
Approximately 110,000 Americans (eighty percent of which are 65 years of age or older) are
fitted with pacemakers each year (Business Word, 2003). In 2003, it was expected that nearly
130,000 implantable defibrillators were to be implanted worldwide (Cannom, 2004). These
statistics justify the need for further research in the field of wireless electromagnetic interactions.
Problem Definition
Hospitals are now utilizing electromagnetic tracking systems (ETS) during surgery as a
more accurate way to view the patient’s area being operated on. This new technology is an
Holmes McKinney5
upgrade from the optical systems that have been widely used for many years. Using GSM
wireless signals, this technology provides the surgeon with a 3D “roadmap” of the patient as well
as the exact location of the probe within the body. Optical systems are unable to provide this
level of precision. A wireless probe is inserted into the patient and communicates with a base
station to relay information through electromagnetic waves. Hence, these waves could possibly
interfere with patients who have implantable cardiac devices if the base station is located too
close to the device. The EMC Center is conducting in vitro testing to find if there is any possible
interaction. Four questions are under consideration:
1) What degree of interaction, if any, is present between a wireless tracking system base
station and various implantable cardiac devises?
2) What is a safe distance between a patient wearing an implantable cardiac device and
the base station during transmission?
3) Is there a relationship in the placement of the implantable cardiac device within the
body and the corresponding interaction that could occur from exposure?
4) Do some implantable cardiac devices have greater interactions with an
electromagnetic tracking system than others?
Due to limited research in the field of medical electromagnetic device interference, a study is
warranted to investigate these issues.
Holmes McKinney6
Analysis of Problem
Electromagnetic fields can be either continuous or pulsed. The ETS uses pulsed signals
to carry digital information in the GSM wireless format at either 800 MHz or 1900 MHz (See
Figure 4). Pulsed electromagnetic fields, such as in Figure 5, have the capability to create “false
heart signals” which can cause AIMDs to misinterpret data. This creates inherent problems
when in close proximity to AIMDs. However, depending on which cardiac device is under
consideration, the pulsed field may have varying effects. The possibility exists that a pacemaker
will be more sensitive to an exposure than an ICD due to their respective functions. Pacemakers
continuously pace a heart based on measured readings that can be easily interfered with during
exposure to pulsed electromagnetic fields. In contrast, an ICD shocks an erratic heart back into a
normal rhythm, which requires the device to charge before treatment may be administered.
Thus, in order for a pulsed electromagnetic field to become dangerous, the ICD must be
subjected to the field for a longer period of time compared to a pacemaker. Therefore, with more
hospitals utilizing ETS technology, the hazards to AIMDs patients cannot be ignored.
Holmes McKinney7
Figure 4: Electromagnetic Spectrum
Figure 5: Pulsed EM Field
Holmes McKinney8
Solution Methods Employed
All testing will be conducted at the EMC Center on the University of Oklahoma’s campus.
Simulation of pacemakers and implantable defibrillators will be accomplished by way of a torso
simulator and testing equipment which generate and monitor electrical signals. Specifically, an
ECG signal injection system will be utilized to simulate heart activity and signal monitoring
equipment for acquiring pacemaker and ICD signals. All equipment will be set up in and around
the torso simulator.
As can be seen in Figure 6, the torso simulator is a plastic box used to represent the
human trunk cavity. The simulator measures 23” x 16.75” x 6” with a volume of 28 quarts. It is
filled with 0.03 molar saline solution which simulates human body conductivity. Two horizontal
grids placed one below the other support both the pacemaker or defibrillator under consideration
and the tracking system base station. The lower grid, which supports the pacemaker or
defibrillator, is completely submersed in the saline solution. The grids can be moved to different
heights, allowing the testing of different separations distances to be measured. The tracking base
station will be moved to every point on the grid to test various interactions.
Figure 6: Torso Simulator
Holmes McKinney9
One important factor of the test procedure is generating an artificial heart signal.
Generating and monitoring a heart rhythm may be accomplished by placing stainless steel plates,
measuring 50 mm x 50 mm x 2 mm, into the torso simulator. One pair of plates (on the long axis
of the tank) monitor the pacemaker or ICD signal, and the second pair (on the short axis) inject a
simulated heart signal into the tank. The signal received from the torso simulator is first applied
to the input of a differential amplifier designed for ECG signal monitoring. After amplification,
the signal is filtered for clarity before being inputted into a two-channel digital storage
oscilloscope. Through observation of the ECG signal on the oscilloscope, any interaction of the
pacemaker or ICD function with the electromagnetic tracking system base station will be
recorded. A summary of this procedural setup can be found in the Appendix (Table 1).
It is important to note that ICD and pacemaker parameters are typically programmed non-
invasively by means of RF signals or pulsed magnetic fields. This ability of the device to
respond to both internal and external magnetic and RF signals allows the device to be
programmed for optimal clinical benefit as the patient’s needs change. Each pacemaker or ICD
company will supply the appropriate instructions necessary for interrogating and programming
its respective units. The programming features vary widely, but all units provide the control
necessary to establish the common parameters needed. Each unit will be programmed for the
worst-case condition (maximum sensitivity and minimum refractory period).
Results
Unfortunately, due to time constraints, testing was unable to commence as of the writing
of this report. Due to limited research concerning this particular type of wireless interaction, it is
unwarranted to attempt any speculation on possible results. Hopefully, future research will
continue with the aforementioned goals presented in consideration.
Holmes McKinney10
Recommendations
Recommendations can be made for future research of implantable cardiac device
interaction with electromagnetic tracking systems in relation to the general in vitro setup.
