Expanding Your HPLC and UHPLC Capabilities with Universal ...
Transcript of Expanding Your HPLC and UHPLC Capabilities with Universal ...
1 The world leader in serving science
Bruce Bailey, Ph.D. Thermo Fisher Scientific, Chelmsford, MA Pittcon™ Conference & Expo 2014 March 2-6, 2014
Expanding Your HPLC and UHPLC Capabilities with Universal Detection: Shedding Light on Compounds That Lack a Chromophore
2
Outline
• Introduction to Charged Aerosol Detection • How Charged Aerosol Technology Works • Comparison with Evaporative Light Scattering Detectors
(ELSD) • Examples of Applications • Inverse Gradient Solution for Uniform Response
3
Introduction to Charged Aerosol Detection
Comparison of Charged Aerosol Detection to UV and MS
• Used to quantitate any non-volatile and many semi-volatile analytes with LC
• Provides consistent analyte response independent of chemical structure and molecule size
• Neither a chromophore, nor the ability to ionize, is required for detection
• Dynamic range of over four orders of magnitude from a single injection (sub-ng to µg quantities on column)
• Mass sensitive detection – provides relative quantification without the need for reference standards
• Compatible with gradient conditions for HPLC, UHPLC, and micro LC
4
The liquid eluent from the LC column enters the detector (1) where it undergoes nebulization by combining with a concentric stream of nitrogen gas or air (2).
The fine droplets are carried by bulk gas flow to the heated evaporation sector (3) where desolvation occurs to form particles, while any larger droplets are drained to waste (4).
The dry particles exit from evaporation (5) and are combined with another gas stream that first passes over a high voltage Corona charger (6). The charged gas then mixes with the dry particles, where excess charge transfers to the particle’s surface (7).
Charged Aerosol Detection – How It Works
Any high mobility species are removed by an ion trap (8) while the remaining charged particles pass to a collector where the passing particles charges are measured with a very sensitive electrometer (9). The resulting signal is then conveyed to a chromatographic data software for quantitation.
Signal is directly proportional to the analyte quantity
1 2
3
4
5 6
7
8
9
6
Particle Charging for Charged Aerosol Detection
Mixing Chamber
• Particle size proportional to mass of analyte + background residue
• Charge per particle proportional to particle size
• Charged particles are measured, not gas phase ions as in MS
Charged particle
Dried particle
Charged gas ion
7
Corona ultra RS vs. Corona Veo RS Detectors
Coaxial N2 flow
Capillary Inlet
Aerosol
FocusJet™ Concentric Nebulizer Tip
Thermo Scientific™ Dionex™ Corona™ Veo™ RS Charged Aerosol Detector
Thermo Scientific™ Dionex™ Corona™ ultra™ RS Charged Aerosol Detector
Cross-flow Nebulizer
Impactor
8
Corona Veo Detector – What's New?
• Radically new concentric nebulization system improves sensitivity and precision
• All new evaporation scheme widens the scope of applications to include low flow capabilities for micro LC, as well as UHPLC
• Usability and serviceability are enhanced by countless improvements, many of which came from our customers
This entirely new detector incorporates many design and performance improvements:
9
Comparison Between Corona Charged Aerosol Detection
vs. ELSD
10
Comparisons Charged Aerosol vs. ELS Detectors
ELSD measures light scattered by the aerosol
ELSD Corona Veo Detection
Charged Aerosol Detection measures the aggregate charge of the aerosol
Evaporating chamber
Siphon
Heated Nebulizer
Light source
Detection chamber
ELSD
11
Detector Response Characteristics R
espo
nse
Mass on Column
Maj
or
resp
onse
err
or
0 1000 2000 3000 4000 5000 6000
Mass on Column
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
Res
pons
e ng
pA*min
Typical ELSD sigmoidal response curve. Typical Charged Aerosol Parabolic Response Curve
12
Comparisons Charged Aerosol vs. ELS Detectors
• A major consequence of ELSD sigmoidal response is that the dynamic range is relatively small and analyte signal rapidly decreases and completely disappears as the amount of analyte decreases.
• Unlike ELSD, Charged Aerosol Detector response does not
simply disappear for the same lower levels of analytes. Subsequently charged aerosol detection performs better for measurement of lower analyte levels and is generally more sensitive and provides a wider dynamic range than ELSD.
13
Calibration of the Charged Aerosol Detector
• Over short ranges, the Charged Aerosol Detector is linear.
• Over wider ranges it is parabolic in behavior. To deal with this, several approaches are available. Which is the most appropriate will depend upon the data. Selection includes: • Log-Log
• Quadratic
• Power function
14
Working with Non-Linear Data
• Limits of Detection (LoD) data by extrapolation from Signal / Noise data is only practical when working with a linear response.
• Both charged aerosol and ELS detector are non-linear. LoDs cannot be extrapolated from the response of high levels of analyte and can only be determined through the generation of calibration curves.
• Extrapolation of non-linear data produces major errors and should be avoided.
