Fifth Generation Femtosecond Laser Flap Thickness

11
FIFTH GENERATION FEMTOSECOND LASER FLAP THICKNESS Mohamed A Guenena, MD Helga P Sandoval, MD, MSCR, Kerry D Solomon, MD Magill research center for vision correction Storm Eye Institute Medical University of South Carolina FINANCIAL DISCLOSURE: None of the authors has any financial interest in any product mentioned herein Mohamed A. Guenena: None Helga P. Sandoval: Alcon Laboratories, Inc. - D; Allergan, Inc. - D; Abbott Medical Optics -D Kerry D. Solomon: Alcon Laboratories, Inc. - A,C,D; Allergan, Inc. - A,C,D; Abbott Medical Optics - A,C,D; Advanced Vision Research - A,C,D; Bausch & Lomb, Inc. - A,C,D; Eyemaginations - A,C,D; QLT, Inc. - A,C,D

description

Mohamed A Guenena, MD Helga P Sandoval , MD, MSCR, Kerry D Solomon, MD Magill research center for vision correction Storm Eye Institute Medical University of South Carolina. Fifth Generation Femtosecond Laser Flap Thickness. Financial Disclosure: - PowerPoint PPT Presentation

Transcript of Fifth Generation Femtosecond Laser Flap Thickness

Page 1: Fifth Generation Femtosecond Laser Flap Thickness

FIFTH GENERATION FEMTOSECOND LASER FLAP THICKNESS

Mohamed A Guenena, MDHelga P Sandoval, MD, MSCR, Kerry D Solomon, MD

Magill research center for vision correctionStorm Eye Institute

Medical University of South CarolinaFINANCIAL DISCLOSURE: None of the authors has any financial interest in any product mentioned hereinMohamed A. Guenena: NoneHelga P. Sandoval: Alcon Laboratories, Inc. - D; Allergan, Inc. - D; Abbott Medical Optics - DKerry D. Solomon: Alcon Laboratories, Inc. - A,C,D; Allergan, Inc. - A,C,D; Abbott Medical Optics - A,C,D; Advanced Vision Research -  A,C,D; Bausch & Lomb, Inc. -  A,C,D; Eyemaginations -  A,C,D; QLT, Inc. -  A,C,D

Page 2: Fifth Generation Femtosecond Laser Flap Thickness

PURPOSE To assess the accuracy and reproducibility of

flap thickness created by the fifth generation femtosecond laser using:Intraoperative pachymetry Postoperative anterior segment OCT

ASCRS 2009

Page 3: Fifth Generation Femtosecond Laser Flap Thickness

METHODS Prospective, unmasked study 24 eyes of 12 patients Had primary LASIK surgery Flaps created with the new 5th generation femtosecond Laser (iFS

Advanced Femtosecond Laser, Abbot Medical Optics ) Attempted Flap depth

100 microns in 4 eyes 110 microns in 20 eyes

Assessment: Intraoperative pachymetry Postoperative anterior segment Fourier-domain optic coherence tomography

(OCT) on a subgroup of patients at an average of 3 month The SAS system software was used to analyze the results

ASCRS 2009

Page 4: Fifth Generation Femtosecond Laser Flap Thickness

OCT SETTINGS:

2 cross scans were used to measure each flap (Figure 1) The cross scan setting entails 2x1024 A scans (8 scans in each directions are then averaged)

The LASIK flap was identified by increased reflectivity at the flap-stromal interface. then the cursor was moved manually to the interface and measurements were

recorded. On each line of the cross scan, the thickness was calculated at four points ( Figure 2)

A 2.5 mm left to the midline B 1 mm left to midline C 1 mm right to midline D 2.5 mm right to midline

Fig. 1 Fig. 2

ASCRS 2009

A B C D

Line 1

Line 2 Line 3Line 4

Page 5: Fifth Generation Femtosecond Laser Flap Thickness

Pre and Postoperative Data

Intraoperative Pachymetry was done by subtraction method

ASCRS 2009

Pre-operative Post-operative day

30Spherical error -2.5 (-6.2—2.2) 0.1 (-0.2―1.2)Cylinder -0.7 ( -3.5―0.5) -0.2(-0.5―0)Spherical equivalent

-2.73 (-7.75― 2.25) 0 (-0.37―0.25)

Pachymetry 565 ( 535―596) NA

100 µm flaps 110 µm flapsNumber of eyes 4 18

Flap thickness by US pachymetry

109 µm (±9.6) 119 µm (±11.5)

RESULTS

Page 6: Fifth Generation Femtosecond Laser Flap Thickness

ASCRS 2009

OCT picture showing the four points measurements for a single line scan in a patient who had a 110 µm flap

Page 7: Fifth Generation Femtosecond Laser Flap Thickness

OCT RESULTS

Attempted Flap thickness

Number of eyes

OCT measurementsMean

Std Dev100 µm 2 117 µm 11.2110 µm 4 123 µm 6.5

ASCRS 2009

Overall flap thickness per group

Comparing the consistency of the flap thickness at different points in the 110 µm group in all line scans

Point A Point B Point C Point DAverage thickness

119 µm (±10.9)

124 µm (±4.4)

124 µm (±7.5)

124 µm (±6.4)

Page 8: Fifth Generation Femtosecond Laser Flap Thickness

OCT RESULTS Comparing the flap thickness by lines for the 110µm

group

ASCRS 2009

Line 1 Line 2 Line 3 Line 4

Average thickness

121 µm (±7.1)

124 µm(±5.5)

124 µm(±7.4)

124 µm(±6)

Page 9: Fifth Generation Femtosecond Laser Flap Thickness

CONCLUSION

The new fifth generation femtosecond laser shows High consistency of flap thickness High reproducibility of flap thickness Accurate flap thickness predictability The flaps are uniform at different points

ASCRS 2009

Page 10: Fifth Generation Femtosecond Laser Flap Thickness

ASCRS 2009

Page 11: Fifth Generation Femtosecond Laser Flap Thickness

ASCRS 2009

Thank you