Previous research at the EMC Center has not made a distinction between left and right side
placement orientation of pacemakers or implantable defibrillators, and this can have a great
effect on the level of interaction with not only electromagnetic tracking systems, but also
technologies such as wireless phones or security systems. Also, previous research has not
considered constant body temperature in the construction of torso simulators. A constant
temperature of 98.6 degrees Fahrenheit would also have an effect on the conductivity of signals
within the simulator, perhaps improving noise levels and accuracy.
Holmes McKinney11
Limitations
The major limitation involved with this particular study is time. It has proven difficult to
obtain the proper implantable device equipment in the timeframe allowed, thus not allowing the
study to commence. However, delays are sometimes unavoidable factors in any form of
research, particularly ones that involve specialized equipment. Hopefully in time this study will
be completed, providing an answer to the safety issue raised in the use of electromagnetic
tracking systems and implantable cardiac devices.
Holmes McKinney12
References
Ashley, Robert J., et al, 1998, Measurement of Potential Magnetic Field Interference with Implanted Cardioverter Defibrillators of Pacemakers, Electro Int. Conf. Proc. of IEEE, 159-170.
Barbaro, V., Bartolini, P., Calcagnini, G., Censi, F., Floris, M., et al, 2003a, In vitro and in vivo evaluation of electromagnetic interference between wireless home monitoring pacemakers and GSM mobile phones, Proc. Ann. Int. Conf. IEEE Engineering in Medicine and Biology Society, 4, 3602-3605.
Barbaro, V., Bartolini, P., Calcagnini, G., Censi, F., Beard, B., et al, 2003b, On the mechanisms of interference between mobile phones and pacemakers: parasitic demodulation of GSM signal by the sensing amplifier, Physics in Medicine and Biology, 48, 1661-1671.
Business Word, 2003, Pacemaker prices still declining, Hospital Materials Management, 28, i6, p1(4).
Cannom, David S., and Prystowsky, Eric N., 2004, The Evolution of the Implantable Cardioverter Defibrillator, NASPE History Series, 27, 419-431.
Grant, Hank and Schlegel, Robert E., 1998, In Vitro Study of the Interaction of Wireless Phones with Cardiac Pacemakers, EMC Report 1998-2.
Hayes, D.L. and Furman, Seymour, 2004, Cardiac Pacing: How it Started, Where We Are, Where We Are Going, NASPE, 27, 693-704.
Hedjiedj, A., Goeury, C., and Nadi, M., 2002, A Methodological approach for the characterization of cardiac pacemaker immunity to low frequency interferences: case of 50 Hz, 60 Hz, 10 kHz and 25 kHz led disruptions, Journal of Medical Engineering and Technology, 26, no.5, 223-227.
Irnich, Werner, 2002, Electronic Security Systems and Active Implantable Medical Devices, Journal of Pacing and Clinical Electrophysiology, 25, no.8, 1235-1258.
Kuriger, Glenn W., Grant, Hank, and Schlegel, Robert E., 1998, In Vitro Study of the Interaction of Wireless Phones with Implantable Cardioverter Defibrillators, EMC Report 1998-1.
Poulin, Francois and Amiot, L.P., 2002, Interference during the use of an electromagnetic tracking system under OR conditions, Journal of Biomechanics, 35, 733-737.
Scholten, A., and Silny, J., 2001, The interference threshold of unipolar cardiac pacemakers in extremely low frequency magnetic fields, Journal of Medical Engineering and Technology, 25, no.5, 185-194.
Witters, Donald, et al, 2001, Medical device EMI: FDA Analysis of Incident Reports, and Recent Concerns for Security Systems and Wireless Medical Telemetry, IEEE International Symposium on Electromagnetic Compatibility, 2, 1289-1291.
Holmes McKinney13
Appendix
Table 1: Test Procedure
Torso SimulatorFill tank with 1.8 g/l saline solutionMeasure saline properties with YSI Model 33 S-C-T Meter Calibrate probe Measure saline temperature Measure salinity (0.18% to 0.21% acceptable) Measure conductivity (nominal value of 3400 micro-mhos per cm)
Position Plastic GridsPosition pacemaker/implantable defibrillator on bottom grid over (0,0) grid pointAdjust height of top grid such that implantable device is 1 cm below test subject
Waveform GeneratorSet frequency to 66.66 HzSet unit to “Burst” modeSet “Burst Rate” to 550 ms for pacemakers, 650 ms for implantable defibrillatorsConnect cable from “Output” jack on generator to stainless steel plates on opposite ends of torso simulator
Signal AmplifierEnable signal pathSet “Gain/Time Constant” to “X10” and “3.2” for implantable pacemakers and “X100” and “3.2” for pacemakersSet “Input” to “On”Set amplifier filter to 10 KHzSet amplifier gain to 500 mV/cmConnect cable from “Input” jack on amplifier to remaining stainless steel plates on torso simulatorConnect cable from “Aux-Out” jack on amplifier to Channel 1 of oscilloscope
OscilloscopeEnsure vertical menu settings on Channel 1 “Coupling” set to “DC” “Invert” set to “Off” “Bandwidth” set to “Full”
“Offset” set to “0 V” “Fine Scale” as needed “Position” as needed
Ensure vertical menu settings for Channel 2 “Coupling” set to “AC” “Invert” set to “Off” “Bandwidth” set to “Full”
“Offset” set to “0 V” “Fine Scale” as needed “Position” as needed
Ensure horizontal menu settings “Time Base” to “Main Only” “Trigger Position” to “50%”
“Sec/Div” set to “250ms”
Set trigger settings “Type” to “Edge” “Source” to “Channel 1” “Coupling” to “DC”
“Slope” to increasing “Level” as needed
Holmes McKinneya
Holmes McKinneyb