Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography...

8
Abstract This Application Note demonstrates the use of Agilent 1260 Infinity Analytical Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study chiral and structurally related isomers with high analysis speed and excellent separation efficiency. The quantitative performance and analytical reproducibility were evaluated and proved to be highly reliable. This study also illustrates that modifier composition has a great effect on both chromatographic separation and MS signal response in ESI positive ion mode, thus optimization of modifier is crucial for successful SFC/MS analysis. Fast Analysis of Chiral and Structurally Related Isomers Using Supercritical Fluid Chromatography Mass Spectrometry Application Note Authors Yuqin Dai and Lisa Zang Agilent Technologies, Inc. Santa Clara, CA, USA

Transcript of Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography...

Page 1: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

AbstractThis Application Note demonstrates the use of Agilent 1260 Infi nity Analytical Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study chiral and structurally related isomers with high analysis speed and excellent separation effi ciency. The quantitative performance and analytical reproducibility were evaluated and proved to be highly reliable. This study also illustrates that modifi er composition has a great effect on both chromatographic separation and MS signal response in ESI positive ion mode, thus optimization of modifi er is crucial for successful SFC/MS analysis.

Fast Analysis of Chiral and Structurally Related Isomers Using Supercritical Fluid Chromatography Mass Spectrometry

Application Note

AuthorsYuqin Dai and Lisa ZangAgilent Technologies, Inc.Santa Clara, CA, USA

Page 2: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

2

IntroductionLC/MS is used extensively in drug discovery applications. However, some limitations including separation effi ciency and speed of analysis for chiral and structural related isomers have been noted. SFC/MS offers an excellent alternative or orthogonal separation to LC/MS. Advantages of SFC include very low system back pressure with liquid CO2 and organic modifi ers, which allow high throughput analysis with high fl ow rate and unique selectivity of SFC for some compound types. Additionally, SFC has proven to be cost effective and is considered a green technique by reducing organic solvent consumption.

This Application Note describes the utility of SFC/MS for fast analysis of a number of chiral and structurally related isomers. The performance of an Agilent 1260 Infi nity SFC coupled to an Agilent 6130B Single Quadrupole MS, in terms of the separation effi ciency, speed of analysis, reproducibility, and quantitative aspect is demonstrated.

ExperimentalChemicals and methodSeveral chiral and structurally related isomers were selected for the evaluation. The structures and molecular information of these compounds are shown in Figure 1. Samples of individual test compounds were prepared in methanol. Chromatographic separation was optimized using the chiral Lux Cellulose-1 column and an Agilent ZORBAX Rx-SIL column respectively. Research grade CO2 gas (from Airgas) was used as the supply for mobile phase A and methanol with 20 mM ammonium formate (mobile phase B) was used as the modifi er of supercritical liquid CO2.

Full MS or SIM

scans in positive ion mode were used for qualitative and quantitative analyses of the analytes of interest.

Figure 1. Structures of test compounds.

WarfarinC19H16O4 (MW = 308.3) C15H25NO3 (MW = 267.4)

C25H35N3O6 (MW = 473.5)

C21H28O5 (MW = 360.4)

C10H15NO (MW = 165.1)

Metoprolol

Atenolol impurity F

Ephedrine hydrochloride (1S, 2R)

Prednisolone and Cortisone

CH3

CH3

OH

•HClHN

(1R, 2S)

CH3

NHCH3

OH

•HCl

O

OOH

O

HOH

H3CO

CH3

CH3

O N

OHHO

O O

H2N NH2

CH3H3C

O ON

HO

H

H H

HO OH

O

OOH

OHH

H HO

OO

Page 3: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

3

InstrumentationAnalyses were performed using the Agilent 1260 Infi nity Analytical SFC system coupled to the Agilent 6130 Single Quadrupole mass spectrometer. The SFC/MS system consists of the following modules:

• Agilent 1260 Infi nity SFC control model (G4301A)

• Agilent 1260 Infi nity SFC binary pump (G3402A)

• Agilent 1260 Infi nity High Performance Degasser (G4225A)

• Agilent 1260 Infi nity SFC Autosampler (G4303A)

• Agilent 1290 Infi nity TCC (G1316C) with a 2-column selection valve

• Agilent 1260 Infi nity isocratic pump,

• Agilent 1260 Infi nity DAD (G1315C) with high pressure SFC fl ow cell

• Agilent 6130B Single Quadrupole Mass Spectrometer with ESI source.

