![Page 1: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/1.jpg)
1
DSUImagingFacilityFEIQuanta250ESEMBasicOperationManual
ForAssistancewithSEMPleaseContact:Dr.MuktiRanaAssociateProfessor&ChairDepartmentofPhysicsandEngineeringSCRoom216302-857-6588,[email protected]
![Page 2: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/2.jpg)
2
The chamber of the SEM is kept under vacuum.
Under Normal condition the system is logged on.
If you do not see that the system is logged on, double click on the xT Microscope icon on the desktop
Then you will see the status bar of xT microscope.
Click on start UI.
![Page 3: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/3.jpg)
3
Log-intotheMicroscopeusingthefollowing- Username:supervisorPassword:Quanta250-9923724
Makesurethatyouareinthebeamcontroltabontherightsideofthescreen.VentthechamberbyclickingontheVENTbuttontorightsideofthewindow
Insert you sample into the chamber by using one of the sample holders and double sidedtapes
ifnecessary.Usingthe“Elephant Trunk”makesurethatyoursample is lessthan10mm. Ifyouneedtochangethebaseofthesampleholderyoucandothatbyunscrewingtheboltandconelikestructure.
![Page 4: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/4.jpg)
4
Determinethetypeofsampleyouhaveandselecttheappropriatevacuumconditionsyouwilluse.Usehighvacuummodefornormalconductivesampleandlowvacuummodeforsoftoranduncoatednon- conductivesamples.YoumayalsousetheESEMmodewithPeltierStageandGSEDdetectorforsampleswhichcontainliquid.Vacuumbuttons
Next,selecttherightaccelerationvoltagefromthepulldowntablocatedjustbelowthemenubarorfromtherightsideofthebeamcontroltab.
Oncetheproperpressurehasbeenachieved,thechamberICONinthelowerrightsideofthescreenwillbefullygreen.
![Page 5: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/5.jpg)
5
TurnontheBeambyclickingontheBEAMONbutton
Thedisplayshouldlooklikethis. Scanspeedcontrol Pausereleasebutton
TopleftSEimage,ToprightBackscatter,BottomLeftisthecombinedSEandBackscatterimageandtheBottomrightisthechamberviewscope.Activateuptothreedisplaysatonetimebyselectingthedisplayandtheclickonthepausereleasebutton.Thedisplaywillstarttoscan.
Toincreaseanddecreasethescanspeedsusethearrowupanddown.
![Page 6: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/6.jpg)
6
Whileviewingthechamberscope(bottomrightscreen)pressandholdthescrollwheelbuttonandpushthemouseupwardtopositionyoursamplenearthe10mmmarkonthescreen.Itmaybenecessarytounpausethechamberscopescreenfirst.
IncreasetheMagnificationandfocustheSEimagebyusingtheUSERinterfacepanelshownbelow.F7displaysareducedscreenforeaserrealtimeadjustments.F5enlargestheactivescreentofittheentirewindow.
OncetheimageisfocusedlinktheZtoFWDbyclickingontheICONonthetoolbar
Youmayalsogotonavigationtabtoseethezdistance.
FormostoperationsitisrecommendedthataZvalueof10mmisnowsetastheworkingdistance.
![Page 7: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/7.jpg)
7
Increasemagnificationtoatleast1k(higherthebetter)andfocustheimageaswellasyoucan.Nextfromthemainmenu,selecttheBeam>DirectadjustmentsandthenstarttheLensalignbyclickingontheLENSALIGN button.
Whileobservingtheimage,placeyoumouseoverthesmallsquareboxintheLensAlignandholditdown.Movethemouseandboxwithinthefieldtoreducemovementintheimage.
TurnoftheLensAlignbycrossingitout.RefocusyouimageandincreasetheMagnificationtoabove20k(thehigherthebetter)andusingtheUserInterfacePanel,StigmatetheimageusingtheXandYcontrolsforsharpestimagepossible.F7displaysareducedscreenforeaserrealtimeadjustments.F5enlargestheactivescreentofittheentirewindow.
WhenusingtheLowVacmode,itmaybenecessarytoadjustthepressureinthechamberforahighcontrastimage.Simplyadjustthepressureuntiltheappropriatecontrasthasbeenachieved.Inmostcasespressurebetween0.83and0.53Torrwillworkwell.
![Page 8: FEI Quanta 250 ESEM Basic instructions - OSCAR...FEI Quanta 250 ESEM Basic Operation Manual For Assistance with SEM Please Contact: Dr. Mukti Rana Associate Professor & Chair Department](https://reader035.fdocuments.net/reader035/viewer/2022081514/602247795d8d045b393964f9/html5/thumbnails/8.jpg)
8
Oncethedesiredimageisachieved,pressF2orgotoscan>Photoorscan>snapshotfortakingandsavingimage.OncethescanhasstoppedadefaultdirectorywillappearintheSharedDataFolder.Storeyour imagesintheSharedFolderonly.Datawillbestoreonthecomputerontheleftsideofthe operator’sconsol.
Turnoffthebeambyclickingthebeamonbutton(itgoesfromyellowbackgroundtopuregrayonceitisturnedoff)
NeverleavetheESEMintheLOWVACorESEMmode,alwaysreturntheSEMtotheHIGHVACMode.