15
Comparisons Charged Aerosol vs. ELS Detectors
Corona Veo
Sedex ELSD LT90
0.00 1.00 2.00 3.00 Time [min]
-2.00
-1.00
0.00
1.00
2.00
C u r
r e n t
[ p A ]
Theophylline
Caffeine
min
pA mV
-1.00
0.00
1.00
2.00
3.00
4.00
5.00
6.00
7.00
8.00
R e s p o n s e [ m
V ]
Theophylline and Caffeine, 2 -31 ng on column
Charged Aerosol Detector
ELSD
16
Comparisons Charged Aerosol vs. ELS Detectors
Theophylline and Caffeine, 8 ng on column
8 ng injected
0.00 1.00 2.00 3.00 4.00 Time [min]
-1.00
0.00
1.00
C u r
r e n t
[ p A
]
theophylline S/N = 238
caffeine S/N = 23
theophylline S/N = 2
-10.0
-5.0
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
40.0
R e s p o n s e [ m
V ]
Charged Aerosol Detector
ELSD
17
Avoid Extrapolation of Non-Linear Data
Medium Level Standard
Avg. SNR for analytes
• Evaporative Light Scattering Detector - 1283
• Charged Aerosol Detector - 230
10-fold Dilution of Medium Level Standard
Avg. SNR for analytes
• Evaporative Light Scattering Detector - 8.5
• Charged Aerosol Detector - 30
0,21 0,50 1,00 1,50 2,00 2,50 3,00 3,50 4,00 4,50 5,00 5,50 6,00 6,50 7,00 7,50 -0,6
5,0
10,0
15,0
20,0
25,0
29,4
min
pA
1 - CAD PF1,5 #5 [manipulated] RPmix 1/10 CAD_1 -768,00
-767,00
-766,00
-765,00
-764,00
-763,00
-762,00
-761,00
min
mV
2 - ELSD #3 [manipulated] RPmix 1/10 ELSD -10,0
-766,75
-766,50
-766,25
-766,00
-765,80 mV
ELSD
Charged Aerosol Detector
0,50 1,00 1,50 2,00 2,50 3,00 3,50 4,00 4,50 5,00 5,50 6,00 6,50 7,00 7,50 7,69 -0,50
0,00
1,00
2,00
3,00
4,00
4,50
min
pA
1 - CAD PF1,5 #6 [manipulated] RPmix 1/100 CAD_1 -767,60 -767,50
-767,25
-767,00
min
2 - ELSD #4 [manipulated] RPmix 1/100 ELSD
ELSD
Charged Aerosol Detector
18
Working with Non-Linear Data
• Charged aerosol detectors performs better for the measurement of low levels of analytes, and have a wide dynamic range of four orders of magnitude. The analyte’s physicochemical properties affect the detector much less than ELSD.
• Charged aerosol detectors uses a single nebulizer to address a wide flow rate range. ELSD requires multiple nebulizers adding to expense and downtime.
19
Working with Non-Linear Data
• The only way to estimate the LoD when response is non-linear is to construct a calibration curve.
• Comparisons are completely meaningless when the response of a non-linear detector to a high concentration of standard is used to imply that the performance of one detector is superior to the other.
20
Comparisons Charged Aerosol vs. ELS Detectors
Charged Aerosol Detector ELSD
Response Curvilinear Sigmoidal Dynamic Range >4 orders 2–3 orders
LoQ and LoD LoQ and LoD often lower (better) than that estimated by SNR
LoQ and LoD often higher (worse) than that estimated by SNR
Sensitivity (LoD) <1 ng >10 ng Semivolatility Range Similar Similar Analyte Response Independent of structure Variable - dependent on compound Ease of Operation Simple Can be complex
21
Charged Aerosol Applications: Shedding Light on Compounds
That Lack a Chromophore
22
Determination of Adjuvants
Column: Thermo Scientific™ Hypersil GOLD™ PFP 1.9 um, 2.1 × 100 mm Mobile Phase A: 0.1% Formic acid in water Mobile Phase B: 0.1% Formic acid in 10:90 acetonitrile:reagent alcohol Gradient: 35% B to 83% B in 6 min to 90% B in 10 min Flow Rate: 0.5 mL/min Inj. Volume: 2 μL Col. Temp: 45 ºC Evap. Temp: 50 ºC
Analysis of Plant Saponins
UV @ 210 nm
23
Glycan Analysis for Bovine Fetuin
Column: Thermo Scientific™ GlycanPac AXH-1™, 1.9 μm, 2.1 × 150 mm Mobile Phase A: 80% Acetonitrile Mobile Phase B: 80 mM Ammonium formate, pH 4.4 Gradient: 2.5% B to 25% B from 1 to 40 min Flow Rate: 0.4 mL/min Inj. Volume: 5 μL Col.Temp: 30 ºC Evap. Temp: 50 ºC