The instrument confi guration is illustrated in Figure 2. Agilent OpenLab ChemStation Edition software version C.01.05 was used to control the SFC/MS instrument. In this study, the UV detector was bypassed. The pre-MS heating was achieved by fl ow through an additional TCC’s 6-µL preheating block with temperature set at 60 °C. As previously described1, the heating device prevents freezing of the lines caused by the expansion of CO2 upon exiting backpressure regulator (BPR). Additionally, a make-up fl ow was introduced prior to the BPR through an Agilent zero dead volume T-connector. This confi guration provides the best retention time and peak area reproducibility, thus it is used for qualitative and quantitative analysis.

The SFC/MS experimental conditions are summarized in Table 1.

Figure 2. Agilent 1260 Infi nity Analytical SFC/6130B Single Quadrupole mass spectrometer confi guration.

Agilent 1260 Isocratic Pump

Pre-MS heating

Agilent 1260 Infinity Analytical SFC System

Aurora A5

Agilent 6130B Single Quadrupole LC/MS

BPR

UV detector

SS capillary tubing 0.12 × 105 mm

60 °C

SS capillary tubing 0.12 × 400 mm

Table 1. SFC/MS conditions.

SFCColumns 1) Agilent ZORBAX Rx-SIL, 4.6 × 100 mm, 1.8 µm (p/n 828975-901)

2) Phenomenex Lux Cellulose-1, 4.6 × 150 mm, 3 µmColumn temperature 40 °CInjection volume (full loop) 5 µLBPR pressure 140 barBPR temperature 60 °CPre-MSD heating 60 °CMake-up fl ow 0.15 mL/min at 100 % BSupercritical fl uid (A) CO2

Modifi er (B) B1) Methanol with 20 mM ammonium formateB2) Methanol with 0.1 % formic acid

SFC fl ow and isocratic conditions

1) 4 mL/min at 20 % B for warfarin2) 2 mL/min at 35 % B for metoprolol3) 4.5 mL/min at 40 % B for impurity F (Cellulose-1 column)4) 4.5 mL/min at 40 % B for impurity F (Qual., Rx-SIL column)5) 4.5 mL/min at 30 % B for impurity F (Quant., Rx-SIL column)6) 3 mL/min at 20 % B for prednisolone and cortisone

SFC fl ow and gradient (for ephedrine mixture)

5 % B (initial), 5–10 % B (0-7 minutes), 10–40 % B (7–7.5 minutes), hold 40 % B for 0.5 minutes, 40–5 % B (8–8.5 minutes)

MSIon mode positiveCapillary voltage 3,000 VDrying gas 12 L/min at 350 °CNebulizer 50 psi

MS scan (200–450) with 20 % cycle time for SIM (warfarin, metoprolol, and the mixture of prednisolone and cortisone)MS scan (100–250) with 20 % cycle time for SIM (Ephedrine)SIM only (ion 474.3) (quantitative analysis for impurity F )

Peak width 0.03 minutesFragmental voltage 120

Page 4: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

4

Results and DiscussionSeparation effi ciencySFC/MS provides several advantages over LC/MS, including high separation effi ciency, shorter analysis time, and reduced organic solvent consumption. Figure 3 shows the isocratic chromatographic separation of three representative chiral and structurally related isomers using either Lux Cellulose-1 or ZORBAX Rx-SIL columns by SFC/MS. All three pairs of isomers were well separated in less than 3.5 minutes.

Warfarin

Warfarin

1.010

20

40

60

80

100

120

140

160×103

1.5 2.0

1.751

2.856

2.5 3.0 3.5 min

A

Atenolol impurity F

Atenolol impurity F

0.60

10

20

30

40

50

60

70

×103

0.8 1.0

1.196

1.211

1.2 1.6 1.8 min

B

1.4

Cortisone

Prednisolone

0.20

10

30405060

8090

×103

0.4 0.6

0.597

0.775

0.8 1.2 1.4 min

C

70

20

1.0

Figure 3. SIM chromatograms of (A) 500 ng/mL of warfarin using the Lux Cellulose-1 column at 20 % B1, (B) 200 ng/mL of atenolol impurity F using the Agilent ZORBAX Rx-SIL column at 40 % B1, and (C) 200 ng/mL of cortisone and prednisolone mixture using the Agilent ZORBAX Rx-SIL column at 20 % B1.

Page 5: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

5

modifi er B2 compared with modifi er B1 (Figure 4C). This modifi er effect was also observed for atenolol impurity F using ZORBAX Rx-SIL column (data not shown). It has been reported that the SFC elution strength of MeOH can be adjusted by acid-salt additives and their concentrations in separation of acidic enantiomer compounds2. Our fi nding showed a similar trend, in that the acidic additive (formic acid) did not help in chiral separation of the compounds studied. We found ammonium formate additive provided better separations.

containing 0.1 % formic acid). It is clear that a modifi er had a signifi cant effect on both chromatographic separation and MS signal response. Overall, modifi er B1 provided higher MS signals than modifi er B2 for all the isomers studied. However, the effect on the chromatographic separation was compound dependent. For example, modifi er B2 signifi cantly reduced separation effi ciency of metoprolol (Figure 4A), atenolol impurity F (Figure 4B), and a mixture of (1R, 2S) and (1S, 2R)-ephedrine isomers (data not shown), while similar chromatographic separation of warfarin was obtained using

Effect of Additive in MethanolChiral discrimination is a very complex phenomenon. It is almost impossible to predict which chiral stationary phase and modifi er combination will provide the best separation. Optimal conditions can vary greatly and are compound specifi c. In this study, two types of modifi ers (B1 and B2) were selected for the evaluation. Figure 4 shows separation profi les of three representative chiral molecules using Lux Cellulose-1 column and mobile phase CO2 in combination with either modifi er B1 (methanol containing 20 mM ammonium formate) or B2 (methanol

Figure 4. SIM chromatographic separation profi les of (A) metoprolol, (B) atenolol impurity F, and (C) warfarin using the Lux Cellulose-1, modifi er B1 (top panel) and B2 (bottom panel)

35 % B1

Metoprolol

00

50

150200250300

0.5 1.0

1.221

1.706

1.5 2.0 2.5 min

A

350400450

100

050

150200250300350400450

100

×103

35 % B2

0 0.5 1.0 1.5 2.0 2.5 min

×103

20 % B1

Warfarin

020

6080

100

1.25 1.75

1.752

2.860

1.711 2.896

2.0 2.5 3.0 min

C

120140160

40

050

150200250300350400450

100

1.5 2.25 2.75 3.25

1.0 1.25 1.75 2.0 2.5 3.0 min1.5 2.25 2.75 3.25

×103

20 % B2

×103

40 % B1

Atenolol impurity F

00

50

150200250300

1.0 1.4

1.406 1.602

1.6 2.0 2.2 min

B

350400450

100

050

150200250300350400450

100

1.2 1.8 2.4

0 1.0 1.4 1.6 2.0 2.2 min1.2 1.8 2.4

×103

40 % B2

×103

Page 6: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

6

ReproducibilityReproducibility is critical for quantitation. Figure 5 shows the overlaid SIM chromatograms of six consecutive injections of warfarin (5A), metoprolol (5B), atenolol impurity F (5C), and a mixture of (1R, 2S) and (1S, 2R) ephedrine isomers (5D). Excellent reproducibility was obtained with RSDs of less than 0.1 % for retention time and less than 5 % for peak area.

Figure 5. Overlay of the SIM chromatograms from six replicate injections of (A) 500 ng/mL of warfarin, (B) 100 ng/mL of metoprolol, (C) 200 ng/mL of atenolol impurity F, and (D) 200 ng/mL of (1R, 2S) and (1S, 2R) ephedrine mixture using the SFC/MS. The separation was performed using the Lux-Cellulose-1 column and the mobile phase CO2 in combination with the modifi er B1.