Separation of Oligosaccharide Alditols
Native Glycans
24
Determination of Carbohydrates in Juice
Column: Amino, 3 μm, 3 × 250 mm Mobile Phase: Acetonitrile:water (92:8) Flow Rate: 0.8 mL/min Inj. Volume: 2 μL Col. Temp: 60 ºC Post-column Temp: 25 ºC Evap. Temp: 75 ºC Sample Preparation: Add 20 mL of 85% acetonitrile to 1 gram juice
Analysis of Simple Sugars
Simplified sample preparation “Dilute-and-shoot” method
25
0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 Time [min]
-1.0
0.0
1.0
2.0
3.0
4.0
5.0
6.0
7.0
8.0
9.0
10.0
11.0
12.0
C u
r r e
n t [
p A
]
Isosteviol
Steviol
Rubusoside
Dulcoside A
Stevioside Steviolbioside
Rabaudioside C
Rabaudioside F
Rebaudioside B
Rebaudioside A
Sodium
Rebaudioside D
Mixture Containing 11 Stevia Glycoside Standards
(n=3)
Column: Thermo Scientific™ Acclaim™ Trinity™ P1, 3 µm, 2.1 × 150 mm Mobile Phase: 88:12 (v/v) Acetonitrile:10 mM ammonium formate, pH 3.1 Flow Rate: 0.8 mL/min Inj. Volume: 2 µL Col. Temp: 30 ⁰C Detection: Corona Veo RS Veo Settings: 2 Hz, 5 second filter, PF 1.0, Evap. Temp 35 ⁰C
26
Characterization of Algae-based Biofuels Column: Thermo Scientific™ Accucore™ C18, 2.6 μm, 3.0 ×150 mm Mobile Phase A: Methanol:water:acetic acid (600:400:4) Mobile Phase B: Tetrahydrofuran:acetonitrile (50:950) Mobile Phase C: Acetone:acetonitrile (900:100) Gradient: Time FlowRate %A %B %C (min) (mL/min) -10.0 1. 00 90 10 0 -0.1 1. 00 90 10 0 0. 0 0. 25 90 10 0 20.0 0. 50 15 85 0 35.0 0. 50 2 78 20 60.0 0. 50 2 3 95 65.0 0. 50 90 10 0 Flow Rate: 1.0 mL/min Inj. Volume: 2 μL Col. Temp: 40 ⁰C Evap. Temp: 40 ⁰C
Analysis of Algal Oils
27
Active Ingredient Composition
Analysis of Gentamicin Standard (200 μg/mL)
Column: Acclaim RSLC PolarAdvantage II, 2.2 μm, 2.1 × 100 mm
Mobile Phase A: 0.025:95:5 HFBA:water:acetonitrile Mobile Phase B: 0.3:95:5 TFA:water:acetonitrile Gradient: 0 to 1.5min,1 to 10%B
1.5 to 7min,10 to 100% B 7 to 10min,100% B 4 min. pre-injection equilibration
Flow Rate: 0.45 mL/min Inj. Volume: 1 μL Col. Temp: 15 ⁰C Evap. Temp: 80 ⁰C
28
Formulation Testing
Column: Acclaim Trinity P1, 3 μm, 3.0 × 50 mm Mobile Phase A: 75% Acetonitrile Mobile Phase B: 25% 200 mM Ammonium acetate pH 4 Flow Rate: 0.8 mL/min Inj. Volume: 5 μL Col. Temp: 30 ⁰C Evap. Temp: 60 ⁰C
Measurement of Chloride Impurity
Analysis of Diclofenac-Sodium Salt (1 mg/mL)
29
Conventional Gradient Elution
Inverse Gradient Compensation
Inverse Gradient Solution for Uniform Response
30
Solution for Uniform Response with Gradients
• Dual-gradient pump is the heart of this exclusive solution
• Inverse gradient fingertight fitting kits are supplied for LC systems
• Furnished with unique eWorkflows Dual Gradient Pump
Inverse Gradient Setup
Thermo Scientific™ Dionex™
Viper™ Fingertight Fitting
31
Effects of Gradient and Mass on Calibration
R² = 0.9997R² = 0.9999R² = 1
0
1
2
3
4
5
6
7
0 500 1000 1500 2000 2500
Mass on Column (ng)
Sulfanilamide
Famotidine
Perphanzine
Inverse gradient extends the consistency of response
R² = 0.9999
R² = 0.9995R² = 0.9998
0
1
2
3
4
5
6
7
0 500 1000 1500 2000 2500
Mass on Column (ng)
Sulfanilamide
Famotidine
Perphanzine
Standard Gradient (Single Pump)
Peak
Are
a (C
harg
ed A
eros
ol D
etec
tor)
Inverse Gradient (Dual Pump)
32
Determination of Drug Discovery Mass Balance
Charged Aerosol
UV
Column: Acclaim 300 C18, 3 μm, 4.6 ×150 mm Mobile Phase A: 20 mM Ammonium acetate, pH 4.5 Mobile Phase B: Acetonitrile Gradient: 2% B to 98% B in 30 min, Inverse Gradient Flow Rate: 0.8 mL/min Inj. Volume: 2 μL Col. Temp: 30 ⁰C Evap. Temp: 35 ⁰C
Corona offers a more uniform response than UV
33