Peak 1

Peak 2

N = 6RSD (RT)RSD (peak area)

Peak 10.051.2

Peak 20.030.8

020

6080

100

1.0 2.0 2.5 3.0 min

A

120140

40

1.5 3.5

×103 Peak 1

Peak 2

N = 6RSD (RT)RSD (peak area)

Peak 10.041.1

Peak 20.031.8

010

304050

1.0 1.0 1.5 2.0 min

B

6070

20

0.5 2.5

×103

Peak 1

Peak 2

N = 6RSD (RT)RSD (peak area)

Peak 10.081.0

Peak 20.000.7

05

152025

0 1.0 1.5 2.0 min

C

3035

10

0.5 2.5

×103

(1R, 2S)

(1S, 2R)

N = 6RSD (RT)RSD (peak area)

(1R, 2S)0.084.9

(1S, 2R)0.073.7

40

80

100

120

4.0 5.0 5.5 6.0 min

D140

60

4.5 6.5

×103

Page 7: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

7

Quantitative performanceThe ZORBAX Rx-SIL column provided similar separation effi ciency for atenolol impurity F as Lux Cellulose-1 column. Since the silica column offers a cost benefi t compared to the chiral column, quantitative performance was further evaluated for atenolol impurity F using the ZORBAX Rx-SIL column and the SFC/MS in SIM mode only. As shown in Figure 6, the lower limit of quantitation (LLOQ) was 1.0 ng/mL with a linear dynamic range of 1.0–250 ng/mL.

ConclusionsThe Agilent 1260 Infi nity Analytical SFC System coupled to the Agilent 6130B Single Quadrupole System provided excellent analytical performance for analyzing chiral and structurally related isomers. The resulting data showed good sensitivity, high separation effi ciency, excellent retention time and peak area reproducibility, and reliable quantitation. Optimization of modifi er composition was important and methanol containing 20 mM ammonium formate proved to enable the successful SFC/MS analyses of all isomers in this study. Furthermore, the Agilent SFC/MS system uses standard grade CO2 gas instead of liquid SFC grade C CO2 gas, which results in 10–15x lower operating costs.

Figure 6. Quantitative analysis of atenolol impurity F using the Agilent ZORBAX Rx-SIL column and 30 % B1 at 4.5 mL/min. (A) calibration curve of Peak 1, (B) calibration curve of Peak 2, (C) SIM chromatogram of impurity F at 1.0 ng/mL (LLOQ).

A

160

×103 ×103

140120100806040200

0

1234

5

6

R2 = 0.99912y = 714.12x – 1,635.2

7

100Amount (ng/mL)

200Ar

ea

C

900

950

850

800

750

700

650

600

550

5001.0

Peak 2Peak 1

1.51.25Time (min)

2.01.75 2.25 2.5 2.75 3.0 3.25

Area

B2001751501251007550250

0

1234

5

6

R2 = 0.99909y = 815.95x – 1,941.6

7

100Amount (ng/mL)

200

Page 8: Fast Analysis of Chiral and Structurally Related Isomers ... · Supercritical Fluid Chromatography coupled to an Agilent 6130B Single Quadrupole mass spectrometer (SFC/MS) to study

www.agilent.com/chem

This information is subject to change without notice.

© Agilent Technologies, Inc., 2014Published in the USA, February 3, 20145991-3859EN

References1. M. Dunkel, G. Vanhoenacker, F. David,

P. Sandra, M. Vollmer “The Agilent 1260 Infi nity SFC/MS Solution” Agilent Technologies Publication Number 5990-7972EN, 2011.

2. T. Zhang, D. Nguyen, P. Franco (Chiral Technologies, Europe IIIkirch, France), M. Vollmer (Agilent Technologies, Waldbronn, Germany) “Enantiomer Separation of Acidic Compounds” Agilent Technology Publication Number 5990-9594EN, 